BPM, Wire-Scanner, OTR/YAG Screens, and Collimator Polarity Conventions

Similar documents
LCLS Injector Diagnostics. Henrik Loos. Diagnostics overview Transverse Beam Properties Longitudinal Beam Properties

Initial Beam Phasing of the SRF Cavities in LCLS-II

ELECTRON BEAM DIAGNOSTICS AND FEEDBACK FOR THE LCLS-II*

Homogeneous Representation Representation of points & vectors. Properties. Homogeneous Transformations

Experience with Insertion Device Photon Beam Position Monitors at the APS

Check the LCLS Project website to verify 2 of 7 that this is the correct version prior to use.

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

Appendix M TERMINOLOGY. Slope of a Line. Slope. Undefined Slope. Slope-Intercept Form

DOSELAB TOMOTHERAPY TG-148 QA QUICK GUIDE TG-148 RECOMMENDED TESTS 1. V.B.1.C. - Y-JAW DIVERGENCE/BEAM CENTERING

FIRST INDIRECT X-RAY IMAGING TESTS WITH AN 88-mm DIAMETER SINGLE CRYSTAL

Section 3. Imaging With A Thin Lens

Active Stereo Vision. COMP 4102A Winter 2014 Gerhard Roth Version 1

AgilEye Manual Version 2.0 February 28, 2007

Supplementary Information

CyberKnife Iris Beam QA using Fluence Divergence

Agilent 10774A Short Range Straightness Optics and Agilent 10775A Long Range Straightness Optics

Physics Requirements for the CXI 0.1 micron Sample Chamber

X-RAY COMPUTED TOMOGRAPHY

10.3 Polar Coordinates

CHAPTER 10 Conics, Parametric Equations, and Polar Coordinates

Receiver Obstacle Railway Transmitter Road Figure 1: Existing obstacle-detecting system (light-interrupting type). Millimetre-wave-based System In ord

Beam Analysis BeamWatch Non-contact, Focus Spot Size and Position monitor for high power YAG, Diode and Fiber lasers. Disruptive Technology

LUSI Pulse Picker System

Nonintercepting Diagnostics for Transverse Beam Properties: from Rings to ERLs

Non-invasive Beam Profile Measurements using an Electron-Beam Scanner

IDEA Connections. User guide

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

PROCEEDINGS OF SPIE. Automated asphere centration testing with AspheroCheck UP

DEVELOPMENT OF OFFNER RELAY OPTICAL SYSTEM FOR OTR MONITOR AT 3-50 BEAM TRANSPORT LINE OF J-PARC

Beam Profiling. Introduction. What is Beam Profiling? by Michael Scaggs. Haas Laser Technologies, Inc.

A high resolution bunch arrival time monitor system for FLASH / XFEL

CHAPTER 10 Conics, Parametric Equations, and Polar Coordinates

Exercise 2-2. Antenna Driving System EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION

Estimation of Absolute Positioning of mobile robot using U-SAT

A machine vision system for scanner-based laser welding of polymers

CXI 1 micron Precision Instrument Stand

Supplementary Figure S1. Schematic representation of different functionalities that could be

PACIFIC LASER SYSTEMS. HVL100 Procedures

Section 8. Objectives

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian

A LARGE COMBINATION HORIZONTAL AND VERTICAL NEAR FIELD MEASUREMENT FACILITY FOR SATELLITE ANTENNA CHARACTERIZATION

Basic Optics System OS-8515C

SOLAR OVENS AND HOT PLATES FOR COOKING AND AVOID FIREWOOD CONSUMPTION. Eduardo A. Rincón Mejía

Test Results using Iris and Airy Disc for the BPM Alignment of SCSS (SPring-8 Compact SASE Source) Prototype Accelerator

DESIGN OF COMPACT PULSED 4 MIRROR LASER WIRE SYSTEM FOR QUICK MEASUREMENT OF ELECTRON BEAM PROFILE

universal Digital Radiography (DR) System

Development of optical imaging system for LIGO test mass contamination and beam position monitoring

1.7 Parallel and Perpendicular Lines

4.5.1 Mirroring Gain/Offset Registers GPIO CMV Snapshot Control... 14

Stretched Wire Test Setup 1)

7. Michelson Interferometer

Supplementary Figure 1

Cost efficient design Operates in full sunlight Low power consumption Wide field of view Small footprint Simple serial connectivity Long Range

Single-Slit Diffraction. = m, (Eq. 1)

Stitching MetroPro Application

Functions of more than one variable

Automatic Laser-Controlled Erection Management System for High-rise Buildings

IDEA Connection 8. User guide. IDEA Connection user guide

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

VTube-LASER Quick Start Guide

Triangle Definition of sin θ and cos θ

Commissioning the Echo-Seeding Experiment ECHO-7 at NLCTA

1. Enumerate the most commonly used engineering materials and state some important properties and their engineering applications.

Powder diffractometer operation instructions D8 Advance with a Cu Kα sealed tube and Lynxeye MARCH 30, 2016

Instruction Manual for HyperScan Spectrometer

Properties of Structured Light

Hartmann wavefront sensing Beamline alignment

Upgrade of the ultra-small-angle scattering (USAXS) beamline BW4

Video-based X-Ray Beam Position Monitoring at CHESS Peter Revesz, Cornell University, CHESS Ithaca, NY 14850

2. Polar coordinates:

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Optical Measurement P-1

PAD Correlator Computer

Transverse Wakefields and Alignment of the LCLS-II Kicker and Septum Magnets

Principles of CT scan

Eric B. Burgh University of Wisconsin. 1. Scope

Sensor production readiness

Undulator K-Parameter Measurements at LCLS

Outline: Introduction: What is SPM, history STM AFM Image treatment Advanced SPM techniques Applications in semiconductor research and industry

Lawrence Berkeley Laboratory UNIVERSITY OF CALIFORNIA

Installation of the Optical Replica Synthesizer (ORS) at FLASH

Design and Correction of optical Systems

ADVANCED OPTICS LAB -ECEN 5606

Camera Calibration Certificate No: DMC II

ISONIC PA AUT Spiral Scan Inspection of Tubular Parts Operating Manual and Inspection Procedure Rev 1.00 Sonotron NDT

Conceptual Design Report. 11 Electron Beam Diagnostics. Synopsis. Chapter 11 - Beam Instrumentation

Radionuclide Imaging MII 3073 RADIONUCLIDE IMAGING SYSTEM

HiFi Radar Target. Kristian Karlsson (RISE)

PANalytical X pert Pro Gazing Incidence X-ray Reflectivity User Manual (Version: )

Engineering & Computer Graphics Workbook Using SolidWorks 2014

Southern African Large Telescope. RSS CCD Geometry

plasmonic nanoblock pair

Feasibility of Steered Shaped Beam Dielectric Lens Antenna for Mobile Communications at mm-waves

Parity and Plane Mirrors. Invert Image flip about a horizontal line. Revert Image flip about a vertical line.

Camera Calibration Certificate No: DMC III 27542

Engineering & Computer Graphics Workbook Using SOLIDWORKS

Instructions for the Experiment

EXPERIENCE WITH AND STUDIES OF THE SNS* TARGET IMAGING SYSTEM

A Beam-Level Delivery Accuracy Study of the Robotic Image Guided Radiosurgery System Using a Scintillator/CCD Phantom

Transcription:

BPM, Wire-Scanner, OTR/YAG Screens, and Collimator Polarit Conventions Beam Position Monitors (BPMs) The Beam Position Monitors (BPMs) in the beamline are used to measure transverse ( and ) electron beam centroid positions on a pulse b pulse basis. The readback coordinate conventions used in the SLC will also be used for the LCLS. If the beam is high with respect to the BPM centerline, the vertical position reading () is positive. If the beam is to the right of the BPM centerline (with beam emerging from paper see Fig. 1 below), the horizontal position reading () is positive. Some of the BPM stripline monitors are rotated b 45 degrees (see right side of Fig. 1). This usuall occurs where the beamline is undergoing a bend. This rotation keeps the electrodes awa from the snchrotron radiation or low-energ dark current, which might otherwise impinge on the electrode. In addition, the BPM strip-lines are calibrated on the bench before installation and this calibration measures the horizontal and vertical offset specific to an one BPM. These bench-measured offsets, and, are entered into the control sstem database, where the are permanentl stored, and are used in the BPM processing software to correct these offsets. In an case, the final position readback reported to the user should be in the standard beamline coordinates, and, as shown in Fig. 1. In addition, the BPM offsets should have been removed and the beam position scaled in millimeters. 2 of 8 that this is the correct version prior to use.

Figure 1. A BPM as seen down its bore with electron beam emerging from paper, showing the beamline coordinates (, ). At left is a BPM installed in the upright position. At right is a BPM installed with an optional 45-degree rotation. The cable color-coding standard is also illustrated. The four ADC s are used to form two difference readings and two sum reading with their ratios forming the raw, un-scaled position readings (the subscripts below refer to the Red, Green, Yellow, and Blue color codes shown in Fig. 1). V V V V G B R Y =, =. (1) VG + VB VR + VY The transformation used to convert these raw readings to final scaled position, with offset and rotation correction, is then cosθ = sinθ sinθ C +, (2) cosθ C + where C is the static calibration factor [1] in millimeters (PCMM), and are the raw readbacks of Eq. (1), and are the static bench-measured offsets (in millimeters) from the database, and θ (= 0 or +45 ) is the specific BPM rotation angle from the database. It is also important to note that the rotation angle (ANGL) will alwas be either zero or +45 degrees, depending on whether the BPM is installed as at left or right in Fig. 1, respectivel. The rotation angle is stored in the database in units of degrees. The specific BPM rotations are listed in Table 2 of PRD 1.1-314 at http://www-ssrl.slac.stanford.edu/lcls/prd/1.1-314-r3.pdf. The choice in Eq. (2) of adding (rather than subtracting) the offset is based on the eisting EPICS standard. Note that these bench-measured offsets are usuall reported in microns and should be first converted to millimeters. These offsets represent the position reading (due to fabrication imperfections) that would be reported (for an un-rotated BPM) if the beam were perfectl centered in the BPM. Therefore these numbers should be subtracted in Eq. (2), but since this is not the EPICS standard, these bench-measured offsets will need to have their signs flipped before being used in Eq. (2). Note also that the special BC1 and BC2 chicane BPMs, with their more rectangular shape, ma require special treatment that is not et outlined here. Wire-Scanners The wire-scanners are used to measure transverse beam size integrated over man pulses. The wires are driven with a stepping motor through the beam, usuall at a ±45-degree motor orientation, as shown in Fig. 2. There are tpicall three wires (, u, and ), which meet in a verte alwas at the top 3 of 8 that this is the correct version prior to use.

of the wire-support card. This means that the -wire (the wire that measures the beam size, but is oriented verticall) alwas intercepts the beam first as the scanner is driven into the beam from its etracted position. The motor drives the scanner along (what we define as) the v-ais (see Fig. 2). This v coordinate is defined to be zero when the u-wire is located at the beamline center (as drawn in Fig. 2) and takes on positive values as the wire is etracted (toward the top of Fig. 2). If the beam position varies (jitter) while the wire is being scanned, the profile data will be scattered and therefore less precise. Several BPMs in the region can be used to calculate the position variation of the beam during the scan, which can be used to correct this scatter (with respect to the average position). This position-jitter correction requires the BPMs and the wire-scanners to have consistent coordinate sstem conventions, which are described here. Figure 2. A wire-scanner as seen down its bore with electron beam emerging from paper, showing the beamline coordinates (, ). At left is a wire-scanner installed with a +45 rotation. At right is a wire-scanner installed in a 45 rotation. The wires meet in a verte alwas at the top of the wire-support card. The transformation from the stepping-motor s ais of motion (v) to the standard and beamline coordinates is = v sinθ, = v cosθ, u = v, (3) where θ is the database rotation angle shown in Fig. 2. The u-ais is alwas the same as the v-ais. The BPM jitter data, B and B, can be subtracted from the wire position data, w, w, and u w (same as,, and u in Eq. (3)) using: c = w - B and c = w - B, and u c = u w - u B, but the u-ais BPM jitter data, u B, should be formed from the and BPM data using the wire rotation angle in the form: 4 of 8 that this is the correct version prior to use.

u = sinθ + cosθ. (4) B B In the present LCLS design, the onl values taken b θ are +45 and 45 degrees. Future installations, however, ma var from these values suggesting that the transformation software should be general, as described in Eq. (3). Each wire-scanner will have an angle parameter in its database description. The angle will be stored in units of degrees and with a sign conforming to Fig 2. Presentl all new LCLS wire scanners will be of the +45 tpe (as shown at left in Fig. 2), with the eception of the three off-ais injector wire-scanners: WS01, WS02, and WS03. These three will be of the 45 tpe (as shown at right in Fig. 2). B OTR and YAG Screens The OTR and YAG screens intercept the electron beam and are used to measure its transverse size within a single beam pulse. The coordinate signs are important so that the beam can be predictabl steered across the screen when necessar, and all camera observation angles are important in order to properl scale the camera piels into millimeters. Columns CCD Image Rows YAG Crstal Figure 3. A YAG crstal (or OTR screen) as seen with the electron beam emerging from the paper, showing the beamline coordinates (, ) in blue and the raw image piel coordinates (indices) in red. 5 of 8 that this is the correct version prior to use.

The imaging optics for all YAG and OTR locations is specified to image the horizontal and vertical beam directions onto the same respective camera directions. Due to reflections of the beam image in the optics and different orientations of the optics assembl with respect to the beam ais for the various profile monitor locations, a reversal in the image from screen to camera, both horizontal and vertical, ma occur. The image processing software shall perform the necessar image transformations to properl displa and store the images so that the horizontal and vertical ais direction in the image represents the respective ais direction in the beam. The ais directions in the beam are defined as shown in Fig. 3 with the beam emerging from the paper and the paper representing the plane of the YAG crstal or the plane normal to the intersection point of the OTR foil and the beam. The fact that the OTR foil has a 45 degree angle w.r.t. the beam makes no difference compared to the YAG monitors. The arra representation of images usuall has the first row of piels at the top with increasing row inde toward the bottom. This is reversed from the scaled vertical beam direction coordinate,, which increases in value toward the top. The stored images shall follow this usual row inde convention. Three pairs (horizontal and vertical) of numbers shall be stored in the database for each profile monitor to be used in the image processing algorithm: A sign (M = ±1, M = ±1) for image reversal (see Table 1) along the respective ais, a row/column inde of the nominal transverse beam ais center (d, d ) [2], and a calibration factor (C, C ), in mm/piel [3], to convert piel inde into metric coordinates. This represents two levels of coordinate processing, from raw image, to converted image, and finall to processed image, with metric beam position and size. The first step applies the reversal sign to each ais to create the converted image to be displaed and stored. The converted column and row indices (p col, p row ) are col col col ( M ) 2 p = M p + N, (5) row row 1 ( M ) 2 p = M p + N, (6) row 1 where (p col, p row ) are the raw image column & row indices, and (N col, N row ) are the number of columns & rows of the image. This means simpl to reverse the order of the rows or columns if the respective M or M is negative. The second steps ields ( M p d ) = C (7) col ( M p d ) row = C + (8) for the scaled beam coordinates (, ), in millimeters. The minus sign in Eq. (8) takes into account the different directions of the row inde, p row, and the vertical beam position,. The calibration factors C and C should in general be identical, ecept for OTRS1 and OTR30, where the CCD is tilted w.r.t. the optical ais. 6 of 8 that this is the correct version prior to use.

Table 1. List of image reversal signs M and M for all YAG and OTR beam profile monitors whose imaging optics has been designed (Table will be completed as the design effort proceeds). Screen M M Screen M M Screen M M YAG01 1 1 OTR1 1 1 OTR21 1 1 YAG02 1 1 OTR2 1 1 OTR_TCAV TBD TBD YAGG1 +1 1 OTR3 1 1 OTR30 TBD TBD YAG03 1 1 OTR4 1 1 OTR33 1 1 YAG04 1 1 OTRS1 TBD TBD OTRDMP TBD TBD YAGS1 1 1 OTR11 1 1 YAGS2 1 1 OTR12 1 1 Collimators The various beam halo collimators along the LCLS beamline are composed of independent left/right or top/bottom rectangular jaws, which are remotel adjustable (see Fig. 4). Figure 4. A horizontal (left) and vertical (right) collimator as seen down its bore with electron beam emerging from paper, showing the beamline coordinates (, ). The horizontal collimators move along the -ais and the vertical collimators move along the -ais. The right jaw moves closer to the beam for smaller positive values of and the top jaw moves 7 of 8 that this is the correct version prior to use.

closer to the beam for smaller positive values of. The settings and readback positions should be in millimeters. References [1] The BPM calibration factor (PCMM) is the stripline radius, in millimeters, divided b two. [2] These transverse beam ais centers, d and d, will be initiall set for the center of the screen, but will likel be adjusted based on laser alignment during commissioning. [3] These calibration factors will be based on optical measurements performed after the camera mounting positions are finalized. 8 of 8 that this is the correct version prior to use.