Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects

Similar documents
Biosignal Data Acquisition and its Post-processing

MULTIPLE PULSE WAVE MEASUREMENT TOWARD ESTIMATING CONDITION OF HUMAN ARTERIES

WRIST BAND PULSE OXIMETER

Low-cost photoplethysmograph solutions using the Raspberry Pi

Frequency Domain Analysis for Assessing Fluid Responsiveness by Using Instantaneous Pulse Rate Variability

Amplitude Modulation Effects in Cardiac Signals

Design of Arterial Blood Pressure, Heart Rate Variability, and Breathing Rate Monitoring Device. Mastan Singh Kalsi

Design of Wearable Pulse Oximeter Sensor Module for Capturing PPG Signals

Measurement of Spatial Pulse Wave Velocity by Using a Clip-Type Pulsimeter Equipped with a Hall Sensor and Photoplethysmography

Chapter 2. Design and development of blood volume pulse sensor and heart rate meter. Abstract

Computer Evaluation of Exercise Based on Blood Volume Pulse (BVP) Waveform Changes

Relation between HF HRV and Respiratory Frequency

Biosignal filtering and artifact rejection, Part II. Biosignal processing, S Autumn 2017

*Notebook is excluded

DESIGN OF A PHOTOPLETHYSMOGRAPHY BASED PULSE RATE DETECTOR

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

understand compatibility of photoplethysmographic pulse rate variability with electrocardiogramic heart rate variability

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

City, University of London Institutional Repository

AN2944 Application note

Design Considerations for Wrist- Wearable Heart Rate Monitors

Design and Validation of an. Arterial Pulse Wave Analysis

Pulse Oximetry. Principles of oximetry

APPLICATION OF HEART PHOTOPLETHYSMOGRAPHY

Biomedical Signal Processing and Applications

Artifact Removal from the Radial Bioimpedance Signal using Adaptive Wavelet Packet Transform

Research Article Human Heart Pulse Wave Responses Measured Simultaneously at Several Sensor Placements by Two MR-Compatible Fibre Optic Methods

Estimating Frequency Response Characteristics of Human Baroreflex System

ELR 4202C Project: Finger Pulse Display Module

Embedded Prototype System for Monitoring Heart Rate

Real Time Heart Attack and Heart Rate Monitoring Android Application

ECG Data Compression

Noninvasive Radial Pressure Waveform Estimation by Transfer Functions Using Particle Swarm Optimization

Doppler in Obstetrics: book by K Nicolaides, G Rizzo, K Hecher. Chapter on Doppler ultrasound: principles and practice by Colin Deane

Development of Electrocardiograph Monitoring System

Biosignal Analysis Biosignal Processing Methods. Medical Informatics WS 2007/2008

PHOTOPLETHYSMOGRAPHIC DETECTOR FOR PERIPHERAL PULSE REGISTRATION

Chapter 5. Frequency Domain Analysis

Next Generation Biometric Sensing in Wearable Devices

Heart Rate Monitoring using Adaptive Noise Cancellation

Laboratory Activities Handbook

6.555 Lab1: The Electrocardiogram

Laboratory Kit for Oscillometry Measurement of Blood Pressure

(51) Int Cl.: A61B 5/00 ( ) G06F 17/00 ( )

VivoSense. User Manual - Equivital Import Module. Vivonoetics, Inc. San Diego, CA, USA Tel. (858) , Fax. (248)

Question 1 Draw a block diagram to illustrate how the data was acquired. Be sure to include important parameter values

Design of Virtual Sphygmomanometer Based on LABVIEWComparison, Reflection, Biological assets, Accounting standard.

Bio-Potential Amplifiers

A Design Of Simple And Low Cost Heart Rate Monitor

Validation of the Happify Breather Biofeedback Exercise to Track Heart Rate Variability Using an Optical Sensor

PHYSIOLOGICAL SIGNALS AND VEHICLE PARAMETERS MONITORING SYSTEM FOR EMERGENCY PATIENT TRANSPORTATION

EKG De-noising using 2-D Wavelet Techniques

An EMFi-film Sensor based Ballistocardiographic Chair: Performance and Cycle Extraction Method

Can Very High Frequency Instantaneous Pulse Rate Variability Serve as an Obvious Indicator of Peripheral Circulation?

City, University of London Institutional Repository

Wireless post-processing and interfacing ECG, blood pressure and blood oxygen measurement systems

An Automated Algorithm for Fast Pulse Wave Detection

Intelligent Pillow for Heart Rate Monitor

HUMAN BODY MONITORING SYSTEM USING WSN WITH GSM AND GPS

Introduction to Medical Electronics Industry Test Analysis and Solution

Fetal ECG Extraction Using Independent Component Analysis

University of Tlemcen

Medical Electronics Dr. Neil Townsend Michaelmas Term 2001 ( Pulse Oximetry: The story so far

HIGH FREQUENCY FILTERING OF 24-HOUR HEART RATE DATA

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

Biomedical and Wireless Technologies for Pervasive Healthcare

ARTICLE IN PRESS. Computers in Biology and Medicine

Wrist Pulse Acquisition and Recording System

Biomedical. Measurement and Design ELEC4623/ELEC9734. Electrical Safety and Performance Standards

Jordan University of Science & Technology Faculty of Engineering. Department of Biomedical Engineering BME 443. Biomedical Instrumentation Lab I

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

CHAPTER 1 INTRODUCTION AND LITERATURE SURVEY

HRV spectrum bands & single peak Coherence

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

Spectral Analysis and Heart Rate Variability: Principles and Biomedical Applications. Dr. Harvey N. Mayrovitz

Crew Health Monitoring Systems

D5.1 Report on the design of a fibre sensor based on NIRS

AC : A HIGH-PERFORMANCE WIRELESS REFLECTANCE PULSE OXIMETER FOR PHOTO-PLETHYSMOGRAM ACQUISITION AND ANALYSIS IN THE CLASSROOM

Masimo Corporation 40 Parker Irvine, California Tel Fax

Your heart in good hands.

Introduction to Ultrasound Physics

P. Robert, K. Kodera, S. Perraut, R. Gendrin, and C. de Villedary

The Removal Of Motion Artifacts From Noninvasive Blood Pressure Measurements

PULSE OXIMETRY MODULE TO IMPLEMENT IN TEAM MONITOR OF VITAL SIGNS

GE Healthcare. Dash 2500 The standard of excellence for sub-acuity monitoring

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION

Portable, Low Cost, Low Power Cardiac Interpreter

Standoff Human Life Sign Detection using High Sensitivity Pulsed Laser Vibrometer

The Application of the Hilbert-Huang Transform in Through-wall Life Detection with UWB Impulse Radar

Detection of Abnormalities in the Functioning of Heart Using DSP Techniques

SFH Photoplethysmography Sensor

Arduino and Raspberry Pi based Efficient Patient Monitoring System

Sensors. CSE 666 Lecture Slides SUNY at Buffalo

Sensor, Signal and Information Processing (SenSIP) Center and NSF Industry Consortium (I/UCRC)

III Lead ECG Pulse Measurement Sensor

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

m+p Analyzer Revision 5.2

Electrode comparison for textile-integrated electrocardiogram and impedance pneumography measurement

A novel biometric signature: multi-site, remote (> 100 m) photo-plethysmography using ambient light.

MASTER THESIS TITLE: Quadrature synchronous sampling for electrical impedance plethysmography implemented on a MSP432 microcontroller

Transcription:

Arterial pulse waves measured with EMFi and PPG sensors and comparison of the pulse waveform spectral and decomposition analysis in healthy subjects Matti Huotari 1, Antti Vehkaoja 2, Kari Määttä 1, Juha Röning 1 1 Oulu University, Oulu, Finland, 2 Tampere University of Technology, Tampere, Finland Matti Huotari, Oulu University, Oulu, FINLAND. Email: matti.huotari@ee.oulu.fi. Abstract The purpose of this study is to show the time domain and frequency domain analysis of signals recorded with Electromechanical Film (EMFi) and Photoplethysmographic (PPG) sensors in arterial elasticity estimation via pulse wave decomposition and spectral components obtained from left forefinger, wrist, and second toe arteries. ECG and pulse waves from the subjects were recorded from 7 persons (30 60 y) in supine position. Decomposition of the pulse waves produces five components: percussion, tidal, dicrotic, repercussion, and retidal waves. Pulse wave decomposition parameters between EMFi and PPG are compared to detect variables for information on person s arterial elasticity. Results show that elasticity information in the form of pulse wave decomposition from PPG and EMFi waves is obtainable and shows clear shortening between percussion wave and tidal wave peak time in PPG waveforms with age. The spectral information obtained with frequency domain analysis could also be valuable in assessment of the arterial elasticity. In addition, both PPG and EMFi measurements are absolutely non invasive and safe. In PPG measurement, the sensors are on the opposite sides of the finger tip, however, EMFi measurement needs the good skilled operator attaching the sensor on the patient s wrist by touching gently to obtain accurate waveforms. Keywords: arterial elasticity measurement, electromechanical film (EMFi), photoplethysmography (PPG), pulse wave decomposition 27.5.2013 FinJeHeW 2013;5(2 3) 57

Introduction Arterial pulse wave is defined as a heart beat driven wave of blood that propagates via each artery into vein through capillaries. Various kind of medical information can be obtained by the detection of pulse wave and its detailed analysis. This information is not obtainable from the blood pressure or electrocardiographic measurements. It could be possible to make early diagnosis on the basis of the pulse wave analysis. Especially atherosclerosis is the main cause of circulatory diseases. Its quantitative assessment is essential for making an early diagnosis of such diseases. In addition, persons with other cardiovascular diseases (CVD) may have decreased arterial elasticity compared with those free of CVDs. Change of arterial elasticity is one of the early markers of accelerated arterial aging and can correlate with many coronary risk factors. Especially arterial elasticity reflects the arterial and aortic expanding during left ventricular contraction. Arterial elasticity can be measured indirectly provided the measurement method is relevant and accurate enough. In this study, biophysical function and structure of the arteries have been measured by photoplethysmographic (PPG) and electromechanical film (EMFi) sensors. For the measurement data, we applied pulse wave decomposition (PWD), which reflects clearly the elasticity of the aorta and its peripheral arteries. The combined PPG & EMFi measurements can establish aortic and arterial elasticity based on PWD of the both signals during a heart cycle. We do not necessarily need distance measurement required for pulse wave velocity estimation, which can be rather inaccurate in the case of the arterial tree [1]. The EMFi and PPG technologies require a few electromechanical or opto electronic components: a EMFi film sensor connected to an amplifier, and a light source to illuminate the tissue (e.g. finger), and a photodetector to measure the small variations in light absorbance amplified also by a transimpedance amplifier. The obtained pulse wave is the peripheral pulse that can be decomposed into five logarithmic normal components. Despite its simplicity, the origins of the different components of the PPG or EMFi signal are not fully understood [2]. We believe that these parallel pulse waves can provide valuable information about the circulatory system with patient friendly and safe means. In the elderly and in the stiffened arteries, the forward and reflected pulse waves travel faster, i.e., pulse wave velocity (PWV) is higher than in a young person s arteries which are elastic. The arterial waves reflected from the periphery of the arterial tree, return earlier merging with the systolic part of the incident wave causing augmentation of the workload of the heart. Favorable softness between coupling of the left ventricle and the arterial tree is thus progressively lost. This loss can be greatest in the aorta, and least in the upper limbs. The wave reflection of the pulse wave due to increase in PWV also increases with age, but can be largely prevented by physical activity and proper diet. The amplitude spectrum of the ECG, EMFi, and PPG are changing by so called integral pulse frequency modulation (IPFM) [3]. This modulation is caused by the autonomic control mechanisms of cardiac functions which are involved in short term fluctuations in the time interval between the consecutive heart cycles. The IPFM reflects cardiac function which is also detected in the periphery of the arterial tree. Healthy modulation in coupling of the left ventricle and the arterial tree is, however, progressively lost. In our study we measure EMFi, two PPGs, and ECG signals, and we apply Fast Fourier Transform on the signals in addition to logarithmic normal function decomposition of the EMFi and PPG pulse waveforms. Methods In this study, the PPG sensors are based on LEDs and photodetector, EMFi sensor is based on a plastic EMFi film, and ECG sensors are standard electrodes. Changes in light absorption, in the pressure, or in electrical potentials are acting on each sensor generating a measurable voltage. The EMFi sensor acts as a sensitive pressure sensor and the PPG sensor as a sensitive absorbance sensor. Signals from ECG, PPG, and EMFi sensors were recorded with the 27.5.2013 FinJeHeW 2013;5(2 3) 58

PC from 7 healthy persons (30 60 y) using a data acquisition card with the sampling frequency of 0 Hz. ECG signal was used as reference in detecting features from the PPG and EMFi related signals. Pulse wave series from EMFi, PPG finger and toe were processed by Origin software as follows. Firstly, the alternating baseline on each signal bottom (minimum) was searched and then subtracted. Secondly, the peak of each signal (maximum) was searched, and the average of the peak values was calculated. The signal was divided by the average to obtain the baseline removed and normalized waveforms which has amplitude from the zero to round about one. Processing continues a pulse by a pulse in the PWD. Results An example pulse wave from each sensor is shown in Figure 1 on which analysis in time domain and FFT in frequency domain was performed. Each pulse wave component from EMFi, PPG finger, and toe were processed by the software. Decomposed pulse waves from the left wrist (EMFi), and from the left forefinger and the left second toe (PPG) are decomposed in time domain and transformed in frequency domain. In Figure 1 it is shown EMFi (solid), PPG finger (dash dot), PPG toe (dashed) pulse waves, and electrocardiogram (ECG) (dot). They have the start time as zero and then decomposed with their residual and confidence intervals, respectively, in the Figure 2. rel 1,2 B EMFi_TTSS_30 D PPG2 (toe) F PPG1 (finger) H ECG 0,6 0,4-0,5 1,5 2,0 2,5 t[s] 3,0 Figure 1. The baseline removed normalized pulse waves: EMFi (left wrist, solid), PPG1 (left forefinger, dash dot), PPG2 (left second toe, dashed), and ECG (dot) (Male 30). 27.5.2013 FinJeHeW 2013;5(2 3) 59

Data: Data24_B TTSS_30-EMFI-2 y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc)) ^2/(2*w^2)) Chi^2/DoF = 0.00007 R^2 = 0.99899 xc1 0.16787 ±0.00041 w1 0.57342 ±0.00164 A1 0.19773 ±0.00031 xc2 0.29496 ±0.00254 w2 0.11755 ±0.00795 A2 0.0053 ±0.00027 xc3 0.47814 ±0.00355 w3 0.17313 ±0.00546 A3 0.07034 ±0.00109 xc4 0.6713 ±0.00765 w4 0.12885 ±0.02126 A4 0.01904 ±0.00192 xc5 0.806 ±0.00518 w5 0.06443 ±0.01196 A5 0.00423 ±0.00153 SCIENTIFIC PAPERS 1,25 a Data: Data24_B TTSS_30-EMFI-2 y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc))^2/(2*w^2)) EMFi rel 0 0,75 0, 0,128 Chi^2/DoF = 0.00007 R^2 = 0.99899 xc1 0.16787 ±0.00041 w1 0.57342 ±0.00164 A1 0.19773 ±0.00031 xc2 0.29496 ±0.00254 w2 0.11755 ±0.00795 A2 0.0053 ±0.00027 xc3 0.47814 ±0.00355 w3 0.17313 ±0.00546 A3 0.07034 ±0.00109 xc4 0.6713 ±0.00765 w4 0.12885 ±0.02126 A4 0.01904 ±0.00192 xc5 0.806 ±0.00518 w5 0.06443 ±0.01196 A5 0.00423 ±0.00153 EMFi rel 4 3 2 1-1 -2-3 0,656 0,176 s 0,192 s 0,128 s+16 s= 0,144 s 0,6 0,160 s_32 s_2,2 0,7 0,9 Resudual (32 s)/2=16 s 0,6 0,7 0,9 5 4 3 2 1-1 -2-3 0,464 0,176 s 0,192 s 0,160 s_32 s_2,2 (32 s)/2=16 s 88 0,656 0,128 s+16 s= 0,144 s 0,1 0,3 0,4 0,5 0,6 0,7 0,9 Resudual 0,1 0,3 0,4 0,5 0,6 0,7 0,9 PPG1 rel 1,1 0,9 0,7 0,6 0,5 0,4 0,3 0,1-0,1 25 0-25 - b 0,14217 0,4739 0536 s 0,17376 s 6854 158 s * 2=316 s 0,64766 0,12637 s_-158 s_1,88887 EMFip-PPG1P=0,14217 s - 0,128 s=-1417 s 0,12637 s 0,77403 Data: Data34_B TTSS_30PPG1_2 y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc))^2/(2*w^2)) Chi^2/DoF = 0.00023 R^2 = 0.99707 xc1 0.26002 ±0.00431 w1 0.79597 ±0.00635 A1 0.34821 ±0.00451 xc2 0.27935 ±0.00414 w2 0.2149 ±0 A2 0.02029 ±0.00115 xc3 0.4971 ±0.0093 w3 0.21556 ±0.01328 A3 0.11562 ±0.00534 xc4 0.6575 ±0.08 w4 0.11953 ±0.02158 A4 0.03082 ±0.0073 xc5 0.77568 ±0.001 w5 0.06467 ±0.00868 A5 0.01035 ±0.00307 0,1 0,3 0,4 0,5 0,6 0,7 0,9 t[s] Residual t[s] 0,1 0,3 0,4 0,5 0,6 0,7 0,9 Figure 2a) A single EMFi pulse wave (solid, measured) which is here decomposed into components: percussion (dash), tidal (dot), dicrotic (dash dot), repercussion (short dash), and retidal wave (short dot). (Insert the two last wave components). The confidence interval (99%) is marketed short dot dot. The residual curve is shown in the lower panel. b) A single PPG1 (finger) pulse wave decomposed, respectively. (Male 30). 27.5.2013 FinJeHeW 2013;5(2 3) 60

EMFi rel 1,1 0,9 0,7 0,6 0,5 0,4 0,3 0,1 3 2 1-1 -2 a Data: Data30_B y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc))^2/(2*w^2)) Chi^2/DoF = 0.00005 R^2 = 0.99943 0,14478 xc1 0.21327 ±0.00091 w1 0.61405 ±0.00213 A1 0.26783 ±0.00078 xc2 0.29341 ±0.00055 w2 0.14128 ±0.00202 A2 0.02649 ±0.00038 xc3 0.49686 ±0.00153 w3 0.19327 ±0.00264 A3 0.0898 ±0.00087 xc4 0.80032 ±0.00409 w4 0.20676 ±0.00389 A4 0.0672 ±0.00069 xc5 1.1614 ±0.00319 w5 0.01339 ±0.00291 A5 0.00052 ±0.0001 9356 0,48485 0,19129 763 0,76115 0,14878 s_4 s_2,028 0,40383 1,16498 PPG1p-EMFip=0,18637-0.14478=4159 0,1 0,3 0,4 0,5 0,6 0,7 0,9 1,1 1,2 1,3 Residual 0,1 0,3 0,4 0,5 0,6 0,7 0,9 1,1 1,2 1,3 PPG1 rel 1,4 1,3 b Data: Data2_B PL_60 1,2 y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc))^2/(2*w^2)) 1,1 Chi^2/DoF = 0.00017 R^2 = 0.9982 xc1 0.28169 ±0.00702 w1 0.6794 ±0.00901 A1 0.30214 ±0.00715 0,9 xc2 0.29168 ±0.00323 w2 0.20137 ±0 A2 0.03543 ±0.00244 xc3 0.55412 ±0.0201 w3 0.29211 ±0.02518 A3 0.2456 ±0.01681 xc4 0.80841 ±0.03208 w4 0.17232 ±0.03695 0,7 0,18241 A4 0.09407 ±0.03278 0,5138 xc5 0.98826 ±0.00982 w5 0.10954 ±0.01568 A5 0.0349 ±0.01 0,6 0,5 2783 6925 0,4 0,3 8597-4142 0,78305 0,96946 0,1 0,18641 0,10356 s_-7885 s_1,568-0,1-0,1 0,1 0,3 0,4 0,5 0,6 0,7 0,9 1,1 1,2 4 Residual 2-2 -4-0,1 0,1 0,3 0,4 0,5 0,6 0,7 0,9 1,1 1,2 Figure 3a) A EMFi pulse wave (solid, measured) which is decomposed into components: percussion (dash), tidal (dot), dicrotic (dash dot), repercussion (short dash), and retidal wave (short dot). The confidence interval (99%) is marketed short dot dot. The residual curve in the lower panel. b) A single PPG1 (finger) pulse wave decomposed, respectively. (Male 60). 27.5.2013 FinJeHeW 2013;5(2 3) 61

Integral of residuals 10 9 8 7 6 5 4 3 2 Integral of Data6_C EMFi Integral of Data16_C PPGfinger Integral of Data27_C Integral of Data37_C Integral of Data48_C Integral of Data58_C abs [Residual EMFi ] 35 30 25 20 15 10 C abs[residual EMFi ] 1 5 0 0,4 0,6 0 0,1 0,3 0,4 0,5 0,6 0,7 0,9t[s] Figure 4. Integral of residuals (EMFi solid) and PPG1 (finger, dash). Insert: Absolute value of residual of EMFi waveform, (Male 30). PPG2 rel 1,1 0,9 0,7 0,6 0,5 0,4 0,3 0,1-0,1-0,1 0,1 0,3 0,4 0,5 0,6 0,7 0,9 4 2 Residual -2 Data: Data15_B TTSS_30_PPG2 y = y0 + A/(sqrt(2*PI)*w*x)*exp(-(ln(x/xc))^2/(2*w^2)) Chi^2/DoF = 0.0001 R^2 = 0.99891 xc1 0.22511 ±0.00103 w1 0.56858 ±0.00198 A1 0.25135 ±0.00199 xc2 0.23982 ±0.00162 w2 0.20243 ±0.00819 A2 0.01611 ±0.00075 xc3 0.60279 ±0.00223 w3 0.20447 ±0.00478 A3 0.03438 ±0.00058-4 -0,1 0,1 0,3 0,4 0,5 0,6 0,7 0,9 Figure 5. The whole single PPG2 pulse wave (black, measured) which is here decomposed into components respectively as in Figure 2, but here only three components are found. (Male 30, Fig. 1). 27.5.2013 FinJeHeW 2013;5(2 3) 62

0,1 1 TTSS_30_EMFi 10 1-0,10 5 FFT2_r[4111]: X = 0.8545 Y = 0.0271 FFT2_r[4135]: X = 2,319; Y = 357 FFT2_r[4146]: X = 2,991; Y = 328 FFT2_r[4149]: X = 3,174; Y = 5 FFT2_r[4152]: X = 3,357; Y = 379 0,1 1 10 FFT2_r[4114]: X = 1.038, Y = 0.1178 FFT2_r[4117]: X = 1.2207, Y = 0.026 FFT2_r[4128]: X = 1,892; Y = 238 FFT2_r[4132]: X = 2,136; Y = 725 0,15 0,1 1 TTSS_30_PPG1 10 1-0,10 5 FFT1_r[4101]: X = 0.244, Y = 0.0254 FFT1_r[4111]: X = 0.854, Y = 0.0388 FFT1_r[4114]: X = 1.0376, Y = 0.1546 FFT1_r[4117]: X = 1.221, Y = 0.0303 FFT1_r[4128]: X = 1.892, Y = 0.0233 FFT1_r[4132]: X = 2.136, Y = 0.0700 FFT1_r[4135]: X = 2.319, Y = 0.0317 FFT1_r[4146]: X = 2.991, Y = 0.0260 FFT1_r[4149]: X = 3.174, Y = 0.0400 FFT1_r[4152]: X = 3.357, Y = 0.0294 0,1 1 10 Figure 6. The pulse wave of EMFi s amplitude spectrum (up), and of PPG1 s amplitude spectrum (down) with the IPFM parameter values for the three first components for 20 s record. In the PPG1 s amplitude spectrum contains the breath rate frequency value (Male 30). 27.5.2013 FinJeHeW 2013;5(2 3) 63

0,1 1 AAVV_31_EMFi 10 1 0,175-0,1 0,125 0, 75 25 FFT1_r[8201]: X = 44; Y = 177 FFT1_r[8217]: X = 0,732; Y = 142 FFT1_r[8225]: X = 0,977; Y = 0,1695 FFT1_r[8233]: X = 1,221; Y = 74 FFT1_r[82]: X = 1,740; Y = 92 FFT1_r[8257]: X = 1,953125; Y = 74377257 FFT1_r[8265]: X = 2,197; Y = 88 FFT1_r[8290]: X = 2,960208; Y = 426239233 0 0,1 1 10 0,1 1 AAVV_31_PPG1 10 1-0,15 0,10 5 FFT2_r[8201]: X = 44; Y = 171 FFT2_r[8217]: X = 0,732; Y = 181 FFT2_r[8225]: X = 0,977; Y = 0,1722 FFT2_r[8233]: X = 1,221; Y = 54 FFT2_r[82]: X = 1,740; Y = 88 FFT2_r[8257]: X = 1,953125; Y = 580455913 0,1 1 10 FFT2_r[8265]: X = 2,197; Y = FFT2_r[8289]: X = 2,9296875; Y = 274294923 Figure 7. The EMFi pulse wave signal s amplitude spectrum (up), and of PPG1 s amplitude spectrum (down) with the IPFM parameter values for the three first components. In the PPG1 s and EMFi s amplitude spectrum contain also the breath rate frequency value for 20 s signal record as a sample length (Male 31). 27.5.2013 FinJeHeW 2013;5(2 3) 64

1-0,1 1 PPLL_60_EMFi 10 0,10 5 FFT3_r[8220]: X = 24; Y = 0,1213 FFT3_r[8247]: X = 1,648; Y = 562 FFT3_r[8274]: X = 2,472; Y = 410 FFT3_r[8301]: X = 3,296; Y = 261 0,1 1 10 0,1 1 PPLL_60_PPG1 10 1 0,175-0,1 0,125 0, 75 25 FFT5_r[8220]: X = 24; Y = 0,1535 0 0,1 1 10 FFT5_r[8247]: X = 1,648; Y = 2 FFT5_r[8274]: X = 2,472; Y = 298 FFT5_r[8301]: X = 3,296; Y = 170 Figure 8. The EMFi pulse wave signal s amplitude spectrum (up), and of PPG1 s amplitude spectrum (down). In the PPG1 s and EMFi s amplitude spectrum contain also the breath rate frequency value (Male 60). 27.5.2013 FinJeHeW 2013;5(2 3) 65

The comparison of the PPG1 pulse waves shows that the tidal wave comes closer to the percussion wave peak value when person s age increases becoming shorter than the percussion wave defined from the start to the wave maximum (Fig. 2b and 3b). The integrals of residual errors for each sensor and in different measurement show good relation. At the first, the integral of each residual overlaps and then EMFi and PPG based residuals differ after the systolic phase (Figure 4). In Figure 5 it is shown a typical toe PPG pulse wave which contains at least three components according to the decomposition. The comparison of both the EMFi s and PPG1 s amplitude spectra shows that the modulation frequency disappears or comes close to the carrier frequency as the person s age increases (Fig. 6, 7, and 8). This study shows also that both the EMFi and PPG pulse waveforms can be decomposed to their component waves, namely, percussion wave, tidal wave, dicrotic wave, repercussion, and retidal waves. Also in frequency domain the amplitude spectra of the respective pulse waves contain at least five or six frequency components. Discussion Clinical research is necessary to quantify the ageing effects in relation to the obtained variables and also to explore pulse waveform changes with subject age. However, based on these optical and mechanical measurement methods it was shown the changes in the pulse waveform. Measurement reliability and repeatability is very good provided the EMFi measurements are done by a skilled operator. Pulse waveform analysis of both the PPG and EMFi offers an alternative means of non invasive cardiovascular monitoring, but further both software and hardware development is required to enable user friendly clinical and preclinical measurement and analysis system. In PPG, the sensors are on the opposite sides of the finger tip, however, EMFi needs the good skilled operator (A.V.) attaching the sensor on the patient s wrist by touching gently to obtain waveforms. The elasticity is obtained quantitatively from both EMFi and PPG pulse signals both in time and frequency domain because of the components time interval changes or the spectral changes clearly inspected. However, from time domain, decomposition into logarithmic normal function, differs from that obtained from frequency domain because the latter contains many pulse waves which are IPFM modulated and they frequency components overlap. This information from the pulse wave propagation analysis can t be received. In studies, EMFi and PPG theory should be described when analyzing arterial pulse wave signals because they contain effects of respiration, autonomous nervous activity, gastric mobility, and also arterial properties, i.e., arterial diseases. References [1] Alametsä J, Palomäki A. Comparison of local pulse wave velocity values acquired with EMFi sensor. Finnish Journal of ehealth and ewelfare 2012;4(2):89 98. [2] Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiological Measurement 7;28:R1 R39. [3] Middleton PM et al. Spectral analysis of finger photoplethysmograpic waveform variability in a model of mild to moderate haemorrhage. Journal of Clinical Monitoring and Computing 8;22:343 353. 27.5.2013 FinJeHeW 2013;5(2 3) 66