DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY. H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli

Similar documents
MICRO XRF OF LIGHT ELEMENTS USING A POLYCAPILLARY LENS AND AN ULTRA THIN WINDOW SILICON DRIFT DETECTOR INSIDE A VACUUM CHAMBER

ANALYTICAL MICRO X-RAY FLUORESCENCE SPECTROMETER

Applications of New, High Intensity X-Ray Optics - Normal and thin film diffraction using a parabolic, multilayer mirror

WIDE ANGLE GEOMETRY EDXRF SPECTROMETERS WITH SECONDARY TARGET AND DIRECT EXCITATION MODES

Spectral distribution from end window X-ray tubes

MICROANALYSIS WITH A POLYCAPILLARY IN A VACUUM CHAMBER

MOXTEK S NEW ULTRA-LITE X-RAY SOURCES: PERFORMACE CHARACTERIZATIONS

X-RAY OPTICS FOR TWO-DIMENSIONAL DIFFRACTION

TOWARDS FAST RECIPROCAL SPACE MAPPING

DEVELOPMENT OF A WAVELENGTH DISPERSIVE X-RAY FLUORESCENCE SPECTROMETER USING A MULTI-CAPILLARY X-RAY LENS FOR X-RAY DETECTION

CONFOCAL GRADED d-spacing MULTILAYER BEAM CONDITIONING OPTICS

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

ON THE ENERGY DEPENDENCE OF THE DETECTOR EFFICIENCY OF A Si-PIN DIODE

GUNSHOT RESIDUE INVESTIGATIONS USING TXRF

ON THE DETECTION LIMIT OF TEY (TOTAL ELECTRON YIELD) Maria F. Ebel, Horst Ebel and Robert Svagera

X-RAY BACKSCATTER IMAGING: PHOTOGRAPHY THROUGH BARRIERS

MICROFOCUSING X-RAY EQUIPMENT FOR THE LAB DIFFRACTOMETER

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

LONG TERM STATISTICS OF X-RAY SPECTROMETERS

USING A CHARGE-COUPLED DEVICE (CCD) TO GATHER X-RAY FLUORESCENCE (XRF)AND X-RAY DIFFRACTION (XRD) INFORMATION SIMULTANEOUSLY

DESIGN AND MEASUREMENT WITH A NEW PORTABLE X-RAY CAMERA FOR FULL-FIELD FLUORESCENCE IMAGING

Copyright -International Centre for Diffraction Data 2010 ISSN

1. Introduction X-ray absorption fine structure (XAFS) is an element-specific powerful technique for chemical analysis. In general, XAFS spectra are o

SIMULTANEOUS XRD/XRF WITH LOW-POWER X-RAY TUBES

Observation of X-rays generated by relativistic electrons in waveguide target mounted inside a betatron

TOWARDS SUB-100 NM X-RAY MICROSCOPY FOR TOMOGRAPHIC APPLICATIONS

TIME-PRESERVING MONOCHROMATORS FOR ULTRASHORT EXTREME-ULTRAVIOLET PULSES

Measuring optical filters

Sources & Beam Line Optics

RIETVELD REFINEMENT OF POWDER DATA FROM MULTILAYER OPTICS

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

Charge Loss Between Contacts Of CdZnTe Pixel Detectors

Nano Beam Position Monitor

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Supplementary Information

Development of X-ray Tool For Critical- Dimension Metrology

New Detectors for X-Ray Metal Thickness Measuring

Niklas Norrby 12/17/2010

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

X-rays. X-rays are produced when electrons are accelerated and collide with a target. X-rays are sometimes characterized by the generating voltage

In-focus monochromator: theory and experiment of a new grazing incidence mounting

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

membrane sample EUV characterization

X-RAY transmission tomography is very useful for the estimation

Performance of the SASE3 monochromator equipped with a provisional short grating. Variable line spacing grating specifications

Spectroscopy in the UV and Visible: Instrumentation. Spectroscopy in the UV and Visible: Instrumentation

A high energy gamma camera using a multiple hole collimator

pcvd diamond beam position monitors for PETRA III

Abstract. 1. Introduction

Design of a light-guide used for the real-time monitoring of LCD-displays

Low Cost Rolled X-ray Prism Lenses to Increase Photon Flux Density in Diffractometry Experiments

BL39XU Magnetic Materials

Supplementary Figure 1

Title detector with operating temperature.

Large-Area CdTe Photon-Counting Pixel Detectors

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

Lesson 2 Diffractometers

Undulator K-Parameter Measurements at LCLS

High granularity scintillating fiber trackers based on Silicon Photomultiplier

Diamond X-ray Rocking Curve and Topograph Measurements at CHESS

POLARIZATION FOR BACKGROUND REDUCTION IN EDXRF --- THE TECHNIQUE THAT DOES INDEED WORK

Performance Comparison of Spectrometers Featuring On-Axis and Off-Axis Grating Rotation

Application Note (A11)

First test experiments with FMB- Oxford direct drive DCM at the Sirius beamline of Synchrotron SOLEIL

AutoMATE II. Micro-area X-ray stress measurement system. Highly accurate micro area residual stress

[4] (b) Fig. 6.1 shows a loudspeaker fixed near the end of a tube of length 0.6 m. tube m 0.4 m 0.6 m. Fig. 6.

880 Quantum Electronics Optional Lab Construct A Pulsed Dye Laser

Applications of Micro XRF for the Analysis of Traditional Japanese "Ainu" Glass Beads and other Artifacts

R-AXIS RAPID. X-ray Single Crystal Structure Analysis System. Product Information

Period 3 Solutions: Electromagnetic Waves Radiant Energy II

Experience of synchrotron sources and optics modelling at Diamond Light Source

Microspot x-ray focusing using a short focal-length compound refractive lenses

The Photoelectric Effect

Advancing EDS Analysis in the SEM Quantitative XRF. International Microscopy Congress, September 5 th, Outline

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

MC SIMULATION OF SCATTER INTENSITIES IN A CONE-BEAM CT SYSTEM EMPLOYING A 450 kv X-RAY TUBE

Full characterization tests of Micromegas with elongated pillars

A PORTABLE X-RAY APPARATUS FOR BOTH STRESS MEASUREMENT AND PHASE ANALYSIS UNDER FIELD CONDITIONS.

Data Collection with. VÅNTEC-2000 Detector

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Sub-50 nm period patterns with EUV interference lithography

Zaidi Embong and Husin Wagiran Physics Department, University Of Technology Malaysia, P.O Box 791, 80990, Johor Baharu

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

Status of the Electron Beam Transverse Diagnostics with Optical Diffraction Radiation at FLASH

THIS IS A NEW SPECIFICATION

X-Ray Beam Position Monitor Based on a Single Crystal Diamond Performing Bunch by Bunch Detection

X-ray backscattering: Variable irradiation geometry facilitates new insights

Conservation of energy during the reflection and transmission of microwaves

Deep multilayer gratings with adjustable bandpass for XRF spectroscopy

State-of-the-art thin film X-ray optics for synchrotrons and FEL sources. Frank Hertlein Incoatec GmbH Geesthacht, Germany

Cr, Co, Cu, Mo, Ag (others on request) Mean Reflectivity: R > 70%

A Novel Multipass Optical System Oleg Matveev University of Florida, Department of Chemistry, Gainesville, Fl

Multi-Element Analysis by Portable Total Reflection X-ray Fluorescence Spectrometer

Kit for building your own THz Time-Domain Spectrometer

Medical Imaging. X-rays, CT/CAT scans, Ultrasound, Magnetic Resonance Imaging

PRELIMINARY TIME SCHEDULE start of commissioning start of user operation April 2000 July plane grating 300 / 1000 lines per mm M 2 M 3

Multi-Element Si Sensor with Readout ASIC for EXAFS Spectroscopy 1

Problems in Color Proofing from the Colorimetric Point of View

X-ray polarimetry with a conventional gas proportional counter through rise-time analysis

Transcription:

Copyright (C) JCPDS International Centre for Diffraction Data 1999 379 DOUBLE MULTILAYER MONOCHROMATOR WITH FIXED EXIT GEOMETRY H.Gatterbauer, P.Wobrauschek, F.Hegediis, P.Biini, C.Streli Atominsitut der ijsterreichischen Universitaten, Schtittelstral3e 115, A- 1020 Wien, Austria Paul Scherrer Institut, Wtirenhngen, Switzerland Abstract To use the power of Total Reflection X-ray Fluorescence Analysis (TXRF) as an analytical method for very low detection limits on surfaces it is necessary to have a monoenergetic, energy adjustable x-ray beam with an intensity as high as possible. Therefore a wide band pass monochromator is desired because it increases the flux of photons impinging on the sample and an energywidth of up to 200 ev does not affect the choice of exciting or suppressing a particular element. The importance of the high flux of photons within the energyband chosen will result in improved sensitivity and finally in better detection limits. Multilayers (ML) produced at PSI have been tested and succesfully used for that purpose. Alignment of the TXRF set-up for a given energy is time consuming and therefore a stable beam geometry is required for the energies adjusted. So in case of only one ML in the beam path this can not be fulfilled and requires a realignment of the TXRF set-up for each energy. A double ML monochromator unit allows the design of a fixed exit geometry for a wide range of energies. The technical solution, the design and construction for such an arrangement is descibed. The experience is discussed and results from experimental series with x-ray tubes, as well as at the HASYLAB Hamburg, Germany, with the newly designed equipment are presented. Finally an inherent advantage of double ML systems is the possibility to suppress higher. Introduction TXRF is an accepted method to determine small amounts of chemical elements on surfaces with extremely good detection limits [ 1,2]. The geometrical arrangement provides a couple of substantial advantages. Efficient excitation by the effect of total reflection. Both primary and reflected beam add to fluorescence radiation with nearly double intensity. As the incident beam scarcely penetrates into the material, scatter radiation is reduced. Additionally the scatter contribution shows a minimum for 90 degree between incident and scattered radiation. Both imply reduced spectral background and limited counting rates for the scatter peaks, an important aspect for energydispersive detection systems where high count throughput is a problem. Another consequence of the small incidence angle is the possibility to mount the detector very close to the sample. This results in large detection solid angles. To exploit the advantages of TXRF and to improve the detection limits several pre-conditions have to be created. TXRF needs a spatially very stable beam because small angle variations cause tremendous changes in excitation conditions. Slight vertical variations as well cause large

This document was presented at the Denver X-ray Conference (DXC) on Applications of X-ray Analysis. Sponsored by the International Centre for Diffraction Data (ICDD). This document is provided by ICDD in cooperation with the authors and presenters of the DXC for the express purpose of educating the scientific community. All copyrights for the document are retained by ICDD. Usage is restricted for the purposes of education and scientific research. DXC Website www.dxcicdd.com ICDD Website - www.icdd.com

Copyright (C) JCPDS International Centre for Diffraction Data 1999 380 horizontal shifts whereby the sample is not illuminated properly. The beam may also show a spatial profile that leads to changes in the excitation conditions if shifts occur. If single elements are of interest it is important to use monoenergetic radiation with an energy just above the absorption edge of the element. Therefore energy tuning by means of the monochromator is necessary. This monochromator should provide a spectral distribution with a FWHM value smaller than the difference of the absorption edge energies of neighboring elements if multiple elements are of concern. It also has to be able to resist very intense beams. Particularly if the excellent properties of synchrotron radiation should be used it is important that the monochromators withstand the high energy densities transferred. The increase of the intensity of the photon flux leads to an increase of the sensitivity and therefore to achieve better detection limits. Experimental Within the experiment for monochromatization multilayers were used. Beside the wide band pass, another advantage of multilayers compared to single crystal monochromators is the possibility to vary their parameters over a wide range. With appropriate chosen parameters (d- spacing, division ratio I, layerpair number N, roughness) high intensity transmission and wide energy tuning can be achieved that match the experimental demands. To provide an energy adjustable monochromator with Fixed Exit it is necessary to arrange two monochromators, in case of two equivalent multilayers, in +n/-n geometry [3-61. Batterman and Bilderback suggest a right angle geometry where the axis of rotation coincides with the point where the normal of the first multilayer intersects the elongation of the surface of the second multilayer [7]. The disadvantage of this method is that the line where the beam impinges on the first multilayer shifts if the energy changes. This can be rectified if the axis of rotation is identical to the intersection line of the incident beam on the surface of the first monochromator (Figure 1). Due to the limited length of the multilayers the second multilayer must be positioned on a translator parallel to its surface. whit direction of translation Figure 1: Fixed Exit Geometry: To select different energy adjustments the whole system has to be rotated around the axis tangent to the surface of the first multilayer. If the single monochromatized beam does not hit the second multilayer the multilayer has to be translated parallel to its surface.

Copyright (C) JCPDS International Centre for Diffraction Data 1999 381 For changing the energy the complete system must be rotated so that the first multilayer matches Braggs law for the according energy. Then the second multilayer must be translated to the point of intersection of the centre of the surface with the monochromatized beam in order to reflect the beam parallel to the incident beam. In fact the so called Fixed Exit is not really perfectly fixed. But in case of hard x-rays, which also means small angles of incidence, the height shift can be neglected. For the reason of the primary adjustment it is necessary that each multilayer is positioned on a separate rotator and vertical translator. Figure 2 gives a schematic overview of the mechanical arrangement. Figure 2: Mechanical Arrangement: Each multilayer is positioned on a separate translator and rotator for the first adjustment. The second multilayer is additionally mounted on a big translator that compensates the limited spatial dimension of the multilayer. All together is rotateable around the axis being tangent to the surface of the first multilayer. Results The experimental results with synchrotron radiation are shown in Figure 3. If the double multilayer monochromator is adjusted once, the energy tuning can be performed in a very simple way. The graphs show two different adjustments at 10 and 15 kev respectively. For higher energies a broadening of the peak of the monochromatized spectrum can be observed. To find an explanation for the high background in the spectra one has to consider some effects that

Copyright (C) JCPDS-International Centre for Diffraction Data 1999 382 are not shown in Figure 1. When the white beam hits the first multilayer only radiation of a certain energy is reflected according to Bragg s Law. The remaining photons of the beam are partly absorbed, especially the high energetic ones are scattered and also transmitted through the multilayer. Due to the small vertical distance between the primary and the double monochromatized beam the cone of the scattered radiation and the monochromatized beam intersect at the point of detection. Figure I 5 IO Ened;(ke ) 25 5 IO Ener;;(ke ) 3: Two different energy adjustments for monochromatization of synchrotron radiation with double multilayer monochromator. The relatively high background is caused by the primary radiation penetrating the first multilayer. The problems caused by scattered ratiation are not so critical when using a beam of an x-ray tube (W-anode) because the intensity of the high energy radiation is relatively small. As energy, the characteristic w-line of the anode was selected (Figure 4). On the left graph the spectrum of the beam reflected by the first multilayer is shown. For the harmonics of higher order the peaks become broader and the intensity decreases. Of course one has to take into consideration that the primary spectrum is not white. On the right graph the spectrum of the double moncxhromatized beam shows really monochromatic properties. Due to inherent slight de-adjustments of the second multilayer the higher harmonics are suppressed and the energy resolution of the first harmonic is determined by the detector properties. 1 1 5 10 15 a P Energy( ev) 30 35 40 5 IO 15 20 P 30 35 40 Energy( ev) Figure 4: Monochromatized x-ray tube spectrum at W-Lp: The left graph shows the spectrum of the single monochromatized beam with the harmonics up to the fourth order. The right one shows the spectrum of the double monochromatized beam. Due to inherent de-adjustment of the second multilayer the higher harmonics are suppressed.

Copyright (C) JCPDS-International Centre for Diffraction Data 1999 383 Figure 5 shows the analogue graphs at a Bragg angle that reflects radiation of the continuous spectrum. For the selected energy of about 28 kev the spectrum of the single monochromatized beam has a quite broad peak. But with slight de-adjustments of the second multilayer the width of the peak can be reduced. Figure 5 IO 15 20 25 Energy(keV) 30 35 40 1 5 IO 15 a R 30 Energy( ev) 5 : Monochromatized x-ray tube spectrum at 28keV: The single monochromatized beam shows a quite broad peak at this energy (left graph). By de-adjusting the second multilayer the width of the peak can be reduced (right graph). 35 Discussion of the results When using synchrotron radiation it is necessary to put a beam stopper behind the first multilayer that absorbs the transmitted and scattered primary radiation to achieve a good background. If the monochromator system is operated at higher energies it might be useful to mount a beam stopper behind the second multilayer. The beam stoppers can also be combined with an extended shielding to absorb the background radiation caused multiple scattered radiation inside the hutch. Another problem represents the damages of the multilayers. If the surrounding gas is air the highly-energetic, intense synchrotron radiation produces radicals that interact with the multilayers. This can be avoided by filling the volume surrounding the multilayers with an inert gas or evacuation. Additionally the heat damages due to energy deposition have to be taken into account. The layers can be destructed permanently if the heat load is too high. Cooling units at the rear side take remedial action. References [l] P.Wobrauschek, H.Aiginger, Anal. Chem. 47(6), 852 (1975) [2] P. Wobrauschek, C.Streli, X-Ray and Inner Shell Processes, AIP Conference Proceedings, Ed. R.L.Johnson et al., 389,233 (1997) [3] J.A.Golovchenko, R.E.Levesque and P.L.Cowan, Rev. Sci. Instrum. 52 (4), 509 (1981) [4] S.M.Heald, Nucl. Instrum. and Meth. A266,457 (1988) [5] T.I.Morrison, N.C.Lien, S.M.Heald, L.S.Fareira and J.Scrofani, Nucl. Instrum. and Meth. A266,467 (1988) [6] J.P.Kirkland, Nucl. Instrum. and Meth. A291, 185 (1990) [7J B.W.Batterman and D.H.Bilderback in Handbook of Synchrotroratiation Vol. 3, 105, Ed. G.S. Brown, D.E.Moncton, North Holland (1991)