The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers

Similar documents
Characterization of Common Electron Multipliers in Harsh Environments

The Development of a Novel Electron Multiplier with an Onboard Integral High Voltage Power Supply for use in Mass Spectrometers

Performance Characterization Of A Simultaneous Positive and Negative Ion Detector For Mass Spectrometry Applications

Residual Gas Analyzers

Residual Gas Analyzers RGA100/200/ amu, 200 amu and 300 amu RGAs

Residual Gas Analyzers XT Series

Electron Multipliers. Introduction to ETP Electron Multipliers Selection by Technique: LC-MS 242

Lecture 12 OPTICAL DETECTORS

Detectors for microscopy - CCDs, APDs and PMTs. Antonia Göhler. Nov 2014

Then the mass spectrometer should go to the field, in your hand, small as possible, light as possible, and operated with light battery.

Transpector 2. Gas Analysis System. NEW-GENERATION RGAs WITH SUPERIOR SENSITIVITY AND PERFORMANCE

Sub-mm Linear Ion Trap Mass Spectrometer Made Using Lithographically Patterned Ceramic Plates

Advantages of Multi-Turn-TOF Technology

Mass Spectrometry and the Modern Digitizer

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

MCPs and MCP based detectors. Raquel Ortega

Development of a Small Residual Gas Analyzer Utilizing the Quadrupole Array Structure Micropole System ~ QL Series ~

830 Vacuum Quality Monitor

835 Vacuum Quality Monitor (VQM TM ) 2012 Brooks Automation, Inc. Proprietary Information

DN 16 ISO-KF. Dimensional drawings THERMOVAC TM 101 (mm)

ELECTRON MULTIPLIERS General Technical Information

O.H.W. Siegmund, Experimental Astrophysics Group, Space Sciences Laboratory, 7 Gauss Way, University of California, Berkeley, CA 94720

Performance of Microchannel Plates Fabricated Using Atomic Layer Deposition

Autoresonant Trap Mass Spectrometry

A New Class of Robust Sub-nanosecond TOF Detectors with High Dynamic Range

Project by: Dr. Jorge A. Diaz. Physics School, University of Costa Rica, National Center for High Technology (CENAT)

PRELIMINARY. Specifications are at array temperature of -30 C and package ambient temperature of 23 C All values are typical

ECEN 4606, UNDERGRADUATE OPTICS LAB

Ion optics innovations for increased sensitivity in hybrid MS systems

The Extrel MAX Systems are UHV-compatible flange mounted Quadrupole Mass

Timing and cross-talk properties of BURLE multi-channel MCP PMTs

The NIRDAPD TEC series photodetector is available in two different active area sizes: Key Features

High collection efficiency MCPs for photon counting detectors

Advances in microchannel plate detectors for UV/visible Astronomy

HIQUAD. New high-end mass spectrometer! Fast, flexible and easy to operate.

Refer to manual. Failure to read message could result in personal injury or serious damage to the equipment or both.

Cylindrical Toroidal Ion Trap Mass Spectrometer. Daniel Austin and Nick Taylor Brigham Young University, Provo, Utah

CRL MASS SPECTROMETRY FACILITY INSTRUMENT USER MANUAL MALDI MICRO. Operating Instructions. Basement Spec Lab:

Homework Set 3.5 Sensitive optoelectronic detectors: seeing single photons

Pixel hybrid photon detectors

Product Range Electronic Units

Performance Quality Value The one source for all your detector needs

Improving Signal-to-Noise by Identifying Sources of Noise in Mass Spectrometer Systems

LI-192 Underwater Quantum Sensor

5. Scintillation counters

5. Scintillation counters

Performance characteristics of a new wide range, fast settling electrometer design for a residual gas analysis mass spectrometer

Physics Requirements for the CXI 0.1 micron Sample Chamber

Ultrasonic Level Detection Technology. ultra-wave

Production of HPDs for the LHCb RICH Detectors

HPR-30 Vacuum Process Gas Analyser A differentially pumped RGA system for vacuum process monitoring

Advanced Materials Research Vol

UNIVERSITY OF CALIFORNIA College of Engineering Department of Electrical Engineering and Computer Sciences

Development of a Low Cost, Low Power, Miniature Sector Mass Spectrometer with IonCCD Detection

Mass spectrometers have evolved into standard

OPTI510R: Photonics. Khanh Kieu College of Optical Sciences, University of Arizona Meinel building R.626

Dr. James McFadden, UC Berkeley

NON-AMPLIFIED HIGH SPEED PHOTODETECTOR USER S GUIDE

Low-cost sensors and high-end instruments for mapping urban NO x pollution

CHAPTER 11 HPD (Hybrid Photo-Detector)

NON-AMPLIFIED PHOTODETECTOR USER S GUIDE

The Benefits of Photon Counting... Page -1- Pitfalls... Page -2- APD detectors... Page -2- Hybrid detectors... Page -4- Pitfall table...

Receiver Signal to Noise Ratios for IPDA Lidars Using Sine-wave and Pulsed Laser Modulation and Direct Detections

Historical. McPherson 15 Mount

Progress towards a 256 channel multianode microchannel plate photomultiplier system with picosecond timing

Simulations Guided Efforts to Understand MCP Performance

8854 Photomultiplier. 129-mm (5-inch) Diameter, 14-stage QUANTACON TM Type Having a Bialkali Photocathode and High- Gain Gallium-Phosphide Dynodes

PrismaPlus. The New Mass Spectrometer with the Added Plus! Modular Design. Powerful Software. Wide Range of Applications.

Thermal Conductivity Sensor for Leak or Pressure Detection MTCS2601. MTCS2601 silicon sensing die in SMD ceramic package

Today s Outline - January 25, C. Segre (IIT) PHYS Spring 2018 January 25, / 26

TRITON Plus Thermal Ionization MS. Another step ahead. in TIMS. Earth and Planetary Sciences Nuclear Sciences

PRISMAPLUS. The precise solution for mass spectrometry. Modular design. Powerful software. Wide range of applications.

Chemical Engineering 412

R & D for Aerogel RICH

Development of dual MCP x-ray imager for 40 ~ 200 kev region

Introduction INTRODUCTION DISCOVER THE NEXT GENERATION IN EDGE WELDED METAL BELLOWS

PSD Characteristics. Position Sensing Detectors

4X150A/7034 Radial Beam Power Tetrode

queensgate a brand of Elektron Technology

SL300 Snow Depth Sensor USL300 SNOW DEPTH SENSOR. Revision User Manual

CHAPTER 3 BASIC OPERATING METHODS OF PHOTOMULTIPLIER TUBES

LC/MS/MS. Page Header. triple quadrupole mass spectrometer.

A Fast Waveform-Digitizing ASICbased DAQ for a Position & Time Sensing Large-Area Photo-Detector System

A new Photon Counting Detector: Intensified CMOS- APS

How Does One Obtain Spectral/Imaging Information! "

PRISMAPRO. The latest generation of compact mass spectrometers. Modular design. Powerful software. Low detection limit.

AccuTOF LC Training Course

PoS(TWEPP-17)025. ASICs and Readout System for a multi Mpixel single photon UV imaging detector capable of space applications

Improvement of the MCP-PMT performance under a high count rate

ICP-MS. plasma 3. Multi-Collector ICP-MS.

SINPHOS SINGLE PHOTON SPECTROMETER FOR BIOMEDICAL APPLICATION

A T u n e d M e a s u r e

[NIGHT VISION TECHNOLOGY] SEMINAR REPORT

Features. Applications. Optional Features

Integrating Spheres. Why an Integrating Sphere? High Reflectance. How Do Integrating Spheres Work? High Damage Threshold

AG SERIES ISOLATED, PROPORTIONAL DC TO HV DC CONVERTERS

Agilent GC-MS Maintenance:

Design of the High Voltage Supply Module of a Prototype Energy Spectrometer for Solar Wind Plasma Measurement

LABORATÓRIUMI GYAKORLAT SILLABUSZ SYLLABUS OF A PRACTICAL DEMONSTRATION. financed by the program

UHV Double-Ended SMA Feedthrough, , Conflat compatible flange

Transcription:

The Development of Miniature Electron Multipliers for Use In Portable Mass Spectrometers *Bruce N. Laprade, Lenny Erickson William G. Dunn and Reginald Farr BURLE Electro-Optics Sturbridge MA Paper 10400-2100 Presented at The 2004 Pittsburgh Conference March 9, 2004 *Presenter

Introduction Easy to use, truly field portable mass spectrometers have been the dream of industrial hygienists, forensic pathologists, law enforcement officials, geologists, arson investigators and a host of others interested in moving lab quality analysis into the field location. Although much progress has been made in reducing room sized instruments to benchtop size, most portable instruments today would best be described as transportable.

Discussion Designers of field portable mass spectrometers face three challenges: Reduce The Size Reduce The Weight Reduce The Power Consumption A quick assessment of most existing systems reveals that the high vacuum system is the most significant contributor in all three categories.

Discussion cont. The purpose of the high vacuum housing and pumping system is simply to provide an environment in which the analyte molecules can be converted into ions, separated by mass, and then detected and identified. All mass separation techniques (quadrupole, time of flight, sector, and ion trap) require the ions to travel some distance in order to be separated and identified. The vacuum environment must therefore be sufficient to prevent a significant number of collisions with residual gas molecules.

Discussion cont. According to the kinetic theory of gases, the mean free path of an ion is described by the following equation: L = 1 2pnσ 2 with n = p kt

Discussion cont. Typical MALDI Time-of-Flight Mass Spectrometers require an ion flight path length of 1 to 2 meters. Modern quadrupole based instruments typically operate at vacuum pressures in the high 10-6 Torr range because ions need to travel 25 cm or more before they reach the detector. If it were possible to shrink the flight path length requirement to less than 5 cm then it should be possible to operate a system in the milli-torr range. milli-torr vacuum levels can be achieved with simple low cost vacuum pumps.

Discussion cont. The sensitivity of a mass spectrometer can be greatly enhanced by the addition of an electron multiplier. The development of micro-machining technology has spawned a whole new generation of small, portable mass spectrometers. The entire source and mass filter can now be reduced to the size of a postage stamp. Miniature mass filters can operate at elevated pressures, enabling less complex pumping systems to be used. Unfortunately, conventional electron multipliers remain relatively large (greater than 8 cm long), and do not operate well above 10-4 torr.

There are 4 main types of Ion Detectors Faraday Cup: Good Linearity, Operates at atmospheric Pressure, Produces no Gain or signal amplification. Discrete Dynode Multipliers: Large, Bulky, Become Noisy when operated above 10-5 Torr Single Channel Electron Multipliers: Channeltrons, Spiraltrons and MAGNUM - Compact, High Gain, Low noise Electron Multipliers, operate well at pressures extending into the 10-4 Torr range. Microchannel Plate (MCP) Detectors: Miniature and subminiature designs produce high gain and low noise, and have demonstrated operation at 3 X 10-2 Torr.

Recent Advances in Electron Multiplier Technology have enabled the development of compact Ion Detectors for use in: Leak Detectors Portable Mass Spectrometers Dedicated Sensors Environmental Sensors Industrial Sensors

The Ideal Detector for Portable Mass Spectrometers Small High Gain and Low Noise Rugged Tolerate poor vacuum conditions Long Life Low Power Consumption Should Tolerate system failures Easily Replaced Excellent Temporal Resolution (TOF only)

Size and Weight Comparison For Typical Ion Detectors Detector Type Weight Height (grams) (mm) 3 X 3 mm MCP 0.016 0.2 Mini Channeltron 0.51 12.5 Quantum MCP Detector 1.39 7.8 Spiraltron 2.6 65 Microtron 4.59 9 Mini TOF 10 15 5900 Magnum 26 35 XPR 25 25 Typical Discrete Dynode Multiplier 35 100

Assorted Miniature Electron Multipliers

Channeltron Electron Multiplier Operation Incident Ions or photons

Miniature Channeltron Electron Multiplier

Miniature Channeltron Electron Multiplier Mechanicals

Miniature Channeltron Electron Multipliers Compact Single Piece Construction Ideal for Portable Low Cost Instruments Can be mounted in various configurations Analog or Pulse Counting Applications Gain Greater than 10 million Low Noise Typical Mechanical and Electrical Characteristics ELECTRICAL CHARACTERISTICS Operation: Maximum Operating Pressure: Maximum Specified Operating Voltage: Bias Current @ 3,000 Volts: Resistance (For Reference Only): SPECIFICATIONS Analog or Pulse Counting 5.0 x 10-5 Torr 3000 Volts 10 to 55 Microamps 55 to 85 Megohms Minimum Gain @ 3,000 Volts: 1.0 x 10 7 Maximum Dark Current @ 3,000 Volts: 1.0 x 10-12 Amps Maximum Dark Count @ 3,000 Volts: 1 ct./sec Maximum Linear Output Current: 10% of Bias Current (Typical) Typical Applications: Leak Detectors Portable Mass Spectrometers Dedicated Sensors Environmental Sensors Industrial Sensors

Typical Spiraltron Performance Specification PHYSICAL CHARACTERISTICS SPECIFICATIONS Cone Mechanical Dimensions: 10 mm round or 6 X 20 mm rectangular Maximum Vacuum Bake Specification: 8 Hours at 300 C at 1.0x10-5 Torr or Lower Operating Temperature Range -50 to 120 C ELECTRICAL CHARACTERISTICS SPECIFICATIONS Operation: Pulse Counting Maximum Operating Pressure: 5X10-4 Torr Maximum Specified Operating Voltage: 3000 Volts Bias Current Range @ 3000 Volts: 10-75 Microamps Resistance (For Reference Only): 40-100 Megaohms Minimum Gain @ 3000 Volts: 1x10 8 Maximum Dark Count @ 3000 Volts: counts in 60 seconds Maximum Linear Output Current: 10% of Bias Current (Typical) Maximum Pulse Height Distribution: 75% Full Width Half Maximum

MAGNUM Family of Products High performance Analog or pulse counting electron multipliers 6-Channel Structure Provides: - Long Life - High Output Current - Improved Linearity - Reduced Ion Feedback - Low Manufacturing Cost - Fewer Components - End User Replacements - Tolerates operation at higher pressure

Miniature Microchannel Plate Electron Multiplier

Sub-Miniature Microchannel Plate Detector

Microtron Detector PHYSICAL CHARACTERISTICS of The Detector SPECIFICATION Package Outside Diameter: 14.2mm (0.56 ) Package Height: 5.8mm (0.23 ) Overall Package Height (Including Leads): 12.4mm (0.49 ) Quality Diameter: 5.5mm (0.217 ) Center-to-Center Spacing: Pore Size: Bias Angle: 8 ± 1 Open Area Ratio: Quality Level: ELECTRICAL CHARACTERISTICS of DETECTOR Electron Gain @ 2000 Volts: Bias Current Range @ 2000 Volts: Resistance: Dark Noise: Linear Output Current Density: (Microamps/cm 2 ) 32µm Nominal 25µm Nominal 45% Minimum Detection SPECIFICATION 2 x 10 6 Minimum 10-30 Microamps 67-200 Megohms Reference < 5 ct. / sec. Typically 10% of Bias Current Density The Microtron detector assembly contains two matched Detection Quality Long-Life Microchannel Plates mounted in stainless steel hardware. The assembly is bakeable to 300 C. Detection Quality detector assemblies are intended for use in applications where only signal detection is required. These economical devices are used as signal detectors and amplifiers and are typically used in applications such as leak detection, residual gas analysis (RGA), or point source detection.

Quantum Detector

Quantum Detector PHYSICAL CHARACTERISTICS of MCP Quality Diameter: Center-to-Center Spacing: Pore Size: SPECIFICATION 3.3mm Minimum 32µm Nominal 25µm Nominal Bias Angle: 8 ± 1 Open Area Ratio: Quality Level: ELECTRICAL CHARACTERISTICS of DETECTOR Electron Gain @ 1000 Volts: Bias Current Range @ 1000 Volts: Resistance: Dark Noise: Linear Output Current Density: (Microamps/cm 2 ) 40% Minimum Detection SPECIFICATION 1 x 10 4 Minimum 0.40 5 Microamps 100 1000 Megaohms (Reference) 0.5 x 10-12 Amps Maximum (Typical) Typically 10% of Bias Current Density The CEMA Model 603.8MA detector assembly contains one Detection Quality Advanced Performance Lon- Life Microchannel Plate and a metal anode readout mounted in ceramic and kovar hardware. The assembly is bakeable to 300 C. Detection Quality detector assemblies are intended for use in applications where image quality is not critical. These economical devices are used as signal detectors and amplifiers and are typically used in applications such as time-of-flight mass spectrometry, residual gas analysis (RGA), or point source detectors

Microchannel Plate Electron Multiplier Faraday Cup Ion Beam MCP Holder Anode Ceramic Base Designed for operation at up to 3 x 10-2 Torr

3. FUNCTIONAL SPECIFICATIONS Figure 2-1. Dimensional Requirements Test Characteristic Quality Diameter: Center-to-Center Spacing: Pore Size: Value 11.02mm (.434 ) x 12.65mm (.498 ) Minimum 32µm Nominal 25µm Nominal Bias Angle: 19 ± 1 Open Area Ratio: Quality Level: Maximum Electron Gain @ 775 Volts: Minimum Electron Gain @ 1225 Volts: Bias Current Range @ 1100 Volts: Resistance: Dark Current: Low Gain 300-500 @ 1x10-2 Torr (Mass 23) High Gain 10,000 @ 1x10-6 Torr (Mass -1) Faraday Cup Mode @ 1x10-6 Torr (Mass -1) 50% Minimum Detection 5 x 10 2 Maximum 1 x 10 4 Minimum 30-80 Microamps 14-37 Megohms Reference (AMPS) -1.0x10-11 to 1.0x10-11 Average 1.0x10-11 Maximum Standard Deviation -1.0x10-12 to 1.0x10-12 Average 5.0x10-14 Maximum Standard Deviation -1.0x10-13 to 1.0x10-13 Average 1.2x10-15 Maximum Standard Deviation

Miniature Time-of-Flight Detector Miniature Time-Of-Flight Detectors for Portable Instruments. These detectors are light weight (less than 10 grams) and Consume less than 0.02 watts. The compact design features 750 ps pulse widths and rise times approaching 300 ps.

PHYSICAL CHARACTERISTICS Microchannel Plate Type: Active Diameter: Center-to-Center Spacing: Pore Size: SPECIFICATIONS Extended Dynamic Range Long-Life 8mm Minimum 6µm Nominal 5µm Nominal Bias Angle: 5 ± 1 Open Area Ratio: 62% Minimum Quality Level: Detection PHYSICAL CHARACTERISTICS of DETECTOR SPECIFICATION Mechanical Dimension Defined by Drawing Number: 31959 Vacuum Bake Specification (Not Operating): Bakeable @ 150 C Operating Temperature Range: -50 to 100 C Vacuum Flange: Input Grid: ELECTRICAL CHARACTERISTICS of DETECTOR Operating Pressure (Maximum): Maximum Specified Operating Voltage: Bias Current Range @ 2000 Volts: Resistance (For Reference Only): Gain (Minimum): Linear Output Current Density (Microamps / cm 2 ) Dark Count (Maximum) 1 : Pulse Width (Minimum): 1.33 Inch Conflat 85% Transmission (Typical) SPECIFICATION 1.0 x 10-5 Torr 2400 Volts 6-30 Microamps 67-333 Megohms 1.0 x 10 6 @ 2000 Volts Typically 10% of Bias Current Density 2.5 cts/second 750 psec @ Half Height (Typical)

Summary Microchannel Plate (MCP) and Single Channel Electron Multipliers have been developed and are in production that offer significant performance advantages Over Faraday Cup and Discrete Dynode Multipliers in field portable Mass Spectrometer applications Including: - Compact, Plug and Play Designs - Light Weight - Tolerant of Poor Vacuum Conditions -High Gain -Low Noise - High Dynamic Range - Long Life - Low Cost of Ownership