This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

Similar documents
1. Preliminary sample preparation

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

ALTURA EDS. Rev. 0915

Using the Hitachi 3400-N VP-SEM

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

Fire CR Calibration Guide

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

SEM Training Notebook

RENISHAW INVIA RAMAN SPECTROMETER

SOP: EDAX Eagle III Microspot XRF

Renishaw InVia Raman microscope

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

RAITH e-line OPERATING INSTRUCTIONS

ThermaViz. Operating Manual. The Innovative Two-Wavelength Imaging Pyrometer

SEM Training Notebook

FEI Quanta 200 ESEM Basic instructions

Figure 1 The Raith 150 TWO

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

JEOL 6500 User Manual

Model SU3500 Scanning Electron Microscope

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

2. Raise HT to 200kVby following the procedure explained in 1.6.

JEOL 6700 User Manual 05/18/2009

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

FEI Tecnai G 2 F20 Operating Procedures

Confocal Raman Microscopy (WITec Alpha 300R)

STEM Spectrum Imaging Tutorial

ZEISS EVO SOP. May 2017 ELECTRON OPTICS

UNIVERSITY OF WATERLOO Physics 360/460 Experiment #2 ATOMIC FORCE MICROSCOPY

This document contains work instructions related to utilizing the dental imaging application, XrayVision version 4.0.

Operating Rausch ScanCam within POSM.

Operating Procedures for MICROCT1 Nikon XTH 225 ST

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

LSM 510 Meta Training Notes

Digital Photography 1

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0

LSM 510 Training Notes

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

FE-SEM SU-8020 Operating manual (Preliminary version)

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Dickinson College Department of Geology

EE 210 Lab Exercise #3 Introduction to PSPICE

Minolta Scanner Plugin

NSS Spectral Imaging System Handbook Version 3.0

Microprobe Power User Part 3: X-Ray Maps

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Photoshop CS2. Step by Step Instructions Using Layers. Adobe. About Layers:

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

ScanArray Overview. Principle of Operation. Instrument Components

AgilEye Manual Version 2.0 February 28, 2007

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Digital Imaging - Photoshop

IBIL setup operation manual for SynerJY software version

Nanoscale Fabrication & Characterization Facility. Raith e-line EBL Users Guide (updated:aug 2 nd, 2017)

Photoshop CS6 Basics. Using Layers to Create a Magazine Cover

Laboratory Experiment #1 Introduction to Spectral Analysis

ISCapture User Guide. advanced CCD imaging. Opticstar

ADOBE PHOTOSHOP CS 3 QUICK REFERENCE

Introduction of New Products

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Olympus LEXT OLS 4000 Confocal Laser Microscope

ISIS A beginner s guide

Toothbrush Holder. A drawing of the sheet metal part will also be created.

Table of Contents. Lesson 1 Getting Started

Generations Automatic Stand-Alone Lace By Bernie Griffith Generations Software

Things to check before start-up.

Creating Actions in Adobe Photoshop Steve Everist, CLPE King County Sheriff s Office Seattle, WA

Lab 3: Low-Speed Delta Wing

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

MicroLab 500-series Getting Started

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Zeiss 880 Training Notes Zen 2.3

Name: Date Completed: Basic Inventor Skills I

Gravitational Lensing Experiment

Introduction to QTO. Objectives of QTO. Getting Started. Requirements. Creating a Bill of Quantities. Updating an existing Bill of Quantities

Reference and User Manual May, 2015 revision - 3

LowDose for JEOL. --Software LowDose and Advanced Data Acquisition. Version 1.1

Bruker D8 HRXRD Collecting X-Ray Reflectivity Data using the PathFinder Detector

Amray 3600 FESEM. Standard Operating Procedure. v2.2 modified by Bryan Cord. General Notes...3. Sample Loading...5. System Loading...

Introduction to Layers

Quick and simple installation and no maintenance needed. 3 Times More affordable Than a normal SEM. Obtaining results in less than 4 minutes

Functions of the SEM subsystems

Publish Drawing Sets with Ease in AutoCAD 2004 Speaker: Heidi Hewett Assistants: Shaan Hurley & Matt Murphy

OPERATING INSTRUCTIONS

This manual describes the Motion Sensor hardware and the locally written software that interfaces to it.

FEI Falcon Direct Electron Detector. Best Practice Document

EPMA General Operations

EinScan-SE. Desktop 3D Scanner. User Manual

DBSP Observing Manual

Copyright 2014 SOTA Imaging. All rights reserved. The CLIOSOFT software includes the following parts copyrighted by other parties:

Transcription:

Tescan MIRA3 SEM: EDS using EDAX TEAM Nicholas G. Rudawski ngr@ufl.edu Cell: (805) 252-4916 Office: (352) 392-3077 Last updated: 12/04/17 This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface. 1. Sample mounting, preparation, and constraints 1.1. Specimens for EDS should be mounted on standard 3 mm-thick SEM pin stubs (Ted Pella #16111, diameter ~13 mm or Ted Pella #16144, diameter ~25 mm). Ideally, the thickness to footprint ratio of the specimen should be kept as small as possible (subject to the above constraints) as this will increase specimen stability. Ideally, the specimens should be flat to prevent shadowing effects, but rough specimens can still be used (at least for qualitative purposes). 1.2. All specimens should be mounted on the aforementioned standard SEM stubs using conductive tape; do not use any type of paint for mounting as this can create a source of contamination. If your sample is non-conductive, it should be given a light C coat (few nm) after being mounted on a stub to ensure a path to ground is produced. 2. Specimen loading 2.1. Log on to MiraTC, vent the chamber, and open the chamber door. 2.2. Specimens should be loaded in the 7 position of the carousel; if you will only be performing EDS, it is not necessary to use any type of stub extender. 2.3. Close the chamber door and evacuate the chamber. 1

3. Beam and detector settings for EDS 3.1. The beam voltage should be set at 2 3 times the energy of the highest energy X-ray you foresee having to analyze in your specimen. The beam intensity should be set to a value of 15 20 to generate a high current probe to produce sufficient counts for EDS. 3.2. The SE or BSE detector may be used when performing EDS. Set the desired detector for Channel A in the SEM Detectors & Mixer panel. Then plug the ACD1 cable into the appropriate channel on the instrument control rack: SIG A for the SE detector (the default) or SIG B for the BSE detector. 3.3. Once appropriate beam and detector settings have been selected, find a region of interest on the specimen (ideally, away from the specimen edges); then focus the image and set WD = 15 mm in the Stage Control panel (this is the optimal WD for performing EDS). 3.4. Perform the basic alignment: auto gun centering, beam centering, and astigmatism correction. When finished, set the magnification as needed and then select to turn off the CCD camera and close the Chamber View panel (otherwise, the EDS detector will not function properly). 4. Starting TEAM and setting up your project file 4.1. You must be logged on to MiraTC before starting TEAM or TEAM will not function properly; do not attempt to open TEAM without being logged on to MiraTC. 4.2. Open Device Manager and expand Network Adapters. 4.3. If any of the adapters displays an error icon, then the adapter needs to be disabled and then enabled again. 2

4.4. Open the TEAM software. A splash screen will pop up as the hardware is initialized (listen for the motor on the EBSD camera to turn on). 4.5. Enter your username into the Log In dialogue box and then select EDS; then enter in your password and select Log In. 4.6. In the Project dialogue box, go to Create and select Project to create a new blank project; this will bring up the Create Project dialogue box, where you can name the project and specimen. 3

5. Cooling the detector 5.1. Expand the Advanced Properties tab (right side of the window); then select EDS detectors ; then Octane Pro, and finally Det 1 Detector Status. Select to start cooling the detector. The button will first turn yellow and finally green and read when cooling is complete. 6. Spectrum Only mode; adjusting spectrum collection settings 6.1. Make sure a live image is actively being acquired from the area of interest. 6.2. Select Spectrum Only from the top menu bar. Then hover over Collect Spectrum to see the options for detector settings and set EV/Chan = 5 and Amp Time = 7.68 ; these settings will give the best energy resolution for the detector and will not need to be adjusted again during the session. 6.3. Continue hovering over Collect Spectrum and set Limit By = Live and then select a desired time (50 s is usually sufficient) then select to start acquiring a spectrum (acquisition will stop automatically and the spectrum will automatically save to the project under Added Spectra ). 6.4. To save copies of the spectrum (both image and excel versions) to your SE FSD designated folder, select from the left side of the spectrum window and then. 4

7. Point Analysis mode 7.1. Select from the top menu bar; then hover over Image Area to see the options for Matrix Size (image resolution). When ready, select to acquire an image and generate an analysis area in the project tree. 7.2. Hover over the button (the same collection settings used for Spectrum Only mode will automatically load); under, you can select options to analyze single or multiple points. If the Single option is selected, just click on the image (or click and drag to define a rectangular region) and analysis will start automatically (the spectrum will save to the project automatically once complete) 5

7.3. If is selected, click on all the points on the image (or click and drag to define rectangular regions) you want to analyze and then select to start the analysis. A spectrum will be collected from each selected point/region in the order of being drawn and will be automatically saved to the project once complete. 7.4. Double clicking on the area in the project tree used for point analysis will pull up the reference SEM image used for point analysis with all spots/regions used for analysis indicated. To save this image to your designated folder, hover over the image and select the button (top left side of the image); the spots/regions used for point analysis will be saved on the image. 6

7.5. To show the spectrum from an individual spot/region, simply double click on the spot/region in the project tree; the SEM image will now only show this analysis spot/region (again, use the button if you want to save the image). Copies of the spectrum (both image and excel versions), can once again be saved to your designated folder by selecting from the left side of the spectrum window and then Send to Folder. 8. Mapping mode 8.1. Select from the top menu bar; the image acquired for the currently activated analysis area (if you just performed point analysis, for example) will appear. If you wish to move to a different area on the specimen and/or change magnification for mapping, do so and then select to acquire a new image and generate a new analysis area in the project. If you wish to map on the same analysis area just used, you do not need to do this as the mapping data will be saved to this analysis area. 8.2. Mapping can be performed over all or part of the imaged area. The default setting is to collect a map over the whole area (shown at left); to adjust this area, click and drag the corners of the green defining box and move it as desired (shown at right). 7

8.3. Hover over the button to see the options for mapping. Select an option for Resolution ; as the resolution increases, more points will be mapped (as indicated in the lower right corner of the green defining box), so the time required to performing mapping will increase. Then select Display Options ; make sure Assigned element colors while mapping and Enable frame count on map are both checked, and make sure under Palette that Color Palette is selected. Select the button to start mapping; the system will auto identify the elements present in the region and then begin mapping. 8.4. A live phase map will be produced as the beam is scanned across the image area; the system will assign a unique to color to each pixel in the image it considers to be of a particular phase (composition). 8.5. To change to a composite map of the individual elements, select E2P from the compass (lower left corner of the window). 8

8.6. To add or remove elements for mapping, first make sure E2P is selected from the compass, then select next to Elements (above the compass). You can then add or remove elements using the interactive periodic table. Additionally, by expanding the ZAF/Peak Selection tab on the right side of the periodic table, you can also choose which peaks you want to use to map for each element (Kα versus Lα, etc.). 8.7. To change the color of a particular element, double click on the element in the pie chart above the compass and then select a new color for the element from the color palette, then click on the section of the pie chart one more time to add the element back to the composite map (the section should expand outward). 9

8.8. If you wish to turn off specific elements in the composite map, simply click once on the expanded section of the pie chart corresponding to the element (if you want to add it back after turning it off, just double click on the section again). 8.9. Mapping will run for the amount of time automatically determined by the system or can be stopped manually prior to completion by selecting manually), select. Once mapping has completed (or has been stopped to save the map. 8.10. Double clicking on the area in the project tree used for mapping will pull up the reference SEM image used for mapping. To save this image to your designated folder, hover over the image and select the side of the image). button (top left 10

8.11. Double clicking on the map in the project tree will pull up the mapping data for this area, where you can then view the phase map, composite element map, and individual element maps. Once a map has been selected, hover over the map and select the button (top left side of the image) to save the map to your designated folder. 8.12. A sum spectrum from the whole analysis area will also come up, which can be saved as described previously. Additionally, by clicking and dragging on any map to create an area, the portion of the sum spectrum isolated from this area will be shown (which can, again, be saved as previously described); the selected area on the image will be indicated. If you want to save the image with the selected area indicated, this must be done via screen capture, rather than by using the button. 11

8.13. The as-collected maps measure intensity without any background subtraction (Phase Map) for the selected elements. However, it is possible to replot the maps as intensity with background subtraction (Net Intensity), atomic/weight % (ZAF), and/or with more/fewer elements. Double click on the map in the project; then hover over and select the desired type of map under Map Type ; select to rebuild the maps using the selected map type; you will be prompted to add/remove any elements you choose before starting. Once the rebuilt map set is complete, it will now show up in the project under the original map (the data can be saved similarly as described for the original map). 9. Line Scan mode 9.1. Select from the top menu bar; the image acquired for the currently activated analysis area (if you just performed mapping, for example) will appear. If you wish to move to a different area on the specimen and/or change magnification for performing a line scan, do so and then select to acquire a new image and generate a new analysis area in the project. If you wish to map on the same analysis area just used, you do not need to do this as the mapping data will be saved to this analysis area. Click and drag on the image to position the line as desired. 9.2. If you want give the line scan a finite width, adjust LineWidth accordingly (usually not necessary); likewise, to decrease the distance between points in the line scan (therefore increasing the resolution of the lines can), reduce the value for Resolution (usually 0.1 µm is sufficient). Keep in mind that increasing (decreasing) LineWidth ( Resolution ) will increase the time necessary to complete the line scan. 12

9.3. Adjusting the Quality setting will change how much signal is automatically collected in the line scan; the higher the quality, the more signal will be collected, but the line scan will take longer to complete. A standard quality line scan with 0.1 µm resolution will usually take 2 3 min to complete. 9.4. Once all settings for the line scan have been determined, select to start collecting the line scan; the system will first collect a survey spectrum to determine which elements are present and then begin mapping. 9.5. When the line scan is complete, select to save the line scan. Keep in mind that the color scheme for the elements will be constrained to be the same throughout the project, so if an element was defined to be a specific color for a previous map/line scan, it will be the same color in the newly acquired line scan. Of course, if a new element is used in the line scan, its color may be adjusted similar to as described for a map. 13

9.6. Double clicking on the line scan in the project tree will pull up the imaged area with the line scans for each element superimposed and the sum spectrum for the line scan (which can be saved as already described). Hover over the image and select the button (top left side of the image) to save the image with the superimposed line scans along with an excel file of the map data to your designated folder. 9.7. Clicking on a point in the line scan will bring up the spectrum for that point (which can be saved as already described); the selected point on the line scan will be indicated in the image. However, if you want to save the image with the superimposed line scans with the selected point indicated, this must be done via screen capture, rather than by using the button. 14

9.8. The as-collected line scan plots measured intensity without any background subtraction (ROI) for the automatically identified elements. However, it is possible to replot the line scan as intensity with background subtraction (Net), atomic %, or weight % or with more/fewer elements. Double click on the line scan in the project; then hover over and select the desired type of line scan under Data Type ; select to rebuild the line scan using the selected data type; you will be prompted to add/remove any elements you choose. Once the rebuilt line scan is complete, it will now show up in the project under the original line scan (the image with the superimposed lines cans and raw data can be saved as described for the original line scan). 10. Changing analysis areas 10.1. If EDS is to be performed on a different area, the new area should be brought to WD = 15 mm as described previously. 11. Changing samples; manipulating project content 11.1. If you want to add another sample to the project, first warm the EDS detector and unload the current specimen as described in the next step (you do not need to close TEAM or MiraTC). Then, load the new sample, bring it to the correct working distance, adjust the SEM alignment (if needed) and re-cool the detector. Select from the bottom of the expanded Project Content tab and follow the instructions to create a new sample. 15

11.2. Generally speaking, if you keep the beam voltage and intensity the same between specimens, the SEM alignment will still be reasonably accurate when the new sample is loaded and brought to WD = 15 mm (though it doesn t hurt to check it again anyways). That being said, if the beam voltage and/or intensity are changed for analyzing a new sample, the complete SEM alignment should be repeated. 11.3. Any sample, area, spot, map, or line scan can be renamed or deleted from the project by selecting the item and then selecting or, respectively. 12. Finishing the session 12.1. In the Main Toolbar panel, select to turn the CCD camera back on and open the Chamber View panel. 12.2. Expand the Advanced Properties tab (right side of the window); then select EDS detectors ; then Octane Pro, and finally Det 1 Detector Status. Select to stop cooling the detector. The button will first turn yellow and finally red and read when cooling is off and Ok to Vent will turn green. Do not vent the SEM unless the detector cooling is off. 12.3. Close the TEAM software (all data is saved automatically). 12.4. If the BSE detector was used when performing EDS, plug the ACD1 cable back into SIG A on the control rack, to switch back to the SE detector. 12.5. Turn off the beam, calibrate the stage, vent the chamber, remove the specimen, pump the chamber back down, and log off MiraTC. 16