These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory.

Similar documents
Note Using the PXIe-5785 in a manner not described in this document might impair the protection the PXIe-5785 provides.

PCI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PCI Oscilloscope Device

8-Channel, 12-Bit, 60 MHz PXI Express Oscilloscope

PXI Contents SPECIFICATIONS. 8-Channel, 12-Bit, 60 MHz PXI Oscilloscope

PXI Contents SPECIFICATIONS. 300 MHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents SPECIFICATIONS. 500 MHz, 2.5 GS/s, 10-Bit PXI Express Oscilloscope

PXIe Contents SPECIFICATIONS. 1.5 GHz, 5 GS/s, 10-Bit PXI Express Oscilloscope

PXI Contents SPECIFICATIONS. 1 GHz Bandwidth, 2 GS/s, 8-Bit PXI Oscilloscope

NI PXI/PCI-5122 Specifications

PXIe, 24-Channel, ±24 V, 50 ma Precision PXI Source Measure Unit

Specifications for the NI PXI/PCI-6552/6551

NI USB Contents DEVICE SPECIFICATIONS. 50 MS/s Bus-Powered USB Oscilloscope

PXIe Contents SPECIFICATIONS. 1 GS/s, 14-bit Reconfigurable Oscilloscope

PXIe Contents SPECIFICATIONS. 125 MHz, 2-Channel, 8-Bit, PXI Express Oscilloscope

PCI Contents SPECIFICATIONS. 100 MHz, 100 MS/s, 14-Bit PCI Oscilloscope Device

PXI Contents SPECIFICATIONS. 100 MHz, 100 MS/s, 14-Bit PXI Oscilloscope

PXI Contents SPECIFICATIONS. 125 MHz, 2-Channel, 8-Bit PXI Oscilloscope

100 MHz Bandwidth, 12-Bit PXI Waveform Generator

This document lists the specifications for the NI PXIe-5186 (NI 5186) 5 GHz digitizer.

This document lists the specifications for the NI PXIe-5185 (NI 5185) 3 GHz digitizer.

PXI Contents SPECIFICATIONS. 150 MHz Bandwidth, 200 MS/s, 12-Bit PXI Oscilloscope

PCI Contents SPECIFICATIONS. 150 MHz Bandwidth, 200 MS/s, 12-Bit PCI Oscilloscope Device

Contents. CALIBRATION PROCEDURE NI PXIe-5698

PXIe Contents SPECIFICATIONS. RF Amplifier

PXIe, 40 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5423.

Power Amplifier. Input Characteristics. SPECIFICATIONS NI PXIe-4610 Power Amplifier. Signal Range

NI PXI/PCI-5412 Specifications 14-Bit 100 MS/s Arbitrary Waveform Generator

80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

NI PXI-4461 Specifications

PXI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PXI Oscilloscope

PXI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PXI Oscilloscope

NI PXIe-5630 Specifications

PXIe, 80 MHz Bandwidth, 16-Bit PXI Waveform Generator. These specifications apply to the one-channel and two-channel PXIe-5433.

PCI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PCI Oscilloscope Device

PCI Contents SPECIFICATIONS. 24-Bit, Flexible Resolution PCI Oscilloscope Device

NI 5421 Specifications NI PXI/PCI Bit 100 MS/s Arbitrary Waveform Generator

CALIBRATION PROCEDURE PXIe-4302/4303 and TB-4302C 32 Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

NI PXI/PCI-5402/5406 Specifications 14/16-Bit, 20/40 MHz Arbitrary Function Generator

DEVICE SPECIFICATIONS NI PXIe-4302/ Ch, 24-bit, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module

DEVICE SPECIFICATIONS NI PXIe Ch, 24-bit, 25.6 ks/s Simultaneous AC LVDT Input Module

NI PXIe Contents CALIBRATION PROCEDURE. Four Channel Precision Source-Measure Unit (SMU)

Analog Characteristics

NI PXIe-5171R. Contents. Required Software CALIBRATION PROCEDURE

NI PXIe-5601 Specifications

Terminology. Output Characteristics. DEVICE SPECIFICATIONS NI PXIe-4463 DSA Analog Output. ni.com/manuals

NI PXI-5422 Specifications 16-Bit 200 MS/s Arbitrary Waveform Generator

PXI Contents SPECIFICATIONS. 1.3 GHz RF Analog Signal Generator

PXIe Contents SPECIFICATIONS. 100 MVector/s PXI Digital Pattern Instrument

PXIe Contents SPECIFICATIONS. 3.3 GHz RF Analog Signal Generator

NI PCI/PXI/PXIe-5114 DEVICE SPECIFICATIONS. 250 MS/s, 8-Bit Oscilloscope

This document lists the specifications for the NI PXI The PXI-5900 is a differential amplifier.

PXIe Contents SPECIFICATIONS. PXIe, 8-Channel, 250 MHz Bandwidth, 14-Bit, Reconfigurable PXI Oscilloscope

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

PXIe Contents SPECIFICATIONS. 32-Channel Digital Pattern Instrument

NI PXIe Contents. Required Software CALIBRATION PROCEDURE. Dual-Output Programmable DC Power Supply

NI PXI/PCI-5421 Specifications 16-Bit 100 MS/s Arbitrary Waveform Generator

Analog Input. DEVICE SPECIFICATIONS NI PXIe-4300

PXIe Contents. Required Software CALIBRATION PROCEDURE

PXIe Contents SPECIFICATIONS. RF Preselector Module

PXI Contents SPECIFICATIONS. 6.6 GHz RF Analog Signal Generator

PXIe Contents. Definitions SPECIFICATIONS. 8.5 GHz Vector Network Analyzer

NI PXIe Contents CALIBRATION PROCEDURE. Single-Channel Precision Source-Measure Unit (SMU)

DEVICE SPECIFICATIONS NI PXIe-4304/ Ch, 24-bit, ±42 V, 5 ks/s or 51.2 ks/s Simultaneous Filtered Data Acquisition Module. ni.

PXIe Contents SPECIFICATIONS. 250 khz to 20 GHz RF Analog Signal Generator

Input Characteristics. Measurement Range. SPECIFICATIONS NI PXIe-4357

100 MHz, 32-Channel, 5 V PXI Digital Waveform Instrument

NI PXIe-5663 Specifications

NI PXI/PXIe-2540 Specifications

250 MS/s, 125 MHz, 8-Bit Digitizers NI PCI-5114, PXI-5114

NI PXI/PXIe-2541 Specifications 300 MHz Ω Matrix

Current ON state current ma minimum, 6 ma typical, 10 ma maximum OFF state current ma maximum

SCXI MHz Quad 8x1 50 Ω Multiplexer

Contents. CALIBRATION PROCEDURE NI PXIe-4463 DSA Analog Output

if it is used in a manner not described in the user documentation.

NI PXI Contents DEVICE SPECIFICATIONS. RF Attenuator

Contents. Software Requirements

NI PXI-2521 Specifications 40-Channel DPST Relay Module

Calibrating the NI 5653 requires you to install one of the following packages on the calibration system. NI-RFSA 2.4 or later NI-RFSG 1.

NI 6143 Specifications

Contents. Software. Documentation. CALIBRATION PROCEDURE NI PXIe-6674T

PXIe, 7½-Digit, ±1,000 V, Onboard 1.8 MS/s Isolated Digitizer, PXI Digital Multimeter

DEVICE SPECIFICATIONS Phase Matrix PXI GHz PXI Vector Signal Analyzer

NI PXI/PCI-6561/6562 Specifications

NI PXI-2520 Specifications 80-Channel SPST Relay Module

is used in a manner not described in the user documentation.

PXI Timing and Synchronization Control

Contents. Software Requirements. CALIBRATION PROCEDURE NI PXIe-5663E

Reconfigurable 6 GHz Vector Signal Transceiver with I/Q Interface

NI PXIe Contents. About These Specifications DEVICE SPECIFICATIONS. 3 GHz Ω Multiplexer (SP16T)

SPECIFICATIONS PXIe ks/s, 119 db, 6 Gains, AC/DC-Coupled, 4-Input PXI Sound and Vibration Module

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

Contents. Software Requirements CALIBRATION PROCEDURE NI PXI-5663

4-Channel, ±6 V, 500 ma Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma Precision PXI Source Measure Unit

PXIe Contents CALIBRATION PROCEDURE. 10 GHz or 20 GHz RF Analog Signal Generator

NI PXI-2557 Specifications

4-Channel ±10 V, 100 ma, Precision PXI Source Measure Unit

4-Channel, ±24 V, 150 ma PXI Source Measure Unit

PXI Contents. Definitions SPECIFICATIONS. ±100 V, 2 W Precision PXI Source Measure Unit. These specifications apply to the PXI-4132.

Transcription:

SPECIFICATIONS PXIe-5113 PXIe, 500 MHz, 3 GS/s, 8-bit PXI Oscilloscope These specifications apply to the PXIe-5113 with 64 MB and 512 MB of memory. Contents Definitions...2 Conditions... 2 Vertical... 3 Analog Input... 3 Impedance and Coupling... 3 Voltage Levels...3 Accuracy... 4 Bandwidth and Transient Response... 5 Spectral Characteristics...7 Horizontal...8 Sample Clock... 8 Phase-Locked Loop (PLL) Reference Clock... 9 Triggers... 9 Analog Trigger... 9 Digital Trigger...10 Programmable Function Interface (PFI)... 10 Probe Compensation... 11 CableSense... 11 Waveform Memory... 12 Calibration...12 External Calibration... 12 Self-Calibration...13 Calibration Specifications... 13 Software... 13 Driver Software...13 Application Software... 13 Interactive Soft Front Panel and Configuration... 13 Synchronization... 14 Synchronization with the NI-TClk API... 14 Power... 15 Physical... 15 Bus Interface... 15 Environmental Characteristics... 15 Product Certifications and Declarations... 16

Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration. Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. Typical specifications describe the performance met by a majority of models. Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing. Measured specifications describe the measured performance of a representative model. Specifications are Nominal unless otherwise noted. Conditions Specifications are valid under the following conditions unless otherwise noted. All vertical ranges, bandwidths, and bandwidth limiting filters Sample rate set to 1.5 GS/s or 3.0 GS/s Onboard sample clock locked to PXI_Clk100 reference clock 15-minute warm-up time at ambient temperature Chassis configured: 1 PXI Express chassis fan speed set to HIGH Foam fan filters removed if present Empty slots contain PXI chassis slot blockers and filler panels Warranted specifications are valid under the following conditions unless otherwise noted. Ambient temperature range of 0 C to 55 C Altitude 2,000 m Calibration cycle maintained Self-calibration run after: Warm-up time has elapsed Module has been power cycled PC or controller has been restarted or wakes from sleep or hibernation modes External calibration performed at 23 C ±3 C 1 For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users available at ni.com/manuals. 2 ni.com PXIe-5113 Specifications

Typical specifications are valid under the following conditions unless otherwise noted. Ambient temperature range of 0 C to 55 C Altitude 2,000 m Vertical Analog Input Number of channels Input type Connectors Two (simultaneously sampled) Referenced single-ended BNC, ground referenced Impedance and Coupling Input impedance Input capacitance (1 MΩ) Input coupling 50 Ω ±1.5%, typical 1 MΩ ±1.0%, typical 15.4 pf AC DC Voltage Levels Table 1. Full-Scale (FS) Input Range and Vertical Offset Range Input Range (V pk-pk ) Vertical Offset Range 50 Ω 1 MΩ 0.04 V ±5 V 0.1 V ±5 V 0.2 V ±5 V 0.4 V ±5 V 1 V ±5 V ±20 V 2 V ±5 V ±20 V 4 V ±5 V ±20 V 10 V ±2 V ±100 V PXIe-5113 Specifications National Instruments 3

Table 1. Full-Scale (FS) Input Range and Vertical Offset Range (Continued) Input Range (V pk-pk ) Vertical Offset Range 50 Ω 1 MΩ Maximum input overload 20 V ±100 V 40 V ±100 V 50 Ω Peaks 7 V 1 MΩ 2 Peaks 250 V DC Accuracy Resolution DC accuracy 3 50 Ω Notice Signals exceeding the maximum input overload may cause damage to the device. Input range: 0.04 V Input range: 0.1 V to 4 V Input range: 10 V 8 bits ±[(2% Reading - Vertical Offset ) + (0.4% Vertical Offset ) + (1% of FS) + 0.2 mv], typical ±[(2% Reading - Vertical Offset ) + (0.4% Vertical Offset ) + (1% of FS) + 0.2 mv], warranted ±[(2% Reading - Vertical Offset ) + (1.1% Vertical Offset ) + (1% of FS) + 0.2 mv], warranted 2 Derate above 250 khz at 20 db/dec until 2.5 MHz, then derate at 5 db/dec. 3 Within ±5 C of self-calibration temperature. 4 ni.com PXIe-5113 Specifications

1 MΩ DC drift 4 Input range: 0.04 V Input range: 0.1 V to 20 V Input range: 40 V AC amplitude accuracy 3 AC amplitude drift 4 Crosstalk Crosstalk 5 Input frequency: 200 MHz Input frequency: 200 MHz to 400 MHz ±[(2% Reading - Vertical Offset ) + (0.4% Vertical Offset ) + (1% of FS) + 0.2 mv], typical ±[(2% Reading - Vertical Offset ) + (0.4% Vertical Offset ) + (1% of FS) + 0.2 mv], warranted ±[(2% Reading - Vertical Offset ) + (1.1% Vertical Offset ) + (1% of FS) + 0.2 mv], warranted ±[(0.2% Reading - Vertical Offset ) + (0.004% Vertical Offset ) + (0.013% of FS)] per C ±0.25 db at 50 khz ±0.0026 db per C at 50 khz <-60 db <-50 db Bandwidth and Transient Response Bandwidth (-3 db) 6 50 Ω 7 Full bandwidth 475 MHz, warranted 500 MHz, typical 350 MHz filter 325 MHz, warranted 350 MHz, typical 4 Used to calculate errors when the onboard temperature changes more than ±5 C from the selfcalibration temperature. 5 Measured on one channel with test signal applied to another channel and the same range setting on both channels. For 1 MΩ path, specifications are valid for input ranges 10 V (V pk-pk ). 6 Normalized to 50 khz. 7 For input ranges 4 V (V pk-pk ) and temperature 0 C to 30 C. PXIe-5113 Specifications National Instruments 5

1 MΩ 8 Full bandwidth 500 MHz, typical 350 MHz filter 350 MHz, typical Figure 1. 50 Ω Full Bandwidth Frequency Response, 3 GS/s, 1 V pk-pk, Measured 6 1 Normalized Amplitude (db) 0 1 2 3 4 5 0 100 200 300 400 500 Frequency (MHz) 600 Bandwidth-limiting filter 20 MHz noise filter Figure 2. 50 Ω 20 MHz Filter Frequency Response, 3 GS/s, 1 V pk-pk, Measured 6 1 Normalized Amplitude (db) 0 1 2 3 4 5 0 5 10 15 20 25 Frequency (MHz) 30 AC-coupling cutoff (-3 db) 10 Hz 8 When used with the NI SP500X passive probe. 6 ni.com PXIe-5113 Specifications

Figure 3. Step Response, 50 Ω, 1 V pk-pk, 500 ps Rising Edge, Measured Volts (V) 0.10 0.05 0 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 5 0 5 10 15 20 25 Time (ns) Figure 4. Step Response, 1 MΩ, 1 V pk-pk, 500 ps Rising Edge, Measured Volts (V) 0.10 0.05 0 0.05 0.10 0.15 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 5 0 5 10 15 20 25 Time (ns) Spectral Characteristics 9 Spurious-free dynamic range (SFDR) 10-45 dbc 9 Excludes ADC interleaving spurs. 10 Input frequencies <100 MHz, input range 4 V pk-pk. -1 dbfs input signal. Includes second through fifth harmonics. PXIe-5113 Specifications National Instruments 7

Table 2. Effective Number of Bits (ENOB) 11 Input Range (V pk-pk ) Filters 20 MHz filter enabled Full bandwidth (Input Frequency <100 MHz) 0.1 V to 4 V 7.3 6.2 0.04 V 6.7 5.8 Total harmonic distortion (THD) 10-45 dbc Noise RMS noise 12 0.04 V pk-pk 0.50% of FS All other ranges 0.33% of FS Horizontal Sample Clock Source Sample rate range, real time 13 Sample rate, time-interleaved sampling (TIS) mode 14 Timebase frequency Timebase accuracy 15 Sample clock jitter 16 Onboard clock (internal oscillator) 22.89 ks/s to 1.5 GS/s 3.0 GS/s 1.5 GHz ±50 ppm 1.1 ps RMS 11 Input frequencies <100 MHz. -1 dbfs input signal corrected to FS. 1 khz resolution bandwidth. 12 Applies to all filter settings and input modes. Verified using a 50 Ω terminator connected to input. 13 Divide by n decimation from 1.5 GS/s. For more information on the sample clock and decimation, refer to the NI High-Speed Digitizers Help. 14 Single channel only. 15 Phase-locked to onboard clock. The default clock is PXI_Clk100. Refer to your chassis specifications for the timebase accuracy of PXI_Clk100. 16 Integrated from 100 Hz to 10 MHz. Includes the effects of converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter. 8 ni.com PXIe-5113 Specifications

Phase-Locked Loop (PLL) Reference Clock Sources Internal External Duty cycle tolerance Triggers Supported triggers Trigger types Onboard clock (internal oscillator) PXI_Clk100 (backplane connector) 45% to 55%, typical Reference (Stop) Trigger Reference (Arm) Trigger Start Trigger (Acquisition Arm) Advance Trigger Edge Glitch Hysteresis Runt Width Window Digital Immediate Software Trigger sources CH 0 CH 1 PFI <0..3> PXI_Trig <0..7> Minimum dead time Interpolator enabled Interpolator disabled Trigger delay Holdoff Analog Trigger 400 ns 400 ns Sources CH 0 CH 1 0 to 7.51 10 14 ns [(2 51-1) * Sample Clock Period] Dead time to 6.15 10 18 ns [(2 64-1) * Sample Clock Period] PXIe-5113 Specifications National Instruments 9

Table 3. Analog Trigger Time Resolution Interpolator Status TIS Enabled Time Resolution TIS Disabled Trigger filters Enabled 0.326 ps 0.651 ps Disabled 0.333 ns 0.667 ns Low frequency (LF) reject High frequency (HF) reject Minimum threshold duration 17 Digital Trigger Sources Time resolution PFI PXI_Trig 100 khz 100 khz Sample clock period PFI <0..3> (front panel HD-BNC connectors) PXI_Trig <0..7> (backplane connector) 1.333 ns 5.333 ns Programmable Function Interface (PFI) Connectors Direction As an input (trigger) Destinations Input impedance V IH V IL Recommended input range Maximum input overload Minimum pulse width PFI <0..3> (front panel HD-BNC connectors) Bidirectional per channel Start Trigger (Acquisition Arm) Reference (Stop) Trigger Reference (Arm) Trigger Advance Trigger 49.9 kω 2 V, typical 0.8 V, typical 0 V to 3.3 V +5 V tolerant 10 ns 17 Data must exceed each corresponding trigger threshold for at least this minimum duration to ensure analog triggering. 10 ni.com PXIe-5113 Specifications

As an output (event) Sources Output impedance Logic type Maximum current drive Maximum frequency Minimum pulse width Ready for Start Start Trigger (Acquisiton Arm) Ready for Reference Reference (Stop) Trigger End of Record Ready for Advance Advance Trigger Done (End of Acquisition) 50 Ω 3.3 V CMOS 12 ma 50 MHz 10 ns Probe Compensation Connectors Output voltage 18 Maximum overload voltage Probe compensation terminal Ground terminal 0 V to 5 V 25 V DC CableSense CableSense pulse voltage 19 CableSense pulse rise time 20 0.4 V 1.3 ns Driver support for CableSense on the PXIe-5113 was first available in NI-SCOPE 18.7. Related Information For more information about CableSense technology, refer to ni.com/cablesense. 18 1 khz, 50% duty cycle square wave. 19 When measured with a high-impedance device. 20 When sourcing into a 50 Ω cable or load. PXIe-5113 Specifications National Instruments 11

Waveform Memory Available onboard memory sizes 21 Minimum record length Number of samples 64 MB 512 MB 1 sample Pretrigger 0 up to (Record Length - 1) Posttrigger Maximum number of records in onboard 100,000 memory 22 0 up to Record Length Table 4. Examples of Allocated Onboard Memory per Record, 512 MB Option Channels Bytes per Maximum Records Record Length Allocated Onboard Sample per Channel Memory per Record 1 1 100,000 1 192 1 1 100,000 1,000 1,200 1 1 52,758 10,000 10,176 1 1 1 536,870,784 536,870,976 2 1 100,000 1 192 2 1 100,000 1,000 2,208 2 1 26,630 10,000 20,160 2 1 1 268,435,392 536,870,976 Calibration External Calibration External calibration corrects the onboard references for gain and offset errors used in selfcalibration and adjusts the compensation attenuator. All calibration constants are stored in nonvolatile memory. 21 Onboard memory is shared among all enabled channels. 22 For 512 MB option. You can exceed this value if you fetch records while acquiring data. For more information, refer to the Enable Records > Memory property in the NI High-Speed Digitizers Help at ni.com/manuals. 12 ni.com PXIe-5113 Specifications

Self-Calibration Self-calibration is done on software command. The calibration corrects for gain, offset, interleaving spurs, and intermodule synchronization errors. Run self-calibration after the specified warm-up time has elapsed and any time the module is power cycled or the PC or controller is restarted or wakes from sleep or hibernation modes. Refer to the NI High-Speed Digitizers Help at ni.com/manuals for more information on when to self-calibrate the device. Calibration Specifications Interval for external calibration Warm-up time 23 2 years 15 minutes Software Driver Software Driver support for this device was first available in NI-SCOPE 18.6. NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5113. NI-SCOPE provides application programming interfaces for many development environments. Application Software NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments: LabVIEW LabWindows /CVI Measurement Studio Microsoft Visual C/C++.NET (C# and VB.NET) Interactive Soft Front Panel and Configuration When you install NI-SCOPE on a 64-bit system, you can use InstrumentStudio to monitor, control, and record measurements from the PXIe-5113. InstrumentStudio is an application that allows you to perform interactive measurements on several different NI device types in a single application. Interactive control of the PXIe-5113 was first available via InstrumentStudio in NI-SCOPE 18.6. InstrumentStudio is included on the NI-SCOPE media. NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5113. MAX is included on the driver media. 23 Warm-up time begins after the chassis and either the controller or PC is powered and NI-SCOPE is loaded. PXIe-5113 Specifications National Instruments 13

Synchronization Channel-to-channel skew, between the channels of a PXIe-5113 50 Ω <60 ps 1 MΩ <60 ps Synchronization with the NI-TClk API 24 NI-TClk is an API that enables system synchronization of supported PXI modules in one or more PXI chassis, which you can use with the PXIe-5113 and NI-SCOPE. NI-TClk uses a shared Reference Clock and triggers to align the Sample Clocks of PXI modules and synchronize the distribution and reception of triggers. These signals are routed through the PXI chassis backplane without external cable connections between PXI modules in the same chassis. Module-to-module skew, between PXIe-5113 modules using NI-TClk 25 NI-TClk synchronization without manual adjustment 26 Skew, peak-to-peak 27 Jitter, peak-to-peak 28 200 ps 120 ps NI-TClk synchronization with manual adjustment 26 Skew, average 27 Jitter, peak-to-peak 28 10 ps 8 ps 24 NI-TClk installs with NI-SCOPE. 25 Although you can use NI-TClk to synchronize non-identical modules, these specifications apply only to synchronizing identical modules. Specifications are valid under the following conditions: All modules installed in the same PXI Express chassis NI-TClk used to align the sample clocks of each module All parameters set to identical values for each module Self-calibration completed Ambient temperature within ±1 C of self-calibration For other configurations, including multi-chassis systems, contact NI Technical Support at ni.com/ support. 26 Manual adjustment is the process of minimizing synchronization jitter and skew by adjusting Trigger Clock (TClk) signals using the instrument driver. 27 Skew is the misalignment between module timing across slots of a chassis and is caused by clock and analog path delay differences. 28 Jitter is the variation in module alignment that can be expected with each call to NI-TClk Synchronize. 14 ni.com PXIe-5113 Specifications

Sample Clock delay/adjustment resolution <1 ps Related Information NI-TClk Overview For more information on manual adjustment, refer to NI-TClk Manual Calibration on NI-SCOPE Devices. Power Current draw +3.3 V DC 1.82 A +12 V DC 1.16 A Power draw +3.3 V DC 6 W +12 V DC 14 W Total Total maximum power allowed 20 W 30 W Physical Dimensions Weight Bus Interface Form factor Slot compatibility 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm 13.0 cm 21.6 cm (0.8 in 5.1 in 8.5 in) 380 g (13.4 oz) Gen 1 x4 module PXI Express or hybrid Environmental Characteristics Temperature and Humidity Temperature Operating 0 C to 55 C Storage -40 C to 71 C PXIe-5113 Specifications National Instruments 15

Humidity Operating 10% to 90%, noncondensing Storage 5% to 95%, noncondensing Pollution Degree 2 Maximum altitude 4,600 m (at 25 C ambient temperature) Shock and Vibration Random vibration Operating Non-operating Operating shock 5 Hz to 500 Hz, 0.3 g RMS 5 Hz to 500 Hz, 2.4 g RMS 30 g, half-sine, 11 ms pulse Product Certifications and Declarations Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/ certification, search by model number or product line, and click the appropriate link in the Certification column. Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015. 2018 2019 National Instruments. All rights reserved. 377464B-01 January 11, 2019