JEOL JEM-1400 Transmission Electron Microscope Operating Instructions

Similar documents
2. Raise HT to 200kVby following the procedure explained in 1.6.

Basic Users Manual for Tecnai-F20 TEM

Instructions for Tecnai a brief start up manual

Operating the Hitachi 7100 Transmission Electron Microscope Electron Microscopy Core, University of Utah

1.3. Before loading the holder into the TEM, make sure the X tilt is set to zero and the goniometer locked in place (this will make loading easier).

User Operation of JEOL 1200 EX II

1. Specimen Holder Removal, Loading, and Insertion

COMPACT MANUAL FOR GI USERS OF THE JEM 1400 FLASH BEGINNERS (For internal use only) Gray means additional information at the end of this mini-manual

Please follow these instructions for use of the Philips CM100 TEM. Adopted from website below.

Protective Equipment Nitrile gloves for handling sample holder and safety glasses for filling liquid nitrogen dewar.

Full-screen mode Popup controls. Overview of the microscope user interface, TEM User Interface and TIA on the left and EDS on the right

Jeol JEM Responsible personell: Endy ( ) Online booking is compulsory!

FEI Tecnai G 2 F20 Operating Procedures

Using the Hitachi 3400-N VP-SEM

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

2 How to operate the microscope/obtain an image

Standard Operating Procedure for the Amray 1810 Scanning Electron Microscope Version: 29 NOVEMBER 2014

1.2. Make sure the viewing screen is covered (exposure to liquid N 2 may cause it to crack).

Tecnai T12 Operating Procedures

STANDARD OPERATING PROCEDURE: JEOL TEM-2100

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the digital camera.

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

JEOL 6700 User Manual 05/18/2009

JSM 6060 LV SCANNING ELECTRON MICROSCOPE STANDARD OPERATING PROCEDURES

1.1. In regular TEM imaging mode, find a region of interest and set it at eucentric height.

SAMUEL ROBERTS NOBLE ELECTRON MICROSCOPY LABORATORY. Operating Procedures for the Zeiss 9 S-2. Transmission Electron Microscope

JEOL JEM 2010 TRAINING TRANSMISSION ELECTRON MICROSCOPE USER MANUAL

SEM Training Notebook

Dickinson College Department of Geology

SEM Training Notebook

Title: Amray 1830 SEM#2 Semiconductor & Microsystems Fabrication Laboratory Revision: D Rev Date: 03/18/2016

OPERATION OF THE HITACHI S-450 SCANNING ELECTRON MICROSCOPE. by Doug Bray Department of Biological Sciences University of Lethbridge

How to use the Jeol 1010 TEM of GI (Liesbeth own GI version)

05/20/14 1. Philips CM200T. Standby Condition

JEOL 6500 User Manual

Section 1: TEM parts and functions... 2

STEM alignment procedures

Basic Operating Instructions for Strata Dual Beam 235 FIB/SEM

SEM OPERATION IN LOW VACUUM MODE

INSTRUCTIONS JEM-2010F FIELD-EMISSION TRANSMISSION ELECTRON MICROSCOPE WITH STEM CAPABILITY

Operating Checklist for using the Scanning Electron. Microscope, JEOL JSM 6400.

Transmission Electron Microscopy 9. The Instrument. Outline

JEOL 2010 FasTEM & DigitalMicrograph User's Guide

FEI Titan Image Corrected STEM

General information. If you see the instrument turned off, notify MIC personnel. MIC personnel will help you insert your samples into the instrument.

This document assumes the user is already familiar with basic operation of the instrument in TEM mode and use of the Microscope Control interface.

Scanning Electron Microscope FEI INSPECT F50. Step by step operation manual

CAPTURING IMAGES ON THE HIGH-MAGNIFICATION MICROSCOPE

FE-SEM SU-8020 Operating manual (Preliminary version)

1. Preliminary sample preparation

Operating Checklist for using the Scanning Electron Microscope, JEOL JSM 6400.

1.1. Log on to the TUMI system (you cannot proceed further until this is done).

Check that the pneumatic hose is disconnected!!!! (unless your using the BSE detector, of course)

Last updated 6/12/18. F20 User Manual at the Simons Electron Microscopy Center

FEI Quanta 200 ESEM Basic instructions

RAITH e-line OPERATING INSTRUCTIONS

Model SU3500 Scanning Electron Microscope

Motorized Axio Observer Start-up instructions

FEI Falcon Direct Electron Detector. Best Practice Document

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

Leica DB LB Research microscope and Studo Lite Imaging software

Instructions for Making On-Line Reservations for Microscopes in NB11-204

Nikon Eclipse Ti A1-A Confocal Operating Manual. Start-up. Microscope

CM20 USER GUIDE. Duncan Alexander, CIME 2010

Things to check before start-up.

Olympus Digital Microscope Camera (DP70) checklist

University of Minnesota Nano Fabrication Center Standard Operating Procedure Equipment Name:

Introduction of New Products

ELECTRON MICROSCOPY PROCEDURES MANUAL

Introduction: Why electrons?

Operating Instructions for Zeiss LSM 510

Standard Operating Procedure Hitachi UHR CFE SU8230 SEM

MAKE SURE YOUR SLIDES ARE CLEAN (TOP & BOTTOM) BEFORE LOADING DO NOT LOAD SLIDES DURING SOFTWARE INITIALIZATION

Zeiss Axio Imager.A1 manual

Nikon Eclipse Ti2-E Widefield/Spinning Disk Confocal Microscope Standard Operation Protocol

Quick Guide for Zeiss 710 Laser Scanning Confocal MGH Cancer Center

Nikon SIM-E & A1-R System

Operating F20/F30 with SerialEM

Zeiss LSM 880 Protocol

Nikon E800 Operating Instructions.

Nikon Ti-E Microscope Manual. Rightmire Hall Ohio State University. Director: Tony Brown Rightmire

Leica DMi8A Quick Guide

LEICA TCS SP5 AOBS TANDEM USER MANUAL

Brightfield Microscopy and Image Acquisition on Spotcam1. by Ryan Taylor/Nancy Kleene Last modified 10/02/05 by Birgit Ehmer

CATHODOLUMINESCENCE IMAGING UNIVERSITY OF ST ANDREWS

Leica SP8 TCS Users Manual

Widefield-NikonEclipseTE200-ORCA Nikon Eclipse TE200 Inverted Microscope with Hamamatsu 1394 Orca-ER Cooled CCD Camera and Micromanager Software

The user should already be familiar with operation of the instrument in STEM mode, use of the Microscope Control interface, and TIA.

Chapter 2 Alignment C. Robert Bagnell, Jr., Ph.D., 2012

USING LEICA AS LASER MICRODISSECTION (LMD6000) MICROSCOPE Written By Jungim Hur

AxioVision 4.5 Brightfield Image Capture Procedure

Manual for BMS E1 eplan series, compound microscope

SPECTRALIS Training Guide

Easy Kohler Illumination Method

Fire CR Calibration Guide

KEYENCE VKX LASER-SCANNING CONFOCAL MICROSCOPE Standard Operating Procedures (updated Oct 2017)

Operation Guide for the Leica SP2 Confocal Microscope Bio-Imaging Facility Hunter College October 2009

RIGAKU VariMax Dual Part 0 Startup & Shutdown Manual

Transcription:

JEOL JEM-1400 Transmission Electron Microscope Operating Instructions Anti-contamination device Objective aperture Objective aperture translation knobs Specimen holder Pump/air switch Left hand control module Side mount camera control switch X deflector knob Viewing screen and cover Right hand control module Mirror for eyepiece viewing Y deflector knob 1. Login to microscope computer with your uniqname and Kerberos (level 1) password. 2. Turn on the two camera systems. Turn on power strip behind Column. If you are using the side mount camera, you must also turn that camera on. 3. Turn on second computer.

4. For negatively stained samples only fill anti-contamination device with liquid nitrogen. Remove white cap from anti-contamination device, insert funnel. Slowly fill device with 1/2 dewar flask of liquid nitrogen from tank in hallway and replace cap. Press white switch upwards to ensure yellow light does not brighten. 5. Remove specimen holder. Pull specimen holder out to first stop (large pull) Turn counterclockwise until a stop (large turn) Pull holder out to second stop (small pull) Turn counterclockwise until next stop (small turn) Set PUMP/AIR switch to AIR WAIT as chamber releases vacuum Pull holder out fully pull switch out before pushing down 6. Mount grid specimen side up (dark/shiny side) in specimen cartridge of specimen holder. Do not touch metal rod with your fingers! 7. Insert specimen holder. Set PUMP/AIR switch to PUMP. Insert specimen holder, aligning gold screw on rod with indentation, to first stop (yellow light will illuminate.) Box pops up asking which holder, single tilt holder press OK. After green light illuminates Turn holder clockwise until first stop (small turn) Control holder as vacuum pulls in (small insertion) Turn holder very slowly clockwise until next stop (large turn) Control holder as vacuum pulls in (large insertion). If the high tension turns off during this procedure, it will have to be turned back on after you seat the holder into place. 8. If you do not see a beam on the phosphor screen check sidemount camera position. Black flag should be down. shiny side up 9. Turn on filament Check that V1 and V2 squares in software Valve/Vacuum monitor are green. Beam current in software Beam controller should be ~half current HT. Push Grey/blue ON button under Filament tab. Filament will saturate automatically. Beam current should be 50-70uA.

10. Center your beam. Lower magnification to 2000x or below by use the magnification knob on the center of the right hand control module. Use the brightness knob on the center of the left hand control module to bring beam to crossover. Crossover is the smallest and brightest spot. As you go to crossover, you may need to adjust the centering X and Y knobs. Use x and y knobs are on the lower inner most left and right hand control modules and are used to center the beam. 11. Spread beam so that it is fairly dim on the phosphor screen. ALWAYS USE BEAM TO THE RIGHT OF CROSSOVER. 12. At this point you will need to bring up the camera. Open AMT602 software on your desktop. Bottom mount camera software will open. If you are using the side mount camera, see other instructions. 13. On camera software in the top right side click for live image. The phosphor screen will come up and an image will appear on the monitor. Adjust the brightness of the histogram using the left hand center brightness control and center the histogram. Caution: Too bright (saturation) can damage the camera. Creating a background correction: You will need to find an empty space by moving the grid to a place that is empty of any contrast or by partial removal of the sample holder. Pull out the sample holder to the end and turn counter clockwise a small amount so that it doesn t go back into the column. Center the histogram with the brightness control. In AMT software, click Corrections Tab. Select Acquire a Background Use brightness knob to center red histogram in chart in software. Press Proceed. When completed, image will say Background Image Correction Complete. Return sample to beam path by turning the holder clockwise and slowly allowing the sample to seat into place. 14. Find an object in your grid. Set magnification to 600x. Control the brightness so that it stays within the histogram. Look for something with contrast. Use the ball on the stage control on right side to change locations on your grid.

15. Eucentric Positioning Put the phosphor screen back down by pressing the button on the right hand controller-à Increase magnification to 15,000x. You will need to control the brightness and the x and y position of the beam as you go up in magnification. Bring the phosphor screen back up. Center the histogram. Press the Standard Focus button on the right hand control module. Press Z-Focus and Image Wobbler buttons on the right hand control module. Use the Focus knob (now set to move through the Z plane) to focus image. Deselect Z-focus and Image Wobble. 16. The camera is now ready to take micrographs and you may cover phosphor screen. 17. Press Click for Live Image button. Press red Speed Live button to change image to Quality Live. 18. Focus image using Image Wobbler and Focus knob. Deselect Image Wobbler when complete. 19. Be sure red histogram is centered by adjusting brightness knob. Image wobbler Focus 20. Take image by pressing Click for Final Image button. 21. Save image as Tif file. Create a folder with your uniqname in the desktop Users Shortcut folder. Label new folder with today s date. 22. To save a location on your grid: Click Memory button in nanospace map (TEM software.) Highlight and click Go to return to memory location.

To remove/switch sample grids 1. Press Abort to stop live image. 2. Turn off filament. Press grey off button in TEM software Beam Controller box. 3. Neutralize the stage Press Stage Neutral in TEM software Stage box. 4. Remove sample holder. Large pull Large turn Small pull Small turn WAIT Pump à air ~20 sec Remove sample holder 5. Replace sample holder Air à pump Insert to first stop WAIT for green light Small turn Small controlled pull in Large turn Large controlled pull in 6. Turn filament on. If there is initially no beam, check side mount camera. For each new sample 1. Perform eucentric positioning (see #15) 2. Focus image (see #18) SHUT DOWN 1. Remove camera from beam path. Press Screen Up button for bottom mount camera. Flip switch for side-mount camera. 2. Turn brightness knob clockwise until it beeps.

3. Turn off filament. 4. Neutralize stage. 5. Remove sample and replace sample holder. 6. Please use M-Box or an encrypted device to transfer your data files. 7. Close AMT software, logoff 2 nd computer. 8. LEAVE COMPUTER ON. If high tension has turned off while reseating the sample holder, you will need to turn it back on again. HT is always left on at 80kV. 9. Turn off monitor. 10. Turn off side-mount camera and power strips (in back.)