Confocal Application Letter No. 13. Sequential Scan for Leica TCS NT/SP systems

Similar documents
Confocal imaging on the Leica TCS SP8. 1) Turn the system on. 2) Use TCS user account. 3) Start LAS X software:

Cell Biology and Bioimaging Core

Leica SPEII confocal microscope. Short Manual

Leica TCS SP8 Quick Start Guide

Leica Sp5 II Confocal User Guide

Leica SP8 TCS Users Manual

Leica_Dye_Finder :53 Uhr Seite 6 Dye Finder LAS AF

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

Leica SP8 TCS Users Manual

Guide to Confocal 5. Starting session

Quick Start Guide. Leica SP5 X

Things to check before start-up.

1 Co Localization and Working flow with the lsm700

Leica TCS SP8 Quick Start Guide

Practical work no. 3: Confocal Live Cell Microscopy

b. Turn the power switch and key to on position for blue laser.

ZEISS LSM510META confocal manual

Microscopy from Carl Zeiss

Imaging Beyond the Basics: Optimizing Settings on the Leica SP8 Confocal

LSM 710 Confocal Microscope Standard Operation Protocol

Operating Instructions for Zeiss LSM 510

LEICA TCS SP5 AOBS TANDEM USER MANUAL

LSM 780 Confocal Microscope Standard Operation Protocol

Leica TCS SL Confocal Training. Neuroscience Imaging Core Staff. Core Director. Facility Manager

MIF ZEISS VIOLET CONFOCAL ZEN 2009 PROTOCOL

Multifluorescence The Crosstalk Problem and Its Solution

OPERATING INSTRUCTIONS

Shreyash Tandon M.S. III Year

ADVANCED METHODS FOR CONFOCAL MICROSCOPY II. Jean-Yves Chatton Sept. 2006

Supplemental Figure 1: Histogram of 63x Objective Lens z axis Calculated Resolutions. Results from the MetroloJ z axis fits for 5 beads from each

Title: Leica SP5 Confocal User Manual

LSM 510 META in Chang Gung University

Nikon SIM-E & A1-R System

Nikon A1Rsi Confocal Start-Up Sequence

Nikon AZ100. Laser Scanning Macro Confocal Microscope. Jordan Briscoe Adam Fries Kyle Marchuk Kaitlin Corbin. May 2017.

Bi Imaging. Multicolor Imaging: The Important Question of Co-Localization. Anna Smallcombe Bio-Rad Laboratories, Hemel Hempstead, UK

Operating Checklist for using the Laser Scanning Confocal Microscope. Leica TCS SP5.

Microscope Confocal Sp2 Upright.

Topics. - How to calibrate the LSM scanner. - How to clean the microscope. - How to adjust the pinhole alignment. - How to adjust the Collimator

Travel to New Dimensions- LSM 880. The Resolution of a Microscope is limited. The Resolution of a Microscope is limited. Image. Image. Object.

Zeiss LSM880 Operating Instructions. UTMB Optical Microscopy Core Jan. 16, 2018

Confocal Application Notes Vol. 5 July 2010

Supplemental Method Information Zeiss LSM710

3 Choose the Channels button and set the Channel Settings. Set the Pinhole to 1 Airy unit.

LSM 800 Confocal Microscope Standard Operation Protocol

Quick Guide. LSM 5 MP, LSM 510 and LSM 510 META. Laser Scanning Microscopes. We make it visible. M i c r o s c o p y f r o m C a r l Z e i s s

User Guide to the IBIF Leica TCS SP8 MP Confocal Microscope

TRAINING MANUAL. Multiphoton Microscopy LSM 510 META-NLO

Zeiss 880 Training Notes Zen 2.3

Microscope Confocal LSM510 META

長庚大學共軛焦顯微鏡課程 長庚大學共軛焦顯微鏡課程. Spot light 長庚大學

Usermanual for Leica SP8 confocal

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Dynamic Confocal Imaging of Living Brain. Advantages and risks of multiphoton microscopy in physiology

We attempted to separate the two dyes by acquiring images using a single excitation wavelength and just two emission wavelengths.

Confocal Microscopy. (Increasing contrast and resolu6on using op6cal sec6oning) Lecture 7. November 2017

Confocal Laser Scanning Microscopy

Why and How? Daniel Gitler Dept. of Physiology Ben-Gurion University of the Negev. Microscopy course, Michmoret Dec 2005

Contents. Introduction

BIO-RAD TECHNICAL NOTE 11. Written by: Anna Smallcombe and Duncan McMillan, Bio-Rad Microscopy Division, Hemel Hempstead, UK

Training Guide for Carl Zeiss LSM 510 META Confocal Microscope

1 Set up the confocal light path for imaging a green dye (Alexa488-EGFP). For example, the

Last updated: May 2014 Y.DeGraaf

Training Guide for Leica SP8 Confocal/Multiphoton Microscope

TRAINING MANUAL. Olympus FV1000

Basics of confocal imaging (part I)

Instructions for the Leica SP5 II laser scanning confocal microscope

Confocal Microscopy. Kristin Jensen

The Zeiss AiryScan System, Confocal Four.

Advanced Optical Microscopy lecture. 03. December 2012 Kai Wicker

Light Microscopy. Upon completion of this lecture, the student should be able to:

CONFOCAL MICROSCOPE (Zeiss LSM 510 META v4.2)

Zeiss 780 Training Notes

1.The Problem LIGHT-LEVEL LEVEL IMAGING. light-level level Cameras. 3. Solutions. 2. Low-light LOW-LIGHT

You won t be able to measure the incident power precisely. The readout of the power would be lower than the real incident power.

Leica SP8 Resonant Confocal. Quick-Start Guide

INTRODUCTION TO MICROSCOPY. Urs Ziegler THE PROBLEM

ZEISS LSM 710 NLO Multiphoton microscope Manual/Quick guide

Training Guide for Carl Zeiss LSM 7 MP Multiphoton Microscope

Training Guide for Carl Zeiss LSM 5 LIVE Confocal Microscope

LSM 510 Meta Training Notes

Leica TCS SP2 User Manual 1.1. Markus Friedrich and James Lee. Department of Biological Sciences Wayne State University

Leica Confocal - 2. Instructions for Leica SP2 Confocal Equipped with Visible Laser Lines

LSM 510 Training Notes

Nikon. King s College London. Imaging Centre. N-SIM guide NIKON IMAGING KING S COLLEGE LONDON

Zeiss LSM 780 Protocol

Bi/BE 227 Winter Assignment #3. Adding the third dimension: 3D Confocal Imaging

Fastest high definition Raman imaging. Fastest Laser Raman Microscope RAMAN

Confocal Microscope. Confocal Microscope C2

Batch Counting of Foci

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

Training Guide for Carl Zeiss LSM 880 with AiryScan FAST

Renishaw InVia Raman microscope

Technology Note ZEISS LSM 880 with Airyscan

ZEN 2012 SP5 black edition Hotfix 12

Ratio Imaging. Dividing one image by another to detect changing conditions. Images collected at different times, wavelengths, polarities, etc

Opterra II Multipoint Scanning Confocal Microscope. Innovation with Integrity

Zeiss LSM 880 Protocol

User manual for Olympus SD-OSR spinning disk confocal microscope

Statistical Evaluation of Confocal Microscopy Images

Transcription:

Confocal Application Letter No. 13 Sequential Scan for Leica TCS NT/SP systems

Leica Microsystems Heidelberg GmbH Im Neuenheimer Feld 518 D-69120 Heidelberg Telephone +49 6221 4148 0 Fax +49 6221 414833 www.llt.de

Sequential Scan The new feature Sequential Scan allows image acquisition of multistained specimens to be performed sequentially for different filter settings using a single scanning procedure. This protocol should help you to create your own macro for the sequential scanning procedure. The Sequential Scan enables: Easy examination of multilabelled samples even with systems which have fewer detectors than needed for simultaneous acquisition (e.g. tripple stained sample with a two-channel-system). Reduction of cross-talk when emission spectra of the fluorochromes have wide overlap (e.g. DAPI and FITC, see Fig. 1). DAPI FITC [nm] 400 500 600 700 800 Figure 1: Emission spectra of DAPI/FITC The emission spectra of DAPI and FITC show a wide range of overlap In addition, Sequential Scan also allows fast changes of the excitation wavelength by controlling the Acousto Optical Tuning Filter (AOTF). This capability is needed for ratioimaging ion measurements or ph measurements with fluorescent indicators such as Fura 2 or SNARF- 1, respectively. ApplicationLetter13 Page 3 of 3

Procedure: 1. Open <Filters> 2. For each label (e.g. DAPI, FITC, TRITC, Cy5), define all filter settings such as filter wheels, mirror devices, selection filters (TCS NT) or mirror positions for the detection band (TCS SP), pinhole, AOTF, PMT (HV), offset, Clut, etc. Deactivate all the PMTs not needed for the actual acquisition. Save the settings for each label in a separate macro in the filter setting dialogue by clicking <save> and assigning a unique name to each of them (e.g. seq_dapi, seq_fitc ). Note: check focal plane for maximal intensity. You have now created Available methods for the various stainings of your specimen needed for step 5. 3. Click on <Seq> within the filter dialogue box to open the dialogue box for the Sequential scan settings (Fig. 2). Figure 2: Dialogue for Filter Settings ApplicationLetter13 Page 4 of 4

4. Mark <Activate sequential scan> on the upper left corner of this dialogue box (Fig. 3). Figure 3: Dialogue for Sequential Scan 5. In <Available methods> a complete list of all saved macros is shown. Select the newly defined macros (seq_dapi, seq_fitc,..) and add them to <Sequential methods> in the desired order. 6. In the field <For each seq. method.> select <add channels> 7. If you only have to scan one focal plane, proceed with step 9 below. If a stack of several optical sections is to be scanned, you have the option to change acquisition channel <between frames> or <between stacks>. <between frames> acquires the image (frame) for an optical section for one channel, switches to the other channel, acquires the image, then moves to the next optical section and so on. This procedure is preferable in general because of the possibility of dynamics in your sample, such as shifting, thermal processes, etc. ApplicationLetter13 Page 5 of 5

The <between stacks> option acquires the images for the complete stack of optical sections for one channel, then changes channels acquires the complete stack of optical sections for the second channel. Figure 4: Workflow of option <between frames> Figure 5: Workflow of option <between stacks> 8. In order to activate the stored method parameters in the macros you created in step 2 above, mark <HV/Offset>, <Pinhole>, <AOTF>, <Filterwheels> and <Mirror Positions (λ)> in the checkbox <Activate stored method parameters>. The images will now be acquired exactly as already defined in your macro. By not marking one of the parameters with a check, the system will use the last setting. If, for example, you wish to adjust the PMT during acquisition of the single images, do not mark <HV/Offset> with a check. 9. The complete set of sequential scan settings may now be saved. It will be saved as type sequential settings (*.seq), and again you must assign to it a unique name. 10. Close the dialogue box. 11. Adjust general parameters such as the focal plane, begin and end of a series, number of sections, accumulation (number of frames to be averaged), zoom factor etc. for one channel. If it is available on your system, adjust these parameters in the simultaneous multi channel scanning mode. 12. Activate the button <series> in order to start the sequential scan procedure. ApplicationLetter13 Page 6 of 6

Note: Do not forget to deactivate sequential scan for simultaneous detection. Remark: If you want to acquire the images as fast as possible, e.g. for Fura 2 measurements, activate as few predefined parameters as possible in the checkbox <Activate stored method parameters> (step 8). Only those parameters checked are then controlled by your Sequential Scan macro during acquisition. Unchecked parameters remain unchanged by the macro, i.e. they are fixed to the values present before the macro was activated. simultaneous acquisition (mixture of red and green = yellow) sequential acquisition (separation of red and green) Figure 6: Sample of DAPI/Cy2/Cy3 tripple stained specimen Parasagittal section of mouse cerebellum immunolabelled with antibodies against glial fiber acid protein for astrocytes (red), with antibodies against calbindin D-28k for Purkinje cells (blue), and counterstained with Hoechst (green). Hoechst anti-gfap/cy2 calbindin-d28k/cy3 ApplicationLetter13 Page 7 of 7