Insulation Test System

Similar documents
Insulation Test System

Combination Wave Test System

Telecom Test System. Lightning Tests

Low Frequency Measuring System

EMC PARTNER. DO-160 & MIL-STD-461G TEST SYSTEM Indirect Lightning Testing

Avionics Test System. Lightning Tests

MILITARY & AEROSPACE TESTING

Electricity Meter Test System

ESD MODULAR HANDHELD TESTER Electrostatic Discharge Testing

ESD Test System. Immunity Tests

Low Frequency Measuring System

Technical Specification

Technical Specification

MIL Standard 461 G. final release December 11 th, EMC PARTNER - Largest range of impulse test equipment up to 100kV and 100kA

Advanced Test Equipment Rentals ATEC (2832)

Schlöder GmbH - EMC Test and Measurement Systems Model #

10. DISTURBANCE VOLTAGE WITHSTAND CAPABILITY

7P Series - Surge Protection Device (SPD) Features 7P P P

Conducted Immunity Testing

Immunity Testing for the CE Mark

INTERNATIONAL STANDARD

Introduction EMC. Filter parameters. Definition of EMC / EMI. X-Capacitor. Sources of EMI. Coupling mechanism. Y-Capacitor.

Intra-system EMI hardening for increased machine reliability. Ray Brett.

Surge Generator for MIL-STD 1275

VERI50 EFT for EFT calibration applicable to all TRA2000 generators and CDN2000 coupling filters

Advanced Test Equipment Rentals ATEC (2832) Lightning Surge Simulator LSS-F02.

Voltage Transient Emission Test

Features. Regulated Converter. RAC02-SE/277 2 Watt Single Output RAC02- SE/277. AC/DC Converter

Technical Requirements for Resistibility of Telecommunications Equipment to. Overvoltage and Overcurrent

EMC standards. Presented by: Karim Loukil & Kaïs Siala

Technical Requirements for Resistibility of Telecommunications Equipment to. Overvoltage and Overcurrent

Transient Protection

CAR - TEST - SYSTEM 14

A Comparison Between MIL-STD and Commercial EMC Requirements Part 2. By Vincent W. Greb President, EMC Integrity, Inc.

2620 Modular Measurement and Control System

E M C T E S T Y O U R SOURCE FOR TOP Q U A L I T Y TEST EQUIPMENT & E M C. w w w. h v t e c h n o l o g i e s. c o m

XY-Capacitor Surge Test (IEC )

CS114 + CS115 + CS116

Regulatory authority use EMC PARTNER's HAR1000 to enforce Harmonic and Flicker CE requirements in Germany

SIOV metal oxide varistors

H V T E S T S O L U T I O N S PA RT N E RS F O R H V & E M C

The Role of the Grounding System in Electronics Lightning Protection

Understanding Design, Installation, and Testing Methods That Promote Substation IED Resiliency for High-Altitude Electromagnetic Pulse Events

satech SynchroStar GPS 200 Series

Electromagnetic and Radio Frequency Interference (EMI/RFI) Considerations For Nuclear Power Plant Upgrades

Center for Quality Engineering. Test Report No.: B0EF0002

PMT/UMT(275) Power Gap Description and Use Application Note

Smart Termination TECHNICAL SPECIFICATIONS CURRENT VOLTAGE INSTRUMENTRANSFORMER

Center for Quality Engineering. Test Report No.: B0JA0002

Safety through Competence. Short Form Catalogue. Measuring Testing Controlling World Class Hardware and Software

Features. Regulated Converter. RAC02-SE/277/W 2 Watt Single Output. RAC02- SE/277/W Wired. AC/DC Converter

50W Constant Current (700mA) Dimming LED Driver

Tektronix: Products > AWG 2040 Arbitrary Waveform Generator

Overview of EMC Regulations and Testing. Prof. Tzong-Lin Wu Department of Electrical Engineering National Taiwan University

Lightning Surge Simulator LSS-F03 series

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

Harmonic control devices. ECE 528 Understanding Power Quality

USER'S MANUAL ACDC-100 TRMS ACDC-100. Versatile AC/DC Clamp-on Multimeter Series

WIRELESS INSULATOR POLLUTION MONITORING SYSTEM

DC/DC power module 3.3V / 4.5A / 14.8W

MEDIUM & HIGH VOLTAGE

Features. Regulated Converters. RAC01-C RAC02-C 1-2 Watt Single Output RAC0_- C. AC/DC Converter

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

Power Quality Issues from an EMC Point of View

400,440Vac Vdc V, A, W, VA, var, Hz, V, Avg. A cos phi, sin phi and phi measurement kwh, kvarh (via Pulse Output)

Type CP-S, CP-C & CP-A Switch mode

Harmonizing the ANSI-C12.1(2008) EMC Tests. Harmonizing the ANSI-C12.1(2008) EMC Tests

EMC Test report for LED Panel Light Models , , , , ,

DC/DC power module 1.8 V / 5A / 9W

Overcurrent Protection / 7SJ45

REK 510 Current injection device for earth-fault protection of a synchronous machine rotor. User s Manual

OLU131P701N1A PSOV131701A(N) PSOV131701B(N) PSOV131701D(N) (for SAMSUNG) PSOV131701E(N) (for SAMSUNG) SOLUM CO., LTD

System description 4. SERVICES ONSITE INSTALLATION AND TRAINING SYSTEM ACCEPTANCE MAINTENANCE... 7

C B. Models. Models 2MOPP

X Series Data Sheet 375, 500 Watt AC-DC and DC-DC DIN-Rail Converters

TECHNICAL REQUIREMENTS FOR ELECTROMAGNETIC DISTURBANCES EMITTED FROM LIGHTING EQUIPMENT INSTALLED IN TELECOMMUNICATION CENTERS

Relion 605 series Self-Powered Feeder Protection REJ603 Product Guide

TEST SUMMARY Seite 2 von 27. Prüfbericht - Nr.: Test Report No HARMONICS ON AC MAINS RESULT: Passed

Power Quality and Reliablity Centre

Avionics Indirect Lightning

Surge Mitigation Component Overview

KOLLMORGEN. Motion Technologies Group. EMC Installation and Application Guidelines for BDS4/5 Goldline Series MB4000H Issue 3

ALP-UDR Universal Detection Relay user-defined protection, automation, and monitoring

Discipline Electrical Testing Issue Date Certificate Number T-2837 Valid Until Last Amended on - Page 1 of 6 LOCATION 1

1587/1577. Insulation Multimeters. Technical Data. Two powerful tools in one.

OLU131P701U1A PSOV131701A(D) PSOV131701B(D) PSOV131701D(D) (for SAMSUNG) PSOV131701E(D) (for SAMSUNG) SOLUM CO., LTD

Note1: Efficiency is tested at nominal input and full load at +25 C ambient Note2: Max Cap Load is tested at nominal input and full resisitive load

704-15K36P RTCA DO-160G COMPLIANT PRODUCT HIGH POWER AIRCRAFT/ VEHICLE DC BUS VOLTAGE SUPPRESSOR DESCRIPTION 704 PACKAGE FEATURES APPLICATIONS

Output Voltage Range ma ma 700 ma 176~305 Vac 160~357 Vdc 250W 93% 0.98 EBD-255S105DV

Series Rev 1.2 Dec. 2015

ABB 1. Multifunctional three-phase monitoring relays. CM-MPS.11, CM-MPS-21, CM-MPS.31 and CM-MPS.41 Data sheet. Features. Approvals. Marks.

MIT510/2, MIT520/2 and MIT1020/2

Accr. 002-Test EN-ISO (2005)

OCS 500N6F SERIES SIMULATOR FOR FAST AND SLOW DAMPED OSCILLATORY WAVES AND RINGWAVE

DC/DC power module 5V / 3A / 15W

PD30CNB25xxPS. Photoelectrics, Background suppression reflective - PointSpot. Main features. Main functions. Description

TEST SUMMARY. Prüfbericht - Nr.: Test Report No.: Seite 2 von 25. Page 2 of 25

By order of American Power Conversion Holdings Inc. at New Taipei City, Taiwan

2520 Pulsed Laser Diode Test System

1587/1577 Insulation Multimeters

Transcription:

Component Tests Insulation Test System Brief Overview of Phenomena............... 2 Applicable Standards................... 3 Test System Overview.................. 3 Generator Specifications................. 5 Accessories and Options................. 9 Software....................... 10 Component Tests: Insulation Test System 1

Brief Overview of Phenomena General Insulation plays a great role in electro-technology, as well as electric power generation, power distribution and loads. Insulation can be stressed by a number of factors. Slowly increasing over voltages can be generated by transmission line switching, turning off heavy loads, switching of capacitive or inductive currents or by lightning strikes on power lines. The disturbance time is in the order of ms. Fast increasing over voltages are generated by switching actions related to failures and lightning in power lines and switch yards. Disturbance times are in the order of µs. Also, lightning in the vicinity of transmission power lines can induce over voltages. Very fast increasing over voltages are generated by switching or failures in gas insulated stations or substations (GIS). The very fast switching occurs as a result of high field levels before breakdown occurs in the gas (ns). Despite the fast switching phenomena, gas insulation has many advantages over conventional methods. The main advantage is that voltage withstand capability does not change with time, whereas oil/paper or solid insulators degrade with time or can even be destroyed by partial discharges. Insulation tests are to ensure that there is no breakdown or repeated flash over. Several test methods can be employed. A.C., D.C. and peak impulse are alternative test methods. It is sufficient that the equipment passes any one of the three. Impulse tests have the advantage of reduced power dissipation in the components and protection devices can be left in circuit. EMC Partner voltage Insulation Test Systems are applied to ensure the safe operation of transformers, cables, switches and a multitude of other products. The 1.2/50 µs voltage wave is used for impulse insulation tests. Impulse testing is performed on components and assemblies either during production or post installation as part of a routine maintenance regime. Accurate monitoring and assessment of the impulse wave shape is an essential part of the test as voltage breakdown is indicated by a change in the waveshape. 1.2/50us voltage impulse Impulse transients can be generated by external events such as lightning or internal events like switching. The result is a voltage impulse with low energy content that appears between windings in transformers, between power lines or across physical gaps between conductors and cases etc. A normalised voltage impulse of 1.2/50us is defined in the standards IEC 60060-1 and IEC61180-1. 1.2/50us voltage & 8/20us current impulses Surge events can be generated by lightning phenomena, switching transients or the activation of protection devices in the power distribution system. A surge itself is influenced by the propogation path taken so that impulses from the same event may have different forms depending upon where a measurement is taken. Combination Wave Generators (CWG) simulate a surge event in power lines close to or within buildings they are energy rich and should breakdown occur, an impulse current will be available that is an additional stress. 250/2500us switching impulse Internal overvoltages as a result of breakdown or switching operations in high voltage networks generate impulses with a long time to peak and much longer duration than lightning events. Pulses to simulate these switching operations are characterized with a risetime of 250us and a duration of 2500us. Typically these long pulses are used in transformer testing. 2 Component Tests: Insulation Test System

Applicable Standards International Electrotechnical Committee (IEC) IEC 61000-4-5 Ed2 (2005) Electromagnetic compatibility (EMC) Part 4-5: Testing and measurement techniques - Surge immunity test IEC 61180-1 Ed1 (1992) High-voltage test techniques for low voltage equipment Part 1: Definitions, test and procedure requirements IEC 60060-1 Ed2 (1989) High-voltage test techniques. Part 1: General definitions and test requirements IEC60335 IEC 61730-2 Ed1 (2004) Photovoltaic (PV) module safety qualification Part 2: Requirements for testing IEC 60664-1 Ed2 (2007) Insulation coordination for equipment within low-voltage systems Part 1: Principles, requirements and tests IEC 60065 Ed7.1 (2005) Audio, video and similar electronic apparatus Safety requirements IEC 61010-1 Ed2 (2002) Safety Requirements for Electrical Equipment for Measurement, Control and Development use. Part 1 - General Requirements Test System Overview Test System Features The Insulation Test System has many unique and outstanding features: - Insulation test 40Ω - Insulation test 500Ω - Insulation test 0.5Joule - Combination Wave Tests - Voltages from 250V up to 96kV - Integrated personal safety - Integrated measurement - Voltage ramp features - Electronic polarity change - Semiconductor switches - Compact designs - High degree of automation without software - 2 year warranty Component Tests: Insulation Test System 3

User Benefits The technical excellence and many unique features of EMC PARTNERS Insulation testers translate directly into benefits for the user. - Range of voltage / impedance combinations to meet multiple Applications - Standard control unit, reduces user training - Measurement system delivers information about insulation breakdown - Integration into existing test facilities saves engineering costs. - Pass / Fail indication for individual samples, speeds up production - High degree of automation, reduces operator workload - Save operator time with the automated test routines and test report facility. - Unparalleled reliability and system up-time Generators A range of generators are available to cover dielectric withstand tests up to 144kV. The generators all employ a patented EMC PARTNER voltage module that enables the same topology to be used for a generator capable of 6kV or 144kV. The same control unit and software is common across the whole range. For all generators, the most significant test parameters can be programmed as fixed values on the instrument front panel or using one of the EMC PARTNER software packages. - MIG0603 Impulse voltage tester 1.2/50µs voltage range 0.5kV up to 6kV. Series resistor 40ohm - MIG1203 and MIG1203SOLAR Impulse voltage tester 1.2/50µs voltage range 1kV up to 12kV. Impedance 40ohm - MIG1203S250 Compact voltage Impulse tester with 1.2/50µs, 250/2500µs and 1/1000µs. Maximum 12kV. Impedance 40 and 500ohm. Photovoltaic modules - MIG1803 Impulse voltage tester 1.2/50µs voltage range 1kV up to 18kV. Series resistor 40ohm - MIG1803-12 Impulse voltage tester 1.2/50µs voltage range 1kV up to 18kV. Series resistor 500ohm. Combination wave 0.25kV up to 18kV. Virtual impedance 12ohm - MIG2403 Impulse voltage tester 1.2/50µs voltage range 2kV up to 24kV. Series resistor 40ohm - MIG3603C Impulse voltage tester 1.2/50µs voltage range 2kV up to 36kV. Series resistor 500ohm. Impulse voltage tester 1.2/50µs voltage range 0.25kV up to 6kV. Series resistor 500ohm. Impulse voltage tester 1.2/50µs voltage range 2kV up to 36kV. Source impedance 12ohm. Combination wave 0.25kV up to 6kV. 0.125kA up to 3kA. Virtual impedance 2ohm - MIG4803 Impulse voltage tester 1.2/50µs voltage range 2kV up to 48kV. Source impedance 50ohm. Combination wave 0.25kV up to 6kV. 0.125kA up to 3kA. Virtual impedance 2ohm - MIG14403 [where flexibility meets accuracy] Insulation tester 1.2/50µs with two source impedances: 48 Ohm and 500 Ohm. Up to 144kV, max 3000A, current waveform approx. 8/20µs. 4 Component Tests: Insulation Test System

Generator Specifications MIG0603 0.25 up to 6kV 10µF 40 Ω 200 Joule MIG0603 MIG1203 0.5 up to 12kV 5µF 40 Ω 420 Joule MIG1203 MIG1203SOLAR 0.5 up to 12kV Selectable Load Ranges 10-183nF ( 7 ranges ) 2.5µF 40 Ω 90 Joule MIG1203SOLAR MIG1203S250, 250/2500µs, 1/1000µs 0.5 up to 12kV 40 Ω and 470 Ω 5µF 420 Joule Pulse front time (1) 1.2 µs Pulse duration (1) 50 µs Pulse front time (2) 250 µs Pulse duration (2) 2500 µs Pulse front time (3) 1 µs Pulse duration (3) 1000 µs MIG1203S250 Component Tests: Insulation Test System 5

MIG1803 MIG1803 0.75 up to 18kV 3.33µF 40 Ω 630 Joule MIG1803-12 MIG1803-12 0.2 up to 18kV 1.67µF 12Ω, 500Ω 280 Joule Impulse Current 8/20µs 0.2 up to 18kV 1.67µF 12Ω 280 Joule Pulse front time 8 µs Pulse duration 20 µs MIG2403 1 up to 24kV 2.5µF 40 Ω 840 Joule MIG2403 MIG3603C 2 up to 36kV 1.67µF 1110 Joule 12 Ω 6 Component Tests: Insulation Test System

Short circuit waveshape 8/20 µs Combination Wave 1.2/50µs (8/20µs) 0.25 up to 6kV 10µF 2 Ω 180 Joule Short circuit waveshape 8/20 µs 0.25 up to 6kV / 2 up to 36kV 500 Ω Short circuit waveshape 8/20 µs MIG3603C MIG4803 2 up to 48kV 10µF 50 Ω Combination Wave 1.2/50µs (8/20µs) 0.25 up to 6kV 10µF 2 Ω 180 Joule Short circuit waveshape 8/20 µs 0.25 up to 6kV 10µF 50 Ω 180 Joule MIG4803 Component Tests: Insulation Test System 7

MIG14403 System 24kV stages starting from 72kV (MIG7203) up to 144kV 2.2uF 48 Ω and 500 Ω 650 Joule MIG14403 MIG14403 4 up to 144kV 4 digit plus comma 375 nf 48 Ω and 500 Ω 3900 Joule positive, negative alternating up to 2 stages (48kV) MIG7203 (72kV), MIG9603 (96kV) and MIG12003 (120kV) are upgradable with STAGE2403 up to 144kV STAGE2403 System Flowcharts Insulation Testing TEMA TEST MANAGER customized solution DSO Option control TEK DSO MIG0603 MIG1203 MIG1803 MIG1803-12 MIG2403 MIG3603C MIG4803 MIG14403 CN-MIG24 TC-MIG24 EUT 8 Component Tests: Insulation Test System

Insulation Testing Applications TEMA TEST MANAGER customized solution control TEK DSO MIG1203 MIGOPTION STEP MIG2403 MIGOPTION UIC EUT Power AC & DC NW-IEC60065-1 NW-IEC60065-1A CN-MIG18AMP MIG-COMAT36 EUT Accessories and Options TC-MIG24 A test cabinet for EUT with maximum dimensions 12 X 15 x 28cm. Can be used together with MIG 1206, MIG1809 and MIG2412. TC-MIG24 is linked to the MIG 1206, MIG1809 and MIG2412 safety circuit. Opening the test cabinet disables test voltages. Safety circuit status is indicated by red and green lamps in the test cabinet. TC-MIG24 CN-MIG-24 Accessory to MIG1206SPD insulation circuit ONLY. Two test pistols with integrated red and green warning lamps. CN-MIG-24 CN-MIG4803 Connection box with 2 connection cables to EUT. Cable length 1.5m one red, one yellow, equipped with alligator clips at the EUT side. Personal safety at user own risk. NW-IEC60065-1 up to 10kV in accordance with IEC / EN 60065. Impulse capacitance 1nF. Requires MIG1203 and MIG OPTION STEP. NW-IEC60065-1A NW-IEC60065-1 up to 10kV in accordance with IEC / EN 60065 >/= Ed.7: without 4 Mohm parallel resistor. Impulse capacitance 1nF. Requires MIG1203 and MIG OP- TION STEP. MIG OPTION UIC to MIG2403 Extends MIG2403 with a coupling network to superimpose the 1.2/50µs impulse to train power supplies: 3000Vd.c.; 1500Vd.c., 1500Va.c. 50Hz, 1000Va.c. 16 2/3Hz in accordance with UIC 550. Component Tests: Insulation Test System 9

COVER BOX SMALL or LARGE Cover box for customized connection of generators with high voltage output on top. SMALL: Dimension 24 x 8 x 8cm. LARGE: Dimension 30 x 23 x 8cm. MIG OPTION STEP OPTION STEP to MIG1203. Must be ordered together with NW-IEC60065-1 or NW- IEC60065-1A. STAGE2403 Extends the generators MIG7203, MIG9603 or MIG12003 up to 144kV. MIG14403 requires 6 stages. Extension of; MIG7303 max. 3 stages, MIG9603 max. 2 stages, MIG12003 max. 1 stage. Generator extention must be made at EMCP. COVER BOX SMALL STAGE2403 COVER BOX LARGE Software For remote control of Insulation Testers, an OPTICAL LINK and one of the following software packages is needed: - GENECS-MIG: This is a relatively simple program that reproduces generator front panel functions on a PC. In addition to remote programming and control of the generators, test report information is available to word processing or other evaluation programs such as EXCEL. - TEMA Software: Comfortable control of EMC PARTNER generators from a PC. Enables up to four generator types to be included in the same test sequence. Generates an enhanced test report. GENECS MIG TEMA software 10 Component Tests: Insulation Test System

EMC PARTNER s Product Range The Largest Range of Impulse Test Equipment up to 100kA and 100kV. Immunity Tests Transient Test Systems for all EMC tests on electronic equipment. ESD, EFT, surge, AC dips, AC magnetic field, surge magnetic field, common mode, damped oscillatory and DC dips. According to IEC and EN 61000-4-2, -4, -5, -8, -9, -10, -11, -12, -16, -18, -19, -29. Lightning Tests Impulse test equipment and accessories for aircraft, military and telecom applications. Complete solutions for RTCA / DO-160 and EUROCAE / ED-14 for indirect lighting on aircraft systems, MIL-STD-461 tests CS106, CS115, CS116 and Telecom, ITU-T.K44 basic and enhanced tests for impulse, power contact and power induction. Component Tests Impulse generators for testing; varistors, gas discharge tubes (GDT), surge protective devices (SPDs), X / Y capacitors, circuit breakers, watt-hour meters, protection relays, insulation material, suppressor diodes, connectors, chokes, fuses, resistors, emc-gaskets, cables, etc. Emission Measurements Measurement of Harmonics and Flicker in 1-phase and 3-phase electrical and electronic products according to IEC /EN 61000-3-2 and 61000-3-3. HARCS Immunity software adds interharmonic tests, voltage variation and ripple on DC tests according to IEC/EN 61000-4-13, -4-14, -4-17. System Automation A full range of accessories enhance the test systems. Test cabinets, test pistols, adapters and remote control software, simplify interfacing with the EUT. Programmable PSU, EMC hardened for frequencies form 16.7Hz to 400Hz. Frequency PS3-SOFT-EXT complies with IEC / EN 61000-4-14 and -4-28. Service Our committment starts with a quality management system backing up our ISO 17025 accreditation. With the SCS number 129, EMC PARTNER provide accredited calibration and repairs. Our customer support team are at your service! Component Tests: Insulation Test System 11

For further information please do not hesitate to contact EMC PARTNER s representative in your region. You will find a complete list of our representatives and a lot of other useful information on our website: www.emc-partner.com The Headquarters in Switzerland EMC PARTNER AG Baselstrasse 160 CH - 4242 Laufen Switzerland Phone: +41 61 775 20 30 Fax: +41 61 775 20 59 Email: sales@emc-partner.ch Web-Site: www.emc-partner.com Your local representative 12 Component Tests: Insulation Test System Version July 2013. Subject to change without notice.