Filling in the MIMO Matrix Part 2 Time Waveform Replication Tests Using Field Data

Similar documents
Recent System Developments for Multi-Actuator Vibration Control

Response spectrum Time history Power Spectral Density, PSD

IMAC 27 - Orlando, FL Shaker Excitation

ME scope Application Note 01 The FFT, Leakage, and Windowing

MULTIPLE INPUT MULTIPLE OUTPUT (MIMO) VIBRATION CONTROL SYSTEM

How to implement SRS test without data measured?

Simulate and Stimulate

ME scope Application Note 02 Waveform Integration & Differentiation

MIL-STD-202G SHOCK (SPECIFIED PULSE)

Development of Random Vibration Profiles for Test Deployers to Simulate the Dynamic Environment in the Poly-Picosatellite Orbital Deployer

A practical guide to using MIMO vibration control for MIL-STD-810 single axis transport testing. of large, resonant land based military payloads

Testing Sensors & Actors Using Digital Oscilloscopes

ME scopeves Application Note #21 Calculating Responses of MIMO Systems to Multiple Forces

Figure 1: The Penobscot Narrows Bridge in Maine, U.S.A. Figure 2: Arrangement of stay cables tested

How to perform transfer path analysis

Separation of Sine and Random Com ponents from Vibration Measurements

Structural Dynamics Measurements Mark H. Richardson Vibrant Technology, Inc. Jamestown, CA 95327

An Introduction to Time Waveform Analysis

Texas Components - Data Sheet. The TX53G1 is an extremely rugged, low distortion, wide dynamic range sensor. suspending Fluid.

Experimental Modal Analysis of an Automobile Tire

Transfer Function (TRF)

Vibration Fundamentals Training System

Advanced High-Frequency 6-DOF Vibration Testing Using the Tensor System

A Comparison of MIMO-FRF Excitation/Averaging Techniques on Heavily and Lightly Damped Structures

EIA STANDARD TP-27B. Mechanical Shock (Specified Pulse) Test Procedure for Electrical Connectors EIA B ELECTRONIC INDUSTRIES ASSOCIATION

Quartz Lock Loop (QLL) For Robust GNSS Operation in High Vibration Environments

3.0 Apparatus. 3.1 Excitation System

Modal Parameter Estimation Using Acoustic Modal Analysis

An Alternative to Pyrotechnic Testing For Shock Identification

A METHOD FOR OPTIMAL RECONSTRUCTION OF VELOCITY RESPONSE USING EXPERIMENTAL DISPLACEMENT AND ACCELERATION SIGNALS

New Features of IEEE Std Digitizing Waveform Recorders

Hybrid Shaker Technology for Wide-band Vibration Test Systems

Resonance Tube Lab 9

DS-2000 Series Measurement of Frequency Response Function

Mechanically Isolated & Electrically Filtered ICP pyroshock Accelerometers. Bob Metz October 2015

The ENDEVCO high-g shock triaxial accelerometer: A smaller, more cost-effective solution to making triaxial measurements. Endevco technical paper 334

PC1141 Physics I. Speed of Sound. Traveling waves of speed v, frequency f and wavelength λ are described by

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

(i) Sine sweep (ii) Sine beat (iii) Time history (iv) Continuous sine

Testing in a Combined Vibration and Acceleration Environment

m+p VibControl Shock Control

EWGAE 2010 Vienna, 8th to 10th September

Experimental Evaluation of Techniques Designed to Reduce Vibration Simulation Test Time

AGN 008 Vibration DESCRIPTION. Cummins Generator Technologies manufacture ac generators (alternators) to ensure compliance with BS 5000, Part 3.

DEPARTMENT OF DEFENSE TEST METHOD STANDARD METHOD 213, SHOCK (SPECIFIED PULSE)

AN5E Application Note

FFT Use in NI DIAdem

VOLD-KALMAN ORDER TRACKING FILTERING IN ROTATING MACHINERY

6.555 Lab1: The Electrocardiogram

Excitation Techniques Do s and Don ts

m+p VibControl Sine Vibration Control

9th National Congress on Theoretical and Applied Mechanics, Brussels, 9-10 May 2012

Resonance Tube. 1 Purpose. 2 Theory. 2.1 Air As A Spring. 2.2 Traveling Sound Waves in Air

Module 4 TEST SYSTEM Part 2. SHAKING TABLE CONTROLLER ASSOCIATED SOFTWARES Dr. J.C. QUEVAL, CEA/Saclay

Proven Sensor Performance for Emerging Shock Environments

A Custom Vibration Test Fixture Using a Subwoofer

Vibration Tests: a Brief Historical Background

SHOCK AND VIBRATION RESPONSE SPECTRA COURSE Unit 4. Random Vibration Characteristics. By Tom Irvine

Earthquake Resistance Test Specifications for Communications Equipment

Real-Time FFT Analyser - Functional Specification

Vibration Transducer Calibration System

System Inputs, Physical Modeling, and Time & Frequency Domains

An Overview of MIMO-FRF Excitation/Averaging Techniques

VIBRATION ANALYZER. Vibration Analyzer VA-12

Active Vibration Isolation of an Unbalanced Machine Tool Spindle

Laboratory 14. Lab 14. Vibration Measurement With an Accelerometer

IADS Frequency Analysis FAQ ( Updated: March 2009 )

The Shocking Truth About the Frequency Domain

CHAPTER 5 FAULT DIAGNOSIS OF ROTATING SHAFT WITH SHAFT MISALIGNMENT

CHAPTER 4 IMPLEMENTATION OF ADALINE IN MATLAB

26 Endevco Isotron accelerometers

Summary. Theory. Introduction

CHAPTER 6 INTRODUCTION TO SYSTEM IDENTIFICATION

Laboratory Experiment #2 Frequency Response Measurements

Good Modal Practices

Sampling and Reconstruction

DETERMINATION OF CUTTING FORCES USING A FLEXURE-BASED DYNAMOMETER: DECONVOLUTION OF STRUCTURAL DYNAMICS USING THE FREQUENCY RESPONSE FUNCTION

Laboratory Experiment #1 Introduction to Spectral Analysis

Detection and characterization of oscillatory transient using Spectral Kurtosis

SHAKER TABLE SEISMIC TESTING OF EQUIPMENT USING HISTORICAL STRONG MOTION DATA SCALED TO SATISFY A SHOCK RESPONSE SPECTRUM Revision C

The rapid evolution of

Correlating HALT & HASS, RS/HALT Vibration and End-Use Environments

ACE. SignalCalc. Ultra-Portable, Cost-Effective Dynamic Signal Analyzer. 4 input channels, 2 sources. High Speed Tachometer. 120 db dynamic range

A Friendly Approach to Increasing the Frequency Response of Piezoelectric Generators

2015 HBM ncode Products User Group Meeting

Summary. Seismic vibrators are the preferred sources for land seismic ( ) (1) Unfortunately, due to the mechanical and

Introduction*to*Machinery*Vibration*Sheet*Answer* Chapter*1:*Vibrations*Sources*and*Uses*

Proposal. Analysis of Parallel Vibration Paths with Potential Application to Vehicle Noise. Reduction. Submitted to. The Engineering Honors Committee

SVENSK STANDARD SS :2014

Fourier Signal Analysis

Modal Parameter Identification of A Continuous Beam Bridge by Using Grouped Response Measurements

SETUP I: CORD. Continuous Systems

The Scientist and Engineer's Guide to Digital Signal Processing By Steven W. Smith, Ph.D.

Channel Emulation Solution

Frequency Response Function Measurements of Disc and Drum Brake With its Verification by CAE

BLADE AND SHAFT CRACK DETECTION USING TORSIONAL VIBRATION MEASUREMENTS PART 1: FEASIBILITY STUDIES

A HARMONIC PEAK REDUCTION TECHNIQUE FOR OPERATIONAL MODAL ANALYSIS OF ROTATING MACHINERY

STEREO IMPACT Solar Energetic Particles Package (SEP) Dynamic Test Plan

A study of Vibration Analysis for Gearbox Casing Using Finite Element Analysis

Fault diagnosis of massey ferguson gearbox using power spectral density

Transcription:

Filling in the MIMO Matrix Part 2 Time Waveform Replication Tests Using Field Data Marcos Underwood, Russ Ayres, and Tony Keller, Spectral Dynamics, Inc., San Jose, California There is currently quite a bit of interest in using field data as much as possible for performing multi-shaker laboratory simulations. With the introduction of MIL-STD-810G in 2008, and especially Method 525, Time Waveform Replication, multi-shaker replication tests are being attempted and performed more than ever. As with any multi-exciter test, it is important to characterize the entire test configuration, including shakers, amplifiers, fixtures, transducers and cables before starting the test. To minimize test errors, especially if long time records are being used, system characterization should be updated during the test, perhaps every control loop. Also, even though the purpose of the test is carefully reproducing time waveforms from field measurements, the control process may take place in the time domain, the frequency domain or both. We show here the development of a unique approach to creating and controlling time records with a seamless transition from block to block. Although multi-shaker testing has been attempted for over 40 years, it is only in the past 8 to 10 years that the correct combination of instrumentation, computers and software have become available to make this technique a reality. In the absence of a set of comprehensive multi-exciter test specifications, many organizations have tried to take the advice of MIL-STD-810 and use measured or borrowed field vibration data to create a laboratory simulation. Some recent reports 1,2 have indicated that when using actual measured field vibration for multi-exciter simulation, care must be taken to assure good phase and coherence duplication, or unexpected results may occur. So it is up to the test engineers to assure that they have taken into account or at least considered: Quality of the multi-channel test data to be used for replication Boundary condition differences, field to lab Number and location of lab shakers Test lab fixturing arrangement and possible effect on system dynamics Control approach to be implemented square, rectangular, over-determined, etc. Position and number of control transducers Approach to error analysis for proof of performance Time Waveform Replication As discussed in Reference 3, time waveform replication (TWR) is a vibration control method whose goal is to cause the response time history at one or more control points to agree with a prespecified vector of waveforms while using one or more exciters. TWR control is typically performed at least partially in the frequency domain. But since the main goal of TWR is to recreate a series of time waveforms as closely as possible, its performance is usually evaluated in the time domain. Figure 1 shows an overlay of the Reference and actual Control waveforms present at Control point 3 in the 4 shaker test described here. In this case, the Control was so good that it is very hard to distinguish between the Control (white) and Reference (blue) traces. This figure shows the first 100 milliseconds of the 4-second records that were used. Note that the peak values shown refer to the entire 4-second record. Figure 2 shows an error display in percent of peak value for the same Control location 3, but this time for the entire 4-second record. Note that one point approaches 0.4%, and a second exceeds 0.3%. But for most of the record the peak error never exceeds 0.2%. When performing TWR control, the Reference and Control response time histories are typically longer than an FFT block. So an indirect convolution method using overlap and add is typically used for Control and Drive signal generation. Figure 1. Overlay of Reference and Control waveforms (expanded). Figure 2. Control error as percentage of peak value shown over full 4-second waveform. Figure 3. Test setup for acquiring and controlling field data. TWR Control Approach While in the frequency domain, the product of two spectra or of a spectrum and a frequency response function, can be taken by direct multiplication. Performing an equivalent operation in the time domain, requires the use of convolution. 4,5 The Control approach described here uses indirect linear convolution between the measured inverse impulse response of the system under test and the Reference and Error time histories. The use of overdeter- 8 SOUND & VIBRATION/SEPTEMBER 2011 www.sandv.com

Figure 4. Control PSDs at four plate corners during field data acquisition phase: (a) 1,1; (b) 2,2; (c) 3,3; (d) 4,4. Figure 5. Control coherence and phase for MIMO random test shown in Figure 4: (a&c) 1,2; (b&d) 2,3 www.sandv.com DYNAMIC TESTING REFERENCE ISSUE 9

Figure 6. Response PSDs at locations: (a) 5, (b) 6, (c) 7 and (d) 8 for Test A. test system includes bolted joints or oil flow, which can change dynamic properties with time, only adaptive control can assure a successful test completion. Figure 7. Four-second Reference and Control waveforms. mined TWR Control is possible with the use of I/O transformations, although this technique was not used in the tests described here. Generally, TWR Control performance is determined by the quality of the system under test and the accuracy of the measured inverse frequency response (impedance) matrix. Adaptive Control Adaptive Control is a patented method [Patent # 5,299,459], that updates the impedance matrix (system model) of virtually every control loop using actual Response and Drive signals. This can minimize the time it takes to match the Reference and Response time histories or spectra. This approach can also be used to update the actual Drive signals in real-time while the test is running to reduce Control errors during a TWR test. Without this capability, Drives are only updated between test sequences or at the start of the test in TWR applications. This approach can simulate and control nonlinear and time-varying environments to a much higher degree of accuracy than using traditional techniques. For example, if the Considerations for using Field Data Before attempting to run any multi-exciter TWR test, it is always good practice to see how good a match there is between the available field data, the goals and objectives of the TWR test and the test facilities at your disposal. At a minimum, the dynamic capabilities of the multi-exciter test system, in terms of load capacity, force rating, bandwidth, acceleration, velocity and displacement capability should match or exceed the test objectives. In addition, the test article s dynamic response characteristics must be compatible with the field data. Is the measured data complete and a good match for the sampling rate and record length parameters of the test control system? Or will resampling be required? Were the field transducers mounted at locations and orientations that will exist and be available in the Lab? Or will the apparatus and fixturing required to mount the test article to the exciters actually eliminate some field measurement locations? Test article dynamic response characteristics must be compatible with the measured field data, and boundary conditions encountered in the field should be satisfied as closely as possible during the lab test. Unless care is taken to understand the dynamic response of the test article when mounted in the test platform configuration, the resulting TWR could be a severe overtest or even an undertest. Using Field Data in a Multi-Exciter TWR Test Test A MIMO Random Control with Zero Coherence Between Control Points. We present several examples of recording field data and using these data to create time waveform records and then to replicate the data on a four-shaker system. In the first example (Test A), the boundary conditions (BC) between the field setup and the test lab simulation are exactly matched. In another case (Test C), they are not matched, but the lab test is still required to 10 SOUND & VIBRATION/SEPTEMBER 2011 www.sandv.com

Figure 8. TWR waveforms with Reference and Control overlaid from Test B: (a) 1; (b) 2; (c) 3; (d) 4. Figure 9. Measured TWR Control error (%) for Test B at locations: (a) 1; (b) 2; (c) 3; (d) 4. reproduce the measured field data. Field data for each case were obtained by exciting an aluminum plate with uncorrelated noise using four shakers and MIMO random control. Response time histories for the selected TWR Control locations were then recorded and used for the subsequent time waveform replication tests. www.sandv.com DYNAMIC TESTING REFERENCE ISSUE 11

Figure 10. Control PSDs derived from TWR time waveforms of Test B: (a) 1; (b) 2; (c) 3; (d) 4. Figure 11. Response PSDs for Test B at locations: (a) 1; (b) 2; (c) 3; (d) 4. 12 SOUND & VIBRATION/SEPTEMBER 2011 www.sandv.com

Figure 3 shows the test setup for controlling the four corners of a lightly damped plate and measuring four field data Responses at four arbitrary locations inboard from the Control locations. Note that in addition to the eight mounted accelerometers, a loose book was placed on the plate to simulate system time variability during the test. Since the test data used for TWR often exceeded 6 g, quite a bit of book motion was experienced during the TWR test. The first phase of the test controlling the four corners was done at a nominal Control level of 1 g rms. This means that, besides having instantaneous acceleration values at the response points that exceeded 5 g, the corners would sometimes also exceed 1 g. This would cause book motion also during the acquisition phase Figure 12. Nuts loosened toward end of Test B. of getting the field data. The Control PSDs experienced during the first-step MIMO random Control are seen in Figure 4. As can be seen, Control was quite good even in the presence of several resonances; this will be shown later. For Test A, the Control coherence was set to zero. This produces completely random phase relationships between Control locations. The coherence for all six upper off-diagonal matrix terms was controlled to zero, resulting in random phase between all Control points for this incoherent situation. While controlling to these parameters, the response PSDs for the auxiliary points, 5 through 8, were also measured. These are seen in Figure 6. While controlling the four corners of the plate with MIMO random excitation, time histories of the four Response locations, 5 thru 8, were recorded to disk. Since stationary data were being recorded, we decided that 4-second time blocks would be sufficient for performing subsequent TWR tests. This also corresponds to one block of the 200 Hz MIMO random test performed with 800 Control lines. A typical 4-second TWR signal from this test showing both Reference and Control is seen in Figure 7. In Figure 7, the instantaneous accelerations at the TWR Control points can easily exceed 6 or 7 g. Therefore, the nonlinearizing loose book on the plate becomes a critical factor in terms of the need for adaptive Control. Summary, Test A.: MIMO Random Control at locations 1, 2, 3 and 4. Response measurements at locations 5, 6, 7 and 8. Time histories of all channels recorded. Test B TWR Control Using Time Records from Test A. In Figure 8, the four channels of time waveforms collected during Test A and used as the basis of Test B are shown. Different record lengths are shown for each controlled channel to illustrate the excellent overlay between the Reference (desired) and actual Control waveforms. At this point in the testing, the system boundary conditions were considered to be the same for both the MIMO random test when waveforms were recorded and the subsequent TWR test. Figure 9 shows the accuracy achieved in the first TWR test and that most errors are less than 0.2% after a few iterations. The error definition used here is instantaneous Control error normalized to peak acceleration. This is quite a stringent definition, compared Figure 13. Control PSDs for Test C with loosened nuts: (a) 1,1 (b) 2,2; (c) 3,3; (d) 4,4. www.sandv.com DYNAMIC TESTING REFERENCE ISSUE 13

Figure 14. Resulting Control coherence and phase during Test C: (a) 1,2 coherence; (b) 2,3 coherence; (c) 1,2 phase; (d) 2,3 phase Figure15. Measured Responses taken during Test C: (a) 5; (b) 6; (c) 7; (d) 8. 14 SOUND & VIBRATION/SEPTEMBER 2011 www.sandv.com

Figure 16. Superimposed Control and Reference waveforms from Test D: (a) 1; (b) 2; (c) 3; (d) 4. Figure 17. TWR error displays for Test D: (a) 1; (b) 2; (c) 3; (d) 4. with other techniques often used. Method 525 of MIL-STD-810G does not require a specific formulation to determine TWR error but does offer 14 different approaches to determine how well the TWR test was performed. Some of these different approaches may be described by Ron Merritt in the future. For the four Control points in Test B, locations 5, 6, 7 and 8, www.sandv.com DYNAMIC TESTING REFERENCE ISSUE 15

Figure 18. PSDs of Control TWR waveforms for Test D: (a) 1; (b) 2; (c) 3; (d) 4. Figure 19. PSDs during TWR of Test D for auxiliary locations: (a) 1; (b) 2; (c) 3; (d) 4. 16 SOUND & VIBRATION/SEPTEMBER 2011 www.sandv.com

power spectral densities were created and are shown in Figure 10. During Test B, locations 1, 2, 3 and 4 (which were the Control points for Test A but are Response locations for Test B) were also analyzed and are shown in Figure 11. Due to reciprocity, the excellent Control results in replicating the Responses at locations 5, 6, 7 and 8 and the nearly linear behavior of the test system, these agree quite well with those shown in Figure 4. Near the end of Test B, it was noted that some unexplained spikes were appearing in the data, especially in the frequency domain representations. An investigation was done, and it was discovered that at the shaker 3 attachment to the plate, the set of attachment nuts had vibrated loose. In effect, this was providing a fifth excitation to the plate but in the form of nonrepetitive shocks. This had the result of exciting the plate resonances to an even higher degree than the random excitation, creating a non-zero coherence with respect to location 3 and introducing a noncontrollable spectrum source. This definitely introduces a sense of realism into an otherwise very clean test. The connection involved here is highlighted in Figure 12. The realization that the single loosened nut added a completely new dynamic into the TWR test led to the creation of additional tests that are described next. Test C MIMO RandomTest Performed with Shaker 3 Attachment Nuts Loosened. The next test performed was done to see the effects of running another controlled MIMO random test, this time with the loose connection at Shaker 3. Figure 13 shows the Control situation for Test C. Note that the Control levels are all very close to 1 g rms. But at the Control location nearest shaker 3, additional spikes, not coming directly from the (linear) shakers are contributing to the measurements. Since these are not linearly related to the complex set of Drive vectors being created, they fall outside the system definition in the spectral density matrix. So the resulting Control coherence is also affected at these spike frequencies (see Figure 14). The Response PSDs of locations 5 thru 8 were also measured during Test C and are shown in Figure 15. In addition to the impacts imparted at shaker 3 during Test C, due to the loosened connection, the lack of a secure tiedown at that location also removed some natural damping imparted by the normally tight connection. This results in a slightly higher Q in the plate responses, as shown in Figure 15. During Test C with the untightened condition at shaker 3, a new set of time waveforms was recorded. This led to Test D. Test D: TWR Test Performed on Retightened Structure Using Waveforms Recorded During Test C. The final test was performed using a set of boundary conditions, (connections tightened at all 4 shakers), which was different from the boundary conditions that existed when the data were taken during Test C. With these different sets of boundary conditions between Test C and Test D, the TWR Control, although quite good, was not as good as that seen during Tests A and B. Examples of the four waveforms replicated during the final test are shown in Figure 16. The poorer boundary condition match in Test D resulted in an increased error during TWR (see Figure 17). But the PSDs measured during the TWR done in Test D (Figure 18), look much more like those of Figure 10 taken during Test B (tight MIMO acquisition, tight TWR control), than those of Figure 15 taken during Test C (loose MIMO acquisition, loose MIMO control). A slight increase in Q is noted, but there are not as many obvious spikes in all four channels as previously. Finally, Figure 19 shows the reciprocity locations 1 thru 4 during the TWR control of Test D; it bears a closer similarity to Figure 11 (no spikes) than Figure 13 (many spikes). Although there are spikes in Figure 19 at the same frequencies as in Figure 13, they are at reduced amplitudes. Conclusions TWR control performance while employing the field Response data discussed here was very good; waveform errors were less than 1% in virtually all cases. Comparing Reference and Control waveforms during TWR testing is an extremely sensitive procedure, much more so than, for example, comparing PSD levels. So extra precautions should be taken to assure that factors such as loose connections or slapping are not present during the test. The use of adaptive Control during TWR testing is very effective in addressing time-varying and nonlinear characteristics in the test setup. Satisfying field boundary conditions in the lab as much as possible may have the largest single effect on assuring a quality laboratory simulation. More work needs to be done on selecting a range of acceptable error metrics when performing TWR tests. However, the point-topoint error comparison used in this article is the most severe. When performing error analysis after a TWR test, it is advantageous to observe results in both the time and the frequency domains. Although our goal is to match the time histories as closely as possible, some results, especially on highly resonant structures, may offer more meaningful comparisons as PSDs. This is evident in Figures 18 and 19. Measuring and comparing noncontrol points in the field and in the lab during TWR tests will help assure a quality TWR simulation. References 1. Lamparelli, Marc, Underwood, Marcos, Ayres, Russell, and Keller, Tony, An Application of ED Shakers to High Kurtosis Replication, ESTECH2009, 55th Annual meeting of the IEST, Schaumburg, IL, May 4-7, 2009. 2. Underwood, Marcos, Ayres, Russell, and Keller, Tony, Filling in the MIMO Matrix, Using Measured Data to run a Multi-Actuator/Multi-Axis Vibration Test, presented at the 80th Shock & Vibration Symposium, San Diego, CA, October 26-29, 2009. 3. MIL-STD-810G Test Method Standard for Environmental Engineering Considerations and Laboratory Tests, http://www.dtc.army.mil/publications/mil-std-810g.pdf, October 31, 2008, Method 525, Time Waveform Replication. 4. Chang, David K., Analysis of Linear Systems, Addison-Wesley, 1959. 5. Schwartz, Mischa, Information Transmission, Modulation and Noise, McGraw-Hill Electrical and Electronic Engineering Series, 1959. The author may be reached at: kellert@sd-star.com. www.sandv.com DYNAMIC TESTING REFERENCE ISSUE 17