A Complete Digital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla)

Similar documents
(N)MR Imaging. Lab Course Script. FMP PhD Autumn School. Location: C81, MRI Lab B0.03 (basement) Instructor: Leif Schröder. Date: November 3rd, 2010

HETERONUCLEAR IMAGING. Topics to be Discussed:

Background (~EE369B)

ISMRM weekend educational course, MR Systems Engineering, Console Electronics

MAGNETIC RESONANCE IMAGING

Twelve voice signals, each band-limited to 3 khz, are frequency -multiplexed using 1 khz guard bands between channels and between the main carrier

MRI SYSTEM COMPONENTS Module One

PHY3902 PHY3904. Nuclear magnetic resonance Laboratory Protocol

USE OF MATLAB IN SIGNAL PROCESSING LABORATORY EXPERIMENTS

On the Design of Software and Hardware for a WSN Transmitter

MRI & NMR spectrometer

Multi Frequency RFID Read Writer System

PROPAGATION CHANNEL EMULATOR : ECP

Field Simulation Software to Improve Magnetic Resonance Imaging

PULSE CODE MODULATION TELEMETRY Properties of Various Binary Modulation Types

M R I Physics Course. Jerry Allison Ph.D., Chris Wright B.S., Tom Lavin B.S., Nathan Yanasak Ph.D. Department of Radiology Medical College of Georgia

Coil Overlook Coil in MRI system TEM Coil Coil Overlook

2015 Spin echoes and projection imaging

Wirelessly Powered Sensor Transponder for UHF RFID

Functional Description / User Manual

Analog & Digital Communication

Encoding of inductively measured k-space trajectories in MR raw data

Chapter 1. 1 The NMR Spectrometer. 1.1 Components of an NMR Spectrometer The Magnet

AC LAB ECE-D ecestudy.wordpress.com

Pulse-Width Modulation (PWM)

- 1 - Rap. UIT-R BS Rep. ITU-R BS.2004 DIGITAL BROADCASTING SYSTEMS INTENDED FOR AM BANDS

1 Introduction. 2 The basic principles of NMR

Appendix B. Design Implementation Description For The Digital Frequency Demodulator

Using Frequency Diversity to Improve Measurement Speed Roger Dygert MI Technologies, 1125 Satellite Blvd., Suite 100 Suwanee, GA 30024

STELAR s.r.l. via E. Fermi, Mede (PV) - Italy tel Uni-Pavia 24/11/06

EE470 Electronic Communication Theory Exam II

AD9772A - Functional Block Diagram

A 3 TO 30 MHZ HIGH-RESOLUTION SYNTHESIZER CONSISTING OF A DDS, DIVIDE-AND-MIX MODULES, AND A M/N SYNTHESIZER. Richard K. Karlquist

Code No: R Set No. 1

RF Integrated Circuits

10. Phase Cycling and Pulsed Field Gradients Introduction to Phase Cycling - Quadrature images

ANALOG COMMUNICATION

Principios Básicos de RMN en sólidos destinado a usuarios. Gustavo Monti. Fa.M.A.F. Universidad Nacional de Córdoba Argentina

Reconfigurable RF Systems Using Commercially Available Digital Capacitor Arrays

EXPERIMENTAL SETUP AIMED TO STUDY THE ELECTRICAL IMPEDANCE VARIATIONS OF A PLASMA COLUMN IN A WIDE FREQUENCY RANGE

Pulsed NMR Experiment Guide Kenneth Jackson Physics 173, Spring 2014 Professor Tsai

Signals and Systems Lecture 9 Communication Systems Frequency-Division Multiplexing and Frequency Modulation (FM)

RADIO RECEIVERS ECE 3103 WIRELESS COMMUNICATION SYSTEMS

Design Implementation Description for the Digital Frequency Oscillator

Digital Audio Broadcasting Eureka-147. Minimum Requirements for Terrestrial DAB Transmitters

Development and Application of 500MSPS Digitizer for High Resolution Ultrasonic Measurements

A 1.9GHz Single-Chip CMOS PHS Cellphone

Technician License Course Chapter 3 Types of Radios and Radio Circuits. Module 7

1 Introduction to Highly Integrated and Tunable RF Receiver Front Ends

ULS24 Frequently Asked Questions

Pre-amplifiers for a 15-Tesla magnetic resonance imager

HF Receivers, Part 2

AN X-BAND FREQUENCY AGILE SOURCE WITH EXTREMELY LOW PHASE NOISE FOR DOPPLER RADAR

A PC-BASED TIME INTERVAL COUNTER WITH 200 PS RESOLUTION

Design of Adaptive RFID Reader based on DDS and RC522 Li Yang, Dong Zhi-Hong, Cong Dong-Sheng

Fully integrated UHF RFID mobile reader with power amplifiers using System-in-Package (SiP)

LUXONDES. See the electromagnetic waves. Product 2018 / 19

Figure 4.1 Vector representation of magnetic field.

Preliminary features of the SDR-X receiver SDR-X , PowerSDR Winrad Winrad DDS SFDR SFDR AD995 AD99 1

ADVANCED EMBEDDED MONITORING SYSTEM FOR ELECTROMAGNETIC RADIATION

A Conceptual Tour of Pulsed NMR*

THE INSTRUMENT. I. Introduction

Analysis of Phase Noise Profile of a 1.1 GHz Phase-locked Loop

Design of Phase Locked Loop as a Frequency Synthesizer Muttappa 1 Akalpita L Kulkarni 2

A NEW GENERATION PROGRAMMABLE PHASE/AMPLITUDE MEASUREMENT RECEIVER

Development of utca Hardware for BAM system at FLASH and XFEL

Gradient Spoiling. Average balanced SSFP magnetization Reduce sensitivity to off-resonance. FFE, FISP, GRASS, GRE, FAST, Field Echo

Eavesdropping Near Field Contactless Payments: A Quantitative Analysis

Lab 8 6.S02 Spring 2013 MRI Projection Imaging

Analog signal generator that meets virtually every requirement

Week 8 AM Modulation and the AM Receiver

PULSED/CW NUCLEAR MAGNETIC RESONANCE

CONSTRUCTION OF A DUAL TUNED COIL

CH85CH2202-0/85/ $1.00

THIS work focus on a sector of the hardware to be used

Analog Filter and. Circuit Design Handbook. Arthur B. Williams. Singapore Sydney Toronto. Mc Graw Hill Education

FLASH rf gun. beam generated within the (1.3 GHz) RF gun by a laser. filling time: typical 55 μs. flat top time: up to 800 μs

Keysight Technologies Using a Network and Impedance Analyzer to Evaluate MHz RFID Tags and Readers/Writers

Wavedancer A new ultra low power ISM band transceiver RFIC

Digital microcontroller for sonar waveform generator. Aleksander SCHMIDT, Jan SCHMIDT

CHAPTER - 6 PIN DIODE CONTROL CIRCUITS FOR WIRELESS COMMUNICATIONS SYSTEMS

two computers. 2- Providing a channel between them for transmitting and receiving the signals through it.

Piezoelectric Discriminators

RFID Systems: Radio Architecture

EMBEDDED DOPPLER ULTRASOUND SIGNAL PROCESSING USING FIELD PROGRAMMABLE GATE ARRAYS

Investigation of a Voltage Probe in Microstrip Technology

German Jordanian University Department of Communication Engineering Digital Communication Systems Lab. CME 313-Lab

APPLICATION NOTE 695 New ICs Revolutionize The Sensor Interface

ETSI Standards and the Measurement of RF Conducted Output Power of Wi-Fi ac Signals

Keywords: GPS, receiver, GPS receiver, MAX2769, 2769, 1575MHz, Integrated GPS Receiver, Global Positioning System

MRI RF-Coils. Innovation with Integrity. Highest sensitivity for your preclinical MRI and MRS applications. Preclinical Imaging

DIGITAL BEAM-FORMING ANTENNA RANGE

Lab 2 Radio-frequency Coils and Construction

The New England Radio Discussion Society electronics course (Phase 4, cont d) Introduction to receivers

Lecture Topics. Doppler CW Radar System, FM-CW Radar System, Moving Target Indication Radar System, and Pulsed Doppler Radar System

Non-linear Digital Audio Processor for dedicated loudspeaker systems

A DSP IMPLEMENTED DIGITAL FM MULTIPLEXING SYSTEM

Costas Loop. Modules: Sequence Generator, Digital Utilities, VCO, Quadrature Utilities (2), Phase Shifter, Tuneable LPF (2), Multiplier

ericssonz LBI-38640E MAINTENANCE MANUAL FOR VHF TRANSMITTER SYNTHESIZER MODULE 19D902780G1 DESCRIPTION

Level 6 Graduate Diploma in Engineering Electronics and telecommunications

Transcription:

IEEE Instrumentation and Measurement Technology Conference Anchorage, AK, USA, 21-23 May 2002 A Complete igital Magnetic Resonance Imaging (MRI) System at Low Magnetic Field (0.1 Tesla) Kosai RAOOF*, IEEE member, Aktham ASFOUR**, and Jean Marc FOURNIER** *Laboratoire des Images et des Signaux (LIS), **Laboratoire d Eectrotechnique de Grenoble (LEG) ENSIEG, BP 46, 38402 Saint Martin d Hères, France email: kosai.raoof@lis.inpg.fr Abstract- A new complete digital MRI system is developed to enhance the signal-to-noise ratio of the detected nuclear magnetic resonance(nmr) signals at low magnetic field (0.1 Tesla). The system is based on digital signal processor (SP) that functions also as a controller and as an interface between the frequency translator and the digital synthesiser. The final system should be portable functioning at low voltage. It is a low cost system, and it can be dedicated to be used in a medical cabinets for examining small members of the human body. Such a system is also useful for quality control of agricultural and food products. The first results obtained with our system are presented in this paper. This includes NMR signal and images of 64x64 pixels. Keywords- Low field MRI, NMR signal detection, igital NMR system, NMR signal to noise ratio enhancement I. INTROUCTION As it is well known, in MRI technique, a substance is subject to a uniform static magnetic field (polarizing field). A net magnetic moment, of the substance, is produced in the direction of the polarizing field. A short radio frequency pulse (RF excitation pulse) is applied at the characteristic Larmor frequency of the processing nuclei which, in turn, emit an RF signal (NMR signal) of the same frequency [1]. This signal is detected and processed to be used for obtaining a fingerprint of the environment of the nucleus being studied. This information is one-dimensional but is converted into-two-dimensional image by adding gradients to the static magnetic field which results in a frequency modulation of the emitted NMR signal. A series of such measurements is analyzed by computer to generate the image [2], [3]. A typical MRI facility consists a magnet which produce the static field, an RF transmitter providing the RF excitation pulse and an RF receiver allowing detection and processing of the NMR signal. The MRI technique requires very expensive workstations because of the high static magnetic fields actually used. In our laboratories, we are working on the development of a dedicated low field (so low cost) MRI imager, at 0.1 tesla. This imager can also be used in medical cabinet for examining small parts of the human body (such as the arm, the wrist, etc. ). The same system is useful for in-line quality control in agricultural and food applications 1 [8]. The major problem associated with the low magnetic field is that the NMR signal-to-noise ratio is very low and enhancement on signal processing are necessary. For doing so, we developed a whole digital MRI system. Recently there are more and more digital parts in the MRI systems. However, the most important parts of the systems like frequency translators, filters, and synthesisers are still analog. In the few last years, the high frequency digital systems was advancing very fast by proposing some new chips especially designed for new generations of digital mobile communications. The demands are so huge so that the enhancement to price ratio are so important especially when the design is needed for base stations. We show, in this paper, that these advances in digital technologies are of great value for improving MRI systems. So, a complete home-built MRI system is presented here. This system allowed us to obtain real signals to validate a nonlinear identification approaches of NMR systems. The interested reader may refer to references [5], [6], and [7] for more details regarding these approaches. II. THE HOME-BUILT MRI SYSTEM Fig. 1 shows our implementation. Apart of the (antenna s RF amplifiers) and the duplexer, the whole system is digital, including the synthesiser and the frequency translator. The direct digital synthesiser (S) produces an analog carrier signal which can be controlled in both frequency and phase. This signal is amplified by the RF amplifier and sent, into the sample under study, via an RF antenna. 1 This project is partially supported by European contract No. QTEPACK QLK1-2000-00936 0-7803-7218-2/02/$10.00 '2002 IEEE

Antenna UPLEXER igital RF Receiver SP Sharc Gradient ACs Gradient amplifiers RF Amplifier S Interface PC PCI BUS Fig.1: Block diagram of the MRI digital system This same antenna is used to receive the NMR signal (we call it FI signal for Free induction ecay). This analog signal is passed to the RF digital receiver for quadratic demodulation. In the next sections we will show the RF systems for transmitter and the receiver. Some first results of NMR signals and images are also shown. II.1. The digital radio frequency transmitter: The core of the digital transmitter is the direct digital synthesiser (S), where the frequency is defined by a parallel interface programmed directly by the SP. Our synthesiser design does not consist of any phase locked-loops that may normally need some analog parts. In the market there are many synthesisers based on the phase locked loops. These synthesisers have problems especially for rapid changing in frequency and suffer from the phase noise. Many synthesisers are tested and finally the A9852 from Analog evices is chosen due to its possibility of high signal-tonoise ratio and the good SFR in the final stage. The programmable gain of 12 bits in the output was very interesting to calibrate the RF signal digitally by the SP. The existence of the on chip comparator and an additional AC was important to synchronise the phase with the digital receiving part. The internal clock was running at 200MHz although good signal to noise ratios was also obtained for 100MHz. Only 32 bits of the frequency tuning word was used due to the receiver internal synthesizer which is limited to 32 bits. A differential RF transformer and amplifier are used to buffer the signal output. The synthesised output ( about 4MHz ) is amplified again by an RF power amplifier designed in the laboratory. ue to the 12-bit AC output an excellent dynamic performance was obtained better than 75dB at 4MHz. The basic clock was derived directly from the SP clock which is running at 40MHz. In order to synchronize the phase of the RF output signal, the synthesized signal is compared to the additional internal AC output that derives a firing start signal. This part was the most delicate part in the design due to the PCB optimisation and screening needed for a high signal-to-noise ratio. Finally the possibility of programming the phase by the 14 bits of the synthesiser was very important to synchronize the receiver signal in real time, by programming a SP loop frequency control system. The final RF power amplifier linearity was assured between 4 and 5MHz. II.2. igital signal acquisition and processing: The NMR signal is detected by a well tuned RF antenna, the signal is matched to the RF preamplifier by a duplexer shown on Fig. 2. This amplifier is directly programmable by the SP (10-30dB). The amplified signal is then sampled by a 12 bit analog-to-digital converter. The bandwidth of the RF amplifier and AC was limited to 10MHz by using a π section low pass filter (LPF). In order to demodulate the NMR signal a complete digital tuner was used. This tuner can provide one channel digital receiver that translates the NMR signal to the base band using a 32 bit synthesiser and multiplier followed by decimation filters. The decimation factor as well as filter s coefficients are completely programmable by the parallel bus of the SP. The receiver was designed as well as the transmitter part as a daughter board plugged on the main SP Sharc PC board, interfaced with the PC via the PCI bus. The SP receives the demodulated base band signals (Real & imaginary parts) through its synchronous serial port clocked at 25Mbit/sec.

ecimation programmable LPF Antenna & uplexer RF Amp 12 Bit A C igital Synthesizer π/2 SP Serial Port BUS ecimation programmable LPF Fig. 2: The RF digital receiver. III. SYSTEM TESTS In order to quantify our results concerning the signal to noise ratio, as well as noise rejection and sensitivity, the digital receiver was tested by applying a known signal level at the RF input. The receiver s digital synthesizer is tuned to 5 MHz, the sampling frequency was 50MHz, and the total decimation factor was 2000. This decimation factor yields base band samples each 40 µs. The decimation factor was calculated from two digital decimation filters. The first filter is a low pass filter which is used to compensate for the droop in the useful bandwidth and assures a flatness better than 0.01dB in the final bandwidth. The second decimation filter is designed to limit the bandwidth to 5kHz, as shown in Fig. 3. This bandwidth defines the usable NMR signal bandwidth during image acquisition which corresponds to about 6x10-3 Tesla/meter of gradient field and 5,8kHz at 3dB. Also, the RF input amplifier is programmed to 30dB gain. The test procedure was as follows : First, a pure sine wave of 5MHz is applied with a variable amplitude from 50dB to 130dB. The observed sensitivity for the digital receiver was 120dB. And the best dynamic range for the digital full scale output was at 96dB. Secondly, a swiped sine wave was applied to valid the cut-off frequency of the whole bandwidth, especially the final decimation filter s frequency response as shown in Fig.3. Finally, the noise rejection was tested by introducing a wide band noise of 4MHz bandwidth on each side band (+/- 50KHz) with SNR=0dB. The results show a total rejection of the noise which was beyond our measuring instrument. Fig.3: The impulse and frequency response of the decimation filter.

IV. REAL NMR SIGNALS AN IMAGES The results are obtained in a 0.1Tesla magnetic field created by an Electro-magnet 2. It has some 10-5 of homogeneity in a volume of about 100 cm 3. The antenna is made of 6 spires coil coupled with a capacitor tuned antenna. The same antenna was used for transmitting and receiving the NMR signals using an appropriate matching circuit. The sample under investigation was a cylindrical tube filled with pure water. The sampling frequency of the NMR was fixed to 50MHz. This frequency was optimum for the digital filters used in the receiving section. The decimation factor was 2000 so that the final signal output sampling rate is 25kHz. Fig. 4 shows the FI signal (real and imaginary parts) obtained as a response to π/2 flip angle pulse of 2ms.. Fig.5: 64x64 pixels images obtained with the gradient-echo imaging sequence (echo time TE=6ms, repetition time TR=1.2s, slice thickness is of 1cm). The cylindrical tube of filled with water (on the top) and the same tube with inserted plastic object in the center. Fig.4: The real and the imaginary parts of the NMR FI signal. In order to define a precise NMR imaging sequence, a Windows98 based sequencer is developed. The sequencer generates a file from the defined graphical interface with which the user can easily define all possible sequence timing parameters including sequence type, image resolution, slice thickness, and the field of view. This generated file will be interpreted by the SP-SHARC to generate the different signals (XYZ gradient, Control signals, RF excitation pulse envelope,.. etc.). Images of 64x64 pixels are obtained using a gradient-echo imaging sequence. Sequences are programmed by SP assembly language. The obtained resolution for the average of 8 successive images is shown in Fig.5. This resolution is about 1.6mm/pixel V. CONCLUSION We demonstrated, in this paper, that the implementation of a digital MRI system is possible by using advances in digital technologies. This system, which is low cost and friendly use, allowed us to obtain high NMR signal to noise ratio for both low field spectroscopy and high resolution imaging applications. As a future work, ultra fast sequences will be developed for real time NMR imaging which is more and more necessary for many industrial applications. ACKNOWLEGEMENTS The authors would like to thank professor A. Constantinesco and r. P. Choquet for their useful discussions and for lending us the electro-magnet which permits verifying our system design. REFERENCES [1] R. R. ERNST, G. BOENHAUSEN, and A. WOKAUN, Principles of nuclear magnetic resonance in one and two dimensions, Clarendon Press, Oxford, 1989. 2 Made by the company RUCH (France).

[2]. P. LIANG and P. C. LAUTHERBURG, Principles of magnetic resonance imaging, a signal processing perspective, IEEE PRESS, INC., New York, USA. [3] P. MANSFIEL and A. P. G. MORRIS, NMR imaging in biomedicine, Academic Press, New York, 1989. [4] P. N. MORGAN and al. Real-time cardiac MRI using SP s, IEEE Transaction on Medical Imaging, Vol. 18, NO. 7, July 1999. [5] A. ASFOUR, K. RAOOF, and J. M FOURNIER, Nonlinear identification of NMR spin systems by adaptive filtering, Journal of Magnetic Resonance, Vol. 145, July 2000, p. 37-51. [6] A. ASFOUR, K. RAOOF, and J. M FOURNIER, Identification of NMR spin systems by Volterra series, IEEE World Congress of Medical Physics and Biomedical Engineering, Chicago, USA, July 2000, NO. 4045-38631. [7] A. ASFOUR, K. RAOOF, and J. M FOURNIER, Filtrage adaptatif RII non linéaire: application à l identification, des systems à résonance magnétique nucléaire (RMN), 18 e Symposium GRETSI 01 on Signal and Image Processing, Toulouse, France, September 2001, NO. 75. [8] European Project- QTEPACK, Advanced Eelectromagnetic Solution for Quality testing of Packaging Processes for Horti-fruit products, No. QLK1-2000-00936.