REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY

Similar documents
HIGH SPEED, HIGH RESOLUTION AND LOW COST DIGITAL RADIOGRAPHY

Amorphous Selenium Direct Radiography for Industrial Imaging

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

Center for Nondestructive Evaluation 304 Wilhelm Hall Iowa State University Ames, Iowa 50010

17th World Conference on Nondestructive Testing, Oct 2008, Shanghai, China

Hardware for High Energy Applications 30 October 2009

SECTION I - CHAPTER 2 DIGITAL IMAGING PROCESSING CONCEPTS

Performance of Image Intensifiers in Radiographic Systems

10/3/2012. Study Harder

Digital Radiography : Flat Panel

Digital Detector Array Image Quality for Various GOS Scintillators

10/26/2015. Study Harder

ULS24 Frequently Asked Questions

Preliminary Modulation Transfer Function Study on Amorphous Silicon Flat Panel System for Industrial Digital Radiography

DAMAGE DETECTION IN PLATE STRUCTURES USING SPARSE ULTRASONIC TRANSDUCER ARRAYS AND ACOUSTIC WAVEFIELD IMAGING

Radiographic sensitivity improved by optimized high resolution X -ray detector design.

ON THE WAY TO DIGITAL RADIOGRAPHY

LWIR NUC Using an Uncooled Microbolometer Camera

SYSTEMATIC NOISE CHARACTERIZATION OF A CCD CAMERA: APPLICATION TO A MULTISPECTRAL IMAGING SYSTEM

IQI-Sensitivity and Applications of Flat Panel Detectors and X-Ray Image Intensifiers A Comparison

DIGITAL IMAGE PROCESSING IN X-RAY IMAGING

Pixel Response Effects on CCD Camera Gain Calibration

Digital Radiology with Photon Counting Detectors

Synchronized electronic shutter system (SESS) for thermal nondestructive evaluation Joseph N. Zalameda

ROBOT VISION. Dr.M.Madhavi, MED, MVSREC

Chromatic X-Ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC

Film Replacement in Radiographic Weld Inspection The New ISO Standard

A TRUE WIENER FILTER IMPLEMENTATION FOR IMPROVING SIGNAL TO NOISE AND. K.W. Mitchell and R.S. Gilmore

WHITE PAPER. Methods for Measuring Flat Panel Display Defects and Mura as Correlated to Human Visual Perception

I. PERFORMANCE OF X-RAY PRODUCTION COMPONENTS FLUOROSCOPIC ACCEPTANCE TESTING: TEST PROCEDURES & PERFORMANCE CRITERIA

DECODING SCANNING TECHNOLOGIES

Photon Counting and Energy Discriminating X-Ray Detectors - Benefits and Applications

TESTING FLAT-PANEL IMAGING SYSTEMS: What the Medical Physicist Needs to Know. JAMES A. TOMLINSON, M.S., D.A.B.R. Diagnostic Radiological Physicist

MICROWAVE THICKNESS MEASUREMENTS OF MAGNETIC COATINGS. D.D. Palmer and V.R. Ditton

X-ray light valve (XLV): a novel detectors technology for digital mammography*

CCD Characteristics Lab

ULTRASONIC SIGNAL CHARACTERIZATIONS OF FLAT-BOTTOM HOLES IN

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

Tomographic 3D-Radiometry for the Visualisation and Measurement of the Defects of Girth Seams

Acquisition, Processing and Display

DIGITALLY ASSISTED ANALOG: REDUCING DESIGN CONSTRAINTS USING NONLINEAR DIGITAL SIGNAL PROCESSING

Equivalent Penetrameter Sensitivity (EPS) for Performance Evaluation of Computed Radiography Systems Muzibur Khan * and Mike Brothers

Essential Parameters and Conditions for Optimum Image Quality in Digital Radiology

Image acquisition. In both cases, the digital sensing element is one of the following: Line array Area array. Single sensor

INCREASING LINEAR DYNAMIC RANGE OF COMMERCIAL DIGITAL PHOTOCAMERA USED IN IMAGING SYSTEMS WITH OPTICAL CODING arxiv: v1 [cs.

Electron-Multiplying (EM) Gain 2006, 2007 QImaging. All rights reserved.

William B. Green, Danika Jensen, and Amy Culver California Institute of Technology Jet Propulsion Laboratory Pasadena, CA 91109

Everything you always wanted to know about flat-fielding but were afraid to ask*

The Noise about Noise

Welcome to: LMBR Imaging Workshop. Imaging Fundamentals Mike Meade, Photometrics

Very short introduction to light microscopy and digital imaging

Camera Test Protocol. Introduction TABLE OF CONTENTS. Camera Test Protocol Technical Note Technical Note

A Short History of Using Cameras for Weld Monitoring

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

Unit thickness. Unit area. σ = NΔX = ΔI / I 0

Receiver Design for Passive Millimeter Wave (PMMW) Imaging

PentaVac Vacuum Technology

Thomas G. Cleary Building and Fire Research Laboratory National Institute of Standards and Technology Gaithersburg, MD U.S.A.

The design and testing of a small scale solar flux measurement system for central receiver plant

IBEX TECHNOLOGY APPLIED TO DIGITAL RADIOGRAPHY

Digital Imaging Rochester Institute of Technology

Real Time Linear Array Imaging. Brian Caccamise

Dental Radiography. One of the problems of dental radiography is having different dimensions than normal.

StarBright XLT Optical Coatings

Essential Parameters for the Visibility of IQIs and Small Indications in Digital Radiography

Gas scintillation Glass GEM detector for high-resolution X-ray imaging and CT

Ray Detection Digital Image Quality and Influential Factors

MINIATURE X-RAY SOURCES AND THE EFFECTS OF SPOT SIZE ON SYSTEM PERFORMANCE

Examination of Pipe Welds by Image Plate Based Computed Radiography System

AUTOMATED EDDY CURRENT DETECTION OF FLAWS IN SHOT-PEENED

An Inherently Calibrated Exposure Control Method for Digital Cameras

OCT Spectrometer Design Understanding roll-off to achieve the clearest images

High Energy Digital Radiography & 3D-CT for Industrial Systems

A Test of non-standard Gain Settings for the NICMOS Detectors

1.Discuss the frequency domain techniques of image enhancement in detail.

DIGITAL RADIOGRAPHY. Digital radiography is a film-less technology used to record radiographic images.

LASER GENERATION AND DETECTION OF SURFACE ACOUSTIC WAVES

Optical Performance of Nikon F-Mount Lenses. Landon Carter May 11, Measurement and Instrumentation

High Resolution BSI Scientific CMOS

Astronomy 341 Fall 2012 Observational Astronomy Haverford College. CCD Terminology

Performance characterization of a novel thin position-sensitive avalanche photodiode-based detector for high resolution PET

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

The HOIS recommended practice for in-service computed radiography of pipes

Recommended Training Curriculum For Digital Radiography Personnel (Level II)

A novel tunable diode laser using volume holographic gratings

A new Photon Counting Detector: Intensified CMOS- APS

The End of Thresholds: Subwavelength Optical Linewidth Measurement Using the Flux-Area Technique

System and method for subtracting dark noise from an image using an estimated dark noise scale factor

Properties of a Detector

Photons and solid state detection

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

SIM University Projector Specifications. Stuart Nicholson System Architect. May 9, 2012

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Temperature Reductions to Mitigate the WF4 Anomaly

Objective Evaluation of Radiographic Contrast- Enhancement Masks

Determining MTF with a Slant Edge Target ABSTRACT AND INTRODUCTION

QUANTITATIVE COMPUTERIZED LAMINOGRAPHY. Suzanne Fox Buchele and Hunter Ellinger

Image Formation: Camera Model

Transcription:

REAL-TIME X-RAY IMAGE PROCESSING; TECHNIQUES FOR SENSITIVITY IMPROVEMENT USING LOW-COST EQUIPMENT R.M. Wallingford and J.N. Gray Center for Aviation Systems Reliability Iowa State University Ames,IA 50011 INTRODUCTION Traditionally, real-time x-ray inspection systems have used standard video camera technology in conjunction with an image intensifier or phosphor screen [1]. Less often, image processing and image analysis capabilities are provided in expensive, turnkey systems. Such systems typically allow for real-time contrast modification, frame shift, smoothing, and edge sharpening. Previously, we have shown that these types of techniques can be implemented using commercially available low-cost image processing hardware that uses arithmetic logic units, hardware multipliers, and a pipelined data flow [2]. Unfortunately, there are several drawbacks of implementing these techniques with low-cost hardware. These include the fundamental limitations of 8-bit DI A and AID converters in the digitization and display process, the low dynamic range of cheap CCD video cameras, and low spatial resolution. Standard, low-cost CCD cameras typically have SIN figures near 54 db, however, with on-chip integration and custom cooling techniques, this figure can be pushed higher (over 70 db). Hence, it is obvious that 8-bit digitizers, with a theoretical SIN limitation of approximately 50 db, cannot capture the entire video dynamic range without sacrificing sensitivity. This limitation can be overcome by segmenting the digitization process under control of the gain and offset parameters of the AID converters [3]. While this technique allows us to approach the full SIN as limited by the camera, the inherent photon counting noise in real-time x-ray systems typically dominates the noise in the system. The most widespread technique for reducing this noise component is temporal, or ensemble averaging. This has the dual advantage of reducing the readout noise and electronic noise in the video camera, thereby effectively improving the SIN of the camera and digitization portion of the imaging system. Review of Progress in Quantitative Nondestructive Evallllltion. Vol. 14 Edited by D.O. Thompson and D.E. Chimenti, Plenum Press, New York, 1995 871

A crucial step in successfully using averaging to reduce noise and improve sensitivity, is to retain the precision in the data after the averaging is completed. It is common to use integer arithmetic and divide by N after the image capture, but quantization noise is introduced that will limit the sensitivity. Ultimately, the sensitivity of the digitization system will be limited by the greater of the quantization noise and the camera noise. In this paper, we present analysis of these noise components from a 16-bit and an 8-bit point of view. We also demonstrate the use of two 16-bit calibration techniques that are designed to improve the measurement sensitivity using low-cost hardware. Finally, we present preliminary results of the use of low-cost imaging hardware with a new scintillating glass detector. NOISE ANALYSIS OF A YERAGING AND INTEGRA non It is of interest to determine the SIN components of the different noise processes in the camera and digitization portion of the imaging system to arrive at a figure for the best achievable SIN and dynamic range, independent of the noise on the input signal. We first consider the noise process in the camera itself. A typical CCD video camera might have a SIN figure of 54 db with a 1 volt peak-to-peak video signal. We define signal-to-noise as ~=1010g[V~m\ l' (1) N cr~ where cr ~ = noise variance on the video signal with a deterministic input. Thus, for the above figures of 54 db and 1 Yp_p, the noise variance is cr~ = 5.0 X 10-7 y2. In a digitization operation, the quantization noise is defined by (2) where s is the quantizer step size [4]. For an 8-bit quantizer (256 levels) and a 1 Y p_p input signal, the quantization noise variance iscr~ = 1.3 xlo- 6 y2 (49.9 db). Notice that for this camera and digitization specification, the quantization noise is the dominant noise process. In other words, if we improve the camera performance through cooling or on-chip integration, the quantization noise component for this digitizer will always limit us to less than 50 db SNR. Ensemble averaging of successive digitized frames is commonly used to reduce the photon counting noise associated with the x-ray detection process. When floating point precision or sufficient word size is used in this process, it also has the benefit of improving the noise performance of the digitizer. Assuming a stochastic input signal with variation significantly above the quantizing step size, we can expect that averaging or integrating 100 successive frames will reduce the quantization noise variance in Eq. 2 to Using this reduced quantization noise variance in Eq. 2, we find that the equivalent quantizer step size is 872

s' = ~12(crn' = 0.39 my. For the 1 V p-p input signal, this yields 2562 quantizer levels, or 12 bits. The corresponding improved quantization SIN is 70 db. If we were capable of performing an integration or average of the analog camera signal, its SIN would also improve by a factor of 100, or 20 db, giving an improved SIN = 54 + 20 = 74 db. So again, the dominant noise component in the image acquisition process, even after frame averaging is quantization noise, albeit the increase in resolution to 12 bits. This figure gives us the upper limit of the intensity resolution of the imaging equipment without consideration of the noise component on the signal incident on the camera or the dynamic range of the image intensifier. IMAGE CALIBRATION TECHNIQUES Several factors, independent of noise, can limit the sensitivity of a real-time x-ray inspection system. Among the most important are the spatial variation in the response of the image intensifier tube and the CCD video camera. In particular, the variation in radiometric sensitivity of the individual CCD pixels can easily mask a low contrast flaw indication. These responses can be largely corrected using a calibration image and subtracting the calibration image in real-time from the incoming measurements [2]. There are two problems with this technique. First, the calibration is performed using 8-bit images, and hence the quantization noise is a problem. Second, the response being corrected is signal dependent, and therefore, a single calibration image is not truly representative of the distortions in a complicated image with wide dynamic range. One of the difficulties of solving the problem with 8-bit calibration is that more expensive imaging hardware would be required, which would include at least three 16-bit accumulation buffers to perform the subtraction. We have adopted the approach of using a single 16-bit accumulation buffer to contain both the measurement signal as well as the corrected signal. We first integrate N frames ofthe measurement into a 16-bit buffer as shown in Fig. 1 with the arithmetic logic unit set to (A+B). Next, a calibration field is established, which is used to decrement the existing accumulation buffer with N frames by setting the ALU function to (A-B), as shown in Fig. 1. The resultant image can then be viewed using hardware divide and offset circuitry for the appropriate scaling to the image display. Note that for this type of calibration to work optimally, both calibration and measurement fields must be relatively flat. In the case where the measurement field varies significantly, division by the calibration field works better but is much more difficult to implement under these hardware constraints. An example of the improvement gained by 16-bit calibration over the 8-bit method using a 1 % thickness penetrameter is shown in Fig. 2. Notice that the 4T and 2T holes are readily detected in the 16-bit image while the 4T hole is barely detected in the 8-bit image. Figure 3 shows another penetrameter image at the 0.6% thickness sensitivity level. In this case, both the 4T and 2T holes are detected. Figure 4 shows an example of detecting casting porosity using this technique. These examples illustrate that this type of calibration, when 16-bit precision is used, can dramatically improve the sensitivity limitations of conventional real-time x-ray inspection systems. 873

Offset & Displ - 16-bit Buffer Divide f-+ ay A Arithmetic Logic Unit A+B v ideo 8 Bit B or Acquisition A-B r--- Figure 1. Block Diagram of measurement integration and calibration circuit As mentioned previously, one limitation of the single image calibration method is that it does not account for the signal-dependent nature of the degradation processes [3]. This can be seen by examining Fig. 5 where the gray-scale response or several image pixels is plotted as a function of x-ray generator current. Because the slopes are not identical, the systematic variation in pixel-to-pixel response is signal dependent. This is not a problem in cases where the image is somewhat uniform, but in cases where there is a wide variation in gray-scale output, the single image correction method will be degraded. A method of calibration which takes into account this signal dependence, is to characterize each pixel's response as a function of the signal level. Each pixel response is roughly linear, as seen in Fig. 4, and therefore, can be described by two parameter slope and intercept. Thus, the modeled response is g;,j (I;) = a;,j ~ + b;,j (3) where ~ is an independent controllable brightness parameter such as generator current. The calibration procedure requires the equalization of the response across the image with uniform illumination. We arbitrarily pick the equalized pixel response to be described by the mean parameters, a;,j and b;,j in Eq. 2. The equalized response is given by g'.. (J:) = p~ +q '.J ":J (4) where 1 M N p= MN~ ~a;,j and 1 M N q = --L L b;,j, M,N = image size, MN ;=1 ;=1 The calibration is performed by measuring the gray-scale, g;,j, solving for the brightness variable using ~ = g;,j - b;,j, a;,j (5) 874

(b) Figure 2 (a) Illustration of l6-bit calibration on a 1 % thickness penetrameter, (b) Illustration of 8-Bit calibration on a 1 % thickness penetrameter Figure 3. Calibrated penetrameter image at 0.6% thickness 875

Ca) Figure 4 (a) Real-time radiograph of casting porosity before calibration, (b) Radiograph after 16-bit calibration (b) 876

and finally, computing the corrected gray scale, g';,j in Eq. 4. Figure 6 shows the result of this technique using a 2% thickness penetrameter on an aluminum bar. Both the 4T and 2T penetrameter holes are readily detected in the processed image after the calibration and contrast enhancement. While this technique is more general than the subtraction method, it does not perform as well in cases of uniform backgrounds. It does, however, perform better in cases of complicated geometry and wide dynamic range and can detect thickness variations on the order of approximately 1 %. 250.00 200.00 -::l a. "S 0 Q) (ij (,.') C/) >-... ca C) 150.00 100.00 50.00 0.00 -+---=;:-----,-----,----,--,--,-----,----, 0.00 0.40 0.80 1.20 Generator Current (ma) 1.60 Figure 5. Calibration characteristic for several pixel locations (a) (b) Figure 6. (a) Real-time radiograph of2% penetrameter, (b) Radiograph after model-based calibration an

One of the drawbacks of this model-based calibration techniques is that it is more computationally and memory intensive due to the requirement of coefficient storage for each pixel and the floating point arithmetic. This can be moderated, however, through the use of piggyback DSP boards which are readily available at low cost. APPLICATIONS USING SCINTllLATING GLASS DETECTOR Among the main limitations of conventional real-time radiographic systems are the spatial resolution and dynamic range of the image intensifier. Image intensifier tubes typically have a spatial frequency limitation of about 2-5Ip/mm. We are presently investigating the use of a new compact scintillating glass material with better resolution and potentially better contrast sensitivity for use in real-time x-ray inspection [5]. One of the properties of the scintillating glass is that the light output at moderate dosages is extremely low. For this reason, most applications to date have involved the use of cooled, low-noise CCD cameras which can integrate long enough to achieve a usable image. These types of cameras are typically very expensive and can offset the savings provided by this low-cost detector. In keeping with the desire to have a low-cost imaging system, we are working on techniques which will allow the use of low-cost imaging cameras with this detector. In particular, the use of 16-bit integration with 8-bit AID converters, the availability of new low-cost integrating CCD video cameras, and the adjustment of AID converter gain have allowed us to obtain usable images. Figure 7 shows an acquired image of a resolution gauge from the scintillating glass using a low-cost Cohu CCD video camera. Figure 8 illustrates the spatial resolution obtainable from this detector by focusing on a smaller area of the detector. CONCLUSIONS While many of the techniques presented in this paper are currently available in expensive turnkey imaging systems, we have demonstrated that these techniques can be implemented using low-cost, off the shelf imaging hardware to dramatically improve the performance of real-time x-ray inspection systems. Among the techniques are AID converter integration which not only reduces the photon process counting noise, but significantly lowers the quantization noise typically associated with 8-bit averaging. This reduction in quantization noise has allowed the sensitivity performance to be improved to approximately 0.5% thickness variation without significant cost. A more generalized calibration procedure has also been presented which accounts for the signal dependent variation in imaging system response when inspecting complicated geometries. While this technique does not outperform the subtraction method in uniform geometries, it does work better for complicated geometries and is more general. Finally, preliminary investigations have been made into the use of low-cost imaging hardware with a new scintillating glass detector. These investigations indicate promise for obtaining usable real-time images with low-cost cameras and suitable image processing. Future study will refine the techniques and yield quantitative figures for the performance in contrast with more expensive scientific grade CCD imagers. 878

Figure 7. Resolution gauge image obtained from the scintillating glass detector using a low-cost CCD camera Figure 8. Illustration of spatial resolution capability of the detector using low-cost camera and zoom lens (10 lp/mm and 20 lp/mm markings shown). REFERENCES 1. Industrial Radiology - Theory and Practice, R. Halmshaw, Applied Science, 1982 2. "Application of real-time image processing and calibration techniques to real-time x-ray NDE", R. M. Wallingford and J. N. Gray, Review of Progress in Quantitative Nondestructive Evaluation, Vol. 13A, pp. 755-762, Plenum Press, 1994. 3. "Tomographic Inspection System Using X-rays", V. R. Kini, MS Thesis, Iowa State University, 1994. 4. Voice and Speech Processing, T. Parsons, McGraw Hill, 1987. 5. High-resolution digital radiography and three-dimensional computed tomography", C. Bueno and M. D. Barker, SPIE Proceedings on X-Ray Detector Physics and Applications II, pp. 179-191,1993. 879