ThermoPLUS Smart Temperature Transmitter

Similar documents
JTD Series of Differential Pressure Transmitters

600T Series Transmitters. Specification Sheet. Model 653T Electronic temperature transmitter (PC programmable)

ST3000 ACE Smart Transmitter JTD Series of Differential Pressure Transmitters. Model JTD910A/920A/930A/960A/921A/931A/961A

TEK-TEMP 2100A. Explosion-Proof Temperature Transmitter. TEMPERATURE. Technology Solutions

PROGRAMMABLE FIELD TEMPERATURE TRANSMITTER

General Specifications

600T Series Transmitters Model 653L Electronic temperature transmitter (PC programmable for RTD and resistance) Specification Sheet

MODEL: B6U-B B6-UNIT. [2] LCD DISPLAY 0: Without 1: With

STT 3000 Smart Temperature Transmitter

General Specifications

STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C

} Max. Load Resistance : 750Ω when power supply voltage is 24V DC (Max. load resistance (Ω) = 50x Power Supply Voltage -450 ).

SEM210 SERIES. Programmable In-head Universal Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

ATT2100/2200 SMART TEMPERATURE TRANSMITTER. Application Areas: ATT 2200 ATT2100

General Specifications

MODEL: 27HU-B 27-UNIT. [3] LED INDICATOR 0: Without 1: With (SIL not available)

General Specifications

0 2 1 â Ó. itemp HART TMT182. Technical Information

PAD - Heavy Duty Differential Pressure Trasmitter

SITRANS T measuring instruments for temperature

MODEL: M3LU2. OUTPUT Field-selectable Current 0 20 ma DC Voltage V DC V DC

OPTITEMP TT 20 Technical Datasheet

The Rosemount 848T Wireless Temperature Transmitter

General Specifications

Alphaline Temperature Transmitters

Temperature Measurement

General Specifications

Datasheet Platinum Resistance Pt100 In-head (Push Button) Temperature Transmitter TX203P mA output, default range supplied C

0 2 1 â Ó. itemp HART TMT182. Technical Information

General Specifications

Two-Wire Programmable Transmitter for:

SMART TEMPERATURE TRANSMITTERS. smar

MODEL: 27HU 27-UNIT MODEL: 27HU [1][2]

MODEL: M3LU. OUTPUT Field-selectable Current 0 20 ma DC Voltage V DC V DC

HYBRID RECORDER OUTLINE

MODEL: B3HU. Space-saving Two-wire Signal Conditioners B3-UNIT

MODEL: R30US4. Remote I/O R30 Series

PRODUCT SPECIFICATION AND MODEL SELECTION GUIDE

General Specifications

SHIMADEN DIGITAL INDICATORS

General Specifications

Miniature resistance thermometer Explosion-protected version Model TR34, thread-mounted

Dissolved Oxygen Transmitter

FKP... 5 PRESSURE TRANSMITTER (DIRECT MOUNT TYPE) FEATURES SPECIFICATIONS. Functional specifications

APT3200L SMART PRESSURE TRANSMITTER. Application Areas: WITH DIAPHRAGM SEAL

DIGITAL PROGRAM CONTROLLER

General Specifications

Advanced Temperature Transmitter

DOC. NO. : C3200-E04A APPROVED. Smart PressureTransmitter for Gauge and Absolute Pressure Measurement MODEL APT3200

LEVEL TRANSMITTER <SANITARY TYPE>

PTG series Smart Pressure Transmitter

SPECIFICATION SHEET DISSOLVED OXYGEN ANALYZER/TRANSMITTER SYSTEM CONFIGURATION

Temperature Input Module for Zone 1 Series 9482/32

STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C

Model 3244MV. MultiVariable Temperature Transmitter with Profibus PA. Product Data Sheet

SHIMADEN DIGITAL INDICATORS

General Specifications

SEM104 SERIES. SEM104P Pt100 Temperature Transmitter. SEM104TC Thermocouple Temperature Transmitter INDEX SECTION CONTENTS PAGE NO.

SIA, SIB, SIC, SID. User's Manual

Series 10-ZSE80/10-ISE80

C330 / C330X PC-Programmable Universal, 2-wire Transmitter

CALYS 150. Advanced documenting multifunction calibrator thermometer

Multifunction Pulse-to-Pneumatic Converter

Temperature Measurement Transmitters for mounting in sensor head

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

C520 / C520X /C520S / C520XS HART Compatible, Universal, Dual-input 2-wire Transmitters

Rosemount 848L Logic Transmitter with FOUNDATION Fieldbus

M1 Series. Humidity - Temperature Transmitter INSTRUCTION MANUAL

Dissolved Oxygen Transmitter (For Low-Concentration Measurement) OBM-165H

Roline L1 Series. Humidity - Temperature Transmitters INSTRUCTION MANUAL

Temperature Measurement

Temperature Measurement Transmitters for mounting in sensor head

GS.No.GBN062E-5. GENERAL SPECIFICATIONS Sensor unit

OEM miniature resistance thermometer Models TR31-3 and TR31-K, thread-mounted

Digital Temperature Transmitters Model T12.10, Universally Programmable, Head Mounting Model T12.30, Universally Programmable, Rail Mounting

Mass and Volume Flow Transmitter

Introduction. Specifications LF620,LF622. Field Intelligent Device Series Electromagnetic Flowmeter Converter

PRESSURE TRANSMITTER (DIRECT MOUNT TYPE)

Temperature transmitter itemp PCP DIN rail TMT 121

Fuzzy Temperature Controllers E5EF

Resistance thermometer Model TR30, compact design

General Specifications

AT9000 Advanced Transmitter Gauge Pressure Transmitters In-line model

FieldIT Temperature Transmitter, field mounted TH202/TH202-Ex

CALYS 150. Advanced documenting multifunction calibrator thermometer

TF 02 / TF 02-Ex. Head mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels

General Specifications

General Specifications

LST100 Ultrasonic level transmitter for upstream oil and gas

CALYS 75. Field documenting multifunction calibrator

CALYS 50. Field multifunction calibrator for basic use

Field IT 2600T Series Pressure Transmitters

FKB... 4 Pressure Transmitter <Sanitary Type> (Remote Seal Type)

Rosemount 148 Temperature Transmitter

Operator Manual. Transmitters Series KAT-... Analog transmitters in 2-wire technique

TT2780. Temperature head transmitter For resistance thermometers Pt100, settable using a PC, for installation in a sensor head Form B, BUZ, BUZ-H

FieldIT Temperature Transmitter, rail mounted TH102/TH102-Ex

General Specifications

MODEL: M6SXU MODEL: M6SXU [1][2] [3][4]

JUMO dtrans T01 HART / T01T HART / T01 Junior / T01 Ex / T01 HART Ex Programmable 2-wire transmitter

Transcription:

No. SS2-TT110-0100 ThermoPLUS Smart Temperature Transmitter Model TT61/71 (Integral Sensor type) OVERVIEW The models TT61/71 Smart Transmitter are temperature transmitters for the model TT90 integral temperature sensor and models TT9 protecting wells for the temperature sensor. It converts and transmits the input signal from a thermocouple or resistance thermometer into a 4~20m C or a digital signal. The SFC (Smart Field Communicator) facilitates remote setting of various parameters, calibration or self-diagnostics. The models TT61/71 Transmitter, consisting of a sensor and a transmitter have been certified to conform with the JIS Flameproof structure standard. For this reason, use of Yamatake's Model TT sensor are highly recommended. Combination of TT71 temperature transmitter and TT90 temperature sensor (Sold separately) *1 FETURES 1. Vastly improves maintenance efficiency. The SFC facilitates setting, calibration and self-diagnostics of the transmitter from the control room 2. Reduces transmitter inventory Single model accepts wide range of sensor types. 3. Reduces installation costs. Two-wire system eliminates the need for compensating lead wire or three-core shield cable from site to control room. 4. Self-diagnostic functions facilitates simple operation: bnormal ambient temperature: If deviates from the range of -40 ~ +85 C (Current ambient temperature can be monitored) bnormal cold junction: If platinum temperature sensor for cold junction becomes abnormal bnormal input: If temperature sensor output shows irregularity Open thermocouple wire: If thermocouple wire becomes open 5. n optimal range for process operation can be selected. The maximum/minimum holding function allows readout if the range is exceeded. 6. Useful integral indicator Integral digital indicator provides both field indication and signal transmission. 7. Explosion-proof model can be used in an acetylene or a hydrogen atmosphere The transmitter has been certified for JIS Flameproof, Exd IIC T6. Note) *1 External appearance may differ from those shown in the photographs in this Specification Sheet depending on type of sensors used. Combination of TT71 temperature transmitter and TT90 temperature sensor with TT92 sensor protecting well (Sold separately) *1 Combination of Temperature TransmitterTT71, Temperature Sensor TT90 (Sold separately) with Sensor Protecting WellbTT95 (Sold separately) *1 SFC000010000P SFC Smart Communicator (Sold separately) Specifications are subject to change without notice. - 1 - First issue: Jul. 2002

No. SS2-TT110-0100 STNR SPECIFICTIONS Converter Input type (Select by SFC) Thermocouple SJ(J), SK(K), ST(T) SE(E) SN(N), SR(R), SS(S), S() Resistance temperature detector (RT) Pt100, Jpt100 mv Measurable range ll temperature ranges listed in official JIS Table (For details, refer to Table 1) -20~120mV urn-out Up-scale (over 21.1m C) or down-scale (under 3.6m C) Cold joint compensation External/internal selectable from an SFC Cold joint compensation accuracy ±0.5 C Conversion accuracy epends on the sensor type and measurement span. (For details refer to Table 2.) Output nalog (4~20m C), or digital (E protocol) selectable by SFC igital indication n optional LC display is in 4.5-digit with an engineering unit (degree C) including 0-100% bar graph. Power supply/load resistance 17.5~42V C/250~1271 (For details, refer to Table 2) Input/output insulation 500V C mbient temperature range Waterproof model: Without LC display: -40~85 C With LC display: -20~70 C JIS Flameproof model: -20~60 C regardless of LC display mbient humidity 5~100% RH (No condensation) dditional features Influence of ambient temperature ±0.5% F.S. (30 C) Influence of ambient humidity ±0.01%F.S. (5~100% RH) Influence of power supply voltage ±0.1%F.S. (@17~42V C) Yamatake Corporation Construction Case luminum alloy Waterproof model JIS C0920 immersion-proof compatible NEM Type 4X, IEC IP6 equivalent JIS Flameproof model JIS Flameproof Ex d II T6 (mbient temperature -20~60 C, Flameproof range: -20~85 C) Installation Horizontal or on 2 in. vertical pipe pedestal Painting Metallic green (Munsell 5G7/8) Weight pprox. 830g (w/integral LC and sensor, w/o bracket) Setting at factory shipment (ll settings can be modified with SFC (sold separately)) Sensor model Pt100 (by specifying 'C' of additional specifications) Temperature range 0~100 C (by specifying 'C' of additional specifications) TG No. XXXXXXXX (up to 8 digits as specified) Output characteristics Linearized amping constant 0.0sec Cold junction compensation Internal cold junction Table 1 Measurable range Type of sensor Guaranteed accuracy range *1 SFC's setting range *2 pplicable standards SJ(J) -50~1200 C -200~1200 C SK(K) -170~1250 C -200~1370 C ST(T) -120~400 C -250~400 C SE(E) -100~1000 C -200~1000 C SN(N) 0~1300 C -200~1300 C JIS C1605-1995 SR(R) 0~1760 C -50~1760 C SS(S) 0~1760 C -50~1760 C S() 400~1820 C 200~1820 C Pt100-200~850 C -200~850 C JIS C1604-1997 JPt100-200~640 C -200~640 C JIS C1604-1989 mv -20~120mV -10~120mV -- Note) 1. Guaranteed accuracy means the range of signal conversion that is guaranteed. Measurable range is governed by sensor that is used. Note) 2.SFC's setting range is the output range that can be set by SFC (sold separately). - 2 -

Yamatake Corporation No. SS2-TT110-0100 Table 2 Conversion accuracy Type of sensor Note) igital accuracy SJ(J) ±0.3 C SK(K) ±0.4 C ST(T) ±0.4 C SE(E) ±0.3 C SN(N) ±0.5 C SR(R) ±0.6 C SS(S) ±0.6 C S() ±0.8 C Pt100 ±0.2 C JPt100 ±0.2 C mv ±0.05%rdg, or ±50µV, whichever is larger Internal cold junction compensation accuracy ±0.5 C Not required nalog accuracy ±0.05%F.S. * Overall accuracy = igital accuracy + analog accuracy (For digital output type, only digital accuracy applies.) Minimum span 25 C 10 C 2mV 1271 R: Road resistance (Ω) 521 250 V-11.5 R= 0.024 Operating range 11.5 17.5 24 42 V: Supply voltage (V) Figure 1 Supply voltage and load resistance characteristics - 3 -

No. SS2-TT110-0100 Example of accuracy calculation When the models TT61/71 are used with an model TT90 sensor (sold separately) the maximum error and accuracy can be calculated as follows: oth models TT61/71 and model TT90 have their own integral errors. When they are integrated, the total sum of the errors must be taken into consideration. Each temperature sensor's error differs one from the other, depending on type of sensor, measuring location and class. Refer to our model TT90 Sensor s Specifications. If another manufacturer s sensor is used with the model TT61, please inquire of the accuracy with that sensor s manufacturer. Calculation of maximum error when SN (N) thermocouple range is set at 0~1000 C analog output [Converter error] Converter accuracy = digital accuracy + analog accuracy + internal cold junction compensator accuracy = 0.5 + 1000 0.05% F.S. + 0.5 = 0.5 + 0.5 + 0.5 = 1.5 ( C) Hence the maximum converter error is within +1.5 C. (ccuracy is 1.5/1000 100 = 0.15% F.S.) [Temperature sensor error] Type of thermocouple: SN(N) Class:2 Operating temperature range: 1000 C [Overall error] Overall error = Converter error + temperature sensor error = 1.5 + 7.5 = 9.0 ( C) The overall maximum error for a combination of a converter and sensor is approximately + 9 C. (Overall accuracy is + 9.0/1000 100 = 0.9% F.S.) Calculation of maximum analog output error when Pt100 resistance temperature sensor range is set at - 20~80 C analog output [Converter error] Converter accuracy = digital accuracy + analog accuracy + internal cold junction compensator accuracy = 0.2 + 100 0.05% F.S. + 0 = 0.2 + 0.05 + 0 = 0.25 ( C) Hence, the maximum converter error is ± 0.25 C (Overall accuracy is 0.25/100 100 = 0.25% F.S.) From model TT90 Specifications + 0.0075 t + 0.0075 1000 = +7.5 ( C) [Temperature sensor error] Resistance temperature detector: Pt100 Class: Temperature range 100 C Yamatake Corporation [Overall accuracy] Overall accuracy = Converter error + temperature sensor error = 0.25 + 0.35 = 0.6 ( C) Hence, the maximum converter and sensor combination error is under ± 0.6 C. (Overall accuracy is 0.6/100 100 = 0.6% F.S.) Calculation of maximum digital output error when Pt100 resistance temperature sensor range is set at 0~200 C analog output [Converter error] Converter accuracy = digital accuracy + analog accuracy + internal cold junction compensator accuracy ( No analog error because output is digital) = 0.2 + 0 + 0 = 0.2 The maximum error is under ± 2 C (ccuracy is 0.2/100 100 = 0.2% F.S.) [Temperature sensor error] Resistance temperature detector: Pt100 Class: Temperature range 200 C [Overall error] Overall error = converter error + temperature sensor error = 0.2 + 1.3 = 1.5 ( C) The maximum overall error for a combination of a converter and temperature sensor is under +1.5 C. (Overall accuracy is 1.5/200 100 = 0.75% F.S.) From model TT90 Specifications, ± 0.15 + 0.002 t ; ± 0.15 + 0.002 100 = ± 0.35 ( C) From model TT90 Specifications, + 0.3 + 0.005 t 0.3 + 0.005 200 = 1.3 ( C) - 4 -

Yamatake Corporation Explanation of converter model number symbols Standard model number Model TT61 Waterproof temperature transmitter (integral type) Model TT71 Flameproof models (integral type) Selectable specifications Output type 4~20 m C analog E protocol Construction Case: luminum alloy 1 Waterproof: JISC0920 immersion-proof equivalent NEM3 and 4X, IEC IP67 equivalent 2 JIS Flameproof: Ex d IIC T6 (mbient temperature -20~60 C) Electric conduit 1 G1/2 thread 2 NPT1/2 thread 3 M20*1.5 thread (for adaptor) 4 Pg13.5 thread (for adaptor) Note) Only G1/2 thread of 1 is available for JIS Flameproof construction. dditional specification selection Integral indicator X M No indicator igital indicator ( C) igital indication: 4.5 digits -1999.9 ~ 1999.9 C Plus 4~20mC level bar graph indication Self-diagnostic status Painting S Standard finish aked acrylic resin Corrosion-proof paint aked acrylic resin, fungus proof finish Corrosion-proof paint aked epoxy resin, fungus proof finish irection of burn-out U Upscale on failure ownscale on failure dditional specifications C Traceability certificate This certificate consists of three copies of the measurement management system's block diagram, calibration certificate, and test report. 2-inch pipe mounting bracket Can be mounted either horizontally or vertically Internal data setting Type of sensor, range setting, Tag. No. (Engraved on name plate and written into EEPROM in eight alpha numerical characters) before shipment. Please specify setting when ordering. No. SS2-TT110-0100 E One electric conduit elbow One elbow is provided to direct the explosion-proof cable gland and cable downward. This must be selected for JIS Flameproof specifications. Material: brass, Wire diameter: Ø8~13 H Tropicalization treatment Fungus proof spray is sprayed on terminal bracket on up to improve tropical climate resistance. T Test report treatment This test report covers testing such as for external appearance, insulation resistance, and voltage withstanding, input/output characteristics. Since input/ output test of thermocouple input model and mv input model are identical, the test is conducted with mv input. Note that this test is for the temperature transmitter itself and not for the integrated temperature transmitter-sensor test. Other common specifications Lightning guard specifications Voltage surge peak amplitude: 100kV Current surge peak amplitude: 1000 nti-vibration specifications 5~9Hz 0~7.5 mmp-p 9~2000Hz 0~29.4 m/s 2 Note) Product specifications except sensor and bracket Finish Metallic green (Munsell 5G7/8) Cautions regarding product selection JIS Flameproof construction consists of temperature transmitter itself and temperature sensor. When specifying a JIS Flameproof model, please order a Yamatake Temperature sensor model TT90. lso, please use a protecting well to protect temperature sheath. ccording to the JIS Flameproof Technical Standard (February 1997) the ambient temperature should be - 20~60 C, and applicable range for explosion-proof is - 20~85 C. lso, explosion-proof standards can be ensured of by using a explosion-proof packing type cable gland. If cable glands of other makes are incorporated, Yamatake cannot guarantee that the unit will comply to explosion-proof standards. Even though ambient temperature is within the specified range, heat may accumulate in converter. If this occurs, provide a shield to keep the converter temperature within the specified range. - 5 -

No. SS2-TT110-0100 Yamatake Corporation MOEL SELECTION Model TT61 Waterproof model asic Model No. Selection1 Selection2 Option TT61 - - - Output nalog (4-20m) igital (E -protocol) Housing Waterproof and dust -proof structure 1 Electrical connection G1 / 2 1 1/2NPT 2 M20 1.5 (with adaptor) 3 Pg13.5 (with adaptor) 4 Integral digital meter Without LC meter for eg. C X M Finish Standard finish S Corrosion -resistant finish Corrosion -proof finish urnout feature Upscale ownscale U Options *1 No selection X Certificate of traceability Mounting bracket Customer configurations (sensor type and measuring range) *2 C Tropicalization treatment H Calibration test report T Note) *1. Up to five options may be selected simultaneously. *2. If it is required to configure sensor type, range and tag No. at the ex factory, please select C and specified specifications separately. If there is no specified specification, the unit will be configured with the default value based on Table 1. - 6 -

Yamatake Corporation No. SS2-TT110-0100 Model TT71 Explosion-proof model asic Model No. Selection1 Selection2 Option TT71 - - - Output nalog (4-20m) igital (E -protocol) Housing JIS Flameproof *5 2 Electrical connection G1 / 2 only for JIS Flameproof 1 Integral digital meter Without LC meter for deg. C X M Finish Standard finish S Corrosion -resistant finish Corrosion -proof finish urnout feature Upscale U ownscale Options *1 No selection X Certificate of conformance and origin Mounting bracket Customer configurations (sensor type and measuring range) *2 C With an elbow for electrical conduit *3 E Tropical finish H Calibration test report T Note) *1. Up to five options may be selected simultaneously. *2. If it is required to configure the sensor type, range and tag No. at the ex factory, please select C and the specified specifications separately. If there is no specified specification, the unit will be configured with the default value based on Table 1 in the Specification sheets. *3. For options, E: with an elbow for electrical conduit must be selected. *5. Use model TT90 (sold separately) for the temperature sensor. JIS Flameproof construction consists of temperature transmitter itself and temperature sensor. - 7 -

No. SS2-TT110-0100 Yamatake Corporation IMENSIONS External dimensions No indicator igital indicator Output Electric conduit Refer to table 1 M4 external Ground terminal Sensor Electric conduit 2-M5 Installation bolt hole 2 in. pipe Mounting bracket (Option) 2 in. pipe Caution: The above drawing are of the external appearance of TT61/71. For external appearance with sensor, refer to TT90 specification, SS2-TT900-0100. Output wiring Internal Ground wiring Sensor input wiring Output wiring 24V C PS urn-out setting w/ shortening bar: Up-scale w/o shortening bar: own-scale SFC Note) The size of the terminal screws are M4. Receiver Totate international uilding 2-12-19 Shibuya Shibuya-ku, Tokyo 150-8316 Japan Tel : Fax : 81-3-3486-2310 81-3-3486-2593 Savemation Saving through utomation - 8 - This has been printed on recycled paper. http://www.yamatake.co.jp/