PEARSON ELECTRONICS POWERLINE RIPPLE DETECTOR MODEL PRD-120 OPERATING MANUAL

Similar documents
CS101. Conducted Susceptibility CS101. CS101 Maximum Current. CS101 Limits. Basis For CS101 Limits. Comparison To MIL-STD Vdc or Less

Specification for Conducted Emission Test

Trees, vegetation, buildings etc.

Model 4402B. Ultra-Pure Sinewave Oscillator 1Hz to 110kHz Typical Distortion of % Serial No. Operating Manual

Model 34A. 3Hz to 2MHz 2-Channel Butterworth/Bessel HP, LP, BP, BR Plug-In Filter Card for Model 3905/3916 Chassis.

Essential Capabilities of EMI Receivers. Application Note

Advanced Test Equipment Rentals ATEC (2832)

Spectrum Analysis with HP-IB Systems

Magnetic-Field Test System / Low-Frequency Test System for Emission and Immunity Tests / MTS-800

Keysight Technologies Essential Capabilities of EMI Receivers. Application Note

Operational Amplifiers

6.101 Project Proposal April 9, 2014 Kayla Esquivel and Jason Yang. General Outline

ECE 6416 Low-Noise Electronics Orientation Experiment

OPERATING AND MAINTENANCE MANUAL

Multi-function Gain-Phase Analyzer (Frequency Response Analyzer) Model 2505

3.2 Measuring Frequency Response Of Low-Pass Filter :

Frequency and Time Domain Representation of Sinusoidal Signals

Model 7000 Low Noise Differential Preamplifier

ELEC 0017: ELECTROMAGNETIC COMPATIBILITY LABORATORY SESSIONS

CALIBRATED IMPULSE GENERATOR MODEL CIG khz 1 GHz

DEPARTMENT OF DEFENSE HANDBOOK

EXPERIMENT 1: Characteristics of Passive and Active Filters

Active Filter. Low pass filter High pass filter Band pass filter Band stop filter

LPF-100 Composite Low Pass Filter

OPERATION & SERVICE MANUAL FOR FC 110 AC POWER SOURCE

The object of this experiment is to become familiar with the instruments used in the low noise laboratory.

UNIVERSITY OF NORTH CAROLINA AT CHARLOTTE Department of Electrical and Computer Engineering

A Simple Notch Type Harmonic Distortion Analyzer

1. General Outline Project Proposal April 9, 2014 Kayla Esquivel and Jason Yang

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

FISCHER CUSTOM COMMUNICATIONS, INC.

COUPLING DECOUPLING NETWORK MODEL CDN-AF4

FFT Spectrum Analyzer

Twelve voice signals, each band-limited to 3 khz, are frequency -multiplexed using 1 khz guard bands between channels and between the main carrier

Agilent N9343C Handheld Spectrum Analyzer (HSA)

MIL-STD 461F Results for M&A Technology Companion epad Computer

ECE3204 D2015 Lab 1. See suggested breadboard configuration on following page!

LBI-30398N. MAINTENANCE MANUAL MHz PHASE LOCK LOOP EXCITER 19D423249G1 & G2 DESCRIPTION TABLE OF CONTENTS. Page. DESCRIPTION...

Lecture 2 Analog circuits. IR detection

ME 365 EXPERIMENT 7 SIGNAL CONDITIONING AND LOADING

Introduction (cont )

INSTRUMENTS, INC. Model 2960AX Disciplined Quartz Frequency Standard 2960AX. Section Page Contents

Testing Power Factor Correction Circuits For Stability

Measurement and Analysis for Switchmode Power Design

Measuring Power Line Impedance

ERICSSONZ LBI-30398P. MAINTENANCE MANUAL MHz PHASE LOCKED LOOP EXCITER 19D423249G1 & G2 DESCRIPTION TABLE OF CONTENTS

The Value of Pre-Selection in EMC Testing. Scott Niemiec Application Engineer

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

GSM Transmitter Modulation Quality Measurement Option

Sallen-Key_High_Pass_Filter -- Overview

Testing Power Sources for Stability

Glossary of VCO terms

Spectrum Analyzer. EMI Receiver

ASC-50. OPERATION MANUAL September 2001

ECE 4670 Spring 2014 Lab 1 Linear System Characteristics

Experiment 1: Instrument Familiarization (8/28/06)

ELC224 Final Review (12/10/2009) Name:

USER. manual. Falco Systems WMA-100. High Voltage Amplifier DC - 500kHz

8 Hints for Better Spectrum Analysis. Application Note

Lecture 2 Analog circuits. Seeing the light..

Broadband Power Amplifier

3100LA Broadband Power Amplifier

STATION NUMBER: LAB SECTION: Filters. LAB 6: Filters ELECTRICAL ENGINEERING 43/100 INTRODUCTION TO MICROELECTRONIC CIRCUITS

Experiment 1: Instrument Familiarization

Lab 9: Operational amplifiers II (version 1.5)

EE 2274 RC and Op Amp Circuit Completed Prior to Coming to Lab. Prelab Part I: RC Circuit

LCM100 USER GUIDE. Line Carrier Modem INDUSTRIAL DATA COMMUNICATIONS

INSTRUCTION MANUAL For LINE IMPEDANCE STABILIZATION NETWORK. Model LI khz to 10 MHz

RIGOL Data Sheet. DG3000 Series Function/Arbitrary Waveform Generator DG3121A, DG3101A, DG3061A. Product Overview. Easy to Use Design.

2100L Broadband Power Amplifier

New Features of IEEE Std Digitizing Waveform Recorders

BUCK Converter Control Cookbook

2. BAND-PASS NOISE MEASUREMENTS

Keysight Technologies N9320B RF Spectrum Analyzer

Sampling and Reconstruction

SC5407A/SC5408A 100 khz to 6 GHz RF Upconverter. Datasheet. Rev SignalCore, Inc.

Agilent N9320B RF Spectrum Analyzer

Agilent 83711B and 83712B Synthesized CW Generators

Hints. for making. Better. Spectrum Analyzer. Measurements. Application Note

MODEL 5002 PHASE VERIFICATION BRIDGE SET

LFR: flexible, clip-around current probe for use in power measurements

GA GHz. Digital Spectrum Analyzer

Guide Version Five techniques for fast, accurate power integrity measurements

Four-Channel Differential AC Amplifier

ELECTRIC FIELD PROBE ANTENNA MODEL PEF-10A. 20 Hz 1 MHz

For the National Voluntary Laboratory Accreditation Program

ROD ANTENNA TESTING Complete article download from: EMI TESTING. Basic RE102 test (2-30 MHz)

Low Pass Filter Introduction

Frequency Selective Circuits

Agilent ESA-L Series Spectrum Analyzers

PXA Configuration. Frequency range

Chambers Accessories Equipment 1 Equipment 2 Amplifiers Antennas Emission

RECOMMENDATION ITU-R SM Method for measurements of radio noise

INSTRUMENTS, INC. Models 2960AR and 2965AR Disciplined Rubidium Frequency Standards. Section Page Contents

UNIVERSITY OF PENNSYLVANIA EE 206

DC/DC Converter Stability Measurement

note application Measurement of Frequency Stability and Phase Noise by David Owen

LIC & COMMUNICATION LAB MANUAL

Agilent 8360B Series Synthesized Swept Signal Generators 8360L Series Synthesized Swept CW Generators Data Sheet

Tuned Radio Frequency Receiver (TRF) The most elementary receiver design, consisting of RF amplifier stages, detector and audio amplifier stages.

Transcription:

PEARSON ELECTRONICS POWERLINE RIPPLE DETECTOR MODEL PRD-120 OPERATING MANUAL Description. The Pearson PRD-120 facilitates the measurement of power line ac ripple using a frequency selective voltmeter, i.e., EMI receiver or spectrum analyzer. In so doing, it greatly simplifies both the manual and automated measurement of injected ripple in such standards as MIL-STD- 461 CS01/CS101, RTCA/DO-160 section 18 and MIL-HDBK- 704-2 through -6. PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 1

Features. The PRD-120 is rated for use on any power bus, ac or dc, up to and including a potential of 125 Volts (dc or rms). Other models are available for other bus potentials. The PRD-120 provides the test engineer with the following built-in functions: Signal conditioning so that both power bus potential and ripple can be monitored on a sensitive frequency selective voltmeter Isolation between power bus and frequency selective voltmeter safety/chassis ground Capability to monitor ripple across both test sample and power source For MIL-STD-461 CS101, a switch selectable function which generates a constant output across the entire CS101 frequency range, even above 5 khz where the limit changes with frequency Principle of Operation. The PRD-120 is a passive signal conditioner, providing dc blocking, selectable ac attenuation, and transformer isolation to allow connection of a sensitive frequency selective voltmeter across the power bus in lieu of the traditional oscilloscope. See Figures 2a and b. Note that in Figure 2b, the spectrum analyzer/emi receiver is not powered through an isolation transformer, because the required isolation is built into the PRD-120. The spectrum analyzer/emi receiver operates with safety ground intact, while maintaining the isolation of the test sample power return above ground. Another difference between the Figure 2a and b drawings is that the PRD-120 connects across the test sample, as required in Figure 2a, and also across the power source. This added feature provides greater test control. By inspection, ripple injected by the coupling transformer must drop either across the test sample PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 2

(desired), or across the power source (undesired). The 10 µf capacitor is an attempt to force injected ripple to drop across the test sample, but it is ineffective below about 10 khz. If the required ripple potential cannot be established across the test sample when driving the required power, a flip of the EUT- Power source switch allows the test engineer to measure the ripple potential across the power source. If the power source ripple level is such that added to the test sample ripple results in the required ripple level, then steps may be taken to reduce the power source impedance, allowing enough of the injected ripple to fall across the test sample to meet the requirement. Stimulation and Monitoring Equipment EUT Isolation XFMR Oscilloscope Power Leads Signal Generator Power Amplifier Coupling XFMR 10 uf LISN high return LISN Figure 2a: Traditional audio frequency injection test set-up Stimulation and Monitoring Equipment EUT Power Leads Spectrum Analyzer/ EMI RCVR Signal Generator Power Amplifier Coupling XFMR PRD 10 uf LISN high return LISN Figure 2b: Audio frequency ripple injection using PRD Figures 3a and b demonstrate that with the PRD-120, ripple is visible on the frequency selective voltmeter display as an entirely separate signal, rather than low-level superimposed ripple riding on the ac bus potential waveform. The PRD-120 has a nominal -66 db transfer function, meaning that 66 db must be added to the signal levels shown in Figure 3a to properly interpret them. Each PRD-120 is individually calibrated and the actual transducer factor supplied with an individual serial number unit must be applied. The PRD-120 makes available not only frequency domain selection and discrimination, but also the higher dynamic range (~80 db) of PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 3

the typical frequency selective voltmeter with logarithmic amplitude display. This is in contrast to about 40 db of usable Figure 3a: View of 6.3 Vrms 800 Hz ripple on a 400 Hz 115 Vrms bus, using the PRD-120 Figure 3b: View of the same ripple on the same bus, using an oscilloscope Figure 4a: View of 6.3 Vrms 100 Hz ripple on a 400 Hz 115 Vrms bus, using the PRD-120 Figure 4b: Envelope of the same 100 Hz ripple on the same bus, using an oscilloscope on-screen dynamic range with an oscilloscope. Figures 4a and b illustrate that it is just as easy to inject ripple below the power bus fundamental frequency as above. This is immediately PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 4

useful for MIL-HDBK-704-2 through -6 testing, and restores to MIL-STD-461 a capability that was lost in 1993.The PRD-120 transducer factor can be either frequency independent, or, at the flip of the right-hand switch visible in Figure 1, the attenuation above 5 khz rolls off at 20 db/decade, to match the slope of the MIL-STD-461D/E/F CS101 limit. With this switch position selection, the output of the PRD-120 into the spectrum analyzer/emi receiver is independent of frequency across the entire CS101 test frequency range, at a nominal 70 dbµv, or 3 mv if testing to MIL-STD-461F Figure CS101-1, Curve 1. This allows for very easy testing, even in manual mode, with a cursor or threshold set at the 70 dbµv target. i Automation of Audio Frequency Conducted Susceptibility Testing. Figure 5a shows 6.3 Vrms ripple at 1 khz using a sweep from 300 Hz to 1.3 khz. The power frequency and second and third harmonics are visible in addition to the injected ripple. If the span is modified to be centered on the ripple frequency, and just wide enough to display the injected waveform, as in Figure 5b, then the same test automation routine that is used with MIL-STD-461 requirement CS114, or RTCA/DO-160 section 20 conducted rf susceptibility, may be applied when automating audio frequency ripple injection. The routine commands both signal source and spectrum analyzer/emi receiver to the desired frequency, with a specified and fixed span for the analyzer/receiver. With only the injected frequency visible, the analyzer/receiver marker-to-peak function is invoked, and then the leveling feedback loop adjusts until the marker is at the desired level. This is how existing automation for rf conducted susceptibility testing is made available for audio frequency conducted susceptibility testing via use of the PRD-120. PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 5

Figure 5a: 1 khz ripple in span to 1300 Hz Figure 5b: 1 khz ripple at center of 100 Hz span The only place this doesn t work well is very close to the power frequency. Prior to 1993, MIL-STD-462 CS01 had an exclusion zone within 10% of the power frequency. Using the MIL-STD-461E/F 6 db bandwidth of 10 Hz, the closest approach to the power frequency at a level 25 db below the bus potential is 15 Hz away. For 400 Hz or wild-frequency power, that is well within the 10% window cited above. ii Also, any span that includes the power frequency won t be able to use the marker-to-peak function to level on the test signal, because marker-to-peak will seek out the power frequency. If it is possible to have the injected signal at the span center frequency, a marker-to-center-frequency command could be used, and if that isn t available at very low frequencies, a zero-span command could tune the receiver to the test signal frequency, and excludes all else. This however relies on extremely accurate and stable tuning so that the peak of the test signal is captured. PRD-120 works on both ac and dc buses. It should be noted that while the PRD-120 s function allowing frequency domain viewing of injected ripple isn t necessary for dc-powered test PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 6

samples, the device works just as well on a dc as an ac bus, and by providing isolation removes the need for an isolation transformer on the monitoring device. While the PRD-120 isn t necessary to monitor audio ripple injected on a dc bus, the fact that it can be used simplifies testing in that the set-up, software, and bookkeeping is identical between ac and dc-powered test samples. Fuse. PRD-120 is protected by a 63 ma (16Ω) fuse, Bussman GMA-63-R. Correct calibration and frequency response requires the use of this specific fuse. A spare fuse is shipped with the PRD-120. Temperature. In use, it is normal for the temperature of the PRD-120 to rise to about 125 F (52 C). Do not block the ventilation holes on the left side and bottom of the case. Warranty and return policy. The PRD-120 is warranted to be free from defects in materials and workmanship for a period of one year after the date of purchase. Pearson Electronics will repair or replace a defective product. The product may be returned for full credit, or exchanged for up to 30 days, provided it has not been damaged by the customer. A restocking fee will apply. i This technique works well except near the 5 khz breakpoint frequency, because the CS101 limit is a Bode plot approximation to the actual response of a parallel RC network. A simple work-around is to use the flat frequency response switch position out to 10 khz, and maintain 70 dbµv, and then at 10 khz switch over to the CS101 position, and continue to maintain 70 dbµv. This results in accurate testing up to 5 khz using the flat setting, and accurate testing from 10 khz up using the CS101 setting. From 5 to 10 khz this technique results in testing a little hot - from 0 db at 5 khz to 6 db at 10 khz (before flipping the switch). If no susceptibility is encountered between 5 and 10 khz, all is well. Running 3 6 db hot is a time-honored technique in EMI testing, to sensitize the measurement a bit and make sure no subtle susceptibilities near the limit are missed. If a susceptibility is PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 7

encountered between 5 and 10 khz, then the actual limit at the failed frequency must be determined and that actual level injected, instead of relying on the 70 dbµv reading. For example, a susceptibility is encountered at 8 khz. The limit at 8 khz is 4 db down from the 136 dbµv flat plateau at and below 5 khz. That means our target reading on the spectrum analyzer/emi receiver is 136 dbµv 4 db - 66 db = 66 dbµv, instead of our hot setting of 70 dbµv. Thus if the TOS (threshold of susceptibility) is between 66 and 70 dbµv as measured using the PRD-120, the test sample actually meets the requirement. ii Using a 3 Hz 6 db BW allows injection within 5 Hz of the power frequency, which meets the 10% window for 50/60 Hz power. A 3 Hz BW is available on many modern EMI receivers, but is not a requirement in MIL- STD-461. PEARSON ELECTRONICS 4009 Transport Street, Palo Alto, CA 94303 Telephone 650-494-6444 FAX 650-494-6716 www.pearsonelectronics.com PRD-120 Operating Manual 01/02/13 2013 Pearson Electronics 8