Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Similar documents
LSM 800 Confocal Microscope Standard Operation Protocol

LSM 780 Confocal Microscope Standard Operation Protocol

Guide to Confocal 5. Starting session

LSM 710 Confocal Microscope Standard Operation Protocol

Simplified Instructions: Olympus Widefield Microscope S1230

Widefield 1. Switching on

Zeiss AxioImager.Z2 Brightfield Protocol

QUICKSTART GUIDE: WIDEFIELD WF3 Zeiss Cell Observer Live Cell Imaging System (SAF, ROOM 409) Imperial College London

SPINNING DISK CSU-X1 USER MANUAL

Leica SP8 TCS Users Manual

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

SHORT INSTRUCTIONS FOR OPERATING LSM1/2 (Zeiss LSM510) AT CIAN Version 1.4, September 2014

BX-61: Brightfield Instruction /Continue to scroll for Fluorescent Instuctions

Zeiss LSM 880 Protocol

3. are adherent cells (ie. cells in suspension are too far away from the coverslip)

Nikon E800 Operating Instructions.

Things to check before start-up.

EPIFLUORESCENCE &/OR BRIGHTFIELD MICROSCOPY

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

QUICKSTART GUIDE: WIDEFIELD HWF1 Zeiss Cell Observer Live Cell Imaging System (HAMMERSMITH, L BLOCK, ROOM 314) Imperial College London

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Zeiss Axio Imager.A1 manual

Olympus Time-lapse Microscope Basic operation

Zeiss LSM 780 Protocol

Title: Nikon A1R Confocal User Manual

Leica SP8 Resonant Confocal. Quick-Start Guide

Leica SPEII confocal microscope. Short Manual

Nikon SIM-E & A1-R System

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

Nikon E800 Operating Instructions.

Zeiss Axiovert 135 Fluorescence Microscope Quick Guide / Operations Manual (v. 1.0 February 09)

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

START-UP PROCEDURE 1 THE MICROSCOPE STAND 3 OBJECTIVES 5 STARTING WITH LAS (SOFTWARE) AND SETTING UP THE MICROSCOPE STAND 7

Leica SP8 TCS Users Manual

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Diskovery Spinning Disk Guide

Title: Leica SP5 Confocal User Manual

User manual for Olympus SD-OSR spinning disk confocal microscope

Using the Nikon TE2000 Inverted Microscope

ZEISS LSM510META confocal manual

Overview. About other software. Administrator password. 58. UltraVIEW VoX Getting Started Guide

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Use of the HSW5 Spinning Disk Confocal Microscope Updated last May 25, 2010 OK

Standard Operating Procedure (SOP) for Shared Equipment: Spinning Disk Confocal Microscope

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

b. Turn the power switch and key to on position for blue laser.

MIF ZEISS LSM510 CONFOCAL USER PROTOCOL

Dante (Microscope) & Beatrice (Guide) Orth Lab

REMEMBER: You have 5GB of disk space on this microscope. Check before you start if you have room for your experiment. If not delete your old data.

Leica Sp5 II Confocal User Guide

Contents STARTUP MICROSCOPE CONTROLS CAMERA CONTROLS SOFTWARE CONTROLS EXPOSURE AND CONTRAST MONOCHROME IMAGE HANDLING

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

personal DELTAVISION (pdv)

Nasmyth Ultraview Vox User Protocol

Olympus Fluoview 1000S Spectral Confocal Microscope Introduction to the NRI-MCDB Microscopy Facility Spectral Confocal Microscope

Zeiss Deconvolution Microscope: A Quick Guide

Nikon C1si Spectral Laser Scanning Confocal Microscope. User Guide

Characterization Microscope Nikon LV150

Quick Start Guide. Leica SP5 X

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

Olympus xcellence Software - basic user guide

Olympus IX71 Microscope and DP71 Camera Instructions

Nikon E800 Microscope. Operating Instructions

Last updated: May 2014 Y.DeGraaf

Swept-Field User Guide

AxioVision 4.5 Brightfield Image Capture Procedure

User manual for Nikon Elements software

LSM 510 Meta Training Notes

Operating Instructions for Zeiss LSM 510

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

DIC Imaging using Laser Scanning Microscopes (LSMs) on Axio Imager Stands

Practical work no. 3: Confocal Live Cell Microscopy

LSM 510 Training Notes

Confocal Raman Microscopy (WITec Alpha 300R)

Fluorescence Imaging with Olympus IX81 Microscope Last updated December 1 st, 2017

OPERATING INSTRUCTIONS

ZEISS LSM 710 CONFOCAL MICROSCOPE USER MANUAL

Nikon TE300 Eclipse Wide-Field Microscope

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

Zeiss 880 Training Notes Zen 2.3

Everest System / Slidebook Operating Procedures

TRAINING MANUAL. Olympus FV1000

Zeiss AxioObserver with ApoTome

Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

Figure 1. Oil-immersion objectives available for use with the Lionheart FX.

Microscopy from Carl Zeiss

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

Axioscan - Startup. 1. Turn on the Axioscan (button to the left) and turn on the computer. 2. Log on and start the ZEN Blue software from the desktop

Zeiss 780 Training Notes

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

NIS-Elements C (For CONFOCAL MICROSCOPE A1) Instructions (Ver. 4.40)

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

SHORT GUIDE TO LASER MICRODISSECTION USING THE PALM COMBI SYSTEM

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

DIC Imaging using Laser Scanning Microscopes (LSM) on Inverted Stands

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Transcription:

Nikon Eclipse Ti-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol Please sign on the log sheet before switching on system. Turn on system Turn on A only if confocal mode or laser is needed. Widefield application can omit this step. Switch on main power control Switch on microscope controller Switch on microscope power Turn on computer power 4 4 Click to log into USER at the startup screen Start the MetaMorph software For widefield users, please click MetaMorph WF icon For Confocal/Laser users, please click MetaMorph CSU icon Set the temperature and CO control for live cell imaging (Only applicable for live cell imaging, please skip this step if it is not needed): Switch on Incubator for temperature and CO control. Switch on the Power of Tokai Hit incubation system controller. Temperature can be altered via pressing the green button of each heating parts on the touch screen. Make sure the CO sliding button is turned ON. Turn on CO tank by turning the main switch anticlockwise. Turn on CO regulator by turning regulator clockwise to set output pressure at 00kPa. Turn on tube switch for TIRF Put on objective heater on objective if oil objective is used. Metal ball floats is an indication of the presence of CO gas. MilliQ water has to be added into the water chamber and covered if overnight(s) acquisition is required. Key in the temperature of interestenter Main Switch Incubation System controller Make sure the metal ball is floating Regulator 00KPa CO indicator

Sample locating and focusing Select objective lens. Apply a drop of immersion oil if 60x oil and 00x oil objective lenses are used. Place your sample, make sure the coverslip bottom is facing down (slide/dish/chamber slide) To view under the microscope, go to Eyepieces select fluorescence channel Current Shutter. The intensity of each channel can be adjusted by clicking one channel and inserting a value of interest. The brightness of the image and the amount of stray light can be adjusted via the aperture diaphragm lever. 4 5 For Brightfield, click Trans Current shutter (can press on Bright Field LED button as an alternative) brightfield brightness can be adjusted using the knob. 4 Press the right arrow of the DISP button to display the XY coordinates and Z position. Move the Stage Controller to adjust XY position (XY speed can be adjusted: ) Focus the sample with the focusing knob Clockwise_Down; Anti-clockwise_Up ( Focusing speed can be adjusted: Switch on the PFS and adjust the focus to lock the focal plane of interest.

Switching to Acquisition mode For Widefield imaging: Click on WF Camera Click Live in the Multi Dimensional Acquisition Select a channel and click Current Shutter to view on the monitor screen. c For Confocal/laser imaging: Click on CSU Camera Click Live in Multi Dimensional Acquisition Select a channel and click Current Shutter to view on the monitor screen. c Image Acquisition Click Multi Dimensional Acquisition on the task bar Go to Main tab to set up acquisition configuration step by step. Check the box(es) of the application(s) as required. Click Saving Select Directory (all data should be saved in E drive/user under your name) Type in the base name of your file (experiment or date or etc.) in Base Name. Do not use digits at the end of the base name, a digit will be added according to the acquisition sequence. Another suffix will be added for record time series image (t, t.) or multi-stage-position image (s, s.).

If multiple fluorescence channels are required, Check the box of Multiple wavelengths in the main menu Click Wavelengths Key in the number of channels in Number of Wavelengths Select each wavelength to set the required Illumination. For Widefield Imaging: - Select WF DAPI Single for Blue emission (such as DAPI) - Select WF GFP Single for Green emission (such as GFP) - Select WF RFP Single for Red emission (such as mcherry) - Select WF Cy5 Single for Far-Red emission (such as mcherry) - Select Trans for brightfield channel - Select DAPI/GFP/RFP/Cy5 Quad channel(s) only when stream application is required. For confocal/laser Imaging: Select CSU DAPI for Blue emission (e.g. BFP) channel Select CSU GFP for Green emission (e.g. GFP) channel Select CSU RFP for Red emission (e.g. mcherry) channel Select CSU CY5 for Far-red emission (e.e. Cy5) channel Select Trans for brightfield channel Channel Settings For Widefiled Imaging: Select W to adjust the first channel Click Live at the bottom of multi-dimensional acquisition panel to have real time image Adjust Gain and Exposure time to have optimum signal intensity. Adjust Gain if necessary (x, x or 4x) Higher Digitizer value gives a higher camera readout speed. Select W and repeat the same procedure to adjust the second channel. Live 4

Timelapse Set up Time interval between each acquisition time point Set the Duration of the entire experiment or Number of time points, either one will do Click Acquire to start the acquisition. Perfect Focus System (PFS) The allowable PFS focusing range refers to the range defined for each objective (where PFS is usable). For glass bottom dish, focus on the sample near to the bottom surface of the sample vessel For plastic dish, focus on the sample near to the bottom surface of the sample vessel, and then move the objective down by about 000um. The status of the PFS is displayed on the LCD of the controller or the front panel of PFS indicator. PFS indicator PFS on/off Shown on the display On PFS on PFS: ON Blinking at slow intervals Blinking at fast intervals PFS on PFS off PFS: DIS PFS: OFF Off PFS off PFS: OFF PFS operating status Perfect focusing is in progress Waiting for interface detection Perfect focusing is off. Perfect focusing is off. Details The PFS will maintain at the focal point if the location of interest is marked. When the interface is detected within the allowable focusing range by moving the focusing position, the PFS is automatically turned on to start perfect focusing. The interface is detected within the allowable focusing range. Turn on the PFS to start perfect focusing. The interface is not detected within the allowable focusing range. In this case, turning on the PFS places it in an interface detection waiting state. 5

Multi stage positions Label the position(s) of interest. Label should be ended with digit so that the number will be automatically updated to record the subsequent position. Click Live to find the right position (x, y) and focus level (z) Click + to add the position (x, y, z) in position list To overwrite recorded stage position, highlight the one to be replaced and click. Make sure the PFS status is ON to ensure every position is in focus. Click Acquire at the bottom to start acquisition. Adjust Focus during Time Lapse Acquisition If amendment is needed halfway through the acquisition process, click Pause Live choose a Position of interest select wavelength click Go to. Adjust the position and focus followed by clicking Set to current click Stop (initially it is Live ) and then Resume to continue the acquisition. 6

Z Series a. Select Z Series in main menu For Spherical object, use Range around current mode: Tick Range around current Focus the center of your object Set up Step Size for distance between each focus plane Set up Number of Steps for the total number of planes Otherwise, use Top and Bottom mode: Tick off Range around current Find any one end of your sample with fine focus, click Set Top To Current Find the other end of your sample with fine focus, click Set Bottom To Current Set up Step Size or Number of Steps for distance between each focus plane 7

Review Acquired Images Click Review Multi Dimensional Data in the Task Bar after Images Acquisition Choose your folder in Select Directory and select an image Data set (base name +suffix. nd) and then click View Select the Wavelength acquired to be displayed. Display a single image by clicking any single grid. Select Stage position in the pull down menu. To review series images, left click the header number of the Row or Column for displaying images of Time series or Z-series respectively. Then click Load Image (s) To export series images as movie, please refer to MetaMorph analysis software protocol. To Overlay images of different channels, check the Color Composite box in the Display tab and then assign corresponding channel to the RGB color to composite a overlay image. To stack all plans in a z-series to create a single D image, choose Maximum projection in Z Projection tab and check the Z Projection box. 8

Turn off system Please check if the equipment will be used by other users. Please switch off system if no one books equipment over two sessions (h) after you. IF 00x/60x objective lens(es) is/are USED, it must be cleaned thoroughly with the LENS PAPER instead of Kimwipes. Oil residue from the objective lens should firstly be removed using a DRY lens tissue. The surface is then wiped with another lens tissue with 00% ethanol. Objective lens is subsequently wiped dry with lens tissue. Switch objective to lowest magnification in the software and press ESC to reach The Lower Z-limit. a. Exit MetaMorph software b. Transfer data to storage server and shut down the computer. c. Switch off microscope power d. Switch off microscope controller e. Switch off main power control f. Switch off laser power A if confocal mode is used. g. Switch off the Power of Tokai Hit incubation system controller. h. Turn off CO regulator by turning regulator clockwise to the end i. Turn off CO tank by turning the main switch clockwise j. Take off objective heater on objective k. Release the valve and remove the water from the chamber by plugging a 50ml syringe (located in the tool box) to the tube. Main Switch Regulator Incubation System controller 9