Power Supply Rejection Ratio Measurement

Similar documents
Power Supply Rejection Ratio Measurement

Power Supply Rejection Ratio Measurement

Power Supply Rejection Ratio Measurement

Battery Impedance Measurement

Invasive and Non-Invasive Stability Measurements

Making Invasive and Non-Invasive Stability Measurements

Solar Cell Impedance Measurement using the Bode 100

DC Biased Impedance Measurement

Measure Low Value Impedance Current Shunt Impedance

Low Value Impedance Measurement using the Voltage / Current Method

Opamp stability using non-invasive methods

DC/DC Converter Stability Measurement

DC/DC Converter Stability Measurement

Measuring Power Line Impedance

How to Measure LDO PSRR

Signal Injectors. Documentation. Version 1.00, October, Picotest Corp. All Rights Reserved.

Lab 9 Frequency Domain

DC/DC Converter Stability Measurement

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

Test No. 1. Introduction to Scope Measurements. Report History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 1

DEPARTMENT OF INFORMATION ENGINEERING. Test No. 1. Introduction to Scope Measurements. 1. Correction. Term Correction. Term...

Exclusive Technology Feature. An Accurate Method For Measuring Capacitor ESL. ISSUE: April by Steve Sandler, Picotest, Phoenix, Ariz.

VVM measurement with E5061B for replacing 8508A vector voltmeter. May 2013 Agilent Technologies

Measuring LDOs requires more bandwidth than you think

Audio Amplifier Frequency Response

Documentation. Voltage Regulator Test Standard. Test Platform for Voltage Regulator and LDO Testing

Measurement of the equivalent circuit of quartz crystals

Test No. 2. Advanced Scope Measurements. History. University of Applied Sciences Hamburg. Last chance!! EEL2 No 2

Input Impedance Measurements for Stable Input-Filter Design

DC Biased Impedance Measurements MOSFET

Smart Measurement Solutions. Bode 100. User Manual

3.2 Measuring Frequency Response Of Low-Pass Filter :

Effectively Using the EM 6992 Near Field Probe Kit to Troubleshoot EMI Issues

Documentation. Voltage Regulator Test Standard. Test Platform for Voltage Regulator and LDO Testing

Background (What Do Line and Load Transients Tell Us about a Power Supply?)

MEASURING HUM MODULATION USING MATRIX MODEL HD-500 HUM DEMODULATOR

Equivalent Circuit Determination of Quartz Crystals

Measuring Impedance with the Bode 100. OMICRON Lab Webinar Nov. 2014

PHYS 3322 Modern Laboratory Methods I AC R, RC, and RL Circuits

When you have completed this exercise, you will be able to determine the frequency response of an

Harmonic Comb Injector

Bode 100. User Manual

Figure E2-1 The complete circuit showing the oscilloscope and Bode plotter.

Equivalent Circuit Determination of Quartz Crystals

Transformer modelling

Virtual Lab 1: Introduction to Instrumentation

Build Your Own Bose WaveRadio Bass Preamp Active Filter Design

ALX-SSB 5 Band Filter Assembly Manual 19 November 2018

Experiment 8 Frequency Response

A Walk Through the MSA Software Vector Network Analyzer Transmission Mode 12/18/09

SIMULATION OF A SERIES RESONANT CIRCUIT ECE562: Power Electronics I COLORADO STATE UNIVERSITY. Modified in Fall 2011

Demo / Application Guide for DSA815(-TG) / DSA1000 Series

Demo Circuit DC550A Quick Start Guide.

S-Parameter Measurements with the Bode 100

Picotest s Power Integrity Workshop

Vector Network Analyzers (VERY) Basics. Tom Powers USPAS SRF Testing Course 19 Jan. 2014

ET 304A Laboratory Tutorial-Circuitmaker For Transient and Frequency Analysis

Lab #5 Steady State Power Analysis

Evaluation Board for ADP2118 EVAL-ADP2118

Simulating Circuits James Lamberti 5/4/2014

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

2. BAND-PASS NOISE MEASUREMENTS

UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE Department of Electrical and Computer Engineering

Low voltage LNA, mixer and VCO 1GHz

Core Technology Group Application Note 6 AN-6

VCC. Digital 16 Frequency Divider Digital-to-Analog Converter Butterworth Active Filter Sample-and-Hold Amplifier (part 2) Last Update: 03/19/14

Preamplifier Options for Reducing Cable-Braid Loop Error

Pre-Lab. Introduction

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Contactless RFID Tag Measurements

ECE 3155 Experiment I AC Circuits and Bode Plots Rev. lpt jan 2013

Methods for Reducing Emissions from Switching Power Circuits. A. McDowell, C. Zhu and T. Hubing

Keysight Technologies Making Accurate Intermodulation Distortion Measurements with the PNA-X Network Analyzer, 10 MHz to 26.5 GHz

ELC224 Final Review (12/10/2009) Name:

Advanced Measurements

Review: The MFJ-223 Vector Impedance Antenna Analyzer Phil Salas AD5X

Filterless 3W Class- D Mono Audio Amplifier

Filters And Waveform Shaping

10 CHANNELS, Coupling & Probes

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Keysight Technologies How to Make the Best Switch Mode Power Supply Measurements. Application Note

PSM Soft. Features and Functions January PC Software Guide. Getting connected and Communication

1000BASE-T1 EMC Test Specification for Common Mode Chokes

Oscilloscope and Function Generators

Low_Pass_Filter_1st_Order -- Overview

12. Output Ripple Attenuator Module (MicroRAM )

Hot S 22 and Hot K-factor Measurements

FREQUENCY RESPONSE AND PASSIVE FILTERS LABORATORY

An LDO Primer. Part III: A Review on PSRR and Output Noise

PDN Probes. P2100A/P2101A Data Sheet. 1-Port and 2-Port 50 ohm Passive Probes

EK307 Passive Filters and Steady State Frequency Response

Keysight E5061B ENA Vector Network Analyzer CONFIGURATION GUIDE

Testing Power Sources for Stability

EXPERIMENT NUMBER 10 TRANSIENT ANALYSIS USING PSPICE

Lab 10: Oscillators (version 1.1)

1GHz low voltage LNA, mixer and VCO

Analysis of RF transceivers used in automotive

Circuit Characterization with the Agilent 8714 VNA

Non Invasive Assessment of Voltage Regulator Phase Margin without Access to the Control Loop

Uncle Sparky s Guide to Voltage, Current, and Resistance Measurements. Spring 2014

Transcription:

Power Supply Rejection Ratio Measurement Using the Bode 100 and the Picotest J2120A Line Injector www.telesplicing.com.tw +886-2-27053146 sales@telesplicing.com.tw

Page 2 of 10 Table of Contents 1 EXECUTIVE SUMMARY... 3 2 MEASUREMENT TASK... 3 3 MEASUREMENT SETUP & RESULTS... 4 3.1 MEASUREMENT SETUP... 4 3.2 DEVICE SETUP... 6 3.3 CALIBRATION... 7 3.4 MEASUREMENT... 8 4 CONCLUSION... 9

Page 3 of 10 1 Executive Summary This application note shows how the Power Supply Rejection Ratio, or PSRR of a linear voltage regulator (TIP120) can be measured using the Bode 100 and additional accessories. The same techniques can be used to measure switching regulators as well. The measurements are performed on the PICOTEST Voltage Regulator Test Standard (VRTS) testing board Rev. 1.5 using the PICOTEST J2120A Line Injector. 2 Measurement Task The PSRR of the TIP120 linear voltage regulator is measured with the Bode 100 and the PICOTEST J2120A line injector. A capacitor is then connected to the output of the regulator and the PSRR is again measured from 10 Hz to 10 MHz. The PICOTEST VRTS 1.5 is used as the basis for the testing. The VRTS 1.5 can be used to help perform most of the common voltage regulator measurements using the Bode 100 in conjunction with the PICOTEST Signal Injectors. The VRTS 1.5 board includes the regulator and output capacitors as well as a load used for the measurements in this application note. Figure 1: Voltage Regulator Test Standard board (VRTS 1.5)

Page 4 of 10 3 Measurement Setup & Results The PSRR describes how a signal on the DC input voltage of the regulator system is transmitted to the regulated output. The PSRR is generally measured in db and defined to be: PSRR = 20 log v OUT v IN Where v OUT and v IN are the AC ripple of the input and output voltage, respectively. PSRR V IN + v IN Voltage Regulator V OUT + v OUT GND Figure 2: Measurement principle Depending on the definition the PSRR can be a negative or positive number. Using the above definition, the PSRR generally is a negative number. 3.1 Measurement Setup The PSRR can be measured by applying a sinusoidal ripple on the supply voltage and measuring the gain from input to output of the regulator. The PICOTEST J2120A Line Injector allows you to add the sinusoidal output voltage of the Bode 100 to the DC-supply voltage of the regulator. The PSRR is then measured by connecting two voltage probes to the input and output of the regulator.

Page 5 of 10 The following figure shows the principle measurement setup: Figure 3: PSRR measurement setup Note: The output impedance of the J2120A is slightly resistive. An input capacitor on the regulator would, therefore, create a low pass network and change the PSRR results. The measurements shown in this application note are performed without an input capacitor! The voltage probes and injectors are connected to the Bode 100 and the VRTS board as shown in the following picture: Figure 4: PSRR example measurement setup

Page 6 of 10 Figure 5: PSRR example measurement setup close-up Note: The resistor R6 has to be enabled by the switch S1 to get a load. Later on, C2 will be enabled too. 3.2 Device Setup The PSRR measurement can be performed directly with the Bode 100 using the Gain / Phase measurement type. Figure 6: Start menu

Page 7 of 10 The Bode 100 is set up as follows: Start Frequency: Stop Frequency: Sweep Mode: Number of Points: Level: Attenuator 1 & 2: Receiver Bandwidth: 10 Hz 10 MHz Logarithmic 401 or more -10 dbm 10 db 100 Hz Note: When the Bode 100 is used with the J2120A the output level should be set in the range from -20 to 10 dbm. The PSRR measurement is a small signal measurement and so the goal is only to maintain a level above the noise floor. Figure 7: Settings Trace 1 Trace 1 format is set to Magnitude (db). 3.3 Calibration Before we start the measurement, we need to calibrate the Bode 100. This will provide the accuracy of the measurement. To do so, press the Full Range calibration button. Figure 8: perform calibration The window called Full Range Calibration opens where you are able to perform the calibration. Figure 9: Full Range Calibration window Now connect OUTPUT, CH1 and CH2 as shown below and perform the calibration by pressing the Start button.

Page 8 of 10 Figure 10: Calibration setup 3.4 Measurement Performing a single sweep leads to the following PSRR curve: Figure 11: PSRR curve At low frequencies the PSRR is very high which results in high suppression of disturbances from the supply line. In the higher frequency area of > 1 MHz the PSRR gets pretty small. To see the influence of an output capacitor on the PSRR the VRTS capacitor C2 is additionally disabled by the switch on the board. This 0.1 µf capacitor changes the PSRR curve as follows:

Page 9 of 10 Figure 12: PSRR curve - comparison Note: The differences between the measurement with and without an output capacitor are even bigger if the PSRR is worse (closer to 0 or even >0). 4 Conclusion The Bode 100 in combination with the J2120A Line Injector offers a test set that enables simple and fast PSRR measurements in a wide frequency range starting at 10 Hz and reaching 10 MHz. No injection transformer is necessary for the test set. This setup allows you to measure the PSRR even in systems with high DC currents up to 5 A without the danger of destroying an expensive transformer.