LIGO PROJECT. Piezo-Electric Actuator Initial Performance Tests. Eric Ponslet April 13, Abstract

Similar documents
Developer Techniques Sessions

Maximizing LPM Accuracy AN 25

Piezomechanik GmbH. PosiCon.an for piezoactuators (low voltage and high voltage actuators) Position Feedback control electronics

PDu150CL Ultra-low Noise 150V Piezo Driver with Strain Gauge Feedback

Non-Contact Capacitance Gauging Instrument & Series 2800 Capacitive Probes

Optimizing Performance Using Slotless Motors. Mark Holcomb, Celera Motion

FFP-TF2 Fiber Fabry-Perot Tunable Filter Technical Reference

Texas Components - Data Sheet. The TX53G1 is an extremely rugged, low distortion, wide dynamic range sensor. suspending Fluid.

Model LIA100. Lock-in Amplifier

PDu150CL Ultra low Noise 150V Piezo Driver with Strain Gauge Feedback

PD32-32 Channel Piezo Driver Manual and Specifications

Shielding. Fig. 6.1: Using a Steel Paint Can

E Charge-controlled amplifier module

TNI mode cleaner/ laser frequency stabilization system

LION PRECISION. TechNote LT February, Capacitive Sensor Operation and Optimization

A SIMPLE METHOD TO COMPARE THE SENSITIVITY OF DIFFERENT AE SENSORS FOR TANK FLOOR TESTING

First Time User Manual

PiezoMike Linear Actuator

Linear Motion Servo Plants: IP01 or IP02. Linear Experiment #0: Integration with WinCon. IP01 and IP02. Student Handout

DRAFT Expected performance of type-bp SAS in bkagra

Contents. CALIBRATION PROCEDURE NI PXIe GHz and 14 GHz RF Vector Signal Analyzer

The VIRGO suspensions

Electronic AC Load ZSAC Series. 400 W up to 21,000 W 260 V up to 440 V 3 A up to 100 A. Rev. 3.03

Tactical grade MEMS accelerometer

Compact Nanopositioning System Family with Long Travel Ranges

Synchronization Control Scheme for Hybrid Linear Actuator Based on One Common Position Sensor with Long Travel Range and Nanometer Resolution

AN5E Application Note

10 Things to Consider when Acquiring a Nanopositioning System

P-611.Z Piezo Z-Stage

Application Note: Precision Displacement Test Stand Rev A

Design Implementation Description for the Digital Frequency Oscillator

EX1000 Series EX1000A EX1000A-TC EX1016A EX1032A EX1048A EX10SC EX1000A-TCDC RELIABLE DATA FIRST TIME EVERY TIME.

Control Servo Design for Inverted Pendulum

Model 176 and 178 DC Amplifiers

Active Vibration Isolation of an Unbalanced Machine Tool Spindle

Contents: Movement & Positioning News M&P 23, 1997, Text Only Version

Large Signal Displacement Measurement with an Asylum SA Atomic Force Microscope Rev B

Wavelength Control and Locking with Sub-MHz Precision

PI piezo Life Time Test Report. A. Bosotti, R. Paparella, F. Puricelli

APPLICATION NOTE POSITION CONTROL OF PIEZO ACTUATORS

Seismic Noise & Vibration Isolation Systems. AIGO Summer Workshop School of Physics, UWA Feb Mar. 2, 2010

Beam Dynamics + Laser Micro Vibrometry 1

queensgate a brand of Elektron Technology

M-041 M-044 Tip/Tilt Stage

Improving the Performance of a Geophone through Capacitive Position Sensing and Feedback. Aaron Barzilai. Stanford University

Aero Support Ltd, 70 Weydon Hill Road, Farnham, Surrey, GU9 8NY, U.K.

S-330 Tip/Tilt Platform

sin(wt) y(t) Exciter Vibrating armature ENME599 1

INDIAN INSTITUTE OF TECHNOLOGY BOMBAY

SC-4/M User manual 4-channel measuring amplifier

Op Amp Booster Designs

Piezo-shakers are covering a different application spectrum than electro-magnetic shakers.

Advanced Test Equipment Rentals ATEC (2832)

A Model Based Digital PI Current Loop Control Design for AMB Actuator Coils Lei Zhu 1, a and Larry Hawkins 2, b

3D Distortion Measurement (DIS)

Getting Started. MSO/DPO Series Oscilloscopes. Basic Concepts

F. Barillot Cedrat Technologies MEFISTO Design & tests of a demonstrator for filet compensation mechanism

A DSP IMPLEMENTED DIGITAL FM MULTIPLEXING SYSTEM

Anthony Chu. Basic Accelerometer types There are two classes of accelerometer in general: AC-response DC-response

DIGITAL FILTERING OF MULTIPLE ANALOG CHANNELS

Analysis of Electrical Noise in Piezoelectric Sensors

BENCHMARK MEDIA SYSTEMS, INC.

FFP-C Fiber Fabry-Perot Controller OPERATING INSTRUCTIONS. Version 1.0 MICRON OPTICS, INC.

Fast Tip/Tilt Platform

ABSTRACT 1 CEBAF UPGRADE CAVITY/CRYOMODULE

ELEC4623 / ELEC9734 BIOMEDICAL ENGINEERING LABORATORY 3: DESIGN, TESTING AND ANALYSIS OF A HIGH QUALITY ISOLATED BIOPOTENTIAL AMPLIFIERS

Four-Channel Differential AC Amplifier

Model 4800 O P E R AT I O N M A N U A L L O A D C E L L S U M M I N G T R A N S M I T T E R

A Prototype Wire Position Monitoring System

Lock-Ins for electrical measurements

Digital LLRF Test on the Renascence Cryomodule

DC MHZ PXI Differential Instrumentation Amplifier

LBI-30398N. MAINTENANCE MANUAL MHz PHASE LOCK LOOP EXCITER 19D423249G1 & G2 DESCRIPTION TABLE OF CONTENTS. Page. DESCRIPTION...

Measurements 2: Network Analysis

5. Transducers Definition and General Concept of Transducer Classification of Transducers

_active vibration isolation desktop unit halcyonics_i4 series

New Technique Accurately Measures Low-Frequency Distortion To <-130 dbc Levels by Xavier Ramus, Applications Engineer, Texas Instruments Incorporated

QUAD PROGRAMMABLE FILTER/AMPLIFIERS For the and Signal Conditioning Systems

Installation, Operation and Maintenance Manual

Angle Encoder Modules

A review of Pound-Drever-Hall laser frequency locking

Signal Conditioning Amplifier

MEC751 Measurement Lab 2 Instrumented Cantilever Beam

Development of Control Algorithm for Ring Laser Gyroscope

Modal Excitation. D. L. Brown University of Cincinnati Structural Dynamics Research Laboratory. M. A. Peres The Modal Shop, Inc Cincinnati, OH

WESTREX RA-1712 PHOTOGRAPHIC SOUND RECORD ELECTRONICS

Operational Amplifiers

ERICSSONZ LBI-30398P. MAINTENANCE MANUAL MHz PHASE LOCKED LOOP EXCITER 19D423249G1 & G2 DESCRIPTION TABLE OF CONTENTS

Digital HF Receiver WJ-8723

Thermal Johnson Noise Generated by a Resistor

Isolator-Free 840-nm Broadband SLEDs for High-Resolution OCT

How Resonance Data is Used

Part 2: Second order systems: cantilever response

thin and flexible probes factory calibration certificate with traceability High precision Analog output: DC 35 khz (depending on probe type)

Description of options, upgrades and accessories for the laser beam stabilization system Compact

UNIT 2. Q.1) Describe the functioning of standard signal generator. Ans. Electronic Measurements & Instrumentation

Moving Test - MT3000

OPERATING AND MAINTENANCE MANUAL

P-500 Series PZT Flexure Stages

Preliminary study of the vibration displacement measurement by using strain gauge

Transcription:

Piezo-Electric Actuator Initial Performance Tests Eric Ponslet April 13, 1998 Abstract This report briefly describes the setup and results from a series of tests performed on a commercially available piezo-electric stack actuator. The tests address dynamic characterization (bandwidth, gain, etc...) as well as mechanical noise under constant commanded deflection of the actuator. ' (6,*1 (1*,1((5,1* $'9$1&(' &20326,7( $33/,&$7,216 8/75$67$%/( 3/$7)2506 ($67*$7( '5 67( /26 $/$026 10 3+21( )$; :::+<7(&,1&&20 LIGO PROJECT

Table of Contents 1. Equipment tested... 3 2. Preliminary Requirements... 5 3. Description of Test Setup... 5 3.1 For response and transfer function measurements...5 3.2 For mechanical noise measurements...6 4. Test Results Equipment as Received from PI.... 8 4.1 Open-loop Response...8 4.1.1 Slew Rate...8 4.1.2 Hysteresis...8 4.1.3 Transfer Function...9 4.2 Closed-Loop Response...9 4.3 Broadband Mechanical Noise @ Factory Settings Open and Closed Loop...10 5. Broadband Mechanical Noise Effect of Additional Filtering and Controller Tuning...11 5.1 Effect of Low-Pass Filtering on Strain Gage Signal Conditioning Board...11 5.2 Effects of Controller Tuning...12 6. Calculated Residual PZT Noise at Test Mass... 12 7. References... 14 2

1. Equipment tested The piezoelectric actuator system that was subjected to the tests described below is commercially available from Polytec PI, Inc., 3152 Redhill Ave, Suite 110, Costa Mesa, CA. The hardware is manufactured in Germany by Physik Instrumente, a well-established company in the micro-positioning arena. The system consists of a long stroke actuator stack with internal preload, instrumented with an internal strain gage bridge to measure actual stack expansion and eliminate hysteresis through closed loop control. The stack is driven by a high voltage amplifier and a closed loop controller mounted in a benchtop rack. All equipment are standard PI catalog items. Actuator P239.97 + P177.10 180 micron High voltage PZT + strain gage sensor Controller E509.S1 Closed loop controller for strain gage sensor Amplifier E507.00 High voltage amplifier (0-1000 V) Rack E501.00 9 ½ rack with power supply Extension Cables P209.10 & P892.10 10 meter extension cables for high voltage & strain gage Table 1: list of PI equipment tested. Figure 1: Overall schematics of PI controller and amplifier electronics [1]. Figures 1 and 2 show schematics of the PI controller and amplifier setup. The command signal (0-10V analog) is supplied to the amplifier module through a BNC connection on the front panel. A multi-turn potentiometer provides an adjustable DC offset of that signal. The resulting control voltage is fed to the servo control module through the bus of the rack. That control module, comprises an adjustable slew rate 3

limiter, an analog closed loop control with PI compensation, and a notch filter. The control signal is sent back to the amplifier module through the Bus of the rack. A switch on the controller module allows enabling and disabling the closed loop control. Control Input TR P401 Servo OFF Overflow Detection Slew Rate Limiter KF ON Control Output (To Amplifier) Sensor Input + KS P407 KL P402 Fine Adjust TI P403 + + + FN P406 Notch Filter Figure 2: Schematics of PI linearizing controller [1]. Figure 3: PI control rack. A number of adjustments to the PI controller are available via multi-turn potentiometers. The adjustments relevant to response and noise are: slew rate (T R ) overall control loop gain (K L ) Integral (I) gain of compensator (K I ) notch filter frequency (F N ) (the notch filter is used to avoid control instability at a primary resonance of the driven system; it was adjusted at the factory based on rough 4

stiffness and mass information provided by Hytec). The Q of that filter is adjustable with an optional potentiometer (not installed on tested hardware). strain gage signal low pass filter (enabled or disabled via jumper); the cutoff frequency of that filter is factory set to 100Hz and is not adjustable. We requested and obtained instructions for component replacements to bring this cutoff point down to 15 Hz. 2. Preliminary Requirements Approximate requirements for the LIGO fine actuation system were derived in Reference [2]. They are briefly summarized in Table 2 and Figure 4 below. Range > 150 microns Bandwidth 0 to 5±1 Hz (-3 db), -20dB/dec min. rolloff Broadband Noise < 40 nm RMS, wideband spectrum limit shown in Fig. 4 Rack-actuator distance ~ 10 meter maximum Drift < 1.5 micron / day Operating Temperature 72 ± 3.5 ºF ( 22 ± 2 ºC) Humidity 20 to 70% RH, non condensing Table 2: preliminary requirements for fine stage actuators. 10-6 10-7 (m/sqrt(hz)) 10-8 10-9 10-10 10-11 10-12 10-2 10-1 10 0 10 1 10 2 Figure 4: broadband displacement noise amplitude spectral density - preliminary requirements [2]. 3. Description of Test Setup 3.1 For response and transfer function measurements The Piezo-electric stack is mounted in a V-shaped cradle and held in place with 2 clamps. The cradle is mounted on a heavy and stiff steel base. A Newport translation table with differential micrometer drive is mounted on the same base, facing the moving end of the PZT. A non-contact eddy current probe (Kaman, SMU9000-15N) is mounted on the translation table and measures the displacement of an aluminum target attached to the moving end of the PZT. The probe has a range of 200µm and a noise floor (HYTEC 5

measurement) around 10-3 µm/ Hz. All measurements are performed using a DSP Technology data acquisition and FFT analysis system (SigLab 20-42). That system also provides a number of outputs with programmable signal generators. Sensor Target PZT stack PZT cradle Sensor bracket & clamps Micro-positioning table Suspension bolts Figure 5: PZT test setup. 3.2 For mechanical noise measurements Because of the very small motions to be detected (picometer level), a very sensitive capacitive displacement probe (Queensgate NS2000-S series, 20µm range) is mounted at the moving end of the PZT as shown in Figure 6. The output sensitivity of the probe is 2V/micron. The broadband noise level from that probe is about 10-12 m/ Hz, which allows detection of mechanical noise at the requirement level up to about 100 Hz. See Section 4.3 for the measured noise floor of that probe. Figure 6: close-up on the test fixture itself, showing the PZT stack, and the capacitive probe mounted on the micrometric translator. To minimize perturbations of the measurements from extraneous noise (acoustic and seismic), the test fixture is suspended from soft bungee cords (about 1 to 3 Hz suspension frequencies) inside an acoustically isolated enclosure (wooden box with 6

acoustic foam barrier). The enclosure itself is sitting on a layer of foam on top of a single stage isolation platform. Figure 7: the test fixture suspended from bungee cords inside the acoustic enclosure; the black box underneath the fixture is the electronics for the capacitive probe. Figure 8: overall lab setup, showing acoustic enclosure on isolation platform and data acquisition system. For noise measurements, the setup is prepared as follows: the control input to the PI controller is shorted with a BNC termination cap. both the sensor cable and high voltage cable are extended with 10 meter extension cords (worse case EMI noise pickup). the actuator is held at a constant expansion of 90 µm (50% of range) controlled via the offset potentiometer on the PI amplifier (see Figs. 1 and 3). 7

the probe is set to its nominal mid range offset from the target (20 µm). the probe power supply is provided by a Tektronix linear power supply (+/- 15V, noise & ripple < 2.5mV RMS / 30 mv P-P). the probe signal is fed to an input channel of the DSP FFT analyzer, set to +/-20 mv sensitivity, DC coupled. 4. Test Results Equipment as Received from PI. 4.1 Open-loop Response 4.1.1 Slew Rate The control signal is band limited by a slew rate limiter (see Figure 2) whether the controller is enabled or disabled. We measured the actual slew rate of the PZT from the open loop step response to a large amplitude command step (4 Volts P-P or approximately 70 µm P-P). The measured response is shown in Fig 9; the slew rate is 2116 µm/second. Note that the high voltage supply to the PZT was also monitored and was found to have the same slow rise as the PZT deflection. Note that the slew rate limiter is factory adjusted to avoid exceeding the current capacity of the high voltage amplifier. The current capacity of the E-507.00 amplifier is between 13 ma (continuous) and 50 ma (peak). The capacitance of the P-239.97 stack is about 3 µf and its expansion range 180 µm. This would lead to a maximum slew rate between 780 and 3000 µm/sec, consistent with the measured rate. 70 60 control signal (V) PZT deflection (microns) 50 40 30 20 10 0-10 0 0.01 0.02 0.03 0.04 0.05 0.06 0.07 0.08 time (sec) Figure 9: open-loop step response (with slew rate limiter active); the initial response rate is 2116 µm/sec. 4.1.2 Hysteresis The hysteresis in open-loop control was not quantitatively recorded; however, PI s calibration certificate shows about 13% hysteresis for full deflection cycles of the PZT. This is consistent with our observations. 8

4.1.3 Transfer Function Note that since the open loop response is non-linear (both because of the hysteresis and the slew rate limitation), the open loop transfer function is not completely meaningfull and probably amplitude dependent. The transfer function shown in Fig. 10 was obtained from white noise excitation between 0 and 500 Hz at 0.5V RMS. 10 0-10 db (µm/µm) -20-30 -40-50 10 0 10 1 10 2 10 3 0-50 phase (deg) -100-150 -200 10 0 10 1 10 2 10 3 Figure 10: open-loop transfer function from control signal (normalized to microns @ 10V/180µm) to PZT expansion (in microns); command signal was white noise with 0.5V RMS (9µm RMS); for indication only since open-loop response is non linear. 4.2 Closed-Loop Response 10 0 db (µm/µm) -10-20 -30-40 -50 10 0 10 1 10 2 10 3 0 phase (deg) -50-100 -150-200 10 0 10 1 10 2 10 3 Figure 11: closed loop transfer function from control signal (normalized to microns @ 10V/180µm) to PZT expansion (in microns); command signal was white noise with 0.5V RMS (9µm RMS). 9

Figure 11 shows a Bode plot of the closed loop transfer function of the PZT actuator. The observed bandwidth is DC to 8 Hz @ -3dB, with a roll-off rate of approximately 30dB/decade above 10 Hz. This compares favorably with the requirement. A sharp anti-resonance is obvious at about 47 Hz and is due to the notch filter (factory adjusted, see Figure 2). The phase lag is about -40º at 5 Hz. 4.3 Broadband Mechanical Noise @ Factory Settings Open and Closed Loop Figure 12 shows amplitude spectral densities of the measured mechanical noise from the PZT. The four curves in the figure show the background noise from the probe (with open circuit PZT), the open-loop noise, the closed loop noise, and the preliminary requirement of Figure 4. All measurements were performed with all hardware as obtained from the factory (no readjustments or modifications). 10-6 10-7 10-8 10-9 noise (m/ Hz) 10-10 10-11 10-12 10-13 10-14 10-15 10-16 10-2 10-1 10 0 10 1 10 2 10 3 10 4 Figure 12: PZT noise levels measured with Queensgate capacitive probe: preliminary requirement (red), probe background (black), open loop PZT at mid range (green), and closed loop PZT at mid range (blue). The following observations can be made: the background noise from the probe is at the expected level (approx. 1 picometer/ Hz). the measured noise to background noise ratio is 10 or more up to about 2000 Hz in closed loop. the closed loop noise is about 30 times higher than the open loop noise (Dc to 100 Hz). 10

a shallow peak (Q~3) can be seen in the closed loop noise spectrum around 47 Hz. That peak is caused by the notch filter (adjustable, factory set to 47 Hz) as also evident in Figure 11. there is a significant amount of 60Hz noise and higher harmonics. That noise is likely a combination of power supply noise, direct pickup by the strain gage and extension cable, and other noise inputs inherent to the test setup and instruments. Substantial day to day variations were observed in the amplitude of those peaks, probably due to changes in the electrical noise environment of the lab. These peaks may not however be a concern as they are significantly attenuated by the isolation stacks (see Figure 14). the noise level is within preliminary requirement up to about 20 Hz in closed loop. Note that the preliminary requirement is unduly strict above 35 Hz, where thermal noise becomes the primary contributor to overall strain noise while PZT and seismic noise rolls off at a very steep 160dB/decade (see Figure 14). except for the 60 Hz and harmonics noise, the measured noise spectrum shown in the figure was found to be remarkably stable from one measurement session to the next. 5. Broadband Mechanical Noise Effect of Additional Filtering and Controller Tuning 5.1 Effect of Low-Pass Filtering on Strain Gage Signal Conditioning Board The strain gage signal conditioning board used by PI includes a low pass filter, factory set to 100Hz and intended to reduce noise in low frequency control applications. That filter can be enabled or disabled via a jumper on the board. The controller was received from PI with the filter disabled. 10-9 10-10 m/sqrt(hz) 10-11 10-12 closed-loop, 15 Hz filter closed-loop, 100 Hz filter closed-loop, no filter 10-13 10 0 10 1 10 2 10 3 Figure 13: PZT noise levels showing the effect of the low pass filtering on the strain gage signal conditioning board. 11

We first measured the effect of the factory-set 100Hz filter (see Figure 13) and found that it effectively reduced noise above 100 Hz by an additional 14 db/decade. This lead us to request information from the PI factory on component exchanges required to lower the roll-off frequency of that filter to about 15 Hz. The noise spectrum was remeasured after exchanging components per PI s instructions. Although the modified filter does attenuate noise by another -20dB/dec between 15 and 30 Hz, it causes a slight increase in the noise at lower frequencies. Also, above 50 Hz, the attenuation is no better than that obtained with the 100Hz filter. Note also that below 35 Hz, the attenuation is no better than about a factor 2, a very small gain when compared to the stack attenuation (about 1/100000 at 35 Hz) as seen in Figure 14. 5.2 Effects of Controller Tuning We experimented with various adjustable parameters of the PI controller: primarily the overall control gain, the integral gain (I), and the notch filter frequency. The effects of those adjustments on step response, transfer function, and noise were qualitatively observed. However, definitive conclusions are hard to draw because of the interdependence of the various adjustments and the fact that all tests so far were performed without mass or stiffness loading on the stack. Any combination of adjustments judged optimal without loads are likely not optimal at all once the actuator is inserted in the system. Note that this is probably truer in this particular application than most other PZT applications, because of the extremely large mass driven by the PZT. In Short, it has become clear that detailed analysis of the effect of those adjustments requires a test setup that simulates as closely as possible the actual mass and stiffness conditions that the actuator will experience in the LIGO detector. Such tests are being planned at time of writing. 6. Calculated Residual PZT Noise at Test Mass The support system, isolation stack, and suspension system filter the largest part of the noise caused by the PZT stack. To evaluate the residual displacement noise at the test mass, we calculated the transfer function from unit deflection of the PZT actuator to U- motion of the optics table surface, using the Matlab models of the SEI. Combining those with approximate transfer functions of the SUS and the measured PZT noise spectrum, we derive the residual noise curves of Figs 14 and 15. Three curves are shown in the figures; they correspond to transmitted noise from closed loop PZT actuators without sensor signal filtering, with 100Hz low-pass filter, and with the modified 15Hz low-pass filter. The seismic residual noise requirement [3] is also shown. Note that the Matlab models use a single rigid body to represent the support tubes and support platform, so that the effects of differential noise inputs from the 4 PZTs cannot be accurately accounted for. The PZT-to-optics-table transfer function calculated with these models implicitly assumes that the inputs from the 4 PZT actuators are synchronized (common-mode transfer function). The noise inputs from the four PZTs are in reality uncorrelated (except probably for the 60Hz line power noise and harmonics). The residual noise at the test mass (D testmass ) was therefore calculated as 12

D testmass 1 = 4 D PZT T T PZT table table testmass where D PZT is the measured amplitude spectral density of the noise from a single PZT stack, T PZT-table is the transfer function from PZT expansion to optics table motion in the U direction, T table-testmass is the transfer function of the SUS system in the U direction, and the 4 factor is an attempt at accounting for uncorrelated inputs from the 4 PZTs. Note also that uncorrelated noise inputs would cause yaw motions of the support platform that are not included in this approximation. 10 0 10-5 seismic noise requirement seismic noise PZT noise, CL, 15Hz filter PZT noise, CL, 100Hz filter PZT noise, CL, no filter amplitude spectral density (m/sqrt(hz)) 10-10 10-15 10-20 10-25 10 0 10 1 10 2 10 3 Figure 14: Residual noise from PZT at the test mass compared to residual seismic noise and LIGO requirement for seismic noise; the effect of the strain gage signal filtering is negligible at this scale. A few observations can be made about those results: The amount of residual noise from the PZT at the test mass is very close to the amount of residual seismic noise. The preliminary requirement of Figure 4 is unnecessarily restrictive above 35 Hz where thermal noise becomes the primary noise source for LIGO. In the critical range from 10 to 35 Hz, the residual PZT noise is roughly 4 to 10 times less than the residual seismic noise (within the assumptions of this analysis). The relative effect of the strain gage filter on the residual test mass noise is small. 13

10-8 10-10 seismic noise requirement seismic noise PZT noise, CL, 15Hz filter PZT noise, CL, 100Hz filter PZT noise, CL, no filter 10-12 amplitude spectral density (m/sqrt(hz) ) 10-14 10-16 10-18 10-20 5 10 15 20 25 30 35 40 45 50 Figure 15: Enlarged view of the residual noise from the PZT at the test mass in the critical frequency range. Residual PZT noise is compared to residual seismic noise and LIGO requirement for seismic noise; 7. References 1. Operating Manual PZ 77E, E-509 Position Servo Control Module, Release 1.0, Physik Instrumente (PI) Gmbh & Co., Waldbronn, Germany, July 25, 1997. 2. D. Coyne, SEI Fine Actuation Subsystem Requirements, LIGO-E970141-00-D (Draft), October 22, 1997. 3. F. Raab, N. Solomonson, and M. Fine, Seismic Isolation Design Requirements Document, LIGO-T960065-03-D, February 20, 1997. 14