Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Similar documents
Zeiss AxioImager.Z2 Brightfield Protocol

Guide to Confocal 5. Starting session

Lab: The Compound Microscope

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Quick Operation Guide

Match the microscope structures given in the left column with the statements in the right column that identify or describe them.

Laboratory Introduction

A BRIEF INTRODUCTION TO MICROSCOPY The two key properties of a microscope that allow you to see microbes are resolution and magnification.

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

MICROSCOPE LAB. Resolving Power How well specimen detail is preserved during the magnifying process.

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Biology 29 Cell Structure and Function Spring, 2009 Springer LABORATORY 1: THE LIGHT MICROSCOPE

Zeiss LSM 880 Protocol

Instructions. To run the slideshow:

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Zeiss LSM 780 Protocol

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Microbiology Laboratory 2

OPELCO OPtical ELements COrporation LB Objective Series for Biological Use

Nikon Instruments Europe

Microscope anatomy, image formation and resolution

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Horiba Jobin-Yvon LabRam Raman Confocal Microscope (GERB 120)

Exercise 2-A MICROSCOPIC TECHNIQUE & EXAMINATION OF MICROORGANISMS

Marine Invertebrate Zoology Microscope Introduction

Exercise 2-A MICROSCOPIC TECHNIQUE & EXAMINATION OF MICROORGANISMS

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Basics of Light Microscopy and Metallography

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

CALIBRATION OF MICROSCOPE EYEPIECE GRATICULE

Media Cybernetics White Paper Spherical Aberration

Basic Microscopy for Plant Biology

SWIFT SERIES M2252DGL MICROSCOPE

LSM 710 Confocal Microscope Standard Operation Protocol

Things to check before start-up.

2/4/15. Brightfield Microscopy! It s all about Magnification..! or is it?!

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Operating Instructions for Zeiss LSM 510

WITec Alpha 300R Quick Operation Summary October 2018

The microscope is useful in making observations and collecting data in scientific experiments. Microscopy involves three basic concepts:

Bringing Answers to the Surface

Education in Microscopy and Digital Imaging

Microscopy: Fundamental Principles and Practical Approaches

Figure 3.4 Approximate size of various types of cells. ~10 um. Red Blood Cells = mm 1500 um. Width of penny Pearson Education, Inc.

THE COMPOUND BRIGHTFIELD MICROSCOPE

LSM 510 Training Notes

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

2 How to operate the microscope/obtain an image

Zeiss LSM 510 Confocor III Training Notes. Center for Cell Analysis & Modeling

Nikon E800 Operating Instructions.

1. Preliminary sample preparation

Swift M10 Series Microscope Use and Care Manual

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Using the Nikon TE2000 Inverted Microscope

Very short introduction to light microscopy and digital imaging

Working Simultaneously. The Next Level of TIRF Microscopy. cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence

Artisan Technology Group is your source for quality new and certified-used/pre-owned equipment

User Manual. Digital Compound Binocular LED Microscope. MicroscopeNet.com

LAB 1 Introduction to Microscopy

How to Use a Microscope

MICROSCOPY MICROSCOPE TERMINOLOGY

Introduction to Microscopes

LSM 510 Meta Training Notes

Swift M3600, M3700 Series Digital & Compound Microscope

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope


LSM 800 Confocal Microscope Standard Operation Protocol

Standard Operating Procedure of Atomic Force Microscope (Anasys afm+)

Microscopy. The dichroic mirror is an important component of the fluorescent scope: it reflects blue light while transmitting green light.

Care and Use of the Compound Light Microscope

Lab 10: Lenses & Telescopes

Tissue Preparation ORGANISM IMAGE TISSUE PREPARATION. 1) Fixation: halts cell metabolism, preserves cell/tissue structure

Nikon E800 Microscope. Operating Instructions

User manual for Olympus SD-OSR spinning disk confocal microscope

Nasmyth Ultraview Vox User Protocol

MICROSCOPY and CELL STRUCTURE

Simplified Instructions: Olympus Widefield Microscope S1230

User Operation of JEOL 1200 EX II

Nature Protocols: doi: /nprot Supplementary Figure 1. Schematic diagram of Kőhler illumination.

CCAM Microscope Objectives

User instructions Compound laboratory microscope

BLIPS perfectly aligned cleaned by any impurity Gently press the lens

Activity 6.1 Image Formation from Spherical Mirrors

Olympus LEXT OLS 4000 Confocal Laser Microscope

Indian Institute of technology Madras Presents NPTEL NATIONAL PROGRAMME ON TECHNOLOGY ENHANCED LEARNING

Image-Pro Plus 7.0 Product Note

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

FLUORESCENCE MICROSCOPY. Matyas Molnar and Dirk Pacholsky

AGES 10+ INSTRUCTION MANUAL. 800x Power Advanced Microscope Biological Experiments Gear. x 2 NOT INCLUDED

LSM 780 Confocal Microscope Standard Operation Protocol

Anatomy: Introduction to the Light Microscope

ML7520 ML7530 DIOPTER ADJUSTMENT RING BINOCULAR BODY, INCLINED 30. (a) Field Iris Control Lever. (c) Filter Slots EYEPIECES, KHW10X

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Physiology Honors Interactive Notebook

1 Co Localization and Working flow with the lsm700

Bio 252: Microscopy Study THE COMPOUND MICROSCOPE

The Next Level of TIRF Microscopy. cell^tirf Illuminator Motorized Total Internal Reflection Fluorescence

Transcription:

Tech Note Oil Objective Introduction The Lionheart FX automated imager is compatible with high numerical aperture oil immersion objectives. These objectives offer magnification up to 100X and significantly improve the optical resolution of the system. Cellular details, as small as a few hundred nanometers, can be resolved on the Lionheart FX using oil-immersion microscopy. Here we describe how to use oil immersion objectives on the Lionheart FX and best practices for producing high quality fluorescent images. Hardware There are two Olympus oil immersion objectives available for use with the Lionheart FX automated cell imager which can deliver magnifications of either 60X or 100X (Figure 1). As the name suggests, these highly specialized objectives must be used with oil. A small drop of immersion oil is deposited between the front lens of the objective and the coverslip linking them together. The transparent oil has a high viscosity that produces an uninterrupted liquid column between the two surfaces. Moreover, the oil is optically similar to glass with a refraction index of 1.515. This increases the maximum angle of the cone of light that can enter the front lens of the objective which is a characteristic that is described by the objective s numerical aperture (NA). The greater the NA of an objective, the more light it is able to collect. Additionally, objectives with large NA values can resolve greater detail in a sample. Oil immersion objectives have much larger NA values compared to dry objectives. Figure 1. Oil-immersion objectives available for use with the Lionheart FX. BioTek Instruments, Inc. P.O. Box 998, Highland Park, Winooski, Vermont 05404-0998 USA COPYRIGHT 2016 Phone: 888-451-5171 Outside the USA: 802-655-4740 Fax: 802-655-7941 Email: customercare@biotek.com www.biotek.com

Detailed specifications of these objectives can be found in Table 1. An important specification to keep in mind when using oil-immersion is the objective s working distance-- the distance between the front lens of the objective and the surface of the coverslip. The oil-immersion objectives used in the Lionheart FX have extremely short working distances (<200 microns). Therefore, great care must be taken to prevent damaging the front lens by collisions with the coverslip. Importantly, the short working distance of oil-immersion objectives make them incompatible with thick plastic bottom microplates (or labware) and must be used with 0.17 mm thick coverslips. Cell culture vessels typically have bottom thickness that range from 0.5 mm to 2.0 mm. Due to the short working distance of oil immersion objectives, attempting to image cells through thick bottom vessels is impossible with these short working distance objectives, as the front lens of the objective would collide with the culture vessel before the sample could be brought into focus. Although the front end of both the 60x and 100x oil objectives are spring loaded to minimize costly damage from inadvertent collisions, they should never be used with incompatible culture vessels. Magnification 100x 60x Numerical aperture 1.40 1.42 Working distance 0.13 mm 0.15 mm Depth of field.28 µm.27 µm Coverslip thickness (mm) 0.17 0.17 Optical correction Plan Super-Apochromat Plan Apochromat Mfg. Olympus Olympus Table 1. Oil-immersion objectives available for the Lionheart Fx. Procedure: Using Oil Immersion Objectives with the Lionheart FX Instructions for installing oil-immersion objectives can be found in the Lionheart FX Operator s manual. Note, oilimmersions objectives are not calibrated during setup and are only compatible with manual mode imaging. 1) Define the bottom elevation of your vessel. Before entering manual mode, it is important to define the bottom elevation of your slide, glass bottom plate or any other vessel you intend to image. Instructions for defining the bottom elevation of a vessel can be found in the Gen5 help menu or on page 39 of the Cell Imaging Training Guide. You will need to use an air ( dry ) objective to complete this process. 2) Find the area and/or objects of interest with a lower power air objective. As usual in microscopy, it is most efficient to begin with a lower magnification to find your objects of interest before switching to a high magnification objective. Using a lower power air objective (such as the 20X) will make it easier to find an area of interest because of its large field of view. Use either the joystick or the navigation buttons in Gen5 to center objects of interest before moving on. 3) Lower the objectives 500 microns BEFORE switching to the oil-immersion objective. The oil-immersion objectives used in the Lionheart FX are not calibrated during instrument setup. Unlike the non-oil objectives installed in the instrument, focal height may not be maintained once you switch from a lower power objective to an oil-immersion objective. It is recommended that you lower the z-height before switching objectives to reduce risk of collisions. 2 P age

4) Change to the oil-immersion objective and follow the oiling procedure. Choose the oil-immersion objective you wish to use and click the oil can icon ( ) to open the Objective Oil Proceduree wizard. Follow the instructions on-screen and click close when you have completed all the steps. Important: Do not add an excessivee amount of oil to the front lens of the objective. One drop should be sufficient to cover the entire lens. Adding too much could cause oil to run down the side of the objective and damage sensitive components in the instrument. 5) Slowly raise the objective until your object of interest comes into sharp focus. Once the Objective Oil Procedure wizard is closed, Gen5 will resume the manual mode session. Slowly increase the focus position by using either the joystick or the focus arrows in the manual mode window. 6) Center the objective of interest and adjust exposure settings. Use the directional arrows to center your object of interest in the field of view. Adjust the exposure by clicking the auto expose button. You can also manually define your exposure settings using the sliders in the Manual Mode window. 3 Page

7) Clean the objective before exiting Gen5 or changing objectives. Gen5 considers an objective to be dirty after completing the steps of the Objective Oil Procedure. Before exiting Gen5 or changing objectives, you will be prompted to clean the objective. Properly removing any residual oil from the front lens of the objective is critical to prevent costly damage to the instrument or the objective itself. To clean the objective, follow the instructions in the Objective Clean-Up window and click close when you are done.! Warning: Only clean the objective with lens paper. Using other cleaning products such as multipurpose lab wipes or paper towels will scratch the lens of the objective causing irreparable damage. Never polish or push hard on the front lens with dry lens paper. Instead, gently wipe the oil off and use multiple sheets of lens paper if needed. Best Practices and Tips Below is a list of best practices and tips for imaging with oil-immersion objectives. Following these recommendations will allow you to make the most of these high magnification objectives and ensure the capture of high quality images. 1) Image your sample through #1.5 (0.17 mm thick) cover slips. As mentioned above, the oil immersion objectives available for use in the Lionheart FX can only image samples through glass 0.17 mm thick. Using coverslips of the correct thickness is crucial to obtaining high quality images and to prevent damage to your objective. Note: Since the Lionheart FX is an inverted microscope by design, microscope slides must be loaded with cover slip facing down. 2) Only use Olympus Immersion Oil Type-F supplied with your objective. Objective manufacturers design immersion oil based on the optical properties of their objectives. Using immersion oil from another manufacturer can have significant effects on image quality due to small variations of the oil s refractive index. Importantly, if you intend to use a slide previously imaged on another microscope, make sure that the coverslip has been cleaned thoroughly. Even a small amount of residual oil from another manufacturer can degrade image quality when mixed with the proper oil. 3) Lower the LED intensity to minimize the effects of photobleaching. Keep in mind that using high magnification oil-immersion objectives will photobleach your sample much faster than an objective with a lower NA. Objectives with high NA collect more light from the sample and therefore are more sensitive to dim signals. However, this also means that your sample is exposed to more light as the objective acts as the source of illumination in fluorescence microscopy. When using these objectives it is good practice to lower the LED intensity while focusing or searching for objectives of interest and turn the LED off when not in use. 4) Be patient while focusing. Quickly raising the objective can result in damaging collisions between the objective s front lens and the coverslip. Slowly bring your specimen into focus using the fine focus arrows or, if available, by gently twisting the joystick. 4 P age

5) Always secure your slide to the holder. Use the clips on the slide holder to secure the sample slide in place. As the objective is brought closer to the sample the high surface tension of immersion oil can push the slide out of its holder if not properly secured. 6) Clean your objective with pure ethanol as needed. If required, pure ethanol can be used to clean excessive oil from the objective. As described above, only use non-abrasive lens paper when cleaning an oil-immersion objective to prevent scratches. 5 P age