OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING

Similar documents
OPEN TEM CELLS FOR EMC PRE-COMPLIANCE TESTING

Failing EMC testing? > 50% of products fail EMC testing first time around

Modulated Wideband Power Amplifier

Radiated noise measurement example using Tekbox TEM Cells

EMC Near-field Probes + Wideband Amplifier

Demo / Application Guide for DSA815(-TG) / DSA1000 Series

Trees, vegetation, buildings etc.

V1.3. TBLC08 50mH AC-LISN TBLC08

AMPLIFIER RESEARCH... APPLICATION NOTE: 23

Immunity Test System RIS 3000 / RIS 6000 acc. to IEC/EN

AMPLIFIER RESEARCH... APPLICATION NOTE: 20

RIGOL Presents: New Solutions for Affordable Pre- Compliance Testing

FFT 3010 EMI TEST RECEIVER

Conducted emission pre compliance measurements

Conducted emission measurement using the Tekbox 5µH LISN TBOH01

INSTRUCTION MANUAL TRI-PLATE LINE MODEL EM-7310

50 μh Line Impedance Stabilisation Network (AC LISN)

Keysight Technologies Essential Capabilities of EMI Receivers. Application Note

Advanced Test Equipment Rentals ATEC (2832)

ER55 EMI TEST RECEIVER Family of automatic test receivers for measurement of electromagnetic interference from 9kHz to 1GHz

Calibration and Validation for Automotive EMC

EXHIBIT 7: MEASUREMENT PROCEDURES Pursuant 47 CFR 2.947

Electromagnetic Compliance: Pre-Compliance Conducted Emissions Testing October 19, 2017

Title: Test on 5.8 GHz Band Outdoor WiFi (802.11b/g) Wireless Base Station

Application Note: Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

7. Transmitter Radiated Spurious Emissions and Conducted Spurious Emission

Swept Return Loss & VSWR Antenna Measurements using the Eagle Technologies RF Bridge

Siglent Technologies SSA3021X Spectrum Analyzer and TG-SSA3000X Tracking Generator Reviewed by Phil Salas AD5X

Conducted emission pre compliance measurements

Spectrum Analyzers 2680 Series Features & benefits

Essential Capabilities of EMI Receivers. Application Note

CONDUCTED RF EQUIPMENT POWER AMPLIFIERS. Practical RF Immunity System Design Considerations

Product Description. Theory of operation

ER55 EMI TEST RECEIVER Family of automatic test receivers for measurement of electromagnetic interference from 9kHz to 2.8GHz.

Specification for Radiated susceptibility Test

AMPLIFIER RESEARCH... APPLICATION NOTE: 19

Utilizzo del Time Domain per misure EMI

DSA700 Series Spectrum Analyzer

Advanced Test Equipment Rentals ATEC (2832)

EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass. EN 55015: 2013 Clause Pass

DSA-815 Demo Guide. Solution: The DSA 800 series of spectrum analyzers are packed with features.

87415A microwave system amplifier A microwave. system amplifier A microwave system amplifier A microwave.

Specification for Conducted Emission Test

Signal Detection with EM1 Receivers

A Study of Conducted-Emission Stable Source Applied to the EMC US and EU Standards

L.S. Compliance, Inc. W66 N220 Commerce Court Cedarburg, WI

Measurement of Digital Transmission Systems Operating under Section March 23, 2005

Chambers Accessories Equipment 1 Equipment 2 Amplifiers Antennas Emission

EMC Amplifiers Going Beyond the Basics to Ensure Successful Immunity Tests

Electromagnetic Compatibility ( EMC )

EMC TEST REPORT For MPP SOLAR INC Inverter/ Charger Model Number : PIP 4048HS

RF Emissions Test Report To Determine Compliance With: FCC, Part 15 Rules and Regulations

Chapter 5 Specifications

Data Sheet SC5317 & SC5318A. 6 GHz to 26.5 GHz RF Downconverter SignalCore, Inc. All Rights Reserved

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2011

DSA800. No.1 RIGOL TECHNOLOGIES, INC.

Understanding Mixers Terms Defined, and Measuring Performance

Page 1 of 51 Report No.: T TEST REPORT FCC ID: 2AGJ5WAP-30. In Accordance with: FCC PART 15, SUBPART C : 2015 (Section 15.

Electromagnetic Compliance: Pre-Compliance Test Basics October 19, 2017

EMC/EMI MEASURING INSTRUMENTS & ACCESSORIES SHORT-FORM CATALOG 2009

Sunlight Supply, Inc.

Federal Communications Commission Office of Engineering and Technology Laboratory Division

Spectrum Analyzer R&S FS300

772D coaxial dual-directional coupler 773D coaxial directional coupler. 775D coaxial dual-directional coupler 776D coaxial dual-directional coupler

Ave output power ANT 1(dBm) Ave output power ANT 2 (dbm)

Microwave. Accessories for Microwave Scalar and System Analyzers

Rigol DSA705 Spectrum Analyzer Reviewed by Phil Salas AD5X

TETRA Tx Test Solution

Xeta4. Occupied Bandwidth. Measurements

Advanced Compliance Solutions, Inc FAU Blvd, Suite 310 Boca Raton, Florida (561)

2.9GHz SPECTRUM ANALYZER

EFFECT OF SHIELDING ON CABLE RF INGRESS MEASUREMENTS LARRY COHEN

Comparison of IC Conducted Emission Measurement Methods

Revision history. Revision Date of issue Test report No. Description KES-RF-14T0042 Initial

FCC Test Report. Report No.: PTCDQ FC01

Palstar, Inc. EMC TEST REPORT FOR. HF LDMOS Amplifier Model: LA-1K. Tested To The Following Standard: FCC Part 97 Subpart D. Report No.

FISCHER CUSTOM COMMUNICATIONS, INC.

PMM 7010 EMI RECEIVERS. The EMI Receiver with built-in LISN

2620 Modular Measurement and Control System

Spectrum Analyzers. 2394A 1 khz to 13.2 GHz Spectrum Analyzer.

PARAMETER CONDITIONS TYPICAL PERFORMANCE Operating Supply Voltage 3.1V to 3.5V Supply Current V CC = 3.3V, LO applied 152mA

For. Unit D16/F. should not use it to claim FCC ID: 2AAIN-MNGLOS

Description RF Explorer RFEAH-25 1 is a 25mm diameter, high performance near field H-Loop antenna.

STUDIO TO TRANSMITTER LINKING SYSTEM

Vector Network Analyzers ZVB

Agilent N9343C Handheld Spectrum Analyzer (HSA)

Signal Hound USB-SA44B 4.4 GHz Spectrum Analyzer and USB-TG44A Tracking Generator

Verifying Simulation Results with Measurements. Scott Piper General Motors

Return Loss Bridge Basics

Features 100A400AM20. 4kHz 400MHz Class A Portable Full VSWR tolerant CE & RoHS compliant High Efficiency

Handheld Spectrum Analyzer R&S FSH khz to 3 GHz

To «Test_Standards» Test of: Radwin Ltd. Outdoor Subscriber Radio Unit. To: FCC CFR 47 Part 15B; ICES-003 Issue 6: 2016

Handheld Spectrum Analyzer R&S FSH 3

TECHNICAL SPECIFICATION FOR RF (TEST &MEASUREMENT) DEVICE

Handheld Spectrum Analyzer R&S FSH3

khz to 2.9 GHz Spectrum Analyzer

Nemko-CCL, Inc West Alexander Street Salt Lake City, UT

Analysis of RF transceivers used in automotive

TEST REPORT : 2AIB7MGS. TEST DATE : to

REMARKS The customer has ordered two measurements mentioned above according to the standard EN

Transcription:

1 Introduction Radiated emission tests are typically carried out in anechoic chambers, using antennas to pick up the radiated signals. Due to bandwidth limitations, several antennas are required to cover the complete frequency range. Furthermore, it requires much space and the cost of the equipment for a standard conformant setup is immense. An engineer of a small or medium size enterprise usually has to rely on his experience and on best practice methods in order to design an EMC compliant product. Nevertheless, it is estimated that > 50% of products fail testing first time around. Anytime an engineer sends a new product for compliance testing, it is a shot in the dark. Failing is very expensive. Not only that re-testing costs are high, but also the project schedule and market introduction gets delayed. What is needed is an affordable laboratory set up to measure radiated emissions in the own lab, prior to compliance testing. A TEM cell is the right piece of equipment for desktop testing of radiated emissions. Tekbox developed open TEM cells to cover the complete frequency range up to 2GHz and with usability even at frequencies beyond. Combined with a spectrum analyzer, products can be tested before and after EMC related design modifications. A set up with a TEM cell will not deliver exactly the same quantitative results as a measurement in a certified test house, however it will give an excellent indication on whether the design suffers from excessive radiated noise or not. The engineer will clearly see, whether his changes improved or deteriorated the EMC performance or whether it remained unchanged. Using TEM cells eliminates the guesswork. 2 TEM cell A TEM cell is a stripline device for radiated emissions and immunity testing of electronic devices. It is not a replacement, but due to its size and cost it is a convenient alternative to measurements in an anechoic chamber. A TEM cell consists of a septum, the conductive strip in the centre section and walls which are connected to ground. The geometry is designed to present a 50Ω stripline. The device under test (DUT) is placed in between the bottom wall and the septum. The TBTC1/2/3 are so called open TEM cells, which got no side walls for convenient placement of the DUT. It may pick up RF background noise, which however can be taken into account by doing a measurement of the cell output signal before powering on the DUT. Tekbox open TEM cells got a better frequency response compared to standard TEM cells of similar size. TEM cells suffer from higher order wave modes which limit the usable bandwidth. A unique design feature of the Tekbox TEM cells implements resistance perpendicular to the desired propagation direction of the wave. Consequently higher order wave modes and resonances are supressed. The device is supplied together with a 50Ω/25W RF termination and a DC block to protect the spectrum analyzer or RF receiver input. Application of TEM cells Radiated emission tests: The septum of the TEM cell picks up radiated noise from the DUT, similar to a broadband antenna, and presents it to the spectrum analyzer or to a receiver input. 1

Radiated immunity tests: The TEM cell will be connected to the output of a swept signal generator + RF amplifier. The septum radiates the RF signal into the DUT. The RF signal is typically amplitude or pulse modulated. Picture 1 radiated emission measurement of a controller board using the TBTC1 TEM cell and a Rigol DSA815 spectrum analyzer Complete solutions Following additional equipment may be used for a complete and low cost setup: Radiated emission testing: Spectrum analyzers such as the RIGOL DSA815 or SIGLENT SSA 3021X RF immunity testing: RF power amplifiers from MINICIRCUITS A suitable amplifier for immunity testing of industrial products is the Mini Circuits, model ZHL-2. It can deliver up to 0.8W (126V/m) and covers the frequency range 10 MHz to 1 GHz. http://www.minicircuits.com/pdfs/zhl-2.pdf Examples for suitable amplifiers for RF immunity testing of automotive products are the Mini Circuits, model ZHL- 20W-13+ and the LZY-22+ http://www.minicircuits.com/pdfs/zhl-20w-13+.pdf http://www.minicircuits.com/pdfs/lzy-22+.pdf For lower electrical field levels, refer to the datasheet of the Tekbox TBMDA1 modulated amplifier, which can be driven by the tracking generator output of a spectrum analyzer. 2

3 EMC pre-compliance testing with the - radiated emissions Combined with a spectrum analyzer, products can be tested before and after EMC related design modifications. The engineer will clearly see, whether his changes improved or deteriorated the EMC performance or whether it remained unchanged. Tem cells eliminate the guesswork. Terminate any of the two ports of the TEM cell with the 50Ω load, connect the DC-Block to the other port and connect it to the input of the spectrum analyzer or test receiver. Place the DUT under the septum and power it on. Reduce the input attenuation of the analyzer to 0dB and if available turn on the pre-amplifier to increase the dynamic range of the measurement. Reduce the resolution bandwidth to 9 khz or whatever specified in the corresponding standard. Set the trace to max hold and let the analyzer run as many sweeps as may be necessary to catch pulsed or sporadic signals from your DUT. EMC measurements are typically started by performing scans using a peak detector to find problem areas. Spurious suspected to exceed the limits should then be measured using a quasi-peak detector, if available on your receiver or spectrum analyzer. A quasi peak detector measures the weighted peak value of the envelope of a signal. Signals are weighted according to their duration and repetition rate. Signals that occur less frequently will result in a lower quasi peak value than frequent signals. Due to the nature of this measurement method, quasi peak measurements consume much more time than simple peak measurements and are often carried out with reduced span in frequency regions with high amplitude spurious. The TEM cell can be very effectively used to identify potential issues that may result in failing the compliance test. As a rule of thumb, given that the PCB is positioned not much higher than 1-2cm above the bottom wall, any spurious with amplitudes higher than 40dµV are a threat. A quick check to ensure that spurious are not picked up from outside is done by powering off the DUT. Despite its construction, the TEM cell is not very sensitive to signals from outside. However cables connected to the DUT may pick up background noise and guide it into the TEM cell. This signals are usually also present with the DUT powered off and consequently can be easily identified. Take a screenshot of the radiated spectrum, modify your design, and check if it improved. Any reduction of spurious level measured in the TEM cell, will also translate into a reduction of the spurious level measured in the test house. When failing a compliance test, measure it in the TEM cell and save the results before carrying out any modifications. Compare the spurious levels documented in the compliance test report versus the spurious levels measured in the TEM cell. Check how many db s are missing to pass the test. Modify/improve the device until the spurious level measured inside the TEM cell dropped accordingly. Picture 2 radiated emission measurement set up 3

Example of radiated emission measurements before and after modification Picture 3 TBTC1, radiated emissions of a LED lamp prototype Picture 4 TBTC1, radiated emissions of a LED lamp prototype after modification. The spurious level in the measured frequency range could be reduced by 10dBµV. The spurious crossing 30dBµV was identified as picked up from outside. 4

4 EMC pre-compliance testing with the TBTC1/2/3 - immunity to radiated signals The E-field (V/m) between septum and lower (upper) wall of the TBTC1 is E = V/d where V is the RMS voltage of the applied signal and d is the distance between septum and lower (upper) wall. This is based on the simplified assumption that the E field would be perfectly homogenous/evenly distributed. A more practical formula is E = V*Cor/d where Cor is a correction factor for the average field strength over the volume of the DUT derived from the analysis of the field distribution over the cross section of the cell. Assuming the DUT is placed in the center of the cell and in the middle between bottom wall and septum, we can however use the simplified formula with sufficient accuracy. d = 2.8 cm d = 5 cm d = 10 cm d = 15 cm E = ( (P*50Ω))*35.7 E = ( (P*50Ω))*20 E = ( (P*50Ω))*10 E = ( (P*50Ω))*6.66 TBTC0, applied RF power Maximum field strength between septum and wall 10W (40 dbm) 799 V/m 1 W (30 dbm) 253 V/m 0.1 W (20 dbm) 82 V/m 0.01 W (10dBm) 25 V/m Table 1 TBTC0, field strength vs. RF power TBTC1, applied RF power Maximum field strength between septum and wall 10W (40 dbm) 447 V/m 1 W (30 dbm) 141 V/m 0.1 W (20 dbm) 44 V/m 0.01 W (10dBm) 14 V/m Table 2 TBTC1, field strength vs. RF power TBTC2, applied RF power Maximum field strength between septum and wall 10W (40 dbm) 224 V/m 1 W (30 dbm) 71 V/m 0.1 W (20 dbm) 22 V/m 0.01 W (10dBm) 7 V/m Table 3 TBTC2, field strength vs. RF power 5

TBTC3, applied RF power Maximum field strength between septum and wall 10W (40 dbm) 148 V/m 1 W (30 dbm) 47 V/m 0.1 W (20 dbm) 14 V/m 0.01 W (10dBm) 5 V/m Table 4 TBTC3, field strength vs. RF power Terminate any of the two ports of the TEM cell with the 50Ω RF-termination, and connect the other port to a swept frequency generator / RF power amplifier. Place the DUT under the septum and power it on. Sweep the frequency and observe the behavior of the DUT. The field decreases with approximately 30dB/m outside the cell, hence it is uncritical to operate it in ordinary facilities. Picture 5 RF immunity measurement set up Picture 3 shows a minimum set-up for immunity testing. Test houses would furthermore insert a directional coupler in between the power amplifier and the TEM Cell and monitor forward and reflected power. EMC standards differ largely with respect to the applied field strength. Standards for industrial electronics such as EN61000-6-4 specify 10V/m (prior to the application of modulation) whereas individual standards of automotive manufacturers may specify a field strength as high as 400V/m. The required field strength determines the requirements for the RF power amplifier as listed in tables 1 to 3. Certain standards specify the RF signal to be amplitude modulated, others specify pulse modulation. 6

5 Technical data TEM Cell TBTC0 TEM cell dimensions: Length: 390 mm Width: 100 mm Height: 62 mm Septum height: 28 mm Rectangular area under the septum: 19 cm x 7 cm x 2.8 cm TEM cell connectors: N-female Nominal cell impedance: 50 Ohm Wave impedance: 377 Ohm Maximum RF input power: 10W (limited by supplied 50 Termination) Input return loss: S11 up to 3.15 GHz < -15dB Transmission loss: up to 3 GHz < 3 db, up to 6 GHz < 4dB TEM Cell TBTC1 TEM cell dimensions: Length: 390 mm Width: 200 mm Height: 108 mm Septum height: 50 mm Rectangular area under the septum: 19 cm x 13 cm x 5 cm TEM cell connectors: N-female Nominal cell impedance: 50 Ohm Wave impedance: 377 Ohm Maximum RF input power: 25W (limited by supplied 50 Termination) Input return loss: S11 up to 1.2 GHz < -20dB, up to 2.1 GHz < -17dB, up to 3GHz < -14dB Transmission loss: up to 1.4 GHz < 1 db, up to 2.1 GHz < 3dB, up to 3 GHz < 6dB TEM Cell TBTC2 TEM cell dimensions: Length: 636 mm Width: 300 mm Height: 205mm Septum height: 100 mm Rectangular area under the septum: 23 cm x 28 cm x 10 cm TEM cell connectors: N-female Nominal cell impedance: 50 Ohm Wave impedance: 377 Ohm Maximum RF input power: 25W (limited by supplied 50 Termination) Input return loss: S11 up to 800 MHz < -15dB, up to 1.5 GHz < -10dB, up to 3GHz < -14dB Transmission loss: up to 800 MHz < 1 db, up to 1.15 GHz < 3dB 7

TEM Cell TBTC3 TEM cell dimensions: Length: 1038 mm Width: 501 mm Height: 305mm Septum height: 150 mm Rectangular area under the septum: 36 cm x 48 cm x 15 cm TEM cell connectors: N-female Nominal cell impedance: 50 Ohm Wave impedance: 377 Ohm Maximum RF input power: 25W (limited by supplied 50 Termination) Input return loss: up to 700 MHz < -16 db Transmission loss: up to 730 MHz < 3dB DC-Block 50V-6GHz-N Connectors: N-Male/Female Nominal impedance: 50 Ohm Max. continuous RF power: 2W Max. continuous RF voltage: 50V RMS Frequency: 500kHz to 6 GHz VSWR: 1.2 Insertion loss: Frequency [MHz] 0.3 0.7 1 10 100 1000 3000 Insertion loss [db] 7.5 3 1.5 0.02 0.02 0.06 0.2 Table 1 DC-Block frequency response RF-Termination 50Ω-3GHz-25W-N Connector: N-Male Nominal impedance: 50 Ohm Max. continuous RF power: 25W Frequency: DC to 3 GHz VSWR: 1.2 Third order intermodulation: 120 dbc RF-Termination 50Ω-6GHz-10W-N Connector: N-Male Nominal impedance: 50 Ohm Max. continuous RF power: 25W Frequency: DC to 6 GHz VSWR: 1.2 Third order intermodulation: 120 dbc 8

Picture 6 TBTC0, input return loss Picture 7 TBTC0, transmission loss 9

Picture 8 TBTC1, input return loss Picture 9 TBTC1, transmission loss 10

Picture 10 TBTC2, input return loss Picture 11 TBTC2, transmission loss 11

Picture 12 TBTC3, input return loss Picture 13 TBTC3, transmission loss 12

6 Warning Keep the DUT insulated from the septum and cell walls. Insert the DC-block to get additional input protection for the spectrum analyzer or measurement receiver. 7 Ordering Information Part Number Description TBTC0 TBTC1 TBTC2 TBTC3 Open TEM cell, 28mm septum height, Termination 50Ω-6GHz-10W-N, DC-Block 50V-6GHz-N, N-Male to N-Male coaxial cable Open TEM cell, 50mm septum height, Termination 50Ω-3GHz-25W-N, DC-Block 50V-6GHz-N, N-Male to N-Male coaxial cable Open TEM cell, 100mm septum height, Termination 50Ω-3GHz-25W-N, DC-Block 50V-6GHz-N, N-Male to N-Male coaxial cable Open TEM cell, 150mm septum height, Termination 50Ω-3GHz-25W-N, DC-Block 50V-6GHz-N, N-Male to N-Male coaxial cable Table 2 Ordering Information 8 History Version Date Author Changes V1.0 14.04.2016 Mayerhofer Creation of the document V1.1 30.06.2016 Mayerhofer Update TBTC3 12.02.2016 Mayerhofer Update TBTC0 Table 3 History 13