This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

Similar documents
1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

FEI Tecnai G 2 F20 Operating Procedures

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

FEI Falcon Direct Electron Detector. Best Practice Document

Transmission Electron Microscopy 9. The Instrument. Outline

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

1. Preliminary sample preparation

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

FEI Helios NanoLab 600 TEM specimen prep recipe Nicholas G. Rudawski (352) (office) (805) (cell) Last updated: 01/19/17

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

STEM alignment procedures

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

Basic Users Manual for Tecnai-F20 TEM

User Operation of JEOL 1200 EX II

2. Raise HT to 200kVby following the procedure explained in 1.6.

2 How to operate the microscope/obtain an image

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Tecnai on-line help manual --

Instructions for Tecnai a brief start up manual

1. Specimen Holder Removal, Loading, and Insertion

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Tecnai T12 Operating Procedures

XTEM. --Software for Complex Transmission Electron Microscopy. Version 1.0

Tecnai on-line help User interface 1 Tecnai F20 Tecnai F30 User interface Software version 2.1.8/3.0

Operating F20/F30 with SerialEM

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

05/20/14 1. Philips CM200T. Standby Condition

Talos on-line help User interface 1 Software version 1.6 and higher

NANO 703-Notes. Chapter 9-The Instrument

Introduction: Why electrons?

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

CM20 USER GUIDE. Duncan Alexander, CIME 2010

Indiana University JEM-3200FS

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center

Titan on-line help manual -- User Interface

STEM Spectrum Imaging Tutorial

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

LowDose for JEOL. --Software LowDose and Advanced Data Acquisition. Version 1.1

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

FEI Titan Image Corrected STEM

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

Leica DMi8A Quick Guide

Confocal Application Notes Vol. 5 July 2010

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

Introduction of New Products

This procedure assumes the user is already familiar with basic operation of the SEM and the MiraTC interface.

Zeiss LSM 780 Protocol

SerialEM Help Index. ABC Amber HLP Converte r T rial version,

Zeiss AxioImager.Z2 Brightfield Protocol

Zeiss LSM 880 Protocol

SCIENTIFIC INSTRUMENT NEWS. Introduction. Design of the FlexSEM 1000

Characterization Microscope Nikon LV150

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

JEOL 6500 User Manual

Section 1: TEM parts and functions... 2

Nikon E800 Operating Instructions.

Model SU3500 Scanning Electron Microscope

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Transmission electron Microscopy

TEM theory Basic optics, image formation and key elements

SEM Training Notebook

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

INSTRUCTIONS FOR COURSE WORK 4 (AxioVert) Instructor: Anne Vaahtokari (MIU) 1. Purpose of the work

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Ultrascan 1000 CCD Camera with CM12

Development of JEM-2800 High Throughput Electron Microscope

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

WITec Alpha 300R Quick Operation Summary October 2018

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Scanning electron microscope

Reference and User Manual May, 2015 revision - 3

Using Autofocus in NIS-Elements

STRUCTURE OF THE MICROSCOPE

Operating Procedures for MICROCT1 Nikon XTH 225 ST

Introduction to Transmission Electron Microscopy (Physical Sciences)

How to align your laser for two-photon imaging

SEM Training Notebook

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

Using Olympus dotslide for polarising microscopy

Horiba LabRAM ARAMIS Raman Spectrometer Revision /28/2016 Page 1 of 11. Horiba Jobin-Yvon LabRAM Aramis - Raman Spectrometer

Leica DB LB Research microscope and Studo Lite Imaging software

Transcription:

FEI Tecnai F20 S/TEM: acquiring diffraction patterns Nicholas G. Rudawski ngr@ufl.edu (805) 252-4916 (352) 392-3077 Last updated: 10/18/17 This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera. 1. Find a region of interest and bring it to eucentric height. 2. Instrument settings 2.1. Select and apply the GL3 FEG register. 2.2. In Microscope Control, select the tab; navigate to the Beam Settings control panel and select spot size = 7. 2.3. Perform the basic alignment of the instrument: center the C2 aperture, correct the condenser astigmatism, balance the deflector coils, and perform rotation centering. 2.4. Set magnification to SA 6200 ; you must be in SA mode if you need to use an SA aperture. 2.5. Expand the beam clockwise from crossover and center it on viewing screen 1

3. Setting up the DP 3.1. Select Diffraction (right-hand control pad) to enter diffraction mode. 3.2. Use the Magnification knob (right-hand control pad) to adjust the camera length as desired for observing the DP. 3.3. Insert an objective aperture into the DP (which one doesn t matter, as long as you can see it); use the Focus knob (right-hand control pad) to focus the objective aperture so the edge is sharp; when finished, retract the objective aperture 3.4. Use the Intensity knob (left-hand control pad) to focus the direct spot (make as small as possible) in the DP. 3.5. Select Diffraction to return to TEM mode; do not adjust the Intensity knob after you return to TEM mode. 4. Centering the beam stop 4.1. Start live imaging with the digital camera (insert the digital camera, flip the viewing screen up, etc.) with an integration time = 0.01 s; move the region of the specimen you wish to take the DP from to the center of the image; then draw 2 diagonal lines in the live image to mark the center of the image. 2

4.2. Insert the beam stop into the image and then center it (this will ensure your acquired diffraction patterns are always centered when collected); if you wish, you can acquire an image to have for reference purposes (the lines will not appear in the acquired image). 5. Inserting the SA aperture (if needed) 5.1. If your specimen is coarsely grained or single-crystal, then you must use an SA aperture when acquiring your DPs to prevent damage to the camera. However, if your specimen consists of very small particles or is very finely grained, an SA aperture may not be necessary. If you are unsure as to whether you should use an SA aperture or not when acquiring DPs from your specific samples, please consult with RSC staff before attempting to collect any DPs. 5.2. If an SA aperture is needed, insert and center the #2 or #1 SA aperture over the area of interest (the beam stop, region of interest, and SA aperture should all now be approximately centered); acquire another image (again, for reference purposes, so you know where in the image the DP originated). 5.3. Flip the viewing screen back down when finished, but continue to leave the camera inserted and acquiring a live image. 3

6. Centering the DP 6.1. Select Diffraction to return to diffraction mode; use the Focus knob to focus the direct spot (if necessary); do not adjust the Intensity knob. 6.2. Use the Multifunction knobs to move the direct spot so it is under the tip of the beam stop (the knobs will automatically be set to shift the DP when diffraction mode is entered) 7. Acquiring the DP 7.1. In DigitalMicrograph, select Analysis from the pull-down menu and then Histogram to view a histogram of the live image. 7.2. Flip the viewing screen up to start collecting a live image of the DP (still using a 0.01 s exposure time); note the right limit of the x-axis of the histogram, which indicates the maximum counts present in the DP. This number should not exceed 2000 3000 counts; do not proceed if the maximum counts in the live DP are higher than this range. 4

7.3. In the Camera Acquire tool, input a value for Exposure (s) of 0.1 s; select Start Acquire to acquire a test exposure of the DP (it should be very noisy). 7.4. In DigitalMicrograph, select Analysis from the pull-down menu and then Histogram to view a histogram of the test exposure; again, noting the right limit of the x-axis, which indicates the maximum counts present in the test exposure. 7.5. Based on the maximum counts in the test exposure, determine a suitable value for Exposure (s) such that the maximum counts in the acquired DP will be in range of 10000 20000 counts. For example, if the maximum counts in the test exposure is ~2000 counts, then a value for Exposure (s) of 1.0 s will be suitable; select Start Acquire to acquire the DP. 7.6. Flip the viewing screen back down after acquisition is finished. 5

7.7. If you wish to acquire the DP at a different camera length, make sure the viewing screen is flipped down, adjust the Magnification knob to change the camera length accordingly, then use the focus knob to focus the direct spot and then use the diffraction alignment to move the direct spot under the beam stop. 7.8. If DP acquisition from a different region is needed, select Diffraction to return to TEM mode, retract the SA aperture, center the new area of interest on the screen, insert the SA aperture again, and repeat the procedure again. 8. When finished recording DPs 8.1. Make sure the instrument is left in TEM mode, the SA aperture is retracted, the viewing screen is flipped down, and the digital camera is retracted. 9. Rotation calibration 9.1. If the indicated camera length is >135 mm (with the viewing screen flipped down), then the DP needs to be rotated clockwise 93 to establish the proper DP-image orientation (assuming the TEM image was collected in SA mode). Left: as collected; right: after clockwise rotation of 93 6

9.2. If the indicated camera length is 135 mm (with the viewing screen flipped down), then the DP needs to be rotated clockwise 152 to establish the proper DP-image orientation (assuming the TEM image was collected in SA mode). Left: as collected; right: after clockwise rotation of 152 7