Photoacoustic tomography imaging based on a 4f acoustic lens imaging system

Similar documents
Wideband Focused Transducer Array for Optoacoustic Tomography

Photoacoustic tomography of biological tissues with high cross-section resolution: Reconstruction and experiment

Capacitive Micromachined Ultrasonic Transducers (CMUTs) for Photoacoustic Imaging

Microwave-induced acoustic imaging of biological tissues

Non-contact Photoacoustic Tomography using holographic full field detection

Ultrasound-modulated optical tomography of absorbing objects buried in dense tissue-simulating turbid media

A Real-time Photoacoustic Imaging System with High Density Integrated Circuit

Photoacoustic Imaging of Blood Vessels in Tissue

Time-Domain Reconstruction for Thermoacoustic Tomography in a Spherical Geometry

Methods for parallel-detection-based ultrasound-modulated optical tomography

Confocal Imaging Through Scattering Media with a Volume Holographic Filter

Transmission- and side-detection configurations in ultrasound-modulated optical tomography of thick biological tissues

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

ABSTRACT 1. INTRODUCTION

Multimodal simultaneous photoacoustic tomography, optical resolution microscopy and OCT system

Curved array photoacoustic tomographic system for small animal imaging

All-optical endoscopic probe for high resolution 3D photoacoustic tomography

Transparent Fabry Perot polymer film ultrasound array for backward-mode photoacoustic imaging

Ultrasound Beamforming and Image Formation. Jeremy J. Dahl

A 3D Profile Parallel Detecting System Based on Differential Confocal Microscopy. Y.H. Wang, X.F. Yu and Y.T. Fei

Shaping light in microscopy:

A miniature all-optical photoacoustic imaging probe

Acousto-optic imaging of tissue. Steve Morgan

Photoacoustic imaging with coherent light

Physics 3340 Spring Fourier Optics

High power visible light emitting diodes as pulsed excitation sources for biomedical photoacoustics

Time-reversal and model-based imaging in a THz waveguide

Synthetic-aperture based photoacoustic re-beamforming (SPARE) approach using beamformed ultrasound data

Diffractive Axicon application note

Reflecting optical system to increase signal intensity. in confocal microscopy

Thermoacoustic tomography (TAT) is a novel, noninvasive,

Multi-depth photoacoustic microscopy with a focus tunable lens

Photoacoustic imaging using an 8-beam Fabry-Perot scanner

Linear arrays used in ultrasonic evaluation

Virtual ultrasound sources

Improving the Quality of Photoacoustic Images using the Short-Lag Spatial Coherence Imaging Technique

Ultrasonic Testing using a unipolar pulse

PHY 431 Homework Set #5 Due Nov. 20 at the start of class

Light emitting diodes as an excitation source for biomedical photoacoustics

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

2. Pulsed Acoustic Microscopy and Picosecond Ultrasonics

Ihor TROTS, Andrzej NOWICKI, Marcin LEWANDOWSKI

Numerical simulation of a gradient-index fibre probe and its properties of light propagation

Digital confocal microscope

Prospects for in vivo blood velocimetry using acoustic resolution photoacoustic Doppler

Parallel Digital Holography Three-Dimensional Image Measurement Technique for Moving Cells

Mirrors and Lenses. Images can be formed by reflection from mirrors. Images can be formed by refraction through lenses.

Suppression of FM-to-AM conversion in third-harmonic. generation at the retracing point of a crystal

STUDIES in oncology have shown that angiogenesis, the

Supplementary Figure 1. GO thin film thickness characterization. The thickness of the prepared GO thin

Ultrasonic Linear Array Medical Imaging System

Conformal optical system design with a single fixed conic corrector

Optical transfer function shaping and depth of focus by using a phase only filter

Imaging obscured subsurface inhomogeneity using laser speckle

Lab Report 3: Speckle Interferometry LIN PEI-YING, BAIG JOVERIA

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Improvement of terahertz imaging with a dynamic subtraction technique

Ultrasound Physics. History: Ultrasound 2/13/2019. Ultrasound

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

Polymer Microring Resonators for High- Sensitivity and Wideband Photoacoustic Imaging

OPTICAL PRINCIPLES OF MICROSCOPY. Interuniversity Course 28 December 2003 Aryeh M. Weiss Bar Ilan University

Theory and Applications of Frequency Domain Laser Ultrasonics

A Virtual Confocal Microscope with Variable Diameter to Improve Resolution

Optimal Pupil Design for Confocal Microscopy

Comparison between optical-resolution photoacoustic microscopy and confocal laser scanning microscopy for turbid sample imaging

COMPUTER PHANTOMS FOR SIMULATING ULTRASOUND B-MODE AND CFM IMAGES

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Atomic Resolution Imaging with a sub-50 pm Electron Probe

Optical coherence tomography

PHYS 202 OUTLINE FOR PART III LIGHT & OPTICS

Dual wavelength laser diode excitation source for 2D photoacoustic imaging.

visibility values: 1) V1=0.5 2) V2=0.9 3) V3=0.99 b) In the three cases considered, what are the values of FSR (Free Spectral Range) and

On Determination of Focal Laws for Linear Phased Array Probes as to the Active and Passive Element Size

Superfast phase-shifting method for 3-D shape measurement

Multi-Element Synthetic Transmit Aperture Method in Medical Ultrasound Imaging Ihor Trots, Yuriy Tasinkevych, Andrzej Nowicki and Marcin Lewandowski

Heisenberg) relation applied to space and transverse wavevector

Aberrated Microlenses to Reduce Crosstalk in Free Space Optical Interconnects Systems

Opto-VLSI-based reconfigurable photonic RF filter

Chapter Ray and Wave Optics

Department of Electrical Engineering and Computer Science

Optics and Lasers. Matt Young. Including Fibers and Optical Waveguides

NEW LASER ULTRASONIC INTERFEROMETER FOR INDUSTRIAL APPLICATIONS B.Pouet and S.Breugnot Bossa Nova Technologies; Venice, CA, USA

Multi-frequency and multiple phase-shift sinusoidal fringe projection for 3D profilometry

Nd:YSO resonator array Transmission spectrum (a. u.) Supplementary Figure 1. An array of nano-beam resonators fabricated in Nd:YSO.

Medical Imaging (EL582/BE620/GA4426)

Practice Problems for Chapter 25-26

Supplementary information, Figure S1A-S1H The thickness and the uniformity of the light sheet at different DOFs. By

Ultrasound Bioinstrumentation. Topic 2 (lecture 3) Beamforming

Chapter 34 The Wave Nature of Light; Interference. Copyright 2009 Pearson Education, Inc.

arxiv: v1 [physics.optics] 19 May 2016

Exp No.(8) Fourier optics Optical filtering

Applied Optics. , Physics Department (Room #36-401) , ,

A New Imaging Technique Combining Diffusive Photon Density Waves

Physics in Modern Medicine Fall 2010

GEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS

X-ray phase-contrast imaging

FIELDS IN THE FOCAL SPACE OF SYMMETRICAL HYPERBOLIC FOCUSING LENS

LOPUT Laser: A novel concept to realize single longitudinal mode laser

Compressive Through-focus Imaging

EUV Plasma Source with IR Power Recycling

Transcription:

Photoacoustic tomography imaging based on a 4f acoustic lens imaging system Zhanxu Chen 1, 2, Zhilie Tang 1*, Wei Wan 2 1 School of Physics and Telecom Engineering, South China Normal University, 510006, Guangzhou, China 2 Department of Electronic Information Engineering, GuangDong Polytechnic Normal University, 510665, Guangzhou, China Corresponding author: tangzhl@scnu.edu.cn Abstract: The theory of photoacoustic tomography (PAT) imaging using a 4f acoustic lens imaging system has been investigated, and the theoretical results show that a 4f acoustic lens has the ability of imaging and guarantees axial and lateral unit magnification of image. A system, a 4f acoustic lens imaging system combining with time-resolved technique, is developed to acquire PAT images. The 4f acoustic lens is able to image initial photoacoustic (PA) pressure distribution, which exactly resembles the absorption distribution, onto an imaging plane. Combining with timeresolved technique, the linear transducer array is adopted to acquire the PA pressure distribution to reconstruct the PAT images. Experimental results demonstrate that the system is able to obtain PAT images and the images contrast sharply with their backgrounds. 2007 Optical Society of America OCIS codes: (170.5120) Photoacoustic imaging; (170.0110) Imaging systems; (170.3010) Image reconstruction techniques; (170.6920) Time-resolved imaging. References and links 1. X. D. Wang, Y. J. Pang, G. Ku, X. Y. Xie, G. Stoica, and L. H. Wang, Non-invasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain, Nat. Biotechnol. 21, 803-806 (2003). 2. R. J. Siphanto, K. K. Thumma, R. G. M. Kolkman, T. G. van Leeuwen, F. F. M. de Mul, J. W. van Neck, L. N. A. van Adrichem and W. Steenbergen, Serial noninvasive photoacoustic imaging of neovascularization in tumor angiogenesis, Opt. Express 13, 89-95 (2005). 3. R. G. M. Kolkman, J. H. G. M. Klaessens, E. Hondebrink, J. C. W. Hopman, F. F. M. de Mul, W. Steenbergen, J. M. Thijssen and T. G. van Leeuwen, Photoacoustic determination of blood vessel diameter, Phys. Med. Biol. 49, 47454756 (2004). 4. A. Oraesky, A. Karabutov, V. S. Solomatin, E. V. Savateeva, V. G. Andreev, Z. Gatalica, H. Singh, and R. Y. D. Fleming, Laser optoacoustic imaging of breast cancer in vivo, Proc. SPIE 4256, 12-22 (2001). 5. C. G. A. Hoelen, F. F. M. de Mul, R. Pongers, and A. Dekker, Three-dimensional photoacoustic imaging of blood vessels in tissue, Opt. Lett. 23, 648-650 (1998). 6. K. Geng, W. Xueding, X. Xueyi, S. George, and W. Lihong, Imaging of tumor angiogenesis in rat brains in vivo by photoacoustic tomography, Appl. Opt., 44, 770-775 (2005). 7. K. H. Song, G. Stoica, and L. H. V. Wang, In vivo three-dimensional photoacoustic tomography of a whole mouse head, Opt. Lett. 31, 2453-2455 (2006). 8. X. D. Wang, X. Y. Xie, G. Ku, and L. H. Wang, G. Stoica, Noninvasive imaging of hemoglobin concentration and oxygenation in the rat brain using high-resolution photoacoustic tomography, J. Biomed. Opt. 11, 024015 1-9 (2006). 9. R. A. Kruger, P. Y. Liu, Y. C. Fang, and C. R. Appledom, Photoacoustic ultrasound (PAUS)- Reconstruction tomography, Med. Phys. 22, 1605-1609 (1995). 10. M. H. Xu, and L. H. Wang, Time-Domain Reconstruction for Thermoacoustic Tomography in a Spherical Geometry, IEEE Trans. Med. Imag. 21, 814-822 (2002). 11. K. P. KÖstli, and P. C. Beard, Two-dimensional photoacoustic imaging by use of Fourier-transform image reconstruction and a detector with an anisotropic response, Appl. Opt. 42, 1899-1908 (2003). 12. Y. Wang, D. Xing, Y. G. Zeng, and Q. Chen, Photoacoustic imaging with deconvolution algorithm, Phys. Med. Biol. 49, 3117-3124 (2004). 13. B. Z. Yin, D. Xing, Y. Wang, Y. G. Zeng, Y. Tan, and Q. Chen, Fast photoacoustic imaging system based on 320-element linear transducer array, Phys. Med. Biol. 49, 1339-1346 (2004). (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4966

14. D. W. Yang, D. Xing, H. M. Gu, Y. Tan, and L. M. Zeng, Fast multielement phase-controlled photoacoustic imaging based on limited-field-filtered back-projection algorithm, Appl. Phys. Lett. 87, 194101 1-3 (2005). 15. J. J. Niederhauser, M. Jaeger, and M. Frenz, Real-time three-dimensional optoacoustic imaging using an acoustic lens system, Appl. Phys. Lett. 85, 846-848 (2004). 16. Z. X. Chen, Z. L. tang, W. Wan, and Y. H. He, Photoacoustic tomography imaging based on an acoustic lens imaging system, Acta. Phys. Sin. 55, 4365-4370 (2006). 17. W. Wan, R. S. Liang, Z. L. Tang, Z. X. Chen, H. C. Zhang, and Y. H. He, The imaging property of photoacoustic Fourier imaging and tomography using an acoustic lens imaging system, J. Appl. Phys. 101, 063103 1-7 (2007). 1. Introduction Recently, there has been a wide interest in photoacoustic tomography (PAT) imaging, which is a noninvasive imaging modality for visualizing both the structural and functional information of biological tissues. It combines both high optical contrast and high acoustic penetration depths. In PAT, a short laser pulse is used to heat biological tissues, and a photoacoustic (PA) effect results in the emission of acoustic signals that can be measured by use of a wide-band ultrasonic transducer. The PA pressure is linearly directly proportional to local optical absorption in the biological tissue, and the PA signals carry information about the geometrical structure and optical properties of biological tissues. Thereby, the PA signals are associated with their physiological and pathological status. The ultrasonic waves are free from the strong scattering and decaying of light in biological tissue, and outside tissue, they can be recorded by an acoustic transducer to map optical absorption distribution. The PAT image can display the structure and function of the biological tissue. In experimental, the imaging modality has been applied to imaging breast cancer, skin cancer, brain tumor, blood concentrations and vasculature structure [1-8]. Many researches have been done to get PAT images [9-14]. However, the original absorption distribution can only be obtained by image reconstruction, so it is hard to get realtime PA images. To get real-time PA images, an acoustic lens, which is able to image the initial PA pressure distribution onto an image space in real time without the necessity of computational reconstructions, has been developed and applied to PAT imaging [15-17]. However, to display the three-dimensional (3D) real object, it requires the images have the same magnification on axial and lateral. In optical imaging, the 4f imaging system guarantees axial and lateral unit magnification of the images. So, the 4f acoustic lens can guarantee that. In this paper, the theory of PAT imaging using a 4f acoustic lens imaging system has been investigated, and the theoretical results show that a 4f acoustic lens has the ability of imaging and guarantees axial and lateral unit magnification of an image. A 4f acoustic lens imaging system combined with time-resolved technique, which is developed to acquire PAT images. 2. Theoretical model 2.1. Principle of 4f acoustic lens imaging The induced PA waves are linearly directly proportional to the local optical absorption and they take on the characteristics of wave, such as interference and diffraction. According to the imaging theory of Fourier, if an acoustic lens allows spatial Fourier transform, then it is capable of imaging. However, to display the three-dimensional real object, the same magnification on axial and lateral is needed. To make sure that, the 4f acoustic lens imaging system is designed, as Fig. 1 shows, which similar to the 4f imaging system in the optical imaging. (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4967

Fig. 1. The 4f acoustic lens imaging system. In this system, the two confocal lenses L 1 and L 2 is combined, the front focal plane of the lens L 1 is the object plane O, the image is inputted from this plane. The back focal plane of the lens L 2 is the imaging plane I, the image is outputted from this plane. The confocal plane is named the transform plane T. The system is named for short the OTI imaging system. When the object is illuminated by laser, the ultrasonic signals induced from the object plane can be considered as coherent signals, the whole OTI imaging system is coherent imaging system. According to Abbe theory of image formation, the imaging process of the system will be discussed as follows. The imaging process of coherent ultrasonic system has two steps as follow: The first step is the Fraunhofer diffraction from the O plane to the T plane, its action is dividing frequency. The second step is also the Fraunhofer diffraction, its action is sum frequency, and that is, the synthesized spectrum is to output images. If the ultrasonic is able to free pass the transform plane, the image can be emersion completely. The amplitude of acoustic pressure on the object plane is supposed as PO ( xy,, ) on the transform plane t ( u, v) = 1, so P 2 ( u, v) = P1 ( u, v), that is P T ( u, v) ; on the imaging plane, the amplitude of acoustic pressure is P I ( x', y' ). As Fig. 2 shows, the transforms of acoustic pressure-amplitude are Fourier transforms in the process of two Fraunhofer diffractions. That is Fig. 2. The principle of coherent imaging system. (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4968

PT ( uv, ) = 1 { PO ( xy},, ) (1) PI ( x', y ') = 2 { P T ( u, v) }. (2) Omitting all coefficients, so ik PT( uv, ) PO( xy, )exp[ ( xu+ yv)] dxdy (3) f ik PI( x ', y ') PT( u, v)exp[ ( ux ' + vy ')] dudv (4) f Substitution of Eq. (3) into Eq. (4) gives + ik PI( x ', y ') PO( x, y)exp{ [ u( x + x ') + v( y + y ')]} dudvdxdy f + + + + ik( x + x ') ik( y + y ') = PO ( x, y){ exp[ u] du} { exp[ v] dv} dxdy f f O P ( xy, ) δ( x+ x') δ( y+ y') dxdy = PO ( x', y') (5) So the outputted image is same to the inputted image; in the formula, the minus in the front of x', y' means the image is inverted. 2.2. The lateral magnification and the axial magnification of the imaging system According to the relation between the lateral magnification and the axial magnification of ideal optical imaging system, in the ideal acoustic imaging system, the lateral magnification and the axial magnification also have the relation as follows: α 2 f ' n' 2 = β = f n β (6) In the formula, α is the lateral magnification of the imaging system, β is the axial magnification, f is the object focal length, f ' is the image focal length, n ' is the refractivity of medium in object space, and n is the refractivity of medium in image space. If the medium in object space is same to the image space, the formula (6) can be simplified as α The equation shows that the image of a cube isn t a cube, except the cube is on the position of β =± 1.In the 4f imaging system, α = 1because of β = 1, the image of a cube is also a cube. Therefore, the 4f acoustic imaging system guarantees axial and lateral unit magnification of the images, a displacement of the object plane by a distance ΔZ results in a displacement of the corresponding focused image plane by a distance Δ Z, just as Fig. 3 shows. In addition, the acoustic lens has the property of long focal depth, which means that the distance ΔZ of the image plane of the lateral magnification α = 1 will be long, about 1-2cm. And it can be described by [17] 2 = β (7) 2 2 λf (0, ) exp( 2 ( ) j a z h z j π f z π Δ Δ = +Δ )[1 exp( )] 2 jπ a Δz λ λ f ( f +Δz) (8) (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4969

Where f is the focal length of an acoustic lens, a is the radius of the acoustic lens, and λ is the wavelength of sound. Fig. 3. The principle of the acoustic lens has the property of long focal depth. 3. Imaging system and method Figure 4 shows the experimental setup. A Q-switched Nd: YAG laser is adopted to provide 1064 -nm laser pulses with a FWHM of 8 ns. The laser beam is expanded to heat the imaging object and produce the PA signal. The distribution of optical absorption generates proportionate PA waves. The acoustic pressure distribution is imaged onto the image planes by a 4f acoustic lens in the scattering media. Into the image plane, a linear detector detects the acoustic pressure distribution. The linear detector consists of 64 PZT probes with a center frequency of 1 MHz frequency. The radius of every probe is about 0.25mm, and the central distance between two neighboring probes is 1.5 mm. And the linear detector is fixed on a scanning stage. The scanning stage, which is controlled by a computer, drivers the linear detector vertically to detect the planar PA signals. The signals are chosen and amplified in turn by the electronic switch of 64 lines and input a Boxcar (made in American Stanford Research System Inc. Company). The Boxcar triggered by the laser Q switch integrates the signals during a fixed delay time, when the peak value of the signal is transformed to the corresponding voltage of dc. These voltage signals are sent to a computer through an acquisition card (ADC, Model: Advantech PCL-818HG) and transformed into corresponding 256 gray levels to display the image. An oscillograph (Model: TDS1002) monitors the PA signals and the sampling gate of the Boxcar. Fig. 4. Experiment setup of PAT imaging based on 4f acoustic lens. (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4970

Similar to an optical imaging system, the PA signals from an object plane need the same delay time to reach the image plane. So the Boxcar is used to hold on a fixed delay time to acquire the PA signal, which guarantees the acquired signals are from the same image plane. So we can reconstruct the different absorption distribution planes by detecting the acoustic pressure on the corresponding focused image planes. However, because of long focal depth of an acoustic lens, a determined object plane can be precisely imaged onto the imaging plane within a range distance inside focal depth by an acoustic lens. The PA signals from different planes require different time to reach a same detecting plane. Therefore, on this same detecting plane, we can use the time-resolved technique to distinguish photoacoustic signals from different planes. So, onto one detecting plane, we can get corresponding photoacoustic tomography images of different object planes without moving the detector. 4. Experimental results and discussion 4.1. Two-dimensional (2D) PAT imaging To demonstrate the feasibility of IOT imaging system, a series of experiments were conducted with turbid phantoms prepared from an insoluble in water milk. It is demonstrated that the 4f acoustic imaging system has the ability of 2D PAT imaging. A sample such as Fig. 5(a), which three black adhesive tape points was adhered to a piece of polymethylmethacrylate, was in 3% milk. The sample was heated by laser, and PA signals from the sample are imaged on the imaging plane by the acoustic lens. The imaging plane is scanned by a detector to signals to reconstruct the PAT image. The PAT image is smaller than the sample because of the limited probes of the detector. The original image is magnified six times as Fig. 5(b). The acoustic image is in a perfect accordance with the sample, it demonstrates the 2D imaging ability of this system. (a) (b) Fig. 5. The sample of three black adhesive tape points adhered to a piece of polymethylmethacrylate hidden inside scattering media (a) and the corresponding PA image (b). As Fig. 6(a) and Fig. 7(a) show the samples made of black adhesive tape are the patterns V and C, and they are glued on the polymethylmethacrylate. The PAT image is obtained as Fig. 6(b) and Fig. 7(b) by scanning on the image plane. The PA images are in a perfect (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4971

accordance with the sample; they are clear and contrast sharply with their backgrounds. All the sizes of the images are approximately equal to the original objects, and it shows the lateral magnifications are unit for the 4f acoustic lens. The PA image is inverted compared to that of object for the acoustic lens forms the inverted real image. Otherwise, the pictures were altered little by the finite aperture and lens aberrations. So, in other to obtain the clear image, the acoustic lens should be revised and a 2D ultrasonic detector array with multiple small elements should be developed. (a) (b) Fig. 6. The sample made of black adhesive tape is the patterns V adhered to a piece of polymethylmethacrylate hidden inside scattering media (a) and the corresponding PA image (b). (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4972

(a) (b) Fig.7. The sample made of black adhesive tape is the patterns C adhered to a piece of polymethylmethacrylate hidden inside scattering media (a) and the corresponding PA image (b). 4.2. The resolution of the imaging system Similar to optical imaging system, the impulse response of the imaging system shows the characteristic of imaging. In experiment, the pulse laser heated a black adhesive tape point and generated the PA waves. The black adhesive tape point s diameter is about 0.8 mm, and it acted as a point acoustic source. Figure 8(a) is the black adhesive tape point, and Fig. 8(b) is the impulse response curve of experimental result. According to Rayleigh criterion for resolution, the lateral resolution equals to the full width at half maximum of the impulse response curve, and it can be worked out about 4mm. (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4973

(a) (b) Fig. 8. The sample of a black adhesive tape point. It acted as a point acoustic source heating by the pulse laser (a) and the impulse response curve (b). In experiment, the lateral resolution can also be measured by adjusting the distance between two points. Experimental result shows that the imaging system can resolve two points when the distance is 3mm between them, and the PA image is as Fig. 9. And the PA image is only one big point when the distance between two points is less than 3mm. There are some error between the above two data but they are approximately equal. Fig. 9. The PA images that two points are 3mm distance. 4.3. The imaging property of the acoustic lens on axis In the imaging modality, the three-dimensional object is sliced, and every slice is a 2D x-y plane. Along the z-axis, the three-dimensional object is segmented some x-y planes. The imaging property of the x-y plane has been shown, and the image is equal to the object. Here, (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4974

the imaging property of the z-axis will be discussed. The sample, which two different patterns are adhered to the front as well as at the back of a piece of polymethylmethacrylate, is adopted in the experiment, just as Fig. 10. The polymethylmethacrylate is about 15mm thick, so the distance between the two object planes is about 15mm. Figure 11 shows the PA signals of two different imaging planes. According to the axial unit magnification of the 4f acoustic lens, the distance between the two object planes must be equal to the distance between two image planes. The distance of image planes can be expressed as D = v t (9) where v is the acoustic velocity, and t is the time difference of two PA signals reached the detector. Fig. 10. The photos of the sample that the two different patterns are adhere to the front as well as at the back of a piece of polymethylmethacrylate. 5 5 6 Figure 11 shows the time difference is t = 8.21 10 s 7.65 10 s = 5.6 10 s = 5.6μs, and v = 1.49 mm/ μ s, so D = v t = 1.49 5.6mm= 8.344( mm). The distance between the two object planes is about 15mm in polymethylmethacrylate, so it must be transformed to the distance in solution. The transforming relation is d = s / n, where s is the distance between the two object planes, n is a relative refraction index, polymethylmethacrylate and n = v = 2.7 mm / μs = 1.8. So the distance vwater 1.49 mm/ μs is d = s/ n = 15mm = 8.333mm. This just proves the distance between the two object planes 1.8 is equal to the distance between two image planes, and the acoustic lens has the property of axial unit magnification. (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4975

Fig.11. The PA signals of two image planes that are from two respective object planes The imaging modality is to slice the three-dimensional object, then how close two slices can be distinguished? It is the depth resolution (DR), and the DR can be expressed as 17 DR = τ v (10) where τ is the temporal impulse width of PA signals and v is the velocity of PA signals in the biological tissue. In general, v is a constant, thus DR is determined byτ, where τ relates to the response characteristic of the detector. 5. Conclusion The PAT imaging theory of a 4f acoustic lens imaging system has been present, and the theoretical results show that a 4f acoustic lens is able to image the PA pressure distribution onto an image space and guarantees axial and lateral unit magnification of image. Experimental results demonstrate that the system is able to obtain PAT images of samples inside optically turbid medium, and the reconstructed images are agree well with the sample. Furthermore, in 3D PAT imaging, the IOT imaging system may provide the PA images with the same magnification on axial and lateral, and the 3D structure of the object can be imaging really. Acknowledgments This work is supported by the National Natural Science Foundation of China (grant No.60377009), National 863 Program Project of China (grant No. 2006AA02Z4B4) and the Natural Science Foundation of Guangdong Province, China (Grant No.05005926). (C) 2007 OSA 16 Apr 2007 / Vol. 15, No. 8 / OPTICS EXPRESS 4976