Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center. Thursday September 8, 2011

Similar documents
Quality control of Gamma Camera. By Dr/ Ibrahim Elsayed Saad 242 NMT

GS Introduction to Medical Physics IV Laboratory 5 Gamma Camera Characteristics

NM Module Section 2 6 th Edition Christian, Ch. 3

Initial Certification

Acceptance Testing and Annual Physics Survey Recommendations for Gamma Camera, SPECT, and SPECT/CT Systems

Investigation of Multiple Head Registration / Center of Rotation for SPECT Gamma Cameras

Parameters Affecting on Intrinsic Uniformity Test For MEDISO

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

Radionuclide Imaging MII Single Photon Emission Computed Tomography (SPECT)

Nuclear Associates , &

Results of the Measurement of the Collimator Hole Angulation for Different Collimators of SPECT with Adaptive Quality Control Phantom

Nuclear medicine is critically dependent on the accurate,

2/14/2019. Nuclear Medicine Artifacts. Symmetric energy windows

T h e P h a n t o m L a b o r a t o r y

Intrinsic and Tomographic Evaluation of Siemens e.cam SPECT System at the Korle-Bu Teaching Hospital (Ghana)

Contrast Evaluation for a Dual Head SPECT System with Different Energy Peaking- Drift

How Gamma Camera s Head-Tilts Affect Image Quality of a Nuclear Scintigram?

Factors Affecting the resolution of SPECT Imaging. h.

J of Nuclear Medicine Technology, first published online July 27, 2011 as doi: /jnmt

QUALITY CONTROL OF IMAGING DEVICES. P.S. Soni

... In vivo imaging in Nuclear Medicine. 1957: Anger camera (X;Y) X Y

CHAPTER 8 GENERIC PERFORMANCE MEASURES

Appendix I: Artefacts and Trouble-shooting

QC Testing for Computed Tomography (CT) Scanner

ACR Update in Nuclear Medicine Accreditation

Quality control in dual head γ- cameras: Comparison between methods and softwares used for image analysis

Acceptance Testing of a Digital Breast Tomosynthesis Unit

C a t p h a n. T h e P h a n t o m L a b o r a t o r y. Ordering Information

LaBr 3 :Ce, the latest crystal for nuclear medicine

CHAPTER 15 DEVICES FOR EVALUATING IMAGING SYSTEMS

The future of nuclear imaging is clear

Reconstruction Filtering in Industrial gamma-ray CT Application

International Journal of Scientific & Engineering Research, Volume 4, Issue 9, September ISSN

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

PHYSICS ADVANCED LABORATORY I COMPTON SCATTERING Spring 2002

First Applications of the YAPPET Small Animal Scanner

Nuclear Associates

Master of Science Thesis. SIMIND Based Pinhole Imaging

2010 Philips BrightView XCT SPECT/CT

Nuclear Associates , &

DESIGN AND OPERATION OF A WIDE RANGE SEGMENTED GAMMA RAY SCANNING ASSAY INSTRUMENT FOR THE MEASUREMENT OF BOTH LOW AND INTERMEDIATE LEVEL WASTE

SPECT Reconstruction & Filtering

Application Note. ipix A Gamma imager to support various applications. Introduction. An easy to carry and deploy instrument

Chiara Secco. PET Performance measurements of the new LSO-Based Whole Body PET/CT. Scanner biograph 16 HI-REZ using the NEMA NU Standard.

Nuclear Associates

PET Detectors. William W. Moses Lawrence Berkeley National Laboratory March 26, 2002

Industry Breakthrough

LaBr 3 :Ce scintillation gamma camera prototype for X and gamma ray imaging

An Activity in Computed Tomography

Primer on molecular imaging technology

Positron Emission Tomography

An Activity in Computed Tomography

Photomultiplier Tube

Slide 1. Slide 2. Slide 3 ACR CT Accreditation. Multi-Slice CT Artifacts and Quality Control. What are the rules or recommendations for CT QC?

Testing and Performance Validation of a Shielded Waste Segregation and Clearance Monitor Designed for the Measurement of Low Level Waste-13043

Pinhole collimator design for nuclear survey system

NON-UNIFORM ATTENUATION CORRECTION USING SIMULTANEOUS TRANSMISSION AND EMISSION CONVERGING TOMOGRAPHY

Nuclear Associates , , CT Head and Body Dose Phantom

Industry Breakthrough

CZT Technology: Fundamentals and Applications

IAEA Coordinated Research Project on Development of Harmonized QA/QC Procedures for Maintenance and Repair of Nuclear Instruments

- Nuclear Radiation Sensor - Detects Beta and Gamma Radiation and X-Rays

3/8 or 1 Free-Geometry Dual-Detector Cameras

STEREOTACTIC BREAST BIOPSY EQUIPMENT SURVEYS

CHAPTER 2 COMMISSIONING OF KILO-VOLTAGE CONE BEAM COMPUTED TOMOGRAPHY FOR IMAGE-GUIDED RADIOTHERAPY

STANDARD OPERATING PROCEDURES SOP:: 2057 PAGE: 1 of 6 REV: 0.0 DATE: 07/11/03

Quality Control for Stereotactic Breast Biopsy. Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

Fundamentals of Positron Emission Tomography (PET)

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

LUDLUM MODEL MODEL AND MODEL GAMMA SCINTILLATORS. June 2017

Changing the Shape of Nuclear Medicine

ipix Gamma Imager Product Introduction Steve Laskos Product Management Director

Surveying and QC of Stereotactic Breast Biopsy Units for ACR Accreditation

Nathan Childress, Ph.D., DABR

TABLE OF CONTENTS. References

Gamma-ray spectral imaging using a single-shutter radiation camera

New Technology in Nuclear Medicine

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

DISC QC/QA Program for Digital Imaging Systems using the DR Radchex Plus Meter

ISO-CART-85. Mobile Low-Level Waste Assay System

16 Instrumentation and Data Acquisition

HISTORY. CT Physics with an Emphasis on Application in Thoracic and Cardiac Imaging SUNDAY. Shawn D. Teague, MD

Exprerimental Evaluation of a Dedicated Pinhole SPECT System for Small Animal Imaging and Scintimammography

Test Equipment for Radiology and CT Quality Control Contents

Technical Description and Operator s Manual. Survey Meter SM 5 D. Sensortechnik und Elektronik Pockau GmbH. Siedlungsstraße 5-7 D Pockau

Victoreen , H, A Gamma Scintillation Detector

Breast Tomosynthesis. Bob Liu, Ph.D. Department of Radiology Massachusetts General Hospital And Harvard Medical School

FRCR Nuclear Medicine

Y11-DR Digital Radiography (DR) Image Quality

236 IEEE TRANSACTIONS ON NUCLEAR SCIENCE, VOL. 59, NO. 1, FEBRUARY 2012

As the role of gamma cameras expands in positron

Ensuring Shielding adequacy in Lead shielded spent fuel transportation casks using gamma scanning

Development of the LBNL Positron Emission Mammography Camera

Practical Aspects of Medical Physics Surveys of Mammography Equipment and Facilities

Cartogam Real-Time Portable Gamma-Ray Imaging System

An Introduction to TG-142 Imaging QA Using Standard Imaging Products. Mark Wiesmeyer, PhD, DABR Technical Product Manager Standard Imaging, Inc.

T h e P h a n t o m L a b o r a t o r y

Chemical Engineering 412

Transcription:

Robert Pagnanelli BSRT(R)(N), CNMT, NCT, FASNC Chief Technologist, Nuclear Imaging Duke University Medical Center Thursday September 8, 2011

Quality Control Quality control should be performed because: Detector damage High Voltage drift Electronic component degradation Quality control is necessary to assure accurate results

Equipment Requiring QC Gamma Camera Survey Meter Dose Calibrator Other lab equipment

Quality Control Procedures- Planar camera Energy Peaking daily Uniformity test daily Linearity and resolution weekly Sensitivity test weekly

Energy Peaking Verifying that the pulse height analyzer s energy window over the photopeak energy Best performed 5 UFOVs from the uncollimated camera with a point source Can use a sheet source with collimated camera Most commonly performed with a collimator Often automated must check!

Uniformity Test Best performed 5 UFOVs from the uncollimated camera Most commonly performed with a Co-57 sheet source on a collimated camera Also checks the collimator MUCH faster Suggestion Perform extrinsically daily Perform intrinsically weekly

Uniformity Test Extrinsic (with collimator) Co-57 radioactive sheet source is placed on top of each detector with the collimator still in place. An image is then taken on each detector for 3-5 million counts. 256x256 matrix A visual assessment (qualitative) of the image is to be made as well as a quantitative assessment.

Co-57 Sheet Source Dimensions 17 x 11 cm - 25 x 18 cm 5 20 mci $3000.

Uniformity Test Integral uniformity Global parameter measuring uniformity over an extended area 100% x (max-min)/(max+min) Differential uniformity Regional parameter that measures uniformity over a small area Performed using all 5x1-pixel area in both the X and Y direction 100% x Largest deviation (max-min)/(max+min)

Performance parameters for detectors- ASNC guidelines PARAMETER STANDARD PREFERRED Integral Uniformity Differential Uniformity <5 <3 <5 <3

Uniformity Test

Uniformity Test Non-Uniformity caused by Malfunctioning PMT Off-peak/Wrong peak Cracked crystal

Uniformity Test ABOVE PEAK BELOW PEAK

Resolution and Linearity Performed to Document spatial resolution and its change over time Evaluate ability to image straight lines Image a flood source through a spatial resolution test phantom (using uniformity flood paramaters) If a 4-quadrant phantom is used, it should be rotated 90 degrees so that every 5 th time the pattern repeats. Evaluation How straight are the lines What are the smallest bars discerned

Resolution and Linearity

Sensitivity Test Documents the system sensitivity as well as the change of sensitivity over time. Cts/pixel/millicurie To measure change in sensitivity: record the time taken to acquire a preset number of counts using a set dose.

INSTRUMENTATION ARTIFACTS NON-UNIFORMITY Ring Artifact Uniform Flood Non-Uniform Flood FBP

Quality Control Procedures- SPECT camera COR weekly or monthly Uniformity correction manufacturer recommendation Plexiglas phantom quarterly or annually

COR Checks for alignment error between the electronic matrix of the detector and the mechanical COR. Always check after camera servicing and software upgrades. Doughnut artifact (360º) or tuning fork artifact (180º) in the transverse images

COR COR is a measure of the alignment of opposite views of camera head position Misalignment (Offset camera and center of acquisition matrix) results in donut or tuning fork artifact

COR A 1 mci point source of Tc99m is placed on camera table SPECT data is acquired over a 360 (or 180) degree orbit 60 projections per head 5-10 sec/frame

COR

Plexiglass SPECT Phantoms

Plexiglass SPECT Phantoms To determine 3D contrast, resolution, uniformity 1. Collimator - LEHR (or LEUHR if used routinely with Tc-99m, whichever is higher resolution) 2. Tc-99m Activity - Between 10-20 mci 3. Count Rate needs to be below 50 kcps 4. Matrix size = 128x128 5. Rotation = 360 degrees 6. Total Views = 120. (Implies 60 steps for dual head, 3-degrees per step) 7. Dual Headed Camera study acquired with camera head in opposed position (180) 8. Acquire for approximately 32 million total counts 9. Radius of Rotation = as close to 20 cm as possible. 10. Orbit = Circular

Plexiglass SPECT Phantoms

High Count Extrinsic Flood-Field Uniformity Correction In SPECT it is assumed that efficiency of photon detection is uniform across the collimated detector Camera systems store flood-field correction maps to correct variations before reconstruction 30 to 100 million ct images for each collimator 256x256 matrix

Image Artifact

Survey Meter Used daily to detect radioactive contamination in laboratory Aka Geiger counter, Geiger-Mueller counter, GM counter

GM Counter Battery Check Daily Ensures that the meter has adequate electrical power. The meter located in the readout scale will be located within the battery region if the power is adequate.

GM Counter Source check Daily Ensure that the meter responds to a radiation source.

GM Counter Calibration Annually Ensures that the meter responds accurately to radiation.

Dose Calibrator Used to measure the activity of the radiopharmaceutical

Dose Calibrator Standard testing Background Voltage Accuracy Constancy Linearity Geometry

Dose Calibrator Background Performed daily Background activity is checked at each setting that will be utilized Contamination Unshielded source in lab

Dose Calibrator Voltage Performed daily Ensures that the operating voltage is within manufacturers recommended range

Dose Calibrator Accuracy Performed Annually Two sealed radiation sources greater than 50 µci and one with an energy between 100 to 500 kev are assayed Isotopes used: Co-57 (122 kev, 271 days) Cs-137 (662 kev, 30 years) Ns-22 Co-60, Ge-68, or Ba-133.

Dose Calibrator The sources must be traceable to The National Institute of Standards and Technology (NIST) Compared to decay corrected calibrated activity Results must be within 5%

Dose Calibrator Geometry Performed at installation, recalibration, etc. Fine print (plug-in) Assess accuracy of the dose calibrators ability to measure consistent activity with different volumes A given amount of radioactivity is assayed in a container and the activity is recorded. Saline is incrementally added to include range typically used in lab Measurements must be within 5%.

Dose Calibrator Constancy Performed daily Designed to show that a long half-life source yields reproducible readings Compared to previous days reading Within 5% of decay corrected accuracy readings Check each setting (isotope) button Activity used should be greater than 50 µci of Co-57 or Cs-137 (or another gamma emitting sealed source)

Dose Calibrator Constancy SETTING MON TUES WED THURS FRI Cs-137 123 124 122 125 124 Tc-99m 243 244 246 244 244 Tl-201 163 164 162 165 164 I-123 194 194 196 195 195

Dose Calibrator Linearity Performed quarterly Assess ability to read both high and low activity Measured values must be within 5% of calculated values Two methods for linearity Decay method Sleeve method

Dose Calibrator Decay method Measure an initial high activity of a short halflife source. Various measurement are taken as it decays. Readings are compared to calculated decay. Elapsed Time (hr) Expected Reading (mci) Measured Reading (mci) 0 300 300 1 267 272 2 238 241 3 212 210 6 150 148 12 75 72.5 24 18.75 19.1 48 1.17 1.19 72 0.073 0.074 78 0.036 0.037

Dose Calibrator Sleeve method Decay is simulated by using lead sleeves of various thickness.

Dose Calibrator Subpart C--General Technical Requirements 35.60 Possession, use, and calibration of instruments used to measure the activity of unsealed byproduct material (a) For direct measurements performed in accordance with 35.63, a licensee shall possess and use instrumentation to measure the activity of unsealed byproduct material before it is administered to each patient or human research subject. (b) A licensee shall calibrate the instrumentation required in paragraph (a) of this section in accordance with nationally recognized standards or the manufacturer's instructions. (c) A licensee shall retain a record of each instrument calibration required by this section in accordance with 35.2060.

Other Lab Equipment Glucometer Daily calibration Accuracy Defibrillator Battery Test Voltage Test

Be Alert All Day! Raw patient data should be carefully reviewed on each patient. All problems do not happen at night Cameras Survey meters Dose calibrator

THE END