PERFORM Operating Document. Use and Maintenance of G.E. Ultrasound LOGIQ E

Similar documents
Diagnostic Ultrasound System. Operation Note

Quick Reference Guide

S S S2 Operation Manual

DC-6. Diagnostic Ultrasound System

4 Working With Scan Modes

DC-6 Expert. Diagnostic Ultrasound System

M7Vet Diagnostic Ultrasound System. Operation Note

M5 Diagnostic Ultrasound System

QUICK GUIDE

Image Optimization: The Sonographer s Responsibility. Prepared by Cathy Daniels, EdD, RTR, RDMS, RDCS, RVT

Ultrasound Diagnostic Scanner INSTRUCTION MANUAL

IDEXX-PACS * 4.0. Imaging Software. Quick Reference Guide

Architecture of Quality Imaging Mary K. Henne, MS, CNMT, RDMS, RVT Ultrasound Education Specialist GE Healthcare

ACUSON X150 Knobology & User Guide

User Manual Veterinary

UGEO H60. Performance in Style. Features

Vevo770 Ultrasound. Quick Operation Guide

Lesson 06: Pulse-echo Imaging and Display Modes. These lessons contain 26 slides plus 15 multiple-choice questions.

Answer: TGC is needed to amplify echoes from deeper structures so that they appear as bright as similar structures located at more shallow depths.

Progeny Imaging. User Guide V x and Higher. Part Number: ECN: P1808 REV. F

Optimisation of Image Acquisition Bordeaux 16th November J.S. McGhie W.B. Vletter R. Frowijn No disclosures

OzE Field Modules. OzE School. Quick reference pages OzE Main Opening Screen OzE Process Data OzE Order Entry OzE Preview School Promotion Checklist

M 5. M5 Hand-carried Diagnostic Ultrasound System. Mindray. A practical, durable ultrasound tool for high-quality imaging where and when you need it.

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Table 1. Placing the Sensor in the Sensor Cradle. Step Instruction Illustration

Index. Page (s) 1 4. Features

Digital Portable Overhead Document Camera LV-1010

Copyright 2014 SOTA Imaging. All rights reserved. The CLIOSOFT software includes the following parts copyrighted by other parties:

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

KoPa Scanner. User's Manual A99. Ver 1.0. SHENZHEN OSTEC OPTO-ELECTRONIC TECHNOLOGY CO.,LTD.

Attenuation and velocity of ultrasound in solid state materials (transmission)

Progeny Imaging Veterinary

Quality Ultrasound Systems

Photo One Digital Photo Shoots and Edits

GE Healthcare LOGIQ P3. Staying ahead of the curve

The ideal K-12 science microscope solution. User Guide. for use with the Nova5000

Equipment for Attenuation and velocity of ultrasound in solid state materials (transmission), experimental set-up

Instructions for Use: Flexible Inspection Scope Camera

Printed in Japan E318

LOGIQ e Ultrasound. GE Healthcare. Product Description

CHROMACAL User Guide (v 1.1) User Guide

DC155 Digital Visualizer. User Manual

EVIS EUS ENDOSCOPIC ULTRASOUND CENTER EU-ME2 Dedicated ultrasound processor with versatile functionality.

FAST TRACK READ ME FIRST! FAST TRACK E300 A Quick-Start Guide to Installing and Using Your COOLPIX 300. Contents. Nikon View... 25

Planmeca Romexis. quick guide. Viewer EN _2

Crossfire Footswitch. User Guide

Vivid S5. Cardiovascular ultrasound system. GE Healthcare. Davis Medical

Happy Link Software INSTRUCTION MANUAL

DS1000E, DS1000D Series Digital Oscilloscope

NextGen LOGIQ e. Ultrasound. GE Healthcare. Product Description

Installation & User Manual Micro-Image Capture 7

Zeiss 880 Training Notes Zen 2.3

Table of Contents 1. Image processing Measurements System Tools...10

Wireless Handy Scanner

ARGUS RADAR Quick Reference Card

DigiScope II v3 TM Aperture Scope User s Manual

INSTRUCTION MANUAL INF Fax: (503)

Breast Ultrasound QA Phantom Recommended by Japan Association of Breast and Thyroid Sonology

Operation Manual. Canon CXDI-1 System Digital Radiography

Artifacts. Artifacts. Causes. Imaging assumptions. Common terms used to describe US images. Common terms used to describe US images

ScanMate. i920 Scanner. Scanning Setup Guide for TWAIN Applications A-61733

Magic Wand Portable Scanner with Auto-Feed Dock. PDSDK-ST470-VP-BX2 User Manual

Instructions for Use: Flexible Inspection Scope Kit

Lesson 06: Pulse-echo Imaging and Display Modes. This lesson contains 22 slides plus 15 multiple-choice questions.

Quick Start Guide for the PULSE PROFILING APPLICATION

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

eflex 75x/300x Digital Microscope with Flexible Neck User Guide

Series. Photo Printer. Direct Printing Guide

General Help. Last revised: Winter When I try to print something on the computer, it appears to work, but nothing comes out of the printer.

ZONARE. z.one pro Ultrasound System. Instructions for Use. (Including Special Procedures interface option)

reference guide reference guide reference guide

R2009/R3016 Radar system Quick Reference Card

RENISHAW INVIA RAMAN SPECTROMETER

KODAK Dental Imaging Software. Quick Start Guide

DIAGNOSTIC SCANNER HS-1600V. Convex/Linear Ultrasonic System

Revosoft Operators Manual

Physics of Ultrasound Ultrasound Imaging and Artifacts รศ.นพ.เดโช จ กราพาน ชก ล สาขาหท ยว ทยา, ภาคว ชาอาย รศาสตร คณะแพทยศาสตร ศ ร ราชพยาบาล

Digital Function Generator

Vivid i Ultra-Portable Echo System

Lesson 12: Doppler Principles. This lesson contains 50 slides plus 26 multiple-choice questions.

KIZLON. Sentinel House,Harvest Crescent, Ancells Business Park, Fleet Hampshire GU51 2UZ, UK Website:

MC3 Motion Control System Shutter Stream Quickstart

RAZER GOLIATHUS CHROMA

SONOACE 6000C MT DIGITAL COLOR. Affordable PC-based Digital CFM Ultrasound System. Specifications.

RING CUTTING SYSTEM OPERATION MANUAL. The lighter your touch, the faster GEM cuts

2012 Monitored Rehab Systems E1201 Manual Kneelax. Installation and user manual

Nikon View DX for Macintosh

GFL-1000 User Manual Ground Fault Locator

CTI Products RadioPro Dispatch User Guide Document # S For Version 8 Software

RVG 5100 and RVG User Guide

Multipurpose Color Ultrasound System IMAGING SYSTEMS

PosiTector 200. Ultrasonic Coating Thickness Gage. Standard and Advanced. INSTRUCTION MANUAL v Simple. Durable. Accurate.

DC162 Digital Visualizer. User Manual. English - 1

Design Your Performance

Excellent Performance; Ease of Use

USER MANUAL ENGLISH 1450 COIN COUNTER & SORTER

MM700. True RMS ENGLISH. INSTRUCTION MANUAL Auto-Ranging. Measurement Technology

Table of Contents. Sound-Eklin rev052511

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

SPECTRALIS Training Guide

Transcription:

PERFORM Operating Document Use and Maintenance of G.E. Ultrasound LOGIQ E PC-POD-IM-001-v02 Revision History Version Reason for Revision Date 01 New POD 22/May/2013 02 Minor changes only 2.2 Change responsibility to Ultrasound Technologist 2.3 Relevant documents to include PC-POD-IM-002 Imaging Participant Tracking 15/September/2015 Summary The content of this PERFORM Operating Document (POD) provides guidelines for the quality control, image acquisition, and maintenance of the G.E. Ultrasound LOGIQ E. PC-POD-IM-001-v02 Printed copies are not controlled. Page 1 of 15

Table of Contents SUMMARY ------------------------------------------------------------------------------ 1 1. DEFINITION OF TERMS -------------------------------------------------------- 3 2. INTRODUCTION ---------------------------------------------------------------- 4 2.1 OVERVIEW OF G.E. ULTRASOUND LOGIQ E... 4 2.2 RESPONSIBILITY... 4 2.3 RELEVANT DOCUMENTS... 4 3. PROCEDURE ----------------------------------------------------------------------- 4 3.1 QUALIFICATION OF PERSONNEL... 4 3.2 TRAINING... 4 3.3 OPERATION... 5 APPENDIX I: PROBE IMMERSION LEVELS APPENDIX II: POD TRAINING RECORD PC-POD-IM-001-v02 Printed copies are not controlled. Page 2 of 15

1. Definition of Terms ARDMS American Registry for Diagnostic Medical Sonography ATO AVI CAMRT CFM CWD DICOM Diode LED JPEG LOGIQ E MSK PDI Principal Investigator (PI) PERFORM operating document (POD) PRF PWD PYROGEN ROI TGC USB Auto Tissue Optimization Audio Video Interleave A national certifying body for medical imaging technologist professionals Color Flow Mode Continuous Wave Doppler Digital Imaging and Communications in Medicine Diode is a two-terminal electronic component with an asymmetric transfer characteristic Light Emitting Diode Joint Photographic Experts Group LOGIQ E Ultrasound brand from GE Healthcare Musculoskeletal Power Doppler Imaging Head researcher that is responsible for all aspects of a given research project or program at PERFORM. Operating documents that are specific to an instrument or technique that require approval by area managers. Pules Repetition Frequency Pulsed Wave Doppler A substance inducing fever Region of Interest Time Gain Compensation A data storage device that includes flash memory with an integrated Universal Serial Bus (USB) interface PC-POD-IM-001-v02 Printed copies are not controlled. Page 3 of 15

2. Introduction This POD is designed to outline the use, maintenance and training needs for the G.E. Ultrasound LOGIQ E. 2.1 Overview of G.E. Ultrasound LOGIQ E An evolving technology with wide applications in imaging through medical/surgical and research practices such as Cardiac, Vascular, Neurological, MSK, OB/GYN, and Urology studies. 2.2 Responsibility It is the responsibility of all users and the Ultrasound Technologist to ensure that this POD is followed. 2.3 Relevant documents PC-POD-GA-001: PERFORM Centre Booking System for Facilities and Equipment PC-POD-IM-002: Imaging Participant Tracking PC-SOP-GA-011: Guidelines for Management of Incidental Findings at PERFORM 3. Procedure 3.1 Qualification of Personnel Scans may be performed by a technologist (certified by ARDMS /CAMRT) or by a physician. Other research personnel performing ultrasound scans may do so after initial and ongoing training by a certified sonographer or physician and at the discretion of the principal investigator. 3.2 Training Basic theoretical training is a pre-requisite for any practical training. All users should receive proper training in applications and instrumentation. Users should never modify the equipment. PC-POD-IM-001-v02 Printed copies are not controlled. Page 4 of 15

3.3 Operation 3.3.1 Important Safety Considerations CAUTION Improper use can result in serious injury. The user must be thoroughly familiar with the instructions and potential hazards involving ultrasound examination before attempting to use the device. Training assistance is available from GE Medical Systems if needed. The equipment user is obligated to be familiar with these concerns and avoid conditions that could result in injury. Incorrect Setting Equipment malfunction or incorrect settings can result in measurement errors or failure to detect details within the image. The equipment user must become thoroughly familiar with the equipment operation in order to optimize its performance and recognize possible malfunctions. 3.3.2 System Power Connect the system to the electrical supply and turn the power on. LED indicators, Indicates hard disk working status. When the LED is flashing, the system is writing or reading from the hard disk (Green). The orange LED will turn on when the battery is low. PC-POD-IM-001-v02 Printed copies are not controlled. Page 5 of 15

LOGIQ E Control Panel Tour TGC. Move slide pots left/right to adjust TGC. New Patient. Press to activate these controls. Additional Feature Keys: Steer, Harmonics, PDI. Mode/Gain/Auto Keys: M Mode, Pulsed Wave Doppler (PW) Modes, Color Flow (CF) Mode and B Mode. Press these key to activate the mode; rotate the key to adjust the Gain; Press Auto to activate/deactivate auto optimization. Imaging/Measurement Key: Cursor, Clear, Body Mark Measure, M/D Cursor, Scan Area, Set / B: Press these keys, as necessary. Depth/Zoom/Ellipse: Adjust these keys, as necessary. Start/Stop. Print Keys: Press the P1, P2, P3 keys to archive, print, or send the image. Freeze: Press Freeze to freeze the image or return to scanning. Keyboard: Use the keyboard to enter patient information and annotations. Press F1, F2, F3, F4, F5 and F12 keys to activate Online help/user Manual, Arrow, Eject, Spooler, Reverse and Touch Panel Mode On/Off functions. The User can define functions for the F6-F11 keys. The following functions are available for F7- F12 Keys: CWD, 3D, LOGIQ View, CrossXBeam, ECG On/Off, Set Home, Text Overlay, Work Sheet, Grab Last, Word Delete and Video DVR. Press [Utility] to enter the Utility function and configure the system. Press [Preset] to preset the system. Press [Report] to Enter the report page. PC-POD-IM-001-v02 Printed copies are not controlled. Page 6 of 15

LOGIQ E Keys, Soft Menu Key Tour In general, there are two types of soft Menu keys: Paddle Switch and adjustable knobs. The Paddle Switch is used to access and adjust the Sub Soft Menu. Press the adjustable knobs to toggle option menu between line one and line two. Rotate the adjustable knobs to adjust the corresponding parameters. Function Keys - Programmable Keys F1 = Help (Enter Online Help / User Manual) F2 = Arrow (Annotation Arrow) F3 = Eject (Eject media) F4 = Spooler (Activates DICOM Job Spooler screen) F5 = Reverse F6-F11 Programmable F12 = Touch Panel Mode On/Off How to Program your programmable keys <Utility> - <Admin> - <Function Key>, <Key Configuration> then use the drop down menu. Choices for program Keys Shortcut keys CWD Ctrl + Alt + R: Restart the system 3D Ctrl + E: Eject LOGIQ View Ctrl + Alt + V: Output Video CrossXBeam ECG On/Off Set Home Text Overlay Grab Last Word Delete Video.DVR How to Program your hot keys <Utility> - <Admin> - <Function Key>, <Hot Key> then use the drop down menu. Choices for hot keys No function - Biopsy Guide - Save as - Active Image - Measurement Select - Auto PC-POD-IM-001-v02 Printed copies are not controlled. Page 7 of 15

Doppler Calculation - Auto Trace - DVD Format - Range Focus - OB Graph Measurement All Clear ATO - Auto TGC - Print4 - Print5 3.3.3 Logon/Starting an Exam Log-on In order to proceed with full system functionality you must log on to the system with the following details Operator: USR Password: USR New Patient To start a new patient s exam: 1. Press Patient. Press the New Patient button on the Patient menu. 2. Select the Exam Category. 3. Type the Patient ID, Patient Name, Birthdate, etc. Please refer to PC-POD-IM-002: Imaging Participant Tracking for naming convention. 4. Press the Register button on the Patient menu (DO NOT press Register if you are automatically generating a patient ID). 5. Press B, Esc, or Exit. Patient Entry Menu Image Management Window [1] Access to this patient s exam history and image management features. Function Selection Window [2] New Patient is used to clear the patient entry screen to input a new patient s data into the database. Register is used to enter new patient information into the database prior to the actual exam being performed. Details displays exam details and additional patient information. PC-POD-IM-001-v02 Printed copies are not controlled. Page 8 of 15

EZBackup, EZMove[3] One-step method to backup (move and delete p patient images) to an external media. Dataflow [4] Selects this exam s dataflow preference. Exit [5] Exits the Patient Menu and returns to scanning. Patient Information [6] Patient ID, Name, Birthdate, Age and Sex. Category Selection and Exam Information [7&8] Select the appropriate category and enter the exam information. Patient View and Exam View [9] Patient View lists the patients in the database. Search key enables searching list by Patient ID, Last Name, First Name, Birthdate, Sex, Exam today, Exam between, Exam date before, Exam date, Exam date after, Accession Number and Exam Description. String field helps define the search parameters and Clear clears the searching condition. Exam View lists the exams of the selected patient. Select the patient or the exam in Patient View and press Exam View. PC-POD-IM-001-v02 Printed copies are not controlled. Page 9 of 15

3.3.4 B/M Mode Image Optimize Power Output Optimizes image quality and allows user to reduce beam intensity. 2% increments between 0-100%. Dynamic Range Dynamic Range controls how echo intensities are converted to shades of gray, thereby increasing the adjustable range of contrast. Focus Number and Position Increases the number of transmit focal zones or moves the focal zone(s) so that you can tighten up the beam for a specific area. A graphic caret corresponding to the focal zone position(s) appears on the right edge of the image. NOTE: Push key to toggle between Focus Number and Focus Position. NOTE: Not available when Auto frequency/autodepth active. Rejection Selects a level below which echoes will not be amplified (an echo must have a certain minimum amplitude before it will be processed). Edge Enhance Edge Enhance brings out subtle tissue differences and boundaries by enhancing the gray scale differences corresponding to the edges of structures. Adjustments to M Mode's edge enhancement affects the M Mode only. Frame Average Temporal filter that averages frames together. This has the effect of presenting a smoother, softer image. Colorize Enables gray scale image colorization. To deactivate, reselect a Gray Map. Gray Map Determines how the echo intensity levels received are presented as shades of gray. Rotation (Up/Down) Rotates the image by selecting the value from the Soft key. Frequency Multi Frequency mode lets you downshift to the probe's next lower frequency or shift up to a higher frequency. PC-POD-IM-001-v02 Printed copies are not controlled. Page 10 of 15

Line density Optimizes B Mode frame rate or spatial resolution for the best possible image. 3.3.5 B Mode Control Panel Controls Auto Optimize Automatic Tissue Optimization optimizes the image based upon a specified Region of Interest (ROI) or anatomy within the display. Zoom Magnifies a zoom region of interest, which is magnified to approximately the size of a full-sized image. An un-zoomed reference image is displayed adjacent to the zoom window. The system adjusts all imaging parameters accordingly. Press Depth/ Zoom/Ellipse key to activate Zoom. Adjust the key to increase or decrease the zoom size. Use the Trackball to position the Zoom ROI. Only when the zoom size reaches the max or press the Depth/ Zoom/Ellipse key again would deactivate the Zoom, and activate Depth. Reverse Flips the image left/right. Range Focus Improves the near/mid field image quality, borders/ interfaces, increases contrast and detail resolution across the image, and allows for less filling in the vessels. 3.3.6 Color Flow/Doppler Image Optimize Baseline Adjusts the baseline to accommodate faster or slower blood flows to eliminate aliasing. PRF/Wall Filter Velocity scale determines pulse repetition frequency. If the sample volume gate range exceeds single gate PRF capability, the system automatically switches to high PRF mode. Multiple gates appear, and HPRF is indicated on the display. Wall Filter insulates the Doppler signal from excessive noise caused from vessel movement. NOTE: Push button to toggle between PRF and Wall Filter. Angle Correct Estimates the flow velocity in a direction at an angle to the Doppler vector by computing the angle between the Doppler vector and the flow to be measured. PC-POD-IM-001-v02 Printed copies are not controlled. Page 11 of 15

Quick Angle Quickly adjust the angle by 60 degrees. Threshold Threshold assigns the gray scale level at which color information stops. Doppler Display Formats Display layout can be preset to have B-Mode and Time-motion side-by-side or overunder. Sample Volume Gate Length Sizes the sample volume gate. Map Allows a specific color map to be selected. After a selection has been made, the color bar displays the resultant map. Packet Size Controls the number of samples gathered for a single color flow vector. Controls in Common with B Mode For more information on Focal Zone, Power Output, Frequency, Frame Average, Dynamic Range, Gray Map, and Colorize, refer to the B/M Mode Image Optimize section in this Quick Guide. 3.3.7 Scanning Hints Line Density Trades frame rate for sensitivity and spatial resolution. If the frame rate is too slow, reduce the size of the region of interest, select a different line density setting, or reduce the packet size. Wall Filter Affects low flow sensitivity versus motion artifact. To improve sensitivity: Increase the Gain. Decrease the PRF. Increase the Power Output. Adjust the Line Density. Decrease the Wall Filter. Increase Frame Average. Increase the Packet Size. PC-POD-IM-001-v02 Printed copies are not controlled. Page 12 of 15

Reduce the ROI to the smallest reasonable size. Position the Focal Zones properly. To decrease motion artifact: Increase the PRF. Increase the Wall Filter. To eliminate aliasing: Increase the PRF. Lower the Baseline. For venous imaging: Ensure that you have selected the vascular exam category. Select a venous application. Select the appropriate probe for very superficial structure. Select two focal zones. Adjust the depth to the anatomy to be imaged. Maintain a low gain setting for gray scale. Activate Color Flow. Maintain the PRF at a lower setting. Increase Frame Averaging for more persistence. PC-POD-IM-001-v02 Printed copies are not controlled. Page 13 of 15

CFM Top/Sub Menu Controls PWD Top/Sub Menu Controls Probe Cleaning, After Each Use Disconnect probe from ultrasound console and remove all coupling gel from probe by wiping with a soft cloth and rinsing with flowing water. Wash the probe with mild soap in lukewarm water. Scrub the probe as needed using a soft sponge, gauze, or cloth to remove all visible residue from the probe surface. Prolonged soaking or scrubbing with a soft bristle brush (such as a toothbrush) may be necessary if material has dried onto the probe surface. Rinse the probe with enough clean potable water to remove all visible soap residue. Air dry or dry with a soft cloth. Probe Disinfection, After Each Use Prepare the germicide solution according to the manufacturer's instructions. Be sure to follow all precautions for storage, use and disposal. Place the cleaned and dried probe in contact with the germicide for the time specified by the germicide manufacturer (follow the germicide manufacturer's recommended time). High-level disinfection is recommended for surface probes and is required for endocavity and intraoperative probes Probes for neuro surgical intra-operative use must NOT be sterilized with liquid chemical disinfectants because of the possibility of neuro toxic residues remaining on the probe. NeuroLOGIQal procedures must be done with the use of legally marketed, sterile, pyrogen free probe sheaths. After removing from the germicide, rinse the probe following the germicide manufacturer's rinsing instructions. Flush all visible germicide residue from the probe and allow to air dry (Appendix I). PC-POD-IM-001-v02 Printed copies are not controlled. Page 14 of 15

Probe Disinfection Agents Ultrasound probes can be disinfected using liquid chemical germicides. The level of disinfection is directly related to the duration of contact with the germicide. Increased contact time produces a higher level of disinfection. Review the probe care card that is packed with each probe. The following website contains the most current and up-to-date recommendations: http://www.gehealthcare.com/usen/ultrasound/ products/probe_case.html 3.3.8 Image Management Clipboard As images are saved by pressing any of the print keys (P1, P2, or P3), the images appear at the bottom of the display on the clipboard as thumbnails of the images saved during the exam. These images remain on the clipboard until the end of the exam. Printing Images Press the appropriate print key (P1, P2, P3). Browsing an Exam s Stored Images From the New Patient menu, open Image History. View the thumbnail images for the past exam or group images. Managing an Exam s Stored Images From the New Patient menu, open Active Images. View active exam images. Deleting an Image Select the image on the clipboard, then press the onscreen Delete shortcut or From the New Patient menu, select the image from the Review Screen of Image Management, then press Delete. Save Image to USB Memory Stick The brand validated is: 1- Kingston 512MB USB To use them the user needs to do the following: 2- Insert them into a free USB port. 3- Click on the Menu option from the right side of the screen. 4- Click Save As. 5- Select the Removable disk drive. 6- Select File Type (DICOM, JPG or AVI). 7- Provide File-Name and press Save. Note: Before pulling out the memory stick, the device should be stopped, by clicking on the Eject- Hardware Icon (see following figure), selecting the device, and pressing Stop. PC-POD-IM-001-v02 Printed copies are not controlled. Page 15 of 15

APPENDIX I Probe Immersion Levels PC-POD-IM-001-v02 Printed copies are not controlled. Appendix I

Probe Immersion Levels 1. Fluid Level 2. Aperture 3. Contact face within patient environment CAUTION In order for liquid chemical germicides to be effective, all visible residue must be removed during the cleaning process. Thoroughly clean the probe, as described earlier before attempting disinfection. You MUST disconnect the probe from the LOGIQ E prior to cleaning/disinfecting the probe. Failure to do so could damage the system.

APPENDIX II POD Training Record Form PC-POD-IM-001-v02 Printed copies are not controlled. Appendix II

POD Title Use and Maintenance of G.E. Ultrasound LOGIQ E SOP Code Ownership Document type Area SOP Number Version PC POD IM 001 v02 Training Record Full Name Institution Contact (email or phone number) Signature Sign here Date