Olympus LEXT OLS 4000 Confocal Laser Microscope

Similar documents
KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

Bringing Answers to the Surface

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Systematic Workflow via Intuitive GUI. Easy operation accomplishes your goals faster than ever.

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Characterization Microscope Nikon LV150

Fire CR Calibration Guide

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Operating Instructions for Zeiss LSM 510

Zeiss 780 Training Notes

Zeiss 880 Training Notes Zen 2.3

Guide to Confocal 5. Starting session

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Zeiss LSM 880 Protocol

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Title: Leica SP5 Confocal User Manual

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

TRAINING MANUAL. Olympus FV1000

User manual for Olympus SD-OSR spinning disk confocal microscope

Zeiss AxioImager.Z2 Brightfield Protocol

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Nikon E800 Operating Instructions.

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Leica SP8 TCS Users Manual

Bruker Optical Profilometer SOP Revision 2 01/04/16 Page 1 of 13. Bruker Optical Profilometer SOP

R I T. Title: Wyko RST Plus. Semiconductor & Microsystems Fabrication Laboratory Revision: A Rev Date: 05/23/06 1 SCOPE 2 REFERENCE DOCUMENTS

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Widefield 1. Switching on

Leica TCS SP8 Quick Start Guide

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Nasmyth Ultraview Vox User Protocol

RENISHAW INVIA RAMAN SPECTROMETER

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

Leica DMi8A Quick Guide

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

OPT3: Operating Procedure for Horiba Jobin Yvon LabRam Aramis Raman/PL System See LabSpec_6_2 General User Quick Start Guide on the computer desktop

b. Turn the power switch and key to on position for blue laser.

Zeiss LSM 780 Protocol

Nikon SIM-E & A1-R System

Things to check before start-up.

Figure 1 The Raith 150 TWO

Renishaw InVia Raman microscope

Very short introduction to light microscopy and digital imaging

Using the Nikon TE2000 Inverted Microscope

Model SU3500 Scanning Electron Microscope

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Leica TCS SP8 Quick Start Guide

Leica SP8 TCS Users Manual

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Resolution. Diffraction from apertures limits resolution. Rayleigh criterion θ Rayleigh = 1.22 λ/d 1 peak at 2 nd minimum. θ f D

Training Guide for Carl Zeiss AxioZoom V16 Stereo Microscope

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Advanced Technology and Manufacturing Institute. Zygo ZeScope

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

LSM 510 Training Notes

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

JEOL 6500 User Manual

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

FEI Quanta 200 ESEM Basic instructions

LSM 710 Confocal Microscope Standard Operation Protocol

SOP: Polyvar Met Light Microscope

EXC500p-- PATHOLOGY MICROSCOPE. EXC500hd -- HD DIGITAL PATHOLOGY MICROSCOPE. EXC500r -- RESEARCH MICROSCOPE EXC500-LABORATORY SCOPE

Brief Operation Manual for Imaging on BX61W1

Microscopy from Carl Zeiss

Using the Hitachi 3400-N VP-SEM

LSM 510 Meta Training Notes

Zeiss Axio Imager.A1 manual

Heidelberg µpg 101 Laser Writer

Imaging Introduction. September 24, 2010

Indian Institute of technology Madras Presents NPTEL NATIONAL PROGRAMME ON TECHNOLOGY ENHANCED LEARNING

Cell Biology and Bioimaging Core

LSM 800 Confocal Microscope Standard Operation Protocol

Instructions for the Experiment

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

3D Measuring Laser Microscope OLS4100. Bringing Answers to the Surface. Precise measurement. Faster operation. High-quality imaging.

Olympus Digital Microscope Camera (DP70) checklist

Swept-Field User Guide

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

EE119 Introduction to Optical Engineering Spring 2003 Final Exam. Name:

Basic Operator Procedure Training NewView 6300 GROWING IN A SHRINKING WORLD

Microscopy. Matti Hotokka Department of Physical Chemistry Åbo Akademi University

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Nikon E800 Microscope. Operating Instructions

3D Scanning Guide. 0. Login. I. Startup

3D Measuring Laser Microscope OLS4100. Bringing Answers to the Surface. More precise measurement. Faster operation. High-quality imaging.

Quick Operation Guide

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

Transcription:

Olympus LEXT OLS 4000 Confocal Laser Microscope The Olympus LEXT OLS4000 is a confocal microscope capable of taking high-resolution 3D images. The magnification (Optical and Digital) of this microscope ranges from 108x 17,280x. It is capable of resolving features 10 nm in size in the z direction (sample height) and 120 nm in the x-y plane. The system is capable of performing a variety of metrology measurements. Step height, surface/line roughness, and area/volume measurements are some of the more commonly used metrology measurements. Figure 1 at the end of the text contains an image with key hardware identified on it. 1. Purpose This manual describes best known methods for capturing 3D images and describes some of the more commonly used measurement capabilities of the system. Users must be trained on the operation of the microscope in order to use it. Only qualified users who have been trained may use the system. There are several pinch points that users must be aware of prior to use. 2. Reference Material a. http://www.olympus-ims.com/en/metrology/ols4000/ b. LEXT User Manual Wizard - contact lab staff for full electronic copy 3. Safety Instructions a. EMO button can be depressed at any time to stop power to the system b. Do Not insert or remove a sample while the laser is on MUST BE IN [COLOR] MODE c. Liquids/wet samples are not permitted on the system at anytime d. Keep hands from of the system there are several pinch points that can occur due to moving objective lenses and stages. e. The system is equipped with a Class III laser i. Users should be familiar with OSHA mandates for this class of laser products: https://www.osha.gov/dts/osta/otm/otm_iii/otm_iii_6.html ii. Maintenance should only be performed by an authorized service representative. 4. System Specifications: Objective Lens Magnification: 5x to 100x Laser 3D image Magnifications: 20X to 100X Digital Magnification: 1 to 8x Total Magnification capability: 5x to 17,280x Color Imaging Mode: White LED Light, 2-megapixel CCD Detector Laser Imaging Mode: 405 nm Laser, Photomultiplier Detector Minimum Z-Resolution: 10 nm Minimum XY-Resolution: 120 nm

5. Logging in a. Login to the computer with User: LEXT Password: Olympus b. Startup: Double click the OLS4000 icon on the Windows desktop c. Click on Allow, I trust this program d. Login with User ID: ADMIN Password: olympus e. Clear off the stage System will ask the user to check the safety around a sample, object lens, and the stage. f. The system will then ask if the user would like to move the reference Z position Select [OK] provided nothing is on the stage. 6. Objectives Information Lens Magnification (optical 8x digital) Field of view (um) Working Distance (mm) Numerical Aperture 5x 108x - 864x 2560-320 20.0 0.15 10x 216x - 1728x 1280-160 11.0 0.30 20x 432x - 3456x 640-80 1.0 0.60 50x 1080x 8640x 640-80 10.6 0.50 50x 1080x - 8640x 256-32 0.35 0.95 100x 2160x - 17280x 128-16 0.35 0.95 7. Loading and Unloading of Samples a. LOAD AND UNLOAD at 5x! i. Loading at higher magnifications can cause damage the objective lenses ii. Loading at higher magnifications can damage your samples. iii. To load press the move to load/unload button iv. Then Press the move to origin v. To Unload Press the load/unload remove sample return stage to home position These may be located on the [MAP SCREEN] tab

b. Do not place your fingers by the objective lenses Potential pinch point. c. Never move the stage or the objective lenses by hand. The system has software features that will move the stage/objective lenses. d. Adjust Coarse Focus (discussed in next section) if sample is large to ensure it does not make contact with the objective lenses. e. At any point in time, an image reference for an area of interest can be set using the [update snapshot] icon to the right of the Map Screen tab f. The green box will show the last known area of interest defined by the user 8. Imaging: a. Select Imaging then Microscope then Color or Laser then then either COLOR or, LASER b. Coarse Focus Generally not needed i. Upon placing the sample on the stage adjust the coarse focus ii. Never force the coarse focus adjustment make sure the coarse focus is unlocked the locking collar is on the right side at the back of the unit. See Picture.

iii.!!!!!!!!!only ADJUST AT THE COARSE FOCUS AT 5X!!!!!!! Damage to the objective lenses and/or your sample may occur if adjusted at higher magnification. c. Auto Focus Select the FOCUS Tab and then select the AF button CLICK [FOCUS] Tab then Auto Focus [AF] Button d. Auto Brightness Select the BRIGHTNESS Tab and then Select AUTO button

e. To manually adjust the brightness up/down user must unlock the Auto Brightness first f. To enable auto brightness - Click Tab then [AUTO] i. Color No Automatic Brightness adjustment ii. Laser Brightness Automatically Adjusts g. Manual Focus can be achieved by moving the objective stage up/down to achieve the correct focal distance.

h. Magnification Change i. Simply select the magnification on the screen. ii. KEEP Hands and Fingers away from objectives while moving Pinch Point iii. Auto Focus should engage and display the live image. i. Image Capturing: 9. Acquiring a 3D Image a. Always uses the LASER to capture an image (405 nm) i. The instrument is only rated to take measurements on the 20x, 50x (high NA), and 100x objectives. ii. An image taken at 5x and 10x may not give you accurate readings b. Move to the desired location on the sample Can be done three ways: i. Moving the sample stage with the Joystick ii. Double Clicking on the small sample captured image iii. Double Clicking on the live image screen c. Under Imaging Microscope Color/Laser select the desired acquisition settings for the sample.

i. Dropdown contains: FINE, FAST, or STEP 1. Fine uses a finer pitch to capture images 10 um 2. Fast uses a coarser pitch less scans and faster 20 um 3. Multi-layer See Manual for details 4. STEP allows users to set the pitch 5. Snap Shot No Pitch change just captures image on screen ii. Color will capture and display a 3D color image iii. DIC stands for Differential Interference Contrast and maybe desired for image enhancements with the use polarizers and a color prism iv. Manual will require the users to set the Top and Bottom Focal Planes v. Stitching Maybe used to quilt an array of images together 1. 25 x 25 max array 2. Figure 2 shows stitched example 3. Contact Lab staff for additional training if needed d. Select Step to setup manually i. Set the bottom 1. Scroll the mouse down past the lowest focal plane on the sample

a. If Red appears turn down the brightness using AB b. Red indicates oversaturated signal on the Photomultiplier Tube 2. Set the TOP button ii. Set the top 1. Scroll the mouse up past the top focal plane on the sample a. If Red appears turn down the brightness using AB b. Red indicates oversaturated signal on the Photomultiplier Tube 2. Set the BOTTOM button iii. Set the pitch (10 nm or 0.010 um min) 1. Number of Steps (Focal Plane Cuts) will be displayed based on the distance between the top and bottom e. Press the Acquisition button to obtain image f. Image will now appear under the [MEASUREMENTS] tab g. Select [PROFILE] h. Select one of the four views i. Color only appears if [COLOR] was selected during acquisition.

10. Image Enhancement a. Several Image enhancements are possible on the Confocal System. The system has three one-shot filter adjustments. i. Level Corrections ii. Noise Removal iii. Brightness and Contrast Correction b. Select [LEVEL ADJUST] surface correction to level sample. See before/after from LEXT Manual in Figure 3. c. Select the [Noise Removal] button and select the desired profile correction

d. Brightness and Contrast Image Enhancement e. The image can be returned to its original state at anytime by selecting the [UNDO] button f. Manual correction of the images can be obtained by using the [IMAGE CORRECTION] button 11. DIC Image Capture a. From Manual: Creates a shadowing effect on color and laser images, creating 3D topologically-inaccurate images with features that pop-out. (See Figure 4) Works by splitting polarized light into two rays that provide information about adjacent points on a sample. Rays are recombined to determine differences in index of refraction and optical path length b. NOTE DIC WILL ALTER THE RESULTS OF ANY METROLOGY MEASUREMENTS IF LEFT IN THE SYSTEM IT IS ONLY TO BE USED TO OBTAIN CLEAR-OPTICAL IMAGES c. Push Polarizer Lens into the system. d. Insert the DIC Lens into the system e. Adjust the color wheel until the clearest image is obtained. f. When acquiring an image Enable [DIC] option so system knows DIC prism is installed.

g. When Completed Remove DIC Filter and Return Polarizer to the retracted position 12. Image Metrology a. Appendix A contains information pertaining to the collection and analysis of the systems measurement capabilities. i. Surface/Line Roughness ii. Profile Heights, Angle, Distance iii. Particle Analysis iv. Film Thickness b. Refer to the user manual for details on data acquisition 13. Auto Save Feature a. Auto Save can be enabled to b. To Enable/Disable Auto-Save select the [ADVANCED SETTINGS] box c. Select the AUTO SAVE box d. Two settings maybe selected e. Find the location of the folder that the images should be saved to: DESKTOP/USER/

Polarizer DIC Prism Objective Lenses Lock Coarse Focus Adjust Sample Stage EMO Joystick Figure 1. Features of the Confocal 3D Microscope

Figure 2.Stitching Example from OEM Website

Figure 3.Images before and after surface level correction

Figure 4.DIC Example from OEM Website

APPENDIX A Reference Material for Metrology Measurements