Biomedical Engineering Evoked Responses

Similar documents
Biomedical Engineering Electrophysiology

Biomedical Instrumentation B2. Dealing with noise

Fundamentals of Pacing Therapy

EDL Group #3 Final Report - Surface Electromyograph System

BME 405 BIOMEDICAL ENGINEERING SENIOR DESIGN 1 Fall 2005 BME Design Mini-Project Project Title

780. Biomedical signal identification and analysis

iworx Sample Lab Experiment AN-2: Compound Action Potentials

CHAPTER 7 INTERFERENCE CANCELLATION IN EMG SIGNAL

Principles of nerve stimulation

BME 3113, Dept. of BME Lecture on Introduction to Biosignal Processing

Electrical noise in the OR

Compound Action Potentials

Ques on (2): [18 Marks] a) Draw the atrial synchronous Pacemaker block diagram and explain its operation. Benha University June 2013

Lab #9: Compound Action Potentials in the Toad Sciatic Nerve

Application Note 175 Using the STMISOC Stimulus Isolator

EE 230 Experiment 10 ECG Measurements Spring 2010

Biomedical Signals. Signals and Images in Medicine Lecture 1 Dr Nabeel Anwar

Hearing and Deafness 2. Ear as a frequency analyzer. Chris Darwin

Perception of pitch. Definitions. Why is pitch important? BSc Audiology/MSc SHS Psychoacoustics wk 4: 7 Feb A. Faulkner.

Imagine the cochlea unrolled

Biomedical Sensor Systems Laboratory. Institute for Neural Engineering Graz University of Technology

Biomedical Instrumentation (BME420 ) Chapter 6: Biopotential Amplifiers John G. Webster 4 th Edition

AUDL GS08/GAV1 Signals, systems, acoustics and the ear. Loudness & Temporal resolution

Florida Atlantic University Biomedical Signal Processing Lab Experiment 2 Signal Transduction: Building an analog Electrocardiogram (ECG)

Complex Sounds. Reading: Yost Ch. 4

Biosignal filtering and artifact rejection. Biosignal processing I, S Autumn 2017

Perception of pitch. Definitions. Why is pitch important? BSc Audiology/MSc SHS Psychoacoustics wk 5: 12 Feb A. Faulkner.

Instrumentation amplifier

EMG. The study of muscle function through the investigation of the electrical signal the muscles produce

AUDL 4007 Auditory Perception. Week 1. The cochlea & auditory nerve: Obligatory stages of auditory processing

EC209 - Improving Signal-To-Noise Ratio (SNR) for Optimizing Repeatable Auditory Brainstem Responses

Electrocardiogram (ECG)

Changing the sampling rate

Special-Purpose Operational Amplifier Circuits

Perception of pitch. Importance of pitch: 2. mother hemp horse. scold. Definitions. Why is pitch important? AUDL4007: 11 Feb A. Faulkner.

Used to overcome ventricular fibrillation may be due to coronary occlusion, shock, or abnormalities in blood chemistry

Physical Therapy. Methods in Physical Therapy. Non-electric heat therapy Therapeutique application of electric current

Signals & Systems for Speech & Hearing. Week 6. Practical spectral analysis. Bandpass filters & filterbanks. Try this out on an old friend

The EarSpring Model for the Loudness Response in Unimpaired Human Hearing

Phase and Feedback in the Nonlinear Brain. Malcolm Slaney (IBM and Stanford) Hiroko Shiraiwa-Terasawa (Stanford) Regaip Sen (Stanford)

describe sound as the transmission of energy via longitudinal pressure waves;

Biomechatronic Systems

Biomechatronic Systems

EXPERIMENT 5 Bioelectric Measurements

Acoustics, signals & systems for audiology. Week 9. Basic Psychoacoustic Phenomena: Temporal resolution

ECG Project. Raphal Blanchet, Axel Boland, Thomas Donnay, Mario Jose Teles Varandas, University of Liege

Psychology in Your Life

Using Rank Order Filters to Decompose the Electromyogram

Acoustics, signals & systems for audiology. Week 4. Signals through Systems

APPLICATION NOTE. Overview

Interface Electronic Circuits

COMMUNICATIONS BIOPHYSICS

RFID Systems: Radio Architecture

BENG 186B Winter 2013 Final

Syllabus Recording Devices

Neurophysiology. The action potential. Why should we care? AP is the elemental until of nervous system communication

Brain-computer Interface Based on Steady-state Visual Evoked Potentials

Kanchan S. Shrikhande. Department of Instrumentation Engineering, Vivekanand Education Society s Institute of.

Limulus eye: a filter cascade. Limulus 9/23/2011. Dynamic Response to Step Increase in Light Intensity

Detection of external stimuli Response to the stimuli Transmission of the response to the brain

F produce muscle activation resembling that which occurs

SOCRATES. Auditory Evoked Potentials

Design on Electrocardiosignal Detection Sensor

Physiological Signal Processing Primer

Biomedical Electrodes, Sensors, and Transducers. Definition of Biomedical Electrodes, Sensors, and Transducers. Electrode: Sensor: Transducer:

ANC: Section 2. Unidirectional Propagation - 1 J Thomas Mortimer & Narendra Bhadra

V Technical Textiles Interconnect PRESENTED BY V TECHNICAL TEXTILES, INC.

*Notebook is excluded

Chapter 2 A Silicon Model of Auditory-Nerve Response

698 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 51, NO. 5, MAY 2004

BIO-ELECTRIC MEASUREMENTS

Psycho-acoustics (Sound characteristics, Masking, and Loudness)

Spectro-Temporal Methods in Primary Auditory Cortex David Klein Didier Depireux Jonathan Simon Shihab Shamma

Results of Egan and Hake using a single sinusoidal masker [reprinted with permission from J. Acoust. Soc. Am. 22, 622 (1950)].

Portable EEG Signal Acquisition System

Biosignal filtering and artifact rejection. Biosignal processing, S Autumn 2012

Analog Circuits and Systems

Product Overview. npi 07/18

Laboratory Activities Handbook

Introduction to Electronic Circuit for Instrumentation

40 Hz Event Related Auditory Potential

Amplitude Modulation Effects in Cardiac Signals

DESIGN OF OTA-C FILTER FOR BIOMEDICAL APPLICATIONS

* Notebook is excluded. Features KL-720 contains nine modules, including Electrocardiogram Measurement, E lectromyogram Measurement,

Applied Electronics II

TNS Journal Club: Efficient coding of natural sounds, Lewicki, Nature Neurosceince, 2002

Analog Circuits and Systems

Basic Electronics Prof. Dr. Chitralekha Mahanta Department of Electronics and Communication Engineering Indian Institute of Technology, Guwahati

Lecture Fundamentals of Data and signals

InstrumentationTools.com

BME 701 Lecture 1. Measurement and Instrumentation

Magnetoencephalography and Auditory Neural Representations

A Comprehensive Model for Power Line Interference in Biopotential Measurements

CHAPTER 3. Instrumentation Amplifier (IA) Background. 3.1 Introduction. 3.2 Instrumentation Amplifier Architecture and Configurations

An optimized layout for multipolar neural recording electrode

Section 6 Chapter 2: Operational Amplifiers

ELECTRICAL CURRENT FOR PAIN CONTROL

Lab E5: Filters and Complex Impedance

Chapter 12. Preview. Objectives The Production of Sound Waves Frequency of Sound Waves The Doppler Effect. Section 1 Sound Waves

DEVELOPMENT OF AN AUDIO EVOKED RESPONSE SYSTEM TO FACILITATE ANAESTHESIA MONITORING

Transcription:

Biomedical Engineering Evoked Responses Dr. rer. nat. Andreas Neubauer andreas.neubauer@medma.uni-heidelberg.de Tel.: 0621 383 5126 Stimulation of biological systems and data acquisition 1. How can biological systems be stimulated? Visually Acoustically Via drugs 2. How to record the response of stimuli? Distinguishing signal from noise Artifacts and responses 3. Computer aided data analysis AI vs. PI Andreas Neubauer I Slide 2 I 15.11.2016 1

Linear Systems Andreas Neubauer I Slide 3 I 15.11.2016 How to describe linear systems Components behave in a proportional manner e. g. electrical systems mechanical systems Consider a system with Input and output Goal: Find a function to describe the system Andreas Neubauer I Slide 4 I 15.11.2016 2

Frequency responses Easiest test for frequency response: function infinitely high and infinitely short Fourier integral of : 1 1 Testing the response of a low pass filter Fourier transform of the input: Andreas Neubauer I Slide 5 I 15.11.2016 Stimuli Andreas Neubauer I Slide 6 I 15.11.2016 3

Neural stimulation I The body is a good conductor Nerves an muscles work electrically neural stimulation with electricity is possible muscle contraction induced by a current between two surface electrodes Andreas Neubauer I Slide 7 I 15.11.2016 Neural stimulation II Apart from an alternating current, one can apply a short pulse between two surface electrodes Pulse duration is between 50 s "# 2 %& to avoid high amplitudes The stimulation occurs under the electrode which sees the pulse as negative pulse the transmembrane potential will be reduced where the current leaves the nerve R: Rheobase C: Chronaxie Andreas Neubauer I Slide 8 I 15.11.2016 4

More about currents and voltages Stimulation of a nerve is achieved if the transmembrane potential is reduced by about 30% Stimulation current must flow for a minimum time generation of an action potential Position of the electrodes Approx. charge required for stimulation Inside a nerve axon 10 ) On the surface of a nerve trunk 10 * ) On the skin several mm from the nerve 10 + ) Andreas Neubauer I Slide 9 I 15.11.2016 Depolarization with an electrode Charge has to be removed from one node of Ranvier depolarization generation of an action potential Membrane capacitance is approx. 10 Applied polarization potential: Approx. 0.1 -./"0 1 2 0%13 10 0.1-10 ) Andreas Neubauer I Slide 10 I 15.11.2016 5

Electrical stimulation a practical example Energy required to stimulate a nerve with electrons placed directly over the nerve: Approx. 10 4 5 corresponds to 10 %6 flowing for 100 s through an electrical resistance of 10 7Ω Energy stored in a battery with an output voltage of 9 - and a capacity of 500 %6/: : ; 500 %6 3600 & 9-16200 5 a small battery is enough to stimulate a nerve more than 10010 + times if the nerve is not superficial a much larger amount of energy may be required several joules for a human heart Andreas Neubauer I Slide 11 I 15.11.2016 Strength duration curves The stimulation energy is not the same for different nerves Large fibers need less energy than small fibers Cutting the innervation of a muscle leads to a slow death of the muscle Denervation Electrical stimuli can be used to stimulate the muscle directly threshold will be increased record the threshold current in dependence of the applied pulse length Rheobase R: Minimum current to stimulate the muscle (independent of pulse length) Chronaxie C: Pulse width with threshold current 2 Rheobase Andreas Neubauer I Slide 12 I 15.11.2016 6

Which other stimuli can be used? Stimuli can be applied to each of our senses mechanically auditory taste/smell (difficult in practice) Andreas Neubauer I Slide 13 I 15.11.2016 visually Andreas Neubauer I Slide 14 I 15.11.2016 7

Auditory stimuli Auditory equivalent of a function? Very short click containing a wide range of frequencies Mostly used in objective hearing tests What if the audiologist wants to know which frequencies have been heard? Hypothesis: Give an auditory stimulus, record the response, apply Fourier transform to the response obtain transfer function of auditory system Unfortunately the brain and the hearing system are nonlinear systems Andreas Neubauer I Slide 15 I 15.11.2016 Auditory stimulus in practice Short burst of single frequency Obtain the response of the ear by testing different frequencies Signal equation: = cos 2@f B Fourier transform leads to: = = B C sin 2@ BF C G sin 2@ BG C 2@ B F C 2@ B G C Andreas Neubauer I Slide 16 I 15.11.2016 8

Visual stimuli Visual equivalent of a function? Flash of light The eye cannot deal with frequencies above 20 H Pattern recognition is much easier with visual stimuli Fourier transformation may be more useful in the spatial than in the temporal domain Neubauer, AADA 2014 Andreas Neubauer I Slide 17 I 15.11.2016 Detour to spatial frequencies Courtesy of M. Ruttorf Neubauer, AADA 2014 Andreas Neubauer I Slide 18 I 15.11.2016 9

Detection of small signals Andreas Neubauer I Slide 19 I 15.11.2016 Bandwidth and signal-to-nose ratios Bandwidth: Frequency range over which the gain remains constant F3 I bandwidth: Frequency range over which the gain is not less than 3 I below the maximum gain Signal Bandwidth [1/&] ECG 0.5F100 EEG Arterial pressure wave 0.5 F75 L) F40 Body temperature L) 1 Respiration L) 10 Electromyograph Nerve action potentials 10 5 10 O 10 10 10 O Smooth muscle potentials 0.05 F 10 Andreas Neubauer I Slide 20 I 15.11.2016 10

Choice of the amplifier Which signal should be amplified? Current? Voltage? Noise will also be amplified! The amplifier will contribute some noise to the signal usually quoted in terms of the equivalent noise at the input, Noise Referred to the Input (RTI) noise (RTI) = (noise at output)/(amplifier gain) RTI can be directly compared with the size of the input signal Example: Measurement of an ECG: 1 %- Required signal to noise ratio: 40 I 100:1 RTI < QR BB 10 - Andreas Neubauer I Slide 21 I 15.11.2016 Input resistance Amplification may be frequency dependent The choice of the input resistance must be related to the recorded signal Source and input resistance form a potential divider If the input resistance is equal to the source resistance the signal will be reduced to half it s value Reducing this error to 1% requires an input resistance 100 times higher than the source resistance => Source resistance must be known! Andreas Neubauer I Slide 22 I 15.11.2016 11

Differential amplifier II In practice almost all bioelectric amplifiers are differential amplifiers ability to reject external signals Use operational amplifiers for the design instrumentation amplifier configuration input voltages Rules for analyzing operational amplifier circuits 1. There will be no voltage between the inputs of an operational amplifier 2. No current will flow into the inputs of an operational amplifier Andreas Neubauer I Slide 23 I 15.11.2016 Differential amplifier II In practice almost all bioelectric amplifiers are differential amplifiers ability to reject external signals Use operational amplifiers for the design instrumentation amplifier configuration input voltages Rules for analyzing operational amplifier circuits 1. There will be no voltage between the inputs of an operational amplifier 2. No current will flow into the inputs of an operational amplifier Andreas Neubauer I Slide 24 I 15.11.2016 12