Vibration studies of a superconducting accelerating

Similar documents
LARGE SCALE TESTING OF SRF CAVITIES AND MODULES

THE CRYOGENIC SYSTEM OF TESLA

XFEL Cryo System. Project X Collaboration Meeting, FNAL September 8-9, 2010 Bernd Petersen DESY MKS (XFEL WP10 & WP13) 1 st stage. Possible extension

Cryogenics for Large Accelerators

Overview of ERL Projects: SRF Issues and Challenges. Matthias Liepe Cornell University

two pairs of dipole steering windings that t inside the quadrupole yoke an RF beam position monitor (BPM) consisting of a pill box RF cavity,

3.9 GHz System (AH1) XFEL WP46

INTRODUCTION. Proceedings of the 12 th International Workshop on RF Superconductivity, Cornell University, Ithaca, New York, USA 558 THP43

RF STATUS OF SUPERCONDUCTING MODULE DEVELOPMENT SUITABLE FOR CW OPERATION: ELBE CRYOSTATS

GROUND MOTION IN THE INTERACTION. ensured that the final focus quadrupoles on both. rms amplitudes higher than some fraction of the

Using Higher Order Modes in the Superconducting TESLA Cavities for Diagnostics at DESY

Tuning systems for superconducting cavities at Saclay

To produce more powerful and high-efficiency particle accelerator, efforts have

LCLS-II TN Vibration measurements across the SLAC site

Texas Components - Data Sheet. The TX53G1 is an extremely rugged, low distortion, wide dynamic range sensor. suspending Fluid.

Status of the European XFEL Accelerator Construction Project. Reinhard Brinkmann, DESY

Amit Roy Director, IUAC

C100 Cryomodule. Seven cell Cavity, 0.7 m long (high Q L ) 8 Cavities per Cryomodule Fits the existing Cryomodule footprint

HIGH POWER COUPLER FOR THE TESLA TEST FACILITY

Tests of the Spoke Cavity RF Source and Cryomodules in Uppsala

R.Bachimanchi, IPAC, May 2015, Richmond, VA

High Power Couplers for TTF - FEL

INSTALLATION AND FIRST COMMISSIONING OF THE LLRF SYSTEM

ANALYSIS OF 3RD OCTAVE BAND GROUND MOTIONS TRANSMISSION IN SYNCHROTRON RADIATION FACILITY SOLARIS Daniel Ziemianski, Marek Kozien

DEVELOPMENT OF A BETA 0.12, 88 MHZ, QUARTER WAVE RESONATOR AND ITS CRYOMODULE FOR THE SPIRAL2 PROJECT

HIGH POWER PULSED TESTS OF A BETA=0.5 5-CELL 704 MHZ SUPERCONDUCTING CAVITY

Results of Vibration Study for LCLS-II Construction in the Research Yard 1

SNS CRYOMODULE PERFORMANCE*

PI piezo Life Time Test Report. A. Bosotti, R. Paparella, F. Puricelli

SADC20 with some geophone noise performance NOISE PERFORMANCE OF THE SADC20 A/D CONVERTER. and

Natural Frequencies and Resonance

A 3 GHz SRF reduced-β Cavity for the S-DALINAC

Mechanical study of the «Saclay piezo tuner» PTS (Piezo Tuning System) P. Bosland, Bo Wu DAPNIA - CEA Saclay. Abstract

EXPERIMENTAL RESULT OF LORENTZ DETUNING IN STF PHASE-1 AT KEK-STF

SRF EXPERIENCE WITH THE CORNELL HIGH-CURRENT ERL INJECTOR PROTOTYPE

Superconducting RF System. Heung-Sik Kang

Cryogenic Operations at SLAC

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

Development of the accelerometer for cryogenic experiments II

FREIA Facility for Research Instrumentation and Accelerator Development Infrastructure and Control Architecture

DEVELOPMENTS AND PROGRESS WITH ESS ELLIPTICAL CRYOMODULES AT CEA-SACLAY AND IPN-ORSAY -

Low-Level RF. S. Simrock, DESY. MAC mtg, May 05 Stefan Simrock DESY

TESLA Quad Package With BPM

Microphonics. T. Powers

5.5 SNS Superconducting Linac

SETUP I: CORD. Continuous Systems

Third Harmonic Superconducting passive cavities in ELETTRA and SLS

A Study of Magnetic Shielding Performance of a Fermilab International Linear Collider Superconducting RF Cavity Cryomodule

Status of superconducting module development suitable for cw operation: ELBE cryostats

Recent ground vibration measurements at CERN (Surface and underground)

QUARTER WAVE COAXIAL LINE CAVITY FOR NEW DELHI LINAC BOOSTER*

Stretched Wire Test Setup 1)

Brett Parker, representing the

Examination of Microphonic Effects in SRF Cavities

INSTRUMENTATION AND CONTROL SYSTEM FOR THE INTERNATIONAL ERL CRYOMODULE

ESS RF Development at Uppsala University. Roger Ruber for the FREIA team Uppsala University

ERL Prototype at BNL. Work supported by Brookhaven Science Associates, LLC under Contract No. DE-AC02-98CH10886 with the U.S. Department of Energy.

Pressure Transducer Handbook

CEBAF Overview June 4, 2010

Update 5/19/11. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton

HIGH POWER INPUT COUPLERS FOR THE STF BASELINE CAVITY SYSTEM AT KEK

R100 Microphonics. Kirk Davis, Mike Drury, Leigh Harwood, Mark Wiseman, etc. Andrew Hutton

CEBAF waveguide absorbers. R. Rimmer for JLab SRF Institute

Recent Progress in the Superconducting RF Program at TRIUMF/ISAC

S. Ghosh On behalf of Linac, IFR, Cryogenics, RF and beam transport group members. Inter University Accelerator Centre New Delhi India

3) For vibrational motion, the maximum displacement from the equilibrium point is called the

Packaging of Cryogenic Components

LORENTZ FORCE DETUNING ANALYSIS OF THE SPALLATION NEUTRON SOURCE (SNS) ACCELERATING CAVITIES *

overview of cryomodules for proton accelerators

Project X Cavity RF and mechanical design. T. Khabiboulline, FNAL/TD/SRF

Frequency Tuning and RF Systems for the ATLAS Energy Upgrade. Gary P. Zinkann

Superconducting RF cavities activities for the MAX project

Methods for reducing unwanted noise (and increasing signal) in passive seismic surveys

Development of a Vibration Measurement Method for Cryocoolers

Advances in CW Ion Linacs

System Integration of the TPS. J.R. Chen NSRRC, Hsinchu

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

KEK ERL CRYOMODULE DEVELOPMENT

VIBRATION MEASUREMENTS IN THE KEKB TUNNEL. Mika Masuzawa, Yasunobu Ohsawa, Ryuhei Sugahara and Hiroshi Yamaoka. KEK, OHO 1-1 Tsukuba, Japan

An Overview of Ground Motion and Vibration Studies

A New 2 K Superconducting Half-Wave Cavity Cryomodule for PIP-II

Borut Baricevic. Libera LLRF. 17 September 2009

Work Package Status Report

HIGHER ORDER MODES FOR BEAM DIAGNOSTICS IN THIRD HARMONIC 3.9 GHZ ACCELERATING MODULES *

Design of the 352MHz, beta 0.50, Double- Spoke Cavity for ESS

A few results [2,3] obtained with the individual cavities inside their horizontal cryostats are summarized in Table I and a typical Q o

Improving seismic isolation in Advanced LIGO using a ground rotation sensor

Operation Status of KEK Accelerator Cryogenic Systems

PROGRESS IN THE ELLIPTICAL CAVITIES AND CRYOMODULE DEMONSTRATORS FOR THE ESS LINAC

Aurélien Ponton. First Considerations for the Design of the ESS Cryo-Modules

Performance of Superconducting Cavities for the European XFEL. Detlef Reschke DESY for the EU-XFEL Accelerator Consortium

Laboratory Report INFN Milano LASA

THE HIGH LUMINOSITY PERFORMANCE OF CESR WITH THE NEW GENERATION SUPERCONDUCTING CAVITY

Initial Beam Phasing of the SRF Cavities in LCLS-II

Engineering Challenges and Solutions for MeRHIC. Andrew Burrill for the MeRHIC Team

4. Superconducting sector magnets for the SRC 4.1 Introduction

Calibration and Processing of Geophone Signals for Structural Vibration Measurements

The ILC Accelerator Complex

TESLA RF POWER COUPLERS DEVELOPMENT AT DESY.

LLRF Plans for SMTF. Ruben Carcagno (Fermilab) Nigel Lockyer (University of Pennsylvania) Thanks to DESY, PISA, KEK, Fermilab, SLAC Colleagues

Transcription:

Vibration studies of a superconducting accelerating module at room temperature and at 4.5 K Ramila Amirikas, Alessandro Bertolini, Wilhelm Bialowons

Vibration studies on a Type III cryomodule at room temperature and at 4.5K Introduction I- Introduction Systematic approach - from room temperature to 4.5K measurements (about 100 hrs of data), covering a large variety of conditions: cold mass during module assembly, warm stand alone, on CMTB warm/cold. Coming next - - Study the influence of the cryomodule support. -Study the behavior of a string of cryomodules in FLASH with implications for mechanical design and beam dynamics. Transfer function measurement using inertial sensors and the ground vibrations as broadband excitation source (quite powerful in Hamburg ), - site independent characterization of the module stability providing a usable input for the ILC and XFEL module design (our data are now being used by FNAL (M.McGee) to validate the ANSYS model of Type-IV cryomodule).

Vibration studies on a Type III cryomodule at room temperature and at 4.5K Introduction II- Z Top E He GRP Quadrupole N Girder Some definition Floor Sensors

Module 6 room temperature measurements quadrupole support stability - Type III design very stiff as expected with no internal resonances up to 100 Hz both in horizontal and in vertical. Vertical / Module 6 on CMTB 25 August 06 Module 6 cold mass on the assembly stand in DESY Hall III Horizontal transverse / CM assembly stand Horizontal transverse / Module 6 on CMTB 01 June 06 25 August 06

Module 6 room temperature measurements vessel top / quadrupole transfer function Horizontal / Module 6 on CMTB 12 February 07 The measurement is limited by the effect of the cryostat support to ~ 40 Hz in horizontal and to ~ 70 Hz in vertical. In this range no resonance pattern was found in the TF confirming i the reliability of the CM inner layout. Vertical / Module 6 on concrete blocks 23 June 06

Module 6 room temperature measurements effect of the cryostat support - Hall III CMTB-installation CMTB-with endcaps 4.7 Hz 6.2 Hz 11 Hz Effect of the support configuration+boundary conditions (pipe weldings+bellows) on the lowest frequency horizontal transverse mode with the clear stiffening after pipe weldings and closing the bellows The cryostat on its support system behaves like a compound pendulum with normal modes (rocking+translational) at low frequencies that dominates the RMS amplitude together with technical noise sources. At present the support design looks the most relevant engineering issue to ensure dynamic stability to the cryomodule.

Module 6 room temperature measurements stability along the module length I- Along the He GRP with geophones /Module 6 on blocks RMS comparison X2 Y2 Y1 24 July 06 X1 Geophones were placed on quad end, on the center and on the back end of the He GRP. The same measurement was repeated on the vessel top with seismometers. Sync_Vert_GRP RMS (nm) >2Hz >10Hz center V front V center V 156 174 156 - +12% - 84 106 78 - +26% - back V 156 +0% 94 +21% Sync_Hor_GRP RMS (nm) >2Hz >10Hz center H front H center H 329 423 226 - +28% - 57 82 50 - +44% - back H 227 +0% 77 +54% Comment RMS larger at the end Vertical: +23% @10Hz believable Horizontal: +50% @10Hz not believable (overestimated) because the signal are not coherent in that region due to the support system. More measurement needed.

Module 6 room temperature measurements stability along the module length II- Module 6 on CMTB with more realistic installation almost no difference between center and module ends vessel center quadrupole side 23 January 07 vessel center quadrupole side 23 January 07 Horizontal transverse Vertical vessel center quadrupole side 23 January 07 vessel center quadrupole side 23 January 07 Horizontal transverse Vertical

Cold quadrupole vibration measurements on Module 6 at CMTB -goals- Test geophone behaviour at 4.5 K on-board seismic sensor with adequate noise level down to below 1 Hz potentially available for quad and cavities. behaviour unknown, never been tested by the manufacturer in these extreme conditions, the company recommends use of the device down to -40 C, only one cryogenic application cited in literature. t the very robust and mature (~30 years) design was encouraging and the test has been successful. Quadrupole vibration measurements at 4.5 K Module 6 on CMTB: quadrupole side Geophones installed on the quad front face chance to give a first quantitative (from 1 Hz) evaluation of the impact of cryogenic plant and high h gradient RF on the quadrupole vibration level, not possible so far because of the lack of sensitivity of cooled piezo accelerometers below ~10 Hz

Cold quadrupole vibration measurements on Module 6 at CMTB experiment setup - Data acquisition measured noise ~ 1 nm/ Hz Vessel top vertical sensor (horizontal transverse companion not visible) Cold/warm Geophone 20 m cable Ground vertical sensor (horizontal transverse companion not visible) preamp Test amplifier 2 2 ωω0 ω0 ω + i Ql 2 2 ωω1 ω1 ω + i Q Inverse e filter Inverse filtering provides equalization of the geophone response down to 0. 5 Hz l 24 bit Güralp digitizer 200 S/s Laptop Data logging Spectral analysis (FFT, PSD, Coherence, RMS,Transfer function,etc.)

Cold quadrupole vibration measurements on Module 6 at CMTB geophones at low temperature- Remote calibration method Accurate remote calibration possible using the signal cable itself; no access to the sensor is necessary. By measuring the electrical impedance vs frequency at the output terminals of the sensor we have access to both electrical l and mechanical parameters. Only the suspended mass has to be known. Z E ( ω ) = R + j ω L + coil 2 0 jω G Geophone equivalent impedance 2 2 / ω ω + m ωω0 j Q l No loss of sensitivity at liquid helium temperature!! Block diagram of the calibration procedure

Cold quadrupole vibration measurements on Module 6 at CMTB reference warm measurement I- General features of the spectra Typical DESY site spectrum at frequencies 1-10 10 Hz. Technical noise dominating > 20 Hz; strongest peak from the insulation vacuum pump at 48 Hz. Effect of the cryostat support well visible: coupling with rocking modes at 11 Hz and 18 Hz and vertical resonance at 27 Hz; quad vs top transfer function almost flat below 40 Hz. Values @ 1 Hz: Ground 76 nm Top 90 nm Quad 103 nm *2 hrs data measured at the end of the 10th thermal cycle

Cold quadrupole vibration measurements on Module 6 at CMTB reference warm measurement II- RMS analysis In the low frequency band the quadrupole motion tracks the ground vibration level. Slight amplitude differences are related mainly to the mechanical transfer function of the module on its support.

Cold quadrupole vibration measurements on Module 6 at CMTB cold measurement I- RMS analysis Ground motion tracking confirmed at low frequencies, with ~10% quad/gnd and top/gnd rms ratios. No difference with warm operation. The refrigeration system doesn t affect the quadrupole vertical stability at low frequency (f<30 Hz). Large vibrations due to the onset of a strong peak above 30Hz. The peak shows slow frequency changes from 30.1Hz up to 32Hz. The amplitude can vary from 200 nm up to > 1 µm. Not a mechanical resonance of the cryomodule; not visible at all in the quad vs top transfer function. Steady state *data measured at the end of the 11th cooldown

Cold quadrupole vibration measurements on Module 6 at CMTB cold measurement II- 08 March in steady state Quad LHe inlet flow: 8.2 g/sec Quad LHe inlet valve: 20% Cavity 2K Inlet valve: 56% Cavity 2K He flow: 5 g/sec Cavity 2K He reservoir level:43% Comments Peak frequency ~ 32Hz in this case. The integrated RMS @1 Hz values are 78 nm (ground), 206 nm (vessel top), 260 nm (quad). The peak is also visible in the ground spectrum. A strong correlation between the average vibration level and the settings of the LHe forward line inlet valve has been found. All the clues point to a thermal acoustic oscillation generated in the diagnostic pipe immediately upward of the same valve.

Cold quadrupole vibration measurements on Module 6 at CMTB cold measurement III- 01 March High power RF Quad LHe inlet flow: 5.3 g/sec Quad LHe inlet valve: 100% Cavity 2K Inlet valve: 50% Cavity 2K He flow: 3.8 g/sec Cavity 2K He reservoir level:43% *1.5 hrs data taken between 6:30 and 8 PM; klystron at 10 Hz, ~27 MV/m average gradient Comments Peak frequency ~ 30.6 Hz in this case. The average integrated RMS @1 Hz values are 50 nm (ground), 215 nm (vessel top), 500 nm (quad). Larger RMS have been measured even with RF off or during LLRF tests. The RF doesn t affect the vibration stability of the module.

Vibration studies on a Type III cryomodule at room temperature and at 4.5K Summary - Geophone test at 4.5K classic 4.5 Hz industrial geophone can operate at 4K without any loss of sensitivity remote high accuracy calibration procedure demonstrated Quadrupole vibration measurements at 4.5K low frequency (1-100 Hz) quadrupole vertical stability is not affected by high gradient RF operation quadrupole vertical stability is not affected by the refrigeration system at frequencies up to 30 Hz; results not conclusive at higher frequency because of the onset of a thermal acoustic oscillation in a diagnostic pipe upward of the LHe forward line (preliminary i result). Next investigation will continue on Module 5 (Type III) on the CMTB, starting ti from end next week, and in the FLASH linac where we could monitor continuously a string of 3 Type III cryomodules during normal operation with geophones (from August) Warm/cold test t on Module 8 (Type III+) with both geophones and Doppler velocimeter. Possible a more detailed test on stability along the module with geophones aboard the cold mass at different positions? *Special thanks to: DESY MKS crew for the continuous support, Markus Kubczigk and Heiko Ehrlichmann for providing software tools