When is it Time to Transition to a Higher Bandwidth Oscilloscope?

Similar documents
Agilent Spectrum Visualizer (ASV) Software. Data Sheet

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

Evaluating Oscilloscope Bandwidths for your Application

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

Keysight Technologies Minimum Required Sample Rate for a 1-GHz Bandwidth Oscilloscope

N2820A/21A High-Sensitivity, High Dynamic Range Current Probes

Educator s Oscilloscope Training Kit for the InfiniiVision 2000 & 3000 X-Series

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

Techniques to Achieve Oscilloscope Bandwidths of Greater Than 16 GHz

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

N9051A Pulse Measurement Software

Agilent N9310A RF Signal Generator. All the capability and reliability of an Agilent instrument you need at a price you ve always wanted

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

Choosing an Oscilloscope with the Right Bandwidth for your Application

N2750A/51A/52A InfiniiMode Differential Active Probes

Agilent 2-Port and 4-Port PNA-X Network Analyzer

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

Keysight Technologies, Inc. Overcome PCB Loss and Deliver a Clean Eye to Your DUT Using Multi-tap De-emphasis

Agilent N4916B De-emphasis Signal Converter

Agilent U1730C Series Handheld LCR Meters

Solar Array Simulation System Integration

Agilent NFA Noise Figure Analyzer

Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note

Agilent 8761A/B Microwave Switches

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

Agilent InfiniiMax III probing system

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

Agilent U1881A and U1882A Power Measurement Application for Agilent InfiniiVision and Infiniium Oscilloscopes

Keysight Technologies Accurate Evaluation of MEMS Piezoelectric Sensors and Actuators Using the E4990A Impedance Analyzer.

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

Evaluating Oscilloscopes for Low-Power Measurements

Two-Way Radio Testing with Agilent U8903A Audio Analyzer

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet

InfiniiMax III probing system


Keysight Technologies RS-232/UART Protocol Triggering and Decode for Infiniium 9000A and 9000 H-Series Oscilloscopes. Data Sheet

Keysight N8803C CAN, LIN, FlexRay, and CAN-FD Protocol Triggering and Decode Software. Data Sheet

Agilent E5061B Network Analyzer. 100 khz to 1.5 GHz/3 GHz 5 Hz to 3 GHz

Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer.

Keysight N8836A PAM-4 Measurement Application For Infiniium S-Series, 90000A, V-Series, X-Series, Q-Series, and Z-Series Oscilloscopes

Introduction. Part 1. Introduction...2

Keysight Technologies Making Current-Voltage Measurement Using SMU

A Time-Saving Method for Analyzing Signal Integrity in DDR Memory Buses

Agilent N9342C Handheld Spectrum Analyzer (HSA)

Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes. Data Sheet

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

MXG X-Series Signal Generator N5183B Microwave Analog

Agilent 8762F Coaxial Switch 75 ohm

Agilent N6780 Series Source/Measure Units (SMUs) for the N6700 Modular Power System

Agilent N9340B Handheld Spectrum Analyzer (HSA)

Keysight Technologies InfiniiScan Event Identification Software

Data Sheet. Agilent M9185A PXI Isolated D/A Converter. DISCOVER the Alternatives... Agilent MODULAR Products. 8/16-Channel 16-bit, ±16 V

Keysight Technologies How to Select the Right Current Probe. Application Note

Keysight Technologies CAN, LIN and FlexRay Protocol Triggering and Decode for Infiniium 9000 and S-Series Oscilloscopes.

Agilent Correlation between TDR oscilloscope and VNA generated time domain waveform

Keysight Technologies Achieving Accurate E-band Power Measurements with E8486A Waveguide Power Sensors. Application Note

Advanced Test Equipment Rentals ATEC (2832)

Solutions for Solar Cell and Module Testing

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option

Keysight Technologies Oscilloscope Probe Loading Experiment

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

Agilent NFA Noise Figure Analyzer

How Offset, Dynamic Range and Compression Affect Measurements

Keysight Technologies Enhance EMC Testing with Digital IF. Application Note

Keysight Technologies Educator s Oscilloscope Training Kit for InfiniiVision X-Series Oscilloscopes. Data Sheet

Evaluating Oscilloscope Sample Rates vs. Sampling Fidelity

Keysight Technologies MATLAB Data Analysis Software Packages

UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer

Keysight Technologies Secondary Radar Transponder Testing Using the 8990B Peak Power Analyzer. Application Note

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Agilent Agilent 34405A Multimeter 5.5 Digit Dual Display, Benchtop DMM More Capabilities at a Value Price. Data Sheet

MIL-STD 1553 Triggering and Hardwarebased Decode (Option 553) for Agilent s InfiniiVision Series Oscilloscopes

Essential Capabilities of EMI Receivers. Application Note

Agilent E8257D PSG Microwave Analog Signal Generator

Keysight Technologies 85072A 10-GHz Split Cylinder Resonator. Technical Overview

Keysight 8762F Coaxial Switch 75 ohm

Agilent First-generation MXG Signal Generators

Keysight Technologies NFC Device Turn-on and Debug

Agilent MXG Signal Generators

Keysight Technologies U9391C/F/G Comb Generators. U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview

Agilent N8480 Series Thermocouple Power Sensors. Technical Overview

Agilent E4438C/E8267D Option 422 Scenario Generator for GPS

Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight Technologies Z9070B Wideband Signal Analysis Solution. Technical Overview

MXG Analog Signal Generator Express Configurations

Agilent U1230 Series Handheld Digital Multimeter (DMM)

Agilent 84904, 6, 7K/L Programmable Step Attenuators

Agilent N9342C Handheld Spectrum Analyzer (HSA)

Agilent U9391C/F/G Comb Generators

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found

Keysight Technologies, Inc. UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer. Application Note

Agilent N9344C/N9343C/N9342C Handheld Spectrum Analyzer (HSA)

Agilent NFA Noise Figure Analyzer

Agilent MXG Signal Generators

Agilent AN Balanced Circuit Measurement with an Impedance Analyzer/LCR Meter/Network Analyzer Application Note

Keysight Technologies Triggering on and Decoding the PSI5 Sensor Serial Bus

Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters. Data Sheet

N8900A InfiniiView Oscilloscope Analysis Software. Data Sheet

Transcription:

When is it Time to Transition to a Higher Bandwidth Oscilloscope? Application Note When purchasing an oscilloscope to test new designs, the primary performance specification that most engineers consider first is the scope s bandwidth. Another key consideration is always price. Unfortunately, there is a fairly linear relationship between these two characteristics of a scope; bandwidth and price. Since most of us must work within constrained capital equipment budgets, engineers tend to purchase test equipment that has just enough performance to meet their current needs. But with each new design-start, performance requirements often increase in order to stay competitive and to deliver newer, faster, and more feature-rich smart products that customers demand. So how do we determine how much bandwidth is required for today s projects, and when do we know when it is time to move up? Although many of today s embedded designs are mixedsignal in nature (contain both analog and digital signals), it is usually the maximum speed of the digital signals (clock rate and edge transition times) within a design s central processing unit (CPU) system that determines how much oscilloscope bandwidth is required. And it is often the timing specifications of CPU memory that must be considered. The most common type of memory used in most of today s embedded designs is double data rate (DDR) memory. This type of memory clocks data into and out of memory on both the rising and falling edges of the clock signal. Let s assume that in your last design project the CPU system in your embedded design was based around DDR1 memory technology with data transfer rates in 200 Mbps range. And if you used a 500-MHz bandwidth scope to capture and verify critical timing parameters of your CPU system running at this speed, this scope probably provided you with sufficient measurement accuracy. But now your next design will be based on higher speed DDR2 technology. Will you need to purchase a higher bandwidth scope? And how much bandwidth will be required?

Capturing Eye-Diagrams on DDR2-667 Memory A very common measurement performed on read and write data signals of DDR memory is an eye-diagram measurement. Eye-diagrams displayed on a scope provide a composite picture of signal quality by overlaying high and low bits to determine when data is valid. Figure 1 shows an example of an eye-diagram measurement on a data signal representative of DDR2-667 memory using an Agilent 500- MHz bandwidth oscilloscope (MSOX3054A). While the clocking rate of DDR2-667 memory is 667 MHz with clocking of the data occurring on both the rising and falling edges of the clock signal, the actual frequency of the clock signal is 333 MHz. And although the maximum specified data transfer rate of this type of memory is 667 Mbps, the maximum toggle rate of the data signal is also 333 MHz based on a maximum repeating 1-0-1-0-1 serial pattern. So one might assume that since the maximum signal switching rate of either the clock or data signal is just 333 MHz, a 500-MHz bandwidth scope would provide sufficient performance to capture and make accurate measurements on these signals. But as you can see in Figure 1, a 500-MHz bandwidth scope basically turns our digital bit stream into what appears to be an overlay of sinusoids. A 500-MHz bandwidth scope is not able to capture much beyond the 1st harmonic. This is why it looks sinusoidal. So clearly, capturing these signals with a 500- MHz bandwidth scope is insufficient. Figure 2 shows the same DDR2-667 data signals captured as an eye-diagram display using Agilent s new MSOX3104A oscilloscope, which has a specified bandwidth of 1 GHz. We can now more clearly see the shapes and details of the overlaid digital bits using this higher bandwidth scope. So for this particular DDR2 measurement application, it is time to move up in bandwidth to 1 GHz. Figure 1: Capturing DDR2-667 data signals using an Agilent 500- MHz bandwidth scope. Figure 2: Capturing DDR2-667 data signals using an Agilent 1-GHz bandwidth scope. 2

Determining Required Bandwidth For digital applications, the amount of required oscilloscope bandwidth depends on the speed of the fastest signal transitions in your designs that need to be measured. Once you know what the fastest edge speeds are in your system, you can then convert these edge speeds into a maximum practical frequency. We will call this f MAX. But don t confuse this f MAX with the f MAX that Nyquist refers to. The reason I say maximum practical frequency is because fast transitions are theoretically composed of an infinite spectrum of frequencies. However, there comes a point where the higher components of frequencies contribute insignificantly into creating the shape of a fast edge. Once we know what f MAX is, we can then determine the required bandwidth. So how do you determine what the fastest edge transitions are? It appears that we have a chicken & egg situation here. Digital devices utilized in today s designs typically have input and/or output rise time specifications. These rise time specifications (RT) have traditionally been specified relative to 10% to 90% threshold criteria. However, more of today s digital devices specify their rise times relative to 20% to 80% threshold criteria. Once you have determined the fastest rise times of your devices, use one of the following simple formulas to compute f MAX 1. Converting Rise Time (RT) to f MAX f MAX = 0.5/RT (10-90) f MAX = 0.4/RT (20-80) Once you ve determined f MAX, you could use a scope with a specified bandwidth of this same frequency to capture and perform rise time measurements with reasonable accuracy perhaps 20% error. But if you wanted to perform rise time measurements with 5% accuracy or better, then you should select a scope with a specified bandwidth that is 1.5 to 2 times higher than f MAX. Converting f MAX to Scope Bandwidth BW = 2 x f MAX Let s now walk through a simple example. Assume that you ve determined that the fastest rise times in your latest design are 800 ps (based on 20% to 80% threshold criteria). f MAX is then 0.4/800ps = 500 MHz. If your budget is limited and you can also tolerate up to 20% timing error when performing rise time measurements, then you could use a 500-MHz bandwidth scope for your measurement applications. In other words, if you attempt to measure the rise time of an 800 ps edge, the scope might measure something more in the range of 1 ns. But if you need more accuracy than this, then perhaps a 1-GHz bandwidth scope would be a better choice. To learn more about how to select the appropriate bandwidth scope for your specific measurement applications, download Agilent s application note titled, Evaluating Oscilloscope Bandwidths for your Application listed at the end of this document. You may discover that some of your devices specify a maximum signal slew rate (SR), as opposed to specifying a rise time. This is true for most DDR memory devices. If this is the case, then you can approximate the 20% to 80% rise time using the following formula: Converting Slew Rate (SR) to Rise Time (RT) RT (20-80) = 0.6 (V H - V L )/SR 3

Summary As you saw in the first part of this application note, if you are attempting to capture and measure high-speed digital signals using your current oscilloscope, but captured waveforms are beginning to look more sinusoidal as opposed to digital in nature, then it s probably time to move up in oscilloscope bandwidth. But there is a more accurate method to determine the amount of bandwidth you need besides just evaluating the shapes of captured waveforms. Based on the fastest edges that you need to measure, you can convert edge speed into maximum practical frequency (f MAX ). You can then select a scope with a bandwidth specification ranging from f MAX to 2X f MAX depending upon the level of accuracy you require as well as the amount capital equipment budget you have. If you think it s time to move up in bandwidth, Agilent s newest 1-GHz bandwidth oscilloscopes in the InfiniiVision 3000 X-Series start at the lowest prices in the industry. In addition, if you already own one of Agilent s lower bandwidth 3000 X-Series scopes (100 MHz, 200 MHz, 350 MHz or 500 MHz bandwidth models); your investment has been protected. The 3000 X-Series oscilloscopes are fully upgradable including bandwidth up to the new 1-GHz performance level. The performance of these oscilloscopes grows as your measurement needs grow from design-start to design-start. 1 High-Speed Digital Design, A Handbook of Black Magic, Howard Johnson, Martin Graham, 1993, Prentice Hall PTD, Prentice-Hall, Inc, Upper Saddle River, New Jersey 07458 4

Related Literature Puiblication Title Publication Type Publication Number Agilent InfiniiVision 3000 X-Series Oscilloscopes Data sheet 5990-6619EN Agilent InfiniiVision 7000 Series Oscilloscopes Data sheet 5990-4769EN Agilent Infiniium 9000 Series Oscilloscopes Data sheet 5990-3746EN Evaluating Oscilloscope Bandwidths for your Applications Application note 5989-5733EN Evaluating Oscilloscope Sample Rates vs Sampling Fidelity Application note 5989-5732EN Understanding Oscilloscope Frequency Response and its Effect on Rise Time Accuracy Application note 5988-8008EN Advantages and Disadvantages of Using DSP Filtering on Oscilloscope Waveforms Application note 5989-1145EN Evaluating Oscilloscopes for Best Waveform Update Rates Application note 5989-7885EN To download these documents, insert the publication number in the URL: http://cp.literature.agilent.com/litweb/pdf/xxxx-xxxxen.pdf Agilent Technologies Oscilloscopes Multiple form factors from 20 MHz to >90 GHz Industry leading specs Powerful applications 5

TM www.agilent.com www.agilent.com/find/3000x-series Agilent Email Updates www.agilent.com/find/emailupdates Get the latest information on the products and applications you select. www.axiestandard.org AdvancedTCA Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Agilent is a founding member of the AXIe consortium. www.lxistandard.org LAN extensions for Instruments puts the power of Ethernet and the Web inside your test systems. Agilent is a founding member of the LXI consortium. www.pxisa.org PCI extensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. Agilent Channel Partners www.agilent.com/find/channelpartners Get the best of both worlds: Agilent s measurement expertise and product breadth, combined with channel partner convenience. Agilent Advantage Services is committed to your success throughout your equipment s lifetime. We share measurement and service expertise to help you create the products that change our world. To keep you competitive, we continually invest in tools and processes that speed up calibration and repair, reduce your cost of ownership, and move us ahead of your development curve. www.agilent.com/find/advantageservices www.agilent.com/quality Windows is a U.S. registered trademark of Microsoft Corporation. For more information on Agilent Technologies products, applications or services, please contact your local Agilent office. The complete list is available at: www.agilent.com/find/contactus Americas Canada (877) 894 4414 Brazil (11) 4197 3600 Mexico 01800 5064 800 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 375 8100 Europe & Middle East Belgium 32 (0) 2 404 93 40 Denmark 45 45 80 12 15 Finland 358 (0) 10 855 2100 France 0825 010 700* *0.125 /minute Germany 49 (0) 7031 464 6333 Ireland 1890 924 204 Israel 972-3-9288-504/544 Italy 39 02 92 60 8484 Netherlands 31 (0) 20 547 2111 Spain 34 (91) 631 3300 Sweden 0200-88 22 55 United Kingdom 44 (0) 118 927 6201 For other unlisted countries: www.agilent.com/find/contactus Revised: January 6, 2012 Product specifications and descriptions in this document subject to change without notice. Agilent Technologies, Inc. 2012 Published in USA, May 18, 2012 5991-0027EN