--> Buy True-PDF --> Auto-delivered in 0~10 minutes. JY/T

Similar documents
NanoSpective, Inc Progress Drive Suite 137 Orlando, Florida

ELECTRON MICROSCOPY. 13:10 16:00, Oct. 6, 2008 Institute of Physics, Academia Sinica. Tung Hsu

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 7. Chapter 2 CHEM Fall L. Ma

Transmission Electron Microscopy 9. The Instrument. Outline

Transmission electron Microscopy

Introduction to Transmission Electron Microscopy (Physical Sciences)

ELECTRON MICROSCOPY. 14:10 17:00, Apr. 3, 2007 Department of Physics, National Taiwan University. Tung Hsu

NANO 703-Notes. Chapter 9-The Instrument

SCANNING ELECTRON MICROSCOPY AND X-RAY MICROANALYSIS

Exam 4. Name: Class: Date: Multiple Choice Identify the choice that best completes the statement or answers the question.

TEM theory Basic optics, image formation and key elements

ELECTRON MICROSCOPY. 09:10 12:00, Oct. 27, 2006 Institute of Physics, Academia Sinica. Tung Hsu

Chapter Ray and Wave Optics

A few concepts in TEM and STEM explained

CHAPTER TWO METALLOGRAPHY & MICROSCOPY

Chapter 4 Imaging Lecture 17

VISUAL PHYSICS ONLINE DEPTH STUDY: ELECTRON MICROSCOPES

Introduction to Electron Microscopy

S200 Course LECTURE 1 TEM

OPTICAL SYSTEMS OBJECTIVES

ECEN 4606, UNDERGRADUATE OPTICS LAB

Transmissions Electron Microscopy (TEM)

ELECTRON MICROSCOPY AN OVERVIEW

Observing Microorganisms through a Microscope LIGHT MICROSCOPY: This type of microscope uses visible light to observe specimens. Compound Light Micros

High Resolution Transmission Electron Microscopy (HRTEM) Summary 4/11/2018. Thomas LaGrange Faculty Lecturer and Senior Staff Scientist

PROCEEDINGS OF A SYMPOSIUM HELD AT THE CAVENDISH LABORATORY, CAMBRIDGE, Edited by

ISO INTERNATIONAL STANDARD

Lab 05: Transmission Electron Microscopy

Basics of Light Microscopy and Metallography

2 How to operate the microscope/obtain an image

Chapter 1. Basic Electron Optics (Lecture 2)

Study on Imaging Quality of Water Ball Lens

Scanning Electron Microscopy Basics and Applications

MSE 595T Transmission Electron Microscopy. Laboratory III TEM Imaging - I

Nanotechnology and material science Lecture V

LEOK-3 Optics Experiment kit

LEO 912 TEM Short Manual. Prepared/copyrighted by RH Berg Danforth Plant Science Center

MSE 460 TEM Lab 2: Basic Alignment and Operation of Microscope

Scanning Electron Microscopy. EMSE-515 F. Ernst

GB/T Translated English of Chinese Standard: GB/T NATIONAL STANDARD OF THE

plasmonic nanoblock pair

Light Microscopy. Upon completion of this lecture, the student should be able to:

25 cm. 60 cm. 50 cm. 40 cm.

ECEN. Spectroscopy. Lab 8. copy. constituents HOMEWORK PR. Figure. 1. Layout of. of the

Experiment 1: Fraunhofer Diffraction of Light by a Single Slit

MSE 460 TEM Lab 4: Bright/Dark Field Imaging Operation

Optical design of a high resolution vision lens

Nanotechnology in Consumer Products

Supplementary Figure 1. Effect of the spacer thickness on the resonance properties of the gold and silver metasurface layers.

Microscope anatomy, image formation and resolution

Introduction of New Products

SUBJECT: PHYSICS. Use and Succeed.

Diffraction. Interference with more than 2 beams. Diffraction gratings. Diffraction by an aperture. Diffraction of a laser beam

COURSE NAME: PHOTOGRAPHY AND AUDIO VISUAL PRODUCTION (VOCATIONAL) FOR UNDER GRADUATE (FIRST YEAR)

JEM-F200. Multi-purpose Electron Microscope. Scientific / Metrology Instruments Multi-purpose Electron Microscope

(Refer Slide Time: 00:10)

Filter & Spectrometer Electron Optics

Introduction to Scanning Electron Microscopy

Procedures for Performing Cryoelectron Microscopy on the FEI Sphera Microscope

Laser Telemetric System (Metrology)

Add CLUE to your SEM. High-efficiency CL signal-collection. Designed for your SEM and application. Maintains original SEM functionality

Chapter 36: diffraction

R.B.V.R.R. WOMEN S COLLEGE (AUTONOMOUS) Narayanaguda, Hyderabad.

Applied Optics. , Physics Department (Room #36-401) , ,

FRAUNHOFER AND FRESNEL DIFFRACTION IN ONE DIMENSION

INTRODUCTION THIN LENSES. Introduction. given by the paraxial refraction equation derived last lecture: Thin lenses (19.1) = 1. Double-lens systems

Test procedures Page: 1 of 5

Recent results from the JEOL JEM-3000F FEGTEM in Oxford

Scanning electron microscope

Chapter 25. Optical Instruments

Characteristics of point-focus Simultaneous Spatial and temporal Focusing (SSTF) as a two-photon excited fluorescence microscopy

IMAGE FORMATION. Light source properties. Sensor characteristics Surface. Surface reflectance properties. Optics

Observing Microorganisms through a Microscope

Microscopy Techniques that make it easy to see things this small.

Cs-corrector. Felix de Haas

Optical Design with Zemax

Introduction: Why electrons?

Chapter 2 The Study of Microbial Structure: Microscopy and Specimen Preparation

New 500 kv Electron Microscope

No part of this material may be reproduced without explicit written permission.

LOS 1 LASER OPTICS SET

From birth to present of hologram.

Introduction to Light Microscopy. (Image: T. Wittman, Scripps)

Applications of Steady-state Multichannel Spectroscopy in the Visible and NIR Spectral Region

Lens Design I. Lecture 3: Properties of optical systems II Herbert Gross. Summer term

General Physics II. Ray Optics

OPTICS DIVISION B. School/#: Names:

FZ/T Translated English of Chinese Standard: FZ/T

BEAM HALO OBSERVATION BY CORONAGRAPH

GEOMETRICAL OPTICS Practical 1. Part I. BASIC ELEMENTS AND METHODS FOR CHARACTERIZATION OF OPTICAL SYSTEMS

Very short introduction to light microscopy and digital imaging

Imaging Systems Laboratory II. Laboratory 8: The Michelson Interferometer / Diffraction April 30 & May 02, 2002

The diffraction of light

X-ray generation by femtosecond laser pulses and its application to soft X-ray imaging microscope

Development of a new multi-wavelength confocal surface profilometer for in-situ automatic optical inspection (AOI)

Phys 531 Lecture 9 30 September 2004 Ray Optics II. + 1 s i. = 1 f

ABC Math Student Copy. N. May ABC Math Student Copy. Physics Week 13(Sem. 2) Name. Light Chapter Summary Cont d 2

Chapter Wave Optics. MockTime.com. Ans: (d)

Point Spread Function. Confocal Laser Scanning Microscopy. Confocal Aperture. Optical aberrations. Alternative Scanning Microscopy

Transcription:

Translated English of Chinese Standard: JY/T011-1996 www.chinesestandard.net Sales@ChineseStandard.net INDUSTRY STANDARD OF THE JY PEOPLE S REPUBLIC OF CHINA General rules for transmission electron microscopy 现代分析仪器分析方法通则 透射电子显微镜方法通则 Issued on: January 22, 1997 Implemented on: April 1, 1997 Approved by: State Education Commission www.chinesestandard.net Page 1 of 15

Table of Contents Foreword... 3 1 Scope... 4 2 Definitions... 4 3 Principle... 4 4 Instruments... 6 5 Specimen... 8 6 Analytical procedures... 8 7 Expression of analysis results... 14 8 Safety precautions... 14 www.chinesestandard.net Page 2 of 15

Foreword The drafting format and method of this Standard complies with the requirements of GB/T 1.1-1993 "Directives for the work of standardization - Unit 1: Drafting and presentation of standards - Part 1: General rules for drafting standards" and GB/T 1.4-88 "Directives for the work of standardization - rules for drafting chemical analysis standards". The drafting organization of this Standard: The State Education Commission. Main drafters of this Standard: Zhang Datong and Wang Yongrui. Participating drafters of this Standard: Wan Derui, Lin Chengyi and Zhou Shanyuan. www.chinesestandard.net Page 3 of 15

General rules for transmission electron microscopy 1 Scope This General Rules specify the conventional analytical method of transmission electron microscopy; it is are applicable to high-voltage transmission electron microscopy, high-resolution transmission electron microscopy, ordinary transmission electron microscopy AND simple transmission electron microscopy (hereinafter referred to as TEM). 2 Definitions This Standard adopts the following definitions: 2.1 Resolution TEM's capability that can clearly distinguish the minimum distance between two object points and crystal faces. 2.2 Astigmatism An image blurring that is caused by non-rotational symmetry of electromagnetic lenses' magnetic field, in which the lenses have different focusing ability in mutually perpendicular directions. 2.3 Electron diffraction The effect whereby electron beams are scattered in crystals; only in the direction of satisfying the Bragg's Law, are there emitting of diffracted beams that are mutually reinforced. 3 Principle The determinant of TEM imaging is the specimen scattering to incident electron, including elastic scattering and inelastic scattering. When thin-specimen is imaging, the un-scattered electrons shall constitute the background. However, the image contrast depends on different scattering characteristics of each part of the specimen to the electrons. Different experimental conditions shall produce different image contrasts. TEM can not only show the microstructure morphology of the specimen, but www.chinesestandard.net Page 4 of 15

3.2 Diffraction modality If the specimen is crystal, its electron diffraction pattern shall be presented on the focal plane of the objective lens. Change the current of intermediate lens to make the image formed on the focal plane of the objective lens. The electron diffraction pattern on the plane is enlarged by the intermediate lens and the projector lens. Then the enlarged image of electron diffraction pattern shall be obtained on the phosphor screen. 4 Instruments 4.1 Instrument structure 4.1.1 Lighting system It is composed of electron gun and condenser. Electron gun provides stable electron source of small size and emits high-brightness electron beams. Condenser converges electron beams and irradiates the specimen. 4.1.2 Specimen room and imaging system The specimen room is under the lighting system. Placed on the specimen holder, the specimen can move or tilt within the specified range. Composed of objective lens, intermediate lens and projector lens, the imaging system enlarges the image and presents it on the phosphor screen. 4.1.3 Observing and recording system The observing room is under the projector lens. The image on the phosphor screen can be observed through windows. The photographic plate is placed under the phosphor screen. When phosphor screen is erected, it shall expose and record the image. 4.1.4 Other systems and accessories Other systems and accessories include vacuum system, power system, security system and cooling system. Some are equipped with scanning accessories, micro diffraction apparatus, electron energy loss spectrometer (EELS) and X-ray energy dispersive spectrometer (EDS), etc. 4.2 Technical indicators www.chinesestandard.net Page 6 of 15

6.2 Test preparation 6.2.1 Centering adjustment Increase high voltage and filament current. After light spot appears on the phosphor screen, center the lighting system and imaging system. Adjustment of current center or voltage center shall be strictly in accordance with the instructions for use. 6.2.2 Astigmatism calibration Use the well-known specimens, for example, micro sieve. Observe the hole at high magnification. Alternately adjust the focus of the objective lens and the anastigmator. When the objective lens is under-focused or over-focused, the uniform and clear Fresnel Fringes image shall be obtained at the edge of the hole. 6.2.3 Magnification calibration Because of hysteresis effect of electromagnetic lens, there are 5%~10% error between the actual magnification and the reading value. If the magnification is less than 50,000 times, use grating replica (2000 pcs/mm) with 50 nm fringe space to calibrate. For greater magnification, use thin crystal with well-known grating space to calibrate. Shoot photos of different magnifications and obtain the calibration curve by calculation. 6.2.4 Camera constant calibration Camera constant is the product of diffraction camera length L and wavelength of the incident electron beams k. Because of uncertainty of the length of TEM camera, it needs to use a gold polycrystalline thin film to calibrate. 6.2.5 Magnetic rotation calibration In the selective-area, when diffraction modality is switched to imaging modality, the change on excitation conditions of intermediate lens shall make the image and the diffraction pattern rotate at different angles, comparing to the actual orientation of specimen crystal. If the magnetic rotation of the image is Φi and the magnetic rotation of the pattern is Φd. then the magnetic rotation of the image to the pattern shall be Φ=Φi Φd. Usually, it utilizes external features to directly reflect the calibrated magnetic rotation of MOO3 thin crystal. Therefore, it shall only need the calibration value of crystal rotation represented by the diffraction pattern to represent the actual rotation of the crystal. 6.2.6 Specimen height calibration During magnification calibration and selective-area diffraction, in order to www.chinesestandard.net Page 9 of 15

them coincide. Switch to diffraction modality. Pull out objective lens aperture. The electron diffraction pattern, which reflects the crystallographic characteristics of specimen in micro-area, can be obtained on phosphor screen. For electron diffraction in selective-area, the size of selective-area is limited by the effect of spherical aberration of objective lens and image focus error. For high-resolution TEM, the diameter of minimum selective-area is about 1 µ. In order to obtain smaller selective-area, it uses micro-diffraction techniques (µ diffraction). In experiment, use scanning transmission operation mode, the beam-spot diameter is reduced to less than 100 nm. The scope of micro-area of specimen that participates in diffraction is limited by the beam-spot diameter. 6.4.3 Bright-field image and dark-field image of diffraction contrast Usually, the analysis on bright-field imaging and dark-field imaging is always combined with electron diffraction of selective-area, so as to determine the phase's microscopic morphology, lattice type and parameters. If objective lens aperture is used to block diffraction beams, and only transmission beams are allowed to go through the aperture hole to imaging, then, the image obtained shall be called bright-field image of diffraction contrast. If objective lens aperture is used to block transmission beams and most of diffraction beams, and only some diffraction beams are allowed to go through the aperture hole to imaging, obviously, the area contributed to the diffraction beams of specimen or the bright contrast represented by phase in dark field shall be called the dark-field image of diffraction contrast. Generally speaking, central dark field imaging produces small aberration. Make selective-area diffraction of specimen first. Tilt the specimen to make some diffraction spot to be the brightest (dual-beam conditions) except the transmission spot. Adjust the lighting electron beam to tilt. Move the transmission beam to the original bright diffraction spot. Then the weak spot which is symmetrical to the original bright spot is moved to optical axis and shall get brighter. Allow only this diffraction beam to get through the aperture hole of objective lens to form the central dark-field image. Use specific diffraction beams to form the central dark-field image of diffraction contrast is one of the effective methods to analyze complicating diffraction figure and display each single-phase crystallographic characteristics. 6.4.4 Dark-field image of weak-beam It is widely used in studies on crystal defects, for example, the bright-field image and central dark-field image have higher resolution. In experiment, make specimen relatively far away from Bragg reflection position, comparing to electron beam. Use the weak first-order diffraction beam to make central darkfield image. Near the center of the crystal dislocation, there are always some www.chinesestandard.net Page 11 of 15

For example, use accessories of convergent diffraction, add the third condenser above the objective lens OR adopt scanning transmission mode to reduce the specimen height. 6.4.7 Lattice type of crystal and determination of lattice parameters For the well-known crystal, it can use the standardization of one or several electron diffraction patterns to determine the phase. However, for the unknown phase with complicated structure, it must determine its three-dimensional structure. The spatial periodicity of crystal structure reflected on electron diffraction pattern is the reciprocal space corresponding to its periodicity. Therefore, if a set of electron diffraction patterns, that are of different crystal belts of same-crystal and that has the correlation, can be obtained, then, according to its correlation, its crystal lattice can be rebuilt by calculation, the lattice parameters can be determined, and the phase can be identified. For this purpose, it shall need dual-tilting or tilting-rotating specimen-holder to make the analyzed crystal to rotate around some low-index crystal orientation, so as to adjust & find the electron diffraction pattern of relevant crystal belt. 6.4.8 Test of small size and shape When testing the size and shape of micro-particle specimen, in order to eliminate systematic errors of TEM and make the tested specimen to be measured precisely, the most effective method shall be the Internal Standard Method. That is, put a standard specimen of well-known size and the tested specimen on a same copper grid carrier. Put them into TEM to observe and shoot photos. It shall not cause magnification error because the standard specimen and the tested specimen are on a same photographic plate under same shooting conditions. The magnification calculated by standard specimen shall be the magnification of tested specimen. The common use specimen for internal standard method at high magnification is polystyrene sphere or crystal of well-known grating space, e.g., beef liver enzyme hydroperoxide (8.6± 0.2nm). When magnification is low, it usually uses grating replica as external standard specimen. In order to eliminate the magnification error, the selected magnification shall not be adjusted during specimen and standard specimen observation AND shooting. 6.4.9 Electron energy loss spectroscopy analysis It is to study the primary process after high-energy electrons are incident to the specimen, motivating electrons of atomic inner shell to ionize. It shall be realized by electron energy loss spectrometer (EELS). EELS is composed of electron energy analyzer, spectrum display, recording system and data www.chinesestandard.net Page 13 of 15

processing system. After the specimen get through the spectrometer, it shall scatter according to amount of electron energy loss, record the appropriate strength and display the spectrum of electron energy loss. Analyze different spectral region, then information about specimen thickness, chemical composition, electronic density and electronic structure can be obtained. 6.4.10 X-ray energy spectrum analysis After high-energy electrons are incident to the specimen, electrons of atomic inner shell are motivated to ionize. Then, X-ray is generated when atoms are reverting to basic state. The energy spectrum analysis is to study the motivation process of this characteristic and provide information about chemical composition in micro-area of specimen. X-ray energy dispersive spectrometer (EDS) collects the characteristic X-ray from specimen, enlarge it, and sent it to multichannel analyzer. Scatter according to its different characteristic energy, record the corresponding strength and display X-ray spectrum. Identify the elements contained in the specimen according to qualitative characteristic energy of spectrum peak. Use data processing system to run thin film quantitative analysis program. Calculate the percentage content of each element according to intensity value. 6.5 Instrument check and power off after testing Take out the specimen. Exit from aperture of objective lens and aperture of selective-area. Adjust to low magnification. Defocus the condenser. Successively cut off filament current, high voltage, main power supply and power supply of other accessories. Cover the back plate of observation room. After 15 mins, turn off cooling water system and regulated power supply. 7 Expression of analysis results 7.1 TEM image is usually provided in photo and interpreted according to massthickness contrast, diffraction contrast and phase contrast imaging theory. 7.2 Photographic plate of electron diffraction is standardized and analyzed according to basic theory of electron diffraction to determine crystallographic orientation relationship or crystal structure. 7.3 Provide results of qualitative or quantitative analysis on chemical composition in micro-area of specimen. 8 Safety precautions 8.1 Operating rules of TEM must be strictly obeyed. www.chinesestandard.net Page 14 of 15

8.2 When power or water is suddenly cut off, it shall sequentially turn off and leave the specimen in original place. After power or water supply is restored, reboot the system to continue the work. END www.chinesestandard.net Page 15 of 15