DAVIS ANEMOMETER MODBUS INTERFACE MODULE DA485. Manual. Pages 10

Similar documents
ROTRONIC HygroClip Digital Input / Output

WIND VANE THIES COMPACT

IP1 Datasheet PWM OUTPUT WITH SINGLE CHANNEL ADC MODULE FEATURES DESCRIPTION CONNECTOR DETAILS

The "FISH" Quad Hand Sensor

ANEMOMETER WINDSENSOR P2546C-OPR

Serial Communications RS232, RS485, RS422

L E C T U R E R, E L E C T R I C A L A N D M I C R O E L E C T R O N I C E N G I N E E R I N G

Lab Exercise 9: Stepper and Servo Motors

EL6483: Sensors and Actuators

CMPS09 - Tilt Compensated Compass Module

JUMO Wtrans Receiver with Wireless Data Transmission

WIND MONITOR ANEMO / WIND VANE YOUNG ALPINE

MT1531 CMOS, Programmable Linear Hall Effect Sensor

MT1531 Series. CMOS, Programmable Linear Hall Effect Sensor. Features. Applications. 1 / 15

FLOW RATE CORRECTION CALCULATOR GDR 1403

Line-to-line RMS Volts, 3 phases 4 digits (XXX.X) Volts

Specification GRUNER YSP 227C

X3M. Multi-Axis Absolute MEMS Inclinometer Page 1 of 13. Description. Software. Mechanical Drawing. Features

JUMO Wtrans Receiver Universal Receiver for JUMO Wireless Measuring Probes

WIND VANE THIES FIRST CLASS 2014 TMR

Voltage-Versus-Speed Characteristic of a Wind Turbine Generator

Single Channel Loop Detector

MADEinUSA OPERATOR S MANUAL. RS232 Interface Rev. A

1 UAT Test Procedure and Report

Channels that are not occupied by temperature sensors, can take over alternative functions:

J. La Favre Using Arduino with Raspberry Pi February 7, 2018

isma-b-w0202 Modbus User Manual GC5 Sp. z o.o. Poland, Warsaw

SHAFTED ROTARY POSITION SENSORS

TWEAK THE ARDUINO LOGO

MV110-8AS. Analog input module 8 channel. User guide

A COMPARISON STUDY OF THE COMMUTATION METHODS FOR THE THREE-PHASE PERMANENT MAGNET BRUSHLESS DC MOTOR

Analog Module D1-15P User Manual

2F. No.25, Industry E. 9 th Rd., Science-Based Industrial Park, Hsinchu, Taiwan Application Note of OGM220, AN001 V1.8

A Super trainer with advanced hardware and software features only found in very expensive equipment.

WIND VANE THIES COMPACT TMR

MV110-8A. Analog input module 8 channel. User guide

GE 320: Introduction to Control Systems

Simple Servo USER Instructions

FLOW RATE CORRECTION CALCULATOR GDR 1403

1W-H3-05 (K)* M12. * Letter K refers to a reader with a common cathode. RFID reader 125 khz Unique. Product Card

Modbus communication module for TCX2: AEX-MOD

Serial Communication AS5132 Rotary Magnetic Position Sensor

Shenzhen ATC Technology CO.,LTD ATC. A-1 Serial Remote I/O Module. User Manual. V1.13 Edit:2018/01/

MECE 3320 Measurements & Instrumentation. Data Acquisition

OHIO SEMITRONICS, INC. DIGITAL AC POWER MONITOR OSI SPECIFICATIONS

RoHS Directive-Compliant Compact Linear Actuators DRL Series. Features

ma/bcd Converter Model MFC-201 ma/bcd The MFC-201 ma/bcd Converter is a device designed to convert a DC current signal into BCD (8421) signals.

5-Phase Stepping Motor and Microstep Driver Package

METEOROLOGICAL INSTRUMENTS

PalmGauss SC PGSC-5G. Instruction Manual

About Security of the RAK DEK

Repeater R-IP12. (Version v ) User Manual

Signal Isolators & Conditioners

1-Wire Addressable Digital Instruments. for Environmental Monitoring

Mercury technical manual

ANEMOMETER VECTOR A100L2 & VECTOR A100LK & VECTOR A100LM

Artifex LIV 110. Laser Diode Characterization System. Engineering

parameter conditions/description min typ max units accuracy 0.2 degrees

Sensors Fundamentals. Renesas Electronics America Inc Renesas Electronics America Inc. All rights reserved.

ioselect Z-NET Z-4RTD 4 Channel RTD Input Module

PMSM Control Using a Three-Phase, Six-Step 120 Modulation Inverter

OPERATING INSTRUCTIONS

VFSC9 ELECTRONIC SPEED CONTROLLER. Mounting and operating instructions

EE 314 Spring 2003 Microprocessor Systems

METEOROLOGICAL INSTRUMENTS

LC-10 Chipless TagReader v 2.0 August 2006

KINAX SR 719 Programmable Transmitter for Position Feedback

ROLL TO ROLL FUNCTION MANUAL FR-A (0.4K)-04750(90K)-R2R FR-A (0.4K)-06830(280K)-R2R FR-A (315K)-12120(500K)-R2R

Compact Actuator DRL Series

Parameter Value Unit Notes

ORIENTAL MOTOR GENERAL CATALOG

Wind Direction Transmitter - compact - with Poti Output xx.712

I-7088, I-7088D, M-7088 and M-7088D User Manual

HD 48V01T... HD 4977T...

Rotary Encoder System Compact Model Range

Data Sheet AS25/AS50. Absolute / Incremental Singleturn Encoder 18 Bit

5. Transducers Definition and General Concept of Transducer Classification of Transducers

Design Implementation Description for the Digital Frequency Oscillator

Instruction Manual ABM HART Gateway Software. Instruction Manual Revision A.1

METEOROLOGICAL INSTRUMENTS

DS Absolute Position, Rotary Electric Encoder

Serial Servo Controller

Features. Description. General Specifications. VS Series Inclinometer : Dual Axis, RS232 and Analogue Output

Operation. Displayed channel. Measuring range. Status indication/ remote control Key lock Measuring mode/ time constant. Scale.

Frequency Converter Fv vector control converters for universal applications

Computer Numeric Control

Absolute multi-turn hollow shaft encoder BOMH Dignalizer SSI

TIP551. Optically Isolated 4 Channel 16 Bit D/A. Version 1.1. User Manual. Issue December 2009

B850 Boiler House Energy Monitor

UMK M A A. RoHS-Compliant. 2-Phase Stepping Motor and Driver Package. Features. Product Number Code. Product Line

SOLAR-360 : 360 Inclinometer, RS232 or RS485 Output

DC Electronic Loads 8500 Series

Measurement, Sensors, and Data Acquisition in the Two-Can System

Technical data. General specifications V DC Power consumption P 0. 1 W Time delay before availability t v

SilverMax Datasheet. QuickSilver Controls, Inc. NEMA 23 Servomotors.

745 Transformer Protection System Communications Guide

SMART Funded by The National Science Foundation

Magnetic Sensor - Incremental / Absolute WMSA50

Current Loop Interface Module M 2025D

VTMS. Valve Train Measurement Solution. Data sheet

Transcription:

DAVIS ANEMOMETER MODBUS INTERFACE MODULE DA485 Manual Pages 10

Сontent 1 Description and Operation of the Product... 3 1.1 Product Designation... 3 1.2 Technical Specifications... 3 1.3 Technology and Operation... 4 2 Supply Package... 7 Appendix A... 8 Computer Communication Protocol DA485... 8 А.1 Data structure for wind processing... 8 А.2 Converter setup registers... 9 А.3 Operation control... 9 А.4 Results registers (Modbus map)... 10 2

The purpose of this Manual is to get the users familiarized with the technology and operation principle of the Davis Anemometer 6410 interface module DA485 (hereinafter converter). An anemometer is a wind speed measuring device. Davis 6410 has included a wind direction vane but still refers to the entire instrument as an anemometer. Davis 6410 is shown in Figure 1. Figure 1 1 Description and Operation of the Product 1.1 Product Designation 1.1.1 The converter is designed to measure electrical signals that are proportional to air flow (wind) direction and speed. A computer with the software via the RS485 converter can be used to display the measured information. 1.1.2 The principle of the converter operation is based on measuring the voltage proportional to the wind direction and the frequancy proportional to the wind speed. The converter provides conversion of signals to physical parameters (wind speed and direction) by means of individual calibration factors. 1.1.3 The RS485 interface that provides access to the data using the Modbus-RTU protocol is implemented in the converter. 1.2 Technical Specifications 1.2.1 The converter provides automatic measurement of electrical parameters under the operating conditions of application within the ranges indicated in Table 1. 3

Table 1 Characteristics Values Frequency measurement range, Hz From 0 to 60 Voltage measurement range, V From 0 to 1,2 Digital interface RS485 A, B Power supply voltage, V Ftom 7 to 24 Useful current, ma 30 Operating conditions: - ambient temperature, С From minus 40 to 50 - relative humidity at the temperature of 25 С, % Up to 98 Average life time, years 8 Length, mm Width, mm Height, mm Mass, kg 95 50 25 0.053 1.3 Technology and Operation 1.3.1 The operation principle of the Davis 6410 anemometer is based on the relationship between the air flow speed and the number of wind cups rotations, as well as between the speed vector direction and the position of the freely turning wind vane. The anemometer and vane is a passive analog device. It is not powered. It responds to a brief direction excitation pulse from the DA485. 1.3.2 Refer to the schematic diagram in Figure 2 for the description. The flat (not twisted) cable has 4 conductors. yellow - direction excitation. This is an excitation pulse sent from the converter to the anemometer and is applied to the high end terminal of the potentiometer (CW). green - direction. This is the direction signal from the anemometer to the converter. This wire is connected to the potentiometer slider terminal of the potentiometer (S). The slider returns a portion of the excitation pulse depending on the angular position of the vane. red - common or analog ground. This wire is connected to the wind speed and the wind direction circuits. This is low end terminal of the potentiometer (CCW). black - wind speed signal. This wire provides the wind speed pulses from the magnetic reed switch to the converter. 4

Figure 2 The wind speed and wind direction functions have separate circuits but the red wire is common to both. The three-cup type impeller spinning induces switchover of the sensor digital output at each rotation, thus resulting in the generation of a sequence of pulses with a frequency proportional to the spin rate. The pulse sequence period is converted into the wind speed using the formula: ν = a/τ 2 + b/τ + с, where: ν wind speed; τ - pulse sequence period (τ = 1/f the value reciprocal of frequency); a, b, с - conversion factors obtained in the course of calibration. The freely spinning wind vane of the anemometer is set on the potentiometer shaft. As the wind direction changes, the vane follows and changes the resistance at the slider terminal of the potentiometer. The potentiometer is a linear resistance type that is free to rotate 360º with no mechanical stop. Rotation of the wind vane axis pin induces variation of electrical voltage that is proportionl to the wind vane direction as shown in Figure 3. Figure 3 The wind direction circuit uses a linear 20K Ohm potentiometer to sense the position of the vane. A voltage pulse is sent from the converter to the pot through the yellow wire. This voltage is applied to one end of the pot. The mechanical slider in the pot picks a portion of that voltage 5

depending on the angular position of the vane/slider. The voltage level of this pulse is determined by the vane/pot slider position. The voltage level is converted into the wind direction by the formula: α = 360 u / U, where: α wind vane direction in degrees; u the voltage of the slider; U an excitation pulse (supply voltage). 1.3.3 DA485 converts the measured values into physical values by the formulas described above, with the coefficients from the non-volatile memory. The formula for conversion of resistance into direction describes an ideal anemometer with a linear characteristic. In order to adapt the real anemometer Davis 6410, write the direction values in degrees at four directions of the wind vane - 0, 90, 180 and 270 degrees, to the converter non-volatile memory. 1.3.4 The external view of the converter is shown in Figure 4. Figure 4 1.3.5 The data from the converter are visualized in the user s data acquisition center. The converter is connected to the computer via the RS485-USB (RS485-Ethernet) converter. The exchange protocol is given in Appendix А. 1.3.6 The program provides for a special mode to setup the converter. In this mode the wind speed is determined independently of the wind direction. To switch to this mode, either set the parameter max to 0 in the setup file, or programmatically give the switch command as described in Appendix A. 1.3.7 Before turning on power, check the converter for external damage, then: connect the cable to the computer through the RS485 converter; connect the power adapter; 6

activate the console application Anemometer service from the supply package, as shown in Figure 5. The software and description of its operation are on the CD. Figure 5 1.3.8 Digital values that qualitatively characterize the ambient conditions of the room should appear on the display in the program window: wind speed - zero; wind direction - zero. The converter performance is tested by rotating the wind cups and changing the wind vane position. 2 Supply Package Table 2 No Component designation Identifier Quantity, pcs. 1 Converter DA485 1 2 Manual - 1 3 Compact disk CD Table 3 describes the configuration of the converter supplied. Table 3 Designation Communications port RS485 ( 19200, 8, 1, no parity) Address 7

А.1 Data structure for wind processing Appendix A Computer Communication Protocol DA485 The Modbus-RTU protocol is used for data exchange in DA485. Functions 3 and 4 are used for data reading, and functions 5 and 16 - for recording. The data structure used for setting up is provided below. The console application Anemometer service is used to display data from the Davis Anemometer 6410 via DA485. All the structure parameters are readable and writeable by means of the Modbus protocol functions. The console application uses the setup file to upload the data. typedef struct { _U8 object; // Modbus address _U8 max; // Maximum determination time (from 30 minutes to 24 hours) // max >100, max-100 minutes, max>200 max-200 hours, if // max<30min or max>24 hours, then - 1 hour; // max=0 for setting up: direction is always displayed // irrespective of the wind speed. _U16 id; //identifier (serial number) //****************************************************** _F32 ac[4]; // the values at the wind vane directions of 0, 90, 180 and 270 degrees _F32 mc[3]; // speed correction factors //****************************************************** _F32 fval[19]; // the values of wind speed and direction } eepromdata; The last 76 bytes of the data structure, 19 floating-point numbers [nineteen], are read-only. Each pair of data structure bytes corresponds to the Modbus protocol register with a shift of 10 registers (20 bytes), if the data are read with function 3. If function 4 is used for reading, the measurement results can be read, starting with the zero register. More details on the cross-reference between the data structure contents and the Modbus protocol registers will be described below in Tables 4, 5. Before using the numbers obtained, check their applicability for processing. Four-byte floating point numbers, where all the bits of all the four bytes are equal to 1, are considered unprocessable (no data, measurement errors, etc..). For verification, it is sufficient to compare the numbers in both registers that are part of the value under verification, with the number 65535 (0xFFFF hexadecimal) or all the 4 bytes with the number 255 (0xFF hexadecimal)). 8

А.2 Converter setup registers Table 4 Register Byte Number Number Structure Description 0 0 max Interval to determine maximum (0 at calibration) 1 object Modbus Address of the converter 1 2 3 id Identifier of the converter (serial number) 2 4 5 3 6 7 ac[0] Converter reading at the wind vane direction of 0 degrees 4 8 5 9 10 ac[1] Converter reading at the wind vane direction of 90 degrees 11 6 12 13 7 14 15 ac[2] Converter reading at the wind vane direction of 180 degrees 8 16 17 9 18 19 ac[3] Converter reading at the wind vane direction of 270 degrees 10 20 21 11 22 23 mc[0] 12 24 Quadratic polynomial coefficients to correct the wind speed module using 25 mc[1] 13 26 the formula: ν = mc[2]/τ 2 + mc[1]/τ + mc[0] (п.1.3.1) 27 14 28 29 15 30 31 mc[2] А.3 Operation control To reset the maxima, use register 22, into which the number 0 should be written by means of function 6, or register 0, into which the number 0 should be written by means of function 5. To switch to the set-up mode and back by means of function 5, write the number 0 to register 3, and to show the ADC code instead of direction, write 0 to register 2. 9

Table 5 А.4 Results registers (Modbus map) Register Number Byte Number Structure Parameter 10 20 21 11 22 23 12 24 25 13 26 27 14 28 29 15 30 31 16 32 33 17 34 35 18 36 37 19 38 39 20 40 41 21 42 43 22 44 45 23 46 47 24 48 49 25 50 51 26 52 53 27 54 55 28 56 57 29 58 59 30 60 61 31 62 63 32 64 65 33 66 67 34 68 69 35 70 71 36 72 73 37 74 75 38 76 77 39 78 79 40 80 81 41 82 83 42 84 85 43 86 87 44 88 89 45 90 91 46 92 93 47 94 95 fval[0] fval[1] fval[2] fval[3] fval[4] fval[5] fval[6] fval[7] fval[8] fval[9] Current wind speed Current wind direction Wind speed averaged over 10 minutes Wind direction averaged over 10 minutes Maximum wind speed during 3 hours Maximum direction during 3 hours Maximum wind speed during 10 minutes Maximum direction during 10 minutes Wind speed averaged over 2 minutes Wind direction averaged over 2 minutes fval[10] Maximum wind speed during the last 2 minutes fval[11] fval[12] fval[13] fval[14] fval[15] fval[16] fval[17] fval[18] Maximum direction during 2 minutes Wind speed averaged over 1 minute Wind direction averaged over 1 minute Maximum wind speed during the last minute Maximum direction during a minute Maximum wind speed since the reset moment Maximum direction since the reset moment Time in seconds from the 3-hour maximum to the current moment 10