Measuring Spectral Performance of Night Vision Devices

Similar documents
Application Note (A13)

Technical Notes. Integrating Sphere Measurement Part II: Calibration. Introduction. Calibration

OL Series 426 Low Light Level lntegrating Sphere Calibration Standards

Application Note 1030

We bring quality to light. Spectro 320 Scanning Spectrometer

Gentec-EO USA. T-RAD-USB Users Manual. T-Rad-USB Operating Instructions /15/2010 Page 1 of 24

Spectroradiometer specbos 1201

TracQ. Basic Data Acquisition and Spectroscopy Software

LINEARPYROMETER LP4. Technical Documentation KE November TN

Broadbandradiometer specbos 1211

High-Accuracy Luminance & Chromaticity Measurement Comparable to Many Spectroradiometers

JETI Specbos Instruments

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

Product tags: VIS, Spectral Data, Color Temperature, CRI, Waterproof, WiFi, Luminous Color, LED, Photometry, General lighting

SPECTRORADIOMETRY METHODS: A GUIDE TO PHOTOMETRY AND VISIBLE SPECTRORADIOMETRY

BTS256-EF. Product tags: VIS, Spectral Measurement, Waterproof, WiFi. Gigahertz-Optik GmbH 1/7

Filter Design for AMLCD Full-Color Displays Compatible with Night Vision Devices

PHOTO RESEARCH, INC. PR-730 / PR-735 and PR-740 / PR-745 SpectraScan

CS-2000/2000A. Spectroradiometer NEW

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

Gamma Scientific Display Measurement Systems

Miniaturized Spectroradiometer

OL 730CV Virtual Radiometer/Photometer PC-Controlled Virtual Instrument

Noise Analysis of AHR Spectrometer Author: Andrew Xiang

Development of a spectrometry system Using lock-in amplification technique

Unprecedented Resolution and Accuracy For Camera & Image Sensor Calibration

Color Measurement with the LSS-100P

Achieving 100,000 : 1 contrast measurement

PICO MASTER 200. UV direct laser writer for maskless lithography

NFMS THEORY LIGHT AND COLOR MEASUREMENTS AND THE CCD-BASED GONIOPHOTOMETER. Presented by: January, 2015 S E E T H E D I F F E R E N C E

We bring quality to light. MAS 40 Mini-Array Spectrometer

LED Tester BTS256-LED

Technological Advances in General Lighting. New Lightmeter for Solid State Lighting. State-of-the-Art LED Illuminance Meter

BTS256-E WiFi - mobile light meter for photopic and scotopic illuminance, EVE factor, luminous color, color rendering index and luminous spectrum.

NEAR EYE DISPLAY (NED) SPECTRORADIOMETER SYSTEMS DATASHEET

DETECTORS Important characteristics: 1) Wavelength response 2) Quantum response how light is detected 3) Sensitivity 4) Frequency of response

BTS2048-UV. Product tags: UV, Spectral Data, LED Binning, Industrial Applications, LED.

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Test procedures Page: 1 of 5

2. Instrumentation Design, Specifications, and Installation of the SUV-100

TECHNICAL MEMORANDUM

SR-80 EXTENDED AREA BLACKBODY

Design of Efficient Filters for Full-Color Displays Used with Night Vision Devices.

New automated laser facility for detector calibrations

Photometry for Traffic Engineers...

Development of 2 Total Spectral Radiant Flux Standards at NIST

GENESYS 20 GENERAL PRACTICES AND INFORMATION

Photometry for Traffic Engineers...

metcon meteorologieconsultgmbh, Instruments for Atmospheric Research W1aa_Feb_2017_1.doc 1 -

University of Wisconsin Chemistry 524 Spectroscopic Components *

GCMS-3 GONIOSPECTROPHOTOMETER SYSTEM

Chemistry Instrumental Analysis Lecture 7. Chem 4631

Model S471-LED Optometer

Instruction manual for Ocean Optics USB4000 and QE65 Pro spectroradiometers

Operating Instructions for ISSI Series LM2X, LM2X-DM, LM2X-DMHP

Dual-FL. World's Fastest Fluorometer. Measure absorbance spectra and fluorescence simultaneously FLUORESCENCE

Calibration of a High Dynamic Range, Low Light Level Visible Source

Optical Power Meter Basics

Instruction manual for Bentham DM-150 and DTMc300 spectroradiometers

LED Lighting Flux and Color Measurement System (LFC)

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

PROPOSED METHOD OF MEASURING DISPLAY SYSTEMS FOR TRAINING WITH STIMULATED NIGHT VISION GOGGLES

EXTERNAL NVG FRIENDLY AND COVERT LIGHTING FOR HELICOPTERS

Digital Image Processing

Radiometric Measurement Traceability Paths for Photovoltaic Calibrations. Howard W. Yoon Physical Measurement Laboratory NIST

Basic Components of Spectroscopic. Instrumentation

Instruction Manual of Luminance and Spectral Radiance Calibrations

MAKING LIGHT WORK OF LIGHT MEASUREMENT. Guide to Photometer & Radiometer. System Configuration

LSST All-Sky IR Camera Cloud Monitoring Test Results

The FTNIR Myths... Misinformation or Truth

Keysight Technologies Optical Power Meter Head Special Calibrations. Brochure

Colorimetry and Color Modeling

High Illuminance Calibration Facility and Procedures

Chapter 2: Digital Image Fundamentals. Digital image processing is based on. Mathematical and probabilistic models Human intuition and analysis

NVIS COMPATIBILITY (A PRIMER)

PCS-150 / PCI-200 High Speed Boxcar Modules

Components of Optical Instruments. Chapter 7_III UV, Visible and IR Instruments

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Airport Visual Aids. Illuminating Engineering Society Airport Lighting Committee. Trade Winds Island Grand Beach Resort St.

ISS-30-VA. Product tags: Integrating Sphere Source. Gigahertz-Optik GmbH 1/5

Improved Radiometry for LED Arrays

Aqualog. CDOM Measurements Made Easy PARTICLE CHARACTERIZATION ELEMENTAL ANALYSIS FLUORESCENCE GRATINGS & OEM SPECTROMETERS OPTICAL COMPONENTS RAMAN

BTS256-PAR. Product tags: VIS, Spectral Data, PAR, Waterproof, General lighting.

A NEW GENERATION PROGRAMMABLE PHASE/AMPLITUDE MEASUREMENT RECEIVER

FFT Spectrum Analyzer

BGS400 Lamp & Luminaire Goniophotometer (C-γ, B-ß Plane)

Current and Future Realizations NRC Photometric and Spectroradiometric Calibration Chains

HF Upgrade Studies: Characterization of Photo-Multiplier Tubes

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs

2. Instrumentation Design, Specifications, and Installation of the SUV-100

Vixar High Power Array Technology

The only simultaneous absorbance and f uorescence system for water quality analysis! Aqualog

UV-VIS-IR Spectral Responsivity Measurement System for Solar Cells

PRODUCTION DATA SHEET

Radiometric and Photometric Measurements with TAOS PhotoSensors

Content Spectrophotometers

The Art of Light Measurement. Avantes BV Apeldoorn, The Netherlands

Ch 5 Hardware Components for Automation

A simulation tool for evaluating digital camera image quality

Transcription:

Specialized Spectral Measurement Equipment And Techniques Dominate Night Vision Applications Since the inception of night vision equipment in the early 1950s the development of night vision goggles has progressed through three generations of image intensifier tubes.the present generation (commonly referred to as GEN III) of goggles with gallium arsenide (GaAs) photocathodes provides many times better performance with nightsky radiance than the second generation AN/PVS-5 Aviators Night Vision Imaging System (ANVIS). Such improved performance of the GEN III ANVIS goggles in the near infrared spectral region from 630 to 930 nanometers (Figure 1), while providing increased capability for night-time flight operations, created a compatibility problem with existing cockpit instrumentation lighting.the instrumentation illumination in certain military cockpits generates enough energy in the near infrared to affect the automatic gain control of the ANVIS goggles. In effect, this incompatibility blinds the goggles to the less-illuminated outside scene.with the pilot's vision essentially blinded, the entire aircraft crew is at risk. To address the ANVIS goggle/cockpit lighting compatibility problem, researchers at the Naval Air Development Center (NADC) started work on a definitive specification for radiance and irradiance levels of cockpit instrumentation. The primary tool used to investigate existing cockpit instrumentation was a computer-controlled spectroradiometer that utilized a cooled photomultiplier tube with an S-1 photocathode thereby allowing measurements over the 380-nm to 1100-nm spectral region.the resulting specification, MIL-L-85762 and the revised MIL-L-85762A, not only sets acceptable radiance and irradiance levels, but it also specifies acceptable measurement equipment and a particular calculation method. Making accurate measurements at these low spectral radiance levels is no easy task. Measurement Equipment MIL-L-85762A specifies a spectroradiometer for the measurement of chromaticity and spectral radiance.that spectroradiometer must have sufficient sensitivity to permit measurement of radiance levels equal to or less than those listed in Table 1.These measurements must be made at a half-power bandwidth of 10 nm and a root-meansquare signal-to-noise ratio of 10:1. In practical application, however, spectroradiometer sensitivities 10 to 100 times higher than this are required to make measurements on actual lighted cockpit instrumentation at a repeatability level of ± 1 percent. Additional Table 1 - Maximum Sensitivity of Spectroradiometer Required by MIL-L-85762A Wavelength Radiance Level 380 to 600 nm 1.0 x 10-10 W/cm 2 sr nm 600 to 900 nm 1.7 x 10-11 W/cm 2 sr nm 900 to 930 nm 1.0 x 10-10 W/cm 2 sr nm 1

Relative Response And Light Energy Level 1.00.90.80.70.60.50.40.30.20.10 CIE Photopic Curve GEN III AN/AVS-6 ANVIS Goggles GEN II AN/AVS-5 Goggles 0 300 400 500 600 700 800 900 1000 Wavelength (nm) Blue Green Red Visible Light Region Infrared Region Light Available From Night Sky Figure 1. Response of GEN III AN/AVS-6 ANVIS night vision goggles with respect to night sky radiation and GEN II goggles. requirements for the measurement system are summarized in Table 2.These system parameters represent the minimum requirements to determine whether the spectral radiance of a lighted cockpit component will interfere with automatic gain control in the GEN III goggles. Calculation Method Once the spectral radiance of the cockpit lighting component has been accurately determined, the calculation of the ANVIS radiance is performed.this process first determines the luminance of the lighting component so a scale factor value can be computed. For example, say the spectral radiance data is analyzed and the luminance turns out to be 5.432 footlamberts (18.610 cd/m 2 ).Then the scale factor, S, to set the luminance to 0.1 fl is S = 0.1/5.432 = 0.01841.The scale factor is then applied to the spectral radiance data, which is then convolved with the relative ANVIS response function G(λ) and integrated over the 450 to 930 nanometer spectral region.the result of this integral is called the ANVIS radiance (AR) of the lighting component and generally must be less than 1.7 x 10-10 AR. A revision of MIL-L-85762 was completed that defines lighting compatibility requirements for other types of night vision devices. MIL-L-85762A addresses this by defining a new relative response function Class B NVIS and redefining the old ANVIS relative response function as Class A NVIS (where NVIS means Night Vision Imaging System).The calculation method of ANVIS radiance for these changes remains unaltered; only the class of NVIS device, either Class A or Class B, needs to be defined (NVIS radiance, Class A = NRA and Class B = NRB in MIL-L- 85762A). Figure 2 shows the difference between these two spectral response functions, where the black curve is the NVIS radiance Class A (NRA) and the colored curve is the NVIS radiance Class B (NRB).The calculation method of ANVIS radiance for these changes remains unaltered, but the luminance used to calculate the scale factor for NVIS radiance Class B (NRB) is 15 fl, instead of the 0.1 fl for ANVIS radiance (AR) and NVIS radiance Class A (NRA). Actual Measurements The key to making accurate and repeatable measurements at these low light levels is a spectroradiometer with a sensitivity of at least 1.7 x 10-11 W/cm 2 -sr-nm with a 10:1 2

signal-to-noise ratio in water-cooled thermoelectric housing. Table 2 - Spectroradiometer Measurement the 600 to 900 nm System Requirements In MIL-L-85762 region. By developing a System Parameter Summarized Requirements Holding the PMT s temperature photomultiplier-based Spectroradiometer Calibration must be traceable to NIST constant by detector, proprietary Sensitivity Calibration Standards at a max. 6 month interval carefully controlling the assembly language algorithms Wavelength Accuracy and Repeatability Accuracy ±1.0 nm, Repeatability ±0.5 nm water bath temperature for a microproces- Wavelength Accuracy Verified at one wavelength in every 150 nm can help to ensure sta- sor-based radiometer, and Repeatability Verification interval from 350 to 950 nm using scanning or non-scanning technique ble PMT operating characteristics. and data acquisition Digital Resolution 11 bit A to D plus sign or ±0.05% of each software, we have produced of the Photocurrent measured scale Included within the a spectrora- Zero Drift Less than 0.2% of full scale reading on cooled PMT housing are most sensitive scale or range diometer that exceeds the optics necessary to Linearity ±1% of full scale, ±2% between scales this requirement by as Linearity Verification Inverse square law, aperture super-position couple the output of the much as two orders of Method or ND s with known transmission spectroradiometer s magnitude.this added Signal Conditioning Controls to improve or change S/N ratio monochromator to the sensitivity allows the use Stray Light Cannot adversely affect accuracy PMT photocathode efficiently.the of input optics on the spectroradiometer, providing Stray Light Verification Measured spectral radiance of NIST traceable standard filtered with Schott BG23 must equal standard spectral radiance * filter transmission ±5% detector assembly also contains a well-defined measurement area. Optics Minimum spot 0.007 with 1.0 fl full scale sensitivity an internal reference lamp, which is controlled The photomultiplier Viewing System Locate measurement spot within 5% of the diameter of the spot to be measured by the micro- tube (PMT) used in this Viewing System X-Y motion of black opaque rear processor-based application was selected Verification illuminated aperture radiometer.the reference for both its spectral Accuracy Spectral radiance within ±5% of NIST traceable standard from 380 to 930 in 5 nm lamp can be used range and low noise increments and UCS u and v coordinates to compensate for to ±0.007 characteristics. Its spectral range of 185 nm up radiance other than the spectroradiometer automatically, changes changes in PMT gain Accuracy Verification Measure NIST traceable standard of spectral calibration standard to more than 950 nm which occur when the meets the requirements imposed by MIL-L-85762A.The S/N criterion can be met by cooling the PMT to -30 C in a high voltage to the PMT is set to different values to expand the system s overall dynamic range. 3

Single- or double-grating monochromators are the spectral selector. Both types use f/3.5 concave holographic diffraction gratings to cover the wavelength range from 300 to 1100 nm. The digital stepper motor that drives the diffraction grating(s) operates over extended periods without drift. Also part of the monochromator are several other components that keep the wavelength position of the grating accurate, control the placement of order-selection filters and operate the electric shutter. The monochromator and the selected detector are both controlled by a microprocessor-based intelligent radiometer and scanning controller that acts as an interface between a host computer and the actual measurement devices. An instruction code set of about 30 ASCII (American Standard Code for Information Interchange) commands enables the system to be used for spectral and spatial measurements. The commands access firmware instructions written in assembly language. These firmware instructions, stored in erasable programmable read-only memory (EPROM), allow fast and precise interaction with the monochromator and the detector. The ASCII commands encoded into the firmware of the radiometer allow optical measurement to be almost entirely automated. With one of these ASCII commands, the radiometer will scan from a low to a high wavelength with a specified wavelength step increment and return detector signal data at each step. Another command sets the number of digital averages taken, from 1 to 255, for each of these detector signal readings. Relative Sensitivity 1.0 3 10-1 2 x 10-1 1 x 10-1 3 x 10-2 2 x 10-2 1 x 10-2 3 x 10-3 2 x 10-3 1 x 10-3 3 x 10-4 2 x 10-4 1 x 10-4 NVIS Class A Max = 1 Min = 1.0000 x 10-4 NVIS Class B Max = 1 Min = 1.0000 x 10-5 3 x 10-5 2 x 10-5 1 x 10-5 450 500 600 700 800 900 Figure 2. Logarithmic plot of NVIS A and NVIS B relative responses. Some of the functions that the device performs automatically include: Scanning a range of wavelengths and returning the data to the host computer. Taking up to 255 readings at each wavelength position and returning the average value to the host computer. Opening and closing the shutter to the detector. Selecting the correct electrometer amplifier range. Subtracting dark current from the detector signal. Setting and reading the high-voltage level to PMT detectors. Calibrating the wavelength scale with up to nine reference wavelengths. Calibrating the detector amplitude scale. Scanning variable intervals. 4

Correcting system (detector and electrometer amplifier) nonlinearity to better than 1%. Turning the reference lamp on or off. Setting or reading the real-time clock. Performing self-test and error-reporting functions. Resetting the input/output buffers and the measurement equipment. The GS-4100 hardware consists of a photometer/radiometer printed circuit board, a central processor board, an RS-232/EPROM board, an analog-to-digital and digital-to-analog board, a motion control board, a stepper control board, and a power supply module. A highly stable power supply, stable to ±0.06 C over a temperature range of 0 to +50 C and mounted on the photometer/radiometer board, drives the PMT. The electrometer amplifier boosts the signal current from the PMT in any of four channels, each differing by a factor of 10. The microprocessor in the radiometer selects the appropriate range for the photoelectrically created current from the PMT. The microprocessor also controls the high voltage applied to the PMT and automatically compensates any zero offset and suppresses the dark current. All these functions take place on the A/D-D/A board, which is matched to the radiometer board. The motion board provides two monochromator motor speeds: fast for long moves, and slow for approaching a wavelength where a detector reading will be made. This two-speed approach eliminates overshoot and hunting. The RS-232/EPROM board contains the main program memory, the baud rate selector and the drivers for communicating with the host computer. Up to 32 kilobytes of assembly language instructions and communications rates up to 9600 baud can be selected on this board. An IBM PC or compatible computer runs the DOS or new Windows based software package that utilizes the firmware commands in the device. This software is menu-driven and allows automatic test sequencing. All system measurement parameters such as high voltage, number of averages, wavelength range and interval, and scan direction can all be selected in advance and stored in test sequence disk files. The program can then execute the test sequence with a single keystroke and run totally unattended for hours or days, acquiring, storing and automatically analyzing the spectral radiance data. Once the test sequence is set, measurements can be taken by semiskilled operators with a minimum of training, providing a great advantage in production-line applications of these low-light-level measurements. The Windows 95/98/NT version of the software also provides a direct link to Microsoft EXCEL spreadsheet. A powerfull macro function allows selection of the data to be linked from the spectral radiance values to the luminance or chromaticity coordinates. The NVIS analysis section of this software package also facilitates the pass/fail criteria of the measurement results with a cockpit lighting application catagory. Figure 3 shows an example of an analysis display of this software function. 5

Typical results are shown in Figure 4. This spectral radiance measurement was made with an effective spot diameter of 3.18 millimeters on a 1.3 fl filtered tungsten source. The spectral radiance values between 600 and 700 nm drops from 1 x 10-10 to less than 5 x 10-12 W/cm 2 - sr-nm. Figure 4 illustrates the effect of applying the NVIS Class A weighting function (the black curve of Figure 2) to the spectral radiance data of Figure 5. The previously dominant visible portion of the spectrum is reduced to about one quarter of the total NVIS Class A radiance value of 1.7 x 10-10 W/cm 2 -sr-nm. The important region turns out to be that between 600 and 700 nm. The low point of 5 x 10-12 W/cm 2 -sr-nm at 685 nm can clearly be seen in the middle of the plot. Figures 6 and 7 show the results of a measurement with the same effective spot diameter on an electroluminescent panel with improper filtration. Once again, the radiance in the 600 to 700 nm span ranges from 1 x 10-10 to almost 5 x 10-12 W/cm 2 -sr-nm. This time, however, the luminance is only 0.66 fl, so that when the NVIS Class A weighting function is applied, there is a larger proportionate amount of radiant energy in the near infrared versus the visible region. So, when the spectral radiance is scaled to give a luminance of 0.1 fl, the AR calculation gives an unacceptably high NVIS Class A value of 3.2 x 10-10 AR. Figure 3. Example of MIL-L-85762A analysis of cockpit display measurement data. Radiance (watts) 1E-7 5E-8 1E-8 5E-9 1E-9 5E-10 1E-10 5E-11 1E-11 5E-12 1E-12 380 500 600 700 800 900 950 Figure 4. Single spectral radiance measurement of lighting component meeting compatibility requirements of MIL-L-85762. Note that the radiance units are in microwatts. 6

ANVIS Radiance (watts) 1.00E-12 8.00E-13 6.00E-13 4.00E-13 2.00E-13 1.60E-15 450 500 600 700 800 900 Luminance: 1.29 fl NVISa Rad: 2.1E-09 AR / Scl (.1): 1.7E-10 AR NVISb Rad: 2.0E-07 AR / Scl (15): 1.6E-08 AR Figure 5. Spectral radiance data from Figure 3, multiplied by the ANVIS relative response curve (the black curve in Figure 2). The data under the plot shows the luminance for the measurement data. Also, the NVIS A and NVIS B radiances are given for scaled and unscaled luminance values. Radiance (watts) 1E-8 5E-9 1E-9 5E-10 1E-10 5E-11 1E-11 5E-12 1E-12 5E-13 Figure 6. Similar data as in Figure 3, but from an improperly filtered electroluminescent display. 1E-13 380 500 600 700 800 900 950 ANVIS Radiance (watts) 4.00E-12 3.00E-12 2.00E-12 1.00E-12 1.49E-15 450 500 600 700 800 900 Luminance: 0.66 fl NVISa Rad: 2.1E-09 AR / Scl (.1): 3.2E-10 AR NVISb Rad: 1.1E-09 AR / Scl (15): 1.7E-10 AR Figure 7. Spectral radiance data from Figure 5, multiplied by the ANVIS relative response curve. The data under the plot shows the luminance and NVIS A and NVIS B radiances for the scaled and unscaled luminance values. Note that the peak radiance in the 600 to 700 nm region is four times that in Figure 4. 7