IR WINDOW TRANSMISSION GUIDEBOOK. Copyright CorDEX Instruments Ltd. ID 4015 Rev A

Similar documents
THERMOGRAPHY. Courtesy of Optris. Fig1 : Thermographic image of steel slabs captured with PI1M

Understanding Infrared Camera Thermal Image Quality

Thermography. White Paper: Understanding Infrared Camera Thermal Image Quality

Use of infrared thermography in electronics

CHAPTER 7. Components of Optical Instruments

Alexandrine Huot Québec City June 7 th, 2016

Energy in Photons. Light, Energy, and Electron Structure

5 FACTORS INFLUENCING RADIOMETRIC TEMPERATURE MEASUREMENTS

Near-IR cameras... R&D and Industrial Applications

INSTRUMENTATION BREADBOARDING (VERSION 1.3)

Lecture 2. Electromagnetic radiation principles. Units, image resolutions.

A Perspective on Wavelength Transformation by Absorptive Optical Filters

What professionals look for in an infrared camera

Application Note (A13)

NAME SECTION PERFORMANCE TASK # 3. Part I. Qualitative Relationships

1.0 PURPOSE AND SCOPE

Synchronized electronic shutter system (SESS) for thermal nondestructive evaluation Joseph N. Zalameda

The Ultimate Infrared Handbook for R&D Professionals

INFRARED ANALYSIS OF SINGLE AND MULTILAYER FILMS IN THE PRODUCTION AREA

Spectral Pure Technology

Black is the absence of all visible light. Black is NOT a color!!!

Testo SuperResolution the patent-pending technology for high-resolution thermal images

Fig Color spectrum seen by passing white light through a prism.

High Performance Thin Film Optical Coatings Technical Reference Document 09/13. Coatings Capabilities. Heat Control - Hot Mirror Filters

Radiometry vs. Photometry. Radiometric and photometric units

Specifications. Optical grade germanium single crystals

THE CCD RIDDLE REVISTED: SIGNAL VERSUS TIME LINEAR SIGNAL VERSUS VARIANCE NON-LINEAR

Term Info Picture. A wave that has both electric and magnetic fields. They travel through empty space (a vacuum).

White Paper on SWIR Camera Test The New Swux Unit Austin Richards, FLIR Chris Durell, Joe Jablonski, Labsphere Martin Hübner, Hensoldt.

746A27 Remote Sensing and GIS

Evaluation of infrared collimators for testing thermal imaging systems

Application Notes Photoconductive Cells

INVESTIGATION OF IMPROVED LABEL CUTTING BY CO 2 LASERS WITH WAVELENGTH OPTIMIZATION Paper #2004

Light waves. VCE Physics.com. Light waves - 2

Measuring the Light Output (Power) of UVC LEDs. Biofouling Control Using UVC LEDs

Geo/SAT 2 INTRODUCTION TO REMOTE SENSING

Development of a shutterless calibration process for microbolometer-based infrared measurement systems

Historical radiometric calibration of Landsat 5

PRODUCTION DATA SHEET

PUV3402 LED multiwave photometer A new approach to online process photometry

Ultraviolet Visible Infrared Instrumentation

Spectral Analysis of the LUND/DMI Earthshine Telescope and Filters

Test 1: Example #2. Paul Avery PHY 3400 Feb. 15, Note: * indicates the correct answer.

USING NIR MOISTURE SENSORS TO IMPROVE DRYER PERFORMANCE

Exercises The Color Spectrum (pages ) 28.2 Color by Reflection (pages )

Improved image processing of road pavement defect by infrared thermography

INFRARED MEASUREMENTS OF THE SYNTHETIC DIAMOND WINDOW OF A 110 GHz HIGH POWER GYROTRON

FluorCam PAR- Absorptivity Module & NDVI Measurement

LWIR NUC Using an Uncooled Microbolometer Camera

Scientific Approach to Cool Room Validation. Garry Ward / July 2016

Calibration of a High Dynamic Range, Low Light Level Visible Source

Life Prediction of Mold Transformer for Urban Rail

Radiometric and Photometric Measurements with TAOS PhotoSensors

Handheld Infrared Thermometers

USER MANUAL TC7000/TC7150 INTRINSICALLY SAFE THERMAL IMAGING CAMERA

Figure 1: Percent reflectance for various features, including the five spectra from Table 1, at different wavelengths from 0.4µm to 1.4µm.

Technical Notes. Introduction. Optical Properties. Issue 6 July Figure 1. Specular Reflection:

Conceptual Physics Fundamentals

ThermaCAM E320 Unparalleled image quality: 320 x 240 pixels. Full-featured infrared camera producing ultra-sharp images

Measurement of the thickness of thin foils and testing of the heat sealing of food and medicinal packaging

Spectrophotometer. An instrument used to make absorbance, transmittance or emission measurements is known as a spectrophotometer :

Texture characterization in DIRSIG

Choosing the Best Optical Filter for Your Application. Georgy Das Midwest Optical Systems, Inc.

USER GUIDE Third Generation For Android and ios

Enhanced LWIR NUC Using an Uncooled Microbolometer Camera

Lab 4 OHM S LAW AND KIRCHHOFF S CIRCUIT RULES

CHEM*3440 Instrumental Analysis Mid-Term Examination Fall Duration: 2 hours

Uses of Electromagnetic Waves

Section Electromagnetic Waves and the Electromagnetic Spectrum

Infrared frequency selective surfaces: design, fabrication and measurement

Human Retina. Sharp Spot: Fovea Blind Spot: Optic Nerve

Third Generation For Android

Light, Color, Spectra 05/30/2006. Lecture 17 1

SPECTROCLICK KIT EXPLORE THE INTERACTION OF LIGHT AND MATTER THE SCIENCE OF SPECTROSCOPY. 101 W. Tomaras Ave. Bldg.

Photoresist erosion studied in an inductively coupled plasma reactor employing CHF 3

LASERS. & Protective Glasses. Your guide to Lasers and the Glasses you need to wear for protection.

Understanding Proprietary Infrared Image Files

See more with the thermal imager testo 880

Absentee layer. A layer of dielectric material, transparent in the transmission region of

TECHNICALBRIEF #9 THE POWDER COATING INSTITUTE 2121 EISENHOWER AVENUE, SUITE 401, ALEXANDRIA, VIRGINIA WEATRERIN G

Metal coatings analysis using the handheld Agilent 4100 ExoScan FTIR

Radiometry vs. Photometry. Radiometric and photometric units

Dr E. Kaplani. Mechanical Engineering Dept. T.E.I. of Patras, Greece

Application Note (A16)

SUPPLEMENTARY INFORMATION

POWER DETECTORS. How they work POWER DETECTORS. Overview

Infrared Imaging of Power Electronic Components

Infrared Camera-based Detection and Analysis of Barrels in Rotary Kilns for Waste Incineration

LlIGHT REVIEW PART 2 DOWNLOAD, PRINT and submit for 100 points

MIRAGE DYNAMIC INFRARED SCENE PROJECTOR. Frequently Asked Questions

I D = I so e I. where: = constant T = junction temperature [K] I so = inverse saturating current I = photovoltaic current

Thermal Imagers and Marine Inspections

CoroCAM FAQ s Answered

Absorption: in an OF, the loss of Optical power, resulting from conversion of that power into heat.

Beehive State Engineers

FLIR Tools for PC 7/21/2016

Fluke 570 Series Infrared Thermometers:

COLOUR INSPECTION, INFRARED AND UV

Sometimes the axis of the I-U-dependence are shown in reverse order. In this case the graph shows the stabilized current and measured voltage.

Thomas G. Cleary Building and Fire Research Laboratory National Institute of Standards and Technology Gaithersburg, MD U.S.A.

Transcription:

IR WINDOW TRANSMISSION GUIDEBOOK ID 4015 Rev A

Content 1. General... Page 3 2. Introduction... Page 4 3. Aims... Page 5 4. What is Infrared Transmission?... Page 7 5. Infrared 101 - R+A+T=1... Page 8 6. Crystal IR Window Transmission Schematic... Page 9 7. CorDEX Crystal IR Window Schematic... Page 10 8. Crystal or Mesh?... Page 11 9. Polymer/Mesh Transmission Schematic... Page 12 10. Crystal Optic vs Polymer/Mesh Optic... Page 13 11. Calibration Basics... Page 15 12. Calibrating for Transmission Traditional... Page 16 13. Coffee Cup Test Results at a Higher Temperature... Page 17 14. Coffee Cup Calibration Applied to a Range... Page 18 15. Multipoint Calibration... Page 20 16. What is Spectral Transmission?... Page 21-22 17. Changes in Camera Response... Page 23 18. CorDEX IW Series Correction Maps... Page 24 19. CorDEX Transmission Correction with Fluke Ti32... Page 25 20. CorDEX Transmission Correction with FLIR A320... Page 26 21. Correction Tools... Page 27 22. Conclusions... Page 28

General Understanding Infrared Transmission Infrared (IR) Windows and their Effects on IR Readings This guidebook explains the factors affecting transmission through an Infrared (IR) window (as used in a practical electrical inspection application). Learn how different IR window types affect your readings and how to correct for transmission losses.

Introduction Infrared thermography is a proven technique for inspecting live electrical equipment. With the installation of one or more IR windows, an infrared thermographer can examine live electrical equipment without removing panel covers. The use of IR windows has become increasingly common over the past ten years. This is due primarily to the increase in electrical safety awareness caused by the widespread implementation of safety standards such as NFPA70E and CSA Z462. However, this guidebook introduces the theory behind infrared transmission. There are two kinds of IR Window: crystal and reinforced polymer. How do these materials affect readings? What are the options available to thermographers to allow correction and more accurate problem diagnosis? IR Windows are not 100% transmissive to infrared radiation and so the thermal imager will be inaccurate when viewing through an IR Window. Thermographers want to measure temperature, if temperature measurement is not important why buy a radiometric camera?

Aims This guidebook will: Explain the theory of infrared transmission as it applied in real life situations to IR Windows Explain the options available to thermographers when it comes to transmission correction Benefits and shortfalls of Single point (or coffee cup) calibration as opposed to multi point calibration Correction in practices determining a calibration range, and how Camera differences can affect Accuracy expectations

Lets get started IR WINDOW TRANSMISSION 1

What is Infrared Transmission? Infrared transmission can be defined as the proportion of infrared radiation emitted from a target which passes through the IR Window and reaches the thermal imager enabling a measurement. IR Window manufacturers are often asked the question: What is the transmission of your IR Window?. Unfortunately, there is no simple answer. Most, if not all IR Windows used for electrical inspection, are spectral in nature: the transmission of the IR Window itself changes with wavelength and therefore target temperature. Additionally, the apparent transmission of the IR Window depends, to some extent, on the spectral response of the camera. Since the majority of predictive maintenance (pdm) thermal imagers operate in the long wave (LW) band of 8-14µm, this guidebook will concentrate on the specific effect on this type of thermal imager. It is not possible to achieve 100% transmission: where does the rest of the IR energy go?

Infrared 101 - R+A+T=1 Reflection and Absorption The total IR Energy an imager sees is made up of three components Reflected, Absorbed and Transmitted energy: R + A + T = 1 Thermographers are accustomed to the first two as these are input into a thermal imager to obtain a valid reading however the third is only used when there is a medium between the target and the camera. In this case an IR Window. When setting up a thermal imager to take a reading without an IR Window, we make an assumption that there is zero transmission component as we have a direct line of sight to the target being measured. The energy measured by the imager (see Figure 1) is made up of: Energy reflected from the target Energy emitted by the target

Crystal IR Window Transmission Schematic When we add an IR Window to the system, we introduce additional radiation sources: Energy reflected from the target and transmitted by the IR Window Energy emitted by the target and transmitted by the IR Window Energy emitted by the IR Window Energy reflected from the IR Window The sum of these four values is known as the total radiance and this is the amount of energy our thermal imager registers and converts into a temperature measurement value. The overall effect can either be positive or negative depending upon the severity of the environment and the composition of the IR Window. This standard crystal IR Window transmission schematic shows the four signals received by the thermal imager. To calibrate an IR window, the goal is to isolate all signals other than the emitted by target signal, shown in green.

CorDEX Crystal IR Window Schematic IW Series IR Windows with HydroGARD coating have almost zero surface reflectivity which means we can remove the Reflected by window signal, shown on the previous page in purple. For real-time measurements we require high emissivity targets, therefore we can remove the Reflected by target signal, shown previously in orange. This leaves only the Emitted by target and Emitted by window signals. As we are generally interested in higher temperatures the effect on reading by energy emitted from the window is minimal. TOP TIP If the IR window is hotter than the original target, then the amount emitted by the IR Window itself will be higher than the camera would originally have seen had the IR Window not been present. Therefore, the camera will read HIGH

Crystal or Mesh? So far we have assumed that we have an uninterrupted path to our target, which is not always the case. Some IR Viewing panes use a combined polymer and mesh as the optic, lets see how this combination affects the camera. In the schematic below, we can see that the opaque mesh (black) has totally different transmission characteristics to semi-opaque polymer (yellow). Inconsistency due to mesh/polymer combination, focus, target temperature and window temperature make polymer/mesh combination impossible to calibrate over standard pdm range.

Polymer/Mesh Transmission Schematic This polymer/mesh IR Window transmission schematic shows multiple signals received by the thermal imager. For this model we have separated the signals to the camera from polymer optic and opaque mesh grid. It is clear that there are now a minimum of six signals reaching the camera; four from the polymer optic and four from the mesh grid. We can see that the mesh and polymer behave very differently. The polymer does allow some energy to pass but also reflects and emits its own energy. The mesh in contrast is totally opaque and therefore the signals it transmits to the camera are reflected and emitted. This behaviour is repeated multiple times due to the mesh honeycomb. Additional problems arise when focus and angle are introduced into the model.

Crystal Optic Vs Polymer/ Mesh Optic Polymer / Mesh Optics Multiple complex signals to thermal imager Severely affected by focus and angle Repeatable transmission correction is not possible Suitable for qualitive (non-measurement) based thermography only Crystal Optics Suitable for quantitive (measurement) based thermography with transmission correction algorithm Simple, repeatable signals to thermal imager Minimal surface reflection Does not affect target focus

We have established it is not possible to correct through mesh SO, WHAT NEXT? 2

Calibration Basics Standard instrument calibrations are performed at set points across the process variable (PV) range. Traditionally, IR window calibrations have been related to a single point calibration also known as a coffee cup calibration. In order to obtain accurate readings across a range, a correction curve must be created. Typically, calibrations across a range of the following 0%, 20%, 40%, 60%, 80%, 100%. This allows the instrument to be accurate at multiple points. Since a thermographer is interested in trending temperature over time to extrapolate a failure before it occurs, calibration over a range is the ultimate goal when it comes to accuracy.

Calibrating for Transmission - Traditional To highlight the limitations of single point calibrations, we will apply the coffee cup test to a range and measure the results using a IW series intelligent IR window and long wave thermal imager. Coffee cup test procedure Results Set the camera emissivity to 1 Point the camera at the hot cup of coffee, make sure it is in focus and record the temperature Insert the window between the cup and the camera and record the reading No window e =1 Alter the emissivity to bring the camera temperature reading back to pre-window Multiply the value shown by the emissivity of the next target and that it the transmission of the window TRY IT FOR YOURSELF! With Window e =0.45

Coffee Cup Test Results at a Higher Temperature Below are two sets of results. One showing the initial calibration temperature where accuracy is 2. The second set of results show the same calibration figure applied to a higher range, in this case 238F. In this case, the camera returns a temperature of 272F which is high by 34. Single point, coffee cup calibration No window 100F Window with calibration e = 0.45 Error @ 100F, 2 No window 238F Window with calibration from above Single point, coffee cup calibration at different temperature e = 0.45 Error @ 238F, 34 Camera reads HIGH

Coffee Cup Calibration Applied to a Range The graph above shows the effect of a single point calibration over a range. At the calibration point, the camera is accurate. However, as the same calibration factor is applied to increasing and decreasing target temperatures, significant errors occur. Coffee cup calibration is only accurate at single point Readings are inaccurate as temperature changes If tested above coffee cup test temperature, camera reads high thus meaning potential false alarms If tested below coffee cup test temperature, camera reads low thus meaning potential missed problems SINGLE POINT CALIBRATIONS ARE INACURATE WHEN APPLIED TO A RANGE

Putting theory into practice MULTIPOINT CORRECTION 3

Multipoint Calibration Goals The ultimate goal is to create a correction map which provides accurate readings over a range of target temperatures. The aim is to achieve a ±5 target accuracy after calibration. Whilst this is outside the standard camera accuracy, it is more than adequate for electrical inspections. When creating the map, it is essential to isolate erroneous signals which will affect the result. Challenges All IR Windows even those using the same material at the same thickness will exhibit slight differences in transmission results. Transmission maps must be individual to each IR Window. A typical electrical pdm range of 85-330F requires a minimum of five calibration points to be in line with standard instrument calibration protocol. Spectral responses of individual cameras change from one to another. This will always create an error but generally will be within the ±5 target accuracy.

What is Spectral Transmission? Plancks curve for blackbody at 95F Total radiant energy detected by thermal imager in 8 14µm band is the area under the curve which e q u ate s to temperature when the imager is calibrated. 8-14µm band With an IR Window b etween the imager and the target, the amount of energy received by the thermal imager in this instance shown in g r e e n. T h e calibration factor is the multiplication value required to increase the green area to orange. 8-14µm band Plancks curve for blackbody at 95F 8-14µm band

What is Spectral Transmission? Hotter temperatures curve moves into the shorter wavelength Colder temperatures curve moves into the longer wavelength Camera monitors 8-14 band but energy changes within this band at different target temperatures. A spectral transmitter modulates the energy as a function of wavelength which means a single point is insufficient to calibrate accurately. Planck's curve moves along the IR Spectrum depending upon target temperature. Since the transmission of the IR Window is better in the shorter wave lengths, more energy is transmitted by the IR Window from hotter targets than for colder targets which transmit more on the longer wavelengths. A material having the ability to change transmission rate as a function of target temperature is known as a spectral transmitter.

Changes in Camera Response Individual thermal imagers all have different spectral responses. This is a characteristic of the individual detectors as they come off the production line. These differences do not affect thermographers in general as the camera is calibrated. The affected only become visible when a spectral transmitter is placed between the thermal imager and the target. Less Energy More Energy All three cameras see slightly different radiant energy levels for the same target temperature. The cameras all read the same as they are calibrated to understand that this specific level of intensity equates to a specific temperature, however this will cause an error when used with a spectrally transmitting IR Window.

CorDEX IW Series Correction Maps Previously, infrared thermographers have requested transmission curves from IR window manufacturers. In reality, a transmission curve has little use as it doesn't provide the thermographer with the information required to obtain an accurate reading. What the thermographer actually needs is a correction curve. A typical CorDEX IR Window correction curve is shown opposite. This curve correlates indicated temperature as shown by the thermal imager and actual temperature of the target across a range. A multipoint correction map is created across the identified pdm range and made available to thermographers Distance to targets, ambient humidity and optic temperature are all considered and corrected for within the map Multiple points are considered when creating the map, far exceeding the standard five instrument calibration requirements TARGET ACCURACY OF ±5 IS THE GOAL, BUT DOES IT WORK...

CorDEX Transmission Correction with Fluke Ti32 Fluke Ti25 Temperature No Correction CorDEX Correction (Without Window) Temperature (With Window) Temperature (With Window) Error (C) Error (%) 30 26 32 2-7% 45 32 44.9-0.1 0% 60 40 61.2 1.2-2% 75 44 78.9 3.9-5% 91 59 96 5-5% 105 63 103-2 2% 119 72 118.5-0.5 0% 134 86 132-2 1% 150 92 151.5 1.5-1%

CorDEX Transmission Correction with FLIR A320 FLIR A320 Temperature (Without Window) No Correction CorDEX Correction Temperature (With Window) Temperature (With Window) Error (C) Error (%) 31 25 29.8 1.2 4% 46 32 44.9 1.1 2% 60 40 61.2-1.2-2% 76 48 76.3-0.3 0% 90 56 90.7-0.7-1% 105 64 104.7 0.3 0% 120 72 118.5 1.5 1% 135 81 133.8 1.2 1% 150 90 148.4 1.6 1%

Correction Tools CorDEX Instruments provide a serials of transmission correction tools which are designer to operate with any longwave (8-14µm) thermal imager. These tools are available at www.irwindows.com or alternatively as a iphone app available on the Apple App Store. CorDEX iphone transmission correction app available on the Apple App Store. Online transmission correction at www.irwindows.com TO TRY THE CORRECTION ALOGORITHM, USE THE DEMO CODE MIEE IN EITHER THE APP OR WEBSITE!

Conclusions To conclude, traditional single point correction is unreliable when applied to a range of target temperatures. Conversely, multipoint correction maps will provide accurate results with different thermal imagers. Typical target accuracy of ±5 is achievable using multipoint correction curves regardless of imager. TEMPERATURE MEASUREMENT THROUGH AN IR WINDOW ACROSS A RANGE IS NOW POSSIBLE

WANT TO LEARN MORE? WWW.IRWINDOWS.COM EMAIL: TRAINING@CORD-EX.COM