Undulator K-Parameter Measurements at LCLS

Similar documents
Seeding at LCLS FEL. J. Welch, (SLAC) J. Welch (SLAC), Joint DESY and University of Hamburg Accelerator Physics Seminar, Feb. 5, 2013, DESY Hamburg

LCLS-II SXR Undulator Line Photon Energy Scanning

3 General layout of the XFEL Facility

FLASH at DESY. FLASH. Free-Electron Laser in Hamburg. The first soft X-ray FEL operating two undulator beamlines simultaneously

Note on the LCLS Laser Heater Review Report

LCLS UNDULATOR COMMISSIONING, ALIGNMENT, AND PERFORMANCE *

Demonstration of exponential growth and saturation at VUV wavelengths at the TESLA Test Facility Free-Electron Laser. P. Castro for the TTF-FEL team

A Study of undulator magnets characterization using the Vibrating Wire technique

R&D Toward Brighter X-ray FELs

Using Higher Order Modes in the Superconducting TESLA Cavities for Diagnostics at DESY

Zhirong Huang. May 12, 2011

ELECTRON BEAM DIAGNOSTICS AND FEEDBACK FOR THE LCLS-II*

Introduction to the Physics of Free-Electron Lasers

Physics Requirements Document Document Title: SCRF 1.3 GHz Cryomodule Document Number: LCLSII-4.1-PR-0146-R0 Page 1 of 7

Set Up and Test Results for a Vibrating Wire System for Quadrupole Fiducialization

The Potential for the Development of the X-Ray Free Electron Laser

arxiv: v1 [physics.acc-ph] 20 Jan 2010

Check the LCLS Project website to verify 2 of 7 that this is the correct version prior to use.

Photon Beamlines and Diagnostics at LCLS

Generation of Coherent X-Ray Radiation Through Modulation Compression

Supplementary Information

Stretched Wire Test Setup 1)

Introduction to High-Resolution Accelerator Alignment Using X-ray Optics

Infrared Single Shot Diagnostics for the Longitudinal. Profile of the Electron Bunches at FLASH. Disputation

FLASH 2. FEL seminar. Charge: 0.5 nc. Juliane Rönsch-Schulenburg Overview of FLASH 2 Hamburg,

THz Pump Beam for LCLS. Henrik Loos. LCLS Hard X-Ray Upgrade Workshop July 29-31, 2009

Physics Laboratory Scattering of Photons from Electrons: Compton Scattering

FLASH II: an Overview

Extending the photon energy coverage of an x-ray self-seeding FEL. via the reverse taper enhanced harmonic generation technique

Paul R. Bolton and Cecile Limborg-Deprey, Stanford Linear Accelerator Center, MS-18, 2575 Sandhill Road, Menlo Park, California

Precision RF Beam Position Monitors for Measuring Beam Position and Tilt Progress Report

Instructions for the Experiment

arxiv:physics/ v1 [physics.acc-ph] 18 Jul 2003

FLASH Operation at DESY From a Test Accelerator to a User Facility

Sources & Beam Line Optics

RANDY W. ALKIRE, GEROLD ROSENBAUM AND GWYNDAF EVANS

X-Ray Transport, Diagnostic, & Commissioning Plans. LCLS Diagnostics and Commissioning Workshop

Status, perspectives, and lessons from FLASH and European XFEL

Bioimaging of cells and tissues using accelerator-based sources

OVERVIEW OF SEEDING METHODS FOR FELS

Commissioning the Echo-Seeding Experiment ECHO-7 at NLCTA

Transverse Wakefields and Alignment of the LCLS-II Kicker and Septum Magnets

X-ray Transport Optics and Diagnostics Commissioning Report

Measurements of MeV Photon Flashes in Petawatt Laser Experiments

On-line spectrometer for FEL radiation at

arxiv: v1 [physics.acc-ph] 6 Apr 2016

FLASH II. FLASH II: a second undulator line and future test bed for FEL development.

FLASH Upgrade. Decrease wavelength and/or increase brilliance

LCLS Injector Diagnostics. Henrik Loos. Diagnostics overview Transverse Beam Properties Longitudinal Beam Properties

Feedback Requirements for SASE FELS. Henrik Loos, SLAC IPAC 2010, Kyoto, Japan

12/08/2003 H. Schlarb, DESY, Hamburg

HIGHER ORDER MODES FOR BEAM DIAGNOSTICS IN THIRD HARMONIC 3.9 GHZ ACCELERATING MODULES *

BLACKBODY RADIATION PHYSICS 359E

LCLS project update. John Arthur. LCLS Photon Systems Manager

Shintake Monitor Nanometer Beam Size Measurement and Beam Tuning

Challenges of Optics for High Repetition Rate XFEL Source

NIST EUVL Metrology Programs

Experimental Physics. Experiment C & D: Pulsed Laser & Dye Laser. Course: FY12. Project: The Pulsed Laser. Done by: Wael Al-Assadi & Irvin Mangwiza

I = I 0 cos 2 θ (1.1)

Beam Diagnostics, Low Level RF and Feedback for Room Temperature FELs. Josef Frisch Pohang, March 14, 2011

CHAPTER 5 FINE-TUNING OF AN ECDL WITH AN INTRACAVITY LIQUID CRYSTAL ELEMENT

Influences of a Beam-Pipe Discontinuity on the Signals of a Nearby Beam Position Monitor (BPM)

Nonintercepting Diagnostics for Transverse Beam Properties: from Rings to ERLs

Nano Beam Position Monitor

Attosecond Diagnostics of Muti GeV Electron Beams Using W Band Deflectors

Nuclear Instruments and Methods in Physics Research A

SIGNAL ELECTRIC FIELD MAGNETIC FIELD # 1 (#2) #3 (# 4) WAVEGUIDE VACUUM CHAMBER BEAM PIPE VACUUM CHAMBER

Experience with Insertion Device Photon Beam Position Monitors at the APS

CHAPTER 9 POSITION SENSITIVE PHOTOMULTIPLIER TUBES

Thermal tuning of volume Bragg gratings for high power spectral beam combining

A Prototype Wire Position Monitoring System

Evidence of High Harmonics from Echo-Enabled Harmonic Generation for Seeding X-ray Free Electron Lasers

Initial Beam Phasing of the SRF Cavities in LCLS-II

Electron Beam Diagnosis Using K-edge Absorp8on of Laser-Compton Photons

An Overview of MAX IV Insertion Devices & Magnetic Measurement System. Hamed Tarawneh On behalf of Insertion Devices Team

Attenuation length in strip scintillators. Jonathan Button, William McGrew, Y.-W. Lui, D. H. Youngblood

Outline of the proposed JLAMP VUV/soft X-ray FEL and the challenges for the photon beamlines and optics

Laser Beam Analysis Using Image Processing

Calibration of Scintillator Tiles with SiPM Readout

KULLIYYAH OF ENGINEERING

ANALYSIS OF 3RD OCTAVE BAND GROUND MOTIONS TRANSMISSION IN SYNCHROTRON RADIATION FACILITY SOLARIS Daniel Ziemianski, Marek Kozien

1-Å FEL Oscillator with ERL Beams

X-ray FEL Oscillator (XFEL-O) Gun Requirements and R&D Overview FLS2010: WG5: High Brightness Guns March 1, 2010

SUPPLEMENTARY INFORMATION

Advanced Beam Instrumentation and Diagnostics for FELs

Mode-locked multichromatic x-rays in a seeded free-electron laser for single-shot x-ray spectroscopy

M. Senoner 1), Th. Wirth 1), W. E. S. Unger 1), M. Escher 2), N. Weber 2), D. Funnemann 3) and B. Krömker 3) INTRODUCTION

Ultra-stable flashlamp-pumped laser *

HOM Based Diagnostics at the TTF

SECOND HARMONIC GENERATION AND Q-SWITCHING

PERFORMANCE OF THE CMS ECAL LASER MONITORING SOURCE IN THE TEST BEAM

National Accelerator Laboratory

BL39XU Magnetic Materials

Spectral characterization of the FERMI pulses in the presence of electron-beam phase-space modulations

Radiation Detection by Cerenkov Emission in. Optical Fibers at TTF

VIBRATING WIRE SENSORS FOR BEAM INSTRUMENTATION Suren Arutunian

H. Weise, Deutsches Elektronen-Synchrotron, Hamburg, Germany for the XFEL Group

Performance of the SASE3 monochromator equipped with a provisional short grating. Variable line spacing grating specifications

TECHNICAL CHALLENGES OF THE LCLS-II CW X-RAY FEL *

Incident IR Bandwidth Effects on Efficiency and Shaping for Third Harmonic Generation of Quasi-Rectangular UV Longitudinal Profiles *

Transcription:

Undulator K-Parameter Measurements at LCLS J. Welch, A. Brachmann, F-J. Decker, Y. Ding, P. Emma, A. Fisher, J. Frisch, Z. Huang, R. Iverson, H. Loos, H-D. Nuhn, P. Stefan, D. Ratner, J. Turner, J. Wu, D. Xiang, SLAC NAL, Menlo Park, Ca. USA R. Bionta, LLNL, Livermore Ca. USA H. Sinn, DESY, Hamburg, Germany Abstract We report our first in-situ measurements of relative undulator segment K-parameters made at the LCLS. The diagnostics, calibration, procedure, and early results are described. Measurement errors and noise are discussed and the outlook for near-term develop presented. INTRODUCTION At the highest design x-ray energy of 8 kev, where undulator field tolerances are most demanding, the LCLS started lasing within a few hours of the time the electron beam was first transported through the undulator, so undulator field quality must be well in hand. Nevertheless, even though advances in magnetic measurements and undulator tuning have produced an undulator of this quality [1], the field quality can change after installation. Field changes are known to occur due to radiation exposure and temperature variation. Similarly, errors in knowledge of the undulator position relative to the beam can occur and have the same net effect as field errors. Position fiducialization errors have occurred when segments of the LCLS undulator were accidentally exposed to temperature excursion of 5 C. Undulator position errors would also result from diagnostic errors, or errors in configurations used to establish the positions of undulator segments. The in-situ undulator measurement techniques described in this paper are being developed to provide periodic assessment of the LCLS undulator field quality. The 130 m long LCLS undulator is composed of 33 nearly identical segments, each separately tuned to a precise strength denoted by its K-parameter. The K- parameter is dimensionless and K = 0.934λ u B [T ], where B is the amplitude of the on-axis magnetic field in Tesla, and λ u is undulator period in centimeters. Tolerance studies of the FEL process led to a tolerance budget based on FEL power for individual segment K parameters of δk K < 1.5 10 4, where δk is the RMS deviation from design. The design K-parameters vary slightly depending on where the segment is to be located, but a typical value is 3.5. We are developing two techniques to measure in-situ segment K parameters, both of which rely on measurements of the x-ray spectrum using a monochromator and Work supported in part by the DOE Contract DE-AC02-76SF00515. This work was performed in support of the LCLS project at SLAC. welch@slac.stanford.edu the ability to remotely move segments onto and off from the beamline. In the one-segment method, all segments are moved off the beamline except for a single reference segment and a x-ray energy spectrum is measured. The reference segment is then removed and a test segment is put in and its x-ray spectrum is measured. The difference in K between the two segments can be inferred from relative shifts in the spectra. At this early stage in commissioning, most of the measurements have been done with the onesegment technique. In the two-segment method, all segments are moved off the beamline except for two adjacent segments. One of the two segments is then stepped horizontally over a range of a few millimeters relative to the other, and a spectrum is taken at each position. By design, the gap of the undulator segments has a slight horizontal taper so that the effective K-parameter varies slightly with horizontal position. Based on simulation [2] we find that when the position is such that the effective K-parameters for the two segments is matched, the x-ray spectrum has the steepest slope at the point of inflection on the high energy side. Thus from the measured position of the match and knowledge of the rate of change of the effective K-parameter with position from magnetic measurements, we can calculate the difference in K-parameters for the two segments. Diagnostics MEASUREMENTS Insertable diagnostics used for K-measurements include a special monochromator called the K-monochromator, followed by a set of photodiodes, and a YAG screen/camera combination called the Direct Imager. The path of the x-rays through K-monochromator involves four Bragg silicon (111) reflections and forms a chicance around a tungsten block. The spontaneous radiation generated by the undulator contains very high energy x-rays, well over 100 kev, which pass through the silicon crystal of the K- monochromator and are absorbed in the tungsten block. Because of the multiple reflections, the K-monochromator will only transmit x-rays with a single angle of incidence, at a single energy and some of its harmonics. The angle of incidence was made to be remotely tunable. The device was tested at SSRL and found to pass x-rays with energy 8193 ev (0.15 nm), essentially the same as the first harmonic x-rays produced by a typical LCLS undulator segment (K = 3.497) when the electron beam is 13.56 GeV. The transmitted bandwidth was found to be 1.2 ev FWHM,

Figure 2: Effect of tuning the K-monochromator angle of incidence on the image of the transmitted x-rays.: left image was taken with angle at 3 µrad, center image at 0 µrad, and right image at +3µrad. Figure 1: The dependence of the photodiode signal (in this early measurement, it is a negative going signal and has an arbitrary scale) on the K-monochromator angle of incidence, also in arbitrary units. and the transmission angle was also measured and related to fiducials on the K-monochromator for installation alignment. Commissioning the K-monochromator included two measurements: (1) finding the electron beam energy, as defined by the strength of the LCLS bend magnets required for the ideal orbit, that generates first harmonic x-rays at 8193 ev; (2) finding the angle of incidence that passes the 8193 ev x-rays. Both are precision measurements. The electron beam energy for transmission was found by scanning the linac output energy ±100 MeV and watching for the appearance of an image on the Direct Imager. It turned out that the bend magnets were well calibrated as we found the machine energy of 13.64 GeV provides maximum transmission. We were also fortunate that the preset angle of incidence was close enough to the ideal that a two dimensional search of angle and energy was not necessary. The optimum angle of incidence was found in two different ways and yielded the same result. One method simply scanned the angle for maximum transmission while measuring the photodiode signal. The result is shown in Figure 1. Alternately, we found that the image of the transmitted x-rays, as seen by the Direct Imager, was very sensitive to the angle of incidence, and by adjusting the angle to symmetrize the image we could set the angle precisely. The images in Figure 2 show the effect of the angle tuning. The images could easily be distinguished to within 30 motor steps, which corresponds to an angular sensitivity of better than 3 microradians. Results The x-ray spectrum was measured by inserting the K- monochromator and the photodiode detector, and scanning the electron beam energy over a range of about 200 MeV. Changing the electron beam energy by a small amount, while keeping the undulator segment K fixed, results in a simple linear shift of the x-ray energy spectrum. Since the K-monochromator only transmits 8192 ev, the x-ray spectrum is effectively scanned by scanning the electron beam energy. Measured and simulated spectra/energy scans are shown in Figure 3, for the case where only the last undulator segment (U33) is on the beamline. The overall shape, width, and resonant electron beam energy, are in reasonable agreement between measurement and simulation. The measured spectrum has noticeably more noise than the simulation.this can come about from either noise in the photodiode signal or fluctuations in the electron beam energy. In fact we believe the latter is responsible for the majority of the excess noise. The effect of energy errors is discussed in the Systematics section below. The spectra in Figure 3 are over a range of 200 MeV. Normally for K-measurements we concentrate on scanning the rising edge of the spectrum as a function of electron beam energy, because it is insensitive to small changes in electron beam orbit angles and solid angle effects. K-measurement results presented in this paper used the one-segment technique referred to in the Introduction. So far the noise levels were too high for consistent twosegment results. First the reference segment is moved into its nominal horizontal position and the electron beam energy is scanned to obtain the x-ray spectrum. Rising edge data from 10% to 90% of the peak is extracted and fit to a third order curve and is plotted as the heavy red line in Figure 4. Next the test segment is moved onto the beamline and set at various equally spaced horizontal positions, typically ranging from 1 to +1 mm. The spectra from each of these positions are plotted as the green curves in Figure 4. Since the undulator gap is slightly non-parallel, with an average angle of 5.5 mrad, the K is slightly different for the different positions and the spectra are shifted in machine energy. Each of the green test spectra is analyzed to extract the electron beam energy that corresponds to the middle of the rising portion where the slope is steepest, Energy at Mid-slope. High noise levels can make the Energy at Mid-slope data sensitive to the exact region of the rising edge data that is included in the fit. In Figure 5, the Energy at Mid-slope data (circles) is fit to a second order polynomial and plotted against the test segment position along with the Energy at Mid-Slope for the reference segment. The energy where the fitted test data

Figure 4: Successive energy scans for different horizontal positions using the one-segment method. Figure 3: Measured (upper plot) and simulated(lower plot) x-ray spectrum as a function of machine energy. Figure 5: One-segment results of a K measurement scan. matches that of the reference segment defines the horizontal position of the test segment at which the two segments have the same spectrum. If there are no significant systematic effects, such as differences in electron energy or the x-ray integration angles, the match position and slope Energy at Mid-slope versus segment position, can be used to determine the difference in K-parameters for any horizontal position. For constant resonant photon energy and angle, dγ γ = K2 dk 2 + K 2 K 0.86dK K where γ is the electron beam energy in units of the electron rest mass. Early results are shown in Figure 5 were for segment 14 and segment 15. Measurements indicate the two segments will give the x-ray spectrum when segment 15 is at 0 mm and segment 14 is at 0.4 mm. The measured slope of the segment 14 mid-point energy versus position is approximately (dγ/γ)/dx = 7.7 10 4 mm 1. Prior to installation the gradient of K with respect to horizontal position (1) had been measured at the magnetic measurement laboratory to be 2.68 10 3 mm 1, which corresponds to an expected (dγ/γ)/dx = 6.6 10 4. The match position and the measured slope imply a K/K = 2.7 10 4 where K is the difference in K between segments when both segments are at 0 mm. A reproducibility measurement of the one-segment method was made and resulted with a standard deviation of the measurement of K/K of 4 10 4. Measurements of 14 pairs of segments were made so far and the results are consistent with all being within the required tolerance. As the techniques are developed and the two-segment technique is fully implemented, we expect the reproducibility to approach1 10 4. SOURCES OF ERRORS The undulator resonance equation is λ = λ u 2γ 2 (1 + K2 /2 + γ 2 θ 2 ), (2)

MeV.jpg Electron Beam Energy Figure 6: The central ray is determined when the electron beam energy is lowered until only the highest energy x-rays are at the passband energy. The electron beam energy for the right/left image is 15 MeV/10 MeV below 13.64 GeV. where θ is the observation angle of the x-rays relative to the direction of the electron beam in the undulator. In it, K, electron energy, and observation angle are related, so that measuring any two allows the calculation of the third, assuming the wavelength λ through the K-monochromator to be constant. This implies that to accurately determine K, one has to accurately know the angle and the energy. The discussion of systematics in this section concentrates on determining limits on the uncertainties of electron beam energy and observation angle. Observation Angle The Central Ray of the x-ray radiation is defined as the radiation emitted at θ = 0. Experimental determination of the Central Ray is necessary to control the K error associated with the observation angle uncertainty. For an ideal comparison of two spectra the central rays from both segments should coincide on the direct imager screen. In general the Central Rays from the different segment will not coincide due to slight deviations in the electron beam trajectory. We mitigated this effect by the technique we used to establish the shift of the x-ray spectrum from the electron beam energy at which the x-ray spectrum has the steepest positive slope. X-rays in this part of the spectrum come only from the smallest observation angles. In practice, if the electron beam is well aligned, we only have to be sure that we capture the Central Rays in the detectror to avoid significant error associated with observation angle. We determined the central ray by observing the image of the x-rays on the Direct Imager while the K- monochromator is inserted. When the electron beam energy is tuned such that only x-rays from very near θ = 0 are at 8193 ev, a very narrow image appears as can be seen in Figure 6. The sensitivity of this method is impressive. The position of the spot for 15 MeV in Figure 6 is determined to a statistical precision of around 3 microns, which at the distance of about 100 m from the last undulator segment is an angular uncertainty of 0.03 microradians. Since the central ray can be used to infer the angle the electron beam makes as it passes through each segment, it can in principle be used to determine electron beam trajectory straightness. Differences in the electron beam energy between the reference segment and the test segment also introduce shifts in the x-ray spectra, and should be accounted for high precision determination of K. Both spontaneous radiation and wakefield energy loss mechanism can contribute to this difference. There is about 0.63 MeV energy loss per electron from spontaneous radiation from a single segment for an electron beam energy of 13.64 GeV. In addition there is a wakefield loss which depends on peak current but not beam energy. For a peak current of 500 A, at which the K measurements are normally done, the total wakefield energy loss was measured to be 5 MeV, or 0.15 MeV per segment. In total, during a normal K measurement, the total energy loss from one segment to the next is about 0.78 MeV. From equation 1, the corresponding apparent K change is 2.3 10 4 ( K/K change of about 7 10 5.), with the downstream segment appearing to be lower K. Pulse-to-pulse fluctuations in the electron beam energy can be measured on a pulse-by-pulse basis and corrected for using beam position monitors in a dispersive region near the end of the linac [2]. In addition, fluctuations in peak bunch current can be measured and the resulting energy fluctuations due to impedance corrected on a pulse-bypulse basis. At the time of these measurements neither of these techniques was quite ready yet, but they are expected to be working soon. OUTLOOK The initial K-measurements made recently are encouraging. The spectrum of the undulator segments seen through the K-monochromator reasonably agrees with the simulation. The photodiode has good sensitivity and the electron energy scans can be done quickly. Systematic effects of observation angle and energy loss are well in hand. Noise levels are higher than expected, but by implementing pulse-by-pulse measurement of energy and peak bunch current, the noise associated with the electron beam energy can be reduced. This will help the fitting algorithm to more reliably find the electron energy with the steepest positive slope and better determine the spectral shift of each measurement. The initial noise levels were too high for good two segment results. When the noise is reduced the two-segment method in conjunction with the one-segment method should improve the confidence of the results as well. We would also like to try to look at K-measurement using different harmonics, not just the first. The higher harmonics should have narrower bandwidth and more sensitive to electron energy change. Higher harmonics can be tuned onto the transmission energy of the K-monochromator by reducing the electron beam energy or using higher order crystal reflections of the K-monochromator.

REFERENCES [1] Z. Wolf, V. Kaplounenko, Y. Levashov and A. Weidemann, LCLS undulator tuning and fiducialization, In the Proceedings of Particle Accelerator Conference (PAC 07), Albuquerque, New Mexico, 25-29 Jun 2007, pp 1320. [2] J. J. Welch et al., Precision measurement of the undulator K parameter using spontaneous radiation, FEL 2006, Berlin Germany, p 548-551.