R&S EB500 Monitoring Receiver Specifications

Similar documents
R&S RSC Step Attenuator Specifications

R&S ZVT Vector Network Analyzer Specifications

R&S EM510 HF Digital Wideband Receiver Efficient and versatile solution for radiomonitoring

R&S ZV-Z135 Calibration Kit Specifications

R&S ZNC Vector Network Analyzer Specifications

R&S ZNB Vector Network Analyzer Specifications

R&S ZV-Z3xx T-Checker Specifications

R&S ZV-Z5x Calibration Units Specifications

R&S ZN-Z5x Calibration Units Specifications. Data Sheet V03.00

R&S ZN-Zxxx Calibration Units Specifications. Data Sheet V04.00

Analog Modulation Analysis (AM/FM/φM) Specifications

R&S FSC Spectrum Analyzer Specifications

R&S ZNBT8 Vector Network Analyzer Specifications

R&S ZV-Z5x Calibration Units Specifications

R&S ZVT Vector Network Analyzer Specifications

R&S FU129 Antenna Filter Unit Antenna switching, rotator control and signal attenuation, amplification and filtering

R&S ESCI/ESCI7 EMI Test Receiver Specifications

R&S MNT100 RF Interference Locator Specifications

Monitoring Receiver R&S ESMB

R&S SMC100A Signal Generator Specifications

Noise Figure Measurement Applications Specifications

R&S EB500 Monitoring Receiver High-performance radiomonitoring from 9 khz to 6 GHz

R&S FPC Spectrum Analyzer Specifications

Spectrum Analyzer FSL

R&S FSH4/8 Spectrum Analyzer Specifications

R&S ESRP EMI Test Receiver Specifications

R&S FS-K112PC NFC Measurement Software Specifications

R&S CBT/R&S CBT32 Bluetooth Tester Specifications

Signal Generator SMA 100A

Vector Network Analyzers ZVB

R&S ZVL Vector Network Analyzer Specifications

R&S SMB100N SIGNAL GENERATOR

EB200 Miniport Receiver

R&S EB500 Monitoring Receiver High-performance radiomonitoring from 9 khz to 6 GHz

R&S FSV Signal and Spectrum Analyzer Specifications

Radiomonitoring & Radiolocation Product Brochure R&S EM100 Digital Compact Receiver Compact radiomonitoring from 9 khz to 7.

R&S FSWP Phase Noise Analyzer Specifications


R&S FSWP Phase Noise Analyzer Specifications

R&S FSV Signal Analyzer Specifications

R&S UMS175 Compact Radiomonitoring System High performance for outdoor deployment

R&S ESR EMI Test Receiver Specifications

ESU EMI Test Receiver

RF Signal Generator SM300

R&S ENV216 Two-Line V-Network For disturbance voltage measurements on single-phase EUTs

R&S FSWP Phase Noise Analyzer Specifications

LabWindows/CVI, VXIpnp driver history for the R&S Radio Tester

R&S ZVH Cable and Antenna Analyzer Specifications

Spectrum Analyzer R&S FS300

Modulation Analyzer FMAB

Spectrum Analyzer FSU

DSA700 Series Spectrum Analyzer

R&S ZNA Vector Network Analyzer Specifications

R&S FSUP Signal Source Analyzer Specifications

Signal Generator SML

Product Brochure Version R&S EB510 HF Monitoring Receiver High-performance radiomonitoring from 8 khz to 32 MHz

Data Sheet Version R&S RT-ZVCxx Multi-Channel Power Probe Specifications

Evolution of the Modern Receiver in a Crowded Spectrum Environment White Paper

Chapter 5 Specifications

R&S ESRP EMI Test Receiver Specifications

Attenuators and Matching Pads, Terminations 75 mw to 1000 W, DC to 18 GHz

Spectrum Analyzer FSU

R&S CMU-Z10/-Z11 Antenna Coupler/ RF Shielding Cover Simple interference-free testing of all mobiles

Voltage Sensors URV5-Z

R&S SMA100A Signal Generator The new standard of excellence in the analog signal generator class

Signal Generator SMA100A

Provided by R&S FSH4/FSH8 Spectrum Analyzer Specifications

Model 865 RF / Ultra Low Noise Microwave Signal Generator

Keysight Technologies N9320B RF Spectrum Analyzer

Receiver requirements for a TDOA-based radiolocation system

Agilent N9320B RF Spectrum Analyzer

R&S ZVA-Zxx Millimeter-Wave Converters Network analysis up to 500 GHz

Spectrum Analyzer R&S FSP

Handheld Spectrum Analyzer R&S FSH khz to 3 GHz

Digital HF Receiver WJ-8723

Signal Generators SMY

R&S NRP USB and LAN Power Sensors Specifications

SMA100A Signal Generator

Power Added Efficiency Measurement with R&S ZNB/ R&S ZVA

Signal Generator SMA100A

Handheld Spectrum Analyzer R&S FSH3

Model 845-M Low Noise Synthesizer

: Triple PLL, lowest reference frequency 10 khz. : ± 5 khz in 10 Hz steps, synthesized.

R&S MNT100 RF Interference Locator

SignalShark Real-Time Remote Analyzer

R&S ETH Handheld TV Analyzer Specifications

R&S ESR EMI Test Receiver Specifications

Noise Figure Measurement in the 60 GHz Range Application Note

R&S ZVH Cable and Antenna Analyzer Specifications

FMU36 Baseband Signal Analyzer

R&S FSWP Phase Noise Analyzer Specifications

R&S NRPM Over-the-Air (OTA) Power Measurement Solution For 5G, WLAN IEEE ad and IEEE ay

SIGNAL GENERATORS. MG3633A 10 khz to 2700 MHz SYNTHESIZED SIGNAL GENERATOR GPIB

Audio Analyzer R&S UPV. Preliminary Specifications

Audio Analyzers UP 300/UP 350

Frequency range 100 khz to 3 GHz 100 khz to 6 GHz 10 MHz to 18 GHz Reference frequency. 2 ppm in addition 2 ppm/10 C. 0 Hz, 100 Hz to 3 GHz

R&S EM100 Digital Compact Receiver Compact radiomonitoring from 9 khz to 7.5 GHz

Technical Data. Digital HF Receiver WJ-8712A WATKINS-JOHNSON. Features

SC5407A/SC5408A 100 khz to 6 GHz RF Upconverter. Datasheet. Rev SignalCore, Inc.

R&S TS8997 Regulatory Test System for Wireless Devices

Transcription:

Radiomonitoring & Radiolocation Data Sheet 01.02 R&S EB500 Monitoring Receiver Specifications

CONTENTS Definitions... 3 Specifications... 4 Frequency...4 Linearity...4 Interference rejection...4 Noise figure120...4 Sensitivity...5 Demodulation...5 IF bandwidth...5 ADC resolution...5 Level and offset measurement...6 IF panorama...6 Modulation measurement (with R&S EB500-IM option)...6 Scan characteristics...6 Antenna input...7 Input and output...7 General data...8 Ordering information... 9 2 Rohde & Schwarz R&S EB500 Monitoring Receiver

Definitions General Product data applies under the following conditions: Three hours storage at ambient temperature followed by 30 minutes warm-up operation Specified environmental conditions met Recommended calibration interval adhered to All internal automatic adjustments performed, if applicable Specifications with limits Represent warranted product performance by means of a range of values for the specified parameter. These specifications are marked with limiting symbols such as <,, >,, ±, or descriptions such as maximum, limit of, minimum. Compliance is ensured by testing or is derived from the design. Test limits are narrowed by guard bands to take into account measurement uncertainties, drift and aging, if applicable. Specifications without limits Represent warranted product performance for the specified parameter. These specifications are not specially marked and represent values with no or negligible deviations from the given value (e.g. dimensions or resolution of a setting parameter). Compliance is ensured by design. Typical data (typ.) Characterizes product performance by means of representative information for the given parameter. When marked with <, > or as a range, it represents the performance met by approximately 80 % of the instruments at production time. Otherwise, it represents the mean value. Nominal values (nom.) Characterize product performance by means of a representative value for the given parameter (e.g. nominal impedance). In contrast to typical data, a statistical evaluation does not take place and the parameter is not tested during production. Measured values (meas.) Characterize expected product performance by means of measurement results gained from individual samples. Uncertainties Represent limits of measurement uncertainty for a given measurement. Uncertainty is defined with a coverage factor of 2 and has been calculated in line with the rules of the Guide to the Expression of Uncertainty in Measurement (GUM), taking into account environmental conditions, aging, wear and tear. Typical data as well as nominal and measured values are not warranted by Rohde & Schwarz. Rohde & Schwarz R&S EB500 Monitoring Receiver 3

Specifications Frequency Frequency range base unit 20 MHz to 3.6 GHz with R&S EB500-HF option 9 khz to 3.6 GHz with R&S EB500-FE option 20 MHz to 6 GHz with R&S EB500-HF and 9 khz to 6 GHz R&S EB500-FE option Frequency resolution 1 Hz BFO 0 Hz to ±8 khz Frequency accuracy 1 10 6 Input for external reference 10 MHz Synthesizer settling time 1 ms, 900 µs (typ.) Oscillator phase noise 9 khz to 32 MHz 130 dbc (1 Hz) at 1 khz offset, 140 dbc (1 Hz) (typ.) 20 MHz to 650 MHz 120 dbc (1 Hz) at 10 khz offset 650 MHz to 6 GHz 103 dbc (1 Hz) at 10 khz offset Linearity Second-order intercept point (SOI) 9 khz to 32 MHz 20 MHz to 6 GHz Third-order intercept point (TOI) 9 khz to 32 MHz 20 MHz to 6 GHz Interference rejection Image rejection IF rejection low distortion mode 1 MHz f 32 MHz 70 dbm, 80 dbm (typ.) 55 dbm, 70 dbm (typ.) low distortion mode 50 dbm (typ.) 40 dbm (typ.) low distortion mode (150 khz spacing at 5 dbm) 1 MHz f 32 MHz 30 dbm, 35 dbm (typ.) 9 khz f < 1 MHz 30 dbm (typ.) (150 khz spacing at 15 dbm) 1 MHz f 32 MHz 20 dbm, 28 dbm (typ.) 9 khz f < 1 MHz 20 dbm (typ.) low distortion mode (2 MHz spacing at 20 dbm) 20 MHz f 650 MHz 15 dbm, 19 dbm (typ.) 650 MHz f 6 GHz 10 dbm, 14 dbm (typ.) (2 MHz spacing at 30 dbm) 20 MHz f 650 MHz 7 dbm, 10 dbm (typ.) 650 MHz f 6 GHz 0 dbm, 5 dbm (typ.) 9 khz f 32 MHz direct reception (no image frequency present) 20 MHz f 6 GHz 80 db, 90 db (typ.) 9 khz f 32 MHz direct reception (no IF present) 20 MHz f 6 GHz 80 db Noise figure 9 khz to 32 MHz 400 khz f 30 MHz 15 db, 12 db (typ.) f > 30 MHz 18 db, 16 db (typ.) low distortion mode 20 db (typ.) 20 MHz to 3.6 GHz 14 db, 10 db (typ.) low distortion mode 20 db (typ.) 3.6 GHz to 6 GHz 20 db, 15 db (typ.) 4 Rohde & Schwarz R&S EB500 Monitoring Receiver

Sensitivity Sensitivity (demodulation) 9 khz to 32 MHz 20 MHz to 3.6 GHz 3.6 GHz to 6 GHz AM: bandwidth = 6 khz, SINAD = 12 db, f mod = 1 khz, m = 0.5 FM: bandwidth = 9 khz, SINAD = 20 db, f mod = 1 khz, deviation = 3 khz SSB: bandwidth = 2.4 khz, SINAD = 12 db, f = 1 khz CW: bandwidth = 600 Hz, SINAD = 12 db AM: bandwidth = 6 khz, SINAD = 12 db, f mod = 1 khz, m = 0.5 FM: bandwidth = 15 khz, SINAD = 20 db, f mod = 1 khz, deviation = 5 khz SSB: bandwidth = 2.4 khz, SINAD = 12 db, f = 1 khz CW: bandwidth = 600 Hz, SINAD = 12 db AM: bandwidth = 6 khz, SINAD = 12 db, f mod = 1 khz, m = 0.5 FM: bandwidth = 15 khz, SINAD = 20 db, f mod = 1 khz, deviation = 5 khz SSB: bandwidth = 2.4 khz, SINAD = 12 db, f = 1 khz CW: bandwidth = 600 Hz, SINAD = 12 db measurement using telephone filter in line with ITU-T 107 dbm ( 1 µv), 400 khz f 32 MHz 107 dbm ( 1 µv), 400 khz f 32 MHz 113 dbm ( 0.5 µv), 400 khz f 32 MHz 119 dbm ( 0.25 µv), 400 khz f 32 MHz 107 dbm ( 1 µv) 107 dbm ( 1 µv) 113 dbm ( 0.5 µv) 119 dbm ( 0.25 µv) 101 dbm ( 2 µv) 101 dbm ( 2 µv) 107 dbm ( 1 µv) 113 dbm ( 0.5 µv) Demodulation Demodulation modes all IF bandwidths AM, FM, PM, PULSE, I/Q IF bandwidths 9 khz LSB, USB, CW IF bandwidths 1 khz ISB Squelch (squelch level) in 1 db steps 30 dbμv to +120 dbμv Gain control AGC, MGC, 130 db AGC mode FAST/DEFAULT/SLOW MGC settable in 1 db steps Automatic frequency control (AFC) automatic retuning for frequency-unstable signals ± ½ IF bandwidth (100 Hz to 5 MHz) Audio filter notch/noise reduction/bandpass IF bandwidth Bandwidth Shape factor ADC resolution Resolution of A/D converter demodulation, level and offset measurement (3 db bandwidth), 29 filters 100/150/300/600 Hz, 1/1.5/2.1/2.4/2.7/3.1/4/4.8/6/9/12/15/30/50/ 120/150/250/300/500/800 khz, 1/1.25/1.5/2/5 MHz 3 db:60 db 1:1.7 for filter up to 2 MHz 3 db:50 db 1:1.6 for 5 MHz filter 16 bit Rohde & Schwarz R&S EB500 Monitoring Receiver 5

Level and offset measurement Signal level Indication error Level detector Measurement modes Measurement time Frequency offset 30 dbμv to 120 dbμv, resolution 0.1 db max. ±3 db, ±1 db (typ.) AVG, PEAK, FAST, RMS Continuous, Periodic 0.5 ms to 900 s or automatic up to ± ½ IF bandwidth (100 Hz to 5 MHz), resolution 1 Hz IF panorama FFT IF panorama IF panorama span Panorama display up to 4096-point FFT operating mode dynamic, overlapping FFT automatic or variable with selectable frequency resolution 0.625/1.25/2.5/3.125/6.25/12.5/25/31.25/ 50/62.5/100/125/200/250/312.5/500/ 625 Hz, 1/1.25/2/2.5/3.125/5/6.25/8.333/10/12.5/ 20/25/50/100/200/500 khz, 1 MHz, 2 MHz 1/2/5/10/20/50/100/200/500 khz, 1/2/5/10/20 MHz ClearWrite, Average, MaxHold, MinHold Modulation measurement (with R&S EB500-IM option) AM (modulation index) Indication error FM (FM deviation) Indication error φm Indication error Bandwidth measurement AM, AM+, AM, m = 0 % to 999.9 %, resolution 0.1 %, f max = 2.5 MHz < 5 % for bandwidth 1 MHz, < 7 % for bandwidth > 1 MHz (S/N > 40 db, AF = 1 khz, measurement time < 1 s) FM, FM+, FM, f = 0 Hz to 2.5 MHz, resolution 1 Hz, f max = 2.5 MHz (f mod + deviation) < 2 % of IF bandwidth used (absolute) (S/N > 40 db, AF = 1 khz, measurement time < 1 s) φ = 0 rad to 12.5 rad, resolution 0.01 rad, f max = 2.5 MHz (f mod + deviation) < (0.1 rad + 3 % of display) (S/N > 40 db, AF = 1 khz, measurement time < 1 s) automatic up to 20 MHz bandwidth, x db and ß% method Scan characteristics Memory scan Frequency scan Panorama scan speed speed with R&S EB500-PS option speed 10000 programmable memory locations up to 500 channel/s user-selectable start/stop frequency and step width up to 500 channel/s RF spectrum with user-selectable start/stop frequency and step width: 100/125/200/250/500/625 Hz, 1/1.25/2/2.5/3.125/5/6.25/8.333/10/12.5/ 20/25/50/100/200/500 khz, 1 MHz, 2 MHz up to 12 GHz/s 6 Rohde & Schwarz R&S EB500 Monitoring Receiver

Antenna input Antenna input VSWR Input level Max. input level (non-destructive) Oscillator reradiation at antenna input Input selection HF 1 input, N female, 50 Ω HF/VHF/UHF combined 1 input, N female, 50 Ω 9 khz to 3.6 GHz 2 3.6 GHz to 6 GHz 2.5 137 dbm to +10 dbm +15 dbm 90 dbm, 100 dbm (typ.) 9 khz to 400 khz lowpass filter 400 khz to 32 MHz highpass/lowpass filter, switchable 20 MHz to 1.5 GHz tracking preselection 1.5 GHz to 6 GHz bandpass filter switchable attenuation manual or automatic 25 db in 5 db steps from 400 khz to 32 MHz (HF tuner) 40 db in 1 db steps from 20 MHz to 6 GHz (VHF/UHF tuner) Input and output Input External reference 10 MHz input level: 0 dbm to 10 dbm Control signals serial, GPS/compass, trigger, blank Output Internal reference 10 MHz output level: 3 dbm to 7 dbm Video A, video B analog video, AM (A) and FM (B), DC to ½ IF bandwidth or analog IF, controlled, 2 channels, adjustable center frequency 0 Hz to 70 MHz, bandwidth equal to IF filter width, level 11 dbm Digital video LAN, bandwidth 5 MHz Digital I/Q LAN, bandwidth 5 MHz Analog audio headphone connector 0 V to 2 V, R i = 100 Ω, f min = 10 Hz to 300 Hz, f max = 12.5 khz (depending on IF filter and modulation) AF line 0.5 V ± 0.2 V (m = 0.5), R i = 100 Ω, f min = 10 Hz to 300 Hz, f max = 12.5 khz (depending on IF filter and modulation) AF balance 0.4 V ± 0.2 V (m = 0.5), R i = 600 Ω, f min = 100 Hz, f max = 12.5 khz Control signals signal > squelch threshold, COR, antenna control LAN for data transfer and remote control 1 Gbit LAN interface (Ethernet 10/100/1000BaseT) USB for external storage 1 USB port (model.03 only) Rohde & Schwarz R&S EB500 Monitoring Receiver 7

General data Environmental data Climatic conditions in line with EN 60068-2-1, EN 60068-2-2 Operating temperature range 0 C to +50 C Permissible temperature range model.02 10 C to + 55 C model.03 0 C to +55 C Storage temperature range model.02 40 C to +70 C model.03 30 C to +70 C with AC/DC adapter 20 C to +70 C Humidity (non-condensing) model.02 max. 95 %, +25 C/+55 C model.03 max. 80 %, +25 C/+40 C Environmental tests Shock in line with EN 60068-2-27, MIL-STD-810E, method 516.4, procedure I Vibration sinusoidal 5 Hz to 150 Hz, in line with EN 60068-2-6 random 10 Hz to 500 Hz, in line with EN 60068-2-64 Electromagnetic compatibility (EMC) in line with EN 300339, ETSI EN 301489-1, ETSI EN 301489-22 MTBF (IEC 1709) model.02 > 35000 h model.03 > 17000 h Power supply AC 100 V to 240 V 50 Hz to 60 Hz, via external AC/DC supply DC 10 V to 32 V, 12 A to 4 A (depending on options installed) Mechanical data Dimensions W H D 213 mm 132 mm 450 mm (8.39 in 5.20 in 17.712 in), ½ 19", 3 HU Weight approx. 7.5 kg (16.53 lb) (depending on options installed) 8 Rohde & Schwarz R&S EB500 Monitoring Receiver

Ordering information Designation Type, description Order No. Monitoring Receiver, with control front R&S EB500 4072.5004.03 panel frequency range from 20 MHz to 3.6 GHz, IF spectrum (max. 20 MHz), remote control software supplied with receiver Monitoring Receiver, without control front R&S EB500 4072.5004.02 panel frequency range from 20 MHz to 3.6 GHz, IF spectrum (max. 20 MHz), remote control software supplied with receiver Hardware options HF Frequency Range Extension R&S EB500-HF 4072.8003.02 frequency range from 9 khz to 32 MHz Software options SHF Frequency Range Extension R&S EB500-FE 4072.9300.02 frequency range from 3.6 GHz to 6 GHz Panorama Scan R&S EB500-PS 4072.9200.02 RF scan, high-speed FFT scan across user-selectable range, selectable spectral resolution ITU Measurement Option R&S EB500-IM 4072.9100.02 ITU-compliant measurement of AM/FM-modulated signals in the R&S EB500 Software Package in line with R&S GX430IS 4071.5817.02 ITU-R SM.1600 (PC-based Software for ITU-compliant measurement of signals with analog or digital modulation) Accessories 19" Rack Adapter R&S ZZA-T04 1109.4187.00 for mounting two R&S EB500 in a 19" rack (both receivers side by side) 19" Rack Adapter R&S ZZA-T02 1109.4164.00 for mounting one R&S EB500 in a 19 rack (including one blind plate) Service options Two-Year Calibration Service Three-Year Calibration Service Five-Year Calibration Service One-Year Repair Service following the warranty period Two-Year Repair Service following the warranty period Four-Year Repair Service following the warranty period R&S CO2EB500 R&S CO3EB500 R&S CO5EB500 R&S RO2EB500 R&S RO3EB500 R&S RO5EB500 Please contact your local Rohde & Schwarz sales office. For product brochure, see PD 5214.3800.12 and www.rohde-schwarz.com. Rohde & Schwarz R&S EB500 Monitoring Receiver 9

10 Rohde & Schwarz R&S EB500 Monitoring Receiver

Rohde & Schwarz R&S EB500 Monitoring Receiver 11

Service you can rely on J Worldwide J Local and personalized J Customized and flexible J Uncompromising quality J Long-term dependability About Rohde & Schwarz Rohde & Schwarz is an independent group of companies specializing in electronics. It is a leading supplier of solutions in the fields of test and measurement, broadcasting, radiomonitoring and radiolocation, as well as secure communications. Established more than 75 years ago, Rohde & Schwarz has a global presence and a dedicated service network in over 70 countries. Company headquarters are in Munich, Germany. Environmental commitment Energy-efficient products Continuous improvement in environmental sustainability ISO 14001-certified environmental management system Certified Quality System ISO 9001 Rohde & Schwarz GmbH & Co. KG www.rohde-schwarz.com Regional contact Europe, Africa, Middle East +49 89 4129 123 45 customersupport@rohde-schwarz.com North America 1 888 TEST RSA (1 888 837 87 72) customer.support@rsa.rohde-schwarz.com Latin America +1 410 910 79 88 customersupport.la@rohde-schwarz.com Asia/Pacific +65 65 13 04 88 customersupport.asia@rohde-schwarz.com R&S is a registered trademark of Rohde & Schwarz GmbH & Co. KG Trade names are trademarks of the owners Printed in Germany (sk) PD 5214.3800.22 Version 01.02 February 2011 R&S EB500 Subject to change 2010-2011 Rohde & Schwarz GmbH & Co. KG 81671 München, Germany 5214380022