Keysight Technologies Oscilloscope Probe Loading Experiment

Similar documents
Keysight Technologies 7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes. Application Note

Keysight Measuring High Impedance Sources Using the U8903B Audio Analyzer. Application Note

Keysight Technologies How to Measure 5 ns Rise/Fall Time on an RF Pulsed Power Amplifier Using the 8990B Peak Power Analyzer.

Keysight Technologies Making Current-Voltage Measurement Using SMU

Introduction. Part 1. Introduction...2

Keysight Technologies Accurate Evaluation of MEMS Piezoelectric Sensors and Actuators Using the E4990A Impedance Analyzer.

Keysight Technologies, Inc. Overcome PCB Loss and Deliver a Clean Eye to Your DUT Using Multi-tap De-emphasis

Keysight Technologies 1 mw 50 MHz Power Reference Measurement with the N432A Thermistor Power Meter. Application Note

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

The Keysight Technologies, Inc. U1730C Series handheld LCR meters allow you to measure at frequencies as high as 100 khz a capability typically found

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies Educator s Oscilloscope Training Kit for InfiniiVision X-Series Oscilloscopes. Data Sheet

Keysight Technologies Minimum Required Sample Rate for a 1-GHz Bandwidth Oscilloscope

Keysight Technologies U1730C Series Handheld LCR Meters Take your expectations higher with the latest LCR meters. Data Sheet

7 Hints That Every Engineer Should Know When Making Power Measurements with Oscilloscopes.

Keysight Technologies N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes. Data Sheet

Keysight Technologies How to Select the Right Current Probe. Application Note

Keysight N8836A PAM-4 Measurement Application For Infiniium S-Series, 90000A, V-Series, X-Series, Q-Series, and Z-Series Oscilloscopes

Keysight Technologies Achieving Accurate E-band Power Measurements with E8486A Waveguide Power Sensors. Application Note

Keysight Technologies RS-232/UART Protocol Triggering and Decode for Infiniium 9000A and 9000 H-Series Oscilloscopes. Data Sheet

Keysight U1882B Measurement Application for Infiniium Oscilloscopes. Data Sheet

Keysight Technologies Secondary Radar Transponder Testing Using the 8990B Peak Power Analyzer. Application Note

Keysight 8762F Coaxial Switch 75 ohm

When is it Time to Transition to a Higher Bandwidth Oscilloscope?

N2820A/21A High-Sensitivity, High Dynamic Range Current Probes

Keysight Technologies DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Keysight DSOXT3FRA/DSOX4FRA/DSOX6FRA Frequency Response Analyzer (FRA) Option

Keysight Technologies Using an External Trigger to Generate Pulses with the B2960A

Educator s Oscilloscope Training Kit for Agilent InfiniiVision X-Series Oscilloscopes

Keysight Technologies

Keysight Technologies Enhance EMC Testing with Digital IF. Application Note

Keysight Technologies N2792A/N2818A 200 MHz and N2793A/N2819A 800 MHz Differential Probes. Data Sheet

Keysight Technologies How to Take Fast, Simultaneous Measurements of Two or More Signals Using BenchVue Software. Application Note

Keysight Technologies

Keysight Technologies, Inc. UWB Antenna Measurements with the 20 GHz E5071C ENA Network Analyzer. Application Note

U1881A and U1882A Power Measurement Application for InfiniiVision and Infiniium Oscilloscopes

Keysight Technologies Differences in Application Between Power Dividers and Power Splitters. Application Note

Keysight Technologies N9398C/F/G and N9399C/F DC Block. Technical Overview

Keysight Technologies InfiniiScan Event Identification Software

Keysight Technologies U9391C/F/G Comb Generators. U9391C (10 MHz to 26.5 GHz) U9391F (10 MHz to 50 GHz) U9391G (10 MHz to 67 GHz) Technical Overview

Keysight Technologies NFA Noise Figure Analyzer. Configuration Guide

Keysight Technologies RF & Microwave Attenuators. Performance you can count on

Educator s Oscilloscope Training Kit for the InfiniiVision 2000 & 3000 X-Series

Keysight N8803C CAN, LIN, FlexRay, and CAN-FD Protocol Triggering and Decode Software. Data Sheet

Keysight Technologies N9063A & W9063A Analog Demodulation

Keysight Technologies 8490G Coaxial Attenuators. Technical Overview

Keysight Technologies Improving the Test Efficiency of MEMS Capacitive Sensors Using the E4980A Precision LCR Meter.

Keysight Technologies Improving Test Efficiency of MEMS Electrostatic Actuators Using the E4980A Precision LCR Meter.

Keysight Technologies Make Better AC RMS Measurements with Your Digital Multimeter. Application Note

Introduction. Part 1. Introduction...2

Keysight Technologies Making Field Effect Transistor Characterization Using SMU

Keysight Technologies 87405C 100 MHz to 18 GHz Preamplifier. Technical Overview

Keysight Technologies Alliance for Wireless Power (A4WP) Measurements Using an Oscilloscope (Part 3)

Keysight Technologies MEMS On-wafer Evaluation in Mass Production

Keysight Technologies 85072A 10-GHz Split Cylinder Resonator. Technical Overview

Keysight Technologies CAN, LIN and FlexRay Protocol Triggering and Decode for Infiniium 9000 and S-Series Oscilloscopes.

Keysight Technologies Simultaneous Measurements with a Digital Multimeter

Keysight Technologies 423B, 8470B, 8472B, 8473B/C Low Barrier Schottky Diode Detectors

Agilent Spectrum Visualizer (ASV) Software. Data Sheet

Keysight DSOX3PWR/DSOX4PWR/DSOX6PWR Power Measurement Options

Evaluating Oscilloscopes for Low-Power Measurements

Keysight Technologies Precise Low Resistance Measurements Using the B2961A and 34420A

Keysight Technologies 87222C/D/E Coaxial Transfer Switches

N2750A/51A/52A InfiniiMode Differential Active Probes

Keysight Technologies Direct Power MOSFET Capacitance Measurement at 3000 V

Keysight Technologies N6850A Broadband Omnidirectional Antenna. Data Sheet

Keysight Technologies Network Analyzer Measurements: Filter and Amplifier Examples. Application Note

Keysight Technologies NFC Device Turn-on and Debug

N2790A 100 MHz, N2791A 25 MHz and N2891A 70 MHz High-voltage Differential Probes

Keysight Technologies Accurate Capacitance Characterization at the Wafer Level

Keysight Technologies FFT and Pulsed RF Measurements with 3000T X-Series Oscilloscopes. Application Note

Keysight Technologies Accurate NBTI Characterization Using Timing-on-the-fly Sampling Mode. Application Note

Keysight 8474B/C/E Planar-Doped Barrier Diode Detectors 0.01 to 50 GHz. Data Sheet

Keysight Technologies N4983A Multiplexer and Demultiplexer. Data Sheet

Keysight Technologies Migrating Balanced Measurements from the

Keysight Technologies N1918A Power Analysis Manager and U2000 Series USB Power Sensors. Demo Guide

Keysight Technologies Using a Scope s Segmented Memory to Capture Signals More Efficiently. Application Note

Keysight Technologies Generating and Applying High-Power Output Signals

Oscilloscope Fundamentals. For Electrical Engineering and Physics Undergraduate Students

Agilent E1412A 6.5-Digit High-Accuracy Multimeter C-Size

Keysight Technologies Power of Impedance Analyzer

Keysight Technologies MATLAB Data Analysis Software Packages

Evaluating Oscilloscope Bandwidths for your Application

Keysight Technologies Migrating from the 4268A/4288A Capacitance Meter to the E4981A Capacitance Meter. Technical Overview

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Keysight Technologies Triggering on and Decoding the PSI5 Sensor Serial Bus

Keysight Technologies P9400A/C Solid State PIN Diode Transfer Switches

Keysight Technologies Measuring Power BJT Electrical Characteristics using the B1505A

Keysight Technologies Eight Hints for Better Scope Probing. Application Note

Keysight Technologies How to Easily Create an Arbitrary Waveform Without Programming. Application Note

Keysight Technologies N9310A RF Signal Generator

Keysight HMMC-1002 DC 50 GHz Variable Attenuator

Keysight Technologies Making Simpler DC Power Measurements with a Digital Multimeter

Keysight Technologies InfiniiMax III/III+ Probing System

InfiniiMax III probing system

Keysight Technologies Enhancing Measurement Performance for the Testing of Wideband MIMO Signals

Keysight Technologies Precise Current Profile Measurements of Bluetooth Low Energy Devices using the CX3300. Application Brief

Keysight Technologies A Flexible Testbed to Evaluate Potential Co-Existence Issues Between Radar and Wireless

Keysight Technologies Automated Receiver Sensitivity Measurements Using U8903B. Application Note

Transcription:

Keysight Technologies Oscilloscope Probe Loading Experiment A hands-on lab experiment and probing tutorial for EE students Demo Guide

When you connect an oscilloscope probe to a test point in a circuit, the probe itself becomes a part of the circuit under test and can affect measured results. This is commonly referred to as probe loading. Using a simple 2-resistor voltage-divider network, this experiment will empirically show how the frequency-dependent impedance of the probe can significantly impact measurement accuracy. Required Equipment and Components 2-channel oscilloscope ( 50 MHz bandwidth) Function generator ( 10 MHz) Two standard 10:1 passive oscilloscope probes Breadboard Two 10 kω resistors Compensating your Probes It is important that you properly compensate your oscilloscope probes prior to building your circuit and performing this experiment, otherwise measurements will be inaccurate. To compensate your probes, connect one probe between the scope s channel-1 input and the probe compensation test terminal located somewhere on the front panel of your scope. Connect the second probe between the scope s channel-2 input and the same probe compensation test terminal. Don t forget to connect both probes ground leads to a ground terminal on the front panel of the scope. Next, set the probe attenuation factors for both input channels to 10:1 (10-to-1). Note that some higher-end scopes will detect that 10:1 probes are connected and will then automatically set the probe attenuation factor for you. Next, set up each channel s V/div setting, as well as the sec/div setting to show one or two periods of the probe compensation signal on the scope s display. The probe compensation signal is typically a 1 khz square wave, so the appropriate sec/div setting should be 200 µsec/div. Using a small flat-blade screw driver, adjust each probe s adjustable compensation capacitor as shown in Figure 1 such that both waveforms have a flat response. You should find this adjustable capacitor either near the probe tip, or on the part of probe closer to where it plugs into the scopes BNC inputs. Figure 1: Adjusting the probe compensation of each passive probe. 2

Figure 2: Using the scope s 1 khz probe compensation signal to compensate 10:1 passive probes. Figure 3: Improperly compensated probes. Figure 2 shows what your channel-1 and channel-2 waveforms should look like if the probe compensation of each probe is properly adjusted. Figure 3 shows an example of the channel-1 probe (yellow waveform) over-compensated, and an example of the channel-2 probe (green waveform) under-compensated. So what is probe compensation all about? We ll find out later. Building the experiment, predicting results, and measuring results Figure 4: 2-resistor voltage-divider network. 3

Using your breadboard and two 10-kΩ resistors, build your 2-resistor voltagedivider network as shown in the schematic of Figure 4. Note that if you do not have a breadboard, then solder your two resistors together as opposed to simply connecting them together with long wires and clips. Long wires will introduce inductance into this experiment, which we want to avoid. Before turning on the function generator and making any measurements with your scope, answer the following questions: V out /V in = If V in = 5 V pp, then V out = Let s now test this circuit against these predicted results. Function Generator Settings and Connections: 1. Set output load impedance to High Z (not 50-Ω) 2. Set waveshape to Sine Wave 3. Set amplitude to 5 Vpp 4. Set offset to 0.0 V 5. Set frequency to 10 khz 6. Connect output of generator to R1. 7. Connect ground of generator to circuit ground. Oscilloscope Settings and Connections: 1. Connect channel-1 probe between V in and ground. 2. Connect channel-2 probe between V out and ground. 3. Begin with a Default Setup condition. Most scopes have a Default Setup selection on the front panel, or perhaps within the Save/Recall menu. 4. Insure that probe attenuation factors are still set to 10:1 for both channels of the scope. 5. Set channel-1 and channel-2 vertical scaling to 1.0 V/div. 6. Center the channel-1 and channel-2 waveforms on-screen using the vertical position/offset controls. 7. Set the horizontal scaling (timebase) to 20.0 µs/div. 8. Set triggering on a rising edge of channel-1 at approximately 0.0 Volts (typical default setting). 9. Measure V in and V out (peak-to-peak) using either manually-placed cursors, automatic measurements, or simply count divisions and multiply by the vertical scaling factor (1.0 V/div). 4

Your scope s display should now look similar to Figure 5. Figure 5: Measuring V in and V out at 10 khz using two channels of the oscilloscope. Record your measurements: V in @ 10 khz= V out @ 10 khz= V out /V in @ 10 khz= Is this pretty close to what you originally predicted? Now change the frequency setting of the function generator to 10 MHz. Also change the scope s horizontal timebase setting to 20.0 ns/div in order to view this faster input signal. Measure V in and V out once again. At this point, your scope s display may look similar to Figure 6. Figure 6: Measuring V in and V out at 10 MHz using two channels of the oscilloscope. Record your measurements: V in @ 10 MHz = V out @ 10 MHz = V out /V in @ 10 MHz = Is this close to what you predicted? If not, why not? 5

Understanding Probe Loading The reason the amplitude of the signal across R2 decreased at 10 MHz was due to capacitive probe and scope loading. In a perfect world, probes would have infinite impedance and would have no affect on your measurements. But any time you connect a probe to a device under test, whether you are using a spectrum analyzer, power meter, multimeter, power meter, network analyzer, or an oscilloscope, the probe and instrument become a part of your circuit under test and can affect the accuracy of measurements. This is especially true when testing higher frequency signals. Now take a close look at the probes that you just used for this experiment near the BNC connection end at the scope s input. You should see a vendor s name and model number associated with these particular probes. You should also see an input impedance specification/characteristc. It probably says something like, 10MΩ/15pF as shown in Figure 7. Figure 7: Oscilloscope probe model number and input impedance characteristics. This means that when the probe is connected to the scope, it has an equivalent input resistance of 10 MΩ in parallel with 15 pf. Figure 8 shows an equivalent probe/scope loading model. This is what is hanging in parallel with R2 (refer to Figure 4). You can assume that the 10 MΩ resistor is so large compared to your 10 kω resistor (R2) that it s not even there. You can also assume that at low frequencies, the 15 pf capacitor will not affect your circuit. But what is the reactance of this capacitance at 10 MHz. Xc = 1/(2πfC) = Figure 8: 10:1 passive probe loading model. 6

Now compute the load impedance that consists of R2 in parallel with Xc. Remember, you can ignore the 10 MΩ resistor. Z Load = (R2)(Xc)/( (R2 2 +Xc 2 ) = Determine the approximate output voltage when the input frequency is set to 10 MHz now based on the voltage-divider network consisting of R1 in series with Z Load. V out = V out /V in = Is this closer to what you measured when the input signal was set at 10 MHz? So we appear to have a dilemma here. We need to measure the output voltage of a circuit, but as soon as we connect an oscilloscope probe to the circuit, it changes the output characteristics. How do we handle this situation? First of all, we used 10 kω resistors in this experiment to illustrate a point. The point was that the capacitive reactance of the probe at higher frequencies can swamp out the impedance of our load resistor (R2). But in reality, most higher frequency designs consists of lower impedance devices/components. But even in low impedance designs, probes can affect the circuit under test when frequencies get high enough, such as hundreds of Mega Hertz, or perhaps Giga Hertz signals. After all, most of today s PCs run on clock rates in the multi-giga Hertz range. Applications such as these typically require special high-frequency active probes. Passive probes, such as the ones you used in this experiment, consists of passive components only; resistors and capacitors. Higher frequency probes usually include active components, such as transistors and amplifiers. And these probes require power in order to operate. The input capacitance of active probes can be in the sub-pico Farads range. This means that they will have less affect but theoretically never zero affect on your circuit at higher frequencies. However, these probes also cost significantly more than standard passive probes, which are usually supplied with the scope. Active probes are almost always a pay for option. If you would like to learn more about oscilloscope probes, download the Keysight Technologies, Inc. Application Note titled, Eight Hints for Better Scope Probing listed at the end of this document. 7

Understanding Probe Loading Figure 9 shows a more detailed but still simplified electrical model of a typical 10:1 passive probe when connected to an oscilloscope using the scope s default 1 MΩ input selection. Although the electrical model of the passive probe and scope includes both inherent/parasitic capacitance (not designed-in) as well as an intentionally designed-in compensation capacitance network, let s ignore these capacitive elements for now and analyze the ideal signal behavior of this probe and scope system under low-frequency conditions. After we remove all of the capacitive components from our electrical model, what remains is just a 9 MΩ probe tip resistor in series with the scope s 1 MΩ input impedance. The net input resistance at the probe tip is then 10 MΩ, which agrees with the probe loading model shown earlier (Figure 8). Using Ohm s law, you can see that the voltage level received at the scope s BNC input is then 1/10th the voltage level that is at the probe tip: V scope = V probe x (1 MΩ/10 MΩ) This is why this type of probe is called a 10:1 (pronounced 10-to-1 ) probe. Once the scope knows that it has a 10:1 probe attached to its input, all measurements and vertical scaling factors are multiplied by 10 in order to reference measurements to the probe tip. The way a scope knows that it has a 10:1 probe attached is either by you manually entering the probe attenuation factor or through automatic detection. If you are using an older analog oscilloscope that does not have probe attenuation factors, then you ve got to do the math yourself in order to reference measurements to the probe tip. For low-frequency or dc applications, ignoring the capacitive elements is appropriate. But if you need to measure dynamic signals, which is the primary measurement application for oscilloscopes, the capacitive elements of this electrical model can t be ignored. So let s now take a closer look at the probe and scope model under dynamic/ac input signal conditions. Figure 9: Simplified model of a typical passive 10:1 probe connected to the scope s 1 MΩ input impedance. 8

Inherent in all oscilloscope probes and scope inputs are parasitic capacitances. These include the probe cable capacitance (C cable ), as well as the scope s input capacitance (C scope ). Inherent/parasitic simply means that these elements of the electrical model are not intentionally designed-in; but are just an unfortunate fact of life in the real world of electronics. And the amount of inherent/parasitic capacitance will vary from scope-to-scope and probe-to-probe. But without additional designed-in capacitive components to compensate for the inherent capacitive elements in the system, the reactance of the system under dynamic signal conditions (non-dc) can change the overall dynamic attenuation of the probing system to something different than the desired 10:1 ratio. The purpose of the additional/designed-in probe tip capacitor (C tip ) along with the adjustable compensation capacitor (C comp ) is to establish a capacitive reactance attenuation network that exactly matches the resistive attenuation of 10:1. In other words, the reactance of C tip must be exactly 9X the reactance of the parallel combination of C comp + C cable + C scope. If this is true, then not only will low-frequency signals be attenuated by a factor of 10 based upon the 9 MΩ probe tip resistance (R tip ) in series with the 1 MΩ scope input resistance (R scope ), but higher frequency signals will also be attenuated by a factor of 10 based upon a similar 10:1 capacitive reactance voltage-divider network. Let s now compute the required amount of compensation capacitance (C comp ) using the following assumptions: C tip = 15 pf C scope = 15 pf C cable = 100 pf C parallel = C scope + C cable + C comp C comp =? Remember, the reactance of C tip must be 9X the reactance of C parallel. Therefore we can use the following formula to compute C comp : 1 1 2πfC = 9 tip 2πfC parallel where C comp = C parallel - C scope - C cable What is the appropriate adjusted value of C comp? Once we know the value of the C comp, we can then compute the overall capacitive loading, which is C tip in series with C parallel. C Load = (C tip )(C parallel ) (C tip +C parallel ) = Is this the same value of loading capacitance shown in the probe loading model of Figure 8? 9

Summary During this hands-on oscilloscope probe loading lab experiment, hopefully you learned that when you connect an oscilloscope probe to a device-under-test, the probe and scope become a part of circuit and can negatively affect the accuracy of measurements especially when probing higher frequency signals. In many of your entry-level teaching labs, this may not be issue that you will need to be concerned about. But in some of your upper-level and graduate-level EE classes and labs that may focus on higher frequency RF applications, or perhaps highspeed digital applications, probe loading is something that you should watch for. And remember that digital signals have higher frequency harmonics far beyond the clock rate of the signal. As a rule-of-thumb, the capacitive reactance of your probing system (including the scope) should be 10X the Théveninequivalent source impedance of your system under test. For the experiment outlined in this document (Figure 4), the Thévenin-equivalent source impedance would be 5 kω. You also learned about the theory of operation of a passive 10:1 oscilloscope probe and how probe compensation works. Although understanding the theory of probe compensation may not be that important, properly compensating your probes is very important even for many of your lower frequency entry-level lab experiments. It is always good practice before making any measurements with a scope to connect your probes to the probe compensation signal on the front panel of the scope and insure that they are properly adjusted. If you would like to learn more about oscilloscopes, you can download a PowerPoint presentation titled, Oscilloscope Fundamentals at www.keysight.com/find/edk. Just click on the Software & Trials tab to download this file. Note that this PowerPoint presentation also includes a complete set of speaker notes. Also available are some application notes on oscilloscopes and probes listed at the end of document in the Related Literature section. Related Literature Publication Title Publication Type Publication Number Eight Hints for Better Scope Probing Application Note 5989-7894EN Evaluating Oscilloscope Fundamentals Application Note 5989-8064EN Evaluating Oscilloscope Bandwidths for Application Note 5989-5733EN your Applications Oscilloscopes in Education Application Note 5989-9166EN To download these documents, insert the publication number in the URL: http://literature.keysight.com/litweb/pdf/xxxx-xxxxen.pdf 10

11 Keysight Oscilloscope Probe Loading Experiment - Demo Guide mykeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.axiestandard.org AdvancedTCA Extensions for Instrumentation and Test (AXIe) is an open standard that extends the AdvancedTCA for general purpose and semiconductor test. Keysight is a founding member of the AXIe consortium. www.lxistandard.org LAN extensions for Instruments puts the power of Ethernet and the Web inside your test systems. Keysight is a founding member of the LXI consortium. www.pxisa.org PCI extensions for Instrumentation (PXI) modular instrumentation delivers a rugged, PC-based high-performance measurement and automation system. www.keysight.com/quality Keysight Electronic Measurement Group DEKRA Certified ISO 9001:2008 Quality Management System Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight s measurement expertise and product breadth, combined with channel partner convenience. For more information on Keysight Technologies products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 112 929 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 0800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-05-23-14) Keysight Technologies Oscilloscopes Multiple form factors from 20 MHz to >90 GHz Industry leading specs Powerful applications This information is subject to change without notice. Keysight Technologies, 2011-2014 Published in USA, August 1, 2014 5990-9175EN www.keysight.com