CAEN Tools for Discovery

Similar documents
Traditional analog QDC chain and Digital Pulse Processing [1]

INDEX. Firmware for DPP (Digital Pulse Processing) DPP-PSD Digital Pulse Processing for Pulse Shape Discrimination

CAEN. Electronic Instrumentation. CAEN Silicon Photomultiplier Kit

nanomca 80 MHz HIGH PERFORMANCE, LOW POWER DIGITAL MCA Model Numbers: NM0530 and NM0530Z

Amptek Silicon Drift Diode (SDD) at High Count Rates

CAEN. Electronic Instrumentation. UM2088 DPHA User Manual. Rev June Rev June Tools for Discovery

nanodpp datasheet I. FEATURES

Summary. Introduction

nanomca datasheet I. FEATURES

AN-DPP-003 Rev A2: Using the DP5 with HPGe USING THE DP5 WITH GERMANIUM DETECTORS

nanomca-sp datasheet I. FEATURES

Amptek Inc. Page 1 of 7

nanomca-ii-sp datasheet

A digital method for separation and reconstruction of pile-up events in germanium detectors. Abstract

Gamma Ray Spectroscopy with NaI(Tl) and HPGe Detectors

CAEN. Electronic Instrumentation DPP-PSD. Rev July Digital Pulse Processing for Pulse Shape Discrimination. User Manual UM2580

ORTEC. Research Applications. Pulse-Height, Charge, or Energy Spectroscopy. Detectors. Processing Electronics

K 223 Angular Correlation

Nyquist filter FIFO. Amplifier. Impedance matching. 40 MHz sampling ADC. DACs for gain and offset FPGA. clock distribution (not yet implemented)

Figure 1: Schematic diagram of Analog Pulse Processing Architecture. Figure 2: Schematic diagram of Digital Pulse Processing (DPP) Architecture

Keyser, Ronald M., Twomey, Timothy R., and Bingham, Russell D. ORTEC, 801 South Illinois Avenue, Oak Ridge, TN 37831s

Electronic Instrumentation for Radiation Detection Systems

Fast first practical help -- detailed instructions will follow- preliminary Experiment F80

The 2017 IEEE NSS-MIC. Industrial Presentation

PX4 Frequently Asked Questions (FAQ)

Real-Time Digital Signal Processors with radiation detectors produced by TechnoAP

TB-5 User Manual. Products for Your Imagination

Chapter 6 Pulse Processing

A NOVEL FPGA-BASED DIGITAL APPROACH TO NEUTRON/ -RAY PULSE ACQUISITION AND DISCRIMINATION IN SCINTILLATORS

Digital Signal Processing for HPGe Detectors

Week 11: Chap. 16b Pulse Shaping

GAMMA-GAMMA CORRELATION Latest Revision: August 21, 2007

NIM INDEX. Attenuators. ADCs (Peak Sensing) Discriminators. Translators Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy)

Real Time Pulse Pile-up Recovery in a High Throughput Digital Pulse Processor

Considerations on the ICARUS read-out and on data compression

Preliminary simulation study of the front-end electronics for the central detector PMTs

NIM. ADCs (Peak Sensing) Analog Pulse Processors Amplifiers (Fast) Amplifiers (Spectroscopy) Attenuators Coincidence/Logic/Trigger Units

ARTICLE IN PRESS. Nuclear Instruments and Methods in Physics Research A

PX5 User Manual and Operating Instructions

DSA-LX. Digital Signal Analyzer. Radiation Safety. Amplified.

GAMMA-RAD5 User Manual

Comparisons Between Digital Gamma-Ray Spectrometer (DSPec) and Standard Nuclear Instrumentation Methods (NIM) Systems

COMPARISON OF A DIGITAL AND AN ANALOGICAL GAMMA SPECTROMETER AT LOW COUNT RATES

CAMAC products. CAEN Short Form Catalog Function Model Description Page

COMPENDIUM OF FRONT-END ELECTRONICS

Silicon Photomultiplier Evaluation Kit. Quick Start Guide. Eval Kit SiPM. KETEK GmbH. Hofer Str Munich Germany.

A MONTE CARLO CODE FOR SIMULATION OF PULSE PILE-UP SPECTRAL DISTORTION IN PULSE-HEIGHT MEASUREMENT

DP5 User Manual and Operating Instructions

Simulation of Algorithms for Pulse Timing in FPGAs

A Readout ASIC for CZT Detectors

A novel acquisition method of nuclear spectrum based on pulse area analysis *

User's Manual Digital Gamma Finder (DGF) Pixie-4

ORTEC Experiment 13. Gamma-Gamma Coincidence with Angular Correlation. Equipment Required

Positron Emission Tomography

Testing of the NSC Electronics Module with the GSI Clover Detector

ORTEC. Time-to-Amplitude Converters and Time Calibrator. Choosing the Right TAC. Timing with TACs

Virtual Laboratory of Nuclear Fission Virtual practicum in the framework of the project Virtual Laboratory of Nuclear Fission

Energy Measurements with a Si Surface Barrier Detector and a 5.5-MeV 241 Am α Source

Beam Condition Monitors and a Luminometer Based on Diamond Sensors

ORTEC. AN34 Experiment 14 Nuclear Lifetimes and the Coincidence Method. Equipment Needed from ORTEC. Equipment Required from Other Manufacturers

Pulse Shape Analysis for a New Pixel Readout Chip

Digital coincidence acquisition applied to portable β liquid scintillation counting device

Instructions for gg Coincidence with 22 Na. Overview of the Experiment

XRF Instrumentation. Introduction to spectrometer

ORTEC. High-Count-Rate Spectroscopy with Ge Detectors: Quantitative Evaluation of the Performance of High-Rate Systems 1. I.

A Modular Readout System For A Small Liquid Argon TPC Carl Bromberg, Dan Edmunds Michigan State University

Verification of a novel calorimeter concept for studies of charmonium states Guliyev, Elmaddin

Digital Pulse Processing in HPGe Gamma-ray Spectroscopy

Mass Spectrometry and the Modern Digitizer

--- preliminary Experiment F80

Dual 500ns ADC User Manual

Data Compression and Analysis Methods for High- Throughput Radiation Detector Systems

A user-friendly fully digital TDPAC-spectrometer

MuLan Experiment Progress Report

ORTEC Experiment 1. Introduction to Electronic Signal Analysis in Nuclear Radiation Measurements. Equipment Required: Purpose. Electronic Circuits

Digital trigger system for the RED-100 detector based on the unit in VME standard

PARTICLE DETECTORS (V): ELECTRONICS

How different FPGA firmware options enable digitizer platforms to address and facilitate multiple applications

Gamma Spectrometer Initial Project Proposal

Model 673 Spectroscopy Amplifier and Gated Integrator Operating and Service Manual

Purpose This experiment will use the coincidence method for time correlation to measure the lifetime in the decay scheme of 57

Photon Counters SR430 5 ns multichannel scaler/averager

Experiment No. 3 Pre-Lab Phase Locked Loops and Frequency Modulation

Analysis and performance Evaluation of trapezoidal filter in pulse shaping

Digital Pulse Processing: A New Paradigm For Nuclear Instrumentation

Shape-based event pileup separation in Troitsk nu-mass experiment

Development of New Peak Detection method for Nuclear Spectroscopy

A PC104 Multiprocessor DSP System for Radiation Spectroscopy Applications

Power electronics CAD: from space applications to industrial applications

APPLICATION NOTE 3671 Data Slicing Techniques for UHF ASK Receivers

Moderne Teilchendetektoren - Theorie und Praxis 2. Dr. Bernhard Ketzer Technische Universität München SS 2013

SPECTROMETRIC CHARACTERISTIC IMPROVEMENT OF CdTe DETECTORS*

Overview 256 channel Silicon Photomultiplier large area using matrix readout system The SensL Matrix detector () is the largest area, highest channel

Development of Personal Dosimeter Using Electronic Dose Conversion Method

Final Results from the APV25 Production Wafer Testing

Technical Information Manual

User Manual WP2081. Digital Pulse Processing in Nuclear Physics. Overview of CAEN DPP algorithms

A high-performance, low-cost, leading edge discriminator

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

Improvement of the Offline Event Reconstruction for the GERDA Experiment

Transcription:

Viareggio 5 September 211 Introduction In recent years CAEN has developed a complete family of digitizers that consists of several models differing in sampling frequency, resolution, form factor and other features. Besides the use of the digitizers as waveform recorders (oscilloscope mode), CAEN offers the possibility to upload special versions of the FPGA firmware that implement algorithms for the Digital Pulse Processing (DPP); when the digitizer runs in DPP mode, it becomes a new instrument that represents a complete digital replacement of most traditional modules such as Multi Channel Analyzers, QDCs, TDCs, Discriminators and many others. In this application note, we describe the capability of the series x724 (14 bit, MSps) to perform Pulse Height Analysis in. The development of this FPGA firmware was based upon digital trapezoidal filters applied to the digitized signals output by a Charge Sensitive Preamplifier. Traditional Analog Approach The traditional electronics for rely upon three fundamental devices: the Charge Sensitive Preamplifier, the Shaping Amplifier and the Peak Sensing ADC. IN (Detector) Charge Sensitive Preamplifier OUT Fig. 1: Charge Sensitive Preamplifier converts the area of the input pulse (charge) to the amplitude of the output. Usually [1] [2], the result of a particle interaction within the detector s sensitive volume is the excitation of the absorber medium, e.g. scintillators, or the release of an observable burst of charge proportional to the energy lost by the particle in the interaction, e.g. semiconductors. In some cases, the value of this charge is sufficient to be managed by the front end electronics, but in many applications, typically where a semiconductor detector is required, a preamplification stage is mandatory. In order to minimize the noise, it is wise to amplify the signal as closest to the detector as possible, and sometimes to insert the very first stage of the preamplifier in the detector s architecture, as happens in HPGe detectors. The Charge Sensitive Preamplifier (Fig. 1) integrates the signal coming from the detector, thus converting the collected charge into a voltage step. Ideally, it is just a simple capacitor; however, in order to avoid saturation, the integrating capacitor is put in parallel with a discharging resistor, so that the preamplifier output will have pulses with a fast rise time and a long exponential tail with decay time τ. The charge information (proportional to the energy released by the particle in the detector) is here represented by the pulse height. Q The charge amplitude proportionality is set by the capacitor value V out = and the decay time of the output signal is τ = RC. C In order to have a good charge amplitude conversion and to minimize the noise, the decay time τ is much larger than the width of the detector signal, typically 5 µs, and for this reason pile up of different particle detections can arise (see Fig. 2). Detector Charge Sensitive Preamplifier Fig. 2: Pile up of detector signals due to the large decay time of the Preamplifier output. Another issue with the output signal of the Charge Sensitive Preamplifier is that the peak is too sharp for the Peak Sensing ADC to be detected with the required precision. In order to avoid these problems in a traditional analog acquisition chain a Shaping Amplifier is requested. This amplifier (see Fig. 3) receives the signal from the Preamplifier output and provides a quasi Gaussian output whose width can be changed selecting different shaping time. The height is still proportional to the energy released by the detected particle. 1

RISE TIME DECAY TIME Preamplifier Shaping Amplifier PEAK AMPLITUDE ~ ENERGY Fig. 3: Shaping Amplifier converts the long tailed Preamplifier output in a quasi Gaussian signal preserving the proportionality between energy and peak amplitude. In this way it s possible to reduce the pile up and feed the Peak Sensing ADC with a smooth signal. Finally, the Peak Sensing ADC is capable to evaluate and digitize the height of the pulses output by the Shaping Amplifier, filling a histogram with these values, i.e. an energy spectrum. CAEN Digital Approach CAEN Digital Pulse Height Analyzer is a self consistent system composed by a digitizer (also know as Flash ADC) of the 724 series ( MS/s, 14 bit) loaded with DPP PHA firmware and managed by the DPP PHA Control Software. In this system the digitizer replaces both the Shaping Amplifier and the Peak Sensing ADC; in fact, the digitizer samples the pulses output by the Charge Sensitive Preamplifier and converts them into a continuous data stream. The pulse shaping is done in digital by means of a trapezoidal filter running online on the digitizer FPGA. The trapezoidal filter [3], also known as moving window deconvolution, can be shortly described as a filter able to transform the typical long tailed exponential signal generated by a Charge Sensitive Preamplifier into a trapezoid whose height is proportional to the amplitude of the input pulse that is to the energy released by the particle in the detector (Fig. 4). It is important to highlight that this trapezoid filter plays more or less the same role of the Shaping Amplifier in a traditional analog acquisition system; for instance, both have a shaping time constant: setting the parameters of the trapezoidal filter is like operating on the potentiometers of the Shaping Amplifier. The control of the Analyzer is managed by DPP PHA Control Software, that allows the user to set the parameters for the acquisition, configure the hardware and perform the data readout, the histogram collection and the spectrum or waveform plotting and saving. The histograms saved in the output files can be easily managed by third part software tools for spectroscopy analysis. Fig. 4: Trapezoidal filter with the relevant parameters. Once the DPP PHA has performed the trapezoidal filtering, the energy of the detected particle can be calculated as the height of the trapezoid in respect to its baseline; this value can be finally saved to the digitizer memory. The original raw data coming from the ADC, i.e. the samples, can be discarded in order to minimize the data throughput from the board to the computer, even though it might be useful, in some cases, to save also a piece of the waveform for further analysis or for monitoring the signals. It is worth noticing that the DPP PHA firmware is also able to calculate and save the time stamp of the input pulses. 2

Measurements All detector and gamma source based tests were performed at the Institute for Nuclear Science and Technology (CEA, Saclay) in collaboration with Mr. Georges Meyer and Mr. Bernard Rousse. The data analysis was performed using ROOT and Visu Gamma. Set up description The measurements shown in this Application Note were performed using an HPGe detector, mod. Ortec GEM 175P, whose preamplified output fed a CAEN DT5724 with DPP PHA firmware. This board is the 4 channel, Desktop version of 724 digitizer series; a dynamic range of 5 mvpp was used. 137 Cs and 6 Co sources were placed close to the detector according to the different measures; different counting rates were reached modifying the distance between source and detector. Trapezoid Rise Time & Input Counting Rate The first set of measures was performed in order to test the system capabilities, mainly Energy Resolution and Peak Shift, as a function of the Input Counting Rate and Trapezoid Rise Time. This parameter is the time needed by the trapezoid to reach its maximum, i.e. flat top, from the baseline. It plays the same role as the shaping time of a analog Shaping Amplifier: for a quick comparison, a value of Trapezoid Rise Time of 1 µs is equivalent in performances to.45 µs of analog Shaping Time. As in the analog case, this value should be greater that the collection time of the detector, but it can be increased without rising the parallel noise being the result of a digital computation; in HPGe detectors the value of this parameter is usually in the range of 1 2 µs. In order to analyze how the system s performances change as a function of Input Counting Rate and Trapezoid Rise Time, the 137 Cs energy spectrum was acquired setting the Rise Time to 1 µs, 5 µs and 9 µs for each rate value. The rate was raised up to kcps. In Fig. 5 the energy spectrum collected with Rise Time = 9 µs and ICR = 1.1 kcps is shown; a resolution of 1.6 kev was obtained for the 661.7 kev photopeak. 137 Cs 18 16 14 12 18 16 14 12 2 / ndf 8.657 / Constant 1766 23.1 Mean 661.5. Sigma.6821.56 8 6 4 8 6 4 2 658 66 662 664 666 668 2 2 3 4 5 6 7 Fig. 5: 137 Cs spectrum obtained with Trapezoid Rise Time = 9 µs, ICR 1.1 kcps and Zoom of the 661.7 kev photopeak. In Fig. 6 the energy resolution of the 661.7 kev photopeak as a function of the ICR is shown for each value of Rise Time. As expected, the higher the value of Rise Time the better the resolution. Nevertheless, rising the ICR, it is more and more probable the pulse pile up, so it is preferable to have shorter shaped signals not to loose resolution significantly and also to reduce the pileup rejection. 3

Resolution vs ICR Resolution (kev) of 137Cs photopeak 4.5 4. 3.5 3. 2.5 2. 1.5 1..5. ICR (cps) RT 1us RT 5us RT 9us Fig. 6: Resolution of the 661.7 kev 137 Cs photopeak as a function of the Input Counting Rate (ICR) for different values of Trapezoid Rise Time (RT). In Fig. 7 it is shown the Peak shift as a function of the ICR for each value of Rise Time; even at 9 µs of Rise Time there is no notable shift of the photopeak up to 3 kcps. Peak Shift vs. Count Rate (ICR) Peak Shift (%) of 137Cs photopeak.6.5.4.3.2.1 -.1 -.2 Baseline Restorer ICR (cps) RT 1us RT 5us RT 9us Fig. 7: Peak position of the 661.7 kev 137 Cs photopeak as a function of the Input Counting Rate (ICR) for different values of Trapezoid Rise Time (RT). In order to correctly evaluate the trapezoid height it is important to carefully estimate its baseline. The baseline, in fact, can fluctuate for a number of reasons as microphone noise, grounding, power supply, etc. The Digital Pulse Height Analyzer can evaluate the baseline level by means of a moving average window whose length can be selected by software. The main effect of a not well compensated baseline fluctuation is clearly noticeable in the peaks collected in an energy spectrum; the base of the peak is larger than what would be expected by a pure Gaussian peak, as shown in Fig. 8. 137 Cs bsl 16 bsl 256 bsl 248 8 6 4 2 652 654 656 658 66 662 664 666 668 67 672 Fig. 8: 137 Cs photopeak collected changing the Baseline parameter. If the baseline fluctuations are not carefully compensated, the base of the peak is larger than the expected. The histograms were drawn with a smoothed line in order to highlight their shape. 4

Advanced Settings: Peaking Holdoff To avoid the loss of energy resolution when the Input Counting Rate increases, DPP PHA allows the user to introduce a Peaking Holdoff. This parameter inhibits the height analysis of pulses closer than a programmable time value. In principle, two trapezoids can be considered valid, i.e. it is possible to correctly calculate their heights, even if the second pulse overlaps the falling edge of the first one. Anyway, because of the overlap, the height of the second trapezoid is evaluated in respect to the baseline value calculated before the first one. This means that the baseline could have been changed in the meanwhile and therefore a loss of energy resolution can be introduced. Peaking Holdoff can be used in order to store the heights of only wellseparated pulses, ensuring a precise baseline calculation for each trapezoid. Of course, the drawback of this technique is to increase the pulse rejection: this parameter sets the minimum separation of the trapezoids before they are considered piled up. In order to evaluate the effectiveness of the Peaking Holdoff, several spectra of the 137 Cs source were acquired, changing the holdoff width. The results are shown in Fig. 9, Fig. and Fig. 11. It can be noticed that the area of the histograms, i.e. total counts, decreases for higher values of Peaking Holdoff because of the pulse rejection. 137 Cs 4 3 PKHO us PKHO 5 us PKHO 3 us PKHO 1.2 us PKHO.4 us 2 1 2 4 6 8 12 14 Fig. 9: 137 Cs spectra acquired with different Peaking Holdoff values. The measurements were performed with Rise Time set to 1 µs. The ICR was about 1 kcps. 137 Cs 4 PKHO us PKHO 5 us PKHO 3 us PKHO 1.2 us PKHO.4 us 3 2 62 64 66 68 7 72 Fig. : Zoom of the 661.7 kev photopeak of Fig. 9. The effect of the Peaking Holdoff is to let a precise evaluation of the trapezoids baseline. 5

Resolution vs Peaking Holdoff Resolution (kev) of 137Cs photopeak 2.8 2.75 2.7 2.65 2.6 2.55 2.5 2.45 2 4 6 8 Peaking Holdoff (us) Fig. 11: Resolution of the 661.7 kev 137 Cs photopeak as a function of the Peaking Holdoff. Dead Time & Pile up Rejection Unlike the analogue Peak Sensing ADCs, CAEN Pulse Height Analyzer is not affected by conversion time, being based on Flash ADCs. This means that the ADCs housed by the board digitize continuously the analogue inputs making the samples always available to the DPP PHA algorithm running on the FPGA. Anyway, the system could be unable to correctly evaluate the height of some pulses for three reasons: dead time due to saturation of the input stage (ADC over range) dead time due to full memories in the digitizer piled up events Concerning the first two points, both the situations are related to a very high ICR, of the order of several Mcps. In fact, being CAEN Pulse Height Analyzer a conversion time less system, the only sources of dead time, i.e. inability of the system to manage the analogue inputs and to analyze the detector s pulses, are the saturation of the input dynamic range, due to an high event pile up, and the inability of the readout link to sustain the data throughput from the memories that contain the energy values to the computer. If the ICR is not so high, the system is never dead to the detector s pulses, but conversely it is always capable to analyze the digitized signals: the system is not affected by dead time, as traditionally defined. However, even at lower rates, the probability to have piled up events is high enough to cause the rejection of a significant number of events. In fact, pulses really close to each other produce an overlap of the related trapezoids. In this situation, the system is unable to disentangle the heights of the piled up pulses so, in order to not lose energy resolution and create false peaks in the spectra, it rejects these events as shown in Fig. 12. Fig. 12: The effect of trapezoids overlapping in three main cases: 1) The second trapezoid starts on the falling edge of the first one. 2) The second trapezoid starts on the rising edge ( T < peaking time) of the first one. 3) The second trapezoid starts on the rising edge of the first one ( T < input rise time). 6

In this case, in addition to an effective capability to recognize and reject the piled up events, it is necessary to know how many pulses have been rejected in order to precisely estimate the Input Counting Rate that is related to the activity of the radiation source and therefore correct the energy spectra. CAEN Pulse Height Analyzer is able both to recognize with high efficiency the piled up pulses and to count how may of them have been rejected; this capability is ensured by the digital implementation of a fast Trigger and Timing Filter. The aim of the Trigger and Timing Filter (TTF) is to identify the input pulses, generate a trigger on them and calculate the time stamp by means of a kind of Constant Fraction Discriminator. To make an analogy with an analog system, the TTF is like a RC CR 2 filter: the integrative component is a smoothing filter based on a moving window averaging filter that reduces the high frequency noise and prevent the trigger logic to generate false triggers on spikes or fast fluctuation of the signals. On the other hand, the purpose of the derivative component is to subtract the baseline, so that the trigger threshold is not affected by the low frequency fluctuation and, more important, by the pile up. As a result of the double derivation (CR 2 ), the output signal of the TTF is bipolar and the zero crossing is independent of the pulse amplitude. This is the same principle of the CFD. The trigger logic uses the threshold to get armed, then waits for the zero crossing to generate the trigger signal and produce a time stamp. Fig. 13: The Trigger and Timing Filter is used by the channels to autotrigger on the input pulses. In Fig. 14, the Output Counting Rate, that is the rate of the pulses whose height was evaluated by the system (i.e. not piled up pulses), versus the Input Counting Rate is shown. As expected, the longer is the Rise Time the sooner the maximum OCR is reached; anyway, an ICR up to kcps can be analyzed with no significant event rejection even with 9 µs of Trapezoid Rise Time. OCR vs ICR OCR (cps) RT 1us RT 5us RT 9us ICR (cps) Fig. 14: Output Counting Rate vs Input Counting Rate for different values of Trapezoid Rise Time. 7

Rise Time Discriminator When the Input Counting Rate is very high, the probability to observe two pulses output by the preamplifier piled up on their rising edge is not negligible. In this situation, the usual pile up rejection methods are ineffective, since the pile up occurs before the first pulse reaches its maximum: the two pulses are not rejected, producing false peaks in the spectrum, as shown in Fig. 12. Anyway, if the two pulses are not exactly simultaneous, there is the possibility to reject the undesired events by analysing the rise time of the piled up signal: CAEN DPP PHA can reject the pulses whose rise time exceed a settable value. In Fig. 15 two measures obtained with an ICR of 1 kcps are shown. In the first measure the Rise Time Discriminator was disabled, leading to the creation of a pile up peak whose energy is twice the 661.7 kev energy of 137 Cs gamma ray. In the second measure, the RTD wad enabled; as a result, the pile up events are strongly suppressed. 137 Cs 4 No RTD With RTD 3 2 1 2 4 6 8 12 14 Fig. 15: 137 Cs spectra with ICR 1 kcps. The black spectrum was collected without Rise Time Discriminator, the red one was obtained using RTD. The measurements was performed with Trapezoid Rise Time set to 1 µs. The value of RTD must be carefully set in order not to reject good pulses. In fact, especially in large detectors, it may happen that not piled up signals show a rise time longer respect to the usual ones, depending on the interaction point within the detector. Unsteady Sources & Live Spectrum Correction Thanks to the high efficiency in the input pulse counting, CAEN digital Pulse Height Analyzer is able to manage a live correction of the energy spectrum in respect to the piled up events. This is very important whenever an unsteady radiation sourced is used, e.g. short living isotopes or activity transients. In these situations, it is not possible to compensate the lost counts simply extending the acquisition time, because in this way the correction is applied only to the source still active in the additional time. What it is done in many traditional systems is to correct the spectrum only at the end of the acquisition run, redistributing the lost counts in respect to the collected histogram; this may lead to correction errors, since this method does not take into account the ICR variations during the data acquisition. As a proof of CAEN Pulse Height Analyzer capabilities to live correct the energy spectra, a measurement was performed acquiring a mid counting ( kcps) 6 Co source for 3 minutes. During this acquisition, an high counting ( kcps) 137 Cs source was added for just seconds. In this situation, the addition of the 137 Cs source produced a sudden transient in the ICR, increasing therefore the piled up events. Because of the higher activity of the 137 Cs source, it is more probable that the piled up pulses involved 137 Cs photons than 6 Co ones, so the correction should be more focused on the former source than the latter. If a traditional correction is applied, the simple redistribution of the lost counts at the end of the acquisition would compensate the lost counts in respect to the final spectrum shape that is, of course, related not only to the activity but also to the collection time. 8

137 6 Cs+ Co 2 18 16 14 b no correction live correction end run correction 12 8 6 4 2 137 6 Cs+ Co 2 18 16 14 12 a 645 65 655 66 665 67 675 68 no correction live correction end run correction 8 6 4 2 2 4 6 8 12 14 137 6 Cs+ Co 9 8 7 6 c no correction live correction end run correction 137 6 Cs+ Co 8 7 6 5 d no correction live correction end run correction 5 4 3 2 4 3 2 1155 116 1165 117 1175 118 1185 119 1315 132 1325 133 1335 134 1345 135 Fig. 16 (a): Energy spectrum acquiring a mid counting ( kcps ICR) 6 Co source for 3 minutes and adding an high counting ( kcps) 137 Cs source for seconds. The red histogram is the uncorrected spectrum, the blue one is the spectrum corrected at the end of the run and the black one is the live corrected spectrum. The measurement was performed with Trapezoid Rise Time set to 1 µs and RTD enabled. (b): Zoom of the 137 Cs photopeak. (c) (d): Zoom of the 6 Co photopeaks. As can be noticed, the traditional correction performed at the end of the run does not take in account the Input Counting Rate variation during the measurement, redistributing the lost counts in respect to the integrals of the 137 Cs and 6 Co spectra. This leads to an undercompensation of the 137 Cs photopeak and an overcompensation of the 6 Co ones. 9

References [1] G. Knoll, Radiation Detection and Measurement John Wiley & Sons, 4th Edition. [2] W.R. Leo, Techniques for Nuclear and Particle Physics Experiments: A How to Approach Springer Verlag, 2nd Revised Edition. [3] V. Jordanov, G. Knoll, Digital synthesis of pulse shapes in real time for high resolution radiation spectroscopy Nuclear Instruments and Methods in Physics Research A 345(2): 337 345. rev. 5 September 211 117 DGT22 ANXX Copyright CAEN SpA. All rights reserved. Information in this publication supersedes all earlier versions. Specifications subject to change without notice. CAEN CAEN SpA Via Vetraia 11 5549 Viareggio Italy Tel +39.584.388.398 Fax +39.584.388.959 info@caen.it www.caen.it www.caen.it