Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals

Similar documents
The Effect of Changing the Applied Sequence Using the TVC on the Accuracy of the AC Signal Calibration

AC-DC TCC with Built-in Tee Connector for Accurate Calibrations

Performance of the Single Junction Thermal Voltage Converter at 1 MHz via Equivalent Circuit Simulation

IMPROVEMENTS IN THE NIST CALIBRATION SERVICE FOR THERMAL TRANSFER STANDARDS

CALIBRATION PROCESS QUANTITY REDUCTION OF THE THERMAL VOLTAGE CONVERTER STANDARD USING A THREE-STAGE BUILD-UP AND BUILD-DOWN METHOD

Study of the Long Term Performance on the Calibration Data of the Coaxial Thermistor Mounts up to 18 GHz

An improvement for dual channel sampling wattmeter

PXIe Contents. Required Software CALIBRATION PROCEDURE

Certification Cover Sheet Accredited Calibration Order Number:

A New Method for the Calibration of the mv Ranges of an AC Measurement Standard

The Metrology Behind Wideband/RF Improvements to the Fluke Calibration 5790B AC Measurement Standard

AC Resistance Thermometry Bridges and their Advantages By Peter Andrews

Realization and traceability of AC power standard at frequency of 50 Hz

DC Voltage Linearity Measurements and DVM Calibration with Conventional and Programmable Josephson Voltage Standards

FINAL REPORT. EUROMET project No Bilateral comparison of DC and AC voltages BEV - NCM. W. Waldmann (BEV, pilot laboratory) P.

50 OHM MULTIJUNCTION THERMAL CONVERTERS FOR AC VOLTAGE MEASUREMENTS UP TO 100 MHZ

Calibration of 100 MΩ Hamon resistor using current-sensing Wheatstone bridge. Ivan Leniček 1, Roman Malarić 2, Alan Šala 3

Establishing a High-precision AC/DC Transfer Standard using ET2001 ADS System

5790A Automated AC Measurement Standard

5500A. Multi-Product Calibrator. Extended Specifications 2005

CCEM-K9 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

For the National Voluntary Laboratory Accreditation Program

AC-DC TRANSFER DIFFERENCE STANDARDS AND CALIBRATIONS AT NRC CANADA

Using Reference Multimeters for Precision Measurements

The Fluke 8508A. Reference Multimeter. Reference standard accuracy and stability, in one functionally versatile, easy to use solution

Fluke 5820A Oscilloscope Calibrator Specifications

Comparison of the Josephson Voltage Standards of the INTI and the BIPM

Certificate of Calibration

TR CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF ELECTRICAL DCLF METROLOGY. Approved By: Senior Manager: Mpho Phaloane Revised By:

Multimeter Selection Guide Fluke 8508A & Agilent 3458/HFL

2-3 Calibration of Standard Voltage and Current Generator

Use of the BVD for traceability of bipolar DC voltage scale from 1 mv up to 1200 V

5520A. Multi-Product Calibrator. Extended Specifications 2005

Comparison of the NIST and NRC Josephson Voltage Standards (SIM.EM.BIPM-K10.b)

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Calibration Laboratory Assessment Service CLAS Certificate Number Page 1 of 10

CERTIFICATE of CALIBRATION

Verifying the Wideband Input of an AC Measurement Standard

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

Cost-Effective Traceability for Oscilloscope Calibration. Author: Peter B. Crisp Head of Metrology Fluke Precision Instruments, Norwich, UK

Precision in Practice Achieving the best results with precision Digital Multimeter measurements

Manual Supplement. This supplement contains information necessary to ensure the accuracy of the above manual.

Comparison of the Josephson Voltage Standards of the SMU and the BIPM

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

5520A. Multi-Product Calibrator. Extended Specifications

Comparison of the Josephson Voltage Standards of the NIMT and the BIPM

CERTIFICATE OF ACCREDITATION ISO/IEC 17025:2005 ANSI/NCSL Z (R2002)

Calibration Technique for SFP10X family of measurement ICs

An Evaluation of Artifact Calibration in the 5700A Multifunction Calibrator

Certificate of Calibration

Standard Resistors Precision DC Shunts Digital Thermometers Non-inductive AC Shunts Intelligent Teraohmmeter Precision Decade Standards Precision Air

Contents. CALIBRATION PROCEDURE NI 5421/ MS/s Arbitrary Waveform Generator

8000 SERIES PRECISION MULTIMETER VERIFICATION AND ADJUSTMENT GUIDE

A COMPLEX CURRENT RATIO DEVICE FOR THE CALIBRATION OF CURRENT TRANSFORMER TEST SETS

Supplementary Information

ADC Based Measurements: a Common Basis for the Uncertainty Estimation. Ciro Spataro

Determination of Uncertainty for Dielectric Properties Determination of Printed Circuit Board Material

Discipline Electro-Technical Calibration Issue Date Certificate Number C-0556 Valid Until Last Amended on - Page 1 of 7

RF-POWER STANDARD FROM AC-DC THERMAL CONVERTER. L. Brunetti, L. Oberto, M. Sellone

Final Report Key Comparison COOMET.EM.BIPM-K10.b. Comparison of the 10 V Josephson Voltage Standards COOMET 542/RU/11. A.S. Katkov 1, P.A.

About the power standards calibration in the National Institute of Metrology, Romania

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

Comparison of the Josephson Voltage Standards of the DMDM and the BIPM

CRITERIA FOR LABORATORY ACCREDITATION IN THE FIELD OF TIME AND FREQUENCY METROLOGY

Keysight Technologies Evaluation of the Performance of a State of the Art Digital Multimeter

Validation Of The New Automatic System For AC Voltage Comparisons. Umberto Pogliano, Gian Carlo Bosco and Marco Lanzillotti

ACLASS Accreditation Services An ANSI-ASQ National Accreditation Board Company

Contents CALIBRATION PROCEDURE NI PXI-5422

Model 1140A Thermocouple Simulator-Calibrator

Power Quality Measurements the Importance of Traceable Calibration

CERTIFICATE OF ACCREDITATION

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSLI Z

Traceability and Modulated-Signal Measurements

Technical Reference for ET2001 Thermal Voltage Converters

Cost Effective Techniques used to Validate the Performance of the microk Resistance Thermometry Instrument with sub mk Uncertainty

A Guide for establishing primary AC-DC transfer standard using ET2001 ADS system

e th2 e th1 e th4 e th3

SCXI 8-Channel Isolated Analog Input Modules

Comparison of the Josephson Voltage Standards of the CENAM and the BIPM (part of the ongoing BIPM key comparison BIPM.EM-K10.b)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z

5790B/5/AF AC Measurement Standard

Multifunction Calibrator - Model 8080

Contents CALIBRATION PROCEDURE NI 5412

Development of thin-film multi-junction thermal converters

Ac-dc transfer standard shunts for frequencies up to 1 MHz. U. Pogliano, C.G. Bosco, M. Lanzillotti, D Serazio

Sept 13 Pre-lab due Sept 12; Lab memo due Sept 19 at the START of lab time, 1:10pm

CALIBRATION AND MEASUREMENT CAPABILITY (CMC) 1,2

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

Report on Bilateral Comparison P1-APMP.EM.BIPM-K11.1 Bilateral Comparison of dc voltage. 1 Introduction and general conditions of the comparison

EUROMET 557 COMPARISON OF AC-DC HIGH VOLTAGE STANDARDS

PXIe Contents. Required Software CALIBRATION PROCEDURE

THERMAL NOISE. Advanced Laboratory, Physics 407, University of Wisconsin. Madison, Wisconsin 53706

Schedule of Accreditation issued by United Kingdom Accreditation Service 2 Pine Trees, Chertsey Lane, Staines-upon-Thames, TW18 3HR, UK

N432A Thermistor Power Meter DATA SHEET

Precision Measurement Solutions

Lab E2: B-field of a Solenoid. In the case that the B-field is uniform and perpendicular to the area, (1) reduces to

1 Ω 10 kω High Precision Resistance Setup to calibrate Multifunction Electrical instruments

Transcription:

Investigation of Two Different Techniques for Accurate Measurements of Sinusoidal Signals Shereen M. El-Metwally 1, Mamdouh Halawa 2 1 Department of Systems and Biomedical Engineering, Cairo University, Giza, Egypt 2 National Institute for Standards (NIS), Giza, Egypt, 1 shereen.elmetwally@yahoo.com 2 mamdouh_halawa@yahoo.com Abstract In this paper, two different techniques for the determination of the rms value of sinusoidal AC voltage at the range of 2V are practically compared: non-thermoelectric technique (via the direct measurement using a precise high sensitive digital voltmeter, and the comparison method using a precise AC voltage calibrator) and thermoelectric technique (via the null thermoelectric technique, and the algorithmic thermoelectric technique using a calibrated multijunction thermal converter (MJTC)). A series of comparisons between the four methods have been discussed and evaluated at frequencies from 20 Hz to 100 khz to investigate the level of performance and competitiveness. The comparison results showed that the four applied methods agreed within values of 80µV ± 15µV/V and 68µV ± 11µV/V at frequencies of 20 Hz and 100 khz, respectively. Keywords- AC voltage measurement, AC-DC thermal transfer, uncertainty budget, proficiency test. I. INTRODUCTION Recently, the accurate AC voltage measurement in the frequency range up to 100 khz becomes more important and commercial parameter in electrical metrology laboratories and national calibration services. The past few years have seen an improvement in AC voltage calibrators and voltmeters, which would make them acceptable as standards in their own right except for the very highest accuracy. As a result, the specification for the most accurate ranges (1V to 10V) in the frequency range from 100 Hz to 10 khz has improved by a factor of two and the frequency range of commercial instruments has been extended up to at least 50 MHz. This gives an annually accuracy of about 20 ppm [1]. This improvement is due to various developments in each of the thermal and non-thermal techniques used for comparing AC and DC voltages and currents. This paper describes four different methods for the accurate determination of the rms value of 2V at frequencies from 20 Hz to 100 khz. As all four methods have their advantages and disadvantages, a set of comparisons including the measurement results associated with the expanded uncertainty, the frequency response and the uncertainty budgets of each method is also investigated in this work. In addition, a proficiency testing methodology, to evaluate the compatibility between the four methods, has been applied. II. NON-THERMOELECTRIC TECHNIQUE The new generation of AC instrument has increased the demand on accuracy. To meet the demand in the AC voltage measurements, a new generation of the high sensitive and very accurate digital multimeter (DMM) has been developed. DMM is one of the most widely used electrical measurement instruments and provides electrical traceability to many industrial users. However, digital multimeters designers have always had to exert efforts to minimize the parameters contributing into the values of errors and associated uncertainty. The AC function has the added dimension of frequency. This complicates calibration by introducing additional test points for each amplitude range. In a typical DMM, the AC measurements are made by an AC converter and available with 8½ digit 100 mv ranges with a resolution of 1 nv. Below this level, noise and linearity errors are likely to dominate the reading. The gain of a DMM's AC function varies with frequency. This is called its "Frequency Response" and requires that measurements are made at the cardinal points throughout each amplitude range [2]. An easy way to comply with the conference paper formatting requirements is to use this document as a template and simply type your text into it. A. Direct Method of Measurement In the technique adopted in this method, a FLUKE 8508A Reference Multimeter is used to measure the rms value of the applied sinusoidal signal in accordance with the corresponding standard [3]. In the 24 hours preceding calibration, the instrument was powered by the mains and placed in the appropriate laboratory ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 12

conditions. A WAVETEK 9100 Calibration System was used to apply the AC voltage signal for the DMM. During this process, the ambient temperature was controlled at 21 ± 1 o C and the relative humidity at 45 ± 5%. Typical ten observations of the supplied signal value from the calibrator is measured, corrected and averaged using the 8½ digit multimeter. As stated in [4], all measurements are tainted by imperfectly known errors, so the significance associated with the result of a measurement must account for this uncertainty. The uncertainty values were estimated taking into account most known contributing factors to the uncertainty of the undertaken measurement. The reported expanded uncertainty of measurement is stated as the standard uncertainty of measurement multiplied by a coverage factor, k = 2, corresponding to coverage probability of approximately 95% [5]. The standard uncertainty of measurement has been determined in accordance with [4]. During this method, the following contributions are taken into account: Uncertainty due to the repeatability of 10 observations (Type A). Uncertainty due to the calibration certificate of the DMM. Uncertainty due to the display resolution of the DMM. Uncertainty due to the drift between two consecutive calibration certificates of the DMM. The calibration results of 2V using this method associated with the corresponding expanded uncertainty are listed in Table I. TABLE I Calibration Results of Direct Technique (khz) Actual Value (V) Deviation (mv) Exp. Uncert. (µv/v) 0.02 1.996277 3.7233 23 0.055 1.996285 3.7152 18 0.4 1.996432 3.5682 19 1 1.996400 3.6 17 2 1.996334 3.6665 17 10 1.996390 3.61 17 30 1.996187 3.813 19 50 1.995424 4.5759 20 100 1.993153 6.8469 20 B. Comparison Method of Measurement In this technique, a calibrated FLUKE 5720A Calibrator has been used to provide the reference value of the 2V at the intended frequencies. On the other hand, WAVETEK 9100 Calibration System is used to apply the nominal value of the 2V, to be measured, at the same frequencies and in the same conditions. A FLUKE 8508A Reference Multimeter is then used, as in Fig. 1, to display and compare the two outputs of each calibrator accordingly. Typical ten observations are also recorded, corrected and averaged associated with the expanded uncertainty (at k=2) as listed in Table II. The following contributions were taken into consideration in this method: Uncertainty due to the repeatability of 10 observations of the two readings of the DMM. Uncertainty due to the calibration certificate of the 5720A Calibrator. Uncertainty due to the display resolution of the DMM. Uncertainty due to the drift between two consecutive calibration certificates of the 5720A calibrator. Since the digital voltmeter is used for comparison only, its display deviation is not relevant. Therefore, the calibration of the DMM did not contribute to the uncertainty budget. ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 13

(khz) Fig. 1. Comparison Method TABLE II Calibration Results of Comparison Technique Actual Value (V) Deviation (mv) Exp. Uncert. (µv/v) 0.02 1.996249 3.7513 21 0.055 1.99624 3.7603 15 0.4 1.996356 3.644 16 1 1.996419 3.581 12 2 1.996338 3.662 12 10 1.996354 3.6457 12 30 1.996171 3.8288 16 50 1.995476 4.5236 17 100 1.993086 6.9142 17 III. THERMOELECTRIC TECHNIQUE Indeed, the thermoelectric technique is one of the most precise methods to determine the rms value of an AC voltage in terms of the SI units accurately. To achieve that, it is necessary to use the AC-DC thermal transfers [6]. In this work, a calibrated Multijunction Thermal Converter (MJTC) at the level of 2V was used during two methods to determine the rms value of the 2V at the same range of frequencies. The AC-DC transfer difference (δ) of the MJTVC is known with an uncertainty less than 4 µv/v. A. Nulling Method The automated setup of AC measurement using this method is shown in Fig. 2. The automation software [7] developed for this method can produce the calibration results along with standard deviation at each point of measurement. Fig. 2. Nulling Method ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 14

In this method, the nominal value of 2V is supplied by the WAVETEK 9100 Calibrator to the MJTC and the corresponding response of 10 typical observations in terms of output emf (E ac ) is recorded and averaged by using a Keithley 182 nanovoltmeter. Then DC positive and DC negative voltage of the value of calibrated 2V is applied to produce the same exact value of the emf (E ac ) then the corresponding applied voltages V dc positive and V dc negative are recorded and the mean value of V dc is then calculated. The actual value of AC voltage at this frequency is then determined by: V = 1 + δ ) (1) ac V dc ( s where δ s is the AC-DC transfer difference of the MJTC that was already determined in PTB, Germany with values of less than 5.5 µv/v within uncertainty less than 4 µv/v. The following contributions are taken into consideration to estimate the expanded uncertainty in this method: Uncertainty due to the repeatability of 10 observations for each effect of AC and DC. Uncertainty due to the calibration certificate of the MJTC and 5720A Calibrator. Uncertainty due to the thermal emf effect of the cables used in this method. Uncertainty due to the drift between two consecutive calibration certificates of the MJTC and 5720A Calibrator. The calibration results associated with the corresponding expanded uncertainty (at k = 2), are listed in Table III. (khz) TABLE III Calibration Results of Nulling Technique Actual Value (V) Deviation (mv) Exp. Uncert. (µv/v) 0.02 1.996197 3.80338 10 0.055 1.996253 3.7474 9 0.4 1.996361 3.63875 9 1 1.996374 3.6264 9 2 1.996382 3.6179 9 10 1.996394 3.605866 9 30 1.996224 3.775652 9 50 1.995459 5.410847 9 100 1.993153 7.547287 11 B. Algorithmic Method This technique was developed by Dr. Ilya Budovisky (NMI Australia) [8]. In the principle of AC-DC Transfer, the AC-DC Difference, δ ac-dc, is usually defined as [7]: V V AC DC δ ac dc = (2) V DC where: V AC = rms value of AC voltage that produce a certain value of output emf. V DC = average of the absolute values of DC voltage applied in positive and negative polarity to produce the same emf. The relation between the input current of the applied voltage on the MJTC and its output emf is given by: E = K V n (3) where K varies somewhat with large changes in heater current but it is constant over a narrow range where nearly equal AC & DC voltage are compared and n is usually 1.6 to 1.9 at the rated heater current [9] (in our ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 15

case n 2). The relationship between a small change in TE heater voltage ( V) and the corresponding change in output ( E) is expressed as: Δ V V = Δ E n. E From (2), (3) and (4), the AC-DC difference can be defined as: E E δ = ac d c n. E dc where E ac is the mean value of the two outputs of the thermoelement due to the AC voltage and E dc is the average of the two outputs of the thermoelement due to the forward and the reverse DC voltage. The absolute AC voltage of this signal at this frequency is then determined using the formula: V = 1+ δ + δ ) (6) ac V dc ( ac dc s where δ s is the AC-DC transfer difference of the calibrated MJTC. The following contributions are taken into consideration to estimate the expanded uncertainty in this method: Uncertainty due to the repeatability of 10 observations for the effect of AC and DC. Uncertainty due to the calibration certificate of the MJTC and 5720A Calibrator. Uncertainty due to the thermal emf of the used cables. Uncertainty due to the drift between two consecutive calibration certificates of the MJTC and 5720A Calibrator. Uncertainty due to the short-term stability of the AC and DC sources (over 3 minutes for both) The calibration results along with corresponding expanded uncertainty (at k = 2), are listed in Table IV. (4) (5) (khz) TABLE IV Calibration Results of Algorithmic Technique Actual Value (V) Deviation (mv) Exp. Uncert. (µv/v) 0.02 1.99622916 3.77084 8 0.055 1.996263021 3.736979 7 0.4 1.996391865 3.608135 7 1 1.99640148 3.59852 7 2 1.996370711 3.629289 8 10 1.996384096 3.615904 8 30 1.996215991 3.784009 8 50 1.995442726 4.557274 8 100 1.99313784 6.86216 9 IV. COMPARISON OF RESULTS Fig. 3 shows the results of the comparison between the four methods: direct method (M1), comparison method (M2), nulling method (M3), and algorithmic method (M4), by computing the relation between the actual measured values taken using the four methods and the corresponding frequency. This relation also reflects the frequency dependence of the four methods at constant voltage. ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 16

Fig. 3. Frequency dependence of the four methods As noted in Fig. 3, the frequency response of the four methods has, approximately, the same trend. In this response, results tend to become much less with the increase in the frequency. The least measured value of the 2 V is achieved at 100 khz. Generally, the calibrated results among the four methods exhibit differences of 80µV and 68µV at 20 Hz and 100 khz respectively. On the other hand, the results of the measured deviation of the 2V along with expanded uncertainty have been compared (as shown in Figures 4 and 5). Referring to this comparison, it is noticed that the algorithmic method (M4) exhibits the lowest expanded uncertainty compared to other methods. In addition, the methods based on the thermoelectric technique exhibit lower values in comparison to the other methods based on non-thermoelectric technique. As shown in Fig. 6, the two techniques agree within less than 15µV/V and 11µV/V at frequencies of 20 Hz and 100 khz, respectively. Fig. 4. Comparison of the four methods at 20 Hz Fig. 5. Comparison of the four methods at 100 khz * Note: A graph scale of 100:1 for the expanded uncertainty and the deviation respectively was used to represent the values in Figs 4 and 5. ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 17

Fig. 6. Expanded Uncertainties of the Four Methods Despite of the longer time consumed as well as the difficult procedures, it has been observed that the thermoelectric technique (null and algorithmic methods) has a smaller standard deviation for the set of 10 measurements in comparison to the other technique. In addition, the values of the expanded uncertainty in the majority of the measuring points are also smaller. This feature, as a result, recommends the use of thermoelectric technique in the high level of precision measurements. V. COMPATIBILITY AND COMPETITIVENESS OF THE FOUR METHODS To confirm the compatibility of the four methods, and according to the previous comparison, the algorithmic method has been recommended to represent the reference values of this comparison. Table V lists the differences between the measured values using methods 1, 2 and 3 and using method 4 (as the reference value) and coded as 1-4, 2-4 and 3-4, respectively. The results agree within a range of 19.5 to 47.5µV and 14.8 to 52µV at 20 Hz and 100 khz respectively. On the other hand, Table VI reports the proficiency testing results of this comparison. Proficiency testing is the determination of the performance by means of comparing and evaluating calibrations by two or more methods in accordance with predetermined conditions [10]. The performance of the three methods against the algorithmic method, (as a reference value) is judged using the equation of: E n X X L R = (7) 2 2 U L + U R where: X L = the value as measured by the compared method, X R = the value as measured by the reference method (algorithmic), U L = the expanded uncertainty of the compared method, U R = the expanded uncertainty of the reference method. En ratio (or En number) should be between -1 and + 1 (or En <1), (the value closer to zero is the better.) TABLE V Differences from the reference value Differences (µv) (khz) 1-4 2-4 3-4 0.02 47.54 19.54-32.54 0.055 21.779-23.321-10.421 0.4 39.935-35.865-30.615 1-1.48 17.52-27.88 2-37.211-32.711 11.389 10 5.904-29.796 10.038 30-28.991-44.791 8.357 50-18.626 33.674 16.427 100 15.26-52.04 14.873 ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 18

Fig. 7. Differences of M1, M2 and M3 vs M4 TABLE VI E n Ratio of the comparison E n Ratio (khz) M 1 M2 M3 0.02 0.95 0.30 0.92 0.055 0.53 0.51 0.34 0.4 0.91 0.73 0.94 1 0.04 0.40 0.89 2 0.92 0.70 0.35 10 0.15 0.67 0.30 30 0.63 0.97 0.24 50 0.42 0.70 0.53 100 0.34 0.95 0.42 Referring to the values of E n in Table VI, it can be noticed that E n values are less than 1 for all the measuring points. This means that all results of the all methods are competent and satisfactory. VI. CONCLUSION The calibration results of a precision AC calibrator at level of 2V using two different techniques: nonthermoelectric technique (via Direct and Comparison methods) and thermoelectric technique (via Null and Algorithmic methods) have been practically compared and evaluated at frequencies from 20 Hz to 100 khz. The four methods agreed within 80µV ± 15µV/V and 68µV ± 11µV/V at 20 Hz and 100 khz respectively. It has been observed that the thermoelectric technique has comparatively smaller combined uncertainty contribution. Although the thermoelectric technique is time consuming and complex, even in automation calibration, it was more accurate and precise. The standard deviation and the expanded uncertainty of the algorithmic method both were less than the same values of the other methods. The four methods agree within expanded uncertainties less than 15µV/V and 11µV/V at frequencies of 20 Hz and 100 khz respectively. The algorithmic method was recommended to represent the reference value and to confirm the performance compatibility and reliability of the other three methods via the proficiency testing analysis. The calculated ratio numbers of all results of the three methods were acceptable and satisfied. REFERENCES [1] R G Jones, "A Review of Precision AC Voltage and Current Measurements," IEE, Savoy Place, London WCPR OBL, UK 1997. [2] Peter B. Crisp, "A Generic DMM Test and Calibration Strategy," IEE, Savoy place, London WCPR 0BL, UK, 1996. [3] Guidelines on the Calibration of Digital Multimeters, EURAMET/cg-15/v.01, (previously EA-10/15, European Association of National Metrology), July 2007. [4] The Expression of Uncertainty and Confidence in Measurement, M 3003, Edition 1, UKAS, December 1997. [5] Expressions of the Uncertainty of Measurements in Calibration, EA-4/02, (previously EAL-R2), Apr. 1997. [6] Marian Kampile, Hector Laiz and Manfred Klonz, "Comparison of Three accurate Methods to Measure AC Voltage at Low Frequency," IEEE Trans. Instrum. Meas., vol. 49, No. 2, April 2000. [7] M. Halawa and Najat El-Rashed, "Performance of the Single Junction Thermal Voltage Converter at 1 MHz via Equivalent Circuit Simulation," Journal of Cal. Lab., USA Sept. 2009. [8] Bijendra Pal, Saood Ahmad, A.k. Govil and P. Banerjee, "Automation and Evaluation of two Difference Techniques to Calibrate Precision Calibrators for LF Voltage & Current using Thermal Devices,", in Proc. of conference on Precision Electromagnetic Measurements, June 2010. [9] Thomas E. Lipe, "A Reevaluation of the NIST Low-Frequency Standards for AC-DC Difference in the Voltage Range 0.6-100 V," IEEE Trans Instrum. Meas., vol. 45, No. 6, DEC. 1996. [10] Jon M. Crickenberger, Calibration Laboratories Technical Guide, NIST HANDBOOK 150-2, Aug., 1995. ISSN : 0975-4024 Vol 5 No 1 Feb-Mar 2013 19