Agilent AN Applying Error Correction to Network Analyzer Measurements

Similar documents
Keysight Technologies Applying Error Correction to Vector Network Analyzer Measurements. Application Note

Network Analysis Basics

Configuration of PNA-X, NVNA and X parameters

For EECS142, Lecture presented by Dr. Joel Dunsmore. Slide 1 Welcome to Network Analyzer Basics.

There is a twenty db improvement in the reflection measurements when the port match errors are removed.

Agilent ENA Series 2, 3 and 4 Port RF Network Analyzers

Keysight Technologies In-Fixture Measurements Using Vector Network Analyzers. Application Note

Agilent Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements. Product Note A

application In-Fixture Measurements Using Vector Network Analyzers Network Analysis Solutions Application Note

ECE 4265/6265 Laboratory Project 7 Network Analyzer Calibration

Agilent PNA Microwave Network Analyzers

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Measurements with Scattering Parameter By Joseph L. Cahak Copyright 2013 Sunshine Design Engineering Services

Network Analysis Seminar. Cables measurement

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

Agilent 8703B Lightwave Component Analyzer Technical Specifications. 50 MHz to GHz modulation bandwidth

FieldFox Handheld Education Series Part 3: Calibration Techniques for Precise Field Measurements

Vector Network Analyzer

Agilent Accurate Measurement of Packaged RF Devices. White Paper

Agilent PNA Microwave Network Analyzers

Agilent 4-Port PNA-L Network Analyzers

Agilent 86030A 50 GHz Lightwave Component Analyzer Product Overview

RF and Microwave Test and Design Roadshow 5 Locations across Australia and New Zealand

Network Analyzer Basics

Application Note 5525

Coaxial TRL Calibration Kits for Network Analyzers up to 40 GHz

Vector Network Analyzer Application note

Waveguide Calibration with Copper Mountain Technologies VNA

SWR/Return Loss Measurements Using System IIA

54100A Series. Network Analyzers. Precision Return Loss Measurements. Application Note

Agilent PNA Series RF Network Analyzers

A New Noise Parameter Measurement Method Results in More than 100x Speed Improvement and Enhanced Measurement Accuracy

RF Characterization Report

Preliminary Users Manual for the Self Contained Return Loss and Cable Fault Test Set with Amplified Wideband Noise Source Copyright 2001 Bryan K.

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler

NATIONAL UNIVERSITY of SINGAPORE

Agilent PN Testing amplifiers and active devices with the Agilent 8510C Network Analyzer. Product Note

Challenges and Solutions for Removing Fixture Effects in Multi-port Measurements

Keysight Technologies Vector Network Analyzer Receiver Dynamic Accuracy

VSWR MEASUREMENT APPLICATION NOTE ANV004.

(a) The insertion loss is the average value of the transmission coefficient, S12 (db), in the passband (Figure 1 Label A)

Keysight Technologies Network Analysis Applying the 8510 TRL Calibration for Non-Coaxial Measurements

Improving TDR/TDT Measurements Using Normalization Application Note

Keysight Technologies In-ixture Microstrip Device Measurements Using TRL * Calibration. Application Note

Today s modern vector network analyzers

TEST EQUIPMENT PLUS. Signal Hound USB-SA44B / USB-TG44A. Application Note 1: The Smith Chart. Rev. 0

ME1000 RF Circuit Design. Lab 4. Filter Characterization using Vector Network Analyzer (VNA)

Agilent 8703A Lightwave Component Analyzer Technical Specifications

Limitations And Accuracies Of Time And Frequency Domain Analysis Of Physical Layer Devices

Comparison of Various RF Calibration Techniques in Production: Which is Right for You? Daniel Bock, Ph.D.

Vector Network Analyzers (VERY) Basics. Tom Powers USPAS SRF Testing Course 19 Jan. 2014

Hot S 22 and Hot K-factor Measurements

Abstract: Stringent system specifications impose tough performance requirements on the RF and microwave cables used in aerospace and defense

HP 8491A/B, 8492A, 8493A/B/C, 11581A, 11582A and 11583A/C Coaxial Attenuators. Product Overview

FABRICATING AND USING A PCB-BASED TRL PATTERN WITH A CMT VNA

Agilent 8491A/B, 8493A/B/C, 11581A, 11582A and 11583C Coaxial Attenuators dc to 26.5 GHz

High Speed Characterization Report

Specifying Calibration Standards and Kits for Agilent Vector Network Analyzers. Application Note

On Wafer Load Pull and Noise Measurements using Computer Controlled Microwave Tuners

Calibration and Accuracy in Millimeter Systems. Keith Anderson

High Speed Characterization Report

Millimeter Signal Measurements: Techniques, Solutions and Best Practices

Agilent 8761A/B Microwave Switches

Termination Insensitive Mixers By Howard Hausman President/CEO, MITEQ, Inc. 100 Davids Drive Hauppauge, NY

Department of Electrical and Computer Engineering ECE332. Lab 3: High Frequency Measurements

PXIe Contents CALIBRATION PROCEDURE

New Ultra-Fast Noise Parameter System... Opening A New Realm of Possibilities in Noise Characterization

Novel Method for Vector Mixer Characterization and Mixer Test System Vector Error Correction. White Paper

Keysight Technologies Ampliier Linear and Gain Compression Measurements with the PNA Microwave Network Analyzers. Application Note

Introduction to On-Wafer Characterization at Microwave Frequencies

Platform Migration 8510 to PNA. Graham Payne Application Engineer Agilent Technologies

Vector network analysis Calibration and advanced measurements

Agilent Upgrade Guide for the 8510 Vector Network Analyzer Product Note

Understanding Mixers Terms Defined, and Measuring Performance

Signal Integrity Testing with a Vector Network Analyzer. Neil Jarvis Applications Engineer

Basics of Using the NetTek YBA250

Making a S11 and S21 Measurement Using the Agilent N9340A

A Noise-Temperature Measurement System Using a Cryogenic Attenuator

System Performance Dimensions

Managing Complex Impedance, Isolation & Calibration for KGD RF Test Abstract

Aries QFP microstrip socket

Agilent Time Domain Analysis Using a Network Analyzer

Experiment 03 - Automated Scalar Reectometry Using BenchVue

On-Wafer Noise Parameter Measurements using Cold-Noise Source and Automatic Receiver Calibration

Electronic Calibration (ECal) Modules for Vector Network Analyzers

Differential Amp DC Analysis by Robert L Rauck

Keysight Technologies Testing Amplifiers and Active Devices with the 8510C Network Analyzer


Advanced Test Equipment Rentals ATEC (2832)

Obtaining Flat Test Port Power with the Agilent 8360 s User Flatness Correction Feature. Product Note

Agilent N4000A, N4001A, N4002A SNS Series Noise Sources 10 MHz to 26.5 GHz

Return Loss Bridge Basics

Overcoming Mixer Measurement Challenges

External Source Control

Impedance 50 (75 connectors via adapters)

Agilent Introduction to the Fixture Simulator Function of the ENA Series RF Network Analyzers: Network De-embedding/Embedding and Balanced Measurement

EXPERIMENT EM3 INTRODUCTION TO THE NETWORK ANALYZER

Agilent 8902A Measuring Receiver Product Note

PNA Family Microwave Network Analyzers (N522x/3x/4xB) CONFIGURATION GUIDE

Cavity Filters. Waveguide Filters

Transcription:

Agilent AN 287-3 Applying Error Correction to Network Analyzer Measurements Application Note 2 3 4 4 5 6 7 8 0 2 2 3 3 4 Table of Contents Introduction Sources and Types of Errors Types of Error Correction One-Port Calibration The Effects of Adapters Two-Port Error Correction Electronic Calibration Estimating Measurement Uncertainty Performing a Transmission Response Calibration Enhanced-Response Calibration for Transmission Measurements Full Two-Port Calibration TRL Calibration Calibrating Noninsertable Devices Swap-Equal-Adapters Method Adapter-Removal Calibration

Introduction Only perfect test equipment would not need correction. Imperfections exist in even the finest test equipment and cause less than ideal measurement results. Some of the factors that contribute to measurement errors are repeatable and predictable over time and temperature and can be removed, while others are random and cannot be removed. The basis of network analyzer error correction is the measurement of known electrical standards, such as a through, open circuit, short circuit, and precision load impedance. The effect of error correction on displayed data can be dramatic (Figure ). Without error correction, measurements on a bandpass filter show considerable loss and ripple. The smoother, error-corrected trace produced by a twoport calibration subtracts the effects of systematic errors and better represents the actual performance of the device under test (). This application note describes several types of calibration procedures, including the popular Short-Open-Load-Through (SOLT) calibration technique, and Through-Reflect-Line (TRL). The effectiveness of these procedures will then be demonstrated in the measurement of high-frequency components such as filters. Calibrations will also be shown for those cases requiring coaxial adapters to connect the test equipment,, and calibration standards. Measuring filter insertion loss CH S 2&M log MAG db/ REF 0 db CH2 MEM log MAG db/ REF 0 db Cor After two-port calibration After response calibration Uncorrected Cor x2 2 START 2 000.000 MHz STOP 6 000.000 MHz Figure. Response versus Two-Port Calibration 2

Sources and Types of Errors All measurement systems, including those employing network analyzers, can be plagued by three types of measurement errors: Systematic errors Random errors Drift errors Systematic errors (Figure 2) are caused by imperfections in the test equipment and test setup. If these errors do not vary over time, they can be characterized through calibration and mathematically removed during the measurement process. Systematic errors encountered in network measurements are related to signal leakage, signal reflections, and frequency response. There are six types of systematic errors: Directivity and crosstalk errors relating to signal leakage Source and load impedance mismatches relating to reflections Frequency response errors caused by reflection and transmission tracking within the test receivers (The full two-port error model includes all six of these terms for the forward direction and the same six (with different data) in the reverse direction, for a total of twelve error terms. This is why two-port calibration is often referred to as twelve-term error correction.) R Directivity A Crosstalk B Frequency response reflection tracking (A/R) transmission tracking (B/R) Source Mismatch Load Mismatch Six forward and six reverse error terms yields 2 error terms for two-port devices Figure 2. Systematic Measurement Errors Random errors vary randomly as a function of time. Since they are not predictable, they cannot be removed by calibration. The main contributors to random errors are instrument noise (e.g., sampler noise, and the IF noise floor), switch repeatability, and connector repeatability. When using network analyzers, noise errors can often be reduced by increasing source power, narrowing the IF bandwidth, or by using trace averaging over multiple sweeps. 3

Drift errors occur when a test system s performance changes after a calibration has been performed. They are primarily caused by temperature variation and can be removed by additional calibration. The rate of drift determines how frequently additional calibrations are needed. However, by constructing a test environment with stable ambient temperature, drift errors can usually be minimized. While test equipment may be specified to operate over a temperature range of 0 C to +55 C, a more controlled temperature range such as +25 C ± 5 C can improve measurement accuracy (and reduce or eliminate the need for periodic recalibration) by minimizing drift errors. Types of Error Correction There are two basic types of error correction response (normalization) corrections, and vector corrections. Response calibration is simple to perform, but corrects for only a few of the 2 possible systematic error terms (namely, reflection and transmission tracking). Response calibration is a normalized measurement in which a reference trace is stored in the network analyzer s memory, and the stored trace is divided into measurement data for normalization. A more advanced form of response calibration for reflection measurements, called open/short averaging, is commonly found on scalar analyzers and averages two traces to derive a reference trace. Vector error correction is a more thorough method of removing systematic errors. This type of error correction requires a network analyzer capable of measuring (but not necessarily displaying) phase as well as magnitude, and a set of calibration standards with known, precise electrical characteristics. The vector-correction process characterizes systematic error terms by measuring known calibration standards, storing these measurements within the analyzer s memory, and using this data to calculate an error model which is then used to remove the effects of systematic errors from subsequent measurements. This calibration process accounts for all major sources of systematic errors and permits very accurate measurements. However, it requires more standards and more measurements than response calibration. The two main types of vector error correction are the one-port and two-port calibrations. One-Port Calibration A one-port calibration can measure and remove three systematic error terms (directivity, source match, and reflection tracking) from reflection measurements. These three error terms are derived from a general equation which can be solved in terms of three simultaneous equations with three unknowns. To establish these equations, three known calibration standards must be measured, such as an open, a short, and a load (the load value is usually the same as the characteristic impedance of the test system, generally either 50 or 75 ohm). Solving the equations yields the systematic error terms and makes it possible to derive the s actual reflection S-parameters. 4

When measuring two-port devices, a one-port calibration assumes a good termination on the unused port of the. If this condition is met (by connecting a load standard, for example), the one-port calibration is quite accurate. However, if port two of the is connected to the network analyzer and its reverse isolation is low (for example, filter passbands or low-loss cables), the assumption of a good load termination is often not valid. In this case, two-port error correction can provide significantly better results than one-port correction. An amplifier is a good example of a two-port device in which the load match presented by the network analyzer does not affect measurements of the amplifier s input match, because the reverse isolation of the amplifier allows one-port calibration to be effective. In Figure 3, a reflection measurement is shown with and without a one-port calibration. Without error correction, the classic ripple pattern appears, which is caused by systematic errors interfering with the test signal. The error-corrected trace is much smoother and better represents the device s actual reflection performance. Return Loss (db) 0 20 40 Data Before Error Correction 2.0..0 VSWR 60 Data After Error Correction.00 6000 2000 MHz Figure 3. Before and After One-Port Calibration The Effects of Adapters Ideally, reflection calibrations should be performed with a calibration kit having the same type connectors as the. If adapters are necessary to make connections, the effects of these adapters must then be considered as part of the measurement uncertainty. An adapter added to a network analyzer test port after a calibration has been done may cause errors that add to or subtract from the desired signal from the (Figure 4). This error is often ignored, which may not be acceptable. Worst-case effective directivity in this case is the sum of the corrected directivity and the reflection (r) of the adapter. An adapter with a VSWR of.5: for example, will reduce the effective directivity of a test coupler to about 4 db, even if the coupler itself has infinite directivity. So with an ideal load on the output of the adapter, the reflected signal appearing at the coupled port will be only 4 db less than the reflection from a short or open circuit. Stacking multiple adapters compounds the problem. If adapters cannot be avoided, the highest-quality types are always the best choice in order to reduce degradation of system directivity. Error correction can mitigate the effects of adapters on the test port, but the test system will be slightly more susceptible to drift because of degraded raw (uncorrected) directivity. 5

leakage signal Coupler directivity = 40 db reflection from adapter desired signal ρ total Adapter = ρ adapter + Termination ρ has SMA (f) connectors Worst-case System Directivity Adapting from APC-7 to SMA (m) APC-7 calibration done here 28 db APC-7 to SMA (m) SWR:.06 7 db 4 db APC-7 to N (f) + N (m) to SMA (m) SWR:.05 SWR:.25 APC-7 to N (m) + N (f) to SMA (f) + SMA (m) to (m) SWR:.05 SWR:.25 SWR:.5 Figure 4. Adapter Considerations Two-Port Error Correction Two-port error correction yields the most accurate results because it accounts for all of the major sources of systematic error. The error model for a two-port device reveals the four S-parameters measured in the forward and reverse directions (Figure 5). Once the system error terms have been characterized, the network analyzer utilizes four equations to derive the actual device S-parameters from the measured S-parameters. Because each S-parameter is a function of all four measured S-parameters, a network analyzer must make a forward and reverse test sweep before updating any one S-parameter. When performing a two-port calibration, the part of the calibration that characterizes crosstalk (isolation) can often be omitted. Crosstalk, which is signal leakage between test ports when no device is present, can be a problem when testing high-isolation devices such as a switch in the open position, and highdynamic-range devices such as filters with a high level of rejection. Forward model Reverse model Port EX Port 2 Port Port 2 ERT' a b E S E D S 2 A S S A 22 A E TT E L a 2 b 2 a b E L' E TT' S2 A S A S22 A E S' ED' S 2 A b 2 a 2 E RT S2 A EX' E = Fwd Directivity D E = Fwd Source Match S E RT = Fwd Reflection Tracking E D' = Rev Directivity E S' = Rev Source Match E RT' = Rev Reflection Tracking E = Fwd Load Match L E = Fwd Transmission Tracking TT E X = Fwd Isolation E = Rev Load Match L' ETT' = Rev Transmission Tracking E X' = Rev Isolation Sm ED S m E D S E E S RT E S E m E X S m E 22 ' 2 2 X ' ( )( + ') L ( )( ) RT ' E TT E TT ' = a S m E D S E m E D S E S E RT E S E L E m E X S m E + + 22 ' 2 2 X ' ( )( ') ' L ( )( ) RT ' E TT E TT ' S2m EX S22 m E D ' ( )( + ( E E S TT E S ' E L )) RT ' 2 = a S m E D S E m E + D E S + 22 ' S ' ( )( E RT E S ') E L ' E ( 2 m E X S )( 2 m E X ) RT ' L E TT E TT ' Notice that each actual S-parameter is a function of all four measured S-parameters Analyzer must make forward and reverse sweep to update any one S-parameter S E ' S E ( 2m X )( + m D ( E ' )) E TT ' E S E L S RT 2 a = S ' ' ( m E D S E )( m E D S ') ' ( )( ) E S E RT E RT ' S E L E m E X S m E 22 2 2 + + X L E TT E TT ' S22 ' ( m E D S )( m E D S ' E ) '( )( ) E RT ' E S E 2 m E X S2 m E X + L ' S RT E TT E 22 a = TT S ( m E + D S E m E D ' S E S E RT E S E L E m E X S m E X ' )( 22 ') ' L ( 2 )( 2 + ) RT ' E TT E TT ' Figure 5. Two-Port Error Correction 6

Unfortunately, a crosstalk calibration can add noise to the error model because measurements are often made near the analyzer s noise floor. If the isolation calibration is deemed necessary, it should be performed with trace averaging to ensure that the test system s crosstalk is not obscured by noise. In some network analyzers, crosstalk can be minimized by using the alternate sweep mode instead of the chop mode (the chop mode makes measurements on both the reflection (A) and transmission (B) channels at each frequency point, whereas the alternate mode turns off the reflection receiver during the transmission measurement). The best way to perform an isolation calibration is by placing the devices that will be measured on each test port of the network analyzer, with terminations on the other two device ports. Using this technique, the network analyzer sees the same impedance versus frequency during the isolation calibration as it will during subsequent measurements of the. If this method is impractical (in test fixtures, or if only one is available, for example), than placing a terminated on the source port and a termination on the load port of the network analyzer is the next best alternative (the and termination must be swapped for the reverse measurement). If no is available or if the will be tuned (which will change its port matches), then terminations should be placed on each network analyzer test port for the isolation calibration. A network analyzer can be used for uncorrected measurements, or with any one of a number of calibration options, including response calibrations and one- or two-port vector calibrations. A summary of these calibrations is shown in Figure 6). UNCORRECTED RESPONSE ONE-PORT FULL TWO-PORT SHORT SHORT SHORT Convenient Generally not accurate No errors removed Other errors: Random (Noise, Repeatability) Drift thru Easy to perform Use when highest accuracy is not required Removes frequency response error ENHANCED-RESPONSE Combines response and -port Corrects source match for transmission measurements OPEN LOAD For reflection measurements Need good termination for high accuracy with two-port devices Removes these errors: Directivity Source match Reflection tracking OPEN LOAD thru OPEN LOAD Highest accuracy Removes these errors: Directivity Source, load match Reflection tracking Transmission tracking Crosstalk Figure 6. Errors and Calibration Standards Electronic Calibration Although Figure 6 shows mechanical calibration standards, Agilent Technologies offers a solid-state calibration solution which makes two-port calibration fast, easy, and less prone to operator errors. The various impedance states in the calibration modules are switched with PIN-diode or FET switches, so the calibration standards never wear out. The calibration modules are characterized at the Agilent factory using a TRL-calibrated network analyzer, making the ECal modules transfer standards (rather than direct standards). ECal provides excellent accuracy, with results generally better than SOLT calibration, but somewhat less than a properlyperformed TRL calibration. 7

Estimating Measurement Uncertainty Figure 7 shows which systematic error terms are accounted for when using analyzers with transmission/reflection test sets (Agilent 872ET family, the 8753ET and the 8720ET family), and S-parameter test sets (Agilent 8753ES, 8720ES, Agilent 850 family and the PNA Series). Some straightforward techniques can be used to determine measurement uncertainty when evaluating two-port devices with a network analyzer based on a transmission/reflection test set. For example, Figure 8 shows measurement of the input match of a filter after a one-port calibration has been performed. The filter has 6 db of return loss and db of insertion loss. The raw load match of an 872ET network analyzer is specified to be 8 db (although it s often significantly better than this). The reflection from the test port connected to the filters output port is attenuated by twice the filter loss in this case, only 2 db. This value is not adequate to sufficiently suppress the effects of this error signal, which illustrates why low-loss devices are difficult to measure accurately. Reflection Test Set (cal type) T/R S-parameter (one-port) (two-port) SHORT Reflection tracking Directivity Source match Load match error can be corrected error cannot be corrected * Agilent 872ET enhanced response cal can correct for source match during transmission measurements Transmission Transmission Tracking Crosstalk Source match Load match OPEN * ( ) LOAD Test Set (cal type) T/R S-parameter (response, (two-port) isolation) Figure 7. Calibration Summary Load match: 8 db (.26) Directivity: 40 db (.00) 6 db RL (.58) db loss (.89).58 (.89)(.26)(.89) =.00 Measurement uncertainty: 20 * log (.58 +.00 +.00) =.4 db ( 4.6dB) 20 * log (.58.00.00) = 26.4 db (+0.4 db) Figure 8. Reflection Example Using a One-Port Cal 8

To determine the measurement uncertainty of this example, it is necessary to add and subtract the undesired reflection signal (with a reflection coefficient of 0.00) with the signal reflecting from the (0.58) (to be consistent with the next example, we will also include the effect of the directivity error signal). The measured return loss of the 6-dB filter may appear to be anywhere from.4 db to 26.4 db, allowing too much room for error. In production testing, these errors could easily cause filters which met specification to fail, while filters that actually did not meet specification might pass. In tuning applications, filters could be mistuned as operators try to compensate for the measurement error. When measuring an amplifier with good isolation between output and input (i.e., where the isolation is much greater than the gain), there is much less measurement uncertainty. This is because the reflection caused by the load match is severely attenuated by the product of the amplifier s isolation and gain. To improve measurement uncertainty for a filter, the output of the filter must be disconnected from the analyzer and terminated with a high-quality load, or a high-quality attenuator can be inserted between the filter and port 2 of the analyzer. Both techniques improve the analyzer s effective load match. As an example (Figure 9), if we placed a 0 db attenuator with a SWR of.05 between port 2 of the network analyzer and the filter used in the previous example, our effective load match would improve to 28.6 db. This value is the combination of a 32.3 db match from the attenuator and a 38 db match from the network analyzer (since the error signal travels through the attenuator twice, the analyzer s load match is improved by twice the value of the attenuator). Our worst-case uncertainty is now reduced to +2.5 db,.9 db, instead of the +0.4 db, 4.6 db we had without the 0 db attenuator. While not as good as what could be achieved with two-port calibration, this level of accuracy may be sufficient for manufacturing applications. Load match: 8 db (.26) Measurement uncertainty: 20 * log (.58 +.039) = 4. db (.9 db) 20 * log (.58.039) = 8.5 db (+2.5 db) Directivity: 40 db (.00).58 0 db attenuator (.36) SWR =.05 (.024) 6 db RL (.58) db loss (.89) (.89)(.36)(.26)(.36)(.89) =.00 (.89)(.024)(.89) =.09 Worst-case error =.00 +.00 +.09 =.039 Low-loss bi-directional devices generally require two-port calibration for low measurement uncertainty Figure 9. Reflection Example using a One-port Cal plus an Attenuator 9

Performing a Transmission Response Calibration Response calibrations offer simplicity, but with some compromise in accuracy. In making a filter transmission measurement using only response calibration, the first step is to make a through connection between the two test ports (with no in place). For this example, test port specifications for the Agilent 872ET network analyzer will be used. The ripple caused by this amount of mismatch is calculated as ±0.22 db, and is now present in the reference data (Figure 0). It must be added to the uncertainty when the is measured in order to compute worst-case overall measurement uncertainty. The same setup and test port specifications for the 872ET can be used to determine the measurement uncertainty with the in place. There are three main error signals caused by reflections between the ports of the analyzer and the (Figure ). Higher-order reflections can be neglected because they are small compared to the three main terms. One of the error signals passes through the twice, so it is attenuated by twice the insertion loss of the. A worst-case condition occurs when all of the reflected error signals add together in phase (0.020 + 0.020 + 0.032 = 0.072). In that case, measurement uncertainty is +0.60/ 0.65 db. Total measurement uncertainty, which must include the 0.22 db of error incorporated into the calibration measurement, is about ±0.85 db. RL = 4 db (.200) RL = 8 db (.26) Thru calibration (normalization) builds error into measurement due to source and load match interaction Calibration Uncertainty = ( ± ρ ρ ) S L = ( ± (.200)(.26) = ± 0.22 db Figure 0. Transmission Example Using a Response Cal Source match = 4 db (.200) 6 db RL (.58) Load match = 8 db (.26) (.26)(.58) =.020 (.58)(.200) =.032 Total measurement uncertainty: +0.44 + 0.22 = +0.66 db 0.46 0.22 = 0.68 db Measurement uncertainty = ± (.020+.032) = ±.052 = + 0.44 db 0.46 db Figure. Transmission Example (continued) 0

Another test example is an amplifier with a port match of 6 db. The test setup and conditions remain essentially the same as in the first two cases (Figure 2), except now the middle error term is no longer present because of the amplifier s reverse isolation. This reduces the measurement error to about ±0.45 db and the total measurement uncertainty to about ±0.67 db (compared to ±0.85 db for the filter). Source match = 4 db (.200) 6 db RL (.58) Load match = 8 db (.26) (.26)(.58) =.020 (.58)(.200) =.032 Total measurement uncertainty: +0.44 + 0.22 = +0.66 db 0.46 0.22 = 0.68 db Measurement uncertainty = ± (.020+.032) = ±.052 = + 0.44 db 0.46 db Figure 2. Measuring Amplifiers with a Response Calibration Enhanced-Response Calibration for Transmission Measurements A feature of the 872ET/ES family of economy network analyzers is their ability to perform an enhanced-response calibration. This calibration requires the measurement of short, open, load, and through standards for transmission measurements. The enhanced-response calibration combines a one-port calibration and a response calibration to allow correction of source match during transmission measurements, something a standard response calibration cannot do. The enhanced-response calibration (Figure 3) improves the effective source match during transmission measurements to about 35 db, compared to 4 db for normal response calibrations with the 872ET. This reduces the calibration error from ±0.22 db to ±0.02 db, and greatly reduces the two measurement error terms that involve interaction with the effective source match. The total measurement error is ±0.24 db instead of the previous value of ±0.85 db for a standard response calibration. While not as good as full two-port error correction, this represents a significant improvement over a standard response calibration and may be sufficient for many applications. Source match = 35 db (.078) Effective source match = 35 db! db loss (.89) 6 db RL (.58) Load match = 8 db (.26) (.26)(.58) =.020 (.26)(.89)(.078)(.89) =.008 Calibration Uncertainty = ( ± ρ ρ ) S L = ( ± (.078) (.26) = ±.02 db (.58)(.078) =.0028 Measurement uncertainty = ± (.020+.008+.0028) = ±.0246 = + 0.2 db 0.26 Total measurement uncertainty: 0.22 +.02 = ± 0.24 db Figure 3. Transmission Measurements using the Enhanced-Response Calibration

Full Two-Port Calibration In an example that calculates the measurement error after a two-port calibration (Figure 4), the worst-case measurement errors for the filter have been reduced to about ±0.5 db for reflection measurements and ±0.05 db for transmission measurements. Phase errors are similarly small. Corrected error terms: (8753D.3-3 GHz Type-N) Directivity = 47 db Source match = 36 db Load match = 47 db Refl. tracking =.09 db Trans. tracking =.026 db Isolation = 00 db db loss (.89) 6 db RL (.58) Reflection uncertainty S m = S a ± (E D + S a 2 E s + S 2a S 2a E L + S a ( E RT )) = 0.58 ± (.0045 + 0.58 2 *.058 + 0.89 2 *.0045 + 0.58*.0022) = 0.58 ±.0088 = 6 db + 0.53 db, 0.44 db (worst-case) Transmission uncertainty S 2m = S 2a ± S 2a (E I / S 2a + S a E S + S 2a S 2a E S E L + S 22a E L + ( E TT )) = 0.89± 0.89 (0 6 / 0.89 + 0.58*.058 + 0.89 2 *0.58*.0045 + 0.58*.0045 +.003) = 0.89 ±.0056 = db ± 0.05 db (worst-case) Figure 4. Calculating Measurement Uncertainty after a Two-Port Calibration TRL Calibration Following SOLT in popularity, the next most common form of two-port calibration is called a Through-Reflect-Line (TRL) calibration. It is primarily used in noncoaxial environments, such as testing waveguide, using test fixtures, or making on-wafer measurements with probes. TRL uses the same 2-term error model as a SOLT calibration, although with different calibration standards. TRL has two variants: True TRL calibration, which requires a network analyzer with four receivers TRL* calibration, developed for network analyzers with only three receivers Other variations of TRL are based on Line-Reflect-Match (LRM) calibration standards or Through-Reflect-Match (TRM) calibration standards. In differentiating TRL and TRL*, the latter assumes that the source and load match of a test port are equal that there is true port-impedance symmetry between forward and reverse measurements. This is only a fair assumption for a three-receiver network analyzer. TRL* requires 0 measurements to quantify 8 unknowns. True TRL requires four receivers (two reference receivers plus one each for reflection and transmission) and 4 measurements to solve for 0 unknowns. Both techniques use identical calibration standards. The Agilent 8720ES network analyzer, which is normally equipped with only three samplers, can perform TRL calibrations when outfitted with Option 400 (which adds a fourth sampler). In noncoaxial applications, TRL achieves better source match and load match corrections than TRL*, resulting in less measurement error. In coaxial applications, SOLT is usually the preferred calibration technique. While not commonly used, coaxial TRL can provide more accuracy than SOLT, but only if very-high quality coaxial transmission lines (such as beadless airlines) are used. 2

Calibrating Noninsertable Devices When performing a through calibration, normally the test ports mate directly. For example, two cables with the appropriate connectors can be joined without a through adapter, resulting in a zero-length through path. An insertable device may substituted for a zero-length through. This device has the same connector type on each port but of the opposite sex, or the same sexless connector on each port, either of which makes connection to the test ports quite simple. A noninsertable device is one that can not be substituted for a zero-length through. It has the same type and sex connectors on each port or a different type of connector on each port, such as waveguide at one end and a coaxial connector on the other end. There are a few calibration choices available for noninsertable devices. The first is to use a characterized through adapter (electrical length and loss specified), which requires modifying the calibration kit definition. This will reduce (but not eliminate) source and load match errors. A high-quality through adapter (with good match) should be used since the match of the adapter cannot be characterized. Swap-Equal-Adapters Method The swap-equal-adapters method requires the use of two precision matched adapters that are equal in performance but have connectors of different sexes. To be equal, the adapters must have the same match, characteristic impedance, insertion loss, and electrical delay. Many of Agilent s calibration kits include matched adapters. The first step in the swap-equal-adapters method is to perform a transmission calibration with the first adapter (Figure 5). Following this, the first adapter is removed and the second adapter is placed on port 2. This second adapter then becomes the effective test port. The reflection calibration is then performed on both test ports. Following this, the is measured with the second adapter in place. The errors remaining after calibration are equal to the difference between the two adapters. The technique provides a high level of accuracy, but not quite as high as the more complicated adapter-removal technique. Port Port 2 Accuracy depends on how well the adapters are matched loss, electrical length, match and impedance should all be equal Port Adapter Port 2 A. Transmission cal using adapter A. Port Adapter Port 2 B 2. Reflection cal using adapter B. Port Adapter Port 2 B 3. Measure using adapter B. Figure 5. Swap-Equal-Adapters Method 3

Adapter-Removal Calibration Adapter-removal calibration provides the most complete and accurate calibration procedure for noninsertable devices (Figure 6). This method uses a calibration adapter that has the same connectors as the noninsertable. The electrical length of the adapter must be specified within one-quarter wavelength at each calibration frequency. Type-N, 3.5-mm, and 2.4-mm calibration kits for the Agilent 850 network analyzer contain adapters specified for this purpose. Two full two-port calibrations are needed for an adapter-removal calibration. In the first calibration, the precision calibration adapter is placed on the analyzer s port 2 and the test results are saved into a calibration set. In the second calibration, the precision calibration adapter is placed on the analyzer s port and the test data is saved into a second calibration set. Pressing the adapter-removal calibration softkey causes the network analyzer to use the two sets of calibration data to generate a new set of error coefficients that remove the effects of the calibration adapter. At this point, the adapter can be removed and the vector analyzer is ready to measure the. Uses adapter with same connectors as Adapter's electrical length must be specified within /4 wavelength adapters supplied with Type-N, 3.5-mm, and 2.4-mm cal kits are already defined for other adapters, measure electrical length and modify cal-kit definition Calibration is very accurate and traceable See product note 850-3 for more details Port Port 2 Cal Port Adapter Adapter Port 2 B Cal Set Cal Port Adapter Adapter Port 2 B Cal Set 2 [CAL] [MORE] [MODIFY CAL SET] [ADAPTER REMOVAL]. Perform two-port cal with adapter on port 2. Save in cal set. 2. Perform two-port cal with adapter on port. Save in cal set 2. 3. Use ADAPTER REMOVAL to generate new cal set. Port Adapter B Port 2 4. Measure without cal adapter. Figure 6. Adapter-Removal Calibration Suggested Reading Understanding the Fundamental Principles of Vector Network Analysis, Agilent application note 287-. Exploring the Architectures of Network Analyzers, Agilent application note 287-2. Network Analyzer Measurements: Filter and Amplifier Examples, Agilent application note 287-4. Specifying Calibration Standards for the Agilent 850 Network Analyzer, Agilent product note 850-5A. Applying the Agilent 850 TRL Calibration for Non-Coaxial Measurements, Agilent product note 850-8A. 4

5

By internet, phone, or fax, get assistance with all your test and measurement needs. Online Assistance www.agilent.com/find/assist Phone or Fax United States: (tel) 800 452 4844 Canada: (tel) 877 894 444 (fax) (905) 206 420 Europe: (tel) (3 20) 547 2323 (fax) (3 20) 547 2390 Japan: (tel) (8) 426 56 7832 (fax) (8) 426 56 7840 Latin America: (tel) (305) 269 7500 (fax) (305) 269 7599 Australia: (tel) 800 629 485 (fax) (6 3) 920 5947 New Zealand: (tel) 0 800 738 378 (fax) (64 4) 495 8950 Asia Pacific: (tel) (852) 397 7777 (fax) (852) 2506 9284 Product specifications and descriptions in this document subject to change without notice. Copyright 997, 2000 Agilent Technologies Printed in U.S.A. July 25, 200 5965-7709E 6