Inductive Conductivity Meter ILM

Similar documents
Temperature Sensor with G1/2" hygienic

Capacitive Limit Switch Pharma NCS

Temperature Sensor with M12 hygienic

Capacitive Limit Switch Pharma NCS

Pharma-Temperature Sensor M12 hygienic

Continuous Level Sensor NSL-M

Temperature Sensor with M12 hygienic

Capacitive Limit Switch Pharma NCS

Temperature Sensor for Build-In System G3/8" FLEXadapt

Temperature Sensor Tri-Clamp

Temperature Sensor for Build-In System G3/8" FLEXadapt

Pharma Temperature Sensor G3/8"

Pharma Temperature Sensor G3/8"

Temperature Sensor Tri-Clamp

Temperature Sensor G1/2" Standard

Temperature Sensor without thread

Continuous Level Sensor NSL-M-00

Temperature Sensor with Thread G1/4" FLEXadapt

Temperature Sensor G1/2" Standard

Pharma Temperature Sensor TFP-EX

Magnetic-Inductive flow Meter fmi

Temperature Sensor Tri-Clamp

Technical Information Smartec CLD18

Technical Information Smartec CLD18

Temperature Sensor. Temperature Transmitter MPU-LCD with Display Product Information TSM and TSB FOOD SENSORS FOR FOOD AND LIFE SCIENCES.

Digital inductive conductivity transmitter

RTD Temperature Sensor omnigrad T -TSM 480, TR 480. Hygienic RTD sensor with Pt 100, class A TSM 480 with electronics programmable via PC

Technical Information Smartec CLD18

The transducer component converts the measured signal and displays the actual value.

ED701 General Industry Pressure Transmitter

omnigrad T - TSM 480, TR 480. Technical Information. RTD Temperature Sensor Hygienic design TSM 480 with electronics programmable via PC

Type PLC Fitting. Valve for Continuous control. Diaphragm valve. Output

P R o D u c T I n F o R m at I o n. TDR Level Sensor LFP. The clean solution

Endress Hauser. SmarTec M CLD 133 Conductivity Measurement

Inductive conductivity transmitter JUMO CTI-Junior

Pressure transmitter For sanitary applications Model SA-11

Level sensor Magnetostrictive, high-resolution measuring principle For sanitary applications, model FLM-H

Operating Instructions for Microwave Level Switch. Model: LNM

ph or ORP Transmitter

Conductivity sensor for hygienic applications

Indumax CLS54D. Technical Information

Ceramax CPS341D. Technical Information

Electromagnetic Flow Transmitter

Conductivity Transmitter

Alfa Laval Conductivity Sensor

Digital conductivity transmitter

Technical Information Memosens CLS82D

Digital flow ELEMENT transmitter for continuous flow measurement

XXXX e. X d.. X X X a

Technical Information Condumax CLS21D/CLS21

Conductivity measurement SmarTec S CLD 132

OPTITEMP TT 20 Technical Datasheet

XXXX e. X d.. X X X a

Magnetic Inductive Flow Sensor induq

Inductive Conductivity/Concentration and Temperature Transmitter with switch contacts. Temperature sensor. Graphics LC display.

Rotary Measurement Technology Incremental Encoders

BATCHFLUX 5500 C Technical Datasheet

Digital flow ELEMENT transmitter for continuous flow measurement

Technical Information Indumax CLS54

ENCODER. Analog Multiturn Encoder. Serie 8.M3661/M3681. Key-Features:

LCI. Inductive Conductivity/ Concentration and Temperature Transmitter. Compact or separate version

Type F-20-3A Sanitary Pressure Transmitter NEMA 4X with Integral Junction Box

Operating manual GTL 241 GTL 251. Temperature probe. Please keep the manual for future use.

OPTIFLUX 5000 Technical Datasheet

Precipition Sensor with analogue Intensity Output ,

JUMO Dtrans T100 Screw-in RTD Temperature Probe with/without Transmitter

Condumax W CLS15 and CLS15D

SMARTPAT COND 1200 Technical Datasheet

Flush pressure transmitter For viscous and solids-containing media Model S-11

INLINE flowmeter for continuous flow measurement

Technical Information Nivector FTC968, FTC968Z

Condumax W CLS19. Technical Information. Conductivity sensors Two-electrode sensors with cell constants k = 0.01 cm 1 or k = 0.

OPTIFLUX 5000 Technical Datasheet

Condumax CLS21 and CLS21D

Cable Glands, Metal, for shielded EMC Cables CG.EM.*.*.*

FLOWave SAW-Flowmeter

JUMO CTI-500. Inductive Conductivity/Concentration and Temperature Transmitter with switch contacts. Type Brief description.

Absolute Encoders Multiturn

Digital flow transmitter for continuous flow measurement

Digital flowmeter for continuous flow measurement

RoHS. High shaft load capacity. Shock / vibration resistant

INLINE flowmeter for continuous flow measurement

Technical Information Condumax CLS15D/CLS15

LF Conductivity Measuring Probe

SensyTemp TSBA (BA R) Resistance thermometer for building automation, machine construction and environmental engineering

INLINE flowmeter for continuous flow measurement

WATERFLUX 3000 Quick Start

Conductivity Measuring Cells for High-Temperature Applications ConduMax H CLS 16

FMG90 Series Electromagnetic Flow Meter

WATERFLUX 3000 Quick Start

Digital flowmeter for continuous flow measurement

KPS Turbidity Meter User s Manual

Precipitation Monitor ,

Absolute Encoders - Singleturn

Pressure transmitter for general industrial applications Model A-10

OEM pressure transmitter for general industrial applications Models O-10 (T), O-10 (5)

Condumax CLS15 and CLS15D

OPTIFLUX 1000 Technical Datasheet

Pressure transmitter For general industrial applications Model A-10

Valve Island with Electronic I/O

Transcription:

1 Product Information ILM-2 ILM-3 Inductive Conductivity Meter ILM f o o d Application / Specified Usage Authorizations Inductive measurement of the specific conductivity of liquid media in the range of 0...999 ms/cm. Designed for hygienic applications in food-, beverage- and pharmaceutical industries. Application Examples Controlling of CIP processes (e. g. phase separation detergents/water) Concentration measurement (e.g. Alkali and acid concentration in remaking) Monitoring of product quality, quality control 74-06 ILM-2 / L20 with EMZ-352 Hygienic design / Process Connection Flow optimized, hygienic and easy sterilizable installation by sleeve EMZ-352 or the build-in system EHG-.../1". CIP/SIP cleaning up to 140 C / 30 minutes maximum Product contacting materials compliant to FDA Sensor made of stainless steel, bobbin case made of PEEK Conforming to 3-A Sanitary Standard 74-06 Additional process connections: Tri-Clamp, Milchrohr (DIN 11851), Varivent, APV, DRD et alli features / Advantages Wearless inductive measurement principle. Contrary to conducitve measurement principle there are no problems caused by corrosion of the electrodes or polarization. Up to 14 measurement ranges selectable, max. four external switchable (ILM-3). Precise measurment by compensation of temperature influence. Each measurement range can be assigned a seperate temperature coefficient (ILM-3). High reproducibility of 1 % of measurement value. Analog output for conductivity and temperature as standard. Installation in pipes from DN 40 possible. ILM-2 / L50 with Tri-Clamp options / Accessories Electrical connection via M12 plug-in connector Version with longer bobbin case for pipes DN 65 or for installation into T-fitting Preassembled cable for M12 plug-in connector Measuring Principle of the Inductive Conductivity Meter Inductive Conductivity An alternating current generates a magnetic field in the primary coil (sender) which induces a current in the circumfluent medium. The current flow in the medium generates another magnetic field in the secondary coil (receiver). The strength of the induced current in the secondary coil depends on the conductivity of the medium. The conductivity of the liquid medium is temperature dependent. To compensate the temperature error, an additional sensor (NTC) in the sensor tip is used for monitoring the temperature of the medium. The temperature coefficient (TC-value) of the liquid can be set up in the electronics of the ILM which is used for automatic compensation of the temperature error. Sender Induced current flow I~ Receiver NTC Medium NEGELE MESSTECHNIK GMBH Raiffeisenweg 7 87743 Egg an der Guenz Phone +49 (0) 83 33. 92 04-0 Fax +49 (0) 83 33. 92 04-49 sales@anderson-negele.com Tech. Support: support@anderson-negele.com Phone +49 (0) 83 33. 92 04-720

f o o d Specification 2 Specification Process connections thread G1" torque sensor, combined with Negele-weld-in sleeves max. 20 Nm Materials Ranges Pressure Protection Class connector head thread connection bobbin case window in lid ambient process CIP/SIP cleaning stainless steel 1.4305 (303), Ø 89 mm stainless steel 1.4305 (303), SW 36 mm PEEK, FDA-number (21CFR177.2415) PMMA -10...+60 C 0...100 C up to 140 C/30 minutes max. 10 bar max. IP 69 K (with PG cable gland and suitable cable) Reproducibility of conductivity 1 % of measurement value Resolution Accuracy Long Term Stability Accuracy of the Output Electrical Connection measurement range < 10 ms/cm 10...50 ms/cm 100...999 ms/cm span offset span offset 100 C 100...150 C cable gland cable connection supply 1 µs/cm 10 µs/cm 100 µs/cm ±2 % of upper range value ±20 µs/cm ±0.5 % of upper range value ±20 µs/cm 0.5 C max. 1.0 C max. 2 x M16 x 1.5 2 x M12 plug 1.4305 18...36 V DC max. 190 ma Inputs range switching E1 and E2 (24 V DC) galvanically isolated Outputs conductivity temperature analog 4...20 ma short-circuit-proof analog 4...20 ma short-circuit-proof LC-Display with backlight 2 x 8-digits Principle wearless inductive Comparison ILM-2 / ILM-3 ILM-2 ILM-3 Ranges Conductivity Ranges Coefficient (TC) 0...2 ms/cm up to 0...999 ms/cm 12 measurement ranges selectable 3 ranges extern switchable 0...+150 C 1 measurement range fix presetted 0...5 %/K, free adjustable 1 TC for all measurement ranges 0...0.5 ms/cm up to 0...999 ms/cm 14 measurement ranges selectable 4 ranges extern switchable -20...+150 C 7 measurment ranges selectable 0...5 %/K, free adjustable 1 TC for each measurement range

3 Mechanical Connection Installation f o o d Dimensioned Drawing Mechanical Connection / Installation The sensor has to be installed in that way that the bobbin case is entirely washed around by media and no bubbles can occure. Installation in a rising pipe is recommended. The inscription FLOW on the bottom side of the sensor has to show in flow direction of the medium. Very heavy vibrations can cause measurement errors (e. g. installation very near a pump). Use Negele CLEANadapt system for safe operation of measuring point! Attention: The maximum tightening torque for mounting is 20 Nm! 60 / 90 Use a welding mandril for correct installation of CLEANadapt weld-in fittings. Please pay attention to the weld-in and installation details in the CLEANadapt product information. 20 / 50 Conditions for a measuring point according to 3-A Sanitary Standard 74-06 The sensors ILM-2 and ILM-3 conforming to the 3-A Sanitary Standard. Sensors are designed for CIP/SIP cleaning. Maximum 140 C/30 minutes. Only with the build-in system CLEANadapt (EMZ, EMK, EHG with pipe diameter > DN25, ISO 20 and 1", Adapter AMC and AMV) allowed. Using the weld in sleeve EMZ, EMK the weld must comply to the requirements of the current 3-A Sanitary Standard. Mounting position, self draining and the position of the leackage hole must be in accordance to current 3-A Sanitary Standard.

f o o d Installation 4 Electrical Connection Handling / Operation Adjustment of Measuring Range Delivery status: measurement range 1: 0...20 ms/cm = 4...20 ma TC-value: 2 %/K Via the external control voltage +24 V DC (18...36 V) range 2 (E1=24 V), range 3 (E2=24 V) or range 4 (E1=E2=24 V) can be selected (see Electrical Connection ). At ILM-3 each measurement range can be assigned a separate temperature coefficient (TC). At ILM-2 one and the same TC is effective for all measurement ranges. At ILM-2 the temperature output is fix presetted to 0...150 C. At ILM-3 the measurement range of the temperature output can be selected from 7 presetted ranges between -20...150 C. Switching the Measuring Range The digital control inputs E1 and E2 are galvanically isolated from supply voltage. E1 E2 Meas. Range 0 0 1 1 0 2 0 1 3 1 1 4* With M12 plug M12 plug left (4 pin) outputs 4...20 ma 1: output conductivity + 2: output temperature + 3: output temperature - 4: output conductivity - M12 plug right (5 pin) supply/control voltages 1: supply +24 V DC 2: digital input E2 3: 0 V (measurement range switching) 4: supply 5: digital input E1 0 0 V DC; 1 24 V DC; Ground: clamp 9 * only ILM-3 Advice Occuring several media with very different conductivity in the application (e. g. CIP cleaning) switching to an adequate measuring range is neccessary for a precise measurement! Detecting the Coefficient of the Medium Delivery status: see Handling/Operation 1. Adjust TC to 0 %/K (see Adjustment). 2. Dip sensor into medium with 25 C. 3. Wait until the measurment value is stable. 4. Metering and note the conductivity value from the display. 5. Warm up the medium to 60 C minimum. Thereby the conductivity value in the display is changing. 6. Wait until the measurment value is stable. 7. Select TC in the operation menue and adjust the temperature coefficient until the measurement value is equal to the value noted at step 4.

5 Operation f o o d Operation Menue ILM-2 Operation Menue ILM-3 Coefficient Level TC Range 3 (E2=24V) Range 2 (E1= 24V) Range 1 (E1, E2 0V) Display Conductivity and Start Status Messages LM-2 / ILM-3 Display Mode After 1 min. the device switches automatically back to the display mode. Output Coefficient 4 Range 4 Coefficient 3 Range 3 Coefficient 2 ^-Symbol Current output conductivity overload, will be displayed if the measured value is higher than the selected measurment range. I out : ca. 22 ma 4 (upper line) currently editable range 1 (lower line) currently activated measurement range ^.^^^ -Symbol the currently measured value is higher than the maximum measurement value (999 ms/cm) I out : ca. 22 ma Range 2 Coefficient 1 Range 1 -Symbol the adjoining value is now editable via arrow buttons vvv -Symbol inductor error/sensor break I out : 2.4 ma Display Conductivity and Display Mode After 1 min. the device switches automatically back to the display mode. Start

Warnings Application Examples food Cleaning / Maintenance Transport / Storage In case of using pressure washers, dont t point nozzle directly to electrical connections! Do not store outside Store in an area that is dry and dust-free Do not expose to corrosive media Protect against solar radiation Avoid mechanical shock and vibration Storage temperature 0...40 C Relative humidity max. 80% Notice on conformity Applicable directives: Electromagnetic Compatibility Equipment Directive 2004/108/EC The CE label confirms compliance of this product with the applicable EC directives. You have to guarantee the compliance of all guidelines applicable for the entire equipement. Disposal Reshipment Sensors and process connection must be clean and must not be contaminated with hazardous media and/or heatconductive paste. Note the cleaning information! Use suitable transport packaging only to avoid damage of the equipment! This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. Phase separation in CIP-equipment with ILM-2 Phase separation in CIP-equipment with ILM-2 6

7 Process Connections f o o d overview of further possible process connections (adapter must be ordered separately!) The complete overview of all available adapters you will find at product information CLEANadapt. ILM-2 ILM-3 Process connection Build-in system EHG (DIN 11850 series 2) Negele weld-in sleeve Negele weld-in sleeve Negele weld-in sleeve Tri-Clamp dn40 EHG-dIN2-40/1" AMC-352/1"-1.5" dn50 EHG-dIN2-50/1" EMZ-352 EMZ-351 EMS-352 AMC-352/2" dn65 EHG-dIN2-65/1" suitable for suitable for pipes suitable for AMC-352/3" dn80 EHG-dIN2-80/1" installation in and vessels with installation in vessels leackage hole pulled-out pipes AMC-352/80 dn100 EHG-dIN2-100/1" AMC-352/100 overview of further possible process connections (adapter must be ordered separately!) ILM-2 ILM-3 Process connection dairy flange (DIN 11851) Varivent APV-Inline Adapter G1½" to G1" dummy flange dn40 AMK-352/40 AMV-352 AMA-352 dn50 AMK-352/50 AMV-352 AMA-352 dn65 AMK-352/65 AMV-352 AMA-352 dn80 AMK-352/80 AMV-352 AMA-352 dn100 AMK-352/100 - AMA-352 AMG-352 suitable for existing G1½" connection BST-350 to close existing measurement points

f o o d Product Information ILM-2 ILM-3 8 order Code ILM-2 ILM-3 (12 measurement ranges, 1 temperature coefficient, 3 measurement ranges external switchable) (14 measurement ranges, 4 temperature coefficients, 4 measurement ranges external switchable) Insertion-Length L20 L50 (20 mm) (50 mm) Electrical Connection PG (cable gland M16x1.5) M12 (M12 plug-in 1.4305 (303)) ILM-2 / L20 / M12 Accessories PVC-cable with M12-connection PVC-cable with M12-connection, 1.4305 (303), IP 69 K, unshielded M12-PVC / 4-5 m PVC-cable 4-pin, length 5 m M12-PVC / 4-10 m PVC-cable 4-pin, length 10 m M12-PVC / 4-25 m PVC-cable 4-pin, length 25 m M12-PVC / 5-5 m M12-PVC / 5-10 m M12-PVC / 5-25 m PVC-cable 5-pin, length 5 m PVC-cable 5-pin, length 10 m PVC-cable 5-pin, length 25 m PVC-cable with M12-connection, brass nickel-plated, IP 67, shielded M12-PVC / 4G-5 m PVC-cable 4-pin, length 5 m M12-PVC / 4G-10 m PVC-cable 4-pin, length 10 m M12-PVC / 4G-25 m PVC-cable 4-pin, length 25 m M12-PVC / 5G-5 m M12-PVC / 5G-10 m M12-PVC / 5G-25 m PVC-cable 5-pin, length 5 m PVC-cable 5-pin, length 10 m PVC-cable 5-pin, length 25 m M12 plug-in screw cap M12-EVK M12 plug-in screw cap, 1.4305 (303), with o-ring, as a protection against humidity and dirt CERT / 2.2 CAL / ILM factory certificate 2.2 acc. to EN10204 (only product contacting surface) factory calibration certificate for ILM 29.08.13 / 2.1 / TB / at-ac.de NEGELE MESSTECHNIK GMBH Raiffeisenweg 7 87743 Egg an der Guenz Phone +49 (0) 83 33. 92 04-0 Fax +49 (0) 83 33. 92 04-49 sales@anderson-negele.com Tech. Support: support@anderson-negele.com Phone +49 (0) 83 33. 92 04-720