Designing practical on-site. on-site calibration protocols for acoustic systems: key elements and pitfalls.

Similar documents
Review of uncertainty sources in indoor PV calibration of c-si, and thin film single junction and multi junction cells and modules

Accurate determination of distribution network losses

The noise radiated by marine piling for the construction of offshore wind farms

COOMET Pilot Comparison 473/RU-a/09: Comparison of hydrophone calibrations in the frequency range 250 Hz to 200 khz

Application Note. Airbag Noise Measurements

Validation of spectral response polychromatic method measurement of full size photovoltaic modules using outdoor measured data

Effect of I-V translations of irradiance-temperature on the energy yield prediction of PV module and spectral changes over irradiance and temperature

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TIME AND FREQUENCY METROLOGY

PXIe Contents. Required Software CALIBRATION PROCEDURE

Why not narrowband? Philip Fontana* and Mikhail Makhorin, Polarcus; Thomas Cheriyan and Lee Saxton, GX Technology

Power Quality Measurements the Importance of Traceable Calibration

Reducing comb filtering on different musical instruments using time delay estimation

Loughborough University Institutional Repository. This item was submitted to Loughborough University's Institutional Repository by the/an author.

Numerical Modeling of a Time Reversal Experiment in Shallow Singapore Waters

Virtual FFT Analyser for identification of harmonics and inter-harmonics metrological aspects

15 th Asia Pacific Conference for Non-Destructive Testing (APCNDT2017), Singapore.

ISO INTERNATIONAL STANDARD

Where DSP meets Measurement Science: A Sound Example. By Andrew Hurrell PhD

Initial solar cell characterisation test and comparison with a LED-based solar simulator with variable flash speed and spectrum

Accessing the performance. light processing projector

Underwater acoustic measurements of the WET-NZ device at Oregon State University s ocean test facility

Organisation Internationale de Métrologie Légale

Noise Measurements Using a Teledyne LeCroy Oscilloscope

Characterizing High-Speed Oscilloscope Distortion A comparison of Agilent and Tektronix high-speed, real-time oscilloscopes

Metrological Characterization of Low Power Instrument Transformer Integrated in MV Recloser

Measuring Instrument Combinations

Downloaded 09/04/18 to Redistribution subject to SEG license or copyright; see Terms of Use at

Power Measurement Basics

3D Intermodulation Distortion Measurement AN 8

Pilot experiments for monitoring ambient noise in Northern Crete

Laboratory Experiment #1 Introduction to Spectral Analysis

Acoustic Calibration Service in Automobile Field at NIM, China

Template Planning Condition on Amplitude Modulation Noise Guidance Notes

An interlaboratory comparison programme on high frequency electromagnetic field measurements in a controllable environment performed in Greece

CHARACTERISATION OF AN AIR-GUN AS A SOUND SOURCE FOR ACOUSTIC PROPAGATION STUDIES

USB Dynamic Signal Acquisition

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Jitter Analysis Techniques Using an Agilent Infiniium Oscilloscope

Errata to Procedural Standard for Sound & Vibration Measurement 2015 Third Edition

Presented on. Mehul Supawala Marine Energy Sources Product Champion, WesternGeco

Loughborough University Institutional Repository. This item was submitted to Loughborough University's Institutional Repository by the/an author.

Transfer Function (TRF)

DIGITAL FILTERING OF MULTIPLE ANALOG CHANNELS

EURO-ASIAN COOPERATION OF NATIONAL METROLOGICAL INSTITUTIONS (COOMET)

Identifying Gas Ultrasonic Meter Problems Using Advanced Techniques

Week 1. Signals & Systems for Speech & Hearing. Sound is a SIGNAL 3. You may find this course demanding! How to get through it:

Removal of Continuous Extraneous Noise from Exceedance Levels. Hugall, B (1), Brown, R (2), and Mee, D J (3)

Assessment of rail noise based on generic shape of the pass-by time history

Signal Processing for Digitizers

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

P a g e 1 ST985. TDR Cable Analyzer Instruction Manual. Analog Arts Inc.

Data Acquisition and Processing of a Distributed 3D Induced Polarisation Imaging system

Cutting tools in finishing operations for CNC rapid manufacturing processes: simulation studies

Bandwidth Enhancement through Fractals and Stacking of Microstrip Antenna for Ku-Band Applications

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

note application Measurement of Frequency Stability and Phase Noise by David Owen

Signals, systems, acoustics and the ear. Week 3. Frequency characterisations of systems & signals

Underwater acoustics Measurement of radiated underwater sound from percussive pile driving

Representing human movement and behaviour in virtual environment using gaming software

Acoustics, signals & systems for audiology. Week 3. Frequency characterisations of systems & signals

Characterization of a Very Shallow Water Acoustic Communication Channel MTS/IEEE OCEANS 09 Biloxi, MS

Federal Communications Commission Office of Engineering and Technology Laboratory Division

ACOUSTIC NOISE MEASUREMENT METHODOLOGY FOR THE BILLIA CROO WAVE ENERGY TEST SITE ANNEX A: SUMMARY OF OPERATIONAL UNDERWATER NOISE FROM A WAVE ENERGY

Implementing Automated Oscilloscope Calibration Systems

This is a preview - click here to buy the full publication

ISO INTERNATIONAL STANDARD. Acoustics Unattended monitoring of aircraft sound in the vicinity of airports

An acoustic emission slope displacement rate sensor: Comparisons with established instrumentation

Localization of underwater moving sound source based on time delay estimation using hydrophone array

Generic noise criterion curves for sensitive equipment

Peter Dack, Vice President Sales & Marketing Radian Research, Inc.

Adaptive notch filters from lossless bounded real all-pass functions for frequency tracking and line enhancing

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Traceability for Oscilloscopes and Oscilloscope Calibrators

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

Phased Array Velocity Sensor Operational Advantages and Data Analysis

HIGH FREQUENCY INTENSITY FLUCTUATIONS

Anthropogenic noise measurements and impacts for assessment of the marine environment

Physikalisch-Technische Bundesanstalt

Agilent PN 4395/96-1 How to Measure Noise Accurately Using the Agilent Combination Analyzers

Microphone calibration service for airborne ultrasound

Voltage Sensors URV5-Z

Bending vibration measurement on rotors by laser vibrometry

EUROMET PROJECT 401 Harmonisation of Audiometry Measurements within Europe

This document is a preview generated by EVS

Non-destructive testing Equipment for eddy current examination. Part 1: Instrument characteristics and verification

RCTrms Technical Notes

Martin Salter Centre for Electromagnetic and Time Metrology, National Physical Laboratory

Capacitive MEMS accelerometer for condition monitoring

Acoustic resolution. photoacoustic Doppler velocimetry. in blood-mimicking fluids. Supplementary Information

Enhanced subsurface response for marine CSEM surveying Frank A. Maaø* and Anh Kiet Nguyen, EMGS ASA

Failure modes and effects analysis through knowledge modelling

Validation of primary hydrophone calibrations by inter-laboratory comparisons and by independent calibration methods

User-friendly Matlab tool for easy ADC testing

Experimental study of traffic noise and human response in an urban area: deviations from standard annoyance predictions

USB-UT350(T) Portable Ultrasonic Pulser/Receiver and Analog to Digital Converter. User s Guide

Contents. CALIBRATION PROCEDURE NI PXIe-5668R 14 GHz and 26.5 GHz Signal Analyzer

Evaluating VNA post-calibration residual errors using the ripple technique at millimetre wavelengths in rectangular waveguide

SENSING OF METAL-TRANSFER MODE FOR PROCESS CONTROL OF GMAW

Project Report Liquid Robotics, Inc. Integration and Use of a High-frequency Acoustic Recording Package (HARP) on a Wave Glider

Transcription:

Loughborough University Institutional Repository Designing practical on-site calibration protocols for acoustic systems: key elements and pitfalls This item was submitted to Loughborough University's Institutional Repository by the/an author. Citation: BUTLER, M., NORRIS, J. and LEPPER, P., 2014. Designing practical on-site calibration protocols for acoustic systems: key elements and pitfalls. IN: UA2014-2nd International Conference and Exhibition on Underwater Acoustics, Rhodes, Greece, 22-27 June 2014, pp.223-230. Additional Information: This is a conference paper. Metadata Record: https://dspace.lboro.ac.uk/2134/18315 Version: Submitted for publication Publisher: IACM Rights: This work is made available according to the conditions of the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International (CC BY-NC-ND 4.0) licence. Full details of this licence are available at: https://creativecommons.org/licenses/by-nc-nd/4.0/ Please cite the published version.

DESIGNING PRACTICAL ON-SITE CALIBRATION PROTOCOLS FOR ACOUSTIC SYSTEMS: KEY ELEMENTS AND PITFALLS Michael Butler a,b, Jennifer Norris a, Paul Lepper b a The European Marine Energy Centre (EMEC), Old Academy Buildings, Stromness, Orkney, KW16 3AW, UK b Loughborough University, Loughborough, UK Contact author: Michael Butler, The European Marine Energy Centre (EMEC), Old Academy Buildings, Stromness, Orkney, KW16 3AW, UK. Tel: 01856 852060 Abstract: Although acoustic systems are increasingly being used for environmental and noise surveys of marine energy devices, there are currently no standard protocols for the on-site full bandwidth calibration of these systems. Reports often include little or no information on the methods of calibration used before, during or after surveys. Without proper calibration, the sound levels may be far from accurate, leading to skewed reporting and inaccurate conclusions. Hydrophone calibrations from internationally recognised standardisation centres, such as NPL, allow providers to reference their systems to international standards. Marine renewable energy devices, however, are often deployed in remote areas and it is not always practical or cost-effective to send every acoustic system to be independently tested before every deployment. On-site referencing of multiple units to a single standardised system can help improve calibration traceability. Although this may at first appear relatively simple, the production of an accurate, full-spectrum calibration, particularly in real-world test sites, is surprisingly difficult. Keywords: Calibration, Marine Energy, Standards, Reporting

1. INTRODUCTION Calibration of acoustic systems can seem complex and is particularly difficult to achieve in remote areas, away from services for standard referencing of equipment. This is true for acoustic surveys of marine energy devices, which are often deployed on remote areas of coast. Setting up a portable calibration system can avoid these difficulties. This paper will describe the key elements involved and pitfalls to avoid when setting up such a system. It will use work at the European Marine Energy Centre as an example, with particular reference to the whole system comparison calibration of Drifting Acoustic Recorder and Tracker (DART) units. 2. KEY ELEMENTS The main focus for ensuring accurate calibrations should be to assess and reduce levels of uncertainty throughout the calibration system. Standards organisations have created guidance for both calibration techniques and the assessment of uncertainty. These provide extensive information to ensure accurate characterisation of acoustic systems in laboratory conditions, including good practice and pitfalls to avoid. It is recommended that these are consulted during detailed design of calibrations [1, 2, 3, 4]. For simplicity, these descriptions of uncertainty are typically grouped into type A and type B uncertainties. Type A are random uncertainties; those which can be calculated through repeated measurement. Type B are systematic uncertainties, which can be predicted and reduced through thoughtful calibration design. This section will provide a brief overview of important considerations for initial calibration design. This will use as an example the comparison calibration design for the DART units, shown in Figure 1. Figure 1: Overview of DARTs comparison calibration, including known data (Orange) from referenced data (Pa/V) and acquired data (units), and unknown physical quantities and conversion factors (Blue) that require further measurement. Elements of the system (I-VII) are described in Table 1.

Whole system comparison calibration Comparison calibration is relatively simple and fast, but requires a large body of water and a stable, calibrated reference hydrophone (see section 3 for details). These requirements can usually be achieved in remote marine areas, through access to the sea and annual reference calibrations. More information on alternative calibration techniques can be found in ISO standards and guidance literature [2]. The technique uses a source, emitting hydrophone and two receiving systems. The first receiver, the reference system, consists of a hydrophone and subsequent signal conditioning pre-calibrated to international standards to provide pressure in Pascals (Pa) from measured voltages, or Pa/V. The reference hydrophone is then placed next to a hydrophone from a second uncalibrated system, for which only Unit values (such as bits in a.wav file) are known. If both hydrophones are located at the same distance from the source, then the acoustic pressure (in Pa) at both hydrophones should be the same. It is therefore possible to use the Pa excitement of the reference system to calculate Pa/Unit values for the un-calibrated system. However, there are a number of unknowns; for example, how is the voltage from the reference hydrophone converted to units? Typically this will involve the analog-to-digital converter (ADC) within a data acquisition unit. Thus the conversion factor of the ADC, the Units/V produced, will need to be estimated before Pa values can be calculated from measurements by the reference system. The conversion factors and physical quantities highlighted in blue in Figure 1 are all unknown; by using additional measurements (see section 3), it is possible to calculate conversion factors, which will then provide a basis for calculating the physical quantities. Estimating uncertainty Uncertainty in measurement of any part of the process, including the measurement of conversion factors, could affect the accuracy of the whole calibration. Therefore, it is important to predict and test those elements that could create uncertainty. Table 1 outlines the expected uncertainties for each aspect of the DART's calibration, as further discussed in section 3. As previously described, type A uncertainties are reduced through repeated measurements, whilst type B are typically reduced by modifying the system design. Often systematic uncertainties will not become obvious until the calibration is trialled, so it is useful to plan test calibrations well in advance of data gathering, in order to allow time for purchasing and testing any additional equipment if required. When estimating uncertainty, it is important to convert all estimates and measurements to the same units. This can be simplest to achieve using percentages of uncertainty values. It is also useful to note that logarithmic units may make certain calculations more difficult; thus it may be helpful to work with linear units.

Uncertainty Type A (random) Type B (Systematic) Mitigation Detail of action taken I. Signal generation Waveform shape and levels may change B -Test signals and make waveforms as similar as possible B - Signals tested Hardware problems causing irregularities in waveforms generated at certain frequencies (e.g. 10kHz) new hardware bought. II. III. IV. Signal transmission Electrical/ acoustic noise Device under test receiving (Pa-unit) V. Reference equipment receiving (Pa-V) VI. ADC accuracy (V-Unit) VII. Analysis accuracy Random changes due to water movement, debris, etc. Random electrical/ acoustic noise Random changes in Acoustic response Random changes in Acoustic response Random changes in electrical response Reflections, mounting vibrations, bubbles, other problems with setup Equipment noise, noisy test area Pa-unit transduction will not be linear & may change over time May not be calibrated accurately & Pa-V transduction may change over time V-unit conversion will not be linear & may change over time Different methods of analysis will create different results for the same analysis (e.g. fft) A repeat B Avoid reflections/rig vibrations/bubbles A repeat B Reduce equipment noise to minimum/avoid noisy areas B - repeated calibration of the whole system B - Re-calibrate often and track change A - repeated calibration B calibration of DAQ B -Compare and decide upon most accurate method A 30 repeats for each frequency calibration, 3 repeats of each set-up. B Deep channel chosen and reflections calculated. Vibration isolation & bubble removal to be included. A Repeats, as above. B Low noise calibration equipment used. New DAQ hardware to be included. Avoiding times of environmental/anthropogenic noise at test site. A 3 repeats of calibration (short-term) and repeats before, during and after surveys. B Calibration of frequency response. Amplitude response to be included. A&B - Reference hydrophone and conditioner calibrated. To be recalibrated again asap. A DAQ output/input calibrated 5 times using a multimeter/calibrated output. B - Calibration of frequency and amplitude response. More accurate electronic testing equipment could be used in future iterations. B Taking into account the transmission environment and sample rate of the devices under test (DARTs), a half pulse Fourier transform was used for amplitude, with frequency bins of 512 providing the best compromise of sharp, but inclusive peaks. Table 1: Example of a simple uncertainty analysis, taking into account expected sources of uncertainty. Detail includes actions that were successfully used to reduce uncertainty (Orange) and actions needed to further reduce uncertainty (Blue).

3. POTENTIAL PITFALLS Signal generation and acquisition The first potential pitfall for a calibration is the calibration tone; both the generation of the electrical signal and the transduction into a physical signal can produce frequency and amplitude artefacts. Therefore, it is important to check the calibration signal and modify the calibration system as necessary. For example, in Table 1, a system with better defined waveforms was required; therefore equipment with improved specifications was purchased. Ideally, the source hydrophone should also be calibrated across the frequency range of interest in order to reduce uncertainties in source amplitude. Generation and acquisition systems also need to be designed to cover all frequencies required for acoustic surveys. For example, if surveys require measurement up to a frequency of 200 khz then it will be necessary to use a generation and acquisition system capable of working at these frequencies, i.e. an acquisition system with a sample rate of at least 400 khz. Signal transmission A major influence upon the transmission of the signal is the environment. Uncertainty due to random changes can be reduced using multiple repeats at each frequency. However, systemic effects, such as the effect of boundaries in the test area, must be designed into the calibration. For example, hydrophones must be placed at a depth such that reflections from the surface and bottom reach the receiving hydrophones a certain period after the direct signal. In Table 1 (VII), the calibration pulse is analysed using a 512 Fourier transform, thus a period of at least 512 samples will be needed before the first reflection is received. The time between direct and reflected signals is related to the distance between the source and receiving hydrophones, since a shorter distance will provide more time before reflections are received. However, this will increase the minimum frequency available for testing. This frequency can be calculated, based upon estimates of the 'far-field' [2], using: c MinFreq = 2 ( ) dis tan ce (10) with distance in meters and c as speed of sound (in m/s) within the local environment. It is therefore important to realise that the typical minimum frequency for a comparison calibration will be low khz and other methods must be used for calibrating low frequencies, such as a pistonphone calibration [2]. It may be necessary to reduce hydrophone distance, and so increase the minimum frequency, in shallow test areas. Speed of sound itself can be calculated using CTD measurements of the calibration site. Guidance on the measurement and calculation of speed of sound can be found on the NPL website [5]. Finally, the positioning, mounting and wetting of the hydrophones can influence transmission. Inaccurate positioning can change the amplitude of signals, whilst mounts

can create vibrational artefacts. Therefore it is important to state and check the hydrophone distances and use guidance to avoid mount vibrations and air bubbles [6, 7]. Additional measurements Additional measurements, such as CTD measurements, will also require calculation of uncertainty, typically through calibration techniques or using standardised calibrated equipment. The importance of these calibrations will depend upon the probable effect on calibration; for example, inaccurate reference hydrophone calibration or data acquisition equipement response calibration could greatly affect values, whereas inaccurate CTD measurements may not have a great effect. Equipment accuracy The accuracy of acquisition equipment and the equipment used to test this equipment must be checked over time and if calibration methods change. For example, data acquisition equipment can be tested with voltage signals calibrated using multi-meter measurements. Systematic uncertainties of electrical equipment can be avoided through regular testing and comparison of voltage and unit values; for example, hydrophone connections may produce different values if connected before or after powering elements of the system. Calculation of values It is useful to check all calculations performed on measurements manually before any automation is introduced into the calibration process. There are a number of steps required for calibrations and it is essential to ensure that each step is calculated correctly. This is particularly important when using analysis packages to calculate the amplitude of received pulses since the values can change depending upon the section of pulse analysed and the length of section analysed. Ideally, the section analysed should be as long as possible, although this may not be feasible in practice. Finally, it is good practice to perform these calculations in SI units before converting to db; as previously mentioned use of logarithmic units can complicate certain calculations. It is also good practice to use RMS amplitudes and clarify this use in reports; this avoids comparison of peak and RMS amplitude values, which, although proportional, are not equivalent. 4. CALCULATING UNCERTAINTY The method for calculating uncertainties differs for each type of uncertainty. In addition, the following methods assume that elements are independent of each other, whereas correlated elements require additional calculation [3]. Calculating type A uncertainties These uncertainties are estimated through the calculation of arithmetic mean and standard deviation of repeated measurements. Guidelines recommend between 4-10 repetitions to cover most measurements, although accuracy can be slightly improved with additional repeats [1].

Arithmetic mean ( χ ) is simply calculated using: χ = n χ i (1) where n is the number of repeats, and χ i is the sum of their values. Standard deviation (σ ) is then calculated using: ( χ i χ ) σ = n 1 (2) Percentage standard deviation can then be calculated: σ σ (%) = χ (3) Finally standard type A uncertainty ( u a ) is calculated using: σ (%) u a = n (4) Calculating type B uncertainties Type B uncertainties are estimated using the best information available and experience of the systems being calibrated. In practice, an estimate is made of the upper and lower bounds of expected error, the difference between these values taken and divided by 2: ( upper lower) a = 2 (5) The mean is typically half-way between these two values: χ = lower + a (6) Standard uncertainty can then be calculated based upon the expected distribution of the error. For example, if all values across the range from the upper to lower estimate are expected to be equally probable, then a rectangular distribution can be used: u b = a 3 (7) whilst errors expected to cluster around mean, i.e. in a bell-shape, can be calculated using:

u b = 1. 48a (8) Finally, the percentage uncertainty can be calculated as above. Calculating combined uncertainties Once the uncertainty for each element has been calculated, it is possible to calculate the combined uncertainties, u c, through the formula: u c = u + u + u... (9) ( 1 2 3 + This formula is useable for uncertainties that combine linearly to produce a total uncertainty. This is not generally the case for calibrations, since each uncertainty is typically independent of others, but it can provide a simple 'worse-case scenario' estimate. Additional, more complicated, means to calculate combined uncertainty are detailed in international standards for uncertainty and measurement [3]. 5. ACKNOWLEGEMENTS This work acknowledges the support of the Knowledge Transfer Partnership (KTP) scheme, organised and funded by the Technology Strategy Board (TSB), UK. REFERENCES [1] S. Bell, A Beginners Guide to Uncertainty in Measurement, Measurement good practice guide, National Physical Laboratory (NPL), No. 11, Issue 2, 1999. [2] International Electrotechnical Commission (IEC), 60565:2006 Underwater Acoustics, Hydrophones, Calibration in the frequency range 0.01Hz to 1MHz. International Electrotechnical Commission (IEC), 2006. [3] Joint Committee for Guides in Metrology (JCGM), 100:2008 Guide to the Expression of Uncertainty in Measurement (GUM), International Organization for Standardization (ISO), Geneva, 2008. [4] S.P. Robinson, P.A. Lepper and R.A. Hazelwood, Good Practice Guide for Underwater Noise Measurement, National Measurement Office, Marine Scotland, The Crown Estate, NPL Good Practice Guide, No. 133, 2014. [5] NPL, http://resource.npl.co.uk/acoustics/techguides/soundseawater/ [6] NPL, Hydrophone mounting, Underwater Acoustics: Technical Guidance Note, NPL, 2012. [7] NPL, Hydrophone wetting, Underwater Acoustics: Technical Guidance Note, NPL, 2012.