Bio-Potential Amplifiers

Similar documents
Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

Lecture 4 Biopotential Amplifiers

Electrocardiogram (ECG)

Biomedical. Measurement and Design ELEC4623. Lectures 9 and 10 Practical biopotential amplifier design and multilead ECG systems

Amplificador de Biopotencial

BIOMEDICAL INSTRUMENTATION PROBLEM SHEET 1

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

Development of Electrocardiograph Monitoring System

ECG Project. Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege

EE 230 Experiment 10 ECG Measurements Spring 2010

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

Indigenous Design of Electronic Circuit for Electrocardiograph

IMPROVEMENTS IN ELECTROCARDIOGRAPHY SMOOTHENING AND AMPLIFICATION

Analog Circuits and Systems

Lab E5: Filters and Complex Impedance

Signal Conditioning Systems

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Introduction to Electronic Circuit for Instrumentation

GBM8320 Dispositifs Médicaux Intelligents

Differential Amplifier : input. resistance. Differential amplifiers are widely used in engineering instrumentation

Development of an alternative method for the calibration of ECG simulators

Bioelectric Signal Analog Front-End Module Electrocardiograph

TRANSDUCER INTERFACE APPLICATIONS

*Notebook is excluded

Operating and Maintenance Instructions. Harvard Isolated Preamplifier

Changing the sampling rate


Biopotential Amplifiers. Hsiao-Lung Chan, Ph.D. Dept Electrical Engineering Chang Gung University, Taiwan

LINEAR IC APPLICATIONS

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

Lecture 14 Interface Electronics (Part 2) ECE 5900/6900 Fundamentals of Sensor Design

EECE Circuits and Signals: Biomedical Applications. Lab ECG I The Instrumentation Amplifier

A Design Of Simple And Low Cost Heart Rate Monitor

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

AD8232 EVALUATION BOARD DOCUMENTATION

Concepts to be Reviewed

Special-Purpose Operational Amplifier Circuits

Wireless Transmission of Real Time Electrocardiogram (ECG) Signals through Radio Frequency (RF) Waves

BENG 186B Winter 2012 Quiz 3. March 7, NAME (Last, First): This quiz is closed book and closed note. You may use a calculator for algebra.

Chapter 4 4. Optoelectronic Acquisition System Design

EXPERIMENT 5 Bioelectric Measurements

Jordan University of Science & Technology Faculty of Engineering. Department of Biomedical Engineering BME 443. Biomedical Instrumentation Lab I

Operational Amplifiers

6.101 Introductory Analog Electronics Laboratory

This is an outline of the subjects that will be touched upon during this presentation.

BENG 186B Winter 2013 Final

STM32 microcontroller core ECG acquisition Conditioning System. LIU Jia-ming, LI Zhi

Crossover Distortion (hole) Amplifier Classification. Amplifiers: Class A, B, AB, D Op-amp Application: ECG. Why is [b] better? - LF356 - LF

EMG Electrodes. Fig. 1. System for measuring an electromyogram.

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

ECE 480 Design Team 6 Electrocardiography and Design

EE105 Fall 2015 Microelectronic Devices and Circuits Multi-Stage Amplifiers. Prof. Ming C. Wu 511 Sutardja Dai Hall (SDH)

Applied Electronics II

ECG Measurement and Analysis

Instrumentation amplifier

Massachusetts Institute of Technology MIT

ELECTRONICS. EE 42/100 Lecture 8: Op-Amps. Rev B 3/3/2010 (9:13 PM) Prof. Ali M. Niknejad

NOISE REDUCTION TECHNIQUES IN ECG USING DIFFERENT METHODS Prof. Kunal Patil 1, Prof. Rajendra Desale 2, Prof. Yogesh Ravandle 3

ELECTRONICS. EE 42/100 Lecture 8: Op-Amps. Rev A 2/10/2010 (6:47 PM) Prof. Ali M. Niknejad

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

AN4995 Application note

DUAL ULTRA MICROPOWER RAIL-TO-RAIL CMOS OPERATIONAL AMPLIFIER

EDL Group #3 Final Report - Surface Electromyograph System

instead we hook it up to a potential difference of 60 V? instead we hook it up to a potential difference of 240 V?

MAHARASHTRA STATE BOARD OF TECHNICAL EDUCATION (Autonomous) (ISO/IEC Certified) WINTER 16 EXAMINATION Subject Code:

2.996/6.971 Biomedical Devices Design Laboratory Lecture 7: OpAmps

Lecture 4. Integrated Electronics

AM Generation High Level Low Level

HUMAN DETECTION AND RESCUE USING BIO POTENTIAL SIGNALS

InstrumentationTools.com

LM2900 LM3900 LM3301 Quad Amplifiers

A PC Based Cost Effective Advanced Cardio Signals Monitoring System

High Common-Mode Voltage Difference Amplifier AD629

Simple Heartbeat Monitor for Analog Enthusiasts

Dimensions in inches (mm) .268 (6.81).255 (6.48) .390 (9.91).379 (9.63) .045 (1.14).030 (.76) 4 Typ. Figure 1. Typical application circuit.

Electrocardiogram (EKG) Data Acquisition and Wireless Transmission

INVITATION FOR QUOTATION. TEQIP-II/2013/UP1G01/Shopping/43

HA-2600, HA Features. 12MHz, High Input Impedance Operational Amplifiers. Applications. Pinouts. Ordering Information

A Comprehensive Model for Power Line Interference in Biopotential Measurements

Fetal ECG Extraction Using Independent Component Analysis

Lecture 2 Analog circuits...or How to detect the Alarm beacon

Removal of Power-Line Interference from Biomedical Signal using Notch Filter

ELG3336 Design of Mechatronics System

Introduction to Op Amps

4. Differential Amplifiers. Electronic Circuits. Prof. Dr. Qiuting Huang Integrated Systems Laboratory

Homework Assignment 03

ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL AMPLIFIERS

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS

ECE 442 Solid State Devices & Circuits. 15. Differential Amplifiers

Medlab GmbH EG04000 User Manual. medlab. Four Lead ECG OEM board EG Technical Manual. Copyright Medlab Version Version 1.

Laboratory Project 1: Design of a Myogram Circuit

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

Biomedical Signal Processing and Applications

UNIVERSITY OF UTAH ELECTRICAL AND COMPUTER ENGINEERING DEPARTMENT ELECTROMYOGRAM (EMG) DETECTOR WITH AUDIOVISUAL OUTPUT

Gechstudentszone.wordpress.com

EXPERIMENT 8 Bio-Electric Measurements

Name Kyla Jackson, Todd Germeroth, Jake Spooler Date May 5, 2010 Lab 3E Group 3 Experiment Title Project Deliverable 3

1. What is the unit of electromotive force? (a) volt (b) ampere (c) watt (d) ohm. 2. The resonant frequency of a tuned (LRC) circuit is given by

Description. Output Stage. 5k (10k) - + 5k (10k)

Transcription:

Bio-Potential Amplifiers

Biomedical Models for Diagnosis Body Signal Sensor Signal Processing Output Diagnosis Body signals and sensors were covered in EE470 The signal processing part is in EE471 Bio-Potential Amplifier Biological Potential difference Signal processing includes amplification and filtering Why Amplification? Why filtering?

Basic Requirements for Amplifiers Type of amplification: Voltage Amplification Current Amplification High input impedance ( 10 MW) why? Isolation and protection circuits why? Low output impedance why? High common mode rejection ration why? The appropriate frequency spectrum, SNR, gain, Calibration input to calibrate the amplifier

Example of Bio-Potential Amplifier ECG Amplifier Origins of the electrocardiogram Blood Cycle The cardiac vector The ECG waveform Indicator of a good ECG 12-lead electrocardiography

Blood Cycle of the Heart Objectives of the cycle: Provide Oxygenated blood to all body cells Remove Carbon-Dioxide accumulating in cells Two Blood Cycles simultaneously Pulmonary cycle add oxygen and remove Co 2 Blood Cycle carry oxygen to body cells

Normal heart as a pump and muscular structure From scienceclarified.com

The Heart Beat (Cardiac cycle) Two Phases 1. Two atriums contract and two ventricles relax 2. Two ventricles contract and two atriums relax 3. Four chambers relax Terminology: Systolic phase contraction Diastolic phase relaxation Control is done via an independent nervous system in the heart though electrical signals

Conduction mechanism of the heart

Sequence of Control Signals SA node (heart pace maker) initiate signal to atriums causing it to contract blood flows through valves to ventricles Impulse flows till it reaches AV node AV node fires a signal along the bundle of his and ventricles contract in top to bottom fashion to push the blood out of the heart http://en.wikipedia.org/wiki/file:ecg_princi ple_slow.gif#file

Origin and propagation of the action potential in the cardiac muscles

Components of ECG The ECG is drawn against time. It contains elements indicating the temporal relationship between different actions taking place during a cardiac cycle. The respective amplitudes of these elements and their respective positions are studied Picture from wikipedia.com

ECG Amplifiers The electrical activity of the heart can be modeled as an Electric dipole in the thorax This dipole is represented with a vector that varies in amplitude and direction (cardiac vector) Defined lead vectors: unit vectors with fixed orientation Vector can be observed from different directions and angles

12-Lead Electrocardiography The electrocardiogram is measured by putting electrodes on specific locations on the body. The electrodes used, their respective wires and resistors for measuring the ECG from a particular direction is called the lead Measurement in Frontal Plane Bipolar limb leads, I, II, III Unipolar limb leads, av R, av L, av F Precordial (chest) leads for transverse plane (V1-V6)

Bipolar limb leads Projection of the cardiac vector in the frontal plane is obtained and the projection vector is studied in three directions that are 60 degrees apart. The directional components are called lead I, II and III as they are measured between LA and RA, LL and RA, and LL and LA resp. In all measurements the right leg (RA) is used as the reference ground for the differential amp. Lead vectors I, II and III form an equilateral triangle. At any given time I II + III = 0

Three limb electrodes are connected together through equal valued resistors to obtain a terminal whose voltage is the average of three voltages. It is equivalent to placing an imaginary electrode to the centre of the Einthoven s triangle. Uni-Polar limp leads Wilson s central terminal

Augmented unipolar limb leads 33% increase in voltage measured by disconnecting the measuring electrode from the Wilson s terminal without affecting the direction of the lead vector. Three new configurations thus obtained as avr, avl and avf. a stands for augmented. A resistance of R/2 is added to balance the input impedances seen by both inputs of the diff. amplifier.

Homework 1 Bio-Potential Amplifiers and ECG leads 1. Explain why Bio-Potential Amplifiers need to have the following requirements: High Input Impedance Low output impedance Frequency response appropriate to the signal Isolation Circuit 2. Show that the voltage at wilson central point (figure 6.4) is the average of the voltages at each node 3. Show that voltage at the augmented lead shown in figure 6.5 increases the output voltage and calculate this increase Notes: Homework due at beginning of Tue Oct 20 th class Late submission are subject to 10% decrease for everyday after the class No late submissions would be accepted after Sun Oct 24 th class

Bio-Potential Amplifier II

Electrode connections in bipolar limb leads

Electrode connections in uni-polar limb leads

Measurement in transverse plane (top view) Six positions are selected from the right of the manibrium to the mid auxiliary line Electrodes are placed over the chest and voltages are measured with respect to the Wilson s central terminal

Electrode connections in unipolar chest leads

Voltage and frequency ranges of some common bio-potential signals

ECG Amplifier requirements Protection circuit (zener diodes, gas-discharge tube) Lead selector switch (can be controlled by a microprocessor) Calibration signal (1mV) Preamplifier: high input impedance, high CMRR, gain selector. Isolation circuit: protect subjects from 50-60 Hz current Driven right leg circuit Driver amplifier: contains BPF to remove dc offset, amplifies signal to appropriate level. Memory system: samples of each lead are stored Microcomputer Recorder-printer: provides hard copy of the signal

Block diagram of an earlier version of an electrocardiograph

Problems frequently encountered in electrocardiography Distortion in the Signal Frequency distortion Saturation or cut-off distortion Ground loops Open lead wires Artifacts from large electrical transients Interference on signal Interference from electrical devices Electromagnetic interference Interference from other biological signals

Frequency distortion The high f components of the signal are attenuated yielding a rounding of sharp edges and dropping in the R-wave magnitude. This is a distortion in the signal due to high f limitation. A distortion due to low f limitation. The signal looks like a differentiated one and the stable baseline needed for the clinical ECG is lost

Saturation or cut-off distortion

Ground loops and their elimination Solution Connect the grounds for both machines together

Artifacts from large electrical transients Example: Defibrillator

Solution: Transient protection A voltage protection scheme at the input of an ECG amplifier

Solution Voltage-limiting devices used for input protection Current-voltage characteristic Parallel silicon diodes How do voltage limiting devices protect the input of the amplifier? 2-20 V 50-90 V Back-to-back silicon zener diodes Gas discharge tube (neon light)

60-Hz power line and electromyography interference 60-Hz power-line interference on the ECG Electromyography interference on the ECG Solution: Proper filter design

Lead Dropping Lead drop = No ECG = Patient is Dead!! Medical staff not EE specialists Training example Solution Lead drop detector circuit

Let go current Effect of Frequency Higher frequency, less dangerous Why don t we generate electricity at high frequencies?

Open lead wire detector Lead wire breaks or poorly contacting the body

2-Generation and effects of magnetic field Protection through alternative path for the current

Currents Picked up in the Body 1-Magnetic field pickup Lead wires for lead I making a closed loop with patient and ECG machine Solution Twisting lead wires together and keeping them close to body minimizes interference

Currents Picked up in the Body 2-Pickup due to displacement current flowing through the patient v A v B = v cm Z Zin + Z in Z Z + in 1 in Z 2 Z 1 and Z 2 << Z in v = Z Z Z 2 1 va B vcm in v = i cm Solution Choose High input impedance Amplifier db Z G

Currents Picked up in the Body 3-Electrical field pickup by connecting wires and instrument What is a Capacitor? Coupling between hot side of the power line and lead wires v A v B = i d1z1 id 2Z2 v A v B = With i d1 i d2 ) i d1( Z1 Z2 How to minimize the current picked up by lead wires and instrument?

Currents Picked up in the Body 4-Electromagnetic interference EM waves generated by Radar facilities X-ray machines Nearby transformers Radio waves EM waves picked-up by patient and lead wires Demodulated by p-n junctions of transistors and/or electrode-electrolyte interfaces Modulating audio signal appears as interference on top of the ECG signal Solution: Can be eliminated by shunting the input terminals of the ECG amplifier with a small capacitor (around 200pF)

Solution: Electrostatic shielding

Driven-right-leg circuit Right leg connected to the output of an OPAMP instead of ground Why? 1. Displacement current flows through output resistance instead of body 2. Patient un-grounded when high voltage appears between patient and ground (R f and R o large values)

Analysis of the driven-right-leg circuit

Bio-Potential Amplifiers 3 Electrical and Computer Engineering (Biomedical 45

2-Generation and effects of magnetic field Protection achieved through alternative low resistance paths for the current Electrical and Computer Engineering (Biomedical 46

Currents Picked up in the Body 1-Magnetic field pickup Lead wires for lead I making a closed loop with patient and ECG machine Electrical and Computer Engineering (Biomedical 47 Solution Twisting lead wires together and keeping them close to body minimizes interference

Currents Picked up in the Body 2-Pickup due to displacement current flowing through the patient v A v B = v cm Z Zin + Z in Z Z + in 1 in Z 2 Z 1 and Z 2 << Z in v v = i cm = Z 2 1 va B vcm in db Z Z Z Solution Choose High input impedance Amplifier G Electrical and Computer Engineering (Biomedical 48

Currents Picked up in the Body 3-Electrical field pickup by connecting wires and instrument What is a Capacitor? Coupling between hot side of the power line and lead wires v A v B = i d1z1 id 2Z2 v A v B = With i d1 i d2 ) i d1( Z1 Z2 How to minimize the current picked up by lead wires and instrument? Electrical and Computer Engineering (Biomedical 49

Currents Picked up in the Body 4-Electromagnetic interference EM waves generated by Radar facilities X-ray machines Nearby transformers Radio waves EM waves picked-up by patient and lead wires Demodulated by p-n junctions of transistors and/or electrode-electrolyte interfaces Modulating audio signal appears as interference on top of the ECG signal Solution: Can be eliminated by shunting the input terminals of the ECG amplifier with a small capacitor (around 200pF) Electrical and Computer Engineering (Biomedical 50

Solution: Electrostatic shielding Electrical and Computer Engineering (Biomedical 51

Driven-right-leg circuit Right leg connected to the output of an OPAMP instead of ground Why? 1. Displacement current flows through output resistance instead of body 2. Patient un-grounded when high voltage appears between patient and ground (R f and R o large values) Electrical and Computer Engineering (Biomedical 52

Analysis of the driven-right-leg circuit Electrical and Computer Engineering (Biomedical 53

KCL at point x Driven Right Leg Circuit i.e. 2v R cm a + v R o f = 0 But Therefore v cm v o 2R = R cm a RL f d v cm v = R i + v R = R 1+ 2 RL f R a i d o What is the effective resistance between right leg and ground? Large transients Saturation chose large R f and R o =~ 5 MΩ Regular operation want v cm as small as possible large R f and small R a Electrical and Computer Engineering (Biomedical 54

Bio-Potential Amplifiers -4

Example of a simple ECG amplifier

Examples of Bio-Potential Amlifiers

Biomedical Signal Processor Examples Cardiac Tachometers Electromyogram integrators Fetal electrocardiography Cardiac monitors Biotelemetry

Averaging type Tachometers

Beat-to-beat Tachometers

Tachometers

EMG integrators Counter P 1 Monostable multivibrator Switch Comparator EMG υ 1 Absolutevalue circuit C R υ 2 υ 3 + v t Integrator

EMG integrators

Fetal electrocardiology

A technique for isolating fetal ECG from maternal

Cardiac monitors Patient Electrodes Preamplifier Isolation Amplifier Communication port RAM Display screen Analog to digital converter Bus Microcomputer CPU Program PROM Chart recorder Storage medium Keyboard Alarm indicator

Frequency modulation Telemetry

Time division multiplexing Telemetry

Digital landline telemetry system

Three-channel time-division multiplexed radiotelemetry transmitter Example of output waveform from commutator

Three-channel time-division multiplexed radiotelemetry receiver

Three-channel frequency-division multiplexed radiotelemetry system