PlaneWave CDK Telescope Instructions. Setting the spacing and collimation for the CDK14/17/20/24

Similar documents
Hubble Optics CDK 17 Collimation Instructions 03/27/2012 Hubble Optics

SIPS instructions for installation and use

User Manual. Trinocular Infinity Compound LED Microscope. MicroscopeNet.com

Step 1 Remove Focus Knob

Replacing the Reciprocator on the SWF Compact Series Machine (601C and 1201C)

AG Optical Systems. Newtonian Astrograph Manual Version AG Optical Systems

Basic steps to time the Gammill quilting machine s rotary sewing hook

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Eric B. Burgh University of Wisconsin. 1. Scope

Thorvie Instructions Always wear eye protection and do not wear loose clothing when operating machinery. FOR AV- 41 ICE AUGER MACHINE

ABM International, Inc.

Removing and Replacing the Y-truck

AR-15 Lower Receiver Assembly Instructions

Fotoman Panoramic Cameras

STEINBERGER TRANSTREM (TYPE 2) TECHNICAL DOCUMENT

Replacing the Reciprocator on an SWF Multi-head.

Howie's Laser Collimator Instructions:

Small Flat Panel Height-Adjustable Dual Swing Arm Wall Mount With Universal Slat Wall Mount (MSP-JS-KWG-110)

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

This manual will aid in the assembly of the FireBall V90 and FireBall X90. The assembly of both machines will be identical, unless specified.

Rim-Lock Door Set Installation Instructions

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

Replacing the build plate clamps

Big & Tall Executive Chair Assembly Instructions

INSTALLATION INSTRUCTIONS Small Flat Panel Height-Adjustable, Extended Pitch Swing Arm Wall Mount Model KWE-110

INSTALLATION INSTRUCTIONS

ECEN 4606, UNDERGRADUATE OPTICS LAB

EmagiKit. Privacy Pod Plus. Quiet. Easy. Affordable. INSTRUCTIONS ASSEMBLY

INSTRUCTION MANUAL. X-Wedge for LX200 and LX600 telescopes _07028_XWedge_InstructionManual.indd 1

7. Michelson Interferometer

User Manual. Trinocular Metallurgical Microscope. MicroscopeNet.com

It is highly recommended that you use a thread lock compound such as Loctite brand on all threads to keep them from vibrating loose.

Quill Stop V2 Installation Guide 11/16/2014

The only coma-corrector for Newtonian telescopes which does not increase the focal length. An f/4 Newton stays an f/4 Newton! Properties of the MPCC:

Kai Installation Instructions

INSTALLATION INSTRUCTIONS

AG Optical Systems LLC

Cook's Corner Kitchen Assembly Instructions

EDGE2 DUAL MONITOR ARM

RULTRACT RETRACTOR CABLE REPLACEMENT INSTRUCTIONS. Rultract, Inc. is the ONLY authorized service center in the U.S.A.

AG Optical Systems. Imaging Dall Kirkham Astrograph Manual Version AG Optical Systems

5X Racing Miata Radio Delete Plate Installation Instructions

Installation and Assembly: In-wall Mount for 32" to 71" Flat Panel Screens

INSTALLATION INSTRUCTIONS

USING THE 2 TELETUBE XLS TM & TELECAT XLS TM ADJUSTABLE SIGHT TUBE

Customer Notice: Congratulations again on your SawStop purchase, and thank you! -SawStop Tualatin, OR

Wagon Vise Retrofit Installation Instructions. American Craft Woodworks. Wagon Vise

ASSEMBLY AND ADJUSTMENT

Complete Dovetail Jig Instructions

Electric Skein Winder

CNC Router Parts. Standard Rack & Pinion Drive Assembly Instructions

Hyperion. 16 f/7.3 Astrograph Operating Instructions

Hollywood Swing Away 2 and 4 Bike Racks Assembly and Installation Guide

Copyright 2009, Seymour Products LLC, All rights reserved. GET THE SCREEN, KEEP THE SOUND

ASSEMBLY AND ADJUSTMENT


AndyMark DART 12.

machines 608 Trestle Point Sanford, FL Phone Fax


Z-Truck Up-and-Down Motion. Y-Truck Side-to-Side Motion. Head. Squaring Plate. Sliding Plate FIGURE 1: THE CARVEWRIGHT MACHINE

Installation Instructions FW8S-STXL / FW8G-STXL Filter Wheel

Service Manual for XLE/XLT Series Laser Engravers

Installation and Assembly: In-wall Mount for 32" to 71" Flat Panel Displays

Olivier Thizy François Cochard

Baby Grande or Grande Crank Shade with Cables and Housing Installation Instructions

Rotary Fixture M/V/X CLASS LASER SYSTEMS. Installation and Operation Instructions

Starizona CGEM Landing Pad

PFW 6875 Installation Guide Installationsanleitung, Guía de Instalacíon, Guida de Installazione, Guide d Installation, Installatie gids

Q-Zone Hoop-Frame. Assembly Instructions. Copyright July 11, 2018 Grace Company (Reproduction Prohibited) Version 1.8

STENCIL MACHINE OPERATION uline.com H-259, H-347 H-408 CUTTING THE OIL BOARD INSERTING THE OIL BOARD

LCD MONITOR/TV WALL MOUNT

INSTALLATION INSTRUCTIONS Flat Panel Static Wall Mount Model: GSM-111

M14 MODULAR CHASSIS SYSTEM (MOD 1) INSTRUCTION MANUAL. West Springfield, MA Phone: (866) FAX: (413)

Installing the Partridge RA Extension on Losmandy G11

PFW 6870 Installation Guide Installationsanleitung, Guía de Instalacíon, Guida de Installazione, Guide d Installation, Installatie gids

VERSION 1.0 JANUARY 5, 2013 R2-ATL MOTOR MOUNT KIT ASSEMBLY GUIDE ASTROMECH DRIVE SYSTEM

Astro-Physics Inc. 400QMD Lubrication/Maintenance Guide

The Phoenix. Professional Quilting Frame. Copyright January 1, 2016 Jim M. Bagley, GraceWood, Inc (Reproduction Prohibited) Version 2.

M14 MODULAR CHASSIS SYSTEM (MOD 1) INSTRUCTION MANUAL. West Springfield, MA Phone: (866) FAX: (413)

Odyssey Elementary Stem Installation

AG Optical Systems. Imaging Harmer Wynne Astrograph Manual Version

Strata. urniture. Adriana Instructions. Parts in the Arm Box: Parts in the Body Box: Watch our assembly videos at

08+ KAWASAKI KLR PD NERF

Motorized or Crank Operated Fortress Zipper Track Shade with Housing and Side Track Installation Instructions

GRILLE GUARD SPRINTER VAN (EXCLUDES X4) INCLUDES MERCEDES, FREIGHTLINER AND DODGE PARTS LIST:


PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

Using a Howie Glatter Laser for Collimation of the Altair Deepfield RC

TABLE OF CONTENTS. Safety notes i. Care and Maintenance. ii. 1. Components Illustration Installation of Components.. 4

Fig. 2 DORMA-Glas Stand/Issue 02/03 Seite/Page 1/7

#4500 BULLET SIZER & LUBRICATOR

MODEL SK61732 COMPRESSOR SERVICE KIT

Flat Panel Dual Swing Arm Wall Mount (FWD-110) MSP-SA

Setup & Operating INSTRUCTIONS. for (FOR PIN FITTING AND ROD RECONDITIONING)

Timing the Millennium, Freedom, Liberty and Discovery

Patton Robotics ESRA II Expressive System for Robotic Animation

WHEN DISASSEMBLING ALUMINUM EXTRUSION, TIGHTEN ALL SETSCREWS AND LOCKS TO PREVENT LOSS DURING SHIPPING

model tsa-sa48 Sliding Crosscut Table installation guide

QB78 CO 2 Pellet Rifle

Installation instructions for FC14S Forward Controls for Shadow ACE and Aero 1100

Transcription:

PlaneWave CDK Telescope Instructions Setting the spacing and collimation for the CDK14/17/20/24

Collimation and Secondary Spacing Procedure The CDK optical design has four optical elements shown in Figure 1. The primary mirror and the two-element lens group are permanently mounted and aligned at PlaneWave Instruments. The secondary mirror is movable to allow fine collimation and to set the spacing between the primary and secondary. To get the best performance out of the CDK optical system, the optics must be collimated and the primary-to-secondary mirror spacing must be set. Please note that the primary-to-secondary spacing is set at the factory and typically does not need to be adjusted unless it was moved after shipping. Figure 1: The optical layout of the CDK system. The system contains 4 elements: the primary mirror, the secondary mirror, and a two-element lens group. Step 1: Collimation / Spacing Adapters for the CDK If your CDK came with the optional Hedrick focuser (200340) attached, You must first remove the attached focuser before attaching the Ronchi Adapter and Spacer. a. Use an Allen wrench and remove the three socket head bolts that holds the focuser retaining ring. Make sure that you are holding the body of the focuser while removing the screws Focuser Retaining Ring Focuser Mounting Screws Figure 2 Focuser and Retaining Ring V20151027 1

b. Remove focuser from the back plate of the telescope. c. Place the Ronchi Adapter in the back plate of the telescope and replace the focuser retaining ring to hold the adapter in place. See Fig 4 Note: For packaging purposes the Ronchi Adapter may come in two pieces; the round mounting plate and the adapter. There should be three 10-32 Flat head screws included to attach the two pieces. d. Insert the Ronchi Spacer into the Ronchi Adapter and lock it down with the thumb screw. e. Place the barrel of a low power eyepiece (for collimation) or the Ronchi Ocular (for spacing) into the Ronchi spacer and lock it down with the thumb screw. Fig3 Adapters required for CDK telescope Fig 4 Ronchi Ocular, Spacer and Adapter Step 2: Rough Collimation: To collimate your CDK you will need to have a low power, 1.25 eyepiece inserted into the Ronchi Spacer. a. Point the telescope at a bright star. b. Defocus the star until it becomes a donut-like ring. If the donut hole appears well centered in the donut, proceed to Step 3. c. If the donut hole is not centered, adjust the collimating screws on the back of the secondary mirror so that the defocused star moves in the direction of the fat side of the donut, as shown in Figure 5. Re-center the star and repeat this process until the donut hole looks centered as shown in Figure 6. See below for details on how to adjust the collimating screws. V20151027 2

Figure 5: An example of what you would see through a low-power eyepiece for an out-of-focus, out-ofcollimation telescope. To collimate this telescope you would adjust the secondary collimating screws so that the out-of-focus star moves in the direction of the arrow. Figure 6: An example of an approximately collimated defocused star. The CDK uses three ¼-20 socket head cap screws for collimation, as shown in Figure 7. These screws require a 3/16 Allen wrench. It may be necessary to slightly loose the two opposing screws in order to tighten the third screw. Remember to always finish tightening with the secondary preloaded securely against all three screws. Collimating is usually a two-person job: one person looks through the eyepiece while one person adjusts the collimating screws. Attempting to do this alone can be quite difficult and frustrating. If necessary, you might try using a webcam with a 1-1/4 nosepiece, such as the Celestron NexImage, in place of an eyepiece. With this setup, you can watch the video stream from the camera while you make adjustments. Note that the central spacing knob is not used until Step 3. V20151027 3

Central Spacing Knob Collimating screw Figure 7: The CDK secondary housing. There are three collimating screws and a central knob. The secondary on the CDK is spring loaded, pulling the secondary into the three collimating screws. Step 3: Setting the Primary-to-Secondary Spacing: The primary-to-secondary spacing is set at the factory. You will typically not need to adjust this unless the secondary has been moved. For your CDK optical system to perform as well as it should, the spacing between the primary mirror and secondary mirror should be set to an accuracy of +/- 1mm! Fortunately, you won t have to directly measure this spacing. When the primary-to-secondary spacing is set correctly, the focal plane will land at an exact known distance behind the fully racked in focuser. The Ronchi Adapter and Spacer places the Ronchi grating of the Ronchi Ocular precisely where the focal plane is supposed to be. About the Ronchi Test Figure 8 illustrates the basic idea behind using a Ronchi grating to identify the position of the focal plane. The light from a single bright star passes through the optical system and converges to a point on the focal plane. When the Ronchi grating is placed in front of or behind the focal plane, it will block some of the light in this light cone. This will cast a series of shadows that Figure 8: The Ronchi grating casts fewer shadows as V20151027 it moves closer to the focal plane. 4

appear as dark, wide lines when the defocused light is viewed through the ocular. As the grating moves closer to the focal plane, fewer lines are visible, and they become wider. When the focal plane and the Ronchi grating are perfectly aligned, no lines (or perhaps one very wide line) are visible. See Figure 9 for an example. In order to set the primary-to-secondary distance, you will be moving the secondary mirror. By adjusting secondary mirror distance, you will be moving the focal plane relative to the fixed position of the Ronchi grating. The job here is to move the secondary mirror such that the focal plane coincides with the Ronchi grating on the Ronchi Ocular. When you have done this, you have accurately set the primary-to-secondary spacing. With reasonable effort, the primary-to-secondary spacing can be set to well within its 1mm tolerance. Figure 9: The left image is a simulation of what a focused star looks like through a Ronchi grating. The right image is a simulation of what an out of focus view looks like through a Ronchi grating. The right image shows 2 full lines. The more lines you see, the farther the Ronchi screen is from the focal plane. Step 4: Setting Up the Test Starting with a telescope that is fairly well collimated, center a bright star and replace the low power eyepiece with the Ronchi Ocular that was provided with the telescope. Make sure the Ocular is seated all the way into the adapter. If the Ronchi is nulled (you can see no lines, or one Ronchi line covers the entire aperture; see Figure 9) This means that your spacing is already set; you may proceed to Step 4. This is not uncommon because the primary-tosecondary spacing is set at PlaneWave and you most likely will not need to adjust it. If you see two or more lines, you should move the secondary toward or away from the primary. Determining the Direction to Move the Secondary To determine which way to move the secondary, gently slide the Ronchi Ocular out of the spacing tube. If the Ronchi reaches null as you do this, the focal plane is too far back and V20151027 5

you need to move the secondary mirror away from the primary. If the number of lines grows as you slide the Ronchi out of the spacing tube, the focal plane is too far forward and you need to move the secondary toward the primary. Remember that the secondary mirror acts as a magnifier. So a small change in secondary spacing can translate into a much larger change at the focal plane. For example, for every 1mm the secondary spacing is changed on the CDK17, the focal plane moves 4.4mm. You want to keep the error at the focal plane to 5mm or better, which equals just under two lines on the Ronchi. This corresponds to an error in primary-to-secondary spacing of about +/- 1.1mm. See table below for spacing tolerance for all CDK models. CDK Focal plane change for every 1mm of secondary spacing change 12.5 7.5mm +/-.7mm 14 4.9mm +/- 1.0mm 17 4.4mm +/- 1.1mm 20 5.1mm +/- 1.0mm 24 4.1mm +/- 1.2mm Allowable primary-to-secondary spacing tolerance to keep focal plane within 5mm of optimal Figure 9 shows a Ronchi simulation with the Ronchi screen set at the focal plane and also with the focal plane 10mm away from the Ronchi screen. The image on the left is what you will see if the primary-to-secondary spacing is perfect. The image on the right is what you will see if the primary-to-secondary spacing is off by about 2mm. Remember, you want better than 1mm spacing accuracy for the CDK17. Adjusting the Secondary Spacing The CDK secondary mirror is not spring loaded, so you must always keep tension in the collimating screws and the central bolt. To move the secondary toward the primary mirror: 1. Loosen the four collimation knobs. Loosen them in equal amounts to maintain your collimation. 2. Hold the secondary housing with one hand. This is to keep the housing from rotating. (The four screws sit in shallow recesses and you would like them to go back in the recesses when the procedure is complete). 3. Take a flat-head screwdriver and place the tip in the central bolt of the secondary assembly. Rotate the central bolt in a clockwise direction. For reference, rotating ¼ turn moves the secondary.4mm, which moves the focal plane 1.76mm in the case of the CDK17. 4. Rotate the four collimation knobs in equal amounts until the assembly is tight again. 5. Check the Ronchi grating and repeat this process as necessary until you get a null. To move the secondary away from the primary: 1. Loosen the four collimation knobs. Loosen them in equal amounts to maintain your collimation. 2. Hold the secondary housing with one hand. This is to keep the housing from rotating. (The four screws sit in shallow recesses and you would like them to go back in the recesses when the procedure is complete). V20151027 6

3. Take a flat-head screwdriver and place the tip in the central bolt of the secondary assembly. Rotate the central bolt in a counter-clockwise direction. For reference, rotating ¼ turn moves the secondary.4mm, which moves the focal plane 1.76mm in the case of the CDK17. 4. Rotate the four collimation knobs in equal amounts until the assembly is tight again. 5. Check the Ronchi screen and repeat this process as necessary until you get a null. Re-checking Collimation After you are satisfied that the secondary spacing is correct, replace the Ronchi ocular with the low power eyepiece and repeat step 1 to verify that the collimation is still close. If recollimation adjustments are necessary, make them and repeat step 3. Step 4: Fine Collimation: Use a high-power eyepiece (5mm or less) and follow the same procedure described in Step 1 (adjust the secondary so that the donut hole is in the center of the donut). You may want to switch to a low-power eyepiece while making adjustments and re-centering the star, and back to high power to check the collimation. Finally, use the Ronchi Ocular to re-check the spacing. As long as only small adjustments were made in Step 3, the spacing should still be fine. If it is not, repeat Step 2 and Step 3. Back Focus The CDK design does not have any moving optical components, so the focal point is fixed. This means that the focal plane of the telescope is a set distance behind the backplate of the telescope. The back focus for the various telescope models in shown in Table 1 below. Backfocus from mounting surface of telescope backplate Back focus from fully racked in focuser CDK12.5 10.45 7.2 CDK14 11.09 8.09 CDK17 10.24 7.24 CDK20 8.81 5.81 CDK24 *14.1 5.81 CDK700 **12 8.5 Table 1: Shows the back focus with and without the focuser installed for the various PlaneWave telescope models. *CDK24 can also use optional spacers to increase back focus to be the same as the CDK17 or CDK14. **The CDK700 is referenced, not from the backplate of the telescope, but from the end of the altitude hub which is the mounting surface for the optical train. Also, the CDK700 is using the Focuser/Rotator 600180, which uses 3.5 of backfocus on the CDK700. V20151027 7

Mounting SecureFit Spacers to the Hedrick Motorized Focuser 2) Face Plate Attachment Screws 3) Face Plate Leveling Screws 1) SecureFit Mounting Screws 1. The face plate of the focuser (200340) has four threaded holes that should be used to mount any of Planewave s SecureFit adapter. Use the ¼-20 screws provided with your SecureFit adapter to mount it on the face of the focuser. 2. The remaining four holes has screws that hold the face plate to the body of the focuser and should not be removed. V20151027 8

3. There are also a series of set screws that are used to level the face plate relative to the image plane. These set screws should not be removed or loosened in any way. V20151027 9