EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow

Similar documents
Experience in implementing continuous arterial spin labeling on a commercial MR scanner

Works-in-Progress package Version 1.0. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B. January 22, 2003

Noninvasive Blood Flow Mapping with Arterial Spin Labeling (ASL) Paul Kyu Han and Sung-Hong Park

Slice profile optimization in arterial spin labeling using presaturation and optimized RF pulses

Multi-Slice Perfusion-Based Functional MRI using the FAIR Technique: Comparison of CBF and BOLD effects

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2009 December 1.

1 Introduction. 2 The basic principles of NMR

H 2 O and fat imaging

MRI Metal Artifact Reduction

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

BOLD fmri: signal source, data acquisition, and interpretation

HETERONUCLEAR IMAGING. Topics to be Discussed:

Pulse Sequence Design and Image Procedures

The SENSE Ghost: Field-of-View Restrictions for SENSE Imaging

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Cover Page. The handle holds various files of this Leiden University dissertation

MR Advance Techniques. Flow Phenomena. Class II

(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

IR/SR TrueFISP. Works-in-Progress package Version 1.2. For the SIEMENS Magnetom. Installation and User s Guide NUMARIS/4VA21B.

2014 M.S. Cohen all rights reserved

IMPROVING THE ROBUSTNESS OF PERFUSION IMAGING WITH ARTERIAL SPIN LABELING MAGNETIC RESONANCE IMAGING HUAN TAN

Cardiac MR. Dr John Ridgway. Leeds Teaching Hospitals NHS Trust, UK

Background (~EE369B)

Encoding of inductively measured k-space trajectories in MR raw data

High-Resolution, Spin-Echo BOLD, and CBF fmri at 4and7T

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

Image Quality/Artifacts Frequency (MHz)

functional MRI: A primer

Applications Guide. Spectral Editing with SVS. (Works-in-Progress) MAGNETOM TaTs and Verio Systems (3T)

2015 Spin echoes and projection imaging

Half-Pulse Excitation Pulse Design and the Artifact Evaluation

Simultaneous Multi-Slice (Slice Accelerated) Diffusion EPI

High Field MRI: Technology, Applications, Safety, and Limitations

MAGNETIC RESONANCE IMAGING

Efficiency of Background Suppression for Arterial Spin Labeling. Dairon Garcia

3T Unlimited. ipat on MAGNETOM Allegra The Importance of ipat at 3T. medical

Hardware. MRI System. MRI system Multicoil Microstrip. Part1

Functional MRI with variable echo time acquisition

Fast Field-Cycling Magnetic Resonance Imaging (FFC-MRI)

SIEMENS MAGNETOM Skyra syngo MR D13

TimTX TrueShape. The parallel transmit architecture of the future. Answers for life.

a. Use (at least) window lengths of 256, 1024, and 4096 samples to compute the average spectrum using a window overlap of 0.5.

Pulse Sequence Design Made Easier

SNR and functional sensitivity of BOLD and perfusion-based fmri using arterial spin labeling with spiral SENSE at 3 T

Advanced MSK MRI Protocols at 3.0T. Garry E. Gold, M.D. Associate Professor Department of Radiology Stanford University

Analysis of spatial dependence of acoustic noise transfer function in magnetic resonance imaging

ISSN X CODEN (USA): PCHHAX. The role of dual spin echo in increasing resolution in diffusion weighted imaging of brain

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2014 May 16.

A. SPECIFIC AIMS: phase graph (EPG) algorithms to cover a wide range of MRI

Magnetic Resonance Imaging Principles, Methods, and Techniques

MRI Summer Course Lab 2: Gradient Echo T1 & T2* Curves

Magnetization transfer attenuation of creatine resonances in localized proton MRS of human brain in vivo

ACRIN 6686 / RTOG 0825

Magnetic Resonance Imaging

High-Field Surface-Coil MR Imaging of Localized Anatomy

Numerical Evaluation of an 8-element Phased Array Torso Coil for Magnetic Resonance Imaging

MR Basics: Module 8 Image Quality

Evaluation of 2D Imaging Schemes for Pulsed Arterial Spin Labeling of the Human Kidney Cortex

MRI AT HIGH MAGNETIC FIELDS. Kâmil Uğurbil. University of Minnesota

Supplementary Figure 1. Scanning Electron Microscopy images of the pristine electrodes. (a) negative electrode and (b) positive electrode.

Masimo Corporation 40 Parker Irvine, California Tel Fax

Inherent Insensitivity to RF Inhomogeneity in FLASH Imaging

Improving high-field MRI using parallel excitation

NIH Public Access Author Manuscript Magn Reson Med. Author manuscript; available in PMC 2010 July 21.

Gradients. Effects of B0 gradients on transverse magnetisation Similar to figure 10 of Sattler review Progr. NMR 34 (1999), 93

Saturated Double-Angle Method for Rapid B 1 Mapping

The development of the RF-pulse for the low level SAR used by the MRI.

SUPPLEMENTARY INFORMATION

Steady-state sequences: Spoiled and balanced methods

In a typical biological sample the concentration of the solute is 1 mm or less. In many situations,

Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment

Implementation of parallel search algorithms using spatial encoding by nuclear magnetic resonance

Magnetic Resonance Research Facility (MRRF) Resources

RF Pulse Toolkit: Application Specific Design

Page 1 of 9. Protocol: adult_other_adni3_study_human_ge_3t_25w_ _ _1. 3 Plane Localizer. 3 Plane Localizer PATIENT POSITION

EE225E/BIOE265 Spring 2012 Principles of MRI. Assignment 7. Due March 16, 2012

Quantitative Measurements of Proton Spin-Lattice (T 1 ) and Spin Spin (T 2 ) Relaxation Times in the Mouse Brain at 7.0 T

Improve Image Quality of Transversal Relaxation Time PROPELLER and FLAIR on Magnetic Resonance Imaging

Selective Arterial Spin Labeling (SASL): Perfusion Territory Mapping of Selected Feeding Arteries Tagged Using Two-Dimensional Radiofrequency Pulses

Liver imaging beyond expectations with Ingenia

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

MR in RTP. MR Data for Treatment Planning: Spatial Accuracy Issues, Protocol Optimization, and Applications (Preview of TG117 Report) Acknowledgements

Downloaded from by on 02/07/18 from IP address Copyright ARRS. For personal use only; all rights reserved

Pulse Sequence for Multislice T 1 -Weighted MRI

Ultrasound physical principles in today s technology

Clear delineation of optic radiation and very small vessels using phase difference enhanced imaging (PADRE)

RAD 229: MRI Signals and Sequences

TITLE: Prostate Cancer Detection Using High-Spatial Resolution MRI at 7.0 Tesla: Correlation with Histopathologic Findings at Radical Prostatectomy

Compensation in 3T Cardiac Imaging Using Short 2DRF Pulses

MR Basics: Module 6 Pulse Sequences

COMMUNICATIONS Volume-Selective Multipulse Spin-Echo Spectroscopy

EE469B: Assignment 1 Solutions

Module 2. Artefacts and Imaging Optimisation for single shot methods. Content: Introduction. Phase error. Phase bandwidth. Chemical shift review

k y 2k y,max k x 2k x,max

MRI physics for SPM users

EE225E/BIOE265 Spring 2014 Principles of MRI. Assignment 6. Due Friday March 7th, 2014, Self Grading Due Monday March 10th, 2014

Complex-Valued Analysis of Arterial Spin Labeling Based Functional Magnetic Resonance Imaging Signals

Improvement of the Image Quality of Black-blood Magnetic Resonance Imaging with the Subtraction Double Inversion Recovery Technique

Passive Tracking Exploiting Local Signal Conservation: The White Marker Phenomenon

Signal Extraction Technology

Transcription:

EPISTAR MRI: Multislice Mapping of Cerebral Blood Flow Robert R. Edelman, Qun Chen A method is described for multislice EPISTAR that perfectly compensates magnetization transfer effects. lnflowing arterial spins are labeled with a 360" adiabatic pulse. Two control tags are applied sequentially at the same location as the labeling pulse, each with a 180" adiabatic pulse so the total RF irradiation, frequency shift, and bandwidth of the labeling and control pulses are identical. Therefore, magnetization transfer effects are the same as for the labeling pulse and cancel with image subtraction for all slices. The method also eliminates tagging of venous spins and concern about asymmetric magnetization transfer effects. INTRODUCTION Radiofrequency (RF) pulses can be used to tag inflowing arterial spins for the purpose of mapping cerebral blood flow (CBF). In the original approach, arterial spins were labeled by continuous adiabatic tagging (1). Subsequently, pulsed RF tagging methods were described, such as EPISTAR (2) and FAIR (3, 4). Potentially, spin labeling methods could have clinical uses, for instance to guide the treatment of stroke or to diagnose Alzheimer's disease (51. A key issue for spin labeling techniques is that the tagging RF pulse produces a magnetization transfer (MT) effect on brain tissue (6). The signal change from MT can be comparable to that from blood flow, so it is necessary to compensate. For EPISTAR imaging, a control tag is applied above the slice to balance MT effects from the inversion tag. Unfortunately, the utility of pulsed labeling methods like EPISTAR and FAIR has been limited because only a single slice can be imaged. The reason is that MT effects are only compensated at the center slice of a multislice acquisition. We demonstrate a straightforward modification of EPISTAR that solves this problem and permits multislice mapping of cerebral blood flow. METHODS In EPISTAR methods, inflowing arterial spins are labeled by an adiabatic RF pulse, after which an inflow time is permitted to elapse to allow the labeled spins to enter the arterial and capillary beds of the tissue of interest. The spins are imaged with an echo planar acquisition. For MRM 40800-805 (1998) From the Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts. Address correspondence to: Robert R. Edelman, Ph.D., Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, MA 02215. Received December 11, 1997; revised June 25, 1998; accepted July 13, I 998. 0740-3194/98 $3.00 Copyright 0 1998 by Lippincott Williams & Wilkins All rights of reproduction in any form reserved. 800 arterial labeling, the spins are inverted over a 90-mm region by a 23-ms duration hyperbolic secant inversion pulse (Fig. 1). The labeling pulse is placed below the slice with a slab-selection gradient of 1.8 millitesla/ meter and has a nominal 360' flip angle (7.4 mtesla, peak voltage = 53.30V). An adiabatic pulse will invert the spins as long as the flip angle is greater than or equal to 180". Use of a larger flip angle than 180' as done here still produces an inversion. In the original single slice EPISTAR method, a control RF pulse equal in amplitude and thickness to the labeling pulse was applied equidistant above the slice on alternate acquisitions to compensate for MT effects. The unique feature of the new EPISTAR method is in the different way that the control pulse is applied (Fig. 2). Instead of a single 360' pulse, two RF pulses are applied in rapid sequence, each with a 180' flip angle (3.7 mtesla, peak voltage = 26.65v) but otherwise identical to the labeling pulse. A 2-ms spoiler gradient is interposed between the two pulses to eliminate any potential interference of the residual transverse magnetization created by the first 180' pulse. Additionally, the control pulses are applied at the same position as the labeling pulse, rather than above the slice. The first control RF pulse inverts the inflowing spins, while the second RF pulse reinverts them to equilibrium magnetization. As a result, there is negligible labeling of the inflowing spins by the control pulse. Since the frequency shift, slice selection gradient, and total off-resonance RF irradiation absorbed by short T, species are identical for the labeling and control pulses, MT effects throughout the brain are exactly identical. Several phantom experiments have been conducted to evaluate the new approach. First of all, we measured the slice profile of the 180' and 360' pulses to demonstrate that the imperfection of RF slice profiles of these pulses do not contribute to signal variations. Secondly, to confirm that MT effects are identical for the labeling and control pulses, we performed experiments using a phantom made of 3% agarose and 97% water to assess the signal decrease caused by the saturation of macromolecular spins. Five healthy volunteers were studied (n = 5, 2 male, 3 female, age = 23-32 years, average age = 27.2). For activation studies, they were instructed to perform repeated, rapid apposition of the thumb and fingers of the right hand. The multislice volume was presaturated immediately before each tagging pulse. The slab thickness of the labeling and control pulses was 90 mm; the slabs were positioned 65 mm below the center of the lowest slice in the multislice stack. The section thickness of each slice = 8-10 mm, field of view = 360-380 mm, matrix size = 96 X 128, 6/8 rectangular FOV. The inflow time was 900 ms, with up to a 1-s delay after each readout

EPISTAR MRI 801 4 1.o 0.8 3 0.6.- A 3 ' I ' 4 0.4 0.2 - nn V.V -12-8 -4 0 4 8 12-12 -8-4 0 4 8 12 Time (msec) Time (msec) FIG. 1. Hyperbolic secant pulse (b = 600, M = 4.0) used for EPISTAR experiments. Tissue volume of interest I\- a Labeling pulse for acq. 1 (360 deg. - TI) Control pulses for acq. 2 (180 deg. - spoil - 180 deg. - TI) b FIG. 2. New EPISTAR method. The labeling pulse (360" flip angle, shaded box) and two control pulses (each with a 180" flip angle, open box) are applied at the same location. seen to produce minimal signal alteration in the tag region as compared with the inversion tag (Fig. 3). Signal was only present along the edges of the tag, where the nutation was less than 180" due to slice profile imperfections. To make sure these imperfections do not interfere with the tissue volume of interest, we have measured the slice profiles of both the 180' and 360' pulses using the same phantom. As demonstrated in Fig. 4, although slice profile imperfections extended from 30 to 60 mm off the center of the inversion slab, these pulses do not affect the spins in the tissue volume of interest. This is because we always place the labeling and control pulses 65 mm below the lowest slice of the multislice stack, thus eliminating their influence to the imaging slices. In addition, there was no observable effect of the slice profile imperfections on CBF maps either. Figure 5 confirmed our hypothesis that the MT effects are identical iising 360 and double 180' pulses. Using phantoms made of 100% water and 97% water/3% agarose, we have evaluated MR signals acquired with 0' (no pulse), 360 and double 180 tagging pulses applied 65 mm below the center of the imag- before the next tagging pulse was applied. The actual inflow time varied for each slice, since a total of 95.5 ms was required to acquire each successive image, For instance, in a five-slice study acquired in a top-clown order, the inflow time varies from 900 ms for the first (top] slice to 1282 ms for the fifth (bottom) slice. RESULTS Using a Siemens resolution phantom made of 1.25 g (NiSO, X 6 H,O) and 1000 g Hzo (measured = 301 ms and Tz = 281 5 2 ms), the control double inversion was FIG. 3. (a) Resolution phantom imaged with new EPISTAR method using sagittally oriented tags for demonstration purposes. Arrow = 360" inversion tag, arrowhead = double 180" control tag. (b) Reference image of the same phantom showing locations of the tagging pulses.

a02 a 2000 $ 0 b -loo0-180deg -4-36ODeg -2000-80 -60-40 -20 0 20 40 M) 80 X (mm) FIG. 4. Measured slice profiles of the 180" and 360" adiabatic inversion pulses. (a) A spherical phantom imaged with the 90-mm pulses applied in the sagittal orientation. (b) Plot of the slice profiles obtained from the experimental data. The patterned bar displays the location of the lowest slice of the multislice stack. Edelman, Chen below the slice, as evidenced in the reduced amount of signal in the superior sagittal sinus. Figure 7 shows an example of a six-slice EPISTAR acquisition. The data were collected with 48 excitations over a period of 1 min 45 s. Cortical enhancement is shown on all slices except the top one (upper left), where most enhancement is still in the arteries. No evidence of subtraction artifact or residual magnetization transfer effects are identified. The absence of venous tagging makes it easier to visualize the posterior aspect of the occipital lobes than would be possible with the old EPISTAR method ing slice. Measurements with no tagging pulse were performed twice to assess the signal variations due to system instability. As shown on Fig. 5, no signal difference was detected in the 100% water phantom using any tagging pulses (Fig. 5c), whereas an averaged 4.6% of signal decrease due to MT effects was observed in the 100% water/3% agarose phantom using either 360' or double 180' tagging pulses [Fig. 5d). The MT effects caused by the 360" pulse is identical to that caused by the double 180' pulses within the system errors of the measurements. In human studies, there was perfect cancellation of MT effects as demonstrated in Fig. 6. For single slice imaging, there was no observable difference in the CBF maps produced by the standard and new EPISTAR methods (Figs. 6a and 6c). Without a control pulse, much of the brain signal intensity in the EPISTAR images was contributed by MT effects (Fig. 6b). Use of two 90" control RF pulses instead of two 180' RF pulses did not eliminate MT effects, as expected (Fig. 6d). One important difference between the old and new EPISTAR methods is the decreased amount of venous tagging when the control pulses are applied.r 0 2600 2400 g 2200 + ci c.@ 2000 m 1800 1600 (c) 100% Water, I m,i,,, 1600 1400 b.- 5 - L 4 1200 B iz 1000 SO! (d) (d) 3% 3% Agarose A! - 180-180 J I. I, I, I. I. I8 38 43 48 53 58 63 Position FIG. 5. Evaluation of MT effects from the 360" and double 180" pulses using phantoms made of 100% water and 97% water/3% agarose. (a) Coronal image showing locations of the slice of interest (solid bar) and the tagging pulse (shaded box); (b) acquired EPISTAR image for the two phantoms. The solid line shows the location where profiles of the phantoms are calculated; (c) slice profiles of the 100% water phantom using different tagging pulses. (d) Slice profiles of the 97% water/3% agarose phantom showing MT effects. Notice the MT effect of the 360" pulse is almost identical to that of the double 180" pulses.

EPISTAR MRl 803 FIG. 6. Perfusion images obtained using (a) conventional single slice EPISTAR method, (b) conventional single slice EPISTAR method without control pulse, (c) new EPISTAR method, and (d) new EPISTAR method with two 90" control pulses. The double inversion (c) eliminates MT effects, whereas the double saturation (d) does not. Figure 8 shows an example of a five-slice acquisition using a motor activation paradigm. Although only simple subtraction of the (activatedbaseline) EPISTAR images was performed, focal CUF changes due to motor cortex activation are seen in all the slices. MT effects since the frequency shifts of the inverting band and control bands are not identical. The technique demonstrated here does not have a high RF duty cycle. Additionally, the cancellation of MT effects is exact, since the labeling and control pulses are applied with the same frequency shift, bandwidth, and total RF irradiation. This exact cancellation of MT effects has been demonstrated using a phantom made of 97% water/3?& agarose to show that the signal decrease caused by the 360" pulse is identical to that caused by the double 180 pulses (Fig. 5d). Compared with the original EPISTAR method, other advantages of the new method are the lack of venous tagging by the control pulses, and the elimination of any concern about asymmetric MT effects. Asymmetric MT effects might he a potential concern when the control pulse is placed above the slice, since the frequency shift of this RF pulse is opposite in sign to that of the labeling pulse. Other multislice imaging methods such as the use of a two-coil system have been proposed previously and used very successfully for assessment of cerebral perfusion (8, 9). The two-coil system consists of one small surface coil for labeling the arterial water spins, and a head coil for MRI that is actively decoupled from the labeling coil. The advantage of using such a method is the complete elimination of MT effects, thus permitting multislice perfiision imaging. However, the need for an extra coil limits the widespread application of such a method at clinical settings. Frank, Wong, and Buxton have proposed another approach for multislice perfusion imaging (10, 11). Their studies suggested that the main contribution of the off- DISCUSSION Several techniques have been proposed for measurement of CBF, hut only recently did Alsop show the capability for multislice imaging using the continuous adiabatic tagging method (7). For their control tag, the RF pulse was sinusoidally oscillated at 250 Hz. This produces two narrow bands of inversion located 250 Hz above and below the unmodulated RF frequency. However, continuous tagging techniques are not feasible on all MRI systems. due to the heavy duty cycle requirements for the RF amplifier. Also, this method does not exactly Compensate FIG. 7. Perfusion images obtained with new EPISTAR method at six different slice locations.

a04 Edelman, Chen FIG. 8. Motor activation study. Multislice EPISTAR images were acquired at five locations without and then with tapping of the right fingers. Top row: Resting EPISTAR images show distribution of cerebral blood flow without a signal contribution from MT effects. Slice thickness = 8 mm, 40 acquisitions, time of acquisition 1 min 45 s. Bottom row: T,-weighted spin-echo images obtained at same levels as the EPISTAR images, showing activated regions in the left motor cortex in all five slices. resonance effects is from thc imperfections in the slice profile of the inversion pulses, and the contribution of MT effects from these pulses are negligible. Thus multislice perfusion imaging in their experiments was performed simply by improving the slice profile of the adiabatic inversions and neglecting MT contribution. Compared to our approach, this method could be limited when MT effects cannot be neglected. It needs to be pointed out, however, that the performance of the double 180' pulses as a perfect control is slightly compromised by slice profile imperfections and motion of arterial water spins. Spins along the edges of the tag will not be reinverted by the second 180' pulse due to the less than 180' nutation in those regions. Some fast-flowing arterial water spins will move out of the tagging slab before experiencing the second 180' pulse. While these factors do not interfere with static spins within the tissue volume of interest, they need to be considered when quantitative perfusion assessment is desired, although the effects are expected to be minor. Two remaining problems with multislice imaging using either continuous or pulsed arterial labeling methods are (1): that the various slices have different inflow times, and (2) the transit time for the labeled arterial spins from the tag region to the capillary bed will vary for each brain tissue location. The problem of varying inflow times can be ameliorated, but not eliminated, by minimizing the data acquisition period for each slice (e.g., using a reduced number of phase encodes or faster gradients). With the new method for compensating MT effects, it is possible to perform a 3D EPISTAR acquisition. A 3D acqui- sition ensures that the inflow times are identical throughout the imaging volume (although arterial transit times will still vary). Three-dimensional imaging was not possible with the original EPISTAR method since the control pulse had a different MT effect than the labeling pulse except at the center of the 3D volume. Finally, thc new method could be modified to permit quantitation of cerebral blood flow, using approaches such as QUIPPS (11, 12). However, it does not appear applicable to the FAIR technique. In conclusion, we have described an improved technique for mapping of CBF that permits multislice imaging. The method should be easily implemented on any MRI system capable of EPI and should greatly increase the practical utility of EPISTAR and related quantitative methods for the study of brain function. REFERENCES 1. D. S. Williams, J. A. Delre, J. S. Leigh, A. P. Koretsky, Magnetic resonance imaging of perfusion using spin inversion of arterial water. Proc. Natl. Acad. Sci. USA 89, 212-216 (1992). 2. R. R. Edelman, B. Siewert, D. G. Darby, V. Thangaraj, A. C. Nobre, M. M. Mesulam, S. Warach, Qualitative mapping of cerebral blood flow and functional localization with echo-planar MR imaging and signal targeting with alternating frequency. Radiology 192, 513-20 (1994). 3. K. K. Kwong, D. A. Cheder, R. M. Weisskoff, B. R. Rosen, Perfusion MR imaging, in Proc., 2nd Annual Meeting ISMRM, 1994,'' p. 100s. 4. S-C. Kim, Quantification of regional cerebral blood flow by flowsensitive alternating inversion recovery (FAIR) technique: application to functional mapping. Magn. Reson. Med. 34, 293-301 (19951. 5. T. A. Sandson, M. O'Connor, R. A. Sperling, R. R. EdslInan, S.

EPISTAR MRI 805 Warach, Noninvasive perfusion MRI in Alzheimer s disease: a preliminary report. Neurology 47, 1339-42 (1996). 6. S. D. Wolff, R. S. Balaban, Magnetization transfer contrast (MTC] and tissue water proton relaxation in vivo. Map. Reson. Med. 10, 135-144 (1988). 7. D. C. Alsop, L. Maccotta, J. A. Detre, Multi-slice perhision imaging using adiabatic arterial spin labeling and an amplitude modulated control, in Proc., 5th Annual Meeting, ISMRM, Vancouver, 1997, p. 81. 8. A. C. Silva, W. Zhang, D. S. Williams, A. P. Koretsky, Multi-slice MRI of rat brain perfusion during amphetamine stimulation using arterial spin labeling. Magn. Reson. Med. 33, 209-14 (1995). 9. G. Zaharchuk, P. J. Ledden, K. K. Kwong, B. R. Rosen, Multislice arterial spin labeling measurements of perfusion territory and functional activation in humans using two coils, in Proc., 5th Annual Mccting, ISMRM, Vancouver, 1998, p. 79. 10. L. R. Frank, E. C. Wong, R. B. Buxton, Slice profile effects in adiabatic inversion: application multislice perfusion imaging. Magn. Reson. Med. 38, 558-64 (1997). 11. E. C. Wong, R. B. Buxton, L. R. Frank, Implementation of quantitative perfusion imaging techniques for functional brain mapping using pulsed arterial spin labeling. NMR Biomsd. 10, 237-49 (1997). 12. E. Wong, L. R. Frank, R. B. Buxton, Quantitative multislice perfusion imaging using QUIPSS 11, EPISTAR, FAIR and PICORE, in Proc., 5th Annual Meeting, ISMRM, Vancouver, 1997, p. 85. Postdoctoral Positions Postdoctoral positions in magnetic resonance imaging and microscopy are available in the Laboratory for Advanced Structural NMR in the Department of Radiology at the University of Pennsylvania Medical Center, for a period of two years, with possible extension to a third year. Research in the host laboratory focuses on the development and application of quantitative MRI methods and NMR microimaging, including image processing and mathematical modeling. Anticipated activities include design and implementation of pulse sequences and image processing algorithms for MR osteodensitometry, in vivo MR microimaging with applications toward analysis of cancellous bone architecture and function, in laboratory animals and patients. Other fields of research include the study of tissue oxygenation by MR oxymetry. The Department of Radiology s Center for Advanced MR Imaging and Spectroscopy (CAMRIS) accomodates 1.5 and 4. OT GE Signa whole-body scanners and a 4.7T small-bore animal imager, and a Bruker 400MHz microimaging system in the Children s Hospital on the University of Pennsylvania Campus is available as well. Preference will be given to candidates with hands-on experience in NMR imaging research and pulse sequence programming. Salary will be commensurate with experience. Interested candidates are encouraged to submit their C.V. and two letters of recommendation to: Felix W. Wehrli, Ph.D., Professor of Radiologic Science and Biophysics, University of Pennsylvania Medical Center, 3400 Spruce Street, Philadelphia, PA 19004, USA, Fax 215-349-5925, email wehrli@oasis.rad.upenn.edu.