ESSENTIAL INSTRUCTIONS Read this page before proceeding!

Similar documents
LIQ_MAN_ABR_5081T-HT/Rev. L January 2015

LIQ_MAN_ABR_5081P-FF-FI/Rev. D January 2015

2-WIRE UNIVERSAL TEMPERATURE TRANSMITTER

FM Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers

CSA Hazardous Area Approvals Fisher FIELDVUE DVC6005 Series Remote Mount Digital Valve Controllers

CSA Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers

TWO-WIRE TRANSMITTERS FOR ph, ORP, CONDUCTIVITY, OXYGEN, OZONE, AND CHLORINE

CERTIFICATE OF COMPLIANCE

TwO-wIRE TRANSmITTERS FOR ph, ORP, CONDUCTIvITy, OxygEN, OzONE, AND CHLORINE

Remote Controlled FOUNDATION Fieldbus Two-Wire ph/orp Transmitter

Loop Powered Indicator

IECEx Hazardous Area Approvals Fisher FIELDVUE DVC6200 Series Digital Valve Controllers

Instruction Manual PN P/rev.H November Model 5081-P. Two-Wire ph/orp Transmitter

IP26 Series Current to Pneumatic Converter

Intrinsic Safety Control Drawing

Instruction Manual PN 51-Xmt-T/rev.C February Model Solu Comp Xmt-T Two-Wire Toroidal Conductivity Transmitter

Instruction Manual PN A-HT/rev.L January Model 5081-A-HT. HART Two-Wire Chlorine, Dissolved Oxygen, and Ozone Transmitter

Specifications. Specifications. Transfer Characteristics Accuracy at 20 C (68 F) Temperature Drift

Instruction Manual LIQ_MAN_5081A-FF/rev.M January A. FouNdAtIoN Fieldbus two-wire Chlorine, dissolved oxygen, and ozone transmitter

MCR-FL-TS-LP-I-EX SIL IEC Programmable Intrinsically Safe Loop-Powered Temperature Measuring Transducer With HART Protocol

Thie. Automation Products Rev. Chang Date Name Supersedes Dwg. : Part Class:

AH71, AH72, and AH73 FM Approved Humidity Transmitter Assemblies Installation and Operating Instructions

KENCO Loop Powered Magnetostrictive Transmitter. KMD SERIES Operation and Installation Manual. Table of Contents

Submersion/Insertion Toroidal Sensor

HT880 and HT885 FM Approved Humidity Transmitter Assemblies Installation and Operating Instructions

STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C

Product manual wire programmable RTD transmitter

STT 3000 Series STT170 SMART TEMPERATURE TRANSMITTER Models STT171, STT173, STT17H, STT17F, STT17C

D1035S INSTRUCTION MANUAL. Frequency-Pulse Isolating Repeater DIN Rail Model D1035S. D Frequency-Pulse Isolating Repeater

Instruction Manual PN 51-Xmt-C/rev.B October Model Solu Comp Xmt-C Two-Wire Contacting Conductivity Transmitter

Fisher 646 Electro Pneumatic Transducer

Instruction Manual PN 51-Xmt-P/rev.C February Model Solu Comp Xmt-P ph, ORP, and Redox Transmitter

MODEL: B6U-B B6-UNIT. [2] LCD DISPLAY 0: Without 1: With

Features and Applications

QN5O, QXBO, QNBO, QX7O, QN7O, QXCO, QNCO

D1061S INSTRUCTION MANUAL. RS422/RS485 Fieldbus Isolating Repeater Din-Rail Model D1061S. D RS422 / RS485 Fieldbus Isolating Repeater

1-830 SERIES DISPLACEMENT TRANSMITTER

Power supply. Calibration. Surface finishing. Seal. Process connection

INSTRUCTION MANUAL. SIL 3 - SIL 2 Digital Output Loop/Bus Powered DIN-Rail Models D1040Q, D1041Q, D1042Q, D1043Q, D1040Q - D1041Q - D1042Q - D1043Q

Certificate of Compliance

Instruction Manual PN A-HT/rev.C July Model 5081-A-HT. HART Smart Two-Wire Chlorine, Dissolved Oxygen, and Ozone Transmitter

JM Enclosure Dimensions (in mm)

[ Rosemount 648 Wireless Temperature Transmitter. Rosemount 648 Wireless. Quick Installation Guide , Rev CA August 2011

1066 Liquid Analytical Fieldbus ph/orp Transmitter Transmitter Instruction Manual LIQ_MAN_1066-P-FF September 2013

Industrial Pressure Transducers

White Paper. Requirements for fieldbus equipment installed in a Zone 2 and Division 2 hazardous area environment. March 24, 2005

Universal Intelligent 2-wire DIN Rail Transmitters

Installation Instructions P/N MMI , Rev. AA July ATEX Installation Instructions for Micro Motion Model 3500 Transmitters

Model LD-LPIX EXPLOSION-PROOF LOOP POWERED INDICATOR

DPharp Differential Pressure and Pressure Transmitters

ATEX Installation Drawings and Instructions

Rosemount 848L Logic Transmitter with FOUNDATION Fieldbus

INSTRUCTION MANUAL. SIL 3 - SIL 2 Digital Output Loop/Bus Powered DIN-Rail Models D1040Q, D1041Q, D1042Q, D1043Q, D1040Q - D1041Q - D1042Q - D1043Q

ABB MEASUREMENT & ANALYTICS DATA SHEET. TB82 2-wire conductivity transmitters

Level Lance Model 5100

INSTRUCTION MANUAL. SIL 3 Switch/Proximity Detector Interface DIN-Rail Models D1034S, D1034D D1034S - D1034D

Universal Intelligent 2-wire DIN Rail Transmitters

TUpH 396RVP and 398RVP Retractable ph/orp Sensors

Rosemount 148 Temperature Transmitter

Fisher LCP100 Local Control Panel

Fisher 646 Electro-Pneumatic Transducers

Rosemount 148 Temperature Transmitter

EN Series Installation Manual

Sarasota CM200 Sarasota HME900 Density Converter Electronics

Wireless Transmitter for Contacting Conductivity

Temperature Input Module for Zone 1 Series 9482/32

600T Series Transmitters Model 653L Electronic temperature transmitter (PC programmable for RTD and resistance) Specification Sheet

Transducer transmitter BILT 4 II 2(1) G. Technical Manual

MEAS CAL ALARM CONF APT2000 TC

Rev ECN Description Approval Date AA Release to Approvals RCS 10/9/12

Universal Intelligent 2-wire In-head Transmitters

Mobrey Ultrasonic. MSP422, MSP400RH, and MSP900GH Level Transmitters. Product Data Sheet February 2015 IP2045, Rev EA

User's Manual Model SE14 Magnetic Flow Converter

MODEL: B3HU. Space-saving Two-wire Signal Conditioners B3-UNIT

FLEX Ex Power Supplies

ControlWave INTRINSICALLY SAFE PROXIMITY SENSOR INTF. BOARD

INTRINSICALLY SAFE DUPLEXER

Rosemount 148 Temperature Transmitter

Expert 1400 / Expert 3400 Submersible Hydrostatic Level Transmitters

ATEX Installation Instructions for Micro Motion CMF400 Sensors with Booster Amplifier

R88000 T (including ATEX)

Mid-West. Instrument. Model 140 Electrical Installation and Operating Instructions. Gauge Front ELECTRICAL

Rosemount 248 Wireless Temperature Transmitter

General Specifications

FICHA TÉCNICA DE PRODUTO

FM Control Drawing Class I Division 2 / Class I Zone 2. Compact version PROMAG 10

Rosemount 148 Temperature Transmitter

D1061S INSTRUCTION MANUAL. RS422/RS485 Fieldbus Isolating Repeater Din-Rail Model D1061S. D RS422 / RS485 Fieldbus Isolating Repeater

2-Wire Toroidal Conductivity Transmitters Series APT2000 Specifications

Digital Input Output Module for Zone 2 Series 9470/33

TEK-TEMP 2100A. Explosion-Proof Temperature Transmitter. TEMPERATURE. Technology Solutions

Analog Input Module HART Ex i / I.S. Inputs, Channels Type 9461/

INTRINSIC SAFETY BASIC PRINCIPLES

600T Series Transmitters. Specification Sheet. Model 653T Electronic temperature transmitter (PC programmable)

User s Manual. Manual Change No V-E. Please use the attached sheets for the pages listed below in IM 01F06F00-01EN (7th).

DOC. NO. : C3200-E04A APPROVED. Smart PressureTransmitter for Gauge and Absolute Pressure Measurement MODEL APT3200

Instruction Manual PN A-FF/rev.H March Model 5081-A FOUNDATION Fieldbus Two-Wire Chlorine, Dissolved Oxygen, and Ozone Transmitter

09746E00. ATEX / IECEx NEC 505 NEC 506 NEC 500 Class I Class I Class II Class III Zone Zone Division Ex i

User manual. Load cell with one built in amplifier KOSD-FA KIMD-FA KEND-FA Load cell with two built in amplifiers KOSD-FAD KIMD-FAD KEND-FAD

Operating Manual. isolate501

Transcription:

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 Remote Controlled HART SMART Two-Wire Conductivity Transmitter ESSENTIAL INSTRUCTIONS Read this page before proceeding! Your purchase from Rosemount Analytical, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in operating and maintaining the instrument. To ensure their continued operation to the design specifications, personnel should read this manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired. Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty invalidation. Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual. For clarification of instructions, contact your Rosemount representative. Follow all warnings, cautions, and instructions marked on and supplied with the product. Use only qualified personnel to install, operate, update, program and maintain the product. Educate your personnel in the proper installation, operation, and maintenance of the product. Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual. Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process. All equipment doors must be closed and protective covers must be in place unless qualified personnel are performing maintenance. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired. WARNING Substitution of components may impair Intrinsic Safety or suitability for division 2. WARNING do not remove or replace while circuit is live unless area is known to be non-hazardous. WARNING explosion Hazard - do not disconnect equipment unless area is known to be non-hazardous. WARNING To prevent ignition of flammable or combustible atmospheres, disconnect power before servicing or understand and adhere to the manufacturer's live maintenance procedures. SpeCIAl CoNdITIoN FoR INTRINSICAlly SAFe USe The 5081 enclosure is made of aluminum alloy and given a protective polyurethane finish; however, care must be taken to protect it from impact, abrasion or friction with other metal surfaces if located in zone 0. WARNING If a Model 475 Universal Hart Communicator is used with these transmitters, the software within the Model 475 may require modification. If a software modification is required, please contact your local Fisher-Rosemount Service Group or National Response Center at 1-800-654-7768.

January 2015 Specifications physical SpeCIFICATIoNS Housing: Cast aluminum with epoxy coating. Type 4X (IP65). Neoprene O-ring cover seals. 160.5 mm x 175.3 mm x 161.3 mm (6.3 in. x 6.9 in. x 6.4 in.) diameter: 155.4 mm (6.1 in.) electrical Conduit openings: 3/4 in. FNPT power Supply and load: A minimum loop resistance (load) of 250 ohms and a minimum power supply of 18 volts DC is required for digital communication as shown in the load/power supply graph below. local Readout: Main Display is 4 digits, 20 mm tall (0.8 in.) Message Display is ten digits, 7 mm tall (0.3 in.) Automatic Temperture Compensation 3-wire Pt100 RTD Conductivity: 0 to 200 C (32 to 392 F) % concentration: 0 to 100 C (32 to 212 F) Low Conductivity: 0 to 100 C (32 to 212 F) Ambient temperature: -20 to 65 C (-4 to 149 F) Relative Humidity: 0-95% with enclosure sealed. Ce: emi/rfi certified: EN61326 HAzARdoUS AReA ClASSIFICATIoN: Intrinsic Safety: Class I, II, III, Div. 1 Groups A-G T4 Tamb = 70 C, = 62 C (-FF) Exia Entity Class I, Groups A-D Class II, Groups E-G Class III T4 Tamb = 70 C IECEx BAS 09.0159X Ex ia IIC T4 Ga ATEX 0600 II 1 G Baseefa03ATEX0099 EEx ia IIC T4 Tamb = -20 C to +65 C Non-Incendive: Class I, Div. 2, Groups A-D Dust Ignition Proof Class II & III, Div. 1, Groups E-G Type 4X Enclosure Class I, Div. 2, Groups A-D Suitable for Class II, Div. 2, Groups E-G T4 Tamb = 70 C explosion-proof: Class I, Div. 1, Groups B-D Class II, Div. 1, Groups E-G Class III, Div. 1 Class I, Groups B-D Class II, Groups E-G Class III Tamb = 65 C max Instruction Sheet LIQ_MAN_ABR_/Rev.G TRANSMITTeR SpeCIFICATIoNS @ 25 C Measured Range*: 0-20,000 µs/cm Accuracy: ± 0.5% of reading and ± 0.001 µs/cm Repeatability: ± 0.25% of reading Stability: 0.25% of output range/month, non-cumulative Ambient Temperature Coefficient: ± 0.05% of reading/ C Temperature Slope Adjustment: 0-5%/ C Other temperature compensation algorithms: ultra-pure water compensation, cation conductivity, or raw (uncompensated) conductivity. Compatible RTd: 100W or 1000W with Automatic Recognition loop SpeCIFICATIoNS loop Accuracy: Under controlled laboratory conditions at 25 C (77 F) with perfectly calibrated ENDURANCE sensor of appropriate cell constant: up to 5,000 µs/cm: ± 1.0% and ± 2 least significant digit from 5,000 to 20,000: ± 2% of reading and ± 2 least significant digit SeNSoR SeleCTIoN GUIdelINeS Cell Constant Suggested Conductivity Range 0.01/cm up to 50 µs/cm 0.1/cm 1.0 to 500 µs/cm 1.0/cm 10 to 20,000 µs/cm Note: The conductivity values shown in the above chart are for UNCOMPENSATED (or RAW) conductivity at 25 C. Maximum range values will vary due to temperature compensation selection, process temperature, and other process conditions. 2

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 INFRARed ReMoTe CoNTRolleR 1. Pressing a menu key allows the user access to calibrate, program, or diagnostic menus. 2. Press ENTER to store data and settings. Press NEXT to move from one submenu to the next. Press EXIT to leave without storing changes. 3. Use the editing keys to scroll through lists of allowed settings or to change a numerical setting to the desired value. 4. Pressing HOLD puts the transmitter in hold and sends the output current to a pre-programmed value. Pressing RESET causes the transmitter to abandon the present operation and return to the main display. FIGURe 1. Infrared remote controller ReSeT - Press RESET to end the current operation and return to the main display. Changes will NOT be saved. ReSeT does NoT return the transmitter to factory default settings. ARRoW KeyS - Use é and ê keys to increase or decrease a number or to scroll through items in a list. Use the ç or è keys to move the cursor across a number. A flashing word or numeral shows the position of the cursor. Hold - Press HOLD to access the prompts used for turning on or off the hold function. enter - Press ENTER to move from a submenu to the first prompt under the submenu. Pressing ENTER also stores changes in memory and advances to the next prompt. NeXT - Press NEXT to advance to the next submenu or to leave a message screen. CAl - Press CAL to access the calibration menu. prog - Press PROG to access the program menu. exit - Press EXIT to end the current operation. Changes are NOT saved. diag - Press DIAG to read diagnostic messages. 3

January 2015 Instruction Sheet LIQ_MAN_ABR_/Rev.G FIGURe 2. Menu Tree CALIbrAtE SEnSOr 0 CAlIbRATIoN temp AdJ Mtr CAL CELL temp OUtPUt process display Process Display Screen program OutPut temp display HArt SEtUP default AbS C µs/cm 5000 25.0C 12.00mA CAL key PROG key DIAG key HOLD key diagnostics 0 air CELL COnSt tslope 1 CAL F 5081-C-Ht SoFt HArd 1 Not available with display in Neutral Salt, Cation, or Resistivity modes. FaULTs program MeNU MNeMoNICS 4 OUtPUt Current output menu header 4MA 4mA current output (setpoint) 20MA 20mA current output (setpoint) HoLd Current output on hold FAULt Fault condition current output setting dpn Current output dampening time test Current output test value temp Temperature menu header tauto Automatic temperature compensation tman Manual temperature compensation input display Display menu header typ Conductivity measurement type temp C / F toggle selection OUtPUt Current (ma) or percent of full scale display COdE Security code

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 ASSeT MANAGeMeNT SolUTIoNS (AMS) Rosemount Analytical AMS windows provide access to all transmitter measurement and configuration variables. The user can read raw data, final data, and program settings and can reconfigure the transmitter from anywhere in the plant. The figures below show two of the many configuration and measurement screens available using HART AMS. FIGURe 3. AMS configuration screen using HART FIGURe 4. AMS measurement screen using HART 5

January 2015 Instruction Sheet LIQ_MAN_ABR_/Rev.G TAble 1. default Settings VARIAble NAMe MNeMoNIC FACToRy SeTTINGS CUSToMeR SeTTINGS program Menu output Output 4 ma 4 ma 0 µs 20 ma 20 ma 20 ms Hold hold 21 ma Fault fault 22 ma Dampening dpn 0 samples/second Test test 04.00 ma Temperature temp Auto temperature compensation tauto on Manual temperature tman 25.0 C (overridden by auto) Temperature compensation COMP (Linear or none) LInEAr algorithm display display Measurement type typ (CondUC or RSTVTY or Cust) Temperature ( C or F) temp C Output (ma or %) output Cur Security Code code 000 Custom Curve SEtUP CuSt Reference temperature t ref 25.0 C Calibrate Menu Cell constant CELL Const 1.00 Temperature slope temp slope 2.000 Output Calibration OUtPUt CAL diagnose Menu diagnose (Each segment displays SAMple ReAdINGS the current value in the transmitter.) Absolute conductivity Abs 1000 µs Off Set 0 AIR 0.0 µs Cell constant (Linear mode) CELL COnSt 3.00/cm Calibration factor CAL F 1.000 Temperature slope tslope 2.000 Software version soft A02.09 Hardware version HArd 01 Show fault warnings FaULTS none 6

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 Wiring Wire sensor as shown below in Figure 5. Keep sensor wiring separate from power wiring. For best EMI/RFI protection, use shielded output signal cable in an earthgrounded metal conduit. Refer to the sensor instruction manual for more details. FIGURe 5. Wiring Model Recommended Transmitter Settings* Software Setting process Stream Type Contacting Conductivity in the range 20 to LInEAr 20,000 ms/cm (default) Ultrapure water n SALt up to 20mS/cm Cation Water CAtiOn *Assumes 1.0 cell constant. WIRING THRoUGH A junction box The sensor can be wired to the analyzer through a remote junction box (PN 23550-00). Wire the extension cable and sensor cable point-to-point. Refer to the sensor instruction manual for more details. Factory-terminated (PN 23747-00) and unterminated (PN 9200275) connecting cable are available. The use of factory-terminated cable is strongly recommended. To prepare unterminated cable for use, follow the instructions in the sensor instruction manual. For maximum EMI/RFI protection, the outer braid of the sensor cable should be connected to the outer braided shield of the extension cable. At the instrument, connect the outer braid of the extension cable to earth ground. power WIRING For general purpose areas, wire power as shown in Figure 6. For hazardous areas, please see hazardous area installation drawings. FIGURe 6. power Supply/Current loop Wiring for 7

January 2015 Installation UNpACKING ANd INSpeCTIoN Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify us immediately. RoTATING THe display The Model display can be rotated 90 left or right. Disengage the cover lock and remove the front cover. Remove the three screws holding the PCB stack and gently lift the display board. Do not disengage the ribbon cable between the display board and the CPU board. Rotate the display. The black infrared sensor will be at the top of the display. Instruction Sheet LIQ_MAN_ABR_/Rev.G INSTAllATIoN See Figure 7 1. Although the analyzer is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperatures. 2. Install the analyzer in an area where vibrations and electromagnetic and radio frequency interference are minimized or absent. 3. Keep the analyzer and sensor wiring at least one foot from high voltage conductors. Be sure there is easy access to the analyzer. 4. The conduit connections on the sides of the 5081 housing should be sealed to prevent moisture from entering the enclosure. 5. The transmitter must not be mounted with both conduit openings at the top. FIGURe 7. Mounting Model INCH MILLIMETER 8

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 FIGURe 8. ATeX label Information ROSEMOUNT ANALYTICAL 1180 MODEL 5081-C-HT-73 IECExBAS09.0159X Baseefa03ATEX0099X Ex ia IIC T4 Ga Tamb = -20 C TO +65 C I.S. SUPPLY Ui = 30 VDC Ii = 200 ma Pi = 0.9 W Ci = 0 µf Li = 0 µh SIGNAL INPUT Uo = 6.51 V Io = 400 ma Po = 642 mw Ci = 24.1 µf Li = 0 µh 9241505-00/F II 1 G 9

January 2015 Instruction Sheet LIQ_MAN_ABR_/Rev.G Figure 9. FMRC explosion-proof Installation 10

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 Figure 10. FM Hazardous Area Approval label 11

January 2015 Instruction Sheet LIQ_MAN_ABR_/Rev.G Figure 11. FM Intrinsically Safe Installation CONDUCTIVITY SENSOR (SEE NOTE 4) INFRARED REMOTE CONTROL UNIT (RMT PN 23572-00) FOR USE IN CLASS I AREA ONLY MODEL 5081-C-HT XMTR HAZARDOUS AREA UNCLASSIFIED AREA UNSPECIFIED POWER SUPPLY 30 VDC MAX 24 V TYP 1 2 3 4 16 15 14 13 5 12 6 11 7 8 9 10 ROSEMOUNT MODEL 275/375/475 HART COMMUNICATOR REMOTE TRANMITTER INTERFACE FOR USE IN CLASS I AREA ONLY (SEE NOTE 2 AND TABLE III) IS CLASS I, II, III, DIVISION 1, GROUPS A, B, C, D, E, F, G; NI CLASS I, DIVISION 2, GROUPS A,B,C,D; SUITABLE CLASS II, DIVISION 2, GROUPS F & G; SUITABLE CLASS III, DIVISION 2 SAFETY BARRIER (SEE NOTE 1 AND 7) LOAD SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY OR SUITABILITY FOR DIVISION 2. NON-INCENDIVE FIELD WIRING CONNECTIONS FOR CLASS 1, DIVISION 2, GROUPS ABCD DIVISION 2 HAZARDOUS AREA WARNING- WARNING- TO PREVENT IGNITION OF FLAMMABLE OR COMBUSTIBLE ATMOSPHERES, DISCONNECT POWER BEFORE SERVICING. 1 2 3 4 16 15 14 13 5 12 6 11 7 8 9 10 CONTACTING CONDUCTIVITY SENSOR FM APPROVED DEVICE IN ACCORDANCE WITH ENTITY REQUIREMENTS OR SIMPLE APPARATUS MODEL 5081-C-HT XMTR 13 14 12 13 D2 UNCLASSIFIED AREA GAS GROUPS A, B 12.05 0.5 C 490.05 2.0 9990.05 3.7 D Ca (µf) TABLE I OUTPUT PARAMETERS La (mh) OUTPUT PARAMETERS Vt It Pt TABLE II MODEL 5081-C-HT TB1-1 THRU 12 6.51 Vdc 280 ma 442 mw 7. THE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS WITH ASSOCIATED APPARATUS WHEN THE FOLLOWING IS TRUE: FIELD DEVICE INPUT ASSOCIATED APPARATUS OUTPUT Vmax OR Ui > Voc, Vt OR Uo Imax OR Ii > Isc, It OR Io Pmax OR Pi > Po Ci+ Ccable < Ca, Ct OR Co Li+ Lcable < La, Lt OR Lo 6. RESISTANCE BETWEEN INTRINSICALLY SAFE GROUND AND EARTH GROUND MUST BE LESS THAN 1.0 Ohm. 5. DUST-TIGHT CONDUIT SEAL MUST BE USED WHEN INSTALLED IN CLASS II AND CLASS III ENVIRONMENTS. 4. SENSORS WITHOUT PREAMPS SHALL MEET THE REQUIREMENTS OF SIMPLE APPARATUS AS DEFINED IN ANSI/ISA RP12.6 AND THE NEC, ANSI/NFPA 70. THEY CAN NOT GENERATE NOR STORE MORE THAN 1.2V, 0.1A, 25mW, AND 20µJ. SEE TABLES I AND II. 3. INSTALLATION SHOULD BE IN ACCORDANCE WITH ANSI/ISA RP12.06.01 "INSTALLATION OF INTRINSICALLY SAFE SYSTEMS FOR HAZARDOUS (CLASSIFIED) LOCATIONS" AND THE NATIONAL ELECTRICAL CODE (ANSI/NFPA 70). 2. INTRINSICALLY SAFE APPARATUS (MODEL 5081-C-HT, IRC TRANSMITTER AND MODEL 275/375/475) AND ASSOCIATED APPARATUS (SAFETY BARRIER) SHALL MEET THE FOLLOWING REQUIREMENTS: THE VOLTAGE (Vmax) AND CURRENT (Imax) OF THE INTRINSICALLY SAFE APPARATUS MUST BE EQUAL TO OR GREATER THAN THE VOLTAGE (Voc OR Vt) AND CURRENT (Isc OR It) WHICH CAN BE DELIVERED BY THE ASSOCIATED APPARATUS (SAFETY BARRIER). IN ADDITION, THE MAXIMUM UNPROTECTED CAPACITANCE (Ci) AND INDUCTANCE (Li) OF THE INTRINSICALLY SAFE APPARATUS, INCLUDING INTERCONNECTING WIRING, MUST BE EQUAL OR LESS THAN THE CAPACITANCE (Ca) AND INDUCTANCE (La) WHICH CAN BE SAFELY CONNECTED TO THE APPARATUS. (REF. TABLES I, II AND III). THE CAPACITANCE AND INDUCTANCE OF THE LOAD CONNECTED TO THE SENSOR TERMINALS MUST NOT EXCEED THE VALUES SPECIFIED IN TABLE I. 1. ANY SHUNT ZENER DIODE SAFETY BARRIER APPROVED BY FM HAVING THE FOLLOWING OUTPUT PARAMETERS: SUPPLY/SIGNAL TERMINALS TB1-15, 16 Voc OR Vt NOT GREATER THAN 30 V Isc OR It NOT GREATER THAN 200 ma Pmax NOT GREATER THAN 0.9 W 13 INSTALATION TO BE IN ACCORDANCE WITH THE NATIONAL ELECTRICAL CODE. 12 DIVISION 2 WIRING METHOD PER THE NEC (EXCLUDING NONINCENDIVE FIELD WIRING). 11. NO REVISION TO DRAWING WITHOUT PRIOR FM APPROVAL. 10. THE ASSOCIATED APPARATUS MUST BE FM APPROVED. D3 9. CONTROL EQUIPMENT CONNECTED TO ASSOCIATED APPARATUS MUST NOT USE OR GENERATE MORE THAN 250 Vrms OR Vdc. 8. ASSOCIATED APPARATUS MANUFACTURER'S INSTALLATION DRAWING MUST BE FOLLOWED WHEN INSTALLING THIS EQUIPMENT. MODEL NO. 5081-C-HT MODEL NO. 275 TABLE III 5081-C-HT ENTITY PARAMETERS SUPPLY / SIGNAL TERMINALS TB1-15, 16 Vmax (Vdc) Imax (ma) Pmax (W) Ci (nf) Li (mh) 30 200 0.9 27.8 0.0 ENTITY PARAMETERS: REMOTE TRANSMITTER INTERFACE Vmax (Vdc) Imax (ma) Pmax (W) 30 300 1.0 Ci (µf) Li (mh).07 0.0 D1 Voc max OUT: Vdc Isc max OUT:mA 1.7 32 375/475 30 200 1.0.00 0.0 1.9 32 14 NON-INCENDIVE FIELD WIRING METHODS MAY BE USED FOR CONNECTING SENSORS TO THE INSTRUMENT. ATTACHED SENSORS MUST BE FM APPROVED AS NON-INCENDIVE FOR CLASS 1, DIVISION 2, GROUPS ABCD WITH INPUT VALUES OF Vmax AND Imax > Voc (Vt) AND Isc (It) LISTED IN TABLE II AND THE Ci AND Li OF THE SENSOR AND INTERCONNECTED WIRING MUST BE < Ca AND La LISTED IN TABLE 1 OR BE CLASSIFIED AS "SIMPLE APPARATUS". SIMPLE APPARATUS ARE DEVICES WHICH ARE INCAPABLE OF GENERATING OR STORING MORE THAN 1.2 V, 0.1 A, 25 mw OR 20 µj (CONTACTING CONDUCTIVITY SENSORS QUALIFY AS SIMPLE APPARATUS). THIS DOCUMENT IS CERTIFIED BY FM REV C REV REV REV REV REV REVISIONS NOT PERMITTED W/O AGENCY APPROVAL 12

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 Figure 12. CSA Hazardous Area Approval label 13

January 2015 Instruction Sheet LIQ_MAN_ABR_/Rev.G Figure 13. CSA Intrinsically Safe Installation 14

Instruction Sheet LIQ_MAN_ABR_/Rev.G January 2015 Engineering Specifications quick-start 1. On the Remote, press PROG, NEXT, NEXT, ENTER. 2. Use the arrow buttons to select Conduc (conductivity), rstvty (resistivity), or CuSt (custom curve) mode. Press ENTER. If you chose ConduC or rstvty, go to step 4. 3. If you selected CuST, you will see the Setup Custom screen. To move to the custom curve configuration menu, press ENTER. You will automatically return to this same Setup Custom screen after configuration is complete. To continue transmitter display programming, press NEXT in the Setup Custom screen. 4. Use the arrow keys to toggle temperature units between Celsius and Farenheit. 5. Press ENTER then RESET. 6. Press PROG, ENTER. 7. Use the arrow buttons to enter the 4 ma value. Press ENTER. 8. Use the arrow buttons to enter the 20 ma value. Press ENTER then RESET. 9. Press PROG, NEXT, ENTER. 10. Use the arrow key to toggle tauto to On or OFF to select using either the process temperature (tauto = On) or a manual temperature (tauto = OFF). Press ENTER. If you selected tauto = OFF, you will be prompted to enter the manual temperature; use the arrow keys, then press ENTER. 11. If you selected CondUC or rstvty in step 2, you will see a CoMp (Temperature Compensation type) screen. Use the arrow keys to select desired temperature compensation: linear (linear), n SAlt (neutral salt), CatIon (cation conductivity), or none (raw or uncompensated conductivity). Press ENTER. If you are in linear mode, you can now enter a particular temperature slope (default is 2%/degC), then press ENTER to apply the slope. 12. Press RESET. 13. Press CAL, NEXT, NEXT, NEXT, ENTER. 14. Use the arrow buttons to enter the cell constant of the sensor. Press ENTER, then EXIT. 15. If you are in rstvty, n SAlt, Cation, or none (raw) mode, quick start is complete, proceed to step 20. 16. If you are in linear mode, press EXIT, CAL, NEXT, ENTER. 17. Hold the sensor in air to zero. Press ENTER, then EXIT. 18. Place the sensor in a solution of known conductivity value. Press CAL, ENTER. 19. Use the arrow buttons to enter the current conductivity value of the solution. Press ENTER. 20. Press RESET. To ReSeT TRANSMITeR To FACToRy default SeTTINGS: 1. Press PROGRAM, NEXT, NEXT, NEXT, NEXT, NEXT. The screen should say default. Press ENTER. 2. Use the arrow keys to toggle between no (retain your configuration and calibration settings) and yes (restore factory default settings to all variables). 3. Press ENTER, then EXIT. 15

LIQ_MAN_ABR_5081C-FF-FI Rev. G January 2015 facebook.com/emersonrosemountanalytical AnalyticExpert.com 8 Credit Cards for U.S. Purchases Only. twitter.com/raihome youtube.com/user/rosemountanalytical emerson process Management 2400 Barranca Parkway Irvine, CA 92606 USA Tel: (949) 757-8500 Fax: (949) 474-7250 rosemountanalytical.com Rosemount Analytical Inc. 2015 2015 Rosemount Analytical, Inc. All rights reserved. The Emerson logo is a trademark and service mark of Emerson Electric Co. Brand name is a mark of one of the Emerson Process Management family of companies. All other marks are the property of their respective owners. The contents of this publication are presented for information purposes only, and while effort has been made to ensure their accuracy, they are not to be construed as warranties or guarantees, express or implied, regarding the products or services described herein or their use or applicability. All sales are governed by our terms and conditions, which are available on request. We reserve the right to modify or improve the designs or specifications of our products at any time without notice.