TBX-1329 AC/DC COUPLING TERMINAL BLOCK

Similar documents
TBX-1329 AC/DC COUPLING TERMINAL BLOCK

SCXI CURRENT INPUT,HIGH-VOLTAGE TERMINAL BLOCK

SCXI High-Voltage Terminal Block

SCXI High-Voltage Terminal Block

SCXI HIGH-ACCURACY ISOTHERMAL TERMINAL BLOCK

TBX-68S Isothermal Terminal Block

The SCB-68 is a shielded board with 68 screw terminals for easy connection to National Instruments 68-pin products.

BNC/TC-2095 RACK-MOUNT ADAPTER

TB-2705 PXI TERMINAL BLOCK FOR ESERIES

GETTING STARTED GUIDE NI AI, ±60 V, 24 Bit, 1 ks/s/ch Simultaneous

TBX-1326 HIGH-VOLTAGE TERMINAL BLOCK

GETTING STARTED GUIDE NI AI Differential/32 AI Single-Ended, ±200 mv to ±10 V, 16 Bit, 250 ks/s Aggregate, Fuel-Cell Measurements

GETTING STARTED GUIDE NI AI, ±10 V, 24 Bit, 50 ks/s/ch Simultaneous

TBX CHANNEL ISOTHERMAL TERMINAL BLOCKINSTALLATION GUIDE

24V dc Analog Block I/O Module

SCC-ACC01 Accelerometer Input Module

FP-DO-401 and cfp-do-401

FP-DO-403 and cfp-do-403

MINI MCR-SL-UI-I-LP-NC

GETTING STARTED GUIDE NI AI, ±5 V, 24 Bit, 51.2 ks/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling

SCC-CO20 Isolated Current Output Module

FP-SG-140 AND CFP-SG-140

SCC-FV01 Frequency Input Module

FP-DO-400 and cfp-do-400

NI Channel SPST Electromechanical Relay Module

FP-RTD-124 AND CFP-RTD-124. Features. FieldPoint Operating Instructions. Eight-Channel Four-Wire RTD and Resistance Input Modules

VXI-TB CHANNEL ISOTHERMAL TERMINAL BLOCK

Instruction Manual ICM 3091N Digital AC Clampmeter EN FR IT DE ES

MS2030 CAT III 600 V A V AUTO RS232

SCXI 8-Channel Isolated Analog Input Modules

FP-RTD-122 and cfp-rtd-122

FP-RTD-122 AND CFP-RTD-122. Features. FieldPoint Operating Instructions. Eight-Channel Three-Wire RTD and Resistance Input Modules

SCXI HIGH-VOLTAGE ATTENUATOR TERMINAL BLOCK

SCXI. SCXI-1120/D User Manual

FP-RLY-420. Features. Installing the FP-RLY-420. FieldPoint Operating Instructions. Eight-Channel SPST Relay Module

OPERATING INSTRUCTIONS NI Port, Low-Speed/Fault Tolerant CAN Module

MODEL: M2DYS SEN TRONIC AG

NI Channel, Universal Digital Input Module

MODEL: M2SES SEN TRONIC AG [3] OUTPUT

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

SmartBlock I/O Module - HE579DAC107/HE579DAC207 8/4 Channel Current and Voltage Output 0-5V / 0-10V / 4-20mA / 0-20mA - CsCAN

8 Input/8 Output AC Block I/O Module

MindWare 2-Slot BioNex Hardware User Reference Guide. Copyright 2011 by MindWare Technologies LTD. All Rights Reserved.

312, 316, 318. Clamp Meter. Users Manual

User's Manual: Series 350T AC Current Input (External Sensor), DC-Powered Transmitters

MODEL: W2DYS SEN TRONIC AG. Mini-MW Series

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

FP-RTD-122. Features OPERATING INSTRUCTIONS. 8-Channel, 3-Wire RTD and Resistance Input Module

MDC4500-4B Amplifier/Combiner Data Sheet & User Manual. February Rev 0.95

Contents 1. General instructions. 1.1 Precautions safety measures Protection mechanisms. 2. Description. 2.1 Instrument Familiarization. 2.

AM-510 Commercial / Residential Multimeter. AM-510-EUR Digital Multimeter. Users Manual

MODEL W Power Amplifier

SCC-DI01 Isolated Digital Input Module

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9211E 4-Channel Thermocouple Input Module

80i-110s AC/DC CURRENT PROBE

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9207E 16-Channel, ±20 ma/±10 V, 24-Bit Analog Input Module

Eight-Channel Thermocouple Input Module

SCXI User Manual. Six-Channel Isolated Digital-to-Analog Converter Module. September 1999 Edition Part Number B-01

Contents NUDRIVE ACCESSORY USER GUIDE

RIGOL. User s Guide. RP5600 Passive Probe. July 2010 RIGOL Technologies, Inc.

DAQ. SCC Series User Manual. SCC Series User Manual. December 1999 Edition Part Number D-01

RAD-IN/OUT-8D. Digital extension modules for the bidirectional wireless system. INTERFACE Data sheet _en_05. 1 Description

User's Manual: Series 450T AC Current Input (External Sensor), AC-Powered Transmitters

Analog Signal Conditioning Accessories

CEC VIBRATION TRANSMITTER

Model MHz Arbitrary Waveform / Function Generator

ETHOS 5030 TRUE-RMS DIGITAL MULTIMETER OPERATION MANUAL

GETTING STARTED GUIDE NI khz to 100 MHz, 50 Ω, AC Coupled RF Receiver Module

PHV RO High Voltage Passive Probe. Instruction Manual

Model 7000 Low Noise Differential Preamplifier

RIGOL. User s Guide. RP1000D Series High Voltage Differential Probe. Feb RIGOL Technologies, Inc

TA MHz x10/x100 Differential Probe User s Manual. This probe complies with IEC , IEC CAT III, Pollution Degree 2.

99 Washington Street Melrose, MA Fax TestEquipmentDepot.com # # AAC Clamp Meter. Instruction Manual

OPERATING INSTRUCTIONS NI Channel, 24-Bit, Universal Analog Input Module

Model: Pro95 TRUE RMS MILLIAMP CLAMP METER

Impact Wrench. 19 mm (3/4 ) MODEL 6906

DATASHEET NI Channel, 250 V Sinking Digital Input Module

Model 4210-MMPC-L. Multi-measurement Prober Cable Kit. Overview. Quick start guide topics. Related documents

PHV 1000-RO High Voltage Passive Probe. Instruction Manual

AM-500 Autoranging Mulitmeter. AM-500-EUR Digital Multimeter. Users Manual

Agilent 1146A AC/DC Oscilloscope Current Probe

N2790A Differential Voltage Probe

Installing the IPS 4240 and the IPS 4255

ProfiScale MULTI Multimeter

IDEAL INDUSTRIES, INC. TECHNICAL MANUAL MODEL: MODEL: Multimeter Service Information

FP-CTR-500 and cfp-ctr-500

Keysight 16440A SMU/Pulse Generator Selector

USB-2404-UI Specifications

OPERATING INSTRUCTIONS AND SPECIFICATIONS NI Channel, ±10 V, 16-Bit Simultaneous Analog Input Module. ni.com/manuals

OPERATING INSTRUCTIONS AND SPECIFICATIONS. NI 9213E 16-Channel Thermocouple Input Module

Model 4210-MMPC-W. Multi-measurement Prober Cable Kit. Overview

CM605. User Manual AC/DC LOW CURRENT CLAMP-ON METER ENGLISH

SCXI TERMINAL BLOCK

INTELLIMETER REGISTER

Pen-type DIGITAL MULTITESTER

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

M302RM OPERATING MANUAL

Operator s Manual. PP022 Passive Probe

Mid-Power Expansion Unit (MxU) Quick Installation Sheet

TETRIS 1000 High Impedance Active Probe. Instruction Manual

Transcription:

INSTALLATION GUIDE TBX-1329 AC/DC COUPLING TERMINAL BLOCK Introduction This guide describes how to install and use the TBX-1329 AC/DC coupling terminal block with the SCXI-1120, SCXI-1120D, and SCXI-1121 modules. The TBX-1329 is a DIN rail-mountable terminal block that consists of a shielded board with screw terminals to connect to the SCXI-1120, SCXI-1120D, or SCXI-1121 front connector. The TBX-1329 mounts on most European standard DIN EN mounting rails. The terminal block has 16 screw terminals that unplug for easy connection. Two screws are provided to connect to the SCXI functional earth terminal via the SH32-32-A cable shield. With the SCXI-1120 and SCXI-1120D, the remaining eight pairs of screw terminals connect signals to the eight SCXI module input channels. With the SCXI-1121, four pairs of screw terminals connect signals to the four SCXI module input channels and four pairs connect to the SCXI module excitation channels. What You Need to Get Started To set up and use your TBX-1329, you will need the following: TBX-1329 AC/DC coupling terminal block TBX-1329 AC/DC Coupling Terminal Block Installation Guide 1/8 in. flathead screwdriver (included in kit) Your SCXI chassis and documentation SCXI-1120, SCXI-1120D, or SCXI-1121 module and documentation SH32-32-A shielded cable assembly Long-nose pliers SCXI is a trademark of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. 321556B-01 Copyright 1997, 1999 National Instruments Corp. All rights reserved. March 1999

3/16 in. wrench Number 1 Phillips-head screwdriver Installation Warning SHOCK HAZARD: Refer to the Signal Connection section in this guide before connecting any signals. If signal wires are connected to the terminal block, dangerous voltages may exist even when the equipment is turned off. Do not operate the device in an explosive atmosphere or where there may be flammable gases or fumes. DO NOT OPERATE DAMAGED EQUIPMENT. The safety protection features built into this device can become impaired if the device becomes damaged in any way. If the device is damaged, turn the device off and do not use it until service-trained personnel can check its safety. If necessary, return the device to National Instruments for service and repair to ensure that its safety is not compromised. Do not operate this equipment in a manner that contradicts the information specified in this document. Misuse of this equipment could result in a shock hazard. Terminals are for use only with equipment that has no accessible live parts. DO NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT. Because of the danger of introducing additional hazards, do not install unauthorized parts or modify the device. Return the device to National Instruments for service and repair to ensure that its safety features are not compromised. Use only National Instruments TBX Series cable assemblies with high-voltage TBX Series terminal blocks. TBX-1329 AC/DC Coupling Terminal Block 2 www.natinst.com

Perform the following steps to mount the SH32-32-A cable assembly and connect the TBX-1329 to your SCXI module. Refer to the figures as needed. 1. Turn off your SCXI chassis. 2. Turn off the computer that contains your data acquisition (DAQ) device or disconnect the device from your SCXI chassis. 3. Connect the TBX cable adapter to the appropriate SCXI module and secure it by tightening both thumb screws as shown in Figure 1. 4 3 2 5 1 3 1 M 2.5 Mounting Screws 2 Cable Assembly 3 Thumb Screws 4 SCXI Chassis 5 TBX Cable Adapter Figure 1. Connecting the SH32-32-A Cable to the SCXI Module National Instruments Corporation 3 TBX-1329 AC/DC Coupling Terminal Block

4. Verify that the four backshell mounting ears on the cable assembly are in the position shown in Figure 1 and Figure 2. If not, remove the backshell mounting ears and install them in the position shown. 3 5 4 1 3 2 1 1 Backshell Mounting Screws and Ears 2 Safety Ground Solder Lugs 3 Captive Cover Screws 4 Terminal Block Connector 5 Signal Wire Entry Note: The SH32-32-A cable is not shown in the exact position for proper connection to the terminal block connector. See Figure 3 for the completed connection. Figure 2. Connecting the SH32-32-A Cable to the TBX-1329 Terminal Block 5. Connect one end of the cable assembly to your SCXI module front connector and secure the SH32-32-A cable by tightening both backshell mounting screws as shown in Figure 1. 6. Connect the other end of the cable assembly to your TBX-1329 terminal block connector and secure the SH32-32-A cable by tightening both backshell mounting screws as shown in Figure 3. 7. Reconnect the DAQ device to your SCXI chassis. TBX-1329 AC/DC Coupling Terminal Block 4 www.natinst.com

See Figure 3 for the completed installation. Figure 3. The Completed Installation Rack Mounting When you have completed the installation, you are ready to mount the TBX assembly in your rack. If you are using the National Instruments TBX Rack-Mount Assembly, refer to the TBX Rack-Mount Installation Guide for instructions. If you are not using this rack-mount assembly, perform the following steps to mount the TBX assembly directly onto your DIN rail. 1. Snap the TBX terminal block onto the DIN rail with a firm push. To remove the TBX terminal block from the DIN rail, place a flathead screwdriver into the slot above the terminal block base and pry it from the rail. 2. Install the SCXI chassis using the appropriate chassis rack-mount kit. National Instruments Corporation 5 TBX-1329 AC/DC Coupling Terminal Block

Signal Connection The following warnings contain important safety information concerning the use of hazardous voltage levels with TBX-1329 terminal blocks. Warning SHOCK HAZARD: Only qualified personnel aware of the dangers involved should use the TBX-1329. KEEP AWAY FROM LIVE CIRCUITS. Do not remove equipment covers or shields unless you are trained to do so. If signal wires are connected to the device, hazardous voltages may exist even when the equipment is turned off. To avoid a shock hazard, do not perform procedures involving cover or shield removal unless you are qualified to do so and disconnect all field power prior to removing covers or shields. National Instruments is NOT liable for any damages or injuries due to misuse of high-voltage signals connected to the terminal block. Equipment described in this document must be used in an Installation Category II environment per IEC 664. This category requires local level supply mains-connected installation. When using the terminal block with high-voltage levels, you must insulate all signal wires appropriately to the highest voltage the terminal block may come in contact with. National Instruments is not liable for any damages or injuries resulting from inadequate signal wire insulation. Any functional earth terminal on your terminal block is for grounding high-impedance sources such as a floating source (1 ma maximum). Do not use these terminals as safety earth grounds. Operating common-mode voltage must not exceed 250 V rms between channels or from any channel to earth. If you are using an SCXI-1121 module, read Voltage Drops and Strain Gauge with the SCXI-1121, before proceeding with the signal connections. Voltage Drops and Strain Gauge with the SCXI-1121 When you use the SCXI-1121 with the TBX-1329 to measure strain gauges, a small amount of voltage drop will develop across the excitation wires in the SH32-32-A cable. This voltage drop is due to the wire resistance and the current flow in these leads when a strain gauge is connected at the TBX-1329 across the excitation outputs provided on the SCXI-1121. To reduce errors in your output voltage, first you need to calculate the voltage drop across the SH32-32-A cable. This voltage drop, which results in an error in your measurement, depends on the cable length and on the TBX-1329 AC/DC Coupling Terminal Block 6 www.natinst.com

strain gauge value and configuration. The SH32-32-A cable has a resistance of 0.21 Ω/m. To determine the amount of error introduced by the cable, do the following: 1. Calculate the total resistance (RTL) of the cable based on the cable length. 2. Refer to your strain gauge data sheet for the bridge resistance of your strain gauge (RSG) connected at the TBX-1329. 3. Use the following formula to determine the total voltage drop (V drop ) in the SH32-32-A cable excitation leads: V drop = [R TL /(R TL + R SG )] x V ex where V ex = excitation voltage. For example, if you have a 1 m SH32-32-A cable, 120 Ω full-bridge strain gauge, and V ex =3.333 V, your calculations from Steps 1 through 3 would be: 1. R TL = 2 x 0.21 Ω/m x 1 m = 0.42 Ω; notice that you multiply the cable length by two to take into consideration both the V ex+ and V ex- lead resistances. 2. R SG = 120 Ω, which is the total equivalent bridge resistance as seen from the V ex± terminals of the TBX-1329. 3. V drop = 11.6 mv 0.3% of 3.333 V excitation. You then need to calculate the voltage drop across the field signal wires you are connecting to the TBX-1329. Perform similar calculations for your field wires as you did for your cable. Resistance may vary depending on your cable and field wires. Add this error amount to the voltage drop across the SH32-32-A cable to get a total voltage drop. You can compensate for this error along with any additional cable lead resistance introduced by the strain gauge connection wires. One simple way of compensation is to calculate the lead resistance as shown in Steps 1 through 3, then input it along with your other strain gauge parameters into the conversion formula in your National Instruments data acquisition software packages. Or, locate the load closer to your SCXI module by using shorter cable lengths, or use heavy-gauge wire to connect to the TBX-1329. National Instruments Corporation 7 TBX-1329 AC/DC Coupling Terminal Block

Current Measurement To measure current with the TBX-1329, you can place a current loop receiver resistor in the sockets on the TBX-1329. Table 1 shows the TBX-1329 socket assignment per channel. Caution Use only National Instruments SCXI current resistors, part number 776582-01, for current measurements. Table 1. Socket Assignments per Channel Channel Number Resistor Designator 0 R0 1 R1 2 R2 3 R3 4 R4 5 R5 6 R6 7 R7 TBX-1329 AC/DC Coupling Terminal Block 8 www.natinst.com

Wiring Instructions Caution When using the device with high common-mode voltages, you MUST insulate your signal wires for the highest input voltage. National Instruments is NOT liable for any damages or injuries resulting from inadequate signal wire insulation. Use only 26-14 AWG wire with a voltage rating of 250 V. Prepare your signal wire by stripping the insulation no more than 7 mm. When connecting or disconnecting signal lines to the SCXI terminal block screw terminals, make sure the lines are powered off. Potential differences between the lines and the SCXI ground create a shock hazard while you connect the lines. Connections, including power signals to ground and vice versa, that exceed any of the maximum signal ratings on the SCXI device can create a shock or fire hazard or can damage any or all of the boards connected to the SCXI chassis, the host computer, and the SCXI device. National Instruments is NOT LIABLE FOR ANY DAMAGES OR INJURIES resulting from incorrect signal connections. If high voltages ( 30 V rms and 42.4 V peak or 60 VDC) are present, YOU MUST CONNECT A SAFETY EARTH GROUND WIRE TO THE TERMINAL BLOCK SAFETY GROUND SOLDER LUG. This complies with safety agency requirements and protects against electric shock when the terminal block is not connected to the chassis. To connect the safety earth ground to the safety ground solder lug, run an earth ground wire in the cable from the signal source to the terminal block. National Instruments is NOT liable for any damages or injuries resulting from inadequate safety earth ground connections. Do not loosen or re-orient the safety ground solder lug hardware when connecting the safety ground wire. To do so reduces the safety isolation between the high voltage and safety ground. National Instruments Corporation 9 TBX-1329 AC/DC Coupling Terminal Block

To connect your field signals to the TBX-1329 terminal block for use with the SCXI modules, follow the labeling on the TBX-1329 indicated along the appropriate row as shown in the parts locator diagram, Figure 4. Connections for the SCXI-1120D and SCXI-1120 are the same. Refer to Figures 2 and 4 as you perform the following instructions: 3 4 5 2 2 1 1 14 13 12 11 10 9 8 7 6 1 Current Loop Receiver Resistor 2 Safety Ground Solder Lugs 3 Product Name 4 Assembly Number 5 Serial Number 6 SC7/SR7 7 SC6/SR6 8 SC5/SR5 9 SC4/SR4 10 SC3/SR3 11 SC2/SR2 12 SC1/SR1 13 SC0/SR0 14 SCXI Module-Type Rows Figure 4. TBX-1329 Parts Locator Diagram TBX-1329 AC/DC Coupling Terminal Block 10 www.natinst.com

AC/DC Coupling Configuration 1. Remove the TBX-1329 terminal block cover by unscrewing the four captive cover screws in the cover corners. These screws stay attached to the cover without falling out. 2. Connect the signal wires to the screw terminals: a. Prepare your signal wire by stripping the insulation no more than 7 mm. b. Connect the signal wires to the screw terminals by inserting the stripped end of the wire fully into the terminals. c. Tighten the terminals to a torque of 5 7 in.-lb (56 79 Newton cm). Refer to your SCXI module user manual for examples of how to connect to field signals and loads. The functional earth terminals are connected to the SCXI chassis via the cable shield. This is not shown in your SCXI module user manual. Allow your signal wires to exit through the TBX-1329 cover opening. 3. Relieve strain on your signal wires by tie-wrapping them to the mounting rails of your rack. 4. Replace the TBX-1329 terminal block cover and tighten the captive cover screws. The installation and signal connection are now complete. Each channel of the TBX-1329 has two switches you can use to switch from AC to DC; thus, each channel can be configured individually. The factory-default setting for the switches is the DC position so that the terminal block will pass all signals, DC and AC, to the SCXI module. In the AC position the switches switch in a capacitor and, depending on which module you use, a referencing resistor. This will block the DC components of the input signal and provide a true AC signal to the SCXI module. Each switch is uniquely marked to make switch identification easier. Each switch that switches the capacitor into the circuit is labeled with SC#. Each switch that switches the resistor into the circuit is labeled with SR#. The # symbol represents the channel number; for example, SC2 is the capacitor National Instruments Corporation 11 TBX-1329 AC/DC Coupling Terminal Block

switch for channel 2 and SR2 is the resistor switch for channel 2. Table 2 shows the TBX-1329 terminal block switch settings. Table 2. TBX-1329 Terminal Block Switch Settings Switch Configuration Switch Settings for Each Channel Description SCXI-1120 and SCXI-1121 SCXI-1120D DC DC coupling selected; removes capacitor and resistor from the circuit. Signals go straight through. Switch both SC# and SR# switches for DC coupling. Switch both SC# and SR# switches for DC coupling. AC DC AC coupling selected; SC# connects the capacitor to the circuit. SR# connects the resistor to the circuit. Switch both SC# and SR# switches for AC coupling. Switch ONLY SC# switch for AC coupling. SR# must be in the DC position. AC Cleaning Instructions Clean devices and terminal blocks by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with deionized water and a stiff nonmetallic brush. The unit must be completely dry and free from contaminants before returning to service. TBX-1329 AC/DC Coupling Terminal Block 12 www.natinst.com

Specifications Maximum working voltage (signal + common mode)... Each input should remain within 250 V rms of ground and any other channel Maximum current... 20 ma 3 db cutoff frequency... 0.072 Hz Capacitors can block up to... 250 VDC Compatible DIN rails... DIN EN 50 022 DIN EN 50 035 Terminal block dimensions... 12.7 7.62 11.18 cm (5 3 4.4 in.) Environmental Operating... 0 to 50 C Storage... 20 to 70 C Relative humidity... 10% to 90% noncondensing Safety... Designed in accordance with IEC/EN 61010-1, UL 3111-1, CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment. Approved at altitudes up to 2,000 meters. Installation Category II Pollution Degree 2 National Instruments Customer Support If you have additional questions, please contact our technical support staff at: Tel: 512 795 8248, Fax: 512 794 5678. Our corporate headquarters address is: 11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100 National Instruments Corporation 13 TBX-1329 AC/DC Coupling Terminal Block