FIRST DATA FROM THE ATLAS INNER DETECTOR FSI ALIGNMENT SYSTEM

Similar documents
First data from the ATLAS Inner Detector FSI Alignment System

The Multi-channel High Precision ATLAS SCT Alignment Monitoring System: A Progress Report

Frequency Scanned Interferometer Demonstration System

Frequency Scanned Interferometer Demonstration System

Frequency Scanned Interferometer for LC Tracker Alignment

Frequency Scanned Interferometry for ILC Tracker Alignment

Frequency Scanned Interferometer for ILC Tracker Alignment

Absolute distance interferometer in LaserTracer geometry

Wavelength Control and Locking with Sub-MHz Precision

Installation and Characterization of the Advanced LIGO 200 Watt PSL

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

R. J. Jones Optical Sciences OPTI 511L Fall 2017

Stability of a Fiber-Fed Heterodyne Interferometer

Optical Vernier Technique for Measuring the Lengths of LIGO Fabry-Perot Resonators

3 General Principles of Operation of the S7500 Laser

arxiv:physics/ v1 [physics.ins-det] 21 Sep 2006

MOI has two main product lines for its component business: 1. Tunable filters (FFP-TF, FFP-TF2, FFP-SI) 2. Fixed filters (FFP-I, picowave)

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

Dispersion measurement in optical fibres over the entire spectral range from 1.1 mm to 1.7 mm

High stability multiplexed fibre interferometer and its application on absolute displacement measurement and on-line surface metrology

Swept Wavelength Testing:

R. J. Jones College of Optical Sciences OPTI 511L Fall 2017

Development of C-Mod FIR Polarimeter*

The VIRGO injection system

Frequency-stepping interferometry for accurate metrology of rough components and assemblies

DEPARTMENT OF THE NAVY. The below identified patent application is available for licensing. Requests for information should be addressed to:

A Multiwavelength Interferometer for Geodetic Lengths

Actively Stabilized Scanning Single-Frequency. Ti:Sa /Dye Ring Laser External Doubling Ring Ti:Sa /Dye Standing Wave Laser

Fast Widely-Tunable CW Single Frequency 2-micron Laser

Testbed for prototypes of the LISA point-ahead angle mechanism

taccor Optional features Overview Turn-key GHz femtosecond laser

7 CHAPTER 7: REFRACTIVE INDEX MEASUREMENTS WITH COMMON PATH PHASE SENSITIVE FDOCT SETUP

High-Coherence Wavelength Swept Light Source

OPTICAL NETWORKS. Building Blocks. A. Gençata İTÜ, Dept. Computer Engineering 2005

PHYS 3153 Methods of Experimental Physics II O2. Applications of Interferometry

LISA and SMART2 Optical Work in Europe

MASSACHUSETTS INSTITUTE OF TECHNOLOGY Department of Electrical Engineering and Computer Science

Doppler-Free Spetroscopy of Rubidium

SA210-Series Scanning Fabry Perot Interferometer

The Compact Muon Solenoid Experiment. Conference Report. Mailing address: CMS CERN, CH-1211 GENEVA 23, Switzerland

Ph 77 ADVANCED PHYSICS LABORATORY ATOMIC AND OPTICAL PHYSICS

History of Velocimetry Technology

Fibre Optic Sensors: basic principles and most common applications

Fast Optical Form Measurements of Rough Cylindrical and Conical Surfaces in Diesel Fuel Injection Components

UNMATCHED OUTPUT POWER AND TUNING RANGE

LightGage Frequency Scanning Technology

The CMS Silicon Strip Tracker and its Electronic Readout

Thermal correction of the radii of curvature of mirrors for GEO 600

SENSOR+TEST Conference SENSOR 2009 Proceedings II

OPTICS IN MOTION. Introduction: Competing Technologies: 1 of 6 3/18/2012 6:27 PM.

Cavity BPMs for the NLC

2003 American Institute of Physics. Reprinted with permission.

Spatially Resolved Backscatter Ceilometer

SUPPLEMENTARY INFORMATION DOI: /NPHOTON

Micro-sensors - what happens when you make "classical" devices "small": MEMS devices and integrated bolometric IR detectors

DIODE LASER SPECTROSCOPY (160309)

BMC s heritage deformable mirror technology that uses hysteresis free electrostatic

Actively Stabilized Scanning Single Frequency. Ti:Sa /Dye Ring Laser

Stabilized Interrogation and Multiplexing. Techniques for Fiber Bragg Grating Vibration Sensors

Stable dual-wavelength oscillation of an erbium-doped fiber ring laser at room temperature

3550 Aberdeen Ave SE, Kirtland AFB, NM 87117, USA ABSTRACT 1. INTRODUCTION

DWDM FILTERS; DESIGN AND IMPLEMENTATION

Kit for building your own THz Time-Domain Spectrometer

IST IP NOBEL "Next generation Optical network for Broadband European Leadership"

arxiv: v1 [physics.optics] 11 Aug 2012

1.6 Beam Wander vs. Image Jitter

TEST AND CALIBRATION FACILITY FOR HLS AND WPS SENSORS

B. Cavity-Enhanced Absorption Spectroscopy (CEAS)

PowerSource TM. Tunable High Power CW Laser Module with Integrated Wavelength Monitoring 1935 TLI. Principle and Setup CONTENTS DESCRIPTION STANDARDS

Keysight Technologies Using a Wide-band Tunable Laser for Optical Filter Measurements

Herwig Schopper CERN 1211 Geneva 23, Switzerland. Introduction

TURNKEY, ULTRA STABLE, OEM LASER DIODE SOURCE OZ-1000 & OZ-2000 SERIES

External cavities for controling spatial and spectral properties of SC lasers. J.P. Huignard TH-TRT

First step in the industry-based development of an ultra-stable optical cavity for space applications

Concepts for High Power Laser Diode Systems

S.R.Taplin, A. Gh.Podoleanu, D.J.Webb, D.A.Jackson AB STRACT. Keywords: fibre optic sensors, white light, channeled spectra, ccd, signal processing.

Numerical analysis of a swift, high resolution wavelength monitor designed as a Generic Lightwave Integrated Chip (GLIC)

Supplementary Figures

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

Development of a Low Cost 3x3 Coupler. Mach-Zehnder Interferometric Optical Fibre Vibration. Sensor

MEASUREMENT APPLICATION GUIDE OUTER/INNER

Laser Alignment System for LumiCal

Status of the ATF extraction line laser-wire.

Module 19 : WDM Components

STRETCHED WIRE OFFSET MEASUREMENTS: 40 YEARS OF PRACTICE OF THIS TECHNIQUE AT CERN

attosnom I: Topography and Force Images NANOSCOPY APPLICATION NOTE M06 RELATED PRODUCTS G

FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS. Version 1.0 MICRON OPTICS, INC.

Chapter 10 WDM concepts and components

It s Our Business to be EXACT

Dual-wavelength Fibre Biconic Tapering Technology

The CMS electromagnetic calorimeter barrel upgrade for High-Luminosity LHC

Interferometer signal detection system for the VIRGO experiment. VIRGO collaboration

Silicon nitride based TriPleX Photonic Integrated Circuits for sensing applications

OPSENS WHITE-LIGHT POLARIZATION INTERFEROMETRY TECHNOLOGY

Lecture 6 Fiber Optical Communication Lecture 6, Slide 1

A novel tunable diode laser using volume holographic gratings

International Conference on Space Optics ICSO 2000 Toulouse Labège, France 5 7 December 2000

The Henryk Niewodniczański INSTITUTE OF NUCLEAR PHYSICS Polish Academy of Sciences ul. Radzikowskiego 152, Kraków, Poland.

THE TUNABLE LASER LIGHT SOURCE C-WAVE. HÜBNER Photonics Coherence Matters.

Transcription:

FIRST DATA FROM THE ATLAS INNER DETECTOR FSI ALIGNMENT SYSTEM S. M. Gibson, P. A. Coe, M. Dehchar, J. Fopma, D. F. Howell, R. B. Nickerson and G. Viehhauser University of Oxford, Denys Wilkinson Building, Keble Road, Oxford, OX1 3RH, UK Abstract As installation of the Large Hadron Collider nears completion in 2008, the focus shifts to commissioning and alignment of the particle detectors. The ATLAS inner detector is equipped with an optical alignment system that monitors micron level motion of the tracker structure in real-time. We present first data from the new Frequency Scanning Interferometry system which has been installed at CERN during 2007. Hundreds of simultaneous, remote distance measurements are combined to reconstruct changes in the tracker shape. We present a novel design, multiple fibre-coupled, Super-Invar reference interferometer and extended analysis techniques. These deliver shortened measurement times and improved system precision. between nodes attached to the support structure, as shown in part in Figure 2. The grid line lengths are measured simultaneously using Frequency Scanning Interferometry (FSI) to a precision of < 1 µm. [2, 3, 4, 5] INTRODUCTION During the past two years the Inner Detector has been installed at the heart of the ATLAS experiment [1], as in Figure 1, in readiness for first collisions at the LHC at CERN later in 2008. As part of the detector commissioning, it is Figure 1: The Inner Detector (Barrel SCT+TRT) during insertion into the heart of the ATLAS experiment. critical to determine the silicon strip positions within the Inner Detector to < 10 µm, in order not to degrade the tracking performance [1]. Since conventional surveys are impossible inside an operational particle tracker, an automated, remotely measured optical alignment system has been developed to monitor real-time shape deformations that are inaccessible with tracks. This system consists of a geodetic grid of 842 fibre-coupled interferometers, formed Work supported by Science and Technology Facilities Council, UK. s.gibson1@physics.ox.ac.uk Figure 2: The FSI grid nodes installed within a support cylinder of the SemiConductor Tracker. Each grid line is an interferometric length measurement precise to < 1 µm. Further to previous IWAA reports [6, 7, 8, 9], this paper focuses on recent developments following the installation of the FSI alignment system at CERN during 2007. After a brief review of the FSI principle, special emphasis is given to the novel design of the evacuated Reference Interferometry System, on which the precise measurements are based. Several improvements to the FSI technique are presented, together with a preliminary analysis of the first commissioning data from the new system. An update is also provided on the light distribution and read-out system necessary for the many hundreds of simultaneous precise, distance measurements. FREQUENCY SCANNING INTERFEROMETRY FSI is a technique for remote, multiple, simultaneous and precise distance measurements. A narrow line-width tunable laser simultaneously illuminates multiple interferometers to be measured and a reference interferometer. As the optical frequency is scanned, a phase shift is induced in all interferometers, at a rate that is proportional to the length of each interferometer. The phase shifts in the interferometers are compared to determine the ratio of interferometer lengths. FR002 1

Figure 3: The FSI laser room at CERN houses the amplified two-colour laser, diagnostic optics and the evacuated reference interferometer system, with associated control and read-out electronics. In the ATLAS FSI system, the Reference Interferometer (RI), is located away from the particle detector in a thermally stabilised room and inside an evacuated chamber. The lengths of the 842 grid line interferometers (GLIs) inside ATLAS are measured with respect to RI, which acts as a reference length. The optical path length in each GLI can be computed by comparing its phase shift to that in the RI. The stability of the reference length is critical to the long term repeatability of the FSI measurements. In practice, any drift in the optical path difference of an interferometer, induces a large error on the phase ratio estimate, for small frequency scans. The error can be greatly reduced by tuning the frequency of two lasers in opposite directions, so that the drift errors cancel to first order. The precision can be further improved by extending the interval of the frequency by linking the phase of several fine tuning subscans, interspersed by periods of rapid coarse tuning. The phase shift between subscans is determined using a pair of Vernier Etlaons. An auxiliary short reference interferometer is used to verify the linking procedure. The details of obtaining precise length measurements using the FSI technique can be found elsewhere [2, 5]. Overview THE INSTALLED FSI SYSTEM The FSI grids within the ATLAS SCT are illuminated from a common laser source housed in remote surface room, shown in Figure 3. The laser source is an amplified two-colour laser system [9], providing up to 600 mw of tunable, high spectral characteristic light, which is distributed via a fibre-splitter-tree to power the on-detector grid. A small fraction of the optical power is diverted into the diagnostic optics and Reference Interferometry System (RIS), which precisely monitor the frequency tuning of the laser during a scan. The RIS and lasers are mounted on pneumatically damped optical tables to isolate from extraneous vibration. The room is air conditioned to thermally stabilize the reference interferometer and to provide humidity control to prevent condensation on the laser diodes, which are cooled by peltier elements (TECs) to a few degrees below the ambient room temperature. The laser control and RIS read-out electronics are synchronised via a 250 m optical link to the multi-channel read-out system which is in the underground counting room adjacent to the ATLAS detector. The multiplexed read-out system uses APDs to register pw return signals from the 842 ondetector grid line interferometers. Evacuated Reference Interferometry System The Reference Interferometry System (RIS) consists of a thermally insulated, evacuated chamber containing two fibre-coupled Michelson inteferometers and a pair of vernier etalons. The RIS provides the essential frequency measurement of the tunable lasers and the reference length against which all SCT grid line interferometers are compared. The stability of the optical path length is paramount to the repeatability and precision of the FSI measurements, FR002 2

and is designed to remain stable to 10 7 K 1 over several years. in length of each interferometer arm, so that the physical path difference between the two arms remains constant, T (C 1 L 1 C 2 L 2 ) = 0. If this requirement is met, the measured optical path difference is to first order, independent of temperature changes, for a system in vacuum. Figure 4: An evacuated chamber contains the Reference Interferometry System, which is read-out by fibre-ribbons routed through custom vacuum feedthroughs to an external array of photodiodes. The evacuated chamber is shown in Figure 4 and was pumped down in situ to an order of magnitude below the 1 mbar vacuum pressure required. The purpose of the low vacuum is threefold: To reduce the systematic errors caused by refractive index differences between the gas in the RIS and the dry N 2 environment of the ATLAS SCT. To eliminate systematic drift errors caused by refractive index changes within the RIS during an FSI scan. Evacuation of the gas removes this source of error. To thermally isolate the RIS from external temperature variations. The vacuum decouples the RIS from heat exchange by conduction and convection currents with the surrounding environment. Any thermal expansion which changes the interferometer physical path difference during a scan can create systematic drift errors. Changes in length of the RI on longer timescales can affect the overall scale and repeatibility of the measurements, which could lead to errors in the FSI grid reconstruction. The RIS was therefore designed to combat these effects using a combination of methods as follows. Both RIs are designed to be extremely thermally stable. As in Figure 5, the RIS is built onto a frame of 6 Super- Invar rods, salvaged from a pair of old Coherent Innovar 90 lasers. Those particular Super-Invar rods were chosen for their documented, ultra low coefficient of thermal expansion (CTE) of 350 ppb K 1. A thermal compensation scheme was developed to use short lengths of materials with larger CTEs to counter the expansion of the long rods. The scheme aims to minimise thermally induced changes in the physical path difference between the two interferometer arms. This requires matching the ratio of lengths of the interferometer arms to the inverse ratio of average CTE of each arm: C 1 /C 2 = L 2 /L 1. A change in temperature, T should then produce the same physical change Figure 5: The Reference Interferometry System is built on a Super-Invar frame with a CTE compensation scheme to reduce thermal expansion. Wherever possible in the layout, care was taken to localise the compensation scheme to reduce unbalanced expansion errors caused by temperature gradients between interferometer components. Achieving this requirement was aided by the Michelson layout, for which the expansion of the short arm counters the expansion of the long arm. The layout also minimises the total amount of higher CTE material required for CTE ratio:length ratio balancing. Furthermore, the frame is supported on a kinematic mount that uses a knife-edge and ball bearings to reduce stiction and therefore decouples the frame from thermal expansion of the evacuated chamber. The contact points are also minimal and use thermally insulating plastics to slow down heat exchange by conduction. The RIS is illuminated via a single mode fibres to provide circular, low M 2 beams, which do not suffer from the higher order modes that could be present with a free space beam input. Both RIs have 6 fibres for read-out, 4 of which are used in the new phase extraction method described below. The long RI also has a piezo mounted mirror for phase stepping, if required. The RIS is instrumented with precise PT100 platinum resistance thermometers. Vernier Etalons The RIS contains a pair of vernier Fabry-Pérot etalons to determine the frequency interval between subscans. An etalon produces a comb of peaks as the frequency is scanned. The two etalons have carefully chosen Free Spectral Ranges that produce a beating pattern over a large frequency range. The FSR choice is limited by the phase resolution of the reference interferometer and peak fitting errors. In previous studies, the phase resolution was estimated to be well below 50 MHz [4]. Therefore a pair of etlaons with FSR 1 = 10.00 GHz and FSR 2 = 10.05 GHz FR002 3

Mirror Long arm Slightly tilted mirror Short arm Light launched from fibre collimator held in a 4-axis manipulator Four-fibres sample interferometer phase across incoming wavefront Non-polarising beam-splitter Figure 6: The reference interferometers use four parallel fibres to directly monitor the interferometer phase. was chosen to provide a beating pattern that can be resolved unambiguously over a cycle of 2010 GHz (Repeat cycle = N 2 FSR 1 = N 1 FSR 2 ). This choice is a trade off between maximising the beat cycle interval and minimising the necessary frequency interval that must be sampled to obtain a peak from each Etlaon (in this case, 10.05 GHz). The beat cycle of 2010 GHz is smaller than the 10 THz coarse tuning range of the lasers, but is sufficient to be easily resolved using a wavemeter. The FSR is small enough to permit corrections within a subscan in the event of a laser mode hop during fine tuning. IMPROVEMENTS TO THE TECHNIQUE Reference Phase Measurement A Frequency Scanning Interferometry measurement relies on precise information about the change in frequency of the laser during a scan. In practice it is not necessary to measure the absolute frequency. Rather is it sufficient to measure the change in phase of the fringes generated in a Reference Interferometer (RI), as the frequency is scanned. If the interferometer length, L, is fixed, then the phase change, Φ, is related to the frequency change, ν, by: Φ = (2π/c)L ν. A direct measure of the reference phase enables corrections of the non-linearities in frequency tuning of the laser with time. Four-fibre Phase Extraction In the prototype ATLAS FSI system, the reference interferometer phase was measured by stepping the position of a mirror mounted on a piezo-actuator. Four intensity measurements at equi-spaced mirror positions (ideally 108 apart in phase) were required for one phase measurement. The local phase value was extracted by a modified Carré algorithm and a series of measurements of the local phase were unwrapped to obtain a tuning curve for the lasers. [4, 5] The phase stepping works reliably, but is limited by the maximum driving frequency of the piezo and four acquisition cycles are required for each phase measurement. These limitations are overcome by use of a four-fibre read out system, which permit a direct measurement of the phase in one data acquisition cycle. The optical layout is shown in Figure 6, which uses four equi-spaced parallel fibres to sample the wavefront across the interferometer beam. By slightly Phase extraction and unwrapping Laser 2 Laser 1 Figure 7: The simultaneous intensity signals from four parallel fibres are used to extract the interferometer phase. FR002 4

Dual Interferometer Phase Extraction A new feature of the installed RIS is that both interferometers have four fibre read-out for direct phase extraction, which was not the case in the prototype FSI system. First commissioning data from the RIS are shown in Figure 8. In this test a single laser simultaneously illuminated both interferometers as the frequency was scanned over 190 GHz. After fringe normalisation, the four-fibre technique is used to extract the phase. The residuals from the best line fit show that both RIs successfully extract the same non-linear frequency tuning curve of the laser. Long Reference Short Reference 0.215536 D/L = 0.2155274 +/- 0.000003 0.215532 Phase ratio tilting the mirror, the phase difference across the four fibres can be adjusted to closely match the 108 required for optimal phase extraction by the modified Carre algorithm. An example of the normalised sinusoidal interferometer fringes sampled by four fibres is shown in Figure 7. These signals were used to extract the phase for two lasers, which were tuning in opposite directions in frequency. The key advantage of the method is that faster scans are possible, which reduces interferometer drift errors and therefore improves the measurement precision. Short intereferometer phase The 10th International Workshop on Accelerator Alignment, KEK, Tsukuba, 11-15 February 2008 0.215528 0.215524 0.215520 0.215516 24 Long inteferometer phase 29 34 39 Subscan / a.u.. Figure 9: The measured length ratio is determined directly from the gradiant of the phase changes extracted from two reference interferometers. for 15 partial subscans to given a preliminary result of, D/L = 0.2155274 ± 0.000003. This is equivalent to a 3 µm measurement precision on the SI length, which is reasonable for a single laser. The resolution is currently limited by several mode hops in the laser data, which restrict the useful frequency interval to only 34 GHz of the 190 GHz subscan. This is thought to be due to a misalignment of the internal laser cavity, after recent shipping. When the laser is realigned and commissioned, the available 190 GHz will improve on this result. Ultimately a reference measurement precision of order a few nanometers is expected by using two laser drift correction technique with multiple subscan linking over a 10 THz coarse tuning range, as demonstrated previously [5]. (a) LIGHT DISTRIBUTION AND READ OUT Fibre Splitter Tree Installation (b) (c) Figure 8: Simultaneous phase extraction from two reference interferometers as a common laser source is frequency scanned. The plots show for each interferometer: (a) the interference pattern recorded by one read-out fibre vs time, (b) the phase extracted from the four fibre measurements vs time, and (c) the residuals from a straight line fit to (b). Importantly the non-linear frequency tuning curve is extracted identically in both interferometers. The fibre coupled light from the lasers must be split between the 842 delivery fibres of the on-detector grid line interferometers (GLIs). A fibre-splitter-tree based on biconic, fused taper couplers was originally envisaged [9]. However a more compact and cost effective solution was identified in the form of Planar Lightwave Cicuits that were mode matched to the required 830nm SM fibre. This technology permits high multiplicity coupling ratios of 1x8, 1x16 or even 1x32 to be created using ion-exchange in glass, in devices of a similar size to a 1x2 bi-conic coupler and with better overall insertion losses. The devices were assembled in to 15 splitter tree modules, one of which is shown in Figure 10. Each 1U module distributes light to upto 64 GLIs via a matrix of interwoven fibres and also conveys the pw return signals to read-out system. The fibre matrix is necessary to deal with the many constraints imposed by optical multiplexing, connectivity and the grid topology. Preliminary Result with Commissioning Data Dual interferometer phase extraction allows the length ratio of the short and long reference interferometers to be determined directly. The phase change in the short RI, Θ = (2π/c)D ν, and the long RI, Φ = (2π/c)L ν, D are compared to give the phase ratio, Θ Φ = L. Thus D the gradient, L in a plot of Figure 9 is the length ratio of the RIs. The length ratio was calculated repeatedly FR002 Read-out System Commissioning The 15 splitter tree modules are assigned to 15 FSI readout cards (FROC). The FROCs use 32 Avalanche Photodiodes to register pw signals from upto 64 multiplexed, optically switched interferometers, as detailed elsewhere [9]. Commissioning tests have shown that after installation, the pedestal data transfer rate is acceptable, as in Figure 11. 5

Figure 10: One of 15 fibre-splitter-tree modules that convey light from the laser to the 842 SCT fibre-coupled grid line interferometers and route the signals to the read-out cards. REFERENCES Figure 11: The 15 FSI Read Out Cards (FROCs) and the measured transfer rate of commissioning data. CONCLUSION The ATLAS Inner Detector FSI alignment system is in place at CERN and the commissioning phase has started. Several improvements have been made to the installed system that were not present in the prototype. A four-fibre phase extraction technique has been developed to improve the precision and the thermally stable Super-Invar dual reference interferometers provide simultaneous phase extraction. First data from the installed reference interferometry system indicate improved performance is possible using these new analysis techniques and extended frequency tuning capabilities of the new lasers. The multi-channel read-out system and fibre-splitter-tree are now being commissioned. We look forward to first measurements of the ATLAS FSI geodetic grid in the coming months, which will reveal the alignment stability of particle tracker. ACKNOWLEDGEMENTS We are grateful for the invaluable, excellent support of technical staff from Oxford Physics Central Electronics and Mechanical Groups. [1] ATLAS Collaboration, The ATLAS Experiment at the CERN Large Hadron Collider JINST, submitted Dec. 2007. [2] P. A. Coe, D. F. Howell, R.B. Nickerson, Frequency scanning interferometry in ATLAS: remote, multiple, simultaneous and precise distance measurements in a hostile environment, Meas. Sci. Technol. 2004; 15(11): 2175-2187. [3] S. M. Gibson, P. A. Coe, A. Mitra, D. F. Howell, R. B. Nickerson, Coordinate Measurement in 2-D and 3-D Geometries Using Frequency Scanning Interferometry, Optics and Lasers in Engineering, 2005; 43(7): 815-813. [4] P.A. Coe, An investigation of Frequency Scanning Interferometry for the alignment of the ATLAS semiconductor tracker, D.Phil. Thesis, Oxford, 2001. CERN-THESIS- 2007-082. [5] S. M. Gibson, The ATLAS SCT alignment system and a comparative study of misalignment at CDF and ATLAS, D.Phil. Thesis, Oxford, 2004. CERN-THESIS-2007-080. [6] P. A. Coe, A. Mitra, S. M. Gibson, D. F. Howell, R. B. Nickerson, Frequency Scanning Interferometry - A versatile high precision, multiple distance measurement technique, 7 th International Workshop on Accelerator Alignment, SPRING8, Japan, Nov. 2002. [7] S. M. Gibson, P. A. Coe, A. Mitra, D. F. Howell, R. B. Nickerson, A study of Geodetic Grids for the Continuous, Quasi Real Time Alignment of the ATLAS Semiconductor Tracker, 7 th International Workshop on Accelerator Alignment, SPRING8, Japan, Nov. 2002. [8] S.M. Gibson et al, Monitoring the heart of ATLAS using Frequency Scanning Interferometry, 8 th International Workshop on Accelerator Alignment, CERN, Geneva, Oct. 2004. [9] S.M. Gibson et al, The Multi-channel High Precision AT- LAS SCT Alignment Monitoring System: A Progress Report, 9 th International Workshop on Accelerator Alignment, SLAC, Menlo Park, California. IWAA-2006-WEPO08. Sept. 2008. FR002 6