CARESTREAM INDUSTREX Digital Imaging Plates

Similar documents
CARESTREAM INDUSTREX Digital Imaging Plates

NON-DESTRUCTIVE EVALUATION UTILIZING IMAGING PLATES FOR FIELD RADIOGRAPHY APPLICATIONS

HPX1-Plus. For Non-Destructive Testing. THE BENCHMARK IN COMPUTED RADIOGRAPHY.

CR Basics and FAQ. Overview. Historical Perspective

INDUSTREX (installed and tested prior to shipping) Windows 7 Ultimate 64 bit Filter

HPX-PRO. For Non-Destructive Testing. THE ANSWER TO PORTABLE DIGITAL NDT.

HPX-PRO. For Non-Destructive Testing.

PRACTICAL CONSIDERATIONS AND EFFECTS OF METALLIC SCREEN FLUORESCENCE AND BACKSCATTER CONTROL IN GAMMA COMPUTED RADIOGRAPHY

DACON INSPECTION SERVICES. Computed RADIOGRAPHY

Moving from film to digital: A study of digital x-ray benefits, challenges and best practices

Experiences of users in Digital Radiography

Examination of Pipe Welds by Image Plate Based Computed Radiography System

Acquisition, Processing and Display

UNDERSTANDING THE CURRENT CAPABILITIES AND LIMITATIONS OF DIGITAL INDUSTRIAL RADIOGRAPHY

Digital radiography: Practical advantages of Digital Radiography. Practical Advantages in image quality

NDT Supply.com 7952 Nieman Road Lenexa, KS USA

STUDENT REVIEW QUESTION SET K CR/DR CONTENT AREA

KODAK INDUSTREX AA400 Film Versatility is the key.

Veterinary Science Preparatory Training for the Veterinary Assistant. Floron C. Faries, Jr., DVM, MS

Advancements In Digital Radiography: CR, DR, and DICONDE

ISO INTERNATIONAL STANDARD

RADIOGRAPHIC EXPOSURE

Gamex CR 2.0 Program description and operating manual

INDUSTREX Film. For Non-Destructive Testing WHEN CLARITY IS CRITICAL

Light Enhancement Film

Computer Output Microfilm Data Sheet

Guide to Manual Processing of NDT Films

Introduction. Chapter 16 Diagnostic Radiology. Primary radiological image. Primary radiological image

Amorphous Selenium Direct Radiography for Industrial Imaging

THE CR SPECIALISTS HD-CR CR 35 NDT

Application Notes! RIGID MATERIALS. Rigid material standards. Printing on Rigid Materials with the Gerber Solara ion & Gerber CAT UV

Radiology Physics Lectures: Digital Radiography. Digital Radiography. D. J. Hall, Ph.D. x20893

COMPUTED RADIOGRAPHY (CR)

Spokane Industries Computed Radiography Implementation Experience. David Jolin Technical Sales

Computed Radiography of Resistance Temperature Sensor for Indian PHWR

AUTOMATED AND QUANTITATIVE METHOD FOR QUALITY ASSURANCE OF DIGITAL RADIOGRAPHY IMAGING SYSTEMS

Film Replacement in Radiographic Weld Inspection The New ISO Standard

September 10, 2009 Page 1 PRODUCT DESCRIPTION PRODUCT FEATURES

A True Innovation in Non-Destructive Testing System FUJIFILM COMPUTED RADIOGRAPHY. Series 4 CR

KODAK EKTASCAN IR Laser Imaging Film / 1356 / EIR

Low V.O.C. 2-Component Polyurethane Game Line Paint System

Diazo C Microfilm 1957, 2957, 3957 Diazo D Microfilm 1956, 2956, 3956

Pilkington OptiView Protect. Pilkington OptiView. Handling and Processing Guidelines

3M Scotchgard Surface Protection Film 2200

Autofluor. Autoradiographic Image Intensifier. Documented... Autofluor is Superior! C 3 PPO-DMSO. Autofluor

Hardware for High Energy Applications 30 October 2009

ScienceDirect. Evaluation of Pipe Wall Thickness Based on Contrast Measurement using Computed Radiography (CR)

Contrast Enhancement Materials CEM 365HR

Conversion, Application and Maintenance of the Avery Dennison Floor Marking System Instructional Bulletin #6.30

This document is originally written in English. Guide Name: CS 7400 User Guide Part Number: 6K3919 Revision Number: 03 Print Date:

9/10/2012. Computed Radiography Chapter 3 Physics and Technology. What is Computed Radiography?

Radiographic Testing (RT) [10]

Standard Guide for Computed Radiology (Photostimulable Luminescence (PSL) Method) 1

TECHNICAL DATA / COLOR PAPER July 2003 E-23

Modelling Computed Radiography Detectors with a Cascaded Linear System Model

Application of 3M Pressure Sensitive Films to 3M Panagraphics Substrates

Section 914. JOINT AND WATERPROOFING MATERIALS

Digital Radiographic Inspection replacing traditional RT and 3D RT Development

VIEWpanel Technical Manual

1. Initial Precautions 2. Technical Precautions and Suggestions 3. General Information and Cure Stages 4. Understanding and Controlling Cure Time

Technical Notes. Introduction. Optical Properties. Issue 6 July Figure 1. Specular Reflection:

Contrast Enhancement Materials CEM 365iS

10/26/2015. Study Harder

RA Duplicating Film X-OMAT 5000 RA Processor; RP X-OMAT Chemicals; 90 Second Cycle; Diffuse Visual Densitometry. Density

X-ray Imaging. PHYS Lecture. Carlos Vinhais. Departamento de Física Instituto Superior de Engenharia do Porto

ON THE WAY TO DIGITAL RADIOGRAPHY

TECHNICAL DATA / COLOR PAPER July 2003 E-19

PolyCarb has good resistance to many chemicals. Some chemicals may harm the PolyCarb sheets. Contact us when in doubt about any chemical.

Tapecoat. Application Guideline

The HOIS recommended practice for in-service computed radiography of pipes

Technical data for Brother TZe Tapes

IMAGELINK Diazo Duplicating Microfilms

KODAK EKTACHROME 64T Professional Film

DIGITAL RADIOGRAPHY ARTIFACTS

Tech Notes. GEL COAT FLEXIBILITY The Flex Strip Test. INTERPLASTIC CORPORATION Thermoset Resins Divison ISSUE 1

Conversion to Digital Radiography from Film Radiography

The HOIS recommended practice for in-service computed radiography of pipes. Dr Stephen F Burch, WCNDT 2012, Durban, SA. April 2012

3M Impact Protection Profile Installation System Instructions

PBWWinc. GEN-U-LINE 4000 Series Application Guide. Technical Data Sheet. General Guidelines:

Digital Imaging Considerations Computed Radiography

ORAFLEX Cushion. for Flexible Packaging. Flexographic Mounting Tapes - Bringing life to your prints. Cushion Mounting Tapes

ELE-3COP-559 Title Installation of /225 Coated Straight and 90 Moulded Parts using a Hot Air Gun.

3/31/2011. Objectives. Emory University. Historical Development. Historical Development. Historical Development

HP Latex Inks and print durability

Surprises with Light JoAnne Dombrowski

EASTMAN PLUS-X Reversal Film / 7276

OPTIMIZATION OF SCANNING PARAMETERS OF COMPUTED RADIOGRAPHY FOR CHARACTERIZATION OF EXPANDED POLYSTYRENE

UNPACK & ASSEMBLY. Done! CAUTION! THE MILL WILL BE VERY HEAVY - GET ASSISTANCE Pepe Tools.

STANDARD INSTALLATION PROCEDURES FOR RAYATEN COATED MOULDED PARTS STRAIGHT, 90 AND 45

Investigations on Fading Characteristics and Ghost Image Formation in Image Plates

PBWWinc. Technical Data Sheet. GERLINE Gameline Paint Instructions for Sport Flooring. General Guidelines:

1. Storage and transportation. Care Instruction

Cleaning & Maintenance Advice Low emissivity (Low E) Coated Glass Clear, Toned and Super Toned.

Industrial and Transportation Business DISTRIBUTOR PRODUCT LINE

COMPARATIVE STUDY BETWEEN COMPUTED RADIOGRAPHY AND CONVENTIONAL RADIOGRAPHY

lb. Coated Inkjet Bond

Issue 2 March Electroluminescent Materials

MultiPrep Procedure. Backside Thinning of a Flip-Chip Device G. D. Liechty, C. A. Smith, Allied High Tech Products, Inc.

PERFORMANCE CHARACTERIZATION OF AMORPHOUS SILICON DIGITAL DETECTOR ARRAYS FOR GAMMA RADIOGRAPHY

Loose Tube Cable End Preparation for Splicing For Series 11D, 1GD, 12D, 1AD, 1DD, 1CD, 11, 1G, 12, 12L, 1A, 1D, 1C, 1NY, 13, 1H, HZD and HZA

Transcription:

2016-09-29 TI-2632 CARESTREAM INDUSTREX Digital Imaging Plates CARESTREAM INDUSTREX Flex Digital Imaging Plates (IPs) are designed for computed radiography in nondestructive testing applications. These plates employ storage phosphors, which capture and retain a latent image in direct proportion to an exposure of ionizing radiation. This latent image is subsequently read out by scanning with a red laser, which stimulates the emission of blue light, in a phenomenon known as photostimulated luminescence. The imaging plate can be loaded into a flexible or rigid cassette and used with metal screens of appropriate thickness. These plates offer the flexibility of film without the need for wet processing, so you can capture and read images quickly and easily, both in the field and in the lab. The images can be optimized, if needed, and can be stored and shared digitally. These imaging plates work with CARESTREAM INDUSTREX Computed Radiography (CR) Systems. CARESTREAM INDUSTREX Flex GP, HR, HR-PRO and XL Blue Digital Imaging Plates are: Flexible Wrap around most shapes. Sharp Better resolution than high-speed films and a suitable replacement for many applications. Available in Multiple Sizes To cover a wide range of applications. Portable With a CARESTREAM INDUSTREX Digital System, image capture in remote areas is easier than ever. Versatile Use in a rigid or flexible cassette. Efficient Phosphor plates can be erased and reused thousands of times, require less exposure than film, and provide faster image access. Certified When used with INDUSTREX Digital Systems, all plates meet requirements for the highest CR System class according to EN 14784-1 and ASTM E 2446. Recommended Uses CARESTREAM INDUSTREX Digital Imaging Plates are available in many sizes, and are designed for a wide variety of NDT applications, including: o Castings o Erosion and corrosion o Pipeline inspection o Welded fabrication o Ordinance Storage and Handling The lifetime of the plates is hundreds or thousands of exposure/readout/erase cycles, determined primarily by the care in handling and environmental conditions. Careless handling and very dirty environments can shorten the useful lifetime of an imaging plate. Inspect phosphor plates every week and replace when wear is evident. Clean as necessary as directed in the cleaning sections of this document. Handling Handling the IP between exposure and reading and operating the CARESTREAM INDUSTREX CR Systems must be done in a darkened room that is free of light contamination. It is not necessary to have complete darkness, but the room must have subdued lighting. Excessive light levels in the processing room during handling or scanning of the exposed imaging plates will result in poor results. Under normal use conditions, phosphor plates will eventually show wear. Plate wear can result in artifacts on radiographs. This may occur from abrasion of the protective overcoat or inadvertent physical damage to the surface or edge. Certain chemical agents, such as non-approved plate cleaners, hand lotions, topical medications, food, and so forth, may also damage the plates. Handle flexible phosphor plates with care. Hold plates by the edges and by the black side. Avoid contact with the plate s white or blue (phosphor) side. Make sure hands are clean and dry. For best results, wear nitrile rubber or latex gloves.

Cutting Cutting the plates is not recommended, as this will compromise the edge seal and make the plates prone to moisture damage at the edges. New imaging plates should be cleaned before the first use. This will impart the necessary anti-static properties and prevent the attraction of dust and dirt, which can result in image artifacts. Temperature Range The normal temperature operating range is 5 to 30 ºC (23 to 86 ºF), although lower temperatures for short periods (several hours) will have no detrimental effect on the performance or physical integrity of the plates. In very hot environments, care should be exercised in bending or flexing the plates, as the phosphor layer could crack if flexed at temperatures exceeding 65 ºC (149 ºF). Between 65 and 100 ºC (149 and 212 ºF), minor distortions of the polyester support may occur, and over 100 ºC (212 ºF), damage to the phosphor layer (blistering/buckling) may occur. When radiography of very hot objects (such as hot welds) is necessary, a thin fire-resistant insulating layer may be placed between the cassette and the object, if experimentation shows it does not negatively affect image quality. Again, as a practical matter, the upper temperature limit will be determined by the point at which the cassette begins to burn or the lead identification letters begin to melt. Cleaning Plates CARESTREAM INDUSTREX Digital Imaging Plates employ a phosphor that is moisture sensitive. When exposed to water (either liquid or vapor form) the phosphor breaks down and liberates iodine. The iodine can become trapped within the phosphor layer at the interface between the overcoat and phosphor layer. The yellow-brown color in the iodine strongly absorbs the blue light emitted by the phosphor. This results in a loss in plate speed, which causes image quality problems. Water, isopropyl alcohol (Isopropanol or rubbing alcohol) and plate cleaners containing isopropyl alcohol are not recommended. Caution Read and follow instructions in Material Safety data Sheets (MSDS) for CARESTREAM Screen Cleaner (CAT No. 845-4977). Note: Plates should be wiped routinely to remove any dust and loose dirt. It is also a good idea to clean the interior of flexible cassettes to avoid transfer of dirt or dust to the plate. Other cleaning solutions can contain chemicals that cause damage to the plate and could result in immediate or future image artifacts. Cleaning Methods For routine cleaning, place the plate with the black (nonphosphor) side down on a clean, dry, non-abrasive surface, such as the cassette s cleaned Tube Side. Wipe the plate gently to remove dust. For other dirt, use CARESTREAM Screen Cleaner as follows: 1. Fold a non-abrasive, lint-free wipe or cloth, and dampen with a small amount of the solution. Do not pour the solution directly onto the plate. Excessive amounts of the screen cleaner may damage the plate. Moisture can cause immediate or future screen damage and image artifacts. Minimize contact with moisture and always dry plates immediately. 2. Wipe the plate thoroughly, one section at a time. Fold the cloth to expose a fresh area for each section of the plate. 3. If necessary, for the removal of persistent dirt, apply pressure while cleaning.

4. Thoroughly dry the plate with a clean and dry, non-abrasive, lint-free cloth to remove residual cleaner. Do not leave the plate to air dry. Never insert a storage phosphor plate into a cassette unless the plate is thoroughly dry. Sharpness Sharpness is typically characterized by Modulation Transfer Function (MTF), a measure of signal or image modulation as a function of spatial frequency. The figure below shows the MTF curves for all imaging plates and highlights the improved sharpness of the HR and XL Blue plates. Modulation Transfer Function Characteristics Physical Properties Nominal Thickness (microns) Total Overcoat Phosphor Support Backing Sharpness (MTF %) 100 90 80 70 60 50 40 30 20 XL Blue HR, HR-PRO GP XL 436 4 160 254 18 HR 436 4 160 254 18 HR-PRO 443 11 160 254 18 GP 584 11 290 254 29 10 0 0 1 2 3 4 5 6 Spacial Frequency (cycles/mm) H192_3008AA CARESTREAM INDUSTREX Digital Imaging Plates are comprised of a layer of barium fluorobromoiodide doped with divalent Europium (BaFBr/I:Eu2+), coated on a 10-mil polyester support. A polymer overcoat provides protection against normal handling such as fingerprints and moisture. For the HR-PRO digital imaging plate, a polyethylene terephthalate (PET) laminate overcoat is applied over an adhesive layer to improve the plate physical properties. A black polycarbonate backing layer provides anti-halation protection and curl control, and also contains slip agents for optimum transport through the CR reader hardware. Flexibility The plates can be bent around pipes and other curved surfaces down to a radius of one inch, although as a practical matter, the bend radius will be much more limited by the configuration of the cassette and the lead screens when used. Signal Retention The stored energy in the imaging plate phosphors will decay over time, resulting in some signal loss. This effect is known as image fading. This loss is also a function of temperature and will be greater at elevated temperatures. In most applications, this signal decay will be of little importance, given the very wide latitude of storage phosphor imaging plates. In critical applications, there are a few measures that can be taken; when the interval between exposing and readout is more than a few hours, the plates can be exposed with a higher radiation dose to compensate for signal decay. For shorter intervals, simply keeping the time between exposing and readout constant will eliminate any decay-induced variability.

The figure below shows the percent signal retention at ambient temperatures after an exposure 120 kv in accordance with ASTM E2445.The plot is representative for Carestream Industrex flex digital imaging plates. The signal retention remains above 75 % at 4 days. Signal Retention Emission The wavelength of light needed to stimulate phosphor luminescence is different from the wavelength of light released by the phosphors. This enables the stimulating light to be filtered from the phosphor luminescence. The emission is typically stimulated with a red laser diode, while the photostimulated emission of the Carestream Industrex Digital Imaging Plates is a narrow band of blue light near 400 nm. Emission GP, HR, HR-PRO, and XL Blue Emission 1.0 0.9 0.8 0.7 Absorption Imaging plates are quite sensitive to low-energy scatter, as shown by their X-ray energy absorption curves below. For this reason, the use of lead screens, especially at energies greater than 100 kv, is important in controlling scatter to produce optimum image quality. Relative (Photon) Absorption 1 0.1 0.01 Absorption X-ray Absorption vs. X-ray Energy GP All others 0 50 100 150 X-ray Energy (kv) Relative Emission 0.6 0.5 0.4 0.3 0.2 0.1 0.0 Disposal 300 350 400 450 500 550 600 650 700 750 800 Wavelength (nm) Due to the presence of barium, this plate may be considered a hazardous waste at the end of its useful life. For disposal or recycling information, please contact local authorities. H192_3011AA To locate the Article Information Sheet (AIS) for each plate, go to www.carestream.com and navigate to Support and Education > Environment Health & Safety. NOTICE: While the data in this publication are typical of production coatings, they do not represent standards which must be met by Carestream Health, Inc. Varying storage, exposure, and operating conditions will affect results. The company reserves the right to change and improve product characteristics at any time. H192_3010AA

Carestream Health, Inc. 150 Verona Street Rochester, NY 14608 United States TI-2632 2016-09-29 Version 4.0 CARESTREAM and INDUSTREX are trademarks of Carestream Health, Inc. Carestream Health, Inc. 2016 Printed in the United States