PXIe Contents SPECIFICATIONS. ±60 V, 3 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4139.

Similar documents
PXIe Contents SPECIFICATIONS. ±60 V, 3 A System PXI Source Measure Unit. These specifications apply to the PXIe-4138.

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

PXIe Contents SPECIFICATIONS. ±200 V, 1 A System PXI Source Measure Unit. These specifications apply to the PXIe-4136.

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

PXIe Contents SPECIFICATIONS. ±200 V, 1 A Precision System PXI Source Measure Unit. These specifications apply to the PXIe-4137.

PXI Contents SPECIFICATIONS. ±20 V, 40 W PXI Source Measure Unit

PXIe Contents SPECIFICATIONS. Precision, 10 fa Sensitivity Low-Current PXI Source Measure Unit. These specifications apply to the PXIe-4135.

PXIe Contents SPECIFICATIONS. 2-Channel, 60 V, 1 A PXI Programmable Power Supply. These specifications apply to the PXIe-4112.

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

NI PXI Contents. About These Specifications DEVICE SPECIFICATIONS GHz Dual Transfer Switch 50 Ω Relay Module

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Device

1- or 2-Port, High-Speed, Flexible Data Rate PXI CAN Interface Module

PXIe Contents SPECIFICATIONS. RF Amplifier

USRP Contents. Definitions. Conditions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

USRP Contents. Definitions SPECIFICATIONS. Software Defined Radio Reconfigurable Device

CDA Contents. Definitions SPECIFICATIONS. Clock Distribution Device

PXIe Contents SPECIFICATIONS. RF Preselector Module

USRP Contents. Definitions SPECIFICATIONS. 50 MHz to 2.2 GHz Tunable RF Transceiver

1- or 2-Port, Software-Selectable PXI CAN Interface Module

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

USRP Contents. Definitions SPECIFICATIONS. 10 MHz to 6 GHz Tunable RF Transceiver

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

PXI Contents. PXI-2585 Specifications. Topology DEVICE SPECIFICATIONS. 10-Channel Multiplexer

Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

High-Speed/Flexible Data-Rate CAN Characteristics 5 VDC

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 40 A Fault Insertion Unit

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXI/PXIe Contents DEVICE SPECIFICATIONS. 10 A Fault Insertion Unit

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

PXIe, 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer,PXI Digital Multimeter

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

PXI Contents. PXI-2584 Specifications DEVICE SPECIFICATIONS. 300 V CAT II, 600 V CAT I Multiplexers

PCI Express, 6½-Digit, ±300 V Digital Multimeter Device

NI PXI/PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. Quad 2 1 Terminated Multiplexer

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

NI USB Contents DEVICE SPECIFICATIONS. 50 MS/s Bus-Powered USB Oscilloscope

6½-Digit, ±300 V, 1.8 MS/s Isolated Digitizer Included, PCI Digital Multimeter Device

Related Information Refer to the NI Switches Help for detailed topology information.

SA1500X. Operating Basics. Cautions DEVICE SPECIFICATIONS. High Impedance Active Probe

Note Using the NI 5751/5751B in a manner not described in this document may impair the protection that the NI 5751/5751B provides.

USRP Contents. Definitions SPECIFICATIONS. USRP Software Defined Radio Stand-Alone Device

NI PXI-2521 Specifications 40-Channel DPST Relay Module

16-Bit, 120 MS/s, 12 Channel DC/16 Channel AC Receiver Adapter Module

100-SPST (latching), 50-DPST. Refer to the NI Switches Help at ni.com/manuals for detailed topology information.

NI roborio. Contents. Processor SPECIFICATIONS. RIO Device for Robotics

NI PXI-2520 Specifications 80-Channel SPST Relay Module

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

Related Information Refer to the NI Switches Help for detailed topology information.

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

NI PXI-2797 Specifications 40 GHz 6 1 Terminated 50 Ω Multiplexer

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

NI PXI/PXIe-2529 Specifications

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

PXI Contents SPECIFICATIONS. 6½-Digit, ±300 V, Onboard 1.8 MS/s Isolated Digitizer, L and C Measurement Support, PXI Digital Multimeter

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

PXIe, 80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

8-Channel, 12-Bit, 60 MHz PXI Express Oscilloscope

NI PXI/PXIe-2540 Specifications

SCXI MHz Quad 8x1 50 Ω Multiplexer

PXIe-6570 DEVICE SPECIFICATIONS. 32-Channel Digital Pattern Instrument

Specifications are typical and valid from -40 C to +85 C unless otherwise noted. Galvanic isolation between channels and to chassis

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

PXI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PXI Oscilloscope

R Series for USB Multifunction RIO with Kintex-7 70T FPGA. may impair the protection the NI USB-7855R provides.

mmwave Transceiver System

NI 2865A 0.3 A Matrix Cards for NI SwitchBlock

NI PXI/PXIe-2527 Specifications

16 AI (16-Bit, 250 ks/s), 4 DI, 4 DO USB Multifunction I/O Device

100 MHz, 32-Channel, 5 V PXI Digital Waveform Instrument

NI PXI-2557 Specifications

Analog Characteristics

NI PXI-2555 Specifications

R Series for USB Multifunction RIO with Kintex-7 70T FPGA

R Series Reconfigurable I/O Module (AI, AO, DIO) for PXI Express, 6 AI, 18 AO, 48 DIO, 1 MS/s AIO, 512 MB DRAM, Kintex-7 325T FPGA

GETTING STARTED GUIDE NI Port DeviceNet Module

NI 9220 DATASHEET. 16 AI, ±10 V, 16 Bit, 100 ks/s/ch Simultaneous. DSUB or push-in springterminal. 250 Vrms, CAT II, channelto-earth

PXI Contents SPECIFICATIONS. 300 MHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

PXIe Contents SPECIFICATIONS. 500 MHz, 2.5 GS/s, 10-Bit PXI Express Oscilloscope

R Series Reconfigurable I/O Module (AI, AO, DIO) for PXI Express, 8 AI, 8 AO, 48 DIO, 500 ks/s AI, Kintex-7 160T FPGA

8-Ch 24 V Sinking DI, 8-Ch 24 V Sourcing DO SIL3 Capable

PXI Contents SPECIFICATIONS. 1 GHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

NI 9216 DATASHEET. 8 RTD, 0 Ω to 400 Ω, 24 Bit, 400 S/s Aggregate, PT100

NI 9881 GETTING STARTED GUIDE. 1-Port CANopen Module

NI 9207 DATASHEET. 16 AI, ±20 ma/±10 V, 24 Bit, 500 S/s Aggregate

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

PXIe Contents SPECIFICATIONS. 1.5 GHz, 5 GS/s, 10-Bit PXI Express Oscilloscope

NI PXIe-5630 Specifications

24-Bit, 6-Channel, 1.25 MS/s/ch PXI Analog Input Module

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

M Series Data Acquisition: 16-Bit, 250 ks/s, 8 AI, 2 AO, 24 DIO Sourcing, Current Input/Output Bank Isolation

Transcription:

SPECIFICATIONS PXIe-4139 ±60 V, 3 A Precision System PXI Source Measure Unit These specifications apply to the PXIe-4139. Contents Definitions...2 Conditions... 2 Cleaning Statement... 2 Device Capabilities... 2 Voltage Programming and Measurement Accuracy/Resolution... 4 Current Programming and Measurement Accuracy/Resolution... 5 Noise... 6 Sinking Power vs. Ambient Temperature Derating... 7 Output Resistance Programming Accuracy... 8 Extended Range Pulsing... 9 Transient Response and Settling Time...9 Load Regulation...10 Measurement and Update Timing Characteristics... 11 Remote Sense...12 Examples of Calculating Accuracy...12 Trigger Characteristics... 15 Protection... 16 Isolation...16 Guard Output Characteristics...16 Calibration Interval... 16 Power Requirement...17 Physical Characteristics... 17 Environment...17 Operating Environment...17 Storage Environment...17 Shock and Vibration...18 Compliance and Certifications...18 Safety... 18 Electromagnetic Compatibility... 18 CE Compliance... 19 Online Product Certification... 19 Environmental Management... 19

Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. Typical specifications describe the expected performance met by a majority of the models. Nominal specifications describe parameters and attributes that may be useful in operation. Specifications are Warranted unless otherwise noted. Conditions Specifications are valid under the following conditions unless otherwise noted. Ambient temperature 1 of 23 C ± 5 ºC Calibration interval of 1 year 30 minutes warm-up time Self-calibration performed within the last 24 hours nidcpower Aperture Time property or NIDCPOWER_ATTR_APERTURE_TIME attribute set to 2 power-line cycles (PLC) Fans set to the highest setting if the PXI Express chassis has multiple fan speed settings Cleaning Statement Caution Clean the hardware with a soft, nonmetallic brush. Make sure that the hardware is completely dry and free from contaminants before returning it to service. Device Capabilities The following table and figure illustrate the voltage and the current source and sink ranges of the PXIe-4139. 1 The ambient temperature of a PXI system is defined as the temperature at the chassis fan inlet (air intake). 2 ni.com PXIe-4139 Specifications

Table 1. Current Source and Sink Ranges DC voltage ranges 600 mv 6 V 60 V DC current source and sink ranges 1 μa 10 μa 100 μa 1 ma 10 ma 100 ma 1 A 3 A 10 A, pulse only Current (A) Figure 1. Quadrant Diagram 11 10 9 50 V, 10 A 20 V, 10 A 20 V, 10 A 50 V, 10 A 8 7 200 W 200 W 6 5 Quadrant II Quadrant I 4 3 50 V, 4 A 4 V, 3 A 6.6 V, 3 A 50 V, 4 A 2 12 W 20 W 1 60 V, 0.2 A 60 V, 0.33 A 0 1 60 V, 0.33 A 60 V, 0.2 A 2 20 W 12 W 3 4 50 V, 4 A 6.6 V, 3 A 4 V, 3 A 50 V, 4 A 5 6 7 8 9 10 11 200 W 50 V, 10 A Quadrant III 20 V, 10 A Quadrant IV 20 V, 10 A 200 W 50 V, 10 A 60 50 40 30 20 10 0 10 20 30 40 50 60 Pulse or DC Pulse only, 1ms, 5% duty cycle Pulse only, 400 µs, 2% duty cycle Voltage (V) DC sourcing power is limited to 20 W, regardless of output voltage. 2 2 Power limit defined by voltage measured between HI and LO terminals. PXIe-4139 Specifications National Instruments 3

Caution Limit DC Power sinking to 12 W. Additional derating applies to sinking power when operating at an ambient temperature of >45 C. If the PXI Express chassis has multiple fan speed settings, set the fans to the highest setting. Related Information Sinking Power vs. Ambient Temperature Derating on page 7 Voltage Programming and Measurement Accuracy/Resolution Table 2. Voltage Programming and Measurement Accuracy/Resolution Range Resolution (noise limited) Noise (0.1 Hz to 10 Hz, peak to peak), Typical Accuracy (23 C ± 5 C) ± Tempco ± (% of (% of voltage + offset) 3 voltage + T cal ± 5 C 4 T cal ± 1 C offset)/ C, 0 C to 55 C 600 mv 100 nv 2 μv 0.02% + 50 μv 6 V 1 μv 6 μv 0.02% + 300 μv 0.016% + 30 μv 0.016% + 90 μv 0.0005% + 1 μv 60 V 10 μv 60 μv 0.02% + 3 mv 0.016% + 900 μv Related Information Noise on page 6 Load Regulation on page 10 Remote Sense on page 12 3 Accuracy is specified for no load output configurations. Refer to Load Regulation and Remote Sense sections for additional accuracy derating and conditions. 4 T cal is the internal device temperature recorded by the PXIe-4139 at the completion of the last selfcalibration. 4 ni.com PXIe-4139 Specifications

Current Programming and Measurement Accuracy/Resolution Table 3. Current Programming and Measurement Accuracy/Resolution Range Resolution (noise limited) Noise (0.1 Hz to 10 Hz, peak to peak), Typical Accuracy (23 C ± 5 C) ± (% of current + offset) T cal ± 5 C 5 T cal ± 1 C Tempco ± (% of current + offset)/ C, 0 C to 55 C 1 μa 100 fa 4 pa 0.03% + 100 pa 10 μa 1 pa 30 pa 0.03% + 700 pa 0.022% + 40 pa 0.022% + 300 pa 0.0006% + 4 pa 0.0006% + 22 pa 100 μa 10 pa 200 pa 0.03% + 6 na 0.022% + 2 na 0.0006% + 200 pa 1 ma 100 pa 2 na 0.03% + 60 na 0.022% + 20 na 0.0006% + 2 na 10 ma 1 na 20 na 0.03% + 600 na 0.022% + 200 na 0.0006% + 20 na 100 ma 10 na 200 na 0.03% + 6 μa 0.022% + 2 μa 0.0006% + 200 na 1 A 100 na 2 μa 0.03% + 60 μa 0.027% + 20 μa 0.0006% + 2 μa 3 A 10 A, 1 μa 20 μa 0.083% + 900 μa pulsing only, typical 0.083% + 600 μa 0.002% + 20 μa 5 T cal is the internal device temperature recorded by the PXIe-4139 at the completion of the last selfcalibration. PXIe-4139 Specifications National Instruments 5

Noise Wideband source noise <20 mv peak-to-peak in 60 V range, device configured for normal transient response, 10 Hz to 20 MHz, typical The following figures illustrate noise as a function of measurement aperture for the PXIe-4139. Figure 2. Voltage Measurement Noise vs. Measurement Aperture, Nominal Measurement Noise, RMS (Volts) 100 m 10 m 1 m 100 μ 10 μ 1 μ 60 V Range 6 V Range 600 mv Range 100 n 1 μ 10 μ 100 μ 1 m 10 m 100 m 555.5 n 2 PLC Aperture (Seconds) Note When the aperture time is set to 2 power-line cycles (PLCs), measurement noise differs slightly depending on whether the nidcpower Power Line Frequency property or NIDCPOWER_ATTR_POWER_LINE_FREQUENCY attribute is set to 50 Hz or 60 Hz. 6 ni.com PXIe-4139 Specifications

Figure 3. Current Measurement Noise vs. Measurement Aperture, Nominal Measurement Noise, RMS (Amps) 10 m 1 m 100 μ 10 μ 1 μ 100 n 10 n 1 n 100 p 10 p 3 A Range 1 A Range 100 ma Range 10 ma Range 1 ma Range 100 μa Range 10 μa Range 1 μa Range 1 p 100 f 1 μ 10 μ 100 μ 1 m 10 m 100 m 555.5 n 2 PLC Aperture (Seconds) Note When the aperture time is set to 2 power-line cycles (PLCs), measurement noise differs slightly depending on whether the nidcpower Power Line Frequency property or NIDCPOWER_ATTR_POWER_LINE_FREQUENCY attribute is set to 50 Hz or 60 Hz. Sinking Power vs. Ambient Temperature Derating The following figure illustrates sinking power derating as a function of ambient temperature for the PXIe-4139. PXIe-4139 Specifications National Instruments 7

Figure 4. Sinking Power vs. Ambient Temperature Derating Sinking Power (W) 13 12 11 10 9 8 7 6 5 4 3 2 1 0 0 5 10 15 20 25 30 35 40 45 50 55 60 Ambient Temperature ( C) Output Resistance Programming Accuracy Table 4. Output Resistance Programming Accuracy Current Level/Limit Range Programmable Resistance Range, Voltage Mode Programmable Resistance Range, Current Mode Accuracy ± (% of resistance setting), T cal ± 5 C 6 1 μa 0 to ±5 MΩ ±5 MΩ to ±infinity 0.03% 10 μa 0 to ±500 kω ±500 kω to ±infinity 100 μa 0 to ±50 kω ±50 kω to ±infinity 1 ma 0 to ±5 kω ±5 kω to ±infinity 10 ma 0 to ±500 Ω ±500 Ω to ±infinity 100 ma 0 to ±50 Ω ±50 Ω to ±infinity 1 A 0 to ±5 Ω ±5 Ω to ±infinity 3 A 0 to ±500 mω ±500 mω to ±infinity 10 A, pulsing only 6 T cal is the internal device temperature recorded by the PXIe-4139 at the completion of the last selfcalibration. 8 ni.com PXIe-4139 Specifications

Extended Range Pulsing 7 Maximum pulse Voltage 50 V Current 10 A On time 8 1 ms Minimum pulse cycle time 5 ms Energy 200 mj Cycle average power 10 W Duty cycle 5% Transient Response and Settling Time Transient response Settling time 9 Voltage mode, 50 V step, unloaded 10 Voltage mode, 5 V step or smaller, unloaded 11 <70 μs to recover within 0.1% of voltage range after a load current change from 10% to 90% of range, device configured for fast transient response, typical <200 μs, typical <70 μs, typical Current mode, full-scale step, 10 A to <50 μs, typical 100 μa ranges 12 Current mode, full-scale step, 10 μa range 12 Current mode, full-scale step, 1 μa range 12 <150 μs, typical <300 μs, typical The following figures illustrate the effect of the transient response setting on the step response of the PXIe-4139 for different loads. 7 Extended range pulse currents fall outside DC range limits. In-range pulse currents fall within DC range limits. In-range pulses are not subject to extended range pulsing limitations. 8 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. 9 Measured as the time to settle to within 0.1% of step amplitude, device configured for fast transient response. 10 Current limit set to 50 μa and 50% of the selected current limit range. 11 Current limit set to 20 μa and 20% of selected current limit range. 12 Voltage limit set to 2 V, resistive load set to 1 V/selected current range. PXIe-4139 Specifications National Instruments 9

Figure 5. 1 ma Range, No Load Step Response, Nominal Amplitude (Volts) 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0.1 0.0 20 40 60 80 Time (Microseconds) Slow, no load Normal, no load Fast, no load Custom, no load 100 Figure 6. 1 ma Range, 100 nf Load Step Response, Nominal Amplitude (Volts) 1.3 1.2 1.1 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 0.1 0.0 0.4 0.8 1.2 1.6 Time (Milliseconds) Slow, 100 nf load Normal, 100 nf load Custom, 100 nf load 2.0 Load Regulation Voltage Device configured for local sense Device configured for remote sense 100 μv per ma of output load change (measured between output channel terminals), typical Load regulation effect included in voltage accuracy specifications, typical 10 ni.com PXIe-4139 Specifications

Current, device configured for local or remote sense Load regulation effect included in current accuracy specifications, typical Related Information Voltage Programming and Measurement Accuracy/Resolution on page 4 Current Programming and Measurement Accuracy/Resolution on page 5 Voltage Programming and Measurement Accuracy/Resolution on page 4 Measurement and Update Timing Characteristics Available sample rates 13 (1.8 MS/s)/N where N = 1, 2, 3, 2 24, nominal Sample rate accuracy Maximum measure rate to host Maximum source update rate 14 Input trigger to Source event delay Source event jitter Measure event jitter Pulse timing and accuracy Minimum pulse on time 15 Minimum pulse off time 16 Pulse on time or off time programming resolution Pulse on time or off time programming accuracy Pulse on time or off time jitter Equal to PXIe_CLK100 accuracy, nominal 1.8 MS/s per channel, continuous, nominal 100,000 updates/s, nominal 10 μs, nominal 1 μs, nominal 1 μs, nominal 50 μs, nominal 50 μs, nominal 100 ns, nominal ±5 μs, nominal 1 μs, nominal 13 When sourcing while measuring, both the nidcpower Source Delay and nidcpower Aperture Time properties affect the sampling rate. When taking a measure record, only the nidcpower Aperture Time property affects the sampling rate. 14 As the source delay is adjusted, maximum source rates vary. 15 Pulse on time is measured from the start of the leading edge to the start of the trailing edge. 16 Pulses fall inside DC limits. Pulse off time is measured from the start of the trailing edge to the start of a subsequent leading edge. PXIe-4139 Specifications National Instruments 11

Remote Sense Voltage accuracy Maximum sense lead resistance Maximum lead drop per lead Add (3 ppm of voltage range + 11 µv) per volt of HI lead drop plus 1 µv per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications. 100 Ω 3 V, characteristic Note Exceeding the maximum lead drop per lead value may result in additional error. Related Information Voltage Programming and Measurement Accuracy/Resolution on page 4 Examples of Calculating Accuracy 17 Example 1: Calculating 5 C Accuracy Calculate the accuracy of 900 na output in the 1 µa range under the following conditions: ambient temperature 28 C internal device temperature within T cal ± 5 C 18 self-calibration within the last 24 hours. Solution Since the device internal temperature is within T cal ± 5 C and the ambient temperature is within 23 C ± 5 C, the appropriate accuracy specification is: 0.03% + 100 pa Calculate the accuracy using the following formula: Accuracy = 900 na * 0.03 % + 100 pa = 270 pa + 100 pa 17 Specifications listed in examples are for demonstration purposes only and do not necessarily reflect specifications for this device. 18 T cal is the internal device temperature recorded by the PXIe-4139 at the completion of the last selfcalibration. 12 ni.com PXIe-4139 Specifications

= 370 pa Therefore, the actual output will be within 370 pa of 900 na. Example 2: Calculating 1 C Accuracy Calculate the accuracy of 900 na output in the 1 µa range. Assume the same conditions as in Example 1, with the following differences: internal device temperature within T cal ± 1 C Solution Since the device internal temperature is within T cal ± 1 C and the ambient temperature is within 23 C ± 5 C, the appropriate accuracy specification is: 0.022% + 40 pa Calculate the accuracy using the following formula: Accuracy = 900 na * 0.022 % + 40 pa = 238 pa Therefore, the actual output will be within 238 pa of 900 na. Example 3: Calculating Remote Sense Accuracy Calculate the remote sense accuracy of 500 mv output in the 600 mv range. Assume the same conditions as in Example 2, with the following differences: HI path lead drop HI sense lead resistance LO path lead drop LO sense lead resistance 3 V 2 Ω 2.5 V 1.5 Ω Solution Since the device internal temperature is within T cal ± 1 C and the ambient temperature is within 23 C ± 5 C, the appropriate accuracy specification is: 0.016% + 30 μv PXIe-4139 Specifications National Instruments 13

Since the device is using remote sense, use the remote sense accuracy specification: Add (3 ppm of voltage range + 11 µv) per volt of HI lead drop plus 1 µv per volt of lead drop per Ω of corresponding sense lead resistance to voltage accuracy specifications. Calculate the remote sense accuracy using the following formula: Accuracy = * 2 Ω + 1 * Ω 500 mv * 0.016 % + 30μV + * 2.5 * 1.5Ω 600 mv * 3ppm + 11μV * 3 + 1 1 of lead drop * Ω * 3 = 80 μv + 30 μv + 12.8μV * 3 + 6μV + 3.8 μv = 158.2 μv Therefore, the actual output will be within 158.2 µv of 500 mv. Example 4: Calculating Accuracy with Temperature Coefficient Calculate the accuracy of 900 na output in the 1 µa range. Assume the same conditions as in Example 2, with the following differences: ambient temperature 15 C Solution Since the device internal temperature is within T cal ± 1 C, the appropriate accuracy specification is: 0.022% + 40 pa Since the ambient temperature falls outside of 23 C ± 5 C, use the following temperature coefficient per degree Celsius outside the 23 C ± 5 C range: 0.0006% + 4 pa Calculate the accuracy using the following formula: TemperatureVariation = 23 5 15 = 3 Accuracy = 900 na * 0.022 % + 40 pa + 900 na * 0.0006 % + 4 * 3 1 14 ni.com PXIe-4139 Specifications

= 238 pa + 28.2 pa = 266.2 pa Therefore, the actual output will be within 266.2 pa of 900 na. Trigger Characteristics Input triggers Types Sources (PXI trigger lines <0...7>) 19 Polarity Minimum pulse width Destinations 20 (PXI trigger lines <0...7>) 19 Polarity Pulse width Output triggers (events) Types Destinations (PXI trigger lines <0...7>) 19 Polarity Pulse width Start, Source, Sequence Advance, Measure, Pulse Configurable 100 ns, nominal Active high (not configurable) >200 ns, typical Source Complete, Sequence Iteration Complete, Sequence Engine Done, Measure Complete, Pulse Complete, Ready for Pulse Configurable Configurable between 250 ns and 1.6 μs, nominal 19 Pulse widths and logic levels are compliant with PXI Express Hardware Specification Revision 1.0 ECN 1. 20 Input triggers can be re-exported. PXIe-4139 Specifications National Instruments 15

Protection Output channel protection Overcurrent or overvoltage Overtemperature Automatic shutdown, output disconnect relay opens Automatic shutdown, output disconnect relay opens Isolation Caution Do not connect to MAINs. Do not connect to signals or use for the measurements within CAT II, III, or IV. Isolation voltage, Channel-to-earth ground 150 VDC, CAT I, verified by dielectric withstand test, 5 s, continuous Note Measurement Categories CAT I and CAT O (Other) are equivalent. These test and measurement circuits are not intended for direct connection to the MAINs building installations of Measurement Categories CAT II, CAT III, or CAT IV. Hazardous Voltage Take precautions to avoid electrical shock when operating this product at hazardous voltages. Note Isolation voltage ratings apply to the voltage measured between any channel pin and the chassis ground connector of the front panel. When operating channels in series or floating on top of external voltage references, ensure that no terminal exceeds this rating. Guard Output Characteristics Cable guard Output impedance Offset voltage 2 kω, nominal 1 mv, typical Calibration Interval Recommended calibration interval 1 year 16 ni.com PXIe-4139 Specifications

Power Requirement PXI Express power requirement 2.5 A from the 3.3 V rail and 2.2 A from the 12 V rail Physical Characteristics Dimensions Weight Front panel connectors 3U, one-slot, PXI Express/CompactPCI Express module; 2.0 cm 13.0 cm 21.6 cm (0.8 in. 5.1 in. 8.5 in.), nominal 419 g (14.8 oz), nominal 5.08 mm (8 position) Environment Maximum altitude Pollution Degree 2 2,000 m (800 mbar) (at 25 C ambient temperature) Indoor use only. Operating Environment Ambient temperature range Relative humidity range Storage Environment Ambient temperature range Relative humidity range 0 C to 55 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 low temperature limit and MIL-PRF-28800F Class 2 high temperature limit.) 10% to 90%, noncondensing (Tested in accordance with IEC 60068-2-56.) -40 C to 70 C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2. Meets MIL-PRF-28800F Class 3 limits.) 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) PXIe-4139 Specifications National Instruments 17

Shock and Vibration Operating shock Random vibration Operating Nonoperating 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. Meets MIL-PRF-28800F Class 2 limits.) 5 Hz to 500 Hz, 0.3 g rms 5 Hz to 500 Hz, 2.4 g rms (Tested in accordance with IEC 60068-2-64. Nonoperating test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) Compliance and Certifications Hazardous Voltage This icon denotes a warning advising you to take precautions to avoid electrical shock. Caution You can impair the protection provided by the PXIe-4139 if you use it in a manner not described in this document. Note If your device is hardware revision F or earlier, a snap-on ferrite bead is required to remain in EMC compliance. Refer to the PXIe-4139 Getting Started Guide included in your original shipping kit for information about this bead. To determine which revision of a device you have, open Measurement & Automation Explorer (MAX) and select the device in question. The hardware revision is displayed in the settings pane on the right-hand side. Safety This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: IEC 61010-1, EN 61010-1 UL 61010-1, CSA C22.2 No. 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification section. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity EN 55011 (CISPR 11): Group 1, Class A emissions 18 ni.com PXIe-4139 Specifications

EN 55022 (CISPR 22): Class A emissions EN 55024 (CISPR 24): Immunity AS/NZS CISPR 11: Group 1, Class A emissions AS/NZS CISPR 22: Class A emissions FCC 47 CFR Part 15B: Class A emissions ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations, certifications, and additional information, refer to the Online Product Certification section. CE Compliance This product meets the essential requirements of applicable European Directives, as follows: 2014/35/EU; Low-Voltage Directive (safety) 2014/30/EU; Electromagnetic Compatibility Directive (EMC) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This page contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/environment/weee. PXIe-4139 Specifications National Instruments 19

电子信息产品污染控制管理办法 ( 中国 RoHS) 中国客户 National Instruments 符合中国电子信息产品中限制使用某些有害物质指令 (RoHS) 关于 National Instruments 中国 RoHS 合规性信息, 请登录 ni.com/environment/rohs_china (For information about China RoHS compliance, go to ni.com/environment/rohs_china.) Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2013 2017 National Instruments. All rights reserved. 376115E-01 Jan17