A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging

Similar documents
Computer applications, PACS, Instrumentation, Equipment, Technology assessment /ecr2012/C-0655

Investigation of Effective DQE (edqe) parameters for a flat panel detector

Fundamental and Clinical Studies for Effectiveness of Zero-filling Interpolation on k-space for Improvement of Sharpness in Magnetic Resonance Imaging

Evaluation of no-grid radiography using the digital scattered x-ray removal processing

Practical guidelines for color calibration and quality assurance of medical displays

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

A study of exposure index value fluctuations in computed radiography and direct digital radiography using multiple manufacturers

Correction of the local intensity nonuniformity artifact in high field MRI

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Studies on reduction of exposure dose using digital scattered X-ray removal processing

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

Improvement of CT image quality with iterative reconstruction idose4

The impact of increasing SID on patient dose in pa abdomen imaging

Dose reduction using Cu-filter for full-spine radiografic examination of patients with adolescent idiopathic scoliosis

Visualization of sources of scattered radiation from x-ray equipment used for interventional radiology

Determining acceptance levels for automatic daily image quality control in magnetic resonance imaging

Lumbar disc height assessment: a comparative study between EOS and digital radiography

Novel cassette-sized, flat-panel digital radiography (DR) system: Initial clinical and workflow results versus computed radiography (CR)

Influence of different iteration levels in fourth generation iterative reconstruction technique on image noise in CT examinations of the neck

Aim. Images for this section: Page 2 of 13

Development of new dosimeter for measuring dose distribution in CT

Difference in signal-to-noise-ratio (SNR) from vertical to horizontal scanner position using a 0,25 Tesla Weightbearing

Worldwide practice of breast MRI: insights from the MIPA study applications - the MIPA study group

Multi-slice computed tomography analysis of bullet trajectory in forensic investigation.

Comparison of high-resolution C-arm cone-beam CT imaging and conventional 3D rotational angiography for the better microvascular visualization

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

PROJECT TIPIRX (integration of teleradiology using Lowcost digitalization of X-ray films): a Brazilian solution

SUSPENSION CRITERIA FOR IMAGE MONITORS AND VIEWING BOXES.

The role of CT in restoration of a very rare Christ wooden statue: a fundamental contribution.

LCD handheld displays characterization by means of the MTF measurement

Image Display and Perception

Suppression of metal artifacts using image-based monoenergetic DECT imaging

White Paper. The Advantages of Hybrid Gamma PXL. 1 Introduction Outline of Hybrid Gamma PXL Study Results...

Optimal speed for dual-energy subtraction neck imaging with flat-panel detector radiography

The feasibility of breath-hold high-resolution 3D-MRCP obtained with 32 channel torso cardiac coil and T2-prepBTFE

Voice recognition versus transcriptionist

Digital Cameras. Consumer and Prosumer

Technical Paper CONSISTENT PRESENTATION OF MEDICAL IMAGES

Display of mammograms on a CRT

Sampling Efficiency in Digital Camera Performance Standards

Assessment of Color Displays. Michael Flynn Radiology Research Henry Ford Health System Detroit, MI

A Novel 12-bit Grayscale Topology PACSmate Revolutionary GrayBoost Technology

Things you may want to know

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

Increasing the Number of Gray Shades in Medical Display Systems How Much is Enough?

A High-Speed Imaging Colorimeter LumiCol 1900 for Display Measurements

Observer Performance of Reduced X-Ray Images on Liquid Crystal Displays

Evaluation of magnetic resonance acoustic noise in 1.5 and 3 Tesla scanners

ABSTRACT 1. PURPOSE 2. METHODS

MTF Analysis and its Measurements for Digital Still Camera

Modified slanted-edge method and multidirectional modulation transfer function estimation

Influence of Image Enhancement Processing on SFR of Digital Cameras

Multi-spectral acoustical imaging

Comparison of new digital mammography systems: physical characterisation and image quality evaluation

The Science Seeing of process Digital Media. The Science of Digital Media Introduction

TI Designs: Biometric Steering Wheel. Amy Ball TIDA-00292

Color Matching Beginning with the Basics. EIZO's Guide to enjoyment. a further. of digital. I see. photography

Preliminary Assessment of High Dynamic Range Displays for Pathology Detection Tasks. CIS/Kodak New Collaborative Proposal

Low Voltage Brushed Motor System

Tomophan TSP004 Manual

Report #17-UR-049. Color Camera. Jason E. Meyer Ronald B. Gibbons Caroline A. Connell. Submitted: February 28, 2017

Clinical and management aspects of digital imaging and PACS

Ideal for display mura (nonuniformity) evaluation and inspection on smartphones and tablet PCs.

Color Science. What light is. Measuring light. CS 4620 Lecture 15. Salient property is the spectral power distribution (SPD)

Image Interpretation System for Informed Consent to Patients by Use of a Skeletal Tracking

There are many types of light sources, ranging from natural sunlight to man-made artificial lightings.

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

Differentiation of Malignant and Benign Masses on Mammograms Using Radial Local Ternary Pattern

Ideal for display mura (nonuniformity) evaluation and inspection on smartphones and tablet PCs.

Color and More. Color basics

k m = 683 lumens/watt Display QC QC Measures for Color Displays joules watts lumens SAM Imaging Education Course

SPECTRAL SCANNER. Recycling

High-Accuracy Luminance & Chromaticity Measurement Comparable to Many Spectroradiometers

Color Measurement with the LSS-100P

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

Multispectral. imaging device. ADVANCED LIGHT ANALYSIS by. Most accurate homogeneity MeasureMent of spectral radiance. UMasterMS1 & UMasterMS2

Digital Photographs, Image Sensors and Matrices

A Study of Slanted-Edge MTF Stability and Repeatability

Term 1 Study Guide for Digital Photography

Achieving 100,000 : 1 contrast measurement

PERFORMANCE MEASUREMENT OF MEDICAL IMAGING SYSTEMS BASED ON MUTUAL INFORMATION METRIC

An Evaluation of MTF Determination Methods for 35mm Film Scanners

Research on 3-D measurement system based on handheld microscope

Fast MTF measurement of CMOS imagers using ISO slantededge methodology

Determination of the MTF of JPEG Compression Using the ISO Spatial Frequency Response Plug-in.

Data Sheet of SAW Components

CAMERA GRAND PRIX 2008 JAPAN

Exploring Texture Pattern Features and Relations to Kansei with 2D FFT - Wallpapers and Dashboard Leather Grain Patterns

Ring-Down versus Comet Tail: Two artifacts uncovered

Texas Instruments. PMP4435 REVA Test Procedure. China Power Reference Design REVA

1 Photo. Bottom side. 11/7/2014 PMP10783 Rev A Test Results

ISO INTERNATIONAL STANDARD. Photography Electronic still-picture cameras Resolution measurements

A Low Noise and High Sensitivity Image Sensor with Imaging and Phase-Difference Detection AF in All Pixels

DISPLAY metrology measurement

Spectroradiometer CS-2000/2000A. The world's top-level capability spectroradiometers make further advances with addition of second model to lineup.

TI Designs TIDA Automotive 1.3M Camera Module Design with OV10640, DS90UB913A and power over Coax Test Data

Small Animal Radiographic Techniques and Positioning COPYRIGHTED MATERIAL

Microtremor Array Measurements and Three-component Microtremor Measurements in San Francisco Bay Area

Data Sheet of SAW Components

SilverFast. Colour Management Tutorial. LaserSoft Imaging

Transcription:

A comparison between medical-grade liquid crystal display (LCD) and ipad color imaging Poster No.: C-1377 Congress: ECR 2014 Type: Scientific Exhibit Authors: K. Yoshimura 1, K. Shibata 2, T. Nihashi 1, H. Ohse 3, H. Toyama 4 ; 1 2 3 4 Nagoya/JP, Toyota/JP, Togo-cho/JP, Toyoake/JP Keywords: DOI: Quality assurance, Technology assessment, PACS, Computer applications 10.1594/ecr2014/C-1377 Any information contained in this pdf file is automatically generated from digital material submitted to EPOS by third parties in the form of scientific presentations. References to any names, marks, products, or services of third parties or hypertext links to thirdparty sites or information are provided solely as a convenience to you and do not in any way constitute or imply ECR's endorsement, sponsorship or recommendation of the third party, information, product or service. ECR is not responsible for the content of these pages and does not make any representations regarding the content or accuracy of material in this file. As per copyright regulations, any unauthorised use of the material or parts thereof as well as commercial reproduction or multiple distribution by any traditional or electronically based reproduction/publication method ist strictly prohibited. You agree to defend, indemnify, and hold ECR harmless from and against any and all claims, damages, costs, and expenses, including attorneys' fees, arising from or related to your use of these pages. Please note: Links to movies, ppt slideshows and any other multimedia files are not available in the pdf version of presentations. www.myesr.org Page 1 of 13

Aims and objectives The aim of this study was to evaluate the performance of colored monitors on which medical colored image was displayed, comparing medical-grade LCD and the ipad. To evaluate the performance of the monitor, we measured luminance [1], chromaticity diagram[2], modulation transfer function (MTF) [3], and noise power spectrum (NPS) [4]. Methods and materials Monitor and Tablet PCs A medical-grade LCD (RX220, Eizo Corp., Ishikawa, Japan) was used for this experiment, and grayscale calibration was adjusted to grayscale standard display function (GSDF). The grayscales of the GSDF using calibration software (RadiCS, Eizo Corp., Ishikawa, Japan) were used and a sensor for the monitor (UX1 Sensor, Eizo Corp., Ishikawa, Japan) on the basis of the gray-scale standard display function curve of digital imaging and communications in medicine (DICOM) part 14 [1]. The ipads (Apple Inc., Cupertino, CA) used were of the first generation (ipad-1) and fourth generation (ipad-4). The manufacturer-specified characteristics of the monitor and tablet PC are summarized in Table 1. Creating Measurement Images The image for measurement of color patterns and MTF were made with Photoshop CS6 (Adobe Systems Inc., San Jose, CA). Each luminance and NPS color image consisted of a square of 450 450 pixels, and red, blue, and green color patterns. The black (TG18- LN8-01) and white (TG18-LN8-18) image is the TG18-LN pattern[5]. The MTF image of bar patterns consists of five segments, including five bars with 1-, 2-, 3-, 4-, and 5-pixel widths, and two uniform areas for measuring the bar and the background intensity located at just over and under the bar patterns. The frequency corresponding to each segment has a 1, 0.5, 0.333, 0.25, and 0.2 Nyquist frequency of the display, respectively [3],[6]. As for the color of the MTF image, the background is black, and the color bar patterns are white, red, blue, and green. Luminance Measurement In the three displays (medical-grade LCD, ipad-1, and ipad-4), luminance performance was measured using a CS-100A luminance and color meter (Konica Minolta Holdings Inc., Tokyo, Japan) pointed at the monitor center in a horizontal direction. The TG18-LN Page 2 of 13

pattern [5] and color patterns were placed at the display center, and the distance between the luminance meter and the display center was set at 100 cm on axis. The values of luminance and chromaticity for each pattern were determined by the averages of five measurements, all of which were performed in a darkroom (ambient lighting of zero lux). The luminance data were analyzed according to the AAPM TG18 methodology [5]. The chromaticity diagram was created with the chromaticity values of color pattern images. MTF and NPS Measurement The MTF and NPS were measured referring to a MTF and NPS image method proposed by Ichikawa et al. [3; 4; 6]. In the three displays (medical-grade LCD, ipad-1, and ipad-4), the MTF and NPS images were captured by a high-resolution single-lens reflex (SLR) digital camera (D90, Nikon Corporation, Tokyo, Japan) used with a close-up lens (Micro- Nikkor 60 mm F2.8D, Nikon Corporation, Tokyo, Japan) positioned at the monitor center in a horizontal direction. The camera has a Complementary Metal Oxide Semiconductor (CMOS) image sensor array with 4288 2848 elements and a contrast resolution of 12 bits. Image acquisition was performed five times, and the data were averaged. The inherent noise characteristics of the medical-grade LCD and ipads used an NPS. A trend removal was applied to the averaged 256-point profiles by second-order polynomial fitting to correct for the background-luminance trend. We calculated NPS in the horizontal and vertical directions using a one-dimensional fast Fourier-transform (FFT) method from the background corrected profiles. Hanning window processing was applied to the FFT values to reduce spectral leakage errors. Images for this section: Page 3 of 13

Table 1 Page 4 of 13

Results Fig. 1 on page 5 shows a comparison of the grayscale curve among the three displays (medical-grade LCD, ipad-1, and ipad-4). The luminance response of the ipad-4 was almost the highest of the three. However, medical-grade LCD was highest in the three displays from Pattern 1 to 3 (TG18-LN8-01 to 03). Specifications for medical-grade liquid crystal display (LCD), along with first generation (ipad-1) and fourth generation (ipad-4) ipads (Apple), are given in Table 1. Fig. 2 on page 6 shows a comparison of the contrast response the three displays (medical-grade LCD, ipad-1, and ipad-4). The LCD fits DICOM curve, but two ipods do not fit it. Because ipads cannot do calibration with GSDF. Fig. 3 on page 7 shows a comparison of the chromaticity diagram among the three displays (medical-grade LCD, ipad-1, and ipad-4). As for the chromaticity diagram, the area of ipad-4 was slightly larger than that of LCD. The characteristic of the chromaticity diagram was similar pattern to the ipad-4 and LCD. On the other hand, the area of ipad-1 was most narrow. Fig. 4 on page 8 shows a comparison of the MTF among the three displays (medicalgrade LCD, ipad-1, and ipad-4), using monochromic bar-pattern. The MTF was highest for ipad-4, followed by ipad-1 and LCD. Fig. 5 on page 9 shows a comparison of the MTF among color images (red, blue, and green), using colored bar-pattern. The MTF of green was the highest among three monitors. The MTF of each color was similar, regarding LCD. Fig. 6 on page 10 shows a comparison of the NPS of a black image among the three displays (medical-grade LCD, ipad-1, and ipad-4). The NPS of ipad-4 was the highest, followed by LCD, and ipad-1. Fig. 7 on page 11 shows a comparison of the NPS among color images (red, blue, green, and white). Among the NPS of different colors, white was the highest for all three monitors. There was no difference not only among the three displays, but also among three colors in each monitor. Images for this section: Page 5 of 13

Fig. 1: This graph showed the comparison of luminance response among three monitor types. Page 6 of 13

Fig. 2: This graph showed the comparison of contrast response among three monitor types. Page 7 of 13

Fig. 3: This graph showed the comparison of chromaticity diagram among three monitor types. Page 8 of 13

Fig. 4: This graph showed the comparison of modulation transfer function (MTF) among three monitor types. The bra pattern color is black. Page 9 of 13

Fig. 5: This graph showed the comparison of modulation transfer function (MTF) among bar pattern colors (red, blue, and green). Page 10 of 13

Fig. 6: This graph showed the comparison of noise power spectrum (NPS) among three monitor types. The color pattern is black. Page 11 of 13

Fig. 7: This graph showed the comparison of noise power spectrum (NPS) among color patterns (red, blue, green, and white). Page 12 of 13

Conclusion The ipad-4 has the better luminance, chromaticity, MTF, and NPS than medical-grade LCD or ipad-1. With these excellent characteristics, it might be available for the diagnosis use of the color image with ipad-4 in clinical situation. However, further studies would be warranted to elucidate the role in clinical situation, using actual clinical images, for example on the receiver operating characteristic (ROC) analysis by several radiologists. Personal information Kumiko Yoshimura, RT., PhD. Graduate School of Medicine, Nagoya University 1-1-20 Daikominami, Higashi-ku, Nagoya 461-8673 Japan References 1. (2011) National Electrical Manufacturers Association. Digital Imaging and Communications in Medicine (DICOM). Part 14. Grayscale standard display function. http://medical.nema.org/dicom/2011/11_14pu.pdf. 2. (1932) Commission Internationale de l'eclairage Proceedings 1931. Cambridge: Cambridge University Press, CIE 3. Ichikawa K, Kodera Y, Fujita H (2006) MTF measurement method for medical displays by using a bar-pattern image. Journal of the Society for Information Display 14:831-837 4. Ichikawa K, Kodera Y, Nishimura A et al (2008) Analysis method of noise power spectrum for medical monochrome liquid crystal displays. Radiol Phys Technol 1:201-207 5. (2005) American Association of Physicists in Medicine Task Group 18 Imaging Informatics Subcommittee. http://www.aapm.org/pubs/reports/ OR_03.pdf. 6. Ichikawa KI, Takayuki (2010) Image quality measurement of digital radiography. Ohmsha Page 13 of 13