OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE RS-01R

Similar documents
OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE TEC-B-01M VOLTAGE CLAMP MODULE FOR EPMS SYSTEMS. VERSION 1.2 npi 2014

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE. ISO-STIM 01D STIMULUS ISOLATION UNIT ±100 V / ±10 ma, bipolar output

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02 B EXTRACELLULAR AMPLIFIER

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE DPA-2FX DIFFERENTIAL AMPLIFIER / FILTER MODULE FOR EPMS SYSTEMS

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-03XS

USING THE AXOCLAMP-2B TUTORIALS

M302RM OPERATING MANUAL

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-16DX EXTRACELLULAR AMPLIFIER

IR-183A & IR-283A Intracellular Recording Amplifier Instruction Manual

Four-Channel Differential AC Amplifier

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION OF THE SEC-05X SINGLE-ELECTRODE CLAMP AMPLIFIER

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE EXT-02F EXTRACELLULAR AMPLIFIER WITH FILTERS

ASK Cards. Features Analog signal isolation Digital signal isolation. Applications

What are the steps to use the Axoclamp-2B for whole-cell patch clamp in continuous single-electrode voltage clamp (csevc) mode?

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE ELC-01X AMPLIFIER FOR EXTRACELLULAR RECORDING AND ELECTROPORATION

INSTRUCTION MANUAL. March 11, 2003, Revision 3

Glass Electrode Meter

Signal Conditioning Amplifier

EPAD OPERATIONAL AMPLIFIER

INSTRUCTION MANUAL FOR CELL SIMULATOR MODEL 2410

Type Ordering Code Package TDA Q67000-A5168 P-DIP-18-5

A Simplified Test Set for Op Amp Characterization

LANGER EMV-TECHNIK. Operating Instructions. A 100 / A 200 / A 300 Optical Fibre Probe

INTEGRATED CIRCUITS. AN109 Microprocessor-compatible DACs Dec

Single Electrode Voltage Clamping

RADIATED IMMUNITY TEST SYSTEM 6 GHz

SL300 Snow Depth Sensor USL300 SNOW DEPTH SENSOR. Revision User Manual

Voltage-to-Frequency and Frequency-to-Voltage Converter ADVFC32

Group: Names: (1) In this step you will examine the effects of AC coupling of an oscilloscope.

Analog-to-Digital-Converter User Manual

Variable Gain Low Frequency Voltage Amplifier

PLL Synchronizer User s Manual / Version 1.0.6

CHAPTER 2 D-Q AXES FLUX MEASUREMENT IN SYNCHRONOUS MACHINES

Radiated Immunity Test System 6 GHz

BIO 365L Neurobiology Laboratory. Training Exercise 1: Introduction to the Computer Software: DataPro

5500A. Multi-Product Calibrator. Extended Specifications 2005

AUTOMOTIVE CURRENT TRANSDUCER HAH1DR 300-S

POT/GAL 15 V 10 A and POT/GAL 30 V 2 A. Electrochemical Impedance Potentiostat Galvanostat Test Interface for Alpha-A Analyzer

UCE-DSO210 DIGITAL OSCILLOSCOPE USER MANUAL. FATIH GENÇ UCORE ELECTRONICS REV1

Technical Service Bulletin DCX AGC Setup Procedure for ADAPT IV Exciters

ELC-03XS. Versatile All-in-One Amplifier

Precision VOLTAGE REFERENCE

Experiment 8 Frequency Response

MGM 3000X Q67000-A5179 P-DSO-20-1 (SMD) MGM 3000X Q67006-A5179 P-DSO-20-1 Tape & Reel (SMD)

Charge Meter. Electronics & Software. Universally Applicable for Piezoelectric Measuring Technology. Type 5015A...

Manual itev 90 Intelligent Two-Electrode Voltage Clamp

PXIe Contents. Required Software CALIBRATION PROCEDURE

Low Noise Variable Gain Low Frequency Voltage Amplifier

TRT HV System Simulations

LM134/LM234/LM334 3-Terminal Adjustable Current Sources

User Guide. SIB Channel MAPMT Interface Board Hamamatsu H7546 series

ATA-2001 Analog LVDT/RVDT Signal Conditioner with digital calibration

Quad 12-Bit Digital-to-Analog Converter (Serial Interface)

SHF Communication Technologies AG

Technical Service Bulletin DCX AGC Setup Procedure for MODAP Exciters

Keywords: GPS, receiver, GPS receiver, MAX2769, 2769, 1575MHz, Integrated GPS Receiver, Global Positioning System

Radiated immunity test system 6 GHz

MEMS On-wafer Evaluation in Mass Production Testing At the Earliest Stage is the Key to Lowering Costs

1510A PRECISION SIGNAL SIMULATOR

USER MANUAL. Ultra-Low Noise High Voltage Amplifier WMA V to +150V output. 300µV rms output noise. 2mV output offset voltage

Operation and Service Manual. 350 MHz Preamplifier SIM914. Stanford Research Systems

Model 25D Manual. Introduction: Technical Overview:

APPENDIX D DISCUSSION OF ELECTRONIC INSTRUMENTS

TesT system for ConduCTed and radiated immunity

Neuroprobe Amplifier

Description of options, upgrades and accessories for the laser beam stabilization system Compact

TETRIS 1000 High Impedance Active Probe. Instruction Manual

DUAL CHANNEL BROADBAND LINEAR AMPLIFIER Model A800D

The latest version offers all the features already available in the IPoD series 8232 to 8238 and 8427 as well as providing several exclusive

VE-2 Voltage Clamp User's Guide. Copyright Alembic Instruments. All rights reserved.

PIEZO FILM LAB AMPLIFIER

Universally Applicable for Piezoelectric Measuring Technology. Technical Data. Charge Input. Instant value. Extreme values ors statistic measurements

SHF Communication Technologies AG

Universally Applicable for Piezoelectric Measuring Technology. Technical Data. Charge Input. Extreme values ors statistic measurements

Agilent 8902A Measuring Receiver

TF 02 / TF 02-Ex. Head mounted temperature transmitters, FOUNDATION Fieldbus (H1), Pt 100 (RTD), thermocouples, 1 or 2 independent channels


2520 Pulsed Laser Diode Test System

Signal Conditioning Amplifier

Isolated, Unregulated DC/DC CONVERTERS

IsoAmp 3000/4000. The Advantages The dynamic range extends into the negative, allowing for strictly linear transmission

Design Implementation Description for the Digital Frequency Oscillator

C 2030 חפר מיכשור ומערכות בע"מ טכנולוגיות לטיפול במים ושפכים מיכשור אנליטי למדידה ובקרת תהליכים OPERATOR'S MANUAL. 2- Wire E.C transmitter Din Rail

Amptek sets the New State-of-the-Art... Again! with Cooled FET

Wireless Neural Loggers

TETRIS User's Guide. High Impedance Active Probe DO177-1

DUAL%CHANNEL%LINEAR%AMPLIFIER %WITH%PHASE%INVERTER Model&A400DI

Ultra Low-Noise Variable Gain Low-Frequency Voltage Amplifier

High Power Piezo Driver

SC-4/M User manual 4-channel measuring amplifier

Low Power, Precision FET-INPUT OPERATIONAL AMPLIFIERS

BROADBAND LINEAR AMPLIFIER Model P150

AMP-13 OPERATOR S MANUAL

NSG Hard keys for important functions. 5.7 color display, easy to use firmware. 3 power meter

CEC VIBRATION TRANSMITTER

HS-xx-mux. User s Manual. Multiplexing Headstage that allows recording on 16 to 64 individual electrodes

When you have completed this exercise, you will be able to determine ac operating characteristics of a

MICROTC LINEAR THERMOCOUPLE AMPLIFIER OPERATOR S MANUAL

Model BSI-1A BIPHASIC STIMULUS ISOLATOR

FSK DEMODULATOR / TONE DECODER

Transcription:

OPERATING INSTRUCTIONS AND SYSTEM DESCRIPTION FOR THE RS-01R INSTRUMENT FOR MEASURING SERIES RESISTANCE IN TEVC RECORDINGS VERSION 1.2 npi 2014 npi electronic GmbH, Bauhofring 16, D-71732 Tamm, Germany Phone +49 (0)7141-9730230; Fax: +49 (0)7141-9730240 e-mail: support@npielectronic.com; http://www.npielectronic.com

Table of Contents 1. Safety Regulations... 3 2. Introduction... 4 2.1. Basic idea of RS measurement... 4 2.2. Solution... 4 2.3. Cell Model... 4 3. Description of the Front Panel... 5 4. Connections... 7 5. Operation... 7 6. Adjustment Procedure Using a Cell Model - Example... 8 7. Links for Additional Information... 12 8. Appendix... 12 version 1.2 page 2

1. Safety Regulations VERY IMPORTANT: Instruments and components supplied by npi electronic are NOT intended for clinical use or medical purposes (e.g. for diagnosis or treatment of humans), or for any other life-supporting system. npi electronic disclaims any warranties for such purpose. Equipment supplied by npi electronic must be operated only by selected, trained and adequately instructed personnel. For details please consult the GENERAL TERMS OF DELIVERY AND CONDITIONS OF BUSINESS of npi electronic, D-71732 Tamm, Germany. 1) GENERAL: This system is designed for use in scientific laboratories and must be operated by trained staff only. General safety regulations for operating electrical devices are to be followed. 2) STATIC ELECTRICITY: Electronic equipment is sensitive to static discharges. Some devices such as sensor inputs are equipped with very sensitive FET amplifiers, which can be damaged with electrostatic charge and must therefore be handled with care. Electrostatic discharge can be avoided by touching a grounded metal surface when changing or adjusting sensors. Always turn power off when adding or removing modules, connecting or disconnecting sensors, headstages or other components from the instrument or 19 cabinet. 3) TEMPERATURE DRIFT / WARM-UP TIME: All analog electronic systems are sensitive to temperature changes. Therefore, all electronic instruments containing analog circuits should be used only in a warmed-up condition (i.e. after internal temperature has reached steady-state values). In most cases a warm-up period of 20-30 minutes is sufficient. 4) HANDLING: Please protect the device from moisture, heat, radiation and corrosive chemicals. version 1.2 page 3

2. Introduction 2.1. Basic idea of RS measurement The instrument is based on the injection of symmetrical current pulses of 10 µa and a few khz around the holding or resting potential of the cell. The membrane potential deviation that appears on the positive slope of the pulse is proportional to RS and is measured using high precision sample-and-hold circuits controlled by a timing unit that is synchronized from the injection pulses applied. RS is displayed on a digital meter with a resolution of 10 and can be also stored on a computer. RS measurement can be started manually or through a TTL input. After the measurement the stored RS value is displayed continuously until the next measurement is started. Therefore, measurement can be automated easily using standard lab software. 2.2. Solution The instrument is based on the injection of symmetrical current pulses of 10 microamperes and a few khz around the holding or resting potential of the cell. The membrane potential deviation that appears on the positive slope of the pulse is proportional to RS and is measured using high precision sample-and-hold circuits controlled by a timing unit that is synchronized from the injection pulses applied. The RS measurement can be started manually or through a TTL input. After the measurement, the stored RS value is displayed continuously until the next measurement is started. To avoid jumps in the membrane potential, a holding current can be set manually or the current supplied by the TEC system is applied automatically from a sample-and-hold circuit. The RS measurement system can be used for automated procedures using standard lab software. One problem is the shape of the injected current (stimulus signal). It should be as square as possible, but in practice there are always parasitic capacitances leading to distortion of the current. Thus, the stimulus signal is conditioned in order to get an optimal square shape. We use a PD (proportional-differential) controller. Time constant and damping of the controller are adjustable to a certain extend. Depending on the frequency a constant measurement window for two sample-and-hold amplifiers is created and the time course of PEL is saved. The first measuring point represents the baseline and the second the value of PEL just before the increase caused by the capacitance of the cell or model cell respectively. The difference of the values of these two measurements corresponds to the value of RS. 2.3. Cell Model The properties of the RS measurement are tested best using a cell model that mimics not only electrode resistance, membrane resistance and membrane capacitance but also the series resistance. A cell model with series resistance is available on request. Important: The series resistance measurement is reliable only on cells with a large membrane capacity (>50 nf). version 1.2 page 4

3. Description of the Front Panel Figure 1: RS-01 R front panel view (the numbers are related to those in the text below) version 1.2 page 5

In the following description of the front panel elements each element has a number that is related to that in Figure 1. The number is followed by the name (in uppercase letters) written on the front panel and the type of the element (in lowercase letters). Then, a short description of the element is given. Each control element has a label and frequently a calibration (e.g. PEL x10 mv). (1) POTENTIAL INPUT PEL x10 mv connector BNC connector for connecting the potential output (potential electrode) of the TEC amplifier. (2) CURRENT RANGE switch (optional) 4-position switch for selecting the CURRENT RANGE. If the TEC amplifier is equipped with a headstage with different current ranges, the CURRENT RANGE selected at the RS-01R must be the same as the CURRENT RANGE selected at the headstage. (3) DIFFERENTIATOR DAMP potentiometer Potentiometer for adjusting the damp factor of the PD controller. (4) FREQUENCY potentiometer Potentiometer for setting the frequency of the test pulses, range: ~2 khz 5 khz. (5) RS MEAS. switch Switch to select the RS measurement. CONT.: RS is measured continuously HOLD: The last measurement of RS is kept MAN.: RS is measured until the switch is released. (6) SERIES RESISTANCE (k ) display LED showing the measured RS in k. (7) BIAS trim pot Trim pot to cancel the BIAS current. (8) HOLD. CURR. potentiometer Potentiometer for setting the HOLDING CURRENT manually, if #8 is in + or - position. (9) MODE OF OPERATION switch 4-position switch for selecting the MODE OF OPERATION (see also #8). +: positive HOLDING CURRENT is applied -: negative HOLDING CURRENT is applied OFF: HOLDING CURRENT is disabled AUTO: the RS-01R applies a HOLDING CURRENT as set at the TEC amplifier. (10) CURRENT HEADSTAGE CONNECTOR TEC 15-pole connector for connecting the TEC amplifier (see also chapter 4). version 1.2 page 6

(11) CURRENT HEADSTAGE 15-pole connector for connecting the current headstage (see also chapter 4). (12) RS IN (TTL) BNC connector to connect a TTL signal for remote operation (HI: RS measurement on, LOW: RS measurement off. (13) RS OUT BNC connector providing a voltage proportional to the series resistance, scaling: 1V / k. (14) TIME BASE potentiometer Potentiometer for setting the TIME BASE of the PD controller (15) POTENTIAL OUTPUT CEL x10 mv connector BNC connector providing the potential at the current electrode. Caution: The voltage at CEL can be up to 150 V!! (16) POTENTIAL OUTPUT PEL x10 mv connector BNC connector providing the potential at the voltage electrode. 4. Connections The RS-01R comes in a 19 rack mount cabinet and can be connected to every npi TURBO TEC amplifier using the supplied cable. Modification of the amplifier is not required. The current headstage is connected to the RS-01R with the 15-pole connector cable to CURRENT HEADSTAGE connector (#11, Figure 1). The 15-pole connector of the RS box (#10, Figure 1)is connected to the CURRENT HEADSTAGE connector at the TEC amplifier. PEL x10 (#1, Figure 1) is connected to PEL x10 at the front panel of TEC and to an oscilloscope. CEL at the front panel of the RS box is connected to the other channel of the oscilloscope. A power supply is not required. Caution: The voltage at CEL can be up to 150 V!! 5. Operation RS measurement is done in CC mode of the amplifier. If currently operating in VC mode and the MODE OF OPERATION switch at the RS-01R is set to AUTO, the TEC amplifier is automatically set to CC mode and the same holding current needed for setting the holding potential in VC mode, is applied to the cell. Thus, there should be only slightly changes in membrane potential, when switching to RS measurement in AUTO position. As a start point use a test pulse frequency of ~3 khz (recommendation of Prof. Conti). Later, the frequency can be adjusted if necessary. Set the RS MEAS. mode switch (#5, Figure 1) to CONT. (continuous) or MAN. (manual). Then, make sure that the CURRENT RANGE switch is in the right position (see #2, Figure 1). You will see a value at the display of the RS box and the course of PEL and CEL at the oscilloscope. With the potentiometers DAMP. (#3, Figure 1)and TIME BASE (#14, Figure 1) adjust the shape of the CEL signal until it is as square as possible. Details of the procedure are version 1.2 page 7

shown in the following figures. After adjustment, the value at the display of the RS-01R corresponds to the series resistance in X.XX k. Notice: Because of the measuring method and the background noise you always will see values of at least 10 to 20, even if no series resistance is present (e.g. when using a cell model). The BNC connector RS (#13, Figure 1) OUT provides a DC voltage proportional to the value of series resistance (scaling: 1V / k ). As mentioned above, the measurement can be started using a TTL signal fed into the BNC connector RS IN (TTL) (#12, Figure 1) at the front panel of the RS box (HI starts the measurement). In principle, after the measurement (when the signal at TTL is low again or the switch is in the OFF position) the stored RS value is displayed continuously until the next measurement is started. 6. Adjustment Procedure Using a Cell Model - Example Important: You can test the RS-01R only with a cell model that mimics the series resistance!! Upper lane: signal of the potential electrode (200 mv / DIV) Lower lane: signal of the current electrode (2 V / DIV) Time base (for Figure 2): 5 ms / DIV 1)Ch1:2Volt5ms 2)Ch2:20mVolt5ms Figure 2: signals without RS box version 1.2 page 8

Time base for Figure 3 to Figure 9): 100 µs / DIV 1)Ch1:2Volt10us 2)Ch2:20mVolt10us Figure 3: Current electrode greatly under compensated (too low setting of TIME CONSTANT) 1)Ch1:2Volt10us 2)Ch2:20mVolt10us Figure 4: Current electrode slightly under compensated (still too low setting of TIME CONSTANT) version 1.2 page 9

1)Ch1:2Volt10us 2)Ch2:20mVolt10us Figure 5: Current electrode over compensated (too high setting of TIME CONSTANT) 1 ) 2 ) Figure 6: Current electrode correctly compensated (perfect setting of TIME CONSTANT) version 1.2 page 10

1)Ch1:2Volt10us 2)Ch2:20mVolt10us Figure 7: Current electrode correctly compensated, but the delay is too high, i.e. the setting of DAMP is too low 1)Ch1:2Volt25us 2)Ch2:20mVolt25us Figure 8: Current electrode correctly compensated, but the delay is too low, i.e. the setting of DAMP is too high version 1.2 page 11

1)Ch1:2Volt25us 2)Ch2:20mVolt25us Figure 9: Current electrode correctly compensated, the delay is good, i.e. the setting of DAMP is perfect 7. Links for Additional Information http://www.npielectronic.de/downloads/poster.html 8. Appendix Series resistance measurement Poster presented at the 29th GÖTTINGEN NEUROBIOLOGY CONFERENCE 2003 (seehttp://www.npielectronic.de/fileadmin/files/rs_poster.zip) version 1.2 page 12