Leica SPEII confocal microscope. Short Manual

Similar documents
Leica SP8 TCS Users Manual

Things to check before start-up.

Leica SP8 TCS Users Manual

Guide to Confocal 5. Starting session

Quick Start Guide. Leica SP5 X

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Cell Biology and Bioimaging Core

Leica TCS SP8 Quick Start Guide

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LSM 710 Confocal Microscope Standard Operation Protocol

Zeiss 880 Training Notes Zen 2.3

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Zeiss 780 Training Notes

Leica TCS SP8 Quick Start Guide

LSM 780 Confocal Microscope Standard Operation Protocol

Leica Sp5 II Confocal User Guide

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

Title: Leica SP5 Confocal User Manual

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

LSM 510 Training Notes

Microscopy from Carl Zeiss

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Leica SP8 Resonant Confocal. Quick-Start Guide

LSM 510 Meta Training Notes

Zeiss LSM 880 Protocol

TRAINING MANUAL. Olympus FV1000

b. Turn the power switch and key to on position for blue laser.

Microscope Confocal LSM510 META

Zeiss LSM 780 Protocol

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Operating Instructions for Zeiss LSM 510

LSM 800 Confocal Microscope Standard Operation Protocol

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Using the Nikon TE2000 Inverted Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

ZEISS LSM510META confocal manual

Simplified Instructions: Olympus Widefield Microscope S1230

The Zeiss AiryScan System, Confocal Four.

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

Everest System / Slidebook Operating Procedures

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Renishaw InVia Raman microscope

Zeiss Axio Imager.A1 manual

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Contents. Introduction

OPERATING INSTRUCTIONS

Microscope Confocal Sp2 Upright.

Nikon SIM-E & A1-R System

Widefield 1. Switching on

SPINNING DISK CSU-X1 USER MANUAL

Zeiss Deconvolution Microscope: A Quick Guide

RENISHAW INVIA RAMAN SPECTROMETER

Swept-Field User Guide

Supplemental Method Information Zeiss LSM710

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

Usermanual for Leica SP8 confocal

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Zeiss LSM 510 Multiphoton Confocal Microscope

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

ScanArray Overview. Principle of Operation. Instrument Components

Practical work no. 3: Confocal Live Cell Microscopy

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Instructions for the Leica SP5 II laser scanning confocal microscope

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Zeiss LSM 510 Multiphoton Confocal Microscope

Diskovery Spinning Disk Guide

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

Zeiss AxioObserver with ApoTome

Title: Nikon A1R Confocal User Manual

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Nikon A1Rsi Confocal Start-Up Sequence

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Nasmyth Ultraview Vox User Protocol

1 Co Localization and Working flow with the lsm700

Nikon E800 Microscope. Operating Instructions

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

Transcription:

Leica SPEII confocal microscope Short Manual Switching ON sequence: 1. Turn on the Workstation under the bench (top, far right). 2. Turn on the Supply Unit - Laser box (big green switch first and then key, turn clockwise). Be careful not to pull the key out. 1

3. Switch on Microscope (if needed for BF). Green switch on left hand side of microscope. 4. Switch on EL6000 fluorescent lamp if using fluorescent samples. Shutter is always open and note position of attenuator. Can open this fully or close down if bright. Wide-field fluorescence set-up: 1. Load slide onto microscope stage. Note that objectives move manually. Start with lowest x10 objective. There are four long pass filters for widefield fluorescence observation: DAPI, FITC, Rhod and CY5. These filters are changed manually using the fluorescence dial (see opposite). Position 5 is for BF and confocal imaging (Scan). 2. Open shutter (left hand side of microscope: down is open) to allow excitation light to reach sample. Close shutter (up position = in) when not observing sample and when ready for confocal imaging. 2

3. When ready for confocal imaging, pull lever from in to out (top left of microscope) and make sure FL shutter closed and BF position (5) selected on filter turret. Confocal imaging: Getting started with LAS software: Log on as TCS user (password; Leicaspe2). Select LAS AF software. Double click. Wait for software to load. Click on OK to start the initialization sequence. Turn on lasers: Select Configuration tab, then lasers. Select lasers lines that you require eg. 405 nm, 488 nm, 532, or 634 nm Image acquisition: Select the Acquire tab and Acquisition window. It will open in xyz mode and 512x512 pixel resolution at 400 Hz (default settings). Details are displayed on a single line. (Note that the Zoom factor is 1.5. This is the lowest factor to use.) 3

NB. Can pan in direction of arrows once zoom is greater than 1.5. Beam Path settings: 1. Note the Beam Path. Select a fluorophore from the Leica Settings drop-down menu. This will set the defaults for the fluorophore you chose. Alternatively, make up your own settings: click on Visible to activate the lasers and select the wavelength you wish to use by moving the slider. Start with 10-20% power. If you select the laser using the slider, you will need to also select the proper dichroic filter and pseudo- color option on the PMT 1 window. 2. IMPORTANT: Manually select the appropriate objective 3. Activate the PMT by clicking on the Active check box in the PMT 1 window. 4. Click on the NONE/SETTINGS/MORE bar to open the drop- down menu and choose the fluorophore emission wavelength to assist in setting the PMT bar. 4

5. Adjust the PMT bar by clicking on it and moving it. Resize the bar by clicking on one edge and dragging it left or right. The slider can also be resized by double clicking on it and entering the values needed. (Note: the minimum wavelength distance from the excitation wavelength should be 10 nm longer). 6. Click on the Live button at the bottom left of the screen or on the second from left button on the panelbox (see below). Optimising the image: Increase the Gain until an image appears. This is done by using the Signal PMT 1 slider, or clicking on the slider and turning the mouse wheel. Adjust the gain and offset using the QLUT (Quick Look Up Table) to change the image to color as intensity values. Set the color values to mostly orange with some blue (saturation level) and some green (offset below which image will be black). If necessary, adjust the laser intensities and PMT bar position. 5

Note: A gain value lower than 400 V would mean that you can lower the laser power and go up the smart gain until about 900-1000 V). A smart gain between 1100-1250 would suggest going up on the laser power. REMEMBER: By enhancing the laser power you will expose your sample to more laser exposition, hence your sample will bleach faster. On the other hand, enhancing the gain won t expose your sample to more laser exposition, and it will protect your sample from too much laser exposition. Thus, in order to protect your sample signal, it is better to first adjust the gain, and then if not enough signal is found, enhance the laser power percent. Click the Stop button, and then click the Capture Image button to acquire an image. When you have optimised each scan parameters, select Start. Changing the quality of your image acquisition: Averaging the Frame number can dramatically improve the quality of an acquired image. Under the acquire tab, change the number of frames to average. This will result in multiple scans where the resulting image is the average of the intensities. Acquisition of a Z Stack Under the Acquire and Acquisition tab, verify the mode is xyz. Expand the z-stack band to show the full selection window. 6

Select the current focal plane (orange diamond) in the z-stack window. Click on Live mode and using the mouse wheel, focus to the top of the sample and click the left arrow begin point to set it. Then again using the mouse wheel, focus on the lowest plane of your sample and click the End point arrow to set it. Select Stop. Set the number of z- steps by either accepting the system optimized button, or selecting the number of steps, or the step size you prefer. Click Start to begin the stack creation. It will end automatically. NB. When using sequential scan in multi-channel imaging, select Stack by Stack for speed and deselect Stack by Stack if looking at colocalisation. REMEMBER to deselect begin and end arrows and put No. of steps as 1 if do not want to re-do a z-stack. Saving Images The stack will be saved under the Experiment tab where it may be renamed (right mouse click and select rename ) and saved by selecting the Save all button. All images are saved as created as.lif (Leica Image Format) files. Be sensible about your file names and create new experiments as needed. 7

Creating a 3-Dimensional Projection: After acquiring a z- stack the images may be processed into a 3- D projection directly or using the 3D projection tool. Without animation: Select the stack to use under the Experiment tab. Click on Process > Tools > Visualization > 3D Projection. Do NOT change the X, Y, or Z planes if you want a simple projection. Click Apply. 8

With animation: Click Create a Movie. Enter the Start Rotation angle (in degrees) corresponding to the start view of the movie and click Set Start. Enter the End Rotation angle corresponding to the end view of the movie and click on Set End. Enter the Number of Frames desired for the rotation. Note that the higher the number of frames, the slower the rotation speed. Click on Apply. The 3D movie may be viewed through the Acquire > Experiment tab. Layers may be turned on or off as can the overlay view. Note: Double clicking on a view will toggle between full frame and multiple frame options. Note the 3D-orthogonal view and the gallery view 9

Switching OFF sequence: This is the reverse of turning on. Check to see if another user is coming on to the system. 1. Save all images. Shut down software: File > Exit. 2. Log hours on EL6000 fluorescent lamp and then switch off. 3. Turn off Microscope (if halogen lamp was on) 4. Turn off Supply unit - Laser box (turn key anti-clockwise to 0 and then turn off big green switch). 5. Shutdown computer. 10